Abstract
Background
This review summarizes the available data on the effectiveness of indocyanine green fluorescence imaging (ICG-FI) for real-time detection of breast cancer (BC) tumors with perioperative imaging technologies.
Methods
PubMed and Scopus databases were exhaustively searched for publications on the use of the real-time ICG-FI evaluation of BC tumors with non-conventional breast imaging technologies.
Results
Twenty-three studies were included in this review. ICG-FI has been used for BC tumor identification in 12 orthotopic animal tumor experiences, 4 studies on animal assessment, and for 7 human clinical applications. The BC tumor-to-background ratio (TBR) was 1.1–8.5 in orthotopic tumor models and 1.4–3.9 in animal experiences.
The detection of primary human BC tumors varied from 40% to 100%. The mean TBR reported for human BC varied from 2.1 to 3.7. In two studies evaluating BC surgical margins, good sensitivity (93.3% and 100%) and specificity (60% and 96%) have been reported, with a negative predictive value of ICG-FI to predict margin involvement intraoperatively of 100% in one study.
Conclusions
The use of ICG-FI as a guiding tool for the real-time identification of BC tumors and for the assessment of tumor boundaries is promising. There is great variability between the studies with regard to timing and dose. Further evidence is needed to assess whether ICG-guided BC surgery may be implemented as a standard of care.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Introduction
Breast cancer (BC) is the leading cause of cancer in women, affecting approximately 2.3 million women annually and accounting for the highest number of cancer-related deaths [1]. The primary treatment modalities for BC include surgery, hormonotherapy, and/or chemotherapy, depending on tumor characteristics [2]. Successful BC surgery involves achieving complete resection of palpable tumors and the resection of small non-palpable (infra-clinical) tumors with microscopically negative margins, while preserving as much normal breast tissue as possible for optimal aesthetic breast reconstruction [3]. Despite significant advancements in preoperative and intraoperative imaging techniques for improved tumor detection, the rate of positive microscopic margins after conservative BC surgery remains high, ranging from 14.9% to 26% in the literature [3,4,5].
Precise localization of the breast tumor lesion and the ability to distinguish between malignant and benign tissue during surgery are critical for the successful surgical treatment of BC patients. Various pathological and imaging techniques have been reported for this purpose [5].
Recently, near-infrared (NIR) fluorescence imaging (FI) has emerged as a promising nonionizing imaging technique for detection of cancerous tissue in different clinical conditions, including BC [4, 6, 7]. NIR-FI utilizes light properties in the NIR spectrum (700–900 nm) to image tissue. NIR fluorescence offers advantages such as high tissue penetration (millimeters to centimeters in depth) and low autofluorescence emitted by natural fluorophores in the human body (e.g., porphyrins), enabling good discrimination between tissues containing fluorophores and those that do not, resulting in a high signal-to-noise ratio (SNR) [8, 9].
Several fluorophores that emit in the NIR spectrum have been investigated for BC detection in preclinical and clinical studies, including specific and nonspecific fluorescent probes [4, 8, 10,11,12,13,14,15,16,17,18,19,20,21,22,23,24,25,26]. Specific fluorophores include pegulicianine [16], bevacizumab–IRDye800CW [17], indocyanine green (ICG) in combination with different particles to enhance tumor avidity and specificity, such as human serum albumin [18], Pseudomonas aeruginosa azurin peptide p28 [19], low molecular weight heparin (LMWH) ICG-loaded liposomes (LMWH-ICG-Lip) [20], ICG-loaded H-Ferritin (HFn) nanoparticles [21], and functionalized erbium-based rare-earth nanoparticles [22]. However, specific fluorescent probes are currently not approved for clinical use and will not be discussed in this review.
Non-specific fluorophores include methylene blue (MB) [24, 25], 5-aminolevulinic acid (5-ALA) [26], and ICG. Among these, ICG is the most popular and widely used fluorophore. ICG is a water-soluble amphiphilic tricarbocyanine with a molecular weight of 775 Da and a hydrodynamic diameter of 1.2 nm, making it an excellent vascular and lymphatic contrast agent when injected intravenously (IV) or into the lymphatic system via subcutaneous injection. Initially developed for photography during World War II, ICG was later utilized for determining cardiac output, hepatic function, and ophthalmic perfusion. Its rapid Food and Drug Administration (FDA) registration was attributed to favorable characteristics, including confinement to the vascular compartment through binding to plasma proteins, fast and almost exclusive excretion into the bile, and low toxicity. In the 1970s, it was discovered that protein-bound ICG emitted fluorescence under illumination with NIR light (750–810 nm), peaking at around 840 nm [27, 28]. ICG is approved by the FDA and the European Medicines Agency (EMA) for clinical applications as a vascular contrast agent. Due to its safety, affordability, and availability, ICG has become the foundation of NIR-FI for tumor detection [29]. Intraoperative indocyanine green fluorescence imaging (ICG-FI) navigation has emerged as a promising technique for detecting cancerous tissue, including liver, colon, ovarian, head and neck, lung, and breast tumors, enabling surgeons to customize surgery based on real-time intraoperative imaging findings.
Since the first successful detection of BC tumors using ICG-FI reported by Ntziachristos and colleagues in the early 2000s, numerous preclinical and a few clinical pilot studies have demonstrated the detectability of BC tumors using ICG-FI after ICG IV injection [4, 11, 13,14,15, 29]. The exact physiological mechanism underlying the preferential uptake of ICG in tumor tissues after intravenous injection is not fully understood. The most plausible hypothesis is the ‘enhanced permeability and retention’ (EPR) effect observed in tumoral tissue due to neoangiogenesis [4, 6, 8, 10, 30]. Following intravenous injection, ICG acts as a macromolecule due to its high binding to plasma proteins. In healthy tissues, macromolecule-bound ICG serves as an excellent contrast agent, remaining in the intravascular compartment. In contrast, according to the EPR effect, these macromolecules are thought to extravasate from abnormal tumor vessels into the malignant tumor’s extracellular space. As the half-life of ICG in blood circulation is 3–5 min, ICG rapidly washes out from the intravascular space. Consequently, under NIR illumination, ICG that has accumulated in tumoral tissue emits a fluorescence signal that can be visualized through 5–10 mm of connective tissue thickness, resulting in the observed hyperfluorescence of tumoral tissue in contrast to surrounding normal tissue [30,31,32,33,34].
The objectives of this study were to conduct a systematic literature review on ICG-FI for real-time detection of BC tumors in preclinical and clinical studies of perioperative imaging technologies and provide a summary of evidence-based data on the effectiveness of ICG-FI in BC.
Methods
This systematic literature review was conducted following the recommendations established by the Preferred Reporting in Systematic Review and Meta-Analysis (PRISMA).
Inclusion and Exclusion Criteria: The focus of our search was on studies that reported real-time perioperative (ex-vivo and in-vivo) ICG-FI in primary breast malignant tumors. This included the following aspects: (1) identification of primary BC tumors using ICG-FI, (2) evaluation of tumor margins after BC surgery using ICG-FI, (3) assessment of fluorescence intensity in BC tumors, and (4) accuracy of ICG-FI in detecting primary BC. In cases where two papers reported on the same population, only the first published study was included. We included either French or English language papers.
The following topics were excluded from the review: Conventional breast imaging for detection of BC, angiographic characterizations of ICG-FI, such as mastectomy flap and breast reconstructive flap vascularization and visualization, sentinel lymph node (SLN) detection, and infraclinical BC tumor marking. Additionally, editorials, reviews, commentaries, letters, and book chapters were excluded.
Sources and Literature Search: A comprehensive search was conducted in the PubMed and Scopus databases with the assistance of a professional medical librarian. The search encompassed articles published before December 2022. Furthermore, the reference lists of the retained articles were analyzed for additional relevant studies that met the inclusion criteria.
The following MeSH terms were used: “Optical Imaging”, “Indocyanine Green”, “Breast Neoplasms”, “breast neoplasms/surgery”, “Mastectomy” and “mammary neoplasms/animal”. Free search terms included: “breast cancer”, “breast neoplasia”, “breast-conserving surgery”, “mastectomy”, “breast surgery”, “fluorescence imaging”, “ICG”, “residual tumor”, “margins” and “animal”. These terms were used in various combinations.
Screening of Titles, Abstracts, and Full Texts: The titles, abstracts, and full texts of relevant studies were screened against the inclusion and exclusion criteria.
Data Extraction and Categorization: Full-text versions of studies that met the inclusion criteria were obtained for comprehensive assessment. The following data were extracted: year of publication, authors, study design, number of subjects, histological cancer type, technical details of ICG-FI (timing, volume/dose of ICG injection, type of FI system, FI intensity analysis program, depth of detection), and, when reported, the accuracy and/or detection rate.
These data were then analyzed and categorized into two groups based on the stage of the experiment: Preclinical experiences (including orthotopic tumor models and animal studies related to BC) or human clinical applications.
Results
The search strategy yielded a total of 2607 studies. After removing duplicates, 23 studies published between 1995 and 2022 fulfilled the inclusion criteria and were included in this review. Of these, 16 studies were animal studies, including 12 studies reporting on orthotopic tumor model experiences and 4 studies reporting on preclinical animal assessments. Additionally, 7 studies reported on human clinical applications of ICG-FI. A PRISMA flow diagram illustrating the study selection process is presented in Fig. 1.
Animal experience in breast cancers with ICG-FI
Preclinical orthotopic tumor models for breast cancer tumor detection by ICG-FI
Twelve studies were included in the analysis of ICG-FI detection of orthotopic BC models [10,11,12, 18,19,20,21, 35,36,37,38,39]. Table 1 provides a summary of the characteristics of these studies. The majority of the studies utilized mouse models with subject numbers ranging from 5 to 60. The 4T1-Luc BC cells were predominantly used [10, 12, 18, 20, 21, 35,36,37,38]. Seven studies reported both ex-vivo and in-vivo FI, while one study reported only ex-vivo imaging. Various FI systems were employed including, for example, Li et al. and Goto et al. who used the PDE system (Hamamatsu) [11, 19], and Sitia et al. and Sevieri et al. who used the STORZ endoscopic system (Karl Storz Se & Co) [21, 38]. The injected dose of ICG ranged from 0.015 to 7.5 mg/kg. The timing for imaging after ICG IV injection varied across studies, with most studies performing imaging at 24 h post-injection. The detection rate of orthotopic BC tumors ranged from 78% to 100%, and the fluorescence intensity reported as tumor-to-background ratio (TBR) varied from 1.1 to 8.5. Notably, even with short delays in ICG injection (< 8 min and 2 h), the fluorescence intensity remained higher in tumors (TBR at 2.5 and 100 arbitrary units (AU)) [11, 18].
ICG-FI and preclinical animal experiences
Four studies were included in this category [13, 14, 40, 41], all conducted on dogs. Table 2 summarizes the principal characteristics of these animal experiences with self-developed BC tumors. The number of subjects varied from 1 to 16, with the evaluation of 1 to 20 malignant breast tumors per dog. All studies employed in vivo evaluation of BC using ICG-FI, with different FI systems utilized. ICG doses ranged from 1 to 5 mg/kg, and the reported TBR of BC varied between 1.3 and 3.9 AU. Two studies utilized the ImageJ software (NIH, Bethesda) for quantification of fluorescence intensity and TBR calculation [14, 41]. Similar to orthotopic model studies, the timing for performing ICG-FI after ICG IV injection varied across studies, ranging from 10–23 min to 48–72 h. Ex-vivo FI identified BC tumors in 24 out of 28 cases, while in-vivo imaging identified tumors in 21 out of 28 cases. The sensitivity of ICG-FI in detecting BC tumors in dogs ranged from 80% to 100%. All three tumors with a short ICG IV injection delay (minutes) were visible by FI [13, 40]. The mean TBR reported for BC tumors in dogs varied from 1.4 to 3.9, with higher values observed in cases with a short ICG injection time (<30 min) [13, 40].
ICG-FI clinical experience in human breast cancers
Breast cancer detection
Clinical experiences with ICG-FI in the detection of BC started in 2016 with a study by Keating et al. [10]. Since then, six other studies have been published on human clinical applications of ICG-FI in BC [4, 15, 16, 31,32,33]. Table 3 summarizes the characteristics of these studies. All studies were pilot studies that aimed to evaluate the feasibility of BC detection using ICG-FI (phase 0–2 studies) and included a limited number of patients, ranging from 8 to 43 patients per study [16, 31]. Patients with both histological adenocarcinoma types, ductal and lobular invasive BC, were included. Histological characteristics of BC were reported in two out of seven studies [4, 31]. Six studies explored primary BC during upfront surgery, while one study investigated neoadjuvant chemotherapy (NAC) [31].
All seven studies included evaluations of ICG-FI using ex-vivo FI, and four studies reported concomitant in-vivo imaging. Different imaging systems were used in each study, except for Veys et al. and Pop et al., who both used the Fluobeam 800 system (Fluoptic, Grenoble, France), and Leiloglou et al. and Kedrzycki et al., who used their own developed system [4, 31,32,33]. For instance, Keating et al. tested three different FI systems in their 12 patients. The injected dose of ICG for BC detection varied from 0.25 mg/kg (in four studies) to 5 mg/kg (in one study). The timing for imaging after IV injection varied, with some studies starting intraoperatively, as soon as 5 min after ICG injection [32, 33], while others ranged from 20 to 135 min [4, 16, 31, 33], and 24 h in two studies [10, 15]. Different programs, including IC-Calc 2.0, Matlab software (Mathworks, Inc., Massachusetts, USA), and ImageJ software (NIH, Bethesda), were used for quantifying fluorescence intensity and calculating TBR.
Ex-vivo imaging was used in all studies to visualize and identify BC. In the preoperative injection setting, the rate of BC detection varied from 40% in patients injected with a low dose (0.25 mg/kg) of ICG 24 h before imaging to 100% in those injected with a high dose (5 mg/kg) the day before FI [10, 15]. In the intraoperative injection setting, the detection rate of BC varied from 72% to 100% [4, 16, 31,32,33]. Notably, in one study that used intraoperative ICG injection with an imaging interval shorter than 5 min, the detection rate of BC was 85% [33]. The mean TBR reported for human BC varied from 2.1 to 3.7. In the preoperative injection setting, TBR values of 2.0 and 3.5 were reported in two out of five patients injected with a low dose of ICG the day before surgery, while a mean TBR of 3.7 was reported in those injected with a high dose [10, 15]. In the group of patients injected intraoperatively, the mean TBR reported was homogeneous, ranging from 2.1 to 3.3 [4, 16, 31,32,33].
Margin evaluation
Only three clinical studies explored the use of ICG-FI for evaluating surgical margins in BC [4, 10, 16]. Two different injection time strategies were used: one using ICG injected 24 h before surgery and the other using ICG injected intraoperatively or shortly before surgery (2 h). In a pilot study of 12 patients injected 24 h before surgery with a dose of 5 mg/kg, Keating et al. reported residual fluorescence in the tumor bed in 6 out of 12 patients, but none of these patients had positive margins on definitive pathology [10]. Pop et al., in a pilot study of 35 patients injected intraoperatively with a dose of 0.25 mg/kg, reported a sensitivity (Se), specificity (Sp), and negative predictive value (NPV) of ICG-FI to predict margin involvement on breast operative specimens of 100%, 60%, and 100%, respectively [4]. Recently, Wang and colleagues reported their data on 43 BC patients who were injected with ICG at 0.5 mg/kg, 2 h before surgery, and found an intraoperative sensitivity and specificity for ICG-FI in distinguishing between normal tissue (clean margins) and tumoral tissue (positive margins) of 93.3% and 96.0%, respectively [16].
Discussion
FI has the potential to be a highly beneficial technique for real-time tumor identification and assessment of tumor boundaries during surgical procedures, particularly in BC. However, its clinical applications for tumor resection are currently limited, with only a few studies conducted in BC [4, 10, 26, 33, 34, 42, 43].
In this systematic review of the literature, 23 studies were included that evaluated the efficacy of ICG-FI for discriminating between benign breast tissue and neoplastic BC. Among these studies, only 7 utilized ICG-FI in clinical settings, and all of them were in the proof-of-concept or feasibility phase [4, 15, 16, 31,32,33]. The results from these studies show promise for the use of ICG-FI in BC surgery. The detection rate of orthotopic BC tumors and BC tumors in dogs using ICG-FI ranged from 78% to 100% and 80% to 100%, respectively [13, 14, 40, 41]. In clinical experiences, ICG-FI was able to detect tumoral disease in approximately 8 out of 10 women (with a sensitivity between 80% and 100%) when ICG was injected shortly before surgery (within 2 h) [4, 15, 16, 31,32,33]. The mean TBR reported for BC tumor identification in these studies varied from 1.1 to 8.5, and in human clinical studies, it ranged from 2.1 to 3.7. These TBR values are higher than the threshold detection value (1.3–1.5) by the human eye to define the tissue as hyperfluorescent regardless of cancer type, evaluation type (in vivo or ex vivo), and the FI camera system used [31, 44].
One major challenge in BC surgery is the intraoperative assessment of breast surgical specimens during breast-conserving surgery to rapidly detect residual tumoral disease [3,4,5]. Early evidence suggests that ICG-FI can be used for the intraoperative evaluation of surgical margin resection after breast conserving surgery, potentially improving surgical treatment outcomes for BC patients [4, 10, 31,32,33,34]. With a reported high negative predictive value (100%), ICG-FI examination of the surgical bed may be able to exclude a positive resection margin with certainty, focusing intraoperative pathological evaluation only on cases where residual fluorescence is observed [4, 31].
However, it’s important to interpret these results with caution due to the variability between studies, especially in terms of research stage (preclinical and clinical phase I settings), ICG dose, timing of ICG-FI, FI camera systems, and fluorescence intensity quantifications. More precision is needed regarding the pathophysiological mechanism of action, dosing, and timing of ICG-FI [42,43,44,45].
Mechanism of action of ICG in breast tumors
The mechanism of preferential uptake of ICG in tumor tissues is not fully understood, but likely involves the EPR effect observed in tumor tissue due to abnormal neoangiogenesis. ICG molecules injected intravenously bind to serum lipoproteins and accumulate in the extravascular space of tumor tissue, emitting fluorescence under NIR illumination. The fluorescence signals can be visualized through connective tissue up to 5–10 mm thick. The rapid clearance of ICG from the intravascular space results in the observed hyperfluorescence of tumoral tissue compared to surrounding normal tissue [7,8,9, 28, 29, 45, 46].
Another lesser-known mechanism of action of ICG in cancer cells, including BC, is its vascular contrast agent properties. During or shortly (3–10 min) after intravenous administration of an ICG bolus, the fluorescence of tumor cells is enhanced due to the binding of ICG to plasma lipoproteins. This mechanism improves the contrast between tumor and normal breast tissue, surpassing pure absorption contrast [7, 28, 29]. Recent studies have shown higher vessel density and increased branch points of the vasculature in breast tumors, which may contribute to the angiographic effect of ICG in breast tumors [47].
A recent study conducted by the group at Imperial College London provides further insight into the mechanism of action of ICG-FI for intraoperative detection of BC tumors [32, 33]. Their findings suggest that the diagnostic accuracy of ICG-FI is improved when the imaging is performed during the angiography phase (within 5 min) compared to longer intervals (over 25 min) after ICG administration. The ex-vivo TBR in the angiography cohort was 3.18 (SD 1.74) compared to 2.10 (SD 0.92) in the later ICG-FI cohort, indicating better tumor detection in the angiography phase [33]. However, it is important to note that larger and adequately powered clinical trials are necessary to confirm these findings.
In addition to the previously discussed pathophysiologic mechanisms of tumor hyperfluorescence, a new pathway for ICG accumulation within tumor cells has been described [48,49,50,51]. This mechanism involves passive tumor cell targeting of ICG through increased uptake via clathrin-mediated endocytosis (CME), facilitated by the high endocytic activity of tumor cells and disruption of tight junctions. This phenomenon was initially observed in a mouse model of colorectal cancer [48, 49] and subsequently confirmed in studies involving sarcoma and BC cell lines [50, 51]. It appears that the affinity of ICG for phospholipid components of the cell membrane, which are altered and enriched in tumor cells, contributes to its ability to bind to and pass intracellularly via CME [52]. Furthermore, tumor cells retain the dye for an extended period (at least 24 h) compared to normal tissues, indicating increased cellular uptake and retention as the primary mechanism of tumor fluorescence, rather than solely relying on the EPR effect. These contrasting findings highlight the complexity of the intra-tumor accumulation of ICG, suggesting that multiple mechanisms, including dysregulation of cancer cell pathways, tumor microvasculature, and the EPR effect, likely contribute to its enhanced uptake in tumors. The specific factors at play may vary depending on the tumor type. Further research is needed to better understand the mechanisms of action of ICG at the cellular level within human tumor tissue, which can provide valuable insights for the clinical use of ICG in fluorescence-guided surgery and potentially other diagnostic and treatment applications. This deeper understanding can help optimize aspects such as dosage and timing of ICG administration, which are still not fully elucidated.
ICG dose and timing
The optimal timing and dose of ICG administration for visualization and delineation of BC tumors through ICG-FI are crucial for accurate diagnosis during surgery. However, reports in the literature vary considerably in terms of protocols and inconsistent findings regarding the effectiveness of different timing and dosing strategies.
Studies have indicated that a low dose of ICG (less than 0.5 mg/kg) administered 24 h before ICG-FI is not effective for BC tumor detection [15, 18, 19]. Even a dose of 1 mg/kg administered the day before surgery does not appear to be sufficient for satisfactory tumor visualization by FI [12, 39, 40]. These findings suggest that preoperative injection, 24 h before surgery, is not the optimal timing for ICG administration.
It is worth noting that most clinical studies, except one, utilized intraoperative ICG injection timing (within ≤120 min) [10]. In contrast, only 4 out of 16 preclinical studies used intraoperative timing. Additionally, studies have explored various injection times ranging from a few minutes to over 24 h, with inconsistent results. However, the interpretation of these findings is challenging due to the variations in ICG dose used across different studies.
Clinical applications of ICG-FI for intraoperative BC detection or discrimination between benign and malignant tissue have mostly been conducted with short delays between ICG injection and FI (ranging from <5 to 143 min) and lower ICG doses (0.25–0.5 mg/kg). These studies reported relatively higher efficacy (sensitivity) ranging from 72% to 100% for BC tumor detection [4, 16, 31,32,33]. The intraoperative injection timing and lower ICG dose used in these studies make this ICG-FI strategy more easily integrated into current clinical workflows with minimal inconvenience for patients.
It is important to note that, although the literature on the application of ICG-FI in human BC is limited, the existing studies demonstrate significant heterogeneity and variation in reporting the efficacy of ICG-FI. Some studies focus solely on TBR, while others report sensitivity, specificity, negative predictive value, and false-positive rate. To ensure accurate evaluation and comparison of different timing, dosing, and FI systems/strategies, future studies should adhere to reporting complete test accuracy data.
In summary, there is a need for standardized protocols and comprehensive reporting of test accuracy data in future studies on ICG-FI for BC. This will facilitate better comparisons and understanding of the optimal timing and dose of ICG administration, as well as the effectiveness of different FI systems and strategies.
Imaging systems and FI quantification
We highlight an important limitation in the field of ICG-FI for BC detection, which is the wide variety of FI systems used in both preclinical and clinical settings. In clinical applications alone, seven different FI systems were utilized across seven different studies, with one study even testing three different FI systems in just 12 patients [10]. This variability in FI systems makes it challenging to compare and interpret the results of these early experiences of ICG-FI in BC.
Furthermore, the handheld camera models used for intraoperative imaging may not be well-suited, especially for evaluating the breast surgical cavity after breast-conserving surgery. Although optical imaging systems may have similar characteristics, there is a lack of direct comparison between these systems. Standardization of functionality and results, along with a checklist of performance criteria, should be required and provided by manufacturers to enable meaningful comparisons between different FI cameras. Perhaps certain FI systems need to be adapted for specific uses such as angiographic assessment, sentinel node detection, or evaluation of different types of tumoral tissue. However, currently, there is no recommended device specifically tailored for ICG-FI in BC tumors, and a comparative evaluation of the existing FI systems is necessary to determine the optimal imaging approach [53].
Another challenge in ICG-FI is the quantification of fluorescence intensity, which further complicates result interpretation and comparison. This issue is not limited to BC tumor detection but is also relevant for assessing tissue viability through vascular assessment [54,55,56]. In the reviewed studies, fluorescence signal quantification and TBR calculations were performed using three different programs across six out of the seven clinical studies that included quantification [4, 10, 15, 16, 31, 33]. Multiple programs with varying algorithms are being evaluated, but efforts should be made by manufacturing companies to develop quantitative imaging systems that are user-friendly and can facilitate the clinical implementation of intraoperative FI in various indications [54].
Perspectives and limitations
Despite the heterogeneity in ICG dose, timing, and fluorescence systems used in preclinical and clinical evaluations of BC tumors with ICG-FI, the results of the few clinical studies available appear promising [4, 16, 31, 33]. However, future prospective controlled studies are still needed to better define the optimal timing and ICG doses for ICG-FI and to strengthen the current evidence supporting its use in guiding BC surgery in clinical practice.
ICG-FI for BC tumor-guided surgery offers several advantages, including the relatively low cost of the fluorescent dye and its safety for patients and the medical team as a non-invasive agent [57].
It is important to address some limitations of the present review. First, the level of evidence for the results obtained and presented in this review is low due to the current literature, which primarily consists of case series with a small number of patients and considerable heterogeneity. Second, the review could not provide clear-cut results regarding the optimal dose and timing of ICG injection for ICG-FI in BC tumor evaluation. One reason for this is the lack of comparison or control groups in clinical studies. The implementation of ICG-FI in BC tumor detection does not seem to follow a reliable translational approach, as most preclinical studies use a high dose of ICG and a preoperative timing of 24 h, while most clinical trials employ low doses and short intervals until FI. This difference may be attributed to the authors’ tendency to use shorter intervals between ICG injection and surgery to better align with clinical settings. Additionally, the discrepant results between preclinical and clinical studies may be explained by differences in the accumulation, distribution, and persistence of fluorescence signals after ICG IV injections between animal orthotopic tumors and true human tumors.
Conclusions and future directions
This systematic review of the literature is the first summarizing the results of ICG-FI in BC surgical procedures. Our findings demonstrate promising evidence that detection of BC tumoral tissue and tumor-margin delineation can be improved in clinical practice with the use of ICG-FI as an adjunctive real-time tool. ICG-FI for BC tumor-guided surgery can offer several complementary advantages, such as the relatively low cost of the fluorescent dye and its safety for the patient and the medical team. The variety of FI systems used in perioperative ICG-FI for BC detection, along with the lack of standardized functionality, result reporting, and quantification methods, poses challenges for comparing and interpreting results. Standardization efforts and comparative evaluations are needed to identify the most suitable FI system and establish consistent quantification approaches. While the current evidence on ICG-FI for BC tumor detection shows promise, further well-designed prospective controlled studies are needed to determine the optimal dose and timing of ICG injection. This will provide stronger evidence to support the clinical use of ICG-FI in BC surgery.
Data availability
Not applicable
References
Sung H, Ferlay J, Siegel RL, Laversanne M, Soerjomataram I, Jemal A, Bray F (2021) Global Cancer Statistics 2020: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries. CA Cancer J Clin 71:209–249. https://doi.org/10.3322/caac.21660
Burstein HJ, Curigliano G, Thürlimann B, Weber WP, Poortmans P, Regan MM, Senn HJ, Winer EP, Gnant M, Panelists of the St Gallen Consensus Conference (2021) Customizing local and systemic therapies for women with early breast cancer: the St. Gallen International Consensus Guidelines for treatment of early breast cancer 2021. Ann Oncol 32:1216–1235. https://doi.org/10.1016/j.annonc.2021.06.023
Bundred JR, Michael S, Stuart B, Cutress RI, Beckmann K, Holleczek B, Dahlstrom JE, Gath J, Dodwell D, Bundred NJ (2022) Margin status and survival outcomes after breast cancer conservation surgery: prospectively registered systematic review and meta-analysis. BMJ 378:e070346. https://doi.org/10.1136/bmj-2022-070346
Pop FC, Veys I, Vankerckhove S, Barbieux R, Chintinne M, Moreau M, Donckier V, Larsimont D, Bourgeois P, Liberale G (2021) Absence of residual fluorescence in the surgical bed at near-infrared fluorescence imaging predicts negative margins at final pathology in patients treated with breast-conserving surgery for breast cancer. Eur J Surg Oncol 47:269–275. https://doi.org/10.1016/j.ejso.2020.09.036
St John ER, Al-Khudairi R, Ashrafian H, Athanasiou T, Takats Z, Hadjiminas DJ, Darzi A, Leff DR (2017) Diagnostic accuracy of intraoperative techniques for margin assessment in breast cancer surgery: a meta-analysis. Ann Surg 265:300–310. https://doi.org/10.1097/SLA.0000000000001897
Liberale G, Vankerckhove S, Caldon MG, Ahmed B, Moreau M, Nakadi IE, Larsimont D, Donckier V, Bourgeois P, Group R&D for the Clinical Application of Fluorescence Imaging of the Jules Bordet Institute (2016) Fluorescence imaging after indocyanine green injection for detection of peritoneal metastases in patients undergoing cytoreductive surgery for peritoneal carcinomatosis from colorectal cancer: a pilot study. Ann Surg 264:1110–1115. https://doi.org/10.1097/SLA.0000000000001618
Corlu A, Choe R, Durduran T, Rosen MA, Schweiger M, Arridge SR, Schnall MD, Yodh AG (2007) Three-dimensional in vivo fluorescence diffuse optical tomography of breast cancer in humans. Opt Express 15:6696–6716. https://doi.org/10.1364/oe.15.006696
Alacam B, Yazici B, Intes X, Nioka S, Chance B (2008) Pharmacokinetic-rate images of indocyanine green for breast tumors using near-infrared optical methods. Phys Med Biol 53:837–859. https://doi.org/10.1088/0031-9155/53/4/002
Desmettre T, Devoisselle JM, Mordon S (2000) Fluorescence properties and metabolic features of indocyanine green (ICG) as related to angiography. Surv Ophthalmol 45:15–27. https://doi.org/10.1016/s0039-6257(00)00123-5
Keating J, Tchou J, Okusanya O, Fisher C, Batiste R, Jiang J, Kennedy G, Nie S, Singhal S (2016) Identification of breast cancer margins using intraoperative near-infrared imaging. J Surg Oncol 113:508–514. https://doi.org/10.1002/jso.24167
Li X, Beauvoit B, White R, Nioka S, Chance B, Yodh AG (1995) Tumor localization using fluorescence of indocyanine green (ICG) in rat models. SPIE Digital Library. Proc SPIE 2389:789–797. https://doi.org/10.1117/12.210021
Wang Z, Ni K, Zhang X, Ai S, Guan W, Cai H, Wang Y, Lu Q, Lane LA (2018) Method for real-time tissue quantification of indocyanine green revealing optimal conditions for near infrared fluorescence guided surgery. Anal Chem 90:7922–7929. https://doi.org/10.1021/acs.analchem.8b00480
Reynolds JS, Troy TL, Mayer RH, Thompson AB, Waters DJ, Cornell KK, Snyder PW, Sevick-Muraca EM (1999) Imaging of spontaneous canine mammary tumors using fluorescent contrast agents. Photochem Photobiol 70:87–94
Newton A, Predina J, Mison M, Runge J, Bradley C, Stefanovski D, Singhal S, Holt D (2020) Intraoperative near-infrared imaging can identify canine mammary tumors, a spontaneously occurring, large animal model of human breast cancer. PLoS ONE 15:e0234791. https://doi.org/10.1371/journal.pone.0234791
Bourgeois P, Veys I, Noterman D, De Neubourg F, Chintinne M, Vankerckhove S, Nogaret JM (2021) Near-infrared fluorescence imaging of breast cancer and axillary lymph nodes after intravenous injection of free indocyanine green. Front Oncol 11:602906. https://doi.org/10.3389/fonc.2021.602906
Hwang ES, Beitsch P, Blumencranz P, Carr D, Chagpar A, Clark L, Dekhne N, Dodge D, Dyess DL, Gold L, Grobmyer S, Hunt K, Karp S, Lesnikoski BA, Wapnir I, Smith BL, INSITE study team (2022) Clinical impact of intraoperative margin assessment in breast-conserving surgery with a novel pegulicianine fluorescence-guided system: a nonrandomized controlled trial. JAMA Surg 157:573–580. https://doi.org/10.1001/jamasurg.2022.1075
Lamberts LE, Koch M, de Jong JS, Adams ALL, Glatz J, Kranendonk MEG, Terwisscha van Scheltinga AGT, Jansen L, de Vries J, Lub-de Hooge MN, Schröder CP, Jorritsma-Smit A, Linssen MD, de Boer E, van der Vegt B, Nagengast WB, Elias SG, Oliveira S, Witkamp AJ, Mali WPTM, Van der Wall E, van Diest PJ, de Vries EGE, Ntziachristos V, van Dam GM (2017) Tumor-specific uptake of fluorescent bevacizumab-IRDye800CW microdosing in patients with primary breast cancer: a phase I feasibility study. Clin Cancer Res 23:2730–2741. https://doi.org/10.1158/1078-0432.CCR-16-0437
Lin X, Liu C, Sheng Z, Gong X, Song L, Zhang R, Zheng H, Sun M (2018) Highly sensitive fluorescence and photoacoustic detection of metastatic breast cancer in mice using dual-modal nanoprobes. ACS Appl Mater Interfaces 10:26064–26074. https://doi.org/10.1021/acsami.8b09142
Goto M, Ryoo I, Naffouje S, Mander S, Christov K, Wang J, Green A, Shilkaitis A, Das Gupta TK, Yamada T (2022) Image-guided surgery with a new tumour-targeting probe improves the identification of positive margins. EBioMedicine 76:103850. https://doi.org/10.1016/j.ebiom.2022.103850
Cao D, Li H, Luo Y, Feng N, Ci T (2021) Heparin modified photosensitizer-loaded liposomes for tumor treatment and alleviating metastasis in phototherapy. Int J Biol Macromol 168:526–536. https://doi.org/10.1016/j.ijbiomac.2020.12.043
Sevieri M, Sitia L, Bonizzi A, Truffi M, Mazzucchelli S, Corsi F (2021) Tumor accumulation and off-target biodistribution of an indocyanine-green fluorescent nanotracer: an ex vivo study on an orthotopic murine model of breast cancer. Int J Mol Sci 22:1601. https://doi.org/10.3390/ijms22041601
Lou KL, Wang PY, Yang RQ, Gao YY, Tian HN, Dang YY, Li Y, Huang WH, Chen M, Liu XL, Zhang GJ (2022) Fabrication of tumor targeting rare-earth nanocrystals for real-time NIR-IIb fluorescence imaging-guided breast cancer precise surgery. Nanomedicine 43:102555. https://doi.org/10.1016/j.nano.2022.102555
Yang RQ, Chen M, Zhang Q, Gao YY, Lou KL, Lin TT, Huang WH, Zeng YZ, Zhang YQ, Dang YY, Ren L, Zhang GJ (2022) Development and preclinical evaluation of a near-infrared fluorescence probe based on tailored hepatitis B core particles for imaging-guided surgery in breast cancer. Int J Nanomedicine 17:1343–1360. https://doi.org/10.2147/IJN.S343479
Tummers QR, Verbeek FP, Schaafsma BE, Boonstra MC, van der Vorst JR, Liefers GJ, van de Velde CJ, Frangioni JV, Vahrmeijer AL (2014) Real-time intraoperative detection of breast cancer using near-infrared fluorescence imaging and methylene blue. Eur J Surg Oncol 40:850–858. https://doi.org/10.1016/j.ejso.2014.02.225
Zhang C, Jiang D, Huang B, Wang C, Zhao L, Xie X, Zhang Z, Wang K, Tian J, Luo Y (2019) Methylene blue-based near-infrared fluorescence imaging for breast cancer visualization in resected human tissues. Technol Cancer Res Treat 18:1533033819894331. https://doi.org/10.1177/1533033819894331
Ottolino-Perry K, Shahid A, DeLuca S, Son V, Sukhram M, Meng F, Liu ZA, Rapic S, Anantha NT, Wang SC, Chamma E, Gibson C, Medeiros PJ, Majeed S, Chu A, Wignall O, Pizzolato A, Rosen CF, Teene LL, Starr-Dunham D, Kulbatski I, Panzarella T, Done SJ, Easson AM, Leong WL, DaCosta RS (2021) Intraoperative fluorescence imaging with aminolevulinic acid detects grossly occult breast cancer: a phase II randomized controlled trial. Breast Cancer Res 23:72. https://doi.org/10.1186/s13058-021-01442-7
Landsman ML, Kwant G, Mook GA, Zijlstra WG (1976) Light-absorbing properties, stability, and spectral stabilization of indocyanine green. J Appl Physiol 40(4):575–583. https://doi.org/10.1152/jappl.1976.40.4.575
Desmettre T, Cohen SY, Devoisselle JM, Gaudric A (2011) Intérêt et indications actuelles de l’angiographie au vert d’indocyanine (ICG). J Fr Ophtalmol 34:568–582
Ntziachristos V, Yodh AG, Schnall M, Chance B (2000) Concurrent MRI and diffuse optical tomography of breast after indocyanine green enhancement. Proc Natl Acad Sci U S A 97:2767–2772. https://doi.org/10.1073/pnas.040570597
Maeda H, Wu J, Sawa T, Matsumura Y, Hori K (2000) Tumor vascular permeability and the EPR effect in macromolecular therapeutics: a review. J Control Release 65:271–284. https://doi.org/10.1016/s0168-3659(99)00248-5
Veys I, Pop CF, Barbieux R, Moreau M, Noterman D, De Neubourg F, Nogaret JM, Liberale G, Larsimont D, Bourgeois P (2018) ICG fluorescence imaging as a new tool for optimization of pathological evaluation in breast cancer tumors after neoadjuvant chemotherapy. PLoS ONE 13:e0197857. https://doi.org/10.1371/journal.pone.0197857
Leiloglou M, Chalau V, Kedrzycki MS, Thiruchelvam PTR, Darzi A, Leff DR, Elson D (2021) Tissue texture extraction in indocyanine green fluorescence imaging for breast-conserving surgery. J Phys D Appl Phys 54:194005. https://doi.org/10.1088/1361-6463/abe571
Kedrzycki MS, Leiloglou M, Chalau V, Chiarini N, Thiruchelvam PTR, Hadjiminas DJ, Hogben KR, Rashid F, Ramakrishnan R, Darzi AW, Elson DS, Leff DR (2021) The impact of temporal variation in indocyanine green administration on tumor identification during fluorescence guided breast surgery. Ann Surg Oncol 28:5617–5625. https://doi.org/10.1245/s10434-021-10503-2
Wang Y, Jiao W, Yin Z, Zhao W, Zhao K, Zhou Y, Fang R, Dong B, Chen B, Wang Z (2022) Application of near-infrared fluorescence imaging in the accurate assessment of surgical margins during breast-conserving surgery. World J Surg Oncol 20:357. https://doi.org/10.1186/s12957-022-02827-4
Holt D, Okusanya O, Judy R, Venegas O, Jiang J, DeJesus E, Eruslanov E, Quatromoni J, Bhojnagarwala P, Deshpande C, Albelda S, Nie S, Singhal S (2014) Intraoperative near-infrared imaging can distinguish cancer from normal tissue but not inflammation. PLoS ONE 9:e103342. https://doi.org/10.1371/journal.pone.0103342
Kumar AT, Carp SA, Yang J, Ross A, Medarova Z, Ran C (2017) Fluorescence lifetime-based contrast enhancement of indocyanine green-labeled tumors. J Biomed Opt 22:40501. https://doi.org/10.1117/1.JBO.22.4.040501
Wojtynek NE, Olson MT, Bielecki TA, An W, Bhat AM, Band H, Lauer SR, Silva-Lopez E, Mohs AM (2020) Nanoparticle formulation of indocyanine green improves image-guided surgery in a murine model of breast cancer. Mol Imaging Biol 22:891–903. https://doi.org/10.1007/s11307-019-01462-y
Sitia L, Sevieri M, Bonizzi A, Allevi R, Morasso C, Foschi D, Corsi F, Mazzucchelli S (2020) Development of tumor-targeted indocyanine green-loaded ferritin nanoparticles for intraoperative detection of cancers. ACS Omega 5:12035–12045. https://doi.org/10.1021/acsomega.0c00244
Wang Z, Chen M, Liu JJ, Chen RH, Yu Q, Wang GM, Nie LM, Huang WH, Zhang GJ (2021) Human serum albumin decorated indocyanine green improves fluorescence-guided resection of residual lesions of breast cancer in mice. Front Oncol 11:614050. https://doi.org/10.3389/fonc.2021.614050
Gurfinkel M, Thompson AB, Ralston W, Troy TL, Moore AL, Moore TA, Gust JD, Tatman D, Reynolds JS, Muggenburg B, Nikula K, Pandey R, Mayer RH, Hawrysz DJ, Sevick-Muraca EM (2000) Pharmacokinetics of ICG and HPPH-car for the detection of normal and tumor tissue using fluorescence, near-infrared reflectance imaging: a case study. Photochem Photobiol 72:94–102. https://doi.org/10.1562/0031-8655(2000)072%3c0094:poiahc%3e2.0.co;2
Favril S, Abma E, Stock E, Devriendt N, Van Goethem B, Blasi F, Brioschi C, Polis I, De Cock H, Miragoli L, Oliva P, Valbusa G, Vanderperren K, de Rooster H (2020) Fluorescence-guided surgery using indocyanine green in dogs with superficial solid tumours. Vet Rec 187:273. https://doi.org/10.1136/vr.105554
Mieog JSD, Achterberg FB, Zlitni A, Hutteman M, Burggraaf J, Swijnenburg RJ, Gioux S, Vahrmeijer AL (2022) Fundamentals and developments in fluorescence-guided cancer surgery. Nat Rev Clin Oncol 19:9–22. https://doi.org/10.1038/s41571-021-00548-3
Tanyi JL, Randall LM, Chambers SK, Butler KA, Winer IS, Langstraat CL, Han ES, Vahrmeijer AL, Chon HS, Morgan MA, Powell MA, Tseng JH, Lopez AS, Wenham RM (2023) A phase III study of pafolacianine injection (OTL38) for intraoperative imaging of folate receptor-positive ovarian cancer (Study 006). J Clin Oncol 41:276–284. https://doi.org/10.1200/JCO.22.00291
Azargoshasb S, Boekestijn I, Roestenberg M, KleinJan GH, van der Hage JA, van der Poel HG, Rietbergen DDD, van Oosterom MN, van Leeuwen FWB (2023) Quantifying the impact of signal-to-background ratios on surgical discrimination of fluorescent lesions. Mol Imaging Biol 25:180–189. https://doi.org/10.1007/s11307-022-01736-y
Wakabayashi T, Cacciaguerra AB, Abe Y, Bona ED, Nicolini D, Mocchegiani F, Kabeshima Y, Vivarelli M, Wakabayashi G, Kitagawa Y (2022) Indocyanine green fluorescence navigation in liver surgery: a systematic review on dose and timing of administration. Ann Surg 275:1025–1034. https://doi.org/10.1097/SLA.0000000000005406
Poellinger A, Burock S, Grosenick D, Hagen A, Lüdemann L, Diekmann F, Engelken F, Macdonald R, Rinneberg H, Schlag PM (2011) Breast cancer: early- and late-fluorescence near-infrared imaging with indocyanine green—a preliminary study. Radiology 258:409–416. https://doi.org/10.1148/radiol.10100258
Gu J, Ternifi R, Sabeti S, Larson NB, Carter JM, Fazzio RT, Fatemi M, Alizad A (2022) Volumetric imaging and morphometric analysis of breast tumor angiogenesis using a new contrast-free ultrasound technique: a feasibility study. Breast Cancer Res 24:85. https://doi.org/10.1186/s13058-022-01583-3
Onda N, Kimura M, Yoshida T, Shibutani M (2016) Preferential tumor cellular uptake and retention of indocyanine green for in vivo tumor imaging. Int J Cancer 139:673–682. https://doi.org/10.1002/ijc.30102
Onda N, Mizutani-Morita R, Yamashita S, Nagahara R, Matsumoto S, Yoshida T, Shibutani M (2017) Fluorescence contrast-enhanced proliferative lesion imaging by enema administration of indocyanine green in a rat model of colon carcinogenesis. Oncotarget 8:90278–90290. https://doi.org/10.18632/oncotarget.21744
Chan CD, Brookes MJ, Tanwani R, Hope C, Pringle TA, Knight JC (2021) Investigating the mechanisms of indocyanine green (ICG) cellular uptake in sarcoma. bioRxiv. https://doi.org/10.1101/2021.04.05.438013
Nicoli F, Saleh DB, Baljer B, Chan CD, Beckingsale T, Ghosh KM, Ragbir M, Rankin KS (2021) Intraoperative near-infrared fluorescence (NIR) imaging with indocyanine green (ICG) can identify bone and soft tissue sarcomas which may provide guidance for oncological resection. Ann Surg 273:e63–e68. https://doi.org/10.1097/SLA.0000000000003857
Kaksonen M, Roux A (2018) Mechanisms of clathrin-mediated endocytosis. Nat Rev Mol Cell Biol 19:313–326. https://doi.org/10.1038/nrm.2017.132
Cho SS, Teng CW, De Ravin E, Singh YB, Lee JYK (2022) Assessment and comparison of three dimensional exoscopes for near-infrared fluorescence-guided surgery using second-window indocyanine-green. J Korean Neurosurg Soc 65:572–581. https://doi.org/10.3340/jkns.2021.0202
Hoogstins C, Burggraaf JJ, Koller M, Handgraaf H, Boogerd L, van Dam G, Vahrmeijer A, Burggraaf J (2019) Setting standards for reporting and quantification in fluorescence-guided surgery. Mol Imaging Biol 21:11–18. https://doi.org/10.1007/s11307-018-1220-0
Van Den Hoven P, Verduijn PS, Van Capelle L, Tange FP, Michi M, Corion LUM, Sibinga Mulder BG, Mureau MAM, Vahrmeijer AL, Van Der Vorst JR (2022) Quantification of near-infrared fluorescence imaging with indocyanine green in free flap breast reconstruction. J Plast Reconstr Aesthet Surg 75:1820–1825. https://doi.org/10.1016/j.bjps.2021.12.004
Hardy NP, Epperlein JP, Dalli J, Robertson W, Liddy R, Aird JJ, Mulligan N, Neary PM, McEntee GP, Conneely JB, Cahill RA (2023) Real-time administration of indocyanine green in combination with computer vision and artificial intelligence for the identification and delineation of colorectal liver metastases. Surg Open Sci 12:48–54. https://doi.org/10.1016/j.sopen.2023.03.004
Somashekhar SP, Kumar CR, Ashwin KR, Zaveri SS, Jampani A, Ramya Y, Parameswaran R, Rakshit S (2020) Can low-cost indo cyanine green florescence technique for sentinel lymph node biopsy replace dual dye (radio-colloid and blue dye) technique in early breast cancer: a prospective two-arm comparative study. Clin Breast Cancer 20:e576–e583. https://doi.org/10.1016/j.clbc.2020.03.013
Acknowledgements
The authors acknowledge the contribution of a medical writer, Sandy Field, PhD, for English language editing and formatting of this manuscript.
Funding
This work was carried out without any funding grant.
Author information
Authors and Affiliations
Contributions
CFP, IV and GL contributed to the design and implementation of the research; CFP, AB and IV contributed to resources; CFP, IV and GL contributed to management of data and formal analysis; CFP, IV and GL contributed to original draft preparation; CFP, IV, AB, DL and GL contributed to writing, review and editing the paper.
Corresponding author
Ethics declarations
Conflict of interest
The authors have no conflicts of interest to report.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Pop, C.F., Veys, I., Bormans, A. et al. Fluorescence imaging for real-time detection of breast cancer tumors using IV injection of indocyanine green with non-conventional imaging: a systematic review of preclinical and clinical studies of perioperative imaging technologies. Breast Cancer Res Treat 204, 429–442 (2024). https://doi.org/10.1007/s10549-023-07199-1
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10549-023-07199-1