As x-ray and electron tomography is pushed further into the nanoscale, the limitations of rotation stages become more apparent, leading to challenges in the alignment of the acquired projection images. Here we present an approach for rapid post-acquisition alignment of these projections to obtain high quality three-dimensional images. Our approach is based on a joint estimation of alignment errors, and the object, using an iterative refinement procedure. With simulated data where we know the alignment error of each projection image, our approach shows a residual alignment error that is a factor of a thousand smaller, and it reaches the same error level in the reconstructed image in less than half the number of iterations. We then show its application to experimental data in x-ray and electron nanotomography.
Similar content being viewed by others
Introduction
In tomography, a series of two-dimensional (2D) projections are acquired as a three-dimensional (3D) object is rotated about one or more axes, after which a 3D reconstruction of the object is obtained. Implicit in the approach is the idea that the only differences between the projections are the known rotational angles, with no additional translations or other motions of the object (Fig. 1). This condition is easy to meet in traditional forms of tomography at millimeter length scales when using precision rotation stages; however, it becomes challenging at sub-100 nm length scale of x-ray or electron nanotomography, where one uses high resolution microscopes to obtain 2D projections revealing nanoscale morphology. At these fine length scales, imperfections in rotation stage motion (often called runout errors, or spindle errors) become more noticeable. While the synchronous (reproducible) component of runout errors can be characterized in advance1 and corrected for, this is not the case for asynchronous errors due to imperfect roundness of bearings and other factors. These random runout errors are typically at the level of about 15 nm in today’s best rotation stages. A limited number of microscopes incorporate sophisticated metrology systems to measure and correct for runout error within a certain limit2,3,4,5; however, experimental constraints such as limited working distance, or cryogenic specimen conditions, make systems of this type the exception rather than the rule.
There is a class of “iterative reprojection” techniques which have been demonstrated to provide a good solution to the uncalibrated tomography problem through a succession of iterative alignment steps following tomographic reconstruction steps. To date, most prior work in this area has relied on filtered backprojection as the tomographic reconstruction algorithm because of its speed and efficiency. However, because many x-ray microscopes take more time to acquire projection images of large objects than electron microscopes do of smaller objects, it is not uncommon in x-ray microscopy to be limited to tomographic datasets with a smaller number of projections than one would have demanded by strict adherence to the Crowther criterion6. In this case of limited number and range of projection angles, iterative reconstruction algorithms such as algebraic7,8 or maximum likelihood9 methods are preferred, because they allow for including a priori information about the data and object, which is essential in obtaining satisfactory reconstructions. Therefore, several authors have developed approaches to using similar iterative reconstruction methods at each step of the reprojection iteration10,11,12,13. However, iterations within iterations are inefficient, and the compute-intensive nature of these iterative methods makes their use less practical for large x-ray datasets.
We report here a new variant to iterative reprojection tomography alignment with a key advance: rather than run N cycles of iterative tomographic reconstruction for each of the M steps of iterative reprojection sequentially (Algorithm 1), we carry out one joint reprojection step after each iteration of the reconstruction algorithm (Algorithm 2). Because the computational complexity of the proposed method (Algorithm 2) is \({\mathscr{O}}(N)\) rather than \({\mathscr{O}}(NM)\) with N and M being the number of reconstruction iterate and alignment iterate, this has the potential for significantly speeding up the combined alignment and reconstruction process. It also avoids the possibility of having separate algorithmic steps traveling for long distances in one direction in parameter space while another step may need to move in an orthogonal direction. We demonstrate the use of this approach in x-ray and electron microscopy using both simulated and experimental datasets, where we have obtained improved reconstructions from difficult-to-align tomographic datasets. We show by example that the proposed joint reprojection algorithm (Algorithm 2) can produce such improved reconstructions with higher convergence rates than the conventional sequential algorithm (Algorithm 1), including with noisy and undersampled datasets. This is shown both in selected 2D slices, and in a 3D rendering.
Related Work
A wide range of approaches for projection alignment are employed in electron tomography14. The most common approach is to use cross-correlation between projections acquired at adjacent rotation angles15,16,17,18, or correlation of vertical variations in the mass of the sample19. However, two features in 3D space can have their apparent separation change in projections as a function of rotation angle, leading to ambiguities on which feature dominates the cross-correlation. These ambiguities exponentiate as the number of features are increased, or as the rotation angles between projection images are widened. As a result, while cross-correlation alignment can remove frame-to-frame “jitter” in tomographic datasets, it cannot be relied upon to find a common rotation axis for complete set of projections20.
When specimens are mounted within semi-transparent capillary holders, one can use high-contrast capillary edges to correct for jitter21. An alternative approach is to place fiducial markers such as small gold beads22 or silica spheres23 on the specimen mount or directly on the specimen, and identify them either manually or automatically24,25; their positions can then be used to correct for alignment errors26. This approach is quite successful, and is frequently employed; however, it comes at the cost of adding objects that can complicate sample preparation, obscure specimen features in certain projections, and add material that may complicate analytical methods such as the analysis of fluorescent x-rays.
For those situations where the addition of fiducial marker materials is problematic, one can instead use a variety of feature detection schemes to identify marker positions intrinsic to the specimen, after which various alignment procedures are applied27,28,29,30 including the use of bundle adjustment31. Among these natural feature selection schemes are object corner detection32, wavelet-based detection33, Canny edge detection34, feature curvature detection35, or common-line approach for registration of features in Fourier space36. One can use Markov random fields37 or the Scale-Invariant Feature Transform (SIFT)38,39 to refine the correspondence of features throughout the tomographic dataset. Finally, in the much simpler case of very sparse and local features, one can simply fit sinusoidal curves onto features in the sinogram representation of a set of line projections, and shift projections onto the fitted sinusoid curve40.
These methods are widely employed with good success in electron tomography, where the mean free path for inelastic scattering is often in the 100–200 nm range even for biological specimens with low atomic number, so that it is rare to study samples thicker than about 1 \(\mu \) m (high angle dark field can allow scanning transmission electron microscopes, or STEMs, to image somewhat thicker materials41,42). The situation can be more challenging in nanotomography with x-ray microscopes, where the great penetration of x-rays means that samples tens of micrometers or more in size can be studied43,44,45,46,47,48. This freedom to work with larger specimens means that while feature-based alignment can still be employed for imaging thin specimens with low contrast10,43, in STEM tomography and especially in hard x-ray nanotomography it becomes increasingly challenging to track fiducials or intrinsic features due to the overlap of a large number of features in depth as seen from any one projection.
In all of the above techniques, the primary strategy is to perform an alignment that is as accurate as possible before tomographic reconstruction. In the last few decades, a new set of automatic alignment techniques have been introduced based on a “bootstrap” process49, now commonly referred to as “iterative reprojection”. These techniques attempt to achieve simultaneous alignment and reconstruction through an iterative refinement process. They are based on the fact that the measurement process (forward model) and object reconstruction (inverse model) should be consistent only for a correct alignment geometry. Already in its initial implementation for electron tomography49, a multiscale approach of the method has been used, where first a downsampled version of the projections is used to generate a lower resolution 3D object reconstruction for a first pass of alignment; this first pass with a smaller dataset can align large features in images and works quickly, after which one can improve the alignment at higher resolution until finally the full resolution dataset is used49,50. A variation on this approach is to generate a low-quality object reconstruction from a single projection and then use all the remaining projections for another object reconstruction, and to then align these 3D objects51. One can use projection cross-correlation and the common-line approach for an initial alignment so as to improve convergence times52. Within an optimization framework, iterative reprojection can incorporate a variety of criteria to seek optimal alignment parameters, including contrast maximization in the overall image49 or in sub-tomogram features53, cross-correlation of reprojection and original projection images54, and for cost function reduction a quasi-Newton distance minimization55 or a Levenberg-Marquardt distance minimization56. While initially developed for electron nanotomography, iterative reprojection schemes have also been applied in x-ray microscopy10,57 and with commercial x-ray microtomography systems58,59. As was noted above, all of these prior approaches used variations on Algorithm 1, whereas our approach using Algorithm 2 produces faster convergence rates and more robust reconstructions, and can yield better accuracy especially in the case of tomograms with a limited set of projection angles.
Results and Discussions
Noise studies with simulated data
We applied the proposed joint algorithm on the simulated data shown in Fig. 2 with different noise levels. We generated an analytical phantom for evaluating the algorithm performance. The phantom consists of three spherical objects with various radii. We computed 100 projection images of the phantom with tilt angles uniformly ranging from 0 to 180 degrees for a 100 × 100 pixel area detector. We then simulated jitter in the measurement process by randomly translating the projections. The parameters for both transverse and axial shifts were generated independently from a uniform distribution \(U(-10,\mathrm{10)}\) pixels in extent for each axis and for each projection. In other words, we randomly translated each projection image by a maximum of 10 pixels both in the transverse (to the axis of rotation) and axial (along the direction of the axis of rotation) directions. This corresponds to 10% jitter of the complete field of view (for a projection image of 100 × 100 size), and is typical for most nanoscale x-ray fluorescence tomography datasets. Finally, we added Gaussian noise to measurements for simulating various experimental conditions and evaluating algorithm robustness. Four different noise levels (i.e., 0%, 5%, 10% and 20%) were used, where the values in parenthesis mean that the standard deviation of noise was selected as a percentage of the maximum value of the signal in the projections.
For joint alignment and reconstruction, we used the simultaneous iterative reconstruction technique (SIRT)9 for the reconstruction step at each iteration of the joint algorithm. The unaligned data and the corresponding reconstructions after convergence of the algorithm are presented in Fig. 3 together with the aligned data and reconstructions. Our joint algorithm could generate satisfactory recovery of all three spheres even for high levels of noise up to 20%, while unaligned data failed to resolve the smallest sphere, and only recovered blurred versions of the larger spheres.
The estimation errors between the true and recovered alignment parameters for both transverse and axial shifts for each projection image are plotted in Fig. 4. The dashed red lines indicate the single pixel bound. The joint algorithm could satisfactorily estimate all axial shifts for up to 10% noise (and most of them for 20% noise) at sub-pixel accuracy. The estimation accuracy is slightly worse for transverse shifts, and as expected, drops with increasing noise levels. We also observed some correlation between axial and transverse estimations. Basically, estimation of the transverse and axial shifts are usually both affected to varying degree, possibly from the image noise or other imaging artifacts appearing in a single projection image.
Comparative studies
Figure 5 demonstrates the comparison of convergence plots of the prior sequential iterative reprojection approach (Algorithm 1), and our joint approach (Algorithm 2). The plots are obtained by calculating the relative mean-square-error between successive iterations. We separately calculated the errors in transverse (solid line) and axial (dashed line) directions. For the sequential approach, we used 40 iterations of the SIRT algorithm to reconstruct an estimate of the object before performing the alignment step, and we re-iterated this process 10 times, providing a total number of 400 reconstruction iterations. For the joint algorithm, we performed 400 SIRT iterations; however, this time we aligned the projection images with cross-correlation after each SIRT iteration. Our joint approach (Algorithm 2) provided superior convergence results for both the alignment estimation and the object reconstruction. We observed that the convergence rate of the axial alignment is more rapid than in transverse alignment. This is probably because of the fact that parallel-beam tomographic data are coupled in the transverse direction (where objects off of the axis of rotation move from projection to projection), while features remain at a constant position in the axial direction.
The initialization of iterative algorithms usually has an impact of the convergence rates and the final solution. In x-ray tomography, it is common practice to use filtered backprojection (FBP) results as the initial object estimate for iterative algorithms. Therefore, we examined the convergence rates of the joint alignment algorithm with different initializations as shown in Fig. 6. Blue and red plots correspond to the relative mean-square error (MSE) in the axial (rapidly convergent plot) and transverse (slowly convergent plot) alignment parameters for each successive iteration with FBP and with null initialized algorithms, respectively. Once again, axial alignment showed faster convergence than transverse alignment for both initializations. We observed that the convergence rate of the FBP-initialized algorithm is slower than the null-initialization. This is potentially because FBP provides strong streaking artifacts with misaligned projections and when too few rotation angles are used according to the Crowther criterion; this can make the algorithm be trapped in a local solution or need more iterations to recover from a false initial point. This is more evident by looking at the initial reconstructed images after 10 iterations, where the FBP-initialized algorithm produces artifacts on spherical objects, while the the Null-initialized algorithm produces smooth representations of spherical objects as expected.
We also evaluated the effect of the choice of the reconstruction algorithm on the convergence rates in Fig. 7. We compared two common iterative methods – SIRT, and maximum-likelihood expectation-maximization (MLEM)60 – and a common non-iterative method, GridRec61. GridRec is a direct Fourier-based method that relies on discrete Fourier transforms of data in a manner similar to the filtered backprojection (FBP) algorithm. The main difference of GridRec with the FBP is that it samples a slice in the Fourier domain on a Cartesian grid before transforming it back to the spatial domain62. The utilization of fast Fourier transforms on a Cartesian grid outperforms other methods in terms of computational speed and is usually desired for quick reconstructions. We used a noise-free simulated object, and performed 400 joint algorithm iterations. MLEM provided the best results and the alignment procedure converged in less than 100 iterations, although subsequent iterations further improved the reconstructed object. The joint estimation of both alignment and reconstruction also cause some cross-talk between the convergence rates. For example, at the 220th and 275th iterations in MLEM we notice sudden increases in MSE due to changes in the alignment estimates, which we believe are due to numerical inaccuracies. The joint algorithm with GridRec failed to estimate both alignment and object, potentially because of the drift of the center of rotation at each iteration step, to the point where after 10 iterations the corrected translations were so large that the reconstructed object started to drift away from the field of view. It is well-known that Fourier-based methods such as standard filtered backprojection and GridRec do not work well when one has too few projection angles (thus violating the Crowther criterion, so that much of Fourier space does not contain projection data), so it is not surprising that both SIRT and MLEM outperform GridRec in our case of using 100 rotation angles with 100 × 100 pixel projection images.
We compared the reconstructions after the convergence of both the sequential (Algorithm 1) and joint (Algorithm 2) iterative reprojection approaches in Fig. 8. Both showed reasonably good quality sinogram reconstructions; however, the sequential algorithm introduced some reconstruction artifacts due to slight mismatch in estimating alignment parameters. These artifacts were almost nonexistent in the joint reprojection algorithm.
Experimental validation
Having characterized the performance improvements of our joint iterative reconstruction approach (Algorithm 2) on simulated data, we then used it to obtain object slice reconstructions from the four experimental datasets: three from various x-ray microscopes, and one from an electron microscope. Sinograms and reconstructed object slices are shown in Fig. 9 both before and after the application of our joint iterative reprojection approach for compensating alignment errors. The severity of the alignment errors in each of the nanotomography datasets can be seen in by the “wiggles” in the unaligned sinograms as compared to the aligned sinograms, and the improvement in reconstructed image quality is evident.
Figure 10 shows the convergence rates of transverse (blue) and axial (red) alignment parameters using the experimental Dataset B. The axial alignment convergence was faster than in the transverse direction; however, both converged after about 50 iterations. The convergence rates are similar to those we observed in simulation studies as shown in Fig. 5.
Discussion
Tomographic reconstruction of the 3D object from the sinogram data requires precise knowledge of the object’s center of rotation with respect to the detector positioning. If this is not the case, systematic ring-like imaging artifacts appear in the reconstructed images, particularly at sharp boundaries, even when there is no geometrical distortion during data acquisition. The center of rotation can typically be determined by using various algorithms at a post-acquisition phase63,64,65,66. With the proposed joint algorithm we did not see any incorrect-centering artifacts in the reconstructions presented in this paper. This is probably because the back/re-projection operator enforces object updates at the center of the detector field-of-view, and automatically corrects the rotation center within its iterations. However, for cases, when the rotation center is significantly off, there is the possibility that the algorithm can fail to recover an object due to excessive blurring in intermediate object updates of the algorithm. In these situations, one of the above-mentioned algorithms can be used either at intermediate steps of the algorithm or after the alignment procedure to ensure that the rotation center used for back/re-projection is in the close proximity of the true rotation center.
In conclusion, we present here a rapid algorithm (Algorithm 2) for jointly aligning datasets and reconstructing the object. The algorithm avoids iterations in iterations, and can converge significantly faster than the previously used approach (Algorithm 1). We have characterized our joint iterative reprojection approach in simulations where the actual alignment errors were known, and we have shown that it works well when applied to experimental data. Given that many x-ray and electron nanotomography datasets are affected by nanometer-scale alignment errors caused by imperfections in rotation stages, this approach provides a path to obtain improved 3D reconstructions. Our initial demonstration presented here considered only translation errors in the individual projection images. We expect that image rotation errors can also be corrected in a similar manner, and that one can also correct for errors between the assumed and actual specimen rotation angles in the tomographic dataset. We plan on studying these extensions of our approach in subsequent work.
Methods
Problem definition
Here, we describe the mathematical basis of the alignment problem for 3D tomography. Let \(f(x,y,z)\) be an unknown 3D object to be recovered. This can be expressed as a vector \(f=[{f}_{1},{f}_{2},\ldots ,{f}_{N}{]}^{T}\), where \(N={N}_{1}\times {N}_{2}\times {N}_{3}\) is the total number of voxels in \(f\). Let the vector \({p}_{i}\) represents a 2D projection image with \(M={M}_{1}\times {M}_{2}\) pixels captured at the \({i}^{th}\) tilt angle \({\theta }_{i}\). Ideally (without any translation error), the measurement process for each projection can be represented by,
where \({p}_{i}\) and \({W}_{i}({\theta }_{i})\) are, respectively, the recorded projection data and the projection matrices for the \({i}^{th}\) tilt angle, and \(e\) is the additive measurement noise during the acquisition process.
Now we consider the translation model. The 3D object f can freely translate in the \(x\), \(y\), and \(z\) directions, which leads to a 2D translation error over the projection images. Therefore, in our model, we apply the translation error to the 2D projection images instead of the 3D object (translation errors in the \(z\) direction have no effect in standard parallel-beam tomography). The measurement geometry and translations are given in Fig. 1. The \({i}^{th}\) projection image can be described as \({p}_{i}={T}_{i}({h}_{i},{v}_{i})W({\theta }_{i})f\), where \({T}_{i}\) represents an \(M\times M\) geometrical transformation operator, and \({h}_{i},{v}_{i}\) are unknown transverse and axial translation errors for \({i}^{th}\) projection image, respectively. If we stack the \(S\) projection images, Eqn. (1) can be rewritten in a compact form as,
where \(p=[{p}_{1},{p}_{2},\ldots ,{p}_{S}{]}^{T}\) is the vector of size \(MS\times 1\), holding the tomographic measurement dataset, \(T(h,v)={\rm{diag}}[{T}_{1}({h}_{1},{v}_{1}),{T}_{2}({h}_{2},{v}_{2}),\ldots ,{T}_{S}({h}_{S},{v}_{S})]\) is the geometrical translation matrix of size \(MS\times MS\), and \(W(\theta )=[{W}_{1}({\theta }_{1}),{W}_{2}({\theta }_{2}),\ldots ,{W}_{S}({\theta }_{S}{)]}^{T}\) is the matrix of size \(MS\times N\) for the measurement process. Here \(h=[{h}_{1},{h}_{2},\ldots ,{h}_{S}{]}^{T}\) and \(v=[{v}_{1},{v}_{2},\ldots ,{v}_{S}{]}^{T}\) are unknown transverse and axial translation errors for all \(S\) projection images respectively, and \(\theta =[{\theta }_{1},{\theta }_{2},\ldots ,{\theta }_{S}{]}^{T}\) denotes \(S\) tilt angles during the acquisition process.
The goal in the iterative reprojection tomography reconstruction is to recover both the translation error \((h,v)=({h}_{1},{v}_{1};{h}_{2},{v}_{2};\ldots ;{h}_{S},{v}_{S})\) and the 3D object \(f\) simultaneously in a joint fashion. The most common way to achieve this is to minimize the distance between the real captured projection data and reprojection data calculated from the current reconstruction. Algorithm 11 summarizes the details of the proposed method.
Implementation details
We implemented our new approach (Algorithm 11) in Python while drawing upon openly available libraries. We used the open-source and documented package TomoPy67 for the reconstruction and reprojection operators, and the open-source Scikit-Image library68 for calculating phase-correlation based sub-pixel image registration69 and image warping operators. The analytical phantoms and corresponding simulated measurements were generated using Xdesign70, which is another open-source Python package for simulating x-ray data.
Experimental datasets
Dataset A is of X-ray fluorescence data collected at the Bionanoprobe (BNP)71 at the Advanced Photon Source at Argonne National Laboratory. This instrument uses a double crystal Si 111 monochromator to produce 10 keV incident x-rays, and a Fresnel zone plate to produce a focus spot with a theoretical Rayleigh resolution of 85 nm. A silicon drift detector was placed at 90° with regards to the incident X-ray beam direction to collect fluorescence signals from which element-specific images were obtained using the analysis program MAPS72. A tomographic dataset was obtained using projection images of 79 × 56 pixels at 60 nm pixel size and a dwell time of 100 msec per pixel, over a rotation angle range of 0° to 135° at 3° increments (0° refers to the sample plane being perpendicular to the incident X-ray beam). This dataset represents the X-ray fluorescence from a cluster of dried bacterial cells with a metallic surface tag imaged at room temperature; the maximum signal was about 50 detected photons per second.
Dataset B is of Gadolinium x-ray fluorescence data collected at the Hard X-ray Nanoprobe (HXN) Beamline73,74 at the National Synchrotron Light Source II (NSLS-II) of Brookhaven National Laboratory. As with the Bionanoprobe, a 10 keV beam was monochromatized, a Fresnel zone plate was used to produce a focus spot with a theoretical Rayleigh resolution of 49 nm, and a three-element fluorescence detector was placed at 90° to the incident beam to collect fluorescence signals from which element-specific images were obtained using PyXRF (https://github.com/NSLS-II/PyXRF), which is an equivalent elemental mapping software such as MAPS. A tomographic dataset was obtained using projection images of \(200\times 120\) pixels at 25 nm pixel size and a dwell time of 30 msec per pixel (with a maximum Gd signal of about 450 detected photons per pixel), over a rotation angle range of −90° to +90° at 3° increments. The sample used was a ceramic composite with mixed ionic and electronic conductivity75, consisting of a Ce0.8Gd0.2O2 (CGO) oxygen ionic conductive phase and a CoFe2O4 (CFO) electronic conductive phase. Nanostructure at the grain boundary is expected to have significant impact on the conductivity of this material75.
Dataset C is of 8 keV x-ray transmission data collected at the Transmission X-ray Microscope (TXM) of sector 32-ID at the Advanced Photon Source76. This microscope uses a specialized diffractive condenser optic to illuminate the sample, and a Fresnel zone plate objective with an experimentally measured resolution of 60 nm to produce a magnified image on a scintillator/lens/CCD image detector system. A tomographic dataset was obtained using projection images of \(2160\times 2560\) pixels each at 15 nm pixel size and a per-projection acquisition time of 1 second, over a rotation angle range of −90° to +90° at 0.25° increments. The dataset was then downsampled \(4\times \) to \(540\times 640\) pixels, from which a \(200\times 200\) pixel subregion was extracted at each projection angle for tomographic reconstruction. The 32-ID TZXM is equipped with an air bearing rotation stage with \( \sim 1.25\) \(\mu \) rad wobble and an eccentricity of \(\pm 100\) nm over 360°, so this dataset had fewer initial alignment errors than the x-ray fluorescence datasets A and B though thermal drifts can still affect specimen translation over the 20 minute data acquisition time. The sample used was a cathode particle agglomerate (Li1.2Co0.1Ni0.15Mn0.55O2) of a Lithium-ion battery with volume of about 6 \(\mu m{}^{3}\).
Dataset D is of 200 kV high-angle annular dark field (HAADF) data acquired using a Hitachi HD2300 scanning transmission electron microscope at the Northwestern University Atomic and Nanoscale Characterization Experimental Center (NUANCE). A tomographic dataset was obtained using projection images of 256 × 256 pixels at 66 nm pixel size and a per-pixel dwell time of 10 \(\mu \)sec, over a rotation angle range of −60° to +60° at 1° increments. A nano-diffraction beam setting was used in order to maintain sufficient depth of focus. The specimen imaged was a human buccal cell collected utilizing a Cytobrush and deposited on a TEM grid. The sample was chemically fixed with glutaraldehyde (2.5%) and formaldehyde (2%) in PBS for 20 minutes. After fixation, the sample was gently washed and then plunge-frozen in liquid ethane (Vitrobot Mark III, FEI) and turbo freeze dried (K775X, Emitech). The dry sample was then plunge-frozen again, transferred to a cryo single tilt holder (Gatan) and kept at liquid nitrogen temperatures during imaging.
Data availability
The algorithms used here are made publicly available in the TomoPy package at https://tomopy.readthedocs.io. The simulated and experimental datasets used are available in TomoBank at https://tomobank.readthedocs.io.
References
Grejda, R., Marsh, E. & Vallance, R. Techniques for calibrating spindles with nanometer error motion. Precision Engineering 29, 113–123 (2005).
Wang, J. et al. Automated markerless full field hard x-ray microscopic tomography at sub-50?nm 3-dimension spatial resolution. Applied Physics Letters 100, 143107–4 (2012).
Holler, M. et al. An instrument for 3D x-ray nano-imaging. Review of Scientific Instruments 83, 073703 (2012).
Kim, J., Lauer, K., Yan, H., Chu, Y. S. & Nazaretski, E. Compact prototype apparatus for reducing the circle of confusion down to 40 nm for x-ray nanotomography. Review of Scientific Instruments 84, 035006 (2013).
Villar, F. et al. Nanopositioning for the ID16A-NI endstation. In MEDSI Mechanical Engineering Design of Synchrotron Radiation Equipment and Instrumentation (Australian Synchrotron, 2014).
Crowther, R. A., DeRosier, D. J. & Klug, A. The reconstruction of a three-dimensional structure from projections and its application to electron microscopy. Proceedings of the Royal Society of London A 317, 319–340 (1970).
Gordon, R., Bender, R. & Herman, G. T. Algebraic reconstruction techniques (ART) for three-dimensional electron microscopy and x-ray photography. Journal of Theoretical Biology 29, 471–481 (1970).
Gilbert, P. Iterative methods for the three-dimensional reconstruction of an object from projections. Journal of Theoretical Biology 36, 105–117 (1972).
Shepp, L. A. & Vardi, Y. Maximum likelihood reconstruction for emission tomography. IEEE Transactions on Medical Imaging 1, 113–122 (1982).
Parkinson, D. Y. et al. Automatic alignment and reconstruction of images for soft x-ray tomography. Journal of Structural Biology 177, 259–266 (2012).
Guizar-Sicairos, M. et al. Quantitative interior x-ray nanotomography by a hybrid imaging technique. Optica 2, 259–266 (2015).
KUPSCH, A. et al. Missing wedge computed tomography by iterative algorithm directt. Journal of Microscopy 261, 36–45 (2016).
Rojbani, H., Baudrier, E., Naegel, B., Mazo, L. & Hamouda, A. Joint 3D alignment-reconstruction multi-scale approach for cryo electron tomography. In Kybic, J. & Sonka, M. (eds.) IEEE 13th International Symposium on Biomedical Imaging (ISBI), 1109–1113, https://doi.org/10.1109/isbi.2016.7493460 (IEEE, 2016).
Amat, F. et al. Alignment of cryo-electron tomography datasets. Methods in Enzymology 482, 343–367 (2010).
Hoppe, W. Zur elektronenmikroskopisch dreidimensionalen Rekonstruktion eines Objektes. Naturwissenschaften 61, 534–536 (1974).
Guckenberger, R. Determination of a common origin in the micrographs of tilt series in three-dimensional electron microscopy. Ultramicroscopy 9, 167–173 (1982).
Dierksen, K., Typke, D., Hegerl, R. & Baumeister, W. Towards automatic electron tomography II. implementation of autofocus and low-dose procedures. Ultramicroscopy 49, 109–120 (1993).
Hayashida, M., Terauchi, S. & Fujimoto, T. Automatic coarse-alignment for TEM tilt series of rod-shaped specimens collected with a full angular range. Micron 41, 540–545 (2010).
Guizar-Sicairos, M. et al. Phase tomography from x-ray coherent diffractive imaging projections. Opt. Express 19, 21345–21357 (2011).
Saxton, W. O., Baumeister, W. & Hahn, M. Three-dimensional reconstruction of imperfect two-dimensional crystals. Ultramicroscopy 13, 57–70 (1984).
Kwon, I.-H., Lim, J. & Hong, C.-K. Runout error correction in tomographic reconstruction by intensity summation method. Journal of Synchrotron Radiation 23, 1237–1240 (2016).
Olins, D. et al. Electron microscope tomography: transcription in three dimensions. Science 220, 498–500 (1983).
Lim, J., Kim, H. & Park, S. Y. Hard x-ray nanotomography beamline 7C XNI at PLS-II. Journal of Synchrotron Radiation 21, 827–831 (2014).
Cao, M., Takaoka, A., Zhang, H.-B. & Nishi, R. An automatic method of detecting and tracking fiducial markers for alignment in electron tomography. Journal of Electron Microscopy 60, 39–46 (2011).
Han, R., Wang, L., Liu, Z., Sun, F. & Zhang, F. A novel fully automatic scheme for fiducial marker-based alignment in electron tomography. Journal of Structural Biology 192, 403–417 (2015).
Jing, Z. & Sachs, F. Alignment of tomographic projections using an incomplete set of fiducial markers. Ultramicroscopy 35, 37–43 (1991).
Ress, D., Harlow, M. L., Schwarz, M., Marshall, R. M. & McMahan, U. J. Automatic acquisition of fiducial markers and alignment of images in tilt series for electron tomography. Journal of Electron Microscopy 48, 277–287 (1999).
Sorzano, C. et al. Marker-free image registration of electron tomography tilt-series. BMC Bioinformatics 10, 124–11 (2009).
Castaño-Diez, D., Al-Amoudi, A., Glynn, A.-M., Seybert, A. & Frangakis, A. S. Fiducial-less alignment of cryo-sections. Journal of Structural Biology 159, 413–423 (2007).
Castaño-Diez, D., Scheffer, M., Al-Amoudi, A. & Frangakis, A. S. Alignator: A GPU powered software package for robust fiducial-less alignment of cryo tilt–series. Journal of Structural Biology 170, 117–126 (2010).
Lawrence, A., Bouwer, J. C., Perkins, G. & Ellisman, M. H. Transform-based backprojection for volume reconstruction of large format electron microscope tilt series. Journal of Structural Biology 154, 144–167 (2006).
Brandt, S., Heikkonen, J. & Engelhardt, P. Multiphase method for automatic alignment of transmission electron microscope images using markers. Journal of Structural Biology 133, 10–22 (2001).
Brandt, S., Heikkonen, J. & Engelhardt, P. Automatic alignment of transmission electron microscope tilt series without fiducial markers. Journal of Structural Biology 136, 201–213 (2001).
Duan, W., Kuester, F., Gaudiot, J.-L. & Hammami, O. Automatic object and image alignment using Fourier descriptors. Image and Vision Computing 26, 1196–1206 (2008).
Phan, S., Bouwer, J., Lanman, J., Terada, M. & Lawrence, A. Non-linear bundle adjustment for electron tomography. In Burgin, M. et al. (eds) 2009 WRI World Congress on Computer Science and Information Engineering, 604–612 (IEEE, 2009).
Liu, Y., Penczek, P. A., McEwen, B. F. & Frank, J. A marker-free alignment method for electron tomography. Ultramicroscopy 58, 393–402 (1995).
Amat, F. et al. Markov random field based automatic image alignment for electron tomography. Journal of Structural Biology 161, 260–275 (2008).
Liu, Y. et al. TXM-Wizard: a program for advanced data collection and evaluation in full-field transmission x-ray microscopy. Journal of Synchrotron Radiation 19, 281–287 (2012).
Han, R. et al. A marker-free automatic alignment method based on scale-invariant features. Journal of Structural Biology 186, 167–180 (2014).
Cheng, C.-C., Chien, C.-C., Chen, H.-H., Hwu, Y. & Ching, Y.-T. Image alignment for tomography reconstruction from synchrotron x-ray microscopic images. PloS ONE 9, e84675 (2014).
Pennycook, S. J. Z-Contrast STEM for materials science. Ultramicroscopy 30, 58–69 (1989).
Ercius, P., Weyland, M., Muller, D. A. & Gignac, L. M. Three-dimensional imaging of nanovoids in copper interconnects using incoherent bright field tomography. Applied Physics Letters 88, 243116–1–3 (2006).
Weiß, D. et al. Computed tomography of cryogenic biological specimens based on x-ray microscopic images. Ultramicroscopy 84, 185–197 (2000).
Wang, Y., Jacobsen, C., Maser, J. & Osanna, A. Soft x-ray microscopy with a cryo scanning transmission x-ray microscope: II. Tomography. Journal of Microscopy 197, 80–93 (2000).
Larabell, C. A. & Le Gros, M. A. X-ray tomography generates 3-D reconstructions of the yeast, saccharomyces cerevisiae, at 60-nm resolution. Molecular Biology of the Cell 15, 957–962 (2004).
Tkachuk, A. et al. X-ray computed tomography in Zernike phase contrast mode at 8 kev with 50-nm resolution using cu rotating anode x-ray source. Zeitschrift für Kristallographie 222, 650–655 (2007).
Dierolf, M. et al. Ptychographic x-ray computed tomography at the nanoscale. Nature 467, 436–439 (2010).
Holler, M. et al. X-ray ptychographic computed tomography at 16-nm isotropic 3D resolution. Scientific Reports 4, 3857 (2014).
Dengler, J. A multi-resolution approach to the 3d reconstruction from an electron microscope tilt series solving the alignment problem without gold particles. Ultramicroscopy 30, 337–348 (1989).
Latham, S. J., Kingston, A. M., Recur, B., Myers, G. R. & Sheppard, A. P. Multi-resolution radiograph alignment for motion correction in x-ray micro-tomography. In Stock, S. R., Müller, B. & Wang, G. (eds) Developments in X-ray Tomography X, vol. 9967, 996710–12 (SPIE, 2016).
Tomonaga, S., Baba, M. & Babal, N. Alternative automatic alignment method for specimen tilt-series images based on back-projected volume data cross-correlations. Journal of Electron Microscopy 63, 279–294 (2014).
Owen, C. H. & Landis, W. J. Alignment of electron tomographic series by correlation without the use of gold particles. Ultramicroscopy 63, 27–38 (1996).
Houben, L. & Sadan, M. B. Refinement procedure for the image alignment in high-resolution electron tomography. Ultramicroscopy 111, 1512–1520 (2011).
Winkler, H. & Taylor, K. A. Accurate marker-free alignment with simultaneous geometry determination and reconstruction of tilt series in electron tomography. Ultramicroscopy 106, 240–254 (2006).
Yang, C., Ng, E. G. & Penczek, P. A. Unified 3-D structure and projection orientation refinement using quasi-newton algorithm. Journal of Structural Biology 149, 53–64 (2005).
Bleichrodt, F. & Batenburg, K. J. Automatic optimization of alignment parameters for tomography datasets. In Kämäräinen, J.-K. & Koskela, M. (eds.) Image Analysis: 18th Scandinavian Conference, vol. 7944 of Lecture Notes in Computer Science, 489–500, doi:10.1007/978-3-642-38886-6_46 (Springer, 2013).
Mayo, S. C., Miller, P., Gao, D. & Sheffield-Parker, J. Software image alignment for x-ray microtomography with submicrometre resolution using a SEM-based x-ray microscope. Journal of Microscopy 228, 257–263 (2007).
Sasov, A., Liu, X. & Salmon, P. L. Compensation of mechanical inaccuracies in micro-CT and nano-CT. In Stock, S. R. (ed.) Developments in X-ray Tomography VI, vol. 7078, 70781C. SPIE (SPIE, 2008).
Liu, X., Laperre, K. & Sasov, A. Practical pseudo-3D registration for large tomographic images. Developments in X-ray Tomography IX 9212 (2014).
Dempster, A. P., Laird, N. M. & Rubin, D. B. Maximum likelihood from incomplete data via the EM algorithm. Journal of the Royal Statistical Society B 39 (1977).
Dowd, B. et al. Developments in synchrotron x-ray computed tomography at the National Synchrotron Light Source. In Editor (ed.) Developments in X-ray Tomography II, vol. 3772, 224–236 (1999).
O’Sullivan, J. D. A fast sinc function gridding algorithm for Fourier inversion in computer tomography. IEEE Transactions on Medical Imaging 4, 200–207 (1985).
Donath, T., Beckmann, F. & Schreyer, A. Automated determination of the center of rotation in tomography data. J. Opt. Soc. Am. A 23, 1048–1057 (2006).
Vo, N. T., Drakopoulos, M., Atwood, R. C. & Reinhard, C. Reliable method for calculating the center of rotation in parallel-beam tomography. Opt. Express 22, 19078–19086 (2014).
Yang, X., De Carlo, F., Phatak, C. & Gürsoy, D. A convolutional neural network approach to calibrating the rotation axis for X-ray computed tomography. Journal of Synchrotron Radiation 24, 469–475 (2017).
Jun, K. & Yoon, S. Alignment solution for CT image reconstruction using fixed point and virtual rotation axis. Scientific Reports 7, 41218 (2017).
Gürsoy, D., De Carlo, F., Xiao, X. & Jacobsen, C. TomoPy: a framework for the analysis of synchrotron tomographic data. Journal of Synchrotron Radiation 21, 1188–1193 (2014).
van der Walt, S. et al. Scikit-Image: Image processing in Python. PeerJ 2 (2014).
Guizar-Sicairos, M., Thurman, S. T. & Fienup, J. R. Efficient subpixel image registration algorithms. Optics Letters 33, 156–158 (2008).
Ching, D. J. & Gürsoy, D. XDesign: an open-source software package for designing x-ray imaging phantoms and experiments. Journal of Synchrotron Radiation 24, 537–544 (2017).
Chen, S. et al. The Bionanoprobe: hard x-ray fluorescence nanoprobe with cryogenic capabilities. Journal of Synchrotron Radiation 21, 66–75 (2014).
Vogt, S. MAPS: A set of software tools for analysis and visualization of 3D x-ray fluorescence data sets. Journal de Physique IV 104, 635–638 (2003).
Nazaretski, E. et al. Pushing the limits: an instrument for hard x-ray imaging below 20 nm. Journal of Synchrotron Radiation 22, 336–341 (2015).
Yan, H. et al. Multimodality hard-x-ray imaging of a chromosome with nanoscale spatial resolution. Scientific Reports 6, 20112 (2016).
Lin, Y., Fang, S., Su, D., Brinkman, K. S. & Chen, F. Enhancing grain boundary ionic conductivity in mixed ionic–electronic conductors. Nature Communications 6 (2015).
De Andrade, V. et al. Nanoscale 3D imaging at the Advanced Photon Source. SPIE Newsroom, doi:https://doi.org/10.1117/2.1201604.006461 (2016).
Acknowledgements
We thank Hanfei Yan, Xiaojing Huang, and Tiffany Victor for sharing experimental data presented in this paper. This research used synchrotron light source beamlines and personnel supported by the U.S. Department of Energy (DOE) Office of Science User Facilities in the DOE Office of Science. For the TXM at beamline 32-ID of the Advanced Photon Source at Argonne and the Bionanoprobe at beamline 21-ID, DOE support was provided to Argonne National Laboratory under Contract No. DE-AC02-06CH11357. For the HXN at beamline 3-ID of the National Synchrotron Light Source II (NSLS-II), DOE support was provided to Brookhaven National Laboratory under Contract No. DE-SC0012704. We thank Northwestern University for support under a Data Science Initiative project, and an Innovative Initiatives Incubators project. The electron tomography work made use of the EPIC facility of Northwestern University’s NUANCE Center, which has received support from the Soft and Hybrid Nanotechnology Experimental (SHyNE) Resource (NSF ECCS-1542205); the MRSEC program (NSF DMR-1121262) at the Materials Research Center; the National Institutes of Health (NIGMS R01 GM104530); the International Institute for Nanotechnology (IIN); the Keck Foundation; and the State of Illinois, through the IIN. Finally, we thank the National Institutes of General Medical Services, National Institutes of Health, for support under R01 GM104530.
Author information
Authors and Affiliations
Contributions
D.G. conceived the joint approach, and C.J. suggested the use of iterative reprojection. Y.P.H., K. He, K. Hujsak, S.C., L.M., M.G., Y.C., and V.D.A. obtained experimental data. D.G., Y.P.H., K. He, S.Y., Y.L., O.C., O.C., A.K., and C.J. contributed to convergence test approaches. All authors reviewed the manuscript.
Corresponding author
Ethics declarations
Competing Interests
The authors declare that they have no competing interests.
Additional information
Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Gürsoy, D., Hong, Y.P., He, K. et al. Rapid alignment of nanotomography data using joint iterative reconstruction and reprojection. Sci Rep 7, 11818 (2017). https://doi.org/10.1038/s41598-017-12141-9
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-017-12141-9
- Springer Nature Limited
This article is cited by
-
Numerical methods for CT reconstruction with unknown geometry parameters
Numerical Algorithms (2023)
-
Accelerating error correction in tomographic reconstruction
Communications Materials (2022)
-
A deep learning approach for semantic segmentation of unbalanced data in electron tomography of catalytic materials
Scientific Reports (2022)
-
A joint alignment and reconstruction algorithm for electron tomography to visualize in-depth cell-to-cell interactions
Histochemistry and Cell Biology (2022)
-
Deep Learning-based Inaccuracy Compensation in Reconstruction of High Resolution XCT Data
Scientific Reports (2020)