Abstract
Herein we introduce a novel water-based graphite ink modified with multiwalled carbon nanotubes, designed for the development of the first wearable self-powered biosensor enabling alcohol abuse detection through sweat analysis. The stencil-printed graphite (SPG) electrodes, printed onto a flexible substrate, were modified by casting multiwalled carbon nanotubes (MWCNTs), electrodepositing polymethylene blue (pMB) at the anode to serve as a catalyst for nicotinamide adenine dinucleotide (NADH) oxidation, and hemin at the cathode as a selective catalyst for H2O2 reduction. Notably, alcohol dehydrogenase (ADH) was additionally physisorbed onto the anodic electrode, and alcohol oxidase (AOx) onto the cathodic electrode. The self-powered biosensor was assembled using the ADH/pMB-MWCNTs/SPG||AOx/Hemin-MWCNTs/SPG configuration, enabling the detection of ethanol as an analytical target, both at the anodic and cathodic electrodes. Its performance was assessed by measuring polarization curves with gradually increasing ethanol concentrations ranging from 0 to 50 mM. The biosensor demonstrated a linear detection range from 0.01 to 0.3 mM, with a detection limit (LOD) of 3 ± 1 µM and a sensitivity of 64 ± 2 μW mM−1, with a correlation coefficient of 0.98 (RSD 8.1%, n = 10 electrode pairs). It exhibited robust operational stability (over 2800 s with continuous ethanol turnover) and excellent storage stability (approximately 93% of initial signal retained after 90 days). Finally, the biosensor array was integrated into a wristband and successfully evaluated for continuous alcohol abuse monitoring. This proposed system displays promising attributes for use as a flexible and wearable biosensor employing biocompatible water-based inks, offering potential applications in forensic contexts.
Graphical Abstract
A novel water-based graphite ink modified with multiwalled carbon nanotubes designed for the development of a wearable self-powered biosensor enabling alcohol abuse detection through sweat analysis.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
In recent decades, there has been a growing interest in the development of affordable wearable electrochemical biosensors for remote sensing [1,2,3,4,5]. This has prompted researchers to explore new technological and research solutions aimed at reducing manufacturing costs and improving the reliability, reproducibility, and stability of these biosensing platforms [6,7,8].
Recently, wearable self-powered biosensors enabled the development of wearable analytical sensing devices without an external electronic power source or an onboard battery, utilizing a biorecognition component to transmit sensing information [9, 10]. The ability to self-generate power is particularly crucial in situations where replacing or recharging a battery is either infeasible or impractical. While the majority of wearable self-powered biosensors primarily depend on enzymatic reactions to generate the necessary energy for their operation, there are also innovative approaches that combine multiple signal transduction techniques to both supply power and produce a detection signal simultaneously [11, 12]. Other researchers are focused on the development of more efficient energy harvesting techniques with higher power output and the creation of additional wearable self-powered biosensor devices equipped with integrated low-power wireless electronics [13, 14]. Most of enzyme-based self-powered biosensors primarily are based on a two-electrode configuration (anode and cathode), generating significant open circuit potential, substantial short circuit currents, and maximum power densities through advancements in catalysts, materials, interfaces, and cell designs [15,16,17,18].
Besides the technological advancement related to wearable self-powered devices, biosensor manufacturing constantly requires innovative materials and electrode preparation techniques to address these challenges [19,20,21]. While conductive inks were initially introduced several decades ago for repairing electrical circuits, they remain expensive and involve specific curing procedures that require significant time and high temperatures. Additionally, these inks are typically dispersed in organic solvents, leading to lower conductivities and potential harm to biological recognition elements such as redox enzymes, antibodies, and DNA [22,23,24]. The recent development of water-based lab-made inks has opened up the possibility of constructing biosensor architectures with electrochemical performance comparable to solid electrodes mentioned in existing literature [25,26,27,28,29].
To create reliable sensor devices, the ink mixtures must have a uniform composition with conductive properties and a moderate drying time. Rapid drying can result in surface cracking, posing challenges for electrode manufacturing, while slow drying hinders scalability and the ability to shape or size electrodes [21, 30, 31].
To ensure high reproducibility and robustness, wearable enzyme-based biosensors need to be tested under real operating conditions, taking into account factors such as the blood/tissue or peripheral bodily fluid ratio, which can be influenced by various factors like hormonal dysfunctions, sweating rate, and age [32,33,34]. The immobilization of bioreceptors significantly affects both reproducibility and robustness of enzyme-based amperometric biosensors [35].
Most ethanol biosensors are developed considering alcohol oxidase and nicotinamide adenine dinucleotide (NAD)–dependent alcohol dehydrogenase [36, 37]. Alcohol oxidase (AOx; alcohol:O2 oxidoreductase, EC 1.1.3.13) is a multi-unit enzyme composed of eight identical sub-units arranged in a quasi-cubic configuration, each containing a firmly bound cofactor, flavin adenine dinucleotide (FAD) molecule [38, 39]. Methylotrophic yeasts (such as Hansenula, Pichia, Candida) produce AOx in subcellular microbodies called peroxisomes while growing on methanol. AOx serves as the initial enzyme in the methanol oxidation pathway of methylotrophic yeasts. While its primary physiological function is the oxidation of methanol, it also has the capability to oxidize other short-chain alcohols like ethanol, propanol, and butanol [40]. For ethanol biosensors relying on ADH, the enzyme catalyzes the oxidation of ethanol to acetaldehyde in the presence of NAD+, leading to the reduction of NAD+ to NADH. The amperometric detection method monitors the release of electrons and protons during the oxidation of NADH. While ADH offers the advantage of being more stable and selective towards ethanol without dependence on oxygen, it faces the drawback of instability and the challenge of relying on the continuous recovery of the coenzyme NAD+ in the assay [41].
In addition to enzymatic detection, various analytical techniques have been suggested for identifying alcohol abuse, including chemiluminescence, high-performance liquid chromatography, and magnetic resonance spectroscopy [42]. Nonetheless, these approaches typically have limitations, such as being time-consuming, costly, and necessitating specialized laboratory equipment and trained staff. Self-powered enzyme-based biosensors offer a promising alternative, especially for creating wearable biosensors for continuous monitoring of metabolites and enabling remote healthcare [43, 44].
This work reports on the formulation of a novel water-based graphite ink modified with multiwalled carbon nanotubes to fabricate the first wearable self-powered biosensor for the detection of alcohol abuse in sweat with a detection limit (LOD) of 3 ± 1 µM, a sensitivity of 64 ± 2 μW mM−1, and excellent storage stability (approximately 93% of initial signal retained after 90 days). The electrodes were printed onto a flexible support and modified by electrodepositing polymethylene blue (pMB) at the anode as catalyst for NADH oxidation, and hemin at the cathode as a catalyst for H2O2 reduction, as displayed in Fig. 1. Notably, alcohol dehydrogenase was further casted onto the anodic electrode, while alcohol oxidase was deposited onto the cathodic one, thus using only the analytical target to trigger the self-powered analytical detection (Fig. 1). After preliminary characterization performed in buffer and artificial sweat, the proposed array was integrated into a wrist band to continuously monitor sweat alcohol levels during daily activities with the results showing promise for future applications in remote forensic medicine. Since there is a 1:1 correspondence between blood alcohol content and sweat alcohol content, it was possible to detect alcohol abuse by triggering skin perspiration using iontophoresis [45].
Materials and methods
Chemicals and reagents
4-(2-Hydroxyethyl)piperazine-1-ethanesulfonic acid (HEPES), acetic acid (CH3COOH), d-glucose, l-lactic acid, potassium chloride (KCl), hydrochloric acid (HCl), sodium hydroxide (NaOH), uric acid, ascorbic acid, pyruvate, d-galactose, d-fructose, dopamine hydrochloride, isopropyl alcohol (IPA), graphite powder (< 20 μm, synthetic), chitosan medium molecular weight, glycerol (ACS grade ≥ 99.5%), multiwalled carbon nanotubes (MWCNTs), oxidized nicotinamide adenine dinucleotide (NAD+), reduced nicotinamide adenine dinucleotide (NADH), bovine serum albumin (BSA), iron(III) chloride (FeCl3), polyvinyl alcohol (PVA), pilocarpine nitrate, agarose, alcohol dehydrogenase (ADH) from Saccharomyces cerevisiae, and alcohol oxidase (AOx) from Pichia pastoris were purchased from Merck Millipore (formerly Sigma-Aldrich).
ADH and AOx were dissolved in 10 mM HEPES buffer pH 8.2 and 10 mM HEPES buffer at pH 7.4, respectively, with enzymatic activity of 300 U/mL for both enzymes.
All solutions were prepared using Milli-Q water (18.2 MΩ cm, Millipore, Bedford, MA, USA).
Water-based conductive ink formulation and SPG electrode preparation
The water-based ink was developed using graphite, chitosan, and glycerol as the conductive material, binder, and stabilizer, respectively. A 2.5% w/v chitosan solution was prepared by dissolving chitosan in 1 M acetic acid, which was then stirred overnight at room temperature. The chitosan solution was subsequently diluted to 1% w/v with distilled water, resulting in a final acetic acid concentration of 0.4 M. The conductive ink was prepared by mixing 5.4 g of graphite powder with 9 mL of the previously prepared chitosan solution and 300 µL of glycerol, as previously optimized [22].
Next, the stencil-printed graphite (SPG) electrodes were fabricated using polyethylene terephthalate (PET) sheets that were cleaned three times with IPA and distilled water, then sanded with fine emery paper (1500 grit) to enhance ink adhesion. A stencil was created on a Smart Vinyl adhesive sheet using a Cricut Explore® 3 equipped with Design Space Software v.7.3.95. After applying the stencil to the PET sheet, 500 µL of the conductive ink was placed onto the PET sheet and spread evenly with a scraper. The prepared electrode was allowed to dry at room temperature for 10 min and then cured in an oven at 100 °C for 1 min. Subsequently, the stencil was removed, and the connecting track between the working electrode and the pad was insulated with nail polish [21].
Anode electrode preparation
To prepare ADH/pMB-MWCNTs/SPG, 3 μL of a MWCNTs EtOH:H2O suspension (70:30, 10 mg mL−1), previously ultrasonicated for 1 week, was drop-cast onto the electrode and allowed to dry at room temperature. Afterwards, the so prepared MWCNTs/SPG were further modified by electrodeposition of pMB sweeping the potential between − 0.4 and + 1.2 V vs. Ag/AgCl in 0.25 mM MB solution (0.02 M borate buffer pH 9.14, supporting electrolyte 0.1 M KNO3, sweep rate 50 mV s−1, from 5 to 35 cycles) [17]. Alternatively, the electrodeposition of pMB was performed by using a potentiostatic pulsed method, adopting the following parameters: + 1.2 V oxidation potential and 1 s oxidation pulse time, − 0.4 V reduction potential, 1 s reduction pulse time, and 0.2 s as interval time, ranging the total number of cycles from 50 to 350 cycles. The electrodeposition was performed in 0.25 mM MB solution (0.02 M borate buffer pH 9.14, supporting electrolyte 0.1 M KNO3). Afterwards, the pMB-MWCNTs/SPG electrodes were modified with 5 µL of ADH. Finally, 5 µL of a BSA solution (10 mg mL−1) was deposited onto the electrode to prevent the electrode biofouling and interferences. The electrodes were further conditioned overnight in 10 mM HEPES buffer pH 8.2 at + 4 °C.
Cathode electrode preparation
To prepare AOx/Hemin-MWCNTs/SPG, 3 μL of a MWCNTs EtOH:H2O suspension (70:30, 10 mg mL−1), previously ultrasonicated for 1 week, was drop-cast onto the electrode and allowed to dry at room temperature [46]. Afterwards, the so prepared MWCNTs/SPG were further modified by drop-casting 5 µL of Hemin solution (1 mg mL−1 in DMF) and allowed to dry at room temperature [47, 48]. Afterwards, the Hemin-MWCNTs/SPG electrodes were modified with 5 µL of AOx. Finally, 5 µL of a BSA solution (10 mg mL−1) was deposited onto the electrode to prevent the electrode biofouling and interferences. The electrodes were further conditioned overnight in 10 mM HEPES buffer pH 7.4 at + 4 °C.
Iontophoretic electrode preparation and protocol
The iontophoretic electrodes were obtained by stencil-printing LOCTITE® ECI 1010 E&C silver ink with a geometric area of 2 cm2 and square shape measuring 20 × 10 mm and cured according to manufacturing instructions [49]. Both electrodes were chlorinized by using 0.1 M FeCl3 solution for 30 s. After cutting, the anode electrode was covered with an anodic iontophoretic hydrogel containing 4% agarose gel, 0.1% oxidized nicotinamide adenine dinucleotide (NAD+), and 2% pilocarpine, while the cathode electrode was situated 0.2 cm away from the anode and coated with a cathodic hydrogel comprising 5% PVA and 1% potassium nitrate. The iontophoresis protocol was optimized to generate sweat volume and ensure skin response to iontophoretic current. Sweat generation was achieved by delivering pilocarpine across the skin using an optimized electrical current density of approximately 0.4 mA cm−2 flowing from the anodic to the cathodic compartment for 10 min.
Equipment and measurements
Cyclic voltammetry (CV), chronoamperometry, open circuit potential (OCP), and linear sweep voltammetry (LSV, polarization curves) experiments were conducted using a PalmSens4 electrochemical workstation equipped with PSTrace 5.6v software. All potentials were referenced to a BASi Ag|AgCl|KCl electrode (all potential values are reported in the paper considering this reference) and a platinum wire was used as the reference and counter electrode, respectively. Stencil-printed graphite (SPG) electrodes with a geometric area of 9 mm2 and square shape measuring 3 × 3 mm were employed as the working electrodes. Polarization curves were recorded using linear sweep voltammetry (LSV) in potentiostatic mode at a scan rate of 1 mV s−1, ranging from the open circuit voltage (OCV) to 0 V. The bioanode, ADH/pMB-MWCNTs/SPG, was employed as the working electrode, while AOx/Hemin-MWCNTs/SPG served as a combined reference and counter electrode. The polarization curves were recorded in 10 mM HEPES pH 7 with different ethanol concentrations (10, 50, 100, 150, 250, 300, 500, 1000, 2500, 5000, 25,000, and 50,000 µM). The temperature-controlled experiments were conducted using a cryostatic bath (LAUDA RM6, Delran, NJ, USA) with a precision of ± 0.01 °C.
Morphological characterization was carried out using a field emission scanning electron microscope (FE-SEM), model Σigma Zeiss (Jena, Germany). The images were captured using the in-lens detector, with a 5-kV acceleration voltage, 4-mm working distance, and 30-µm aperture, in top-view mode without any additional sample treatment.
Volunteer for experiments in sweat
Three male and three female volunteers, aged 30 and seemingly in good health, participated in all measurements and procedures described herein, posing minimal health risks. Written informed consent was obtained from the volunteers. Personal data handling adhered to the regulations outlined in the GDPR law 675/1996, in line with Directive 95/46/EC, aimed at safeguarding personal integrity during data processing. The research project and data collection present no risk of harm, and all information collected is publicly available.
Results and discussion
Electrochemical and SEM characterization of pMB modified MWCNTs/SPG electrodes
The influence of the redox mediator on the analytical performance of the anodic electrode was investigated by electrodepositing pMB by cyclic voltammetry and potentiostatic pulse method. Notably, Fig. 2A reports CVs for pMB electrodeposition performed by scanning SPG electrodes between − 0.4 and + 1.2 V at 50 mV s−1 as scan rate (10 cycles). Two redox peaks are observed: a cathodic peak at − 0.22 V and an anodic peak at − 0.20 V in curve 1. These redox peaks are attributed to the reduction and oxidation of methylene blue. The current of the anodic peak at − 0.20 V decreases rapidly, and a new anodic peak emerges at approximately − 0.06 V in the second cycle, with its current increasing gradually over subsequent potential cycles. The potential of this latter peak shifts to 0.10 V by the seventh cycle. Following electrolysis, a thin film with a greenish hue is observed on the working SPG electrode [50,51,52]. Alternatively, the potentiostatic pulse method was adopted as described in Fig. 2B. The electrodeposition of pMB was conducted using the following parameters: an oxidation potential of + 1.2 V and an oxidation pulse duration of 1 s, a reduction potential of − 0.4 V, a reduction pulse duration of 1 s, and an interval time of 0.2 s in both steps. Figure 2C reports the potentiostatic pulsed electrodeposition for 250 cycles, where the pulses achieved steady-state current values after 8 cycles probably due pMB homogeneous surface coverage (additional cycles lead to multilayer deposition and increased nanostructuration). Both types of anodic electrodes, namely pMB-MWCNTs/SPG obtained with cyclic voltammetry and potentiostatic pulsed method, were investigated in 10 mM HEPES buffer pH 7.2 (containing 100 mM KCl as supporting electrolyte) as non-turnover experiment, and in the presence of 1 mM NADH to evaluate their catalytic efficiency (Fig. 2D). Notably, pM-MWCNTs/SPG electrodes obtained by cyclic voltammetry and potentiostatic pulsed method reported a similar CV in 10 mM HEPES buffer pH 7.2 (containing 0.1 M KCl), highlighting the presence of a couple of redox peaks with a formal potential (E0′) of − 0.06 V (black curve). After adding 1 mM NADH, pM-MWCNTs/SPG obtained by cyclic voltammetry (red curve, modified with 10 electrodeposition cycles) and potentiostatic pulsed method (blue curve, 200 potentiostatic pulsed cycles) reported a catalytic current (Icat) of 524 ± 23 µA and 342 ± 18 µA at + 0.2 V, respectively, probably due to higher polymer surface coverage.
Furthermore, both electrodeposition methods were optimized keeping constant oxidation/reduction potentials and times, as reported in the literature, but varying the number of CVs (Fig. 2E) and potentiostatic pulsed method (Fig. 2F) [53]. Notably, pMB-MWCNTs/SPG electrodes obtained by cyclic voltammetry showed the highest catalytic current for NADH oxidation after 10 cycles (524 ± 23 µA). Notably, the catalytic current was decreasing with the number of electrodeposition cycles (112 ± 9 µA with 35 cycles). This can be ascribed to pMB multilayer formation that might not be electrically connected with the electrode surface. pMB-MWCNTs/SPG electrodes obtained by the potentiostatic pulsed method showed a similar trend with the highest catalytic current for NADH oxidation at 200 cycles (342 ± 18 µA). The optimized pMB-MWCNTs/SPG electrodes obtained by cyclic voltammetry (10 cycles) were further characterized by scanning electron microscopy (SEM) observing a bundle-like structure covered by pMB layers (Fig. 2G). Moreover, the optimized pMB-MWCNTs/SPG electrodes obtained by potentiostatic pulsed method (200 cycles) showed a homogeneously dispersed layer of MWCNTs covered by pMB layers. The different morphology can be ascribed to the different electrodeposition methods that led to MWCNTs reorganization onto the electrode surface and different pMB layers growth process. Hence, the optimized pMB-MWCNTs/SPG electrodes obtained by cyclic voltammetry (10 cycles) were selected as potential bioanode electrodes for the development of ethanol self-powered biosensor.
Electrochemical characterization of bioanode ADH/pMB-MWCNTs/SPG
The pMB-MWCNTs/SPG electrodes, prepared as described, were further modified with alcohol dehydrogenase (ADH) from Saccharomyces cerevisiae to create second-generation ethanol biosensors. In Fig. 3A, cyclic voltammograms (CVs) are depicted under both non-turnover (black curve) and turnover conditions (with the addition of 5, 10, 15, 20, and 50 mM ethanol, represented by purple, green, magenta, blue, and red curves, respectively). Under non-turnover conditions, the ADH/pMB-MWCNTs/SPG electrodes exhibited a pair of redox peaks with a formal potential E0′ = − 0.067 V attributed to the pMB layer deposited onto MWCNTs/SPG electrodes (Fig. 3A, black curve). Upon substrate addition, a significant, mass-transfer-limited electrocatalytic response commencing at EONSET = − 0.08 V with a maximum current of 112 µA (5 mM ethanol, purple curve), 174 µA (10 mM ethanol, purple curve), 307 µA (15 mM ethanol, purple curve), 434 µA (20 mM ethanol, purple curve), and 518 µA (50 mM ethanol, purple curve) at E = + 0.2 V was observed (Fig. 3A, purple, green, magenta, blue, and red curves, respectively). The limitation in mass transfer is associated with the formation of bundles consisting of pMB-MWCNTs, which improves the catalytic efficiency for NADH oxidation. Meanwhile, the roughness and porosity of SPG electrodes regulate the diffusion occurring at the electrode surface.
The modified electrodes underwent chronoamperometry testing by incrementally increasing substrate concentration within the range of 0–50 mM for ethanol (Fig. 3B).
The calibration curve derived from the chronoamperometric response of the ADH/pMB-MWCNTs/SPG electrodes (spanning overall the 1 × 10−5–5 × 10−2 M), as depicted in Fig. 3C, revealed a linear range from 0.01 to 0.3 mM (Fig. 3C, inset), with a detection limit (LOD) of 3 ± 1 µM (calculated using LOD = 3.3 × (σ/S) where σ represents the absolute standard deviation of the intercept, and S denotes the slope of the calibration curve), and a sensitivity of 326 ± 28 μA mM−1, accompanied by a correlation coefficient of 0.99 (RSD 7.2%, n = 10). Furthermore, the calibration curve was fitted to ascertain the classical Michaelis–Menten kinetic parameters, resulting in an Imax of 653 ± 24 µA and an apparent Michaelis–Menten constant (KMapp) of 19 ± 4 mM (approximately 200 times higher than KM measured in solution) [54]. This discrepancy could be attributed to the controlled diffusion of enzymatic product (i.e., NADH) through the rough electrode surface, typically leading to an extended linear range. However, the presence of pMB-MWCNTs bundles may have impeded their availability for surface catalytic reactions, where the enzyme is adsorbed. The analytical performance of the proposed electrode platform could be attributed to the low quantity of enzyme effectively immobilized onto the electrode surface and the limited accessibility of the pMB layer for catalytic NADH oxidation. The influence of pH and temperature on the performance of ADH/pMB-MWCNTs/SPG electrodes was investigated. To cover a broad pH spectrum from 4.5 to 10, three different buffers, namely acetate, MOPS, and HEPES, were utilized. The current signal exhibited an upward trend with increasing pH, peaking at pH 6.5 (218 ± 11 µA), and subsequently declining as the pH surpassed 6.5 (Fig. 3D). Similarly, the impact of temperature was evaluated in 10 mM HEPES buffer at pH 6.5 (containing 100 mM KCl). The findings revealed an optimal temperature range of 25–35 °C for ADH/pMB-MWCNTs/SPG electrodes (Fig. 3E), consistent with previously reported literature. The selectivity of the ADH/pMB-MWCNTs/SPG electrodes was assessed to gauge the effect of potential interfering compounds on their response (Fig. 3F). The signal obtained at a fixed ethanol concentration was compared with signals obtained with equivalent quantities of various potential interfering compounds, including d-fructose, d-galactose, d-mannitol, d-glucose, l-lactate, ascorbic acid, uric acid, and dopamine. The outcomes indicated that there was no noteworthy current response for the potential interfering compounds due to the presence of a BSA protecting layer mainly acting as diffusional barrier and its isoelectric point.
Electrochemical characterization of biocathode AOx/Hemin-MWCNTs/SPG
Moreover, Hemin-MWCNTs/SPG electrodes were investigated in 10 mM HEPES buffer pH 7.2 (containing 100 mM KCl as supporting electrolyte) as non-turnover condition, and in the presence of 1 mM H2O2 to evaluate their catalytic efficiency towards H2O2 reduction to H2O (Fig. 4A). Particularly, Hemin-MWCNTs/SPG electrodes showed a CV in 10 mM HEPES buffer pH 7.2 (containing 0.1 M KCl), without reporting any redox wave (black curve). After adding 1 mM H2O2, Hemin-MWCNTs/SPG electrodes reported a catalytic current (Icat) of − 58 ± 6 µA at E = + 0.1 V. The Hemin-MWCNTs/SPG electrodes, prepared as described, were further modified with alcohol oxidase (AOx) from Pichia pastoris to create first-generation ethanol biosensors able to work as biocathode in an enzymatic fuel cell due to the high operational voltage of Hemin as catalyst for H2O2 reduction [47, 48]. In Fig. 4B, cyclic voltammograms (CVs) are depicted under both non-turnover (black curve) and turnover conditions (with the addition of 10 mM ethanol, red curve). Under non-turnover conditions, the AOx/Hemin-MWCNTs/SPG electrodes did not exhibit any redox wave (Fig. 4B, black curve). Upon addition of 10 mM ethanol, a catalytic response commencing at EONSET = + 0.36 V with a maximum current of − 19 ± 2 µA at E = + 0.1 V was observed (Fig. 4B, red curve).
The modified electrodes were subjected to chronoamperometry testing by gradually increasing substrate concentration within the range of 0–50 mM for ethanol (Fig. 4C). The calibration curve derived from the chronoamperometric response of the AOx/Hemin-MWCNTs/SPG electrodes (spanning overall the 1 × 10−5–5 × 10−2 M), as illustrated in Fig. 4D, showed a linear range from 0.01 to 0.3 mM (Fig. 4D, inset), with a detection limit (LOD) of 3 ± 1 µM, and a sensitivity of 7.2 ± 0.3 μA mM−1, accompanied by a correlation coefficient of 0.99 (RSD 5.8%, n = 10). Additionally, the calibration curve was fitted to determine the classical Michaelis–Menten kinetic parameters, resulting in an Imax of 16 ± 2 µA and an apparent Michaelis–Menten constant (KMapp) of 8.2 ± 0.9 mM, similar to other KMapp data reported in the literature. The analytical performance of the proposed electrode platform could be attributed to the intimate contact between the enzyme and the catalyst specifically acting towards H2O2 reduction. The impact of pH and temperature on the performance of AOx/Hemin-MWCNTs/SPG electrodes was examined. To cover a wide pH range from 4.5 to 10, three different buffers, namely acetate, MOPS, and HEPES, were employed. The current signal exhibited an upward trend with increasing pH, peaking at pH 7.5 (8.3 ± 0.7 µA), and subsequently declining as the pH surpassed 7.5 (Fig. 4E). Similarly, the effect of temperature was assessed in 10 mM HEPES buffer at pH 7.5 (containing 100 mM KCl). The findings indicated an optimal temperature of 35 °C for AOx/Hemin-MWCNTs/SPG electrodes (Fig. 4F), consistent with previously reported literature [55]. The selectivity of the AOx/Hemin-MWCNTs/SPG electrodes was evaluated to assess the influence of potential interfering compounds on their response (Fig. 4G). The signal obtained at a fixed ethanol concentration was compared with signals obtained with equivalent amounts of various potential interfering compounds, including d-fructose, d-galactose, d-mannitol, d-glucose, l-lactate, ascorbic acid, uric acid, and dopamine. The results indicated no significant current response for the potential interfering compounds due to the presence of a BSA protecting layer mainly acting as diffusional barrier and its isoelectric point [56].
Characterization of enzymatic fuel cell as ethanol self-powered biosensor
Finally, the selected bioelectrodes, ADH/pMB-MWCNTs/SPG and AOx/Hemin-MWCNTs/SPG, were assembled as bioanode (connected as the working electrode) and biocathode (connected as the reference-counter electrode) to obtain an enzymatic fuel cell (EFC) working as an ethanol self-powered biosensor [57]. The EFC was characterized by running polarization curves gradually increasing substrate concentration within the range of 0–50 mM for ethanol (Fig. 5A). ADH/pMB-MWCNTs/SPG||AOx/Hemin-MWCNTs/SPG EFC reported an OCV of 0.052 ± 0.007 V with a power output of 0.3 ± 0.1 µW at 0.044 V in non-turnover conditions (not shown) and 0.42 ± 0.004 V with a power output of 25.4 ± 1.1 µW at 0.33 V (green curve, 50 mM ethanol). The calibration curve derived from the LSV curves of the AOx/Hemin-ADH/pMB-MWCNTs/SPG||AOx/Hemin-MWCNTs/SPG EFC (spanning overall the 1 × 10−5–5 × 10−2 M), as illustrated in Fig. 5B, showed a linear range from 0.01 to 0.3 mM (Fig. 5B, inset), with a detection limit (LOD) of 3 ± 1 µM, and a sensitivity of 64 ± 2 μW mM−1, accompanied by a correlation coefficient of 0.98 (RSD 8.1%, n = 10 couple of electrodes).
Both the anode and cathode electrodes, namely ADH/pMB-MWCNTs/SPG and AOx/Hemin-MWCNTs/SPG, were characterized by open circuit potentiometry (OCP) to assess their individual operational stability over 2800 s, as shown in Fig. 5C (anode: red curve and cathode: black curve). After the addition of 5 mM ethanol as substrate, ADH/pMB-MWCNTs/SPG electrodes showed an OCP of − 0.168 ± 0.008 V and AOx/Hemin-MWCNTs/SPG electrodes showed an OCP of 0.341 ± 0.004 V, displaying an increased stability over 2800 s, probably due to the porosity of the electrodes, which affects the amount of enzyme molecules immobilized and its stability. After assembling the EFC, the OCV resulted to be 0.516 ± 0.002 V (blue curve), which is in good agreement with the OCV that can be calculated as a difference between the OCP curves measured individually, namely 0.509 V.
The impact of pH and temperature on the performance of ADH/pMB-MWCNTs/SPG||AOx/Hemin-MWCNTs/SPG EFC was examined. To cover a wide pH range from 4.5 to 10, three different buffers, namely acetate, MOPS, and HEPES, were employed. The power output signal exhibited an upward trend with increasing pH, peaking at pH 7 (24.3 ± 1.7 µW), and subsequently declining as the pH surpassed 7 (Fig. 5D). Similarly, the effect of temperature was assessed in 10 mM HEPES buffer at pH 7 (containing 100 mM KCl). The findings indicated an optimal temperature of 35 °C for ADH/pMB-MWCNTs/SPG||AOx/Hemin-MWCNTs/SPG EFC (Fig. 5E), consistent with previously reported literature [57]. The selectivity of the ADH/pMB-MWCNTs/SPG||AOx/Hemin-MWCNTs/SPG EFC was evaluated to assess the influence of potential interfering compounds on their response (Fig. 5F). The signal obtained at a fixed ethanol concentration was compared with signals obtained with equivalent amounts of various potential interfering compounds, including d-fructose, d-galactose, d-mannitol, d-glucose, l-lactate, ascorbic acid, uric acid, and dopamine. The results indicated no significant current response for the potential interfering compounds due to the presence of a BSA protecting layer mainly acting as diffusional barrier and its isoelectric point [56]. Besides the preliminary analytical features, the storage stability of the proposed platforms was tested by recording the power output response for 20 consecutive measurements every day over a period of 90 days. The stability measurements were performed by continuously supplying 5 mM ethanol (Fig. 5G) dispensed through a FIA system. In particular, ADH/pMB-MWCNTs/SPG||AOx/Hemin-MWCNTs/SPG EFC reported a response drop of its initial signal of 7% after 25 days, probably because of the enzyme intrinsic stability and the porosity/roughness of SPG electrodes, which is stabilizing the enzymatic layer. Table 1 summarizes the results obtained for other self-powered ethanol biosensors. In this context, one must also take into account the variability in composition observed in peripheral bodily fluids, especially concerning electrolytes, which could impact the consistency of sweat analyses. Other platforms reported wider linear ranges (mM concentrations) but higher LOD compared to the platform herein proposed [57,58,59,60,61]. The proposed self-powered biosensor was integrated within a wristband and integrated with a system to induce/control the perspiration rate and applying a Nafion membrane to control the diffusion at the electrode surface reporting a LOD of 0.1 g/L and the linear range to 0.5–14 g/L that enables the quantification at the threshold EBAC in most of the countries in the world. In this regard, the proposed biosensor showed higher LOD but wider linear range compared with other platforms previously reported probably due to the integration of a iontophoretic system to induce/control the sweating rate [62,63,64].
Alcohol self-powered biosensor integrated in a wrist band
After preliminary analytical characterization of both bioanode and biocathode electrodes separately and assembled as EFC, they were integrated within a wristband to perform continuous alcohol abuse monitoring in sweat. As shown in Fig. 6, both ADH/pMB-MWCNTs/SPG and AOx/Hemin-MWCNTs/SPG electrodes were placed within the rubber wrist band together with printed silver AgCl electrodes to induce skin perspiration through iontophoretic process. The recess within the rubber wristband created an electrochemical cell with a thickness of 2 mm, which enabled sweat accumulation. The wrist band was worn by three voluntary healthy male and female patients.
Figure 6 shows the power output recording for ethanol detection (red curve) in sweat measured for three healthy male volunteers (blue curve) and three healthy female volunteers (red curve), which was correlated with estimated blood alcohol content in g L−1 (scatter point black lines). A method formulated by Swedish professor Erik Widmark in the 1920s can be utilized to approximate the estimated blood alcohol content (EBAC) [65]. This was calculated using Eq. 1:
where A represents the quantity of alcohol ingested (in g), T signifies the duration for which alcohol remained in the bloodstream (typically the time elapsed since consumption commenced), measured in hours, β denotes the elimination rate of alcohol, usually averaging about 0.15 g/L/h, and Vd indicates the volume of distribution (in L), typically calculated as body weight (in kg) multiplied by 0.71 L/kg for males and 0.58 L/kg for females. Considering three healthy volunteer male patients with a weight of approximately 80 kg and drinking a total of 28 g of alcohol, the EBAC profile showed a maximum of 0.50 g L−1 decreasing over 2 h (black scattered curve). Similarly, three healthy volunteer female patients with a weight of approximately 60 kg and drinking a total of 15 g of alcohol, the EBAC profile showed a maximum of 0.43 g L−1 decreasing over 2 h (black scattered curve). A lag time of 15 min was observed in sweat EBAC output recording probably due to the activation of iontophoretic process and metabolites transport within the peripheral blood microcirculation system [45].
Conclusions
In this study the first example of self-powered ethanol biosensor based on water-based graphite inks modified with MWCNTs and using ethanol as input for both EFC compartments (bioanode and biocathode) was reported. The self-powered biosensor was assembled using ADH/pMB-MWCNTs/SPG||AOx/Hemin-MWCNTs/SPG. The self-powered biosensor was characterized by running polarization curves gradually increasing substrate concentration within the range of 0–50 mM for ethanol, reporting a linear range from 0.01 to 0.3 mM, with a detection limit (LOD) of 3 ± 1 µM, and a sensitivity of 64 ± 2 μW mM−1, accompanied by a correlation coefficient of 0.98 (RSD 8.1%, n = 10 couple of electrodes). The biodevice showed a good operational stability (over 2800 s with continuous ethanol turnover) and an excellent storage stability (~ 93% of initial signal retained after 90 days). Finally, the proposed array was integrated in a wristband and successfully tested for the continuous monitoring of alcohol abuse.
The suggested system exhibits encouraging characteristics for utilization as a versatile and wearable biosensor using biocompatible water-based inks, which could be applied in forensic scenarios.
References
Kim J, Campbell AS, de Ávila BE-F, Wang J. Wearable biosensors for healthcare monitoring. Nat Biotechnol. 2019;37:389–406.
Reid RC, Mahbub I. Wearable self-powered biosensors. Curr Opin Electrochem. 2020;19:55–62.
Jeerapan I, Sempionatto JR, Wang J. On-body bioelectronics: wearable biofuel cells for bioenergy harvesting and self-powered biosensing. Adv Funct Materials. 2020;30:1906243. https://doi.org/10.1002/adfm.201906243.
Zeng X, Peng R, Fan Z, Lin Y. Self-powered and wearable biosensors for healthcare. Mater Today Energy. 2022;23:100900.
Song Y, Mukasa D, Zhang H, Gao W. Self-powered wearable biosensors. Acc Mater Res. 2021;2:184–97. https://doi.org/10.1021/accountsmr.1c00002.
Kim J, Kumar R, Bandodkar AJ, Wang J. Advanced materials for printed wearable electrochemical devices: a review. Adv Elect Materials. 2017;3:1600260. https://doi.org/10.1002/aelm.201600260.
Yáñez-Sedeño P, Campuzano S, Pingarrón JM. Screen-printed electrodes: promising paper and wearable transducers for (bio) sensing. Biosensors. 2020;10:76.
Bollella P. Enzyme-based amperometric biosensors: 60 years later… Quo Vadis? Anal Chim Acta. 2022;1234:340517.
Grattieri M, Minteer SD. Self-powered biosensors. ACS Sensors. 2018;3:44–53.
Conzuelo F, Ruff A, Schuhmann W. Self-powered bioelectrochemical devices. Curr Opin Electrochem. 2018;12:156–63.
Katz E, Bückmann AF, Willner I. Self-powered enzyme-based biosensors. J Am Chem Soc. 2001;123:10752–3.
Zhang S, Cicoira F. Flexible self-powered biosensors. Nature. 2018;561:466–7.
Imbriano A, Tricase A, Macchia E, Torsi L, Bollella P. Self-powered logically operated fluorescent detection of hepatitis B virus (HBV). Anal Chim Acta. 2023;1252:341037.
Sailapu SK, Macchia E, Merino-Jimenez I, Esquivel JP, Sarcina L, Scamarcio G, Minteer SD, Torsi L, Sabaté N. Standalone operation of an EGOFET for ultra-sensitive detection of HIV. Biosens Bioelectron. 2020;156:112103. https://doi.org/10.1016/j.bios.2020.112103.
Bollella P, Gorton L. Enzyme based amperometric biosensors. Curr Opin Electrochem. 2018;10:157–73.
Gonzalez-Solino C, Bernalte E, Bayona Royo C, Bennett R, Leech D, Di Lorenzo M. Self-powered detection of glucose by enzymatic glucose/oxygen fuel cells on printed circuit boards. ACS Appl Mater Interfaces. 2021;13:26704–11. https://doi.org/10.1021/acsami.1c02747.
Zumpano R, Lambertini L, Tortolini C, Bollella P, Favero G, Antiochia R, Mazzei F. A glucose/oxygen enzymatic fuel cell exceeding 1.5 V based on glucose dehydrogenase immobilized onto polymethylene blue-carbon nanotubes modified double-sided screen printed electrodes: proof-of-concept in human serum and saliva. J Power Sources. 2020;476:228615.
Bollella P, Boeva Z, Latonen R-M, Kano K, Gorton L, Bobacka J. Highly sensitive and stable fructose self-powered biosensor based on a self-charging biosupercapacitor. Biosens Bioelectron. 2021;176:112909.
Camargo JR, Silva TA, Rivas GA, Janegitz BC. Novel eco-friendly water-based conductive ink for the preparation of disposable screen-printed electrodes for sensing and biosensing applications. Electrochim Acta. 2022;409:139968.
Camargo JR, Orzari LO, Araujo DAG, de Oliveira PR, Kalinke C, Rocha DP, dos Santos AL, Takeuchi RM, Munoz RAA, Bonacin JA. Development of conductive inks for electrochemical sensors and biosensors. Microchem J. 2021;164:105998.
Tricase A, Imbriano A, Valentino M, Ditaranto N, Macchia E, Franco CD, Kidayaveettil R, Leech D, Piscitelli M, Scamarcio G, Perchiazzi G, Torsi L, Bollella P. Water‐based conductive ink formulations for enzyme‐based wearable biosensors. Adv Sensor Res. 2023;2300036. https://doi.org/10.1002/adsr.202300036.
Marchianò V, Tricase A, Caputo M, Farinini E, Leardi R, Imbriano A, Leech D, Kidayaveettil R, Gentile L, Torsi L, Macchia E, Bollella P. Tailoring water-based graphite conductive ink formulation for enzyme stencil-printing: experimental design to enhance wearable biosensor performance. Chem Mater. 2024;36:358–70. https://doi.org/10.1021/acs.chemmater.3c02229.
Marchianò V, Tricase A, Ditaranto N, Macchia E, d’Ingeo S, Di Franco C, Scamarcio G, Torsi L, Bollella P. High voltage flexible glucose/O2 fully printed hydrogel-based enzymatic fuel cell. J Phys D Appl Phys. 2024;57:135503.
Silveri F, Paolini D, Della Pelle F, Bollella P, Scroccarello A, Suzuki Y, Fukawa E, Sowa K, Di Franco C, Torsi L, Compagnone D. Lab-made flexible third-generation fructose biosensors based on 0D-nanostructured transducers. Biosens Bioelectron. 2023;237:115450. https://doi.org/10.1016/j.bios.2023.115450.
Bollella P, Katz E. Enzyme-based biosensors: tackling electron transfer issues. Sensors. 2020;20:3517.
Gao W, Emaminejad S, Nyein HYY, Challa S, Chen K, Peck A, Fahad HM, Ota H, Shiraki H, Kiriya D, Lien D-H, Brooks GA, Davis RW, Javey A. Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis. Nature. 2016;529:509–14. https://doi.org/10.1038/nature16521.
Gao W, Ota H, Kiriya D, Takei K, Javey A. Flexible electronics toward wearable sensing. Acc Chem Res. 2019;52:523–33. https://doi.org/10.1021/acs.accounts.8b00500.
Min J, Sempionatto JR, Teymourian H, Wang J, Gao W. Wearable electrochemical biosensors in North America. Biosens Bioelectron. 2021;172:112750.
Paolini D, Della Pelle F, Scroccarello A, Silveri F, Bollella P, Ferraro G, Fukawa E, Suzuki Y, Sowa K, Torsi L, Compagnone D. Enhanced electron transfer efficiency of fructose dehydrogenase onto roll-to-roll thermal stamped laser-patterned reduced graphene oxide films. ACS Appl Mater Interfaces. 2024;16:22443–54. https://doi.org/10.1021/acsami.4c03339.
Araujo DA, Camargo JR, Pradela-Filho LA, Lima AP, Munoz RA, Takeuchi RM, Janegitz BC, Santos AL. A lab-made screen-printed electrode as a platform to study the effect of the size and functionalization of carbon nanotubes on the voltammetric determination of caffeic acid. Microchem J. 2020;158:105297.
Qin Y, Ouyang X, Lv Y, Liu W, Liu Q, Wang S. A review of carbon-based conductive inks and their printing technologies for integrated circuits. Coatings. 2023;13:1769. https://doi.org/10.3390/coatings13101769.
Xu J, Fang Y, Chen J. Wearable biosensors for non-invasive sweat diagnostics. Biosensors. 2021;11:245.
Xuan X, Pérez-Ràfols C, Chen C, Cuartero M, Crespo GA. Lactate biosensing for reliable on-body sweat analysis. ACS Sens. 2021;6:2763–71. https://doi.org/10.1021/acssensors.1c01009.
Jo S, Sung D, Kim S, Koo J. A review of wearable biosensors for sweat analysis. Biomed Eng Lett. 2021;11:117–29.
Semenova D, Gernaey KV, Morgan B, Silina YE. Towards one-step design of tailored enzymatic nanobiosensors. Analyst. 2020;145:1014–24.
Bartlett PN, Simon E, Toh CS. Modified electrodes for NADH oxidation and dehydrogenase-based biosensors. Bioelectrochemistry. 2002;56:117–22.
Adachi T, Miyata T, Makino F, Tanaka H, Namba K, Kano K, Sowa K, Kitazumi Y, Shirai O. Experimental and theoretical insights into bienzymatic cascade for mediatorless bioelectrochemical ethanol oxidation with alcohol and aldehyde dehydrogenases. ACS Catal. 2023;13:7955–65. https://doi.org/10.1021/acscatal.3c01962.
Stasyuk N, Demkiv O, Gayda G, Zakalska O, Nogala W, Gonchar M. Amperometric biosensors based on alcohol oxidase and peroxidase–like nanozymes for ethanol determination. Microchim Acta. 2022;189:474.
Smutok O, Kavetskyy T, Prokopiv T, Serkiz R, Wojnarowska-Nowak R, Šauša O, Novák I, Berek D, Melman A, Gonchar M. New micro/nanocomposite with peroxidase-like activity in construction of oxidases-based amperometric biosensors for ethanol and glucose analysis. Anal Chim Acta. 2021;1143:201–9.
Ribeaucourt D, Bissaro B, Guallar V, Yemloul M, Haon M, Grisel S, Alphand V, Brumer H, Lambert F, Berrin J-G, Lafond M. Comprehensive insights into the production of long chain aliphatic aldehydes using a copper-radical alcohol oxidase as biocatalyst. ACS Sustain Chem Eng. 2021;9:4411–21. https://doi.org/10.1021/acssuschemeng.0c07406.
Carucci C, Raccis F, Salis A, Magner E. Specific ion effects on the enzymatic activity of alcohol dehydrogenase from Saccharomyces cerevisiae. Phys Chem Chem Phys. 2020;22:6749–54.
Mishra RK, Sempionatto JR, Li Z, Brown C, Galdino NM, Shah R, Liu S, Hubble LJ, Bagot K, Tapert S. Simultaneous detection of salivary Δ9-tetrahydrocannabinol and alcohol using a wearable electrochemical ring sensor. Talanta. 2020;211:120757.
Olmedo-Aguirre JO, Reyes-Campos J, Alor-Hernández G, Machorro-Cano I, Rodríguez-Mazahua L, Sánchez-Cervantes JL. Remote healthcare for elderly people using wearables: a review. Biosensors. 2022;12:73.
Vaghasiya JV, Mayorga-Martinez CC, Pumera M. Wearable sensors for telehealth based on emerging materials and nanoarchitectonics. npj Flex Electron. 2023;7:26.
Min J, Tu J, Xu C, Lukas H, Shin S, Yang Y, Solomon SA, Mukasa D, Gao W. Skin-interfaced wearable sweat sensors for precision medicine. Chem Rev. 2023;123:5049–138. https://doi.org/10.1021/acs.chemrev.2c00823.
Bollella P, Hibino Y, Kano K, Gorton L, Antiochia R. Enhanced direct electron transfer of fructose dehydrogenase rationally immobilized on a 2-aminoanthracene diazonium cation grafted single-walled carbon nanotube based electrode. ACS Catal. 2018;8:10279–89.
Koushanpour A, Gamella M, Katz E. A biofuel cell based on biocatalytic reactions of lactate on both anode and cathode electrodes – extracting electrical power from human sweat. Electroanalysis. 2017;29:1602–11. https://doi.org/10.1002/elan.201700126.
Koushanpour A, Gamella M, Guz N, Katz E. A biofuel cell based on biocatalytic reactions of glucose on both anode and cathode electrodes. Electroanalysis. 2017;29:950–4. https://doi.org/10.1002/elan.201600706.
Bolat G, De la Paz E, Azeredo NF, Kartolo M, Kim J, de Loyola e Silva AN, Rueda R, Brown C, Angnes L, Wang J. Wearable soft electrochemical microfluidic device integrated with iontophoresis for sweat biosensing. Anal Bioanal Chem. 2022;414:5411–21.
Liu J, Mu S. The electrochemical polymerization of methylene blue and properties of polymethylene blue. Synth Met. 1999;107:159–65.
Karyakin AA, Karyakina EE, Schuhmann W, Schmidt H-L, Varfolomeyev SD. New amperometric dehydrogenase electrodes based on electrocatalytic NADH-oxidation at poly (methylene blue)-modified electrodes. Electroanalysis. 1994;6:821–9.
Karyakin AA, Karyakina EE, Schmidt H-L. Electropolymerized azines: a new group of electroactive polymers. Electroanalysis Int J Devoted Fundam Pract Asp Electroanal. 1999;11:149–55.
Karyakin AA, Bobrova OA, Karyakina EE. Electroreduction of NAD+ to enzymatically active NADH at poly(neutral red) modified electrodes. J Electroanal Chem. 1995;399:179–84. https://doi.org/10.1016/0022-0728(95)04300-4.
Razumien J, Gurevičien V, Laurinavičius V, Gražulevičius JV. Amperometric detection of glucose and ethanol in beverages using flow cell and immobilised on screen-printed carbon electrode PQQ-dependent glucose or alcohol dehydrogenases. Sens Actuators B Chem. 2001;78:243–8.
Das M, Goswami P. Direct electrochemistry of alcohol oxidase using multiwalled carbon nanotube as electroactive matrix for biosensor application. Bioelectrochemistry. 2013;89:19–25.
Ma C, Yang H, Zhou X, Wu B, Zhang G. Polymeric material for anti-biofouling. Colloids Surf B. 2012;100:31–5.
Franco JH, Minteer SD, De Andrade AR. Ethanol biofuel cells: hybrid catalytic cascades as a tool for biosensor devices. Biosensors. 2021;11:41. https://doi.org/10.3390/bios11020041.
Ramanavicius A, Kausaite A, Ramanaviciene A. Enzymatic biofuel cell based on anode and cathode powered by ethanol. Biosens Bioelectron. 2008;24:761–6.
Bonfin CS, Franco JH, de Andrade AR. Ethanol bioelectrooxidation in a robust poly (methylene green-pyrrole)-mediated enzymatic biofuel cell. J Electroanal Chem. 2019;844:43–8.
Franco JH, Klunder KJ, Lee J, Russell V, de Andrade AR, Minteer SD. Enhanced electrochemical oxidation of ethanol using a hybrid catalyst cascade architecture containing pyrene-TEMPO, oxalate decarboxylase and carboxylated multi-walled carbon nanotube. Biosens Bioelectron. 2020;154:112077. https://doi.org/10.1016/j.bios.2020.112077.
Ruff A, Pinyou P, Nolten M, Conzuelo F, Schuhmann W. A self-powered ethanol biosensor. ChemElectroChem. 2017;4:890–7. https://doi.org/10.1002/celc.201600864.
Li B, Scott Downen R, Dong Q, Tran N, LeSaux M, Meltzer AC, Li Z. A discreet wearable IoT sensor for continuous transdermal alcohol monitoring—challenges and opportunities. IEEE Sens J. 2021;21:5322–30. https://doi.org/10.1109/JSEN.2020.3030254.
Khemtonglang K, Chaiyaphet N, Kumsaen T, Chaiyachati C, Chuchuen O. A smart wristband integrated with an IoT-based alarming system for real-time sweat alcohol monitoring. Sensors. 2022;22:6435. https://doi.org/10.3390/s22176435.
Sun M, Gu Y, Pei X, Wang J, Liu J, Ma C, Bai J, Zhou M. A flexible and wearable epidermal ethanol biofuel cell for on-body and real-time bioenergy harvesting from human sweat. Nano Energy. 2021;86:106061. https://doi.org/10.1016/j.nanoen.2021.106061.
Posey D, Mozayani A. The estimation of blood alcohol concentration: Widmark revisited. Forensic Sci Med Pathol. 2007;3:33–9.
Acknowledgements
The following funding agencies are acknowledged: NoOne-A binary sensor with single-molecule digit to discriminate biofluids enclosing zero or at least one biomarker, ERC Stg2021, GA:101040383; Centro di Innovazione in Single-Molecule Digital Assay – “Digital Assay” from Regione Puglia DGR N. 218 of 21/02/2022; 1LIVEXYLELLA – ID 01—Trasmissione D. M. n. 664519 del 28/12/2022.
Funding
Open access funding provided by Università degli Studi di Bari Aldo Moro within the CRUI-CARE Agreement.
Author information
Authors and Affiliations
Contributions
P.B. conceived the project and wrote the manuscript. V.M. and A.T. performed all electrochemical measurements. P.B. and E.M. directly supervised V.M. and A.T. during the project. L.T., E.M., and P.B. revised the manuscript and are responsible for funding acquisition. The final version was approved by all authors.
Corresponding author
Ethics declarations
Conflict of interest
The authors declare no competing interests.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
ABC Highlights: authored by Rising Stars and Top Experts.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Marchianò, V., Tricase, A., Macchia, E. et al. Self-powered wearable biosensor based on stencil-printed carbon nanotube electrodes for ethanol detection in sweat. Anal Bioanal Chem 416, 5303–5316 (2024). https://doi.org/10.1007/s00216-024-05467-7
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00216-024-05467-7