Abstract
Purpose
New energy-based sutureless vessel ligation devices, such as the Thunderbeat (Olympus Medical Systems Corp., Tokyo, Japan), could reduce operative time and limit blood loss in head and neck surgery; however, efficacy and safety in major head and neck surgery have not been investigated in a prospective, randomized study.
Methods
This prospective, double-arm, randomized controlled trial consisted of two parts: total laryngectomy (TL) and neck dissection (ND). Thirty patients planned for TL were randomized in two groups. For the ND part, forty-two operative sides were likewise randomized. In both parts, Thunderbeat was used in addition to the standard instrumentation in the intervention groups, while only standard instrumentation was used in the control groups. Primary outcome values were blood loss, operative time and complication rate.
Results
For the TL part there was no difference in mean blood loss (p = 0.062), operative time (p = 0.512) and complications (p = 0.662) between both hemostatic techniques. For the neck dissection part, there was a reduction in blood loss (mean 210 mL versus 431 mL, p = 0.046) and in operative time (median 101 (IQR 85–130) minutes versus 150 (IQR 130–199) minutes, p = 0.014) when Thunderbeat was used. There was no difference in complication rate between both hemostatic systems (p = 0.261).
Conclusion
The Thunderbeat hemostatic device significantly reduces operative blood loss and operative time for neck dissections, without increase in complications. In TL, blood loss using Thunderbeat was comparable with the standard technique, but the operative time tended to be shorter.
Trial registration
UMCG Research Register, Reg. no. 201700041, date of registration: 18/1/2017
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
A total laryngectomy (TL) with additional neck dissection (ND) is considered as major head and neck surgery. During head and neck surgery, precise hemostasis ensures a clear view on the surgical field. Minimalizing blood loss and operative time are beneficial both for the patient and cost-effectiveness [1]. Traditional instruments for dissecting and controlling hemostasis during head and neck surgery are bipolar forceps and monopolar scalpel, with the additional ligation or clipping for larger vessels. Some operative time is lost due to switching between operative instruments. Reduced operative time can be expected if the same device is used for both dissection and hemostasis.
The rate of overall complications during TL is estimated at 45% [2]. Because of the development of organ preserving strategies, an increased number of patients undergo a salvage TL after initial (chemo)radiation instead of primary TL. This is associated with a higher risk in complications of 42–67.5% [3, 4]. Post-operative complications entail bleeding, dysphagia, stomal complications and airway issues. The most common, serious and difficult to treat complication of TL is pharyngocutaneous fistula. Its incidence has a wide range from 3–65% [5]. It has a grave effect on postoperative functional recovery and quality of life. It usually leads to increased hospitalization and can lead to re-operations, decreased speech and swallowing function.
Complication rates for ND vary greatly according to the extent of the surgery and the levels involved [6], especially in salvage surgery [7]. Beside the general surgical complications, like bleeding and postoperative wound infection, ND can lead to loss of range of motion and muscle impairments of the shoulder due to sacrificing or damaging accessory nerve, loss of sensation, lymphedema and pain [4, 6, 8]. More effective hemostasis and a consequently better view on the surgical field may lead to less complications.
To achieve effective hemostasis during operation, new energy-based devices (EBD) have been developed, e.g., the advanced bipolar LigaSure Small Jawclamp (Medtronic) and the ultrasonic Harmonic Focus scalpel (Ethicon Endo-Surgery, Cincinnati, OH, USA). These devices are multi-purpose. Dissection, hemostasis and cutting can be performed with the same instrument, saving operative time. THUNDERBEAT Open Fine Jaw (TB, Olympus Medical Systems Corp., Tokyo, Japan), however, has a hybrid blade that combines both ultrasonically cutting, with bipolar coagulation. This combination enables coagulation and cutting at the same time [9,10,11]. It can be used for sealing and ligation of vessels up to a diameter of 7 mm.
EBD have been clinically evaluated for minimal invasive gynecological surgery, open thyroidectomy and free flap reconstructive surgery for the head and neck area. These studies suggest a decrease in hospital stay, blood loss and postoperative drainage [12,13,14]. However, there are no clinical studies evaluating TL and ND. We have, therefore, conducted a randomized controlled study to evaluate the safety and efficacy of the TB during total laryngectomy surgery and neck dissection in comparison with standard electrocautery devices (SED).
Methods
Ethical considerations
The study was checked by the Institutional Review Board of the University Medical Center Groningen (UMCG) and judged as according to the Dutch Medical Research Law, there is no need for Institutional Review Board approval; therefore, a waiver was released. The study was performed in accordance with Good Clinical Practice principles and the Consolidated Standards of Reporting Trials (CONSORT) statement.
Patient selection
For the TL part, consecutive patients with primary or recurrent laryngeal or hypopharyngeal cancer, scheduled for TL between Feb 2018 and Jun 2019 at a single tertiary center (UMCG, the Netherlands), were eligible for inclusion. During the same period at the same center, consecutive patients with head or neck cancer, scheduled for Radical Neck Dissection (RND), Modified Radical Neck Dissection (MRND) or Selective Neck Dissection (SND) were eligible for inclusion in the ND part. Selective neck was performed only in 2 patients in both groups and the surgery involved level 2–5 in all four patients (only level 1 was spared).
Randomization and blinding
For both parts of the study, computer-generated randomization was used to allocate patients equally in two groups. After allocation, the intervention group was operated on with the additional use of TB, while the control group was operated on with use of SED only.
Patients who underwent a ND in combination with TL were enrolled in both parts. Surgical staff and data assessors were not blinded.
Pre-operative data
The following characteristics were acquired and analyzed: gender, age, comorbidity based on the Adult Comorbidity Evaluation 27 (ACE-27) score [15], use of anti-coagulants pre-operatively and operatively (except for the use of prophylactic low molecular weight heparin), tumor site, previous treatment on surgical site, salvage surgery, pTNM-stage and disease stage.
Surgical technique
Patients were treated in accordance with the treatment protocol of the medical center. All procedures were performed by four senior head and neck oncologic surgeons.
Standard devices for cutting and hemostasis included monopolar and bipolar energy devices (ERBE GmbH, Germany), sutures and surgical clips. In the TB arm, the TB device was set at SEAL/CUT ‘1’ and SEAL ‘3’. Intra-operative blood loss was measured by weighing the gauzes per-operatively and subtracting the weight of the gauzes. One gram was assumed equal to one mL blood. Operative time of the whole procedure was recorded, as well as operative time of (hemi) thyroidectomy during TL.
Outcomes measures
Primary outcome measures were operative time, blood loss and post-operative complications (according to the Clavien–Dindo (CD) classification [16]). Major bleeding was defined as a bleeding that requires surgical intervention. The CD system is a validated classification system, grading postoperative surgical complications from grade I (minor deviation from normal postoperative course), through grade III (complications requiring a surgical intervention) to grade V (death of a patient). Secondary outcome measures were time interval between surgery and drain removal, post-operative drainage volume and length of hospital stay.
Statistical methods
Data was clustered for analysis if possible. For both parts, comorbidity score was clustered as ‘None/Mild’ (ACE-27 0–1) and ‘Moderate/Severe’ (ACE-27 2–4); pathological tumor stage was clustered as ‘Early disease’ (T1–T2) and ‘Advanced disease’ (T3–T4). For the TNM classification, the 8th edition of the UICC was used [17]. In all cases the pathological TNM (pTNM) was used; if it was not available, the clinical TNM (cTNM) was used. Results of complications (based on CD-classification) was clustered into no complications (0–1) and mild/major (2–5) complications.
For the TL part the site of primary tumor was clustered as larynx or pharynx; salvage surgery included cases with recurrent/residual disease after (chemo) radiation and also surgery for a dysfunctional larynx without signs of malignancy. For the ND part no clusters were made.
The distribution of continuous variables was assessed by the Shapiro–Wilk test. For normal distributed variables a mean and a standard deviation (SD) were assessed, for variables without normal distribution, the median and inter quartile range (IQR). Significance was determined for all continuous variables using the Mann–Whitney U test. For dichotomous nominal or categorical variables, a Chi-squared test was used to assess significance. In tests with a count > 20% below 5, the Fisher’s exact test was used. A p value of less than 0.05 was considered significant. All analysis were performed using IBM SPSS Statistics for Mac OS, 64-bit edition [18].
Results
TL part results
Figure 1a shows the process of patient enrolment and analysis. This part prospectively randomized thirty patients with the indication for total laryngectomy, with or without additional neck dissection and reconstructive surgery, in two groups. As shown in Fig. 1a, the intervention group consisted of fifteen patients and eleven patients in the control group. The demographic characteristics of the patients, tumor site and stage for both groups are listed in Table 1. The two groups were comparable for gender, age, comorbidity score, pre- and perioperative use of anticoagulants, tumor site, salvage surgery, T stage, N stage and disease stage. Significance for ‘previous treatment’ was not calculated due to small sample size.
Operative results
As shown in Table 2, no statistically significant differences of the primary and secondary outcome measures were found between the two cohorts. However, intraoperative blood loss was about half for the TB group compared to the SED group, for both the complete TL procedure (p = 0.06) and the (hemi) thyroidectomy procedure (p = 0.755). After excluding an outlier in the SED group, this result is not significantly different (TB + : 214 ± 203 mL, TB-: 368 ± 324 mL, p = 0.123 ±). In patients undergoing TL, with use of the TB, operative time had a slightly higher median and IQR compared to the SED group. For the thyroid surgery, a lower, but insignificant median and IQR for operative time was observed for the TB group.
Postoperative results
As shown in Table 2 the complication rate in the TB group was comparable with the SED group (53.3% compared to 61.5%). Hospital stay and time interval between surgery and first drain removal did not differ significantly. The post-operative drainage volume was insignificantly higher in the TB group; however, it did not result in statistically significant longer time until drain removal or extended hospitalization.
ND part results
This part prospectively randomized forty-two ND procedures of thirty-five patients, with or without additional TL, parotidectomy or reconstructive surgery, in two groups (Fig. 1b). The intervention group consisted of twenty-three and the control group of nineteen NDs. Due to incomplete intra-operative datasheets for three patients, the surgical data was analyzed for twenty-one patients in the intervention group and eighteen in the control group.
The demographic characteristics of the patients, tumor characteristics and operative procedure details for both groups are listed in Table 3. The two groups were comparable for gender, age, comorbidity score, anti-coagulation use, disease stage and primary versus salvage surgery. Statistical significance for ‘histology’, ‘site of primary tumor’, ‘previous treatment’, ‘nodal stage’ and ‘type of ND’ was not calculated due to low sample size.
Operative results
Regarding the primary outcomes, a significant difference was found between the two cohorts for operative time and blood loss during ND. Intra-operative blood loss was more than halved for the intervention group compared to the control group (TB + : 210 ± 209 mL, TB−: 431 ± 409 mL, p = 0.046). However, after exclusion of an outlier in the SED group, this result is not statistically significant (TB + : 210 ± 209 mL, TB−: 357 ± 269 mL, p = 0.078 ±). In patients undergoing ND with the use of TB, the median for operative time was forty-five minutes less than the control group [TB + : 101 (85–130) minutes, TB−: 150 (130–199), p = 0.014]. There were slightly more complications within the control group than the intervention group; however, this was not a statistically significant difference.
Postoperative results
Hospital stay, time interval between surgery and first drain removal and drainage volume did not differ statistically between the thirteen patients in the intervention group and fourteen patients in the control group (Table 4).
Discussion
This is the first study to analyze the effect of the TB device during major head and neck surgery, compared to SED in a prospective, randomized double-arm trial.
This study showed statistically significant reduction in operative time and blood loss when TB was used during ND. During TL, blood loss was reduced, but operative time was increased, although both not significantly when TB was used. As there was no difference in post-operative complications for both procedures, the TB can be regarded as a safe instrument for both TL and ND.
Perioperative blood loss and post-operative complications may lead to prolonged hospitalization, re-operation, delay in adjuvant therapy, etc. [1]. Studies by both Gambardella et al. concerning hemostasis in axillary lymph node dissection and Suzuki et al. concerning short-term outcomes for neck dissection, show a trend toward limiting blood loss with use of the TB and Ligasure in comparison with electrocautery [9, 10]. In line with these studies, we experienced a decrease in blood loss, however, statistically not significant. The differences within the TL part were likewise striking in term of blood loss and very likely statistically insignificant due to the small sample size.
Literature shows shorter operative time when using the TB compared to conventional electrosurgery in laparoscopic procedures, such as hysterectomy and lymphadenectomy [9, 19]. The contradictory results for the present TL part for operative time could be explained by several factors: limited sample size and the fact that the TB is less suitable in the most crucial part of the surgery, namely during the mucosal incisions. To ensure optimal healing of the mucosal wound, ultrasonic or bipolar energy is avoided when the pharyngeal mucosa in incised. However, during a specific part of the surgery, namely during dissection of the thyroid gland from the larynx, a beneficial effect of TB could be observed. For the ND, however, a clear benefit is shown in favor of TB. The difference in median operative time was almost fifty minutes, which could benefit the patient due to diminished anesthetics and increase cost-effectiveness. However, performing a reliable cost-effectiveness study was not possible, due to several factors. The costs of the operating room time highly differs among hospitals and the costs of the TB depends on an agreement with the distributor. However, despite of not giving exact cost calculations, the data of the present study may be used for cost calculations. Each professional can calculate the cost-effectiveness of purchasing such a device, by calculating the OR time costs and asking an offer from the local distributor.
Several studies have concluded that the TB can be as safe as other energy-based devices when used in laparoscopic surgery [20,21,22], thyroid surgery [11] and neck dissections [10]. This study has indicated that the TB is at least as safe as the SED during TL and ND, as the statistically insignificant difference was found in favor of the TB group.
The outcomes of this study regarding hospitalization seem comparable to earlier study results. In other studies, analyzing new energy-based devices within a range of surgical procedures, the duration of hospitalization was either the same or less. In a prospective trial regarding the efficacy of the UltraCision Harmonic scalpel in thyroidectomy by Papavramidis et al. a statistically significant shortening of hospitalization was observed of almost a day [13]. However, a meta-analysis of three RCTs also analyzing hospitalization after neck dissection with use of the Harmonic scalpel versus SED, concluded that the marginal shortening they found was not statistically significant [23]. In a retrospective non-inferiority study by van Slycke et al. investigating the hospital stay for patients undergoing a TB-assisted thyroidectomy versus SED, stated that the TB was non-inferior to SED for the aspect of hospitalization [11]. Fagotti et al. prospectively compared patients undergoing laparoscopic hysterectomy TB-assisted with SED, with no difference (p = 0.82) in duration hospitalization [19]. In this study, there was no statistically significant differences in hospital stay for the TB group in both arms. Interestingly, the remarkably lower blood loss had no measurable impact on hospital stay or complication, which is also very likely due to the limited sample size. To establish a certain non-inferiority or superiority, a larger sample size is required. However, the presented data prove that TB is safe (no more bleeding than operating with SED). On the other hand, it also gives honest results to prevent too high expectations of using TB.
Strengths and limitations
The strengths of the present study lie in its prospective, randomized design, evaluating two of the most common and elaborate procedures in head and neck surgery. During the study, treatment protocols were not changed. Furthermore, the same four skilled surgeons performed the surgery in all patients, limiting the learning curve for the use of the TB device to a minimum. Unfortunately, due to unforeseen circumstances, there was a loss of intra-operative data in several cases. The results have to be interpreted with caution, as this unblinded study included a limited number of patients and there is high deviation of the data (for instance for the blood loss).
Conclusions
This study demonstrates the safety of Thunderbeat compared to standard electrocautery devices in TL and ND. TB significantly reduces operative time and operative blood loss during ND. This effect could not be seen in TL surgery; however, the blood loss was reduced with more than 50% during surgery with TB versus SED. Due to the limited sample size, the results need to be carefully interpreted and larger, randomized, powered studies are necessary to confirm the suggested trend for blood loss in TL and establish whether the TB could further diminish operative time in major head and neck surgery.
Data availability
The authors affirm that this manuscript is an honest, accurate, and transparent account of the study being reported; that no important aspects of the study have been omitted; and that any discrepancies from the study as planned have been explained. The data that support the findings of this study are available on request from the corresponding author.
References
Farwell DG, Reilly DF, Weymuller J, Ernest A, Greenberg DL, Staiger TO, Futran NA (2002) Predictors of perioperative complications in head and neck patients. Arch Otolaryngol Head Neck Surg 128(5):505–511
Serbanescu-Kele CM, Halmos GB, Wedman J, van der Laan BF, Plaat BE (2015) Early feeding after total laryngectomy results in shorter hospital stay without increased risk of complications: a retrospective case-control study. Clin Otolaryngol 40(6):587–592
Genden EM, Ferlito A, Rinaldo A, Silver CE, Fagan JJ, Suárez C, Langendijk JA, Lefebvre JL, Bradley PJ, Leemans CR (2008) Recent changes in the treatment of patients with advanced laryngeal cancer. Head Neck 30(1):103–110
Freiser ME, Ojo RB, Lo K, Saint-Victor S, Bollig C, Nayak CS, Sargi ZB (2016) Complications and oncologic outcomes following elective neck dissection with salvage laryngectomy for the N0 neck. Am J Otolaryngol 37(3):186–194
Wang M, Xun Y, Wang K, Lu L, Yu A, Guan B, Yu C (2020) Risk factors of pharyngocutaneous fistula after total laryngectomy: a systematic review and meta-analysis. Eur Arch Otorhinolaryngol 277(2):585–599
Gane EM, Michaleff ZA, Cottrell MA, McPhail SM, Hatton AL, Panizza BJ, O’Leary SP (2017) Prevalence, incidence, and risk factors for shoulder and neck dysfunction after neck dissection: a systematic review. Eur J Surg Oncol 43(7):1199–1218
van den Bovenkamp K, Noordhuis MG, Oosting SF, van der Laan B, Roodenburg JL, Bijl HP, Halmos GB, Plaat BEC (2017) Clinical outcome of salvage neck dissections in head and neck cancer in relation to initial treatment, extent of surgery and patient factors. Clin Otolaryngol 42(3):693–700
Deng J, Ridner SH, Dietrich MS, Wells N, Wallston KA, Sinard RJ, Cmelak AJ, Murphy BA (2012) Prevalence of secondary lymphedema in patients with head and neck cancer. J Pain Symptom Manage. 43(2):244–252
Gambardella C, Clarizia G, Patrone R, Offi C, Mauriello C, Romano R, Filardo M, Conzo A, Sanguinetti A, Polistena A (2019) Advanced hemostasis in axillary lymph node dissection for locally advanced breast cancer: new technology devices compared in the prevention of seroma formation. BMC Surg 18(1):125
Suzuki K, Shimizu M, Sakagami T, Yagi M, Fujisawa T, Sawada S, Kanda A, Kobayashi Y, Iwai H (2018) A comparison of short-term outcomes of neck dissection for head and neck cancers using ThunderbeatTM, LigaSureTM or treatment without an energy-based device: a case controlled study. Int J Surg 58:60–64
Van Slycke S, Gillardin J-P, Van Den Heede K, Minguet J, Vermeersch H, Brusselaers N (2016) Comparison of the harmonic focus and the thunderbeat for open thyroidectomy. Langenbecks Arch Surg 401(6):851–859
Melck AL, Wiseman SM (2010) Harmonic scalpel compared to conventional hemostasis in thyroid surgery: a meta-analysis of randomized clinical trials. Int J Surg Oncol 2010:396079
Papavramidis TS, Sapalidis K, Michalopoulos N, Triantafillopoulou K, Gkoutzamanis G, Kesisoglou I, Papavramidis ST (2010) UltraCision harmonic scalpel versus clamp-and-tie total thyroidectomy: a clinical trial. Head Neck 32(6):723–727
Chang JW, Na G, Shin SH, Kim YS, Koo BS, Ahn SH, Chang Choi E, Kim WS (2020) Is an ultrasonic and bipolar integrated energy device more useful than a conventional electric device in head and neck free flap reconstruction? A prospective comparison. J Oral Maxillofac Surg 78(8):1437.e1431-1437.e1439
Piccirillo JF, Feinstein AR (1996) Clinical symptoms and comorbidity: significance for the prognostic classification of cancer. Cancer 77(5):834–842
Dindo D, Demartines N, Clavien P-A (2004) Classification of surgical complications: a new proposal with evaluation in a cohort of 6336 patients and results of a survey. Ann Surg 240(2):205
Brierley JD, Gospodarowicz MK, Wittekind C (2017) TNM classification of malignant tumours. Wiley
IBM Corp. (2019) IBM SPSS Statistics for Mac OS vol 64-bit edition. Armonk, NY
Fagotti A, Vizzielli G, Fanfani F, Gallotta V, Rossitto C, Costantini B, Gueli-Alletti S, Avenia N, Iodice R, Scambia G (2014) Randomized study comparing use of THUNDERBEAT technology vs standard electrosurgery during laparoscopic radical hysterectomy and pelvic lymphadenectomy for gynecologic cancer. J Minim Invasive Gynecol 21(3):447–453
Allaix ME, Arezzo A, Giraudo G, Arolfo S, Mistrangelo M, Morino M (2017) The thunderbeat and other energy devices in laparoscopic colorectal resections: analysis of outcomes and costs. J Laparoendosc Adv Surg Tech A 27(12):1225–1229
Patrone R, Gambardella C, Romano RM, Gugliemo C, Offi C, Andretta C, Vitiello A, Tartaglia E, Flagiello L, Conzo A, Mauriello C, Conzo G (2019) The impact of the ultrasonic, bipolar and integrated energy devices in the adrenal gland surgery: literature review and our experience. BMC Surg 18(Suppl 1):123
Suhardja TS, Norhadi S, Ee E, Hodgkins B (2018) Comparison of the thunderbeat and other energy devices in laparoscopic colorectal resection: a aingle-center experience. J Laparoendosc Adv Surg Tech A 28(12):1417–1421
Ren ZH, Xu JL, Fan TF, Ji T, Wu HJ, Zhang CP (2015) The harmonic scalpel versus conventional hemostasis for neck dissection: a meta-analysis of the randomized controlled trials. PLoS One 10(7):e0132476
Funding
Olympus Europe (OLYMPUS EUROPE SE and CO. KG, Amsinckstrasse 63, 20097 Hamburg, Germany) provided the THUNDERBEAT Open Fine Jaw disposable devices and the generator for free of charge during the study. No other financial support was provided.
Author information
Authors and Affiliations
Contributions
All authors made substantial contributions to the conception or design of the work; or the acquisition, analysis, or interpretation of data; or the creation of new software used in the work; drafted the work or revised it critically for important intellectual content, approved the version to be published; and agree to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved.
Corresponding author
Ethics declarations
Conflict of interest
B. E. C. Plaat has a consultancy role for and has received research funding by Olympus Europa SE and CO. KG for a different research project. Other authors have no conflict of interest to declare.
Ethical approval
The study was checked by the Institutional Review Board of the University Medical Center Groningen (UMCG) and a waiver was released. The study was performed in accordance with Good Clinical Practice principles and the Consolidated Standards of Reporting Trials (CONSORT) statement.
Consent to participate
Not applicable.
Consent for publication
All authors are aware of the content of the manuscript and agree publishing it.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Kuipers, N.C., de Kleijn, B.J., Wedman, J. et al. A comparison of the Thunderbeat and standard electrocautery devices in head and neck surgery: a prospective randomized controlled trial. Eur Arch Otorhinolaryngol 278, 4987–4996 (2021). https://doi.org/10.1007/s00405-021-06739-z
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00405-021-06739-z