Abstract
To assess whether phantomless calcium-hydroxyapatite (HA) specific bone mineral density (BMD) measurements with dual-layer spectral computed tomography are accurate in phantoms and vertebral specimens. Ex-vivo human vertebrae (n = 13) and a phantom containing different known HA concentrations were placed in a semi-anthropomorphic abdomen phantom with different extension rings simulating different degrees of obesity. Phantomless dual-layer spectral CT was performed at different tube current settings (500, 250, 125 and 50 mAs). HA-specific BMD was derived from spectral-based virtual monoenergetic images at 50 keV and 200 keV. Values were compared to the HA concentrations of the phantoms and conventional qCT measurements using a reference phantom, respectively. Above 125 mAs, errors for phantom measurements ranged between −1.3% to 4.8%, based on spectral information. In vertebral specimens, high correlations were found between BMD values assessed with spectral CT and conventional qCT (r ranging between 0.96 and 0.99; p < 0.001 for all) with different extension rings, and a high agreement was found in Bland Altman plots. Different degrees of obesity did not have a significant influence on measurements (P > 0.05 for all). These results suggest a high validity of HA-specific BMD measurements based on dual-layer spectral CT examinations in setups simulating different degrees of obesity without the need for a reference phantom, thus demonstrating their feasibility in clinical routine.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Introduction
Osteoporosis is estimated to affect 28 million patients in the European Union with increasing prevalence, due to the aging population1,2. Reducing bone stability and thus increasing the risk of fractures, osteoporosis substantially adds to morbidity and mortality not only of the elderly, but also of younger subjects with impaired bone metabolism as present in patients with cancer or eating disorders3,4. However, up to 70% of eligible women and far more men do not undergo bone mineral density (BMD) screening with dual-energy x-ray absorptiometry (DXA), the clinical standard5,6, although BMD is a predictor for future fracture risk and all-cause mortaliy7,8. Subsequently, only 10–22% of patients receive adequate therapy for osteoporosis1,2,9. In addition to this, DXA values have shown to be affected by several confounding factors, e.g. spondylosis, patient size, and vascular calcifications, likely causing DXA to correctly diagnose osteoporosis in only about 44% of women and 21% of men with prevalent osteoporotic fractures10.
qCT allows for volumetric BMD measurements, commonly applied to the spine, and calibrated to a reference phantom with known hydroxyapatite concentrations, may outperform DXA-based BMD measurements11, and also allows more advanced models for fracture risk assessment such as finite element analysis (FEA)12,13,14. In order to reduce radiation doses, ongoing research also focuses on how to opportunistically use multidetector computed tomography (MDCT) data obtained for various reasons, e.g. for morphologic imaging, disease staging for osteoporosis screening or therapy monitoring15. For example, BMD-equivalent values derived from MDCT have previously been used to predict screw loosening in subjects undergoing spondylodesis16 and incidental osteoporotic vertebral fractures17, and to assess the fracture risk in subjects with inflammatory bowel disease and prostate cancer18,19. However, there are limitations to all CT-based BMD measurements, since they can be affected by scan parameters such as different tube voltages, the use of intravenous contrast medium, the fat fraction within vertebral bone marrow, and beam hardening or scatter artefacts20,21,22,23,24.
One CT technique gaining more and more attention is acquiring spectral information using dual-energy CT (DECT). There are different approaches to exploit the spectral information such as fast kV-switching or using two x-ray sources with different characteristics, known as dual-source CT25,26,27. Recently a new technique with one x-ray tube and two different detector layers mounted upon each other and absorbing different energy spectra of the polychromatic X-ray spectrum, known as dual-layer spectral CT, was introduced27,28. While DECT requires the pre-selection of specific dual-energy CT protocols, spectral CT allows for routinely reconstructing spectral information without the use of a specific protocol. Those methods enable the estimation of object composition by exploiting material- and energy-dependent x-ray absorption of various materials. For example, the quantification of iodine is already commercially available and can be applied to clinical routine. DECT has shown to be feasible for volumetric BMD assessment of the lumbar spine using DXA as a standard of reference29. A recent study reported a high accuracy of density measurements in phantoms with known hydroxyapatite (HA) concentrations30, however, to our best knowledge, the validity of HA-specific BMD measurements has not been assessed previously in vertebral specimens.
Aims of this study therefore were to (i) validate spectral CT based, HA-specific mineral density quantification with a phantom with known HA densities as reference standard, and (ii) to validate spectral CT based BMD quantification in vertebral specimens with qCT as standard reference; both in a setup with an anthropomorphic phantom simulating different degrees of obesity.
Results
Radiation doses, SNR and CNR
CT Dose Index (CTDIvol) ranged between 1.2 mGy at 10 mAs and 63.1 at 500 mAs. At 500 mAs, the signal-to-noise was 29.0 and the contrast-to-noise was 32.6 (Table 1).
Spectral-based BMD quantification in phantoms with known HA densities
Measurements of the four known HA densities of a dedicated phantom produced accurate values, with measurement errors between 0.5% and 4.8% of the specified densities for scans acquired with 125 mAs, which are comparable to the radiation exposures used in most clinical examination protocols, or more (Table 2). At our lowest exposure level, measurement errors increased up to 14.8% (lowest HA concentration assessed with 38 mAs).
Spectral-based BMD quantification in vertebral specimens
While BMD measurements in mid-vertebral specimens derived from conventional qCT ranged between 141 and 144 mg/cm3, estimation of HA density based on spectral data ranged between 134 and 136 mg/cm3, scanned at 50 mAs or more (Table 3). All spectral estimations of HA density acquired with an exposure of 50 mAs or more were highly correlated with those estimated using qCT (r > 0.96 for all; Fig. 1). No statistical significant differences were found between the spectral and qCT measurements from 50 mAs to 500 mAs (p > 0.02), however, at 10 mAs significant differences were found between qCT and spectral CT values.
In addition, agreement between BMD results based on spectral CT and qCT were assessed with a Bland-Altman plot (Fig. 2 ). At 50 mAs and above, mean differences were between ±1.96 SD in most cases, indicating a good agreement between both measurements. At 50 mAs and more, measurements from spectral CT were consistently lower than qCT measurements by a small margin (mean differences, 6.7–8.6 mg/cm3 for 50 mAs and above; Table 3).
When comparing BMD results from spectral CT acquired with two different setups simulating two different degrees of obesity, no significant differences were found (Table 4, p > 0.02).
In linear regression models with the simulated degree of obesity as an additional independent variable, significant correlations were found between qCT- and spectral CT-based BMD measurements for all exposure levels, with high correlation coefficients for 50–500 mAs (Betas; 500 mAs: 0.99; 250 mAs: 0.99; 125 mAs: 0.97; 50 mAs: 0.98) and only a moderate correlation for 10 mAs (Beta, 0.27). For 50 mAs, a small but significant association between the degree of obesity and the BMD measurements based on spectral CT were found (Beta, 0.08; p < 0.001). For all other exposures, the degree of obesity was not significantly associated with the measurements (p > 0.001).
Discussion
In our study, HA-specific density measurements based on spectral data obtained with a dual-layer spectral CT showed high accuracy when performed on a phantom with different known HA concentrations in scans with radiation exposures comparable to clinical protocols. Moreover, phantomless HA-specific spectral BMD measurements showed high correlations and good agreement with conventional qCT-based measurements, which are the current standard of reference for assessing volumetric BMD. To our best knowledge, this is the first time the potential value of phantomless BMD measurements based on dual-layer spectral CT scans has been assessed.
While osteoporosis is associated with a substantial morbidity and mortality, the condition is underdiagnosed and thus undertreated1,2,5,6,31,32. Therefore, patients would highly benefit from a reliable opportunistic BMD assessment and thus fracture risk assessment with non-dedicated imaging data.
Several approaches for opportunistic measurements based on CT data have been investigated recently, such as obtaining BMD information from contrast-enhanced routine abdominal examinations17,33 and CT colonography18 as well as non-contrast enhanced or post-myelography CT examinations16, and BMD-equivalent values derived from those examinations have shown to highly correlate with values obtained from phantom-calibrated qCT examinations. However, there are limitations to these opportunistic measurements: First, patient size potentially affects BMD values34. A larger abdomen volume is able to alter the x-ray spectrum received at the detector, increasing the possibility of beam hardening. The absorption of bone mineral is energy-dependent and thus when estimating BMD from the full spectrum, a HA calibration phantom should be used which is often not placed in the scanner couch during routine scans. Second, it has been shown that the application of intravenous contrast agents substantially affects BMD measurements, presumably due to iodine altering attenuation in measurement ROIs within the vertebrae23,24. It can be assumed, though, that a spectral-based HA-specific measurement is more robust against a possible bias by other materials affecting attenuation. Material decomposition based on spectral information is already commercially available, e.g. iodine concentration can be quantified and used for the imaging of myocardial perfusion or pulmonary embolism35. However, while previous studies reported BMD quantification in cadaveric specimen and patients based on dual-source spectral CT examinations to be feasible29,36, material decomposition and thus HA-specific quantification were not considered in these previous studies. This is noteworthy since it has been shown that marrow adipose tissue may add a substantial bias to CT-based trabecular BMD measurements37.
Monoenergetic images can be generated not only based on dual-layer spectral CT, but also based on other DECT approaches such as set-ups generating two different x-ray spectra38. However, such systems require pre-specified dual-energy scan protocols using both x-ray spectra. In addition, in case of dual-source CT, two x-ray sources generate scatter radiation on each other, which needs to be corrected for, as shown previously39. Moreover, due to the compact design and the perpendicular set-up of systems with dual x-ray sources, one of the energies has a smaller field of view, which limits spectral analysis and image quality for larger patients. Finally, spectral information and thus material decomposition can only be post-computed in image space — in contrast to dual-layer spectral CT, which allows for direct material decomposition in projection space — which reduces the accuracy of the material decomposition and the effectiveness of beam-hardening corrections.
Just recently van Hamersvelt et al. investigated BMD quantification in phantoms with known HA-concentrations based on examinations with a dual-layer spectral CT system identical to the one used in this study, describing measurements to be highly accurate30. In their study, the authors postulated exact concentrations of water and HA within resin phantoms and applied attenuation profiles from the NIST database and estimated BMD values. However, fitting tabulated database values may be inaccurate for several reasons: Firstly, in a setup with a clinical scanner, virtual monoenergetic are not equivalent and only approximations to actual monoenergetic measurements. Secondly, the specific detector response and the spectral decomposition algorithm need to be taken into account, as they may differ from system to system and particularly if reference values were transferred from a different DECT approach. In contrast to the mentioned study using NIST database attenuation profiles, we modelled the full imaging chain for our specific setup by acquiring calibration measurements using a phantom with known HA concentrations. Using this approach, we were able to create scanner-specific attenuation profiles, which enabled us to overcome software and hardware inconsistencies, assuming there is proprietary information about the system used in this study that is unknown. In the study of van Hamersvelt et al. measurement errors ranged between 0.5 and 5.2% for 100–200 mAs, which is comparable to the errors of 0.0 to 4.8% we observed in our analyses for exposures of 125 mAs (a tube exposure typically used in clinical protocols) and above. Moreover, this is in accordance with the study by Hofmann et al. presenting a three-material decomposition method for BMD quantification on DECT, finding a mean measurement error of about 3.5% over all HA concentrations38.
Interestingly, Hofmann et al. found that measurements of known HA densities obtained from qCT overestimated actual known densities, which also is in accordance with our findings that qCT values were slightly higher than values from spectral CT. As discussed before, this may indicate that HA-specific measurements based on spectral analyses may be more accurate than qCT measurements30,38.
Measurement errors at 50 and 38 mAs were more substantial in lower HA concentrations (14.8% for 100 mg/cm3 as assessed with 38 mAs). This may be caused by the fact that a lowered tube current decreases the number of photons emitted, thus influencing the final number of photons reaching the layers of the detector and ultimately the accuracy of spectral signal and BMD measurements. This is a known limitation of all dual-layer detectors and should be taken into account when assessing the clinical value of dedicated ultra-low-dose CT protocols. Other means of dose reduction such as sparse sampling should be considered40.
BMD measurements in cadaveric vertebral specimens specific for HA based on spectral CT imaging showed high correlations and agreement with measurements obtained using phantom-calibrated qCT as standard of reference. Correlations were significant based on scans with 50–500 mAs, with correlation coefficients increasing with increasing mAs. This suggests that HA-specific BMD quantification is valid in vertebral specimens, and should also be feasible in-vivo.
Note, when comparing BMD measurements in vertebral specimens based on spectral CT in setups simulating two different degrees of obesity, we did not find any significant differences. Moreover, in linear regressions, no significant associations were found between degrees of obesity and BMD based on spectral CT, which may be particularly beneficial to obese patients, in which therapeutic decisions should be based on accurate measurements as well. It is noteworthy since DXA as the current clinical standard and qCT have been reported to be significantly affected by body composition and mAss8,11,14,15, and it suggests that spectral CT may be more robust to such external confounders.
Our study has limitations. First, we used monoenergetic images at 50 and 200 keV to assess HA characteristics and concentrations. However, in a dual-layer setup, we can assume that two monoenergetic levels can represent the complete spectrum information obtained in the scanner. Second, we assessed both known HA concentrations in phantoms and BMD in vertebral specimens in standardized setups, including the two different degrees of obesity simulated by an anthropomorphic phantom with two different extension rings; however, other factors that have been described to affect the BMD quantification such as different tube voltages, the use of intravenous contrast medium, the fat fraction within vertebral bone marrow were not taken into account in this setup20,21,22,23,24. Therefore, further studies are necessary assessing clinical patient scans with different scan protocols. Finally, plain thresholding was applied to define HA-containing pixels, with a 20 HU threshold. Since in this study in both, the setup with phantoms with known HA concentrations as well as the setup with vertebral specimens, only HA and water were present in the ROIs, therefore different threshold values (assessed for 20 to 100 HU) did not measurably affect final BMD values. However, in future in-vivo studies additional materials such as iodine may be present and need to be taken into account and different thresholding techniques may be needed in order to reliably identify HA.
In summary, we have shown that known HA concentrations can be quantified accurately by using spectral CT imaging information obtained with a dual-layer spectral CT scanner. In addition, high correlations and agreement between HA-specific BMD measurements in vertebral specimens and phantom-calibrated qCT measurements in the same specimens in setups simulating different degrees of obesity were found. This finding suggests that phantomless BMD quantification based on dual-layer spectral CT imaging is feasible and may be applied in clinical studies.
Methods
Specimens and phantoms
Institutional review board approval was obtained prior to this study (Ethikkommission der medizinischen Fakultät, Technical University of Munich, Germany).
Four cylindrical phantoms (QRM) with known calcium hydroxyapatite (HA) concentrations were examined in an anthropomorphic phantom (Anthropomorphic Abdomen Phantom, QRM, Möhrendorf, Germany; Fig. 3). The exact HA concentrations were 809, 405, 209 and 100 mg/ml. The composition of the phantoms was CT water (0 HU, 80–120 kV) and HA. The base materials of the abdomen phantom were tissue equivalent resin (about 35 to 50 HU at 120 keV) imitating muscle and liver, etc. The axial diameter was 300 × 200 mm.
In a second setup, 13 mid-vertebral specimens harvested from human donors without history of pathological bone changes other than osteoporosis were examined in this study. The donors had dedicated their bodies for educational and research purposes to our institution and gave their informed consent prior to death, in compliance with local institutional and legislative requirements. All experiments were performed in accordance with relevant institutional and legislative guidelines and regulations.
Specimens consisted of mid-vertebral, 10-mm-thick axial slice of a thoracic vertebra between Th5 and Th12. Specimens were preserved in formalin, after complete removal of surrounding soft tissue. Before the scan, specimens were immersed in water bath and air inside the trabecular was eliminated using a vacuum machine. Specimens were then placed in the same anthropomorphic abdomen phantom analogous to their physiologic orientation in the human body (Fig. 4). Successively, two extension rings simulating fat were placed around the phantom, with the smaller one increasing the diameters by 350 × 250 mm, and the larger one by 400 × 300 mm, simulating patients with waist circumferences 98.5 cm and 114.3 cm, respectively.
CT imaging protocols
Spectral CT images were acquired by using a dual-layer spectral CT scanner (IQon Spectral CT, Philips Healthcare, Best, the Netherlands).
An abdomen protocol was used for all examinations, with a fixed tube voltage of 120 kV and a pitch of 1.058. HA phantom scans were performed with different exposures of 1000, 500, 250, 125, 50 and 38 mAs; mid-vertebrae specimens were scanned with 500, 250, 125, 50 and 10 mAs. For scans with 500 mAs, tube current was 700 mA and exposure time was 709 ms. For the other exposure levels, exposure time remained the same while tube current was adjusted accordingly.
All vertebral specimens scans were performed twice, with the use of two different extension rings of the anthropomorphic simulating two degrees of obesity.
CT Dose Indices (CTDIvol) were recorded from the dose report generated by the scanner. Scan lengths were 50 mm per mid-vertebral specimen. Estimated effective doses (ED) were calculated directly from Dose Length Product using the k factor approach41. Signal- and Contrast-to-Noise-Ratio (SNR/CNR) were computed using mean and standard deviation in the homogeneous regions of the qCT calibration phantom.
We used the same reconstruction protocol for both qCT and spectral CT based BMD quantifications. Reconstruction kernel was spectral with the level set at 4, the slice thickness was 0.67 mm. The field of view was 200 mm by 200 mm with a grid size of 512 by 512 pixels. HA phantom scans at 1000 mAs were used to determine the spectral absorption behaviour of HA. The remaining scans were used to validate the BMD quantification method based on spectral information.
BMD quantification based on spectral imaging data
For determining the spectral absorption behaviour of HA, the HA phantoms with four different known HU concentrations were scanned at 1000 mAs. Theoretically higher radiation exposure increases the number of photons, reducing image noise without changing the HA absorption nature in the spectrum. Virtual monoenergetic images at 50 (HU L ) and 200 keV (HU H ) were reconstructed. ROIs were manually drawn by a radiologist in the four regions in each phantom (Fig. 3). ROIs were cylinders with a diameter of 5 mm and a height of 33 mm. CT numbers (Hounsfield Units (HU)) inside the ROIs were extracted. A line function depicting the relationship with two monoenergetic HA HUs was fitted:
HU L and HU H are CT numbers for HA measured using monoenergetic 50 and 200 keV images. a is a fitting parameter specific for HA. b is a fitting parameter adjusting for water. In this set-up, pixels in the ROIs only contain HA and water, therefore pixels in the ROIs (pairs of HU L and HU H for HA) are always located along or close to the line described by equation (1) (blue line in Fig. 5). The higher the HU L and HU H are, the further of the points on the line to the origin are, the denser the HA concentration is. HU L is proportional to the exact concentration of HA (BMD):
The fitting parameters u, v were obtained from known BMD values of the four HA phantoms and the corresponding measured mean CT numbers in the monoenergetic image at 50 keV derived from the calibration scan. u is determined by the mAss attenuation coefficient of HA at monoenergetic 50 keV. v is the CT number determined by water. a, b, u and v are scanner specific variables computed once prior to specimen scans.
Next, the calibration lines from equation (1) and equation (2) were used to identify and quantitate BMD values in virtual spectral images. Scans with lower exposures (500, 250, 125, 50 and 38 mAs) were included: a measurement of two CT numbers (HU′ H , HU′ L ) at 50 and 200 keV were extracted from the corresponding monoenergetic images. The distance d from this measurement (HU′ H , HU′ L ) to the line as specified by equation (1) was computed:
We applied plain thresholding (d) to identify HA. A measurement was considered bone if it is closer to the regression line as defined in equation (1), i.e. d was less than 20 HU. If d was greater than 20 HU, the measurement was likely to represent non-HA material, such as soft tissue. For each measurement of (HU′L, HU′H) that was considered to represent bone, its BMD value was estimated. In order to minimize possible measurement errors, each measurement was projected to the HUL/HUH regression line as specified in equation (1). For this, the closest point on the line described with equation (1) (HUL, HUH) to the actual measurement (HU′L, HU′H) was determined with the following equation:
The resulting HUL was finally converted to a BMD value with calibrated equation (2). This procedure was performed pixel-by-pixel in a given ROI, and resulting BMDs were averaged.
Analogously, equations (3) and (4) were used to determine HA content and equation (4) to quantitate BMD values for vertebral specimens.
BMD quantification in vertebral specimens based on qCT
ROIs were placed manually by a radiologist in the centre of vertebral specimens. These ROI were 2D circles with diameter 5.9 mm or 7.8 mm, depending on the vertebra size, and were placed on the 3 slices in the middle of the vertebrae (Fig. 4).
For comparison purpose, BMD was also calculated with the conventional qCT method (QCT Pro Bone Mineral Densitometry Software, Phantom Module. Version 4.0, Mindways, TX, USA). For these scans, a calibration phantom (Mindways) was placed beneath the anthropomorphic phantom containing 5 rods with known density equivalents for phosphate and water. For each scan, mean values of the five rods in the QCT phantom were measured in the conventional images generated by the dual-layer spectral scanner. The ROI selection was similar to the one described in the previous section (Fig. 4). Conversion functions for the scanner were computed in a least square manner with MATLAB.
Statistical analysis
To describe measurement errors for the quantification of known HA concentrations based on spectral information, descriptive statistics were used containing absolute values and relative values compared to the known concentration as given by the manufacturer of the phantom. Descriptive statistics were also used for describing BMD measurements in vertebral specimens both spectral-based and using qCT with different exposures and different setups simulating different degrees of obesity. Normality of BMD measurement values was confirmed by Shapiro-Wilk tests.
To assess correlations between qCT and spectral CT measurements, both Pearson and Spearman correlations were used. Since Pearson’s r was consistently higher than Spearman’s ρ, linear correlations were assumed. Agreement between BMD measurements based on spectral CT and qCT was assessed with Bland-Altman plots42, and paired-samples t-tests were used to assess values for statistically significant differences.
In order to assess the influence of different degrees of simulated obesity by using extension rings around the abdomen phantom, multivariable linear regression models were performed with BMD based on spectral data as dependent and BMD based on qCT and simulated degrees of obesity as independent variables, separately for different mAs. Differences between values from two setups simulating two different degrees of obesity were compared by using paired-samples t-tests.
Statistical analyses were performed with SPSS 24 (IBM; Armonk, NY, USA), with two-sided P < 0.05 indicating a significant difference.
References
Haussler, B. et al. Epidemiology, treatment and costs of osteoporosis in Germany–the BoneEVA Study. Osteoporos Int 18, 77–84, https://doi.org/10.1007/s00198-006-0206-y (2007).
Hernlund, E. et al. Osteoporosis in the European Union: medical management, epidemiology and economic burden. A report prepared in collaboration with the International Osteoporosis Foundation (IOF) and the European Federation of Pharmaceutical Industry Associations (EFPIA). Arch Osteoporos 8, 136, https://doi.org/10.1007/s11657-013-0136-1 (2013).
Bredella, M. A. et al. Determinants of bone mineral density in obese premenopausal women. Bone 48, 748–754, https://doi.org/10.1016/j.bone.2010.12.011 (2011).
Rizzoli, R. et al. Cancer-associated bone disease. Osteoporos Int 24, 2929–2953, https://doi.org/10.1007/s00198-013-2530-3 (2013).
Curtis, J. R. et al. Longitudinal trends in use of bone mAss measurement among older americans, 1999–2005. J Bone Miner Res 23, 1061–1067, https://doi.org/10.1359/jbmr.080232 (2008).
Zhang, J. et al. Central DXA utilization shifts from office-based to hospital-based settings among medicare beneficiaries in the wake of reimbursement changes. J Bone Miner Res 27, 858–864, https://doi.org/10.1002/jbmr.1534 (2012).
Buckens, C. F. et al. Osteoporosis markers on low-dose lung cancer screening chest computed tomography scans predict all-cause mortality. Eur Radiol 25, 132–139, https://doi.org/10.1007/s00330-014-3361-0 (2015).
Cosman, F. et al. Clinician’s Guide to Prevention and Treatment of Osteoporosis. Osteoporos Int 25, 2359–2381, https://doi.org/10.1007/s00198-014-2794-2 (2014).
Kanis, J. A. et al. Intervention thresholds for osteoporosis in men and women: a study based on data from Sweden. Osteoporos Int 16, 6–14, https://doi.org/10.1007/s00198-004-1623-4 (2005).
Schuit, S. C. et al. Fracture incidence and association with bone mineral density in elderly men and women: the Rotterdam Study. Bone 34, 195–202 (2004).
Yu, W. et al. Spinal bone mineral assessment in postmenopausal women: a comparison between dual X-ray absorptiometry and quantitative computed tomography. Osteoporos Int 5, 433–439 (1995).
Kopperdahl, D. L. et al. Assessment of incident spine and hip fractures in women and men using finite element analysis of CT scans. J Bone Miner Res 29, 570–580, https://doi.org/10.1002/jbmr.2069 (2014).
Dall’Ara, E., Pahr, D., Varga, P., Kainberger, F. & Zysset, P. QCT-based finite element models predict human vertebral strength in vitro significantly better than simulated DEXA. Osteoporos Int 23, 563–572, https://doi.org/10.1007/s00198-011-1568-3 (2012).
Oei, L., Koromani, F., Rivadeneira, F., Zillikens, M. C. & Oei, E. H. Quantitative imaging methods in osteoporosis. Quant Imaging Med Surg 6, 680–698, https://doi.org/10.21037/qims.2016.12.13 (2016).
Link, T. M. Osteoporosis imaging: state of the art and advanced imaging. Radiology 263, 3–17, https://doi.org/10.1148/radiol.12110462 (2012).
Schwaiger, B. J. et al. Bone mineral density values derived from routine lumbar spine multidetector row CT predict osteoporotic vertebral fractures and screw loosening. AJNR Am J Neuroradiol 35, 1628–1633, https://doi.org/10.3174/ajnr.A3893 (2014).
Baum, T. et al. Converted lumbar BMD values derived from sagittal reformations of contrast-enhanced MDCT predict incidental osteoporotic vertebral fractures. Calcif Tissue Int 90, 481–487, https://doi.org/10.1007/s00223-012-9596-3 (2012).
Fidler, J. L. et al. Comprehensive Assessment of Osteoporosis and Bone Fragility with CT Colonography. Radiology 278, 172–180, https://doi.org/10.1148/radiol.2015141984 (2016).
Schwaiger, B. J. et al. Vertebral and femoral bone mineral density and bone strength in prostate cancer patients assessed in phantomless PET/CT examinations. Bone 101, 62–69, https://doi.org/10.1016/j.bone.2017.04.008 (2017).
Engelke, K. et al. Clinical Use of Quantitative Computed Tomography-Based Advanced Techniques in the Management of Osteoporosis in Adults: the 2015 ISCD Official Positions-Part III. J Clin Densitom 18, 393–407, https://doi.org/10.1016/j.jocd.2015.06.010 (2015).
Garner, H. W., Paturzo, M. M., Gaudier, G., Pickhardt, P. J. & Wessell, D. E. Variation in Attenuation in L1 Trabecular Bone at Different Tube Voltages: Caution Is Warranted When Screening for Osteoporosis With the Use of Opportunistic CT. AJR Am J Roentgenol 208, 165–170, https://doi.org/10.2214/AJR.16.16744 (2017).
Goodsitt, M. M. & Rosenthal, D. I. Quantitative computed tomography scanning for measurement of bone and bone marrow fat content. A comparison of single- and dual-energy techniques using a solid synthetic phantom. Invest Radiol 22, 799–810 (1987).
Pompe, E. et al. Intravenous contrast injection significantly affects bone mineral density measured on CT. Eur Radiol 25, 283–289, https://doi.org/10.1007/s00330-014-3408-2 (2015).
Toelly, A. et al. Influence of Contrast Media on Bone Mineral Density (BMD) Measurements from Routine Contrast-Enhanced MDCT Datasets using a Phantom-less BMD Measurement Tool. Rofo, https://doi.org/10.1055/s-0043-102941 (2017).
Vetter, J. R., Perman, W. H., Kalender, W. A., Mazess, R. B. & Holden, J. E. Evaluation of a prototype dual-energy computed tomographic apparatus. II. Determination of vertebral bone mineral content. Med Phys 13, 340–343, https://doi.org/10.1118/1.595951 (1986).
Kalender, W. A., Perman, W. H., Vetter, J. R. & Klotz, E. Evaluation of a prototype dual-energy computed tomographic apparatus. I. Phantom studies. Med Phys 13, 334–339, https://doi.org/10.1118/1.595958 (1986).
Mallinson, P. I. et al. Dual-Energy CT for the Musculoskeletal System. Radiology 281, 690–707, https://doi.org/10.1148/radiol.2016151109 (2016).
Altman, A. & Carmi, R. TU-E-210A-03: A Double-Layer Detector, Dual- Energy CT—Principles, Advantages and Applications. Medical Physics 36, 2750–2750 (2009).
Wichmann, J. L. et al. Dual-energy CT-based phantomless in vivo three-dimensional bone mineral density assessment of the lumbar spine. Radiology 271, 778–784, https://doi.org/10.1148/radiol.13131952 (2014).
van Hamersvelt, R. W. et al. Accuracy of bone mineral density quantification using dual-layer spectral detector CT: a phantom study. Eur Radiol, https://doi.org/10.1007/s00330-017-4801-4 (2017).
Ballane, G., Cauley, J. A., Luckey, M. M. & El-Hajj Fuleihan, G. Worldwide prevalence and incidence of osteoporotic vertebral fractures. Osteoporos Int, https://doi.org/10.1007/s00198-017-3909-3 (2017).
Flais, J. et al. Low prevalence of osteoporosis treatment in patients with recurrent major osteoporotic fracture. Arch Osteoporos 12, 24, https://doi.org/10.1007/s11657-017-0317-4 (2017).
Mueller, D. K. et al. Phantom-less QCT BMD system as screening tool for osteoporosis without additional radiation. Eur J Radiol 79, 375–381, https://doi.org/10.1016/j.ejrad.2010.02.008 (2011).
Yu, E. W., Thomas, B. J., Brown, J. K. & Finkelstein, J. S. Simulated increases in body fat and errors in bone mineral density measurements by DXA and QCT. J Bone Miner Res 27, 119–124, https://doi.org/10.1002/jbmr.506 (2012).
Rajiah, P., Abbara, S. & Halliburton, S. S. Spectral detector CT for cardiovascular applications. Diagn Interv Radiol 23, 187–193, https://doi.org/10.5152/dir.2016.16255 (2017).
Wichmann, J. L. et al. Quantitative dual-energy CT for phantomless evaluation of cancellous bone mineral density of the vertebral pedicle: correlation with pedicle screw pull-out strength. Eur Radiol 25, 1714–1720, https://doi.org/10.1007/s00330-014-3529-7 (2015).
Bredella, M. A. et al. Marrow Adipose Tissue Quantification of the Lumbar Spine by Using Dual-Energy CT and Single-Voxel (1)H MR Spectroscopy: A Feasibility Study. Radiology 277, 230–235, https://doi.org/10.1148/radiol.2015142876 (2015).
Hofmann, P. et al. Phantom-less bone mineral density (BMD) measurement using dual energy computed tomography-based 3-material decomposition. in Proc. SPIE 9785, Medical Imaging 2016: Computer-Aided Diagnosis. (eds GD Tourassi & SG Armato)(International Society for Optics and Photonics).
Petersilka, M., Stierstorfer, K., Bruder, H. & Flohr, T. Strategies for scatter correction in dual source CT. Med Phys 37, 5971–5992, https://doi.org/10.1118/1.3504606 (2010).
Mei, K. et al. [Accepted] Is multidetector CT-based bone mineral density and quantitative bone microstructure assessment at the spine still feasible using ultra-low tube current and sparse sampling? Eur Radiol, https://doi.org/10.1007/s00330-017-4904-y (2017).
McCollough, C. H. CT dose: how to measure, how to reduce. Health Phys 95, 508–517, https://doi.org/10.1097/01.HP.0000326343.35884.03 (2008).
Bland, J. M. & Altman, D. G. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1, 307–310 (1986).
Acknowledgements
We acknowledge support through the European Research Council (ERC-StG-2014 637164), the DFG (DFG BA4085/2–1 and DFG BA 4906/1–1 and as part of the Gottfried Wilhelm Leibniz program), by the German Department of Education and Research (BMBF) under grant IMEDO (13GW0072C) and the support of the TUM Institute for Advanced Study, funded by the German Excellence Initiative. Furthermore, this work was supported by the German Research Foundation (DFG) and the Technical University of Munich (TUM) within the Open Access Publishing Program.
Author information
Authors and Affiliations
Contributions
Literature research: All authors. Conception and design of the study: K.M., B.J.S., A.S.G., J.S.K., E.J.R., F.P., T.B., P.B.N. Experimental studies: K.M., B.J.S., F.K.K., S.E., P.B.N. Statistical analysis: K.M., B.J.S., A.S.G., T.B. Manuscript editing: All authors. Guarantors of integrity of entire study: K.M., B.J.S., P.B.N. All authors reviewed and approved of the final version of this manuscript.
Corresponding author
Ethics declarations
Competing Interests
The authors declare that they have no competing interests.
Additional information
Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Mei, K., Schwaiger, B.J., Kopp, F.K. et al. Bone mineral density measurements in vertebral specimens and phantoms using dual-layer spectral computed tomography. Sci Rep 7, 17519 (2017). https://doi.org/10.1038/s41598-017-17855-4
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-017-17855-4
- Springer Nature Limited
This article is cited by
-
Comparison of volumetric and areal bone mineral density in CT and scout scans using spectral detector technology
European Radiology Experimental (2023)
-
The impact of imaging time and contrast agent dose on screening for osteoporosis with contrast-enhanced CT
European Radiology Experimental (2022)
-
Opportunistic osteoporosis screening: contrast-enhanced dual-layer spectral CT provides accurate measurements of vertebral bone mineral density
European Radiology (2021)
-
Quantitative imaging of the spine in adolescent idiopathic scoliosis: shifting the paradigm from diagnostic to comprehensive prognostic evaluation
European Journal of Orthopaedic Surgery & Traumatology (2021)
-
Accuracy and precision of volumetric bone mineral density assessment using dual-source dual-energy versus quantitative CT: a phantom study
European Radiology Experimental (2021)