Abstract
A new slim modiolar electrode (CI532/632) has been reported to ensure better modiolar proximity than conventional electrodes. Better modiolar proximity has been proposed to yield better electrode discrimination capability and potentially better speech outcomes, necessitating its efficient measurement. Currently, intracochlear positional index (ICPI), the most reliable indicator for evaluating modiolar proximity, has been measured exclusively through ‘metal artifact-less’ cone beam CT. However, popular use of this index is precluded due to lack of cone beam CT in many institutions. Thus, eyes are now on elucidation of easy-to-measure indicators of modiolar proximity derived from conventional CT, which is accessible in all centers. We observed that enhanced tomographic resolution significantly reduces partial volume artifacts, providing better visualization of modiolus-electrode distance. Aided by ultra-high kernel specification with high-resolution index, we developed a novel and easy-to-measure, conventional CT-specific indicator, “modified ICPI”, for evaluation of modiolar proximity. Further, we showed that it closely correlates with the previously proposed parameter of modiolar proximity, the spiral diameter, measured from post-insertion radiograph, reiterating the value of X-ray-based spiral diameter. Through this study, we have taken a step toward the stage of immediate visual feedback regarding modiolar proximity and changes in insertion technique intraoperatively, ensuring optimal modiolar proximity.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Introduction
The new slim modiolar electrodes (e.g., CI532 or CI632) combine the slim electrode diameter of slim-straight electrodes and the modiolus-hugging feature of conventional perimodiolar electrodes. This new commercially available slim modiolar electrode ensures better modiolar proximity1,2 and provides substantial preservation of residual hearing3,4,5,6,7,8,9,10,11. For modiolar hugging electrodes, better auditory performance depends on the final intracochlear positioning of the electrode array, such as scalar location and modiolar proximity12,13. The slim modiolar electrodes have been reported to provide a consistently higher scalar tympani position than conventional perimodiolar electrodes6,14. In addition, 4.1–4.6% rate of tip-fold over has been reported in the literature9,14, but tend to gradually decrease with experience15. Thus, specifically for slim modiolar electrodes, degree of modiolar proximity may be one of the critical determinants of better auditory performance. Recently, cochlear implant recipients with slim modiolar electrodes (i.e., CI532/632) led to either similar or statistically superior audiological results compared with those with straight electrodes (i.e., CI422/522) strictly matched for age and preoperative hearing thresholds2. Furthermore, a growing body of evidence suggests that the enhanced modiolar proximity of slim modiolar electrodes would lead to improved place-pitch spectral discrimination16, and speech perception outcomes14,17, as compared with conventional perimodiolar electrodes. However, not all reports in literature fully support this phenomenon9,18,19, probably due to confounding variables such as heterogeneous degrees of modiolar proximity even with slim modiolar electrodes20. Indeed, better modiolar proximity has been proposed to attenuate the substantial overlap (i.e., spread of excitation) in the electrical field between electrode contacts, resulting in better electrode discrimination capability16,21, and potentially better speech outcomes than conventional electrodes21. In addition, positioning of the electrodes close to the spiral ganglion neurons can reduce the effects of neural-electrode interaction by improving channel discrimination16, lowering threshold levels22, and reduction of power consumption23. Therefore, robust and efficient measurement of modiolar proximity is mandatory for slim modiolar electrodes which were designed to achieve the best modiolar proximity amongst all electrodes.
To date, there are different approaches to measure the electrode position relative to the modiolar wall distance. One of the widely accepted measurements is the wrapping factor to determine how the electrode array is located relative to the lateral wall24. The wrapping factor has been suggested to be inversely correlated with the degree of modiolar proximity. However, the wrapping factor has been reported to yield an inconsistency to determine the relative relationship of the electrode array against the lateral wall between the cone-beam CT and the histologic finding, requiring a more accurate and refined radiological value for modiolar proximity25. Recently, intracochlear positioning of the electrode array in close proximity to the modiolus—with good modiolar hugging, as evidenced by the wrapping factor normalized or intracochlear positioning index (ICPI)25—would lead to better hearing outcomes for cochlear implantees with the slim modiolar electrodes.
With the advent of cone-beam CT based on flat-panel volumetric technology, significantly better image quality enables clear delineation of the intracochlear electrode array with more definition of the fine osseous structures surrounding the temporal bone26,27. Although a previous ex vivo study showed that the average electrode-to-modiolus distance value measured by two imaging modalities (conventional CT vs. cone-beam CT) was not different28, measurement of electrode-to-modiolus distance using conventional CT is not feasible due to metal artifacts blurring the region of interest. A recent study proposed the most reliable and finely tuned indicator reflecting the degree of modiolar proximity, called ICPI25, which was obtained exclusively through ‘metal artifact-less’ cone-beam CT. However, lack of such cone-beam CT in many institutions preclude widespread use of this index. Thus, eyes are now on the elucidation of easy-to-measure indicators of modiolar proximity derived from conventional CT that is accessible to all cochlear implantation centers.
Herein, we observed that higher tomographic resolution significantly attenuates imaging blurring and enhances imaging sharpness, thereby enabling elaborate measurement of electrode-to-modiolus distance even in conventional CT. Aided by ultra-high kernel specification with a high-resolution index, we came up with a novel and easy-to-measure, conventional CT-specific indicator, modified ICPI, for evaluation of modiolar proximity. Further, we evaluated whether or not modified ICPI obtained from conventional CT correlates with the previously proposed parameter of modiolar proximity, the spiral diameter, measured from a post-insertion radiograph during or after the operation. If that is the case, simple X-ray would eventually be a nice substitute for conventional CT in terms of obtaining modiolar proximity and comparison of it just for practice within a single center. This should be the case especially for pediatric patients, where radiation matters. However, the role of ICPI obtained from conventional CT does not seem to swiftly wane, especially for comparison of the parameter across centers, especially for adults. Simple X-ray technique is not standardized across centers, precluding comparison of modiolar proximity values across centers. Therefore, CT measurement should not be underestimated as just an interim approach, but it would better regard the use of CT as a valuable independent approach per se. Collectively, these results offer otologists a novel and easy-to-calculate indicator for modiolar proximity through conventional CT and reiterates the value of X-ray-based measurement of the spiral diameter, enabling immediate visual feedback and changes in insertion technique intraoperatively, thereby ensuring optimal modiolar proximity.
Methods
Participants
We retrospectively reviewed cochlear implant recipients in whom slim modiolar electrodes, such as the CI532 or CI632, were implanted by a single surgeon (B.Y.C) using the round window approach, exclusively with the pull-back technique. Recently, the insertion technique that ensures better modiolar proximity, called the pull-back maneuver, has been introduced for slim modiolar electrodes, based on a human cadaveric temporal bone study29. Only patients with 0.4 mm slice thickness on temporal bone CT and a high-resolution filter specification were included. Subjects with the following conditions were also excluded from this study: (1) history of explantation or reimplantation, (2) severe cochlear ossification, and (3) obvious cochlear anomalies on radiological images based on the classification of inner ear anomalies. Ultimately, a total of 30 recipients (33 ears) of CI532 (N = 15, 16 ears) or CI632 (N = 16, 17 ears) cochlear implants were enrolled. In our current study, two of 33 ears were implanted using the extended pull-back maneuver20, due to short cochlear duct length measured preoperatively. The study protocol and a waiver of consent for this retrospective chart review were approved by the review board of the Clinical Research Institute at Seoul National Bundang Hospital (approval no. IRB-B-2004/604-119). All methods employed in this study were in accordance with the approved guidelines and the Declaration of Helsinki.
CT protocol
High-resolution temporal bone CT scans were performed on the day after the surgery, particularly in adult cochlear implant recipients, whenever available. Axial images were obtained with 0.4 mm slice thickness using 120 kV, 64 × 0.6 mm collimation, 1-s rotation, pitch factor of 0.85, and 205 mAs, in accordance with the age of the subjects by 256-channel multi-detector computed tomography (SOMATOM Force, Siemens Healthineers, Forchheim, Germany)30. All data were reconstructed using two sets of ultra-high resolution kernels with different resolution indices (Uh59, resolution index of 8.3 line-pair (lp)/cm; Ur81, 19.7 lp/cm). Importantly, no difference in effective dose was found between the two different filters (Uh59 vs. Ur81), as they are based on the difference in the reconstruction algorithm of the raw data obtained from the imaging scans, regardless of the scan parameters such as kVp and mAs in the conventional temporal bone CT.
Quantitative and qualitative assessments
As shown in Fig. 1a, high-resolution temporal bone CT images were reformatted to create the “Cochlear View,” following multiplanar reconstruction (MPR) algorithm31,32. A board-certified neuroradiologist (Y.J.B, with 10 years of experience) and an otologic surgeon (B.Y.C) independently performed visual inspection of the “Cochlear View” on the conventional CT with ultra-high kernel specification of Uh59 (low resolution index) and Ur81 (high resolution index). As modified from the version of Kidoh et al.33, the degree of image sharpness under the bone window setting (window level and width of 300 and 4000 Hounsfield units, respectively,) was assessed using a 4-point visual score, and the overall subjective image quality was separately examined according to a 5-point visual score (Table 1). The inter-class correlation was excellent with regard to imaging sharpness and overall subjective image qualities (Table S1).
As shown in Fig. 1b, a neuroradiologist (Y.J.B) obtained an axial scan to compare the noise level between the two different resolution indexes. The reader allocated the region of interest (ROI) in the slice where the middle turn of the cochlea was clearly visible. The noise level was defined as the standard deviation of the Hounsfield units in each ROI of the structures. Values were measured twice on the same image, and the average of the two values was used for further analysis.
Intracochlear position index
With reference to a recent study by Miguel et al.25, ICPI(i) refers to the ratio of the Euclidean distance between the modiolus (M) and the electrode (Ei) relative to the distance between the modiolus and the lateral wall (LWi). Specifically, the modiolus (M) value was determined as a crossing point in relation to the two lines associated with the modiolus and the round window. In this study, we defined the “modified ICPI,” a novel parameter that reflects the degree of modiolar proximity. The modified ICPI is the average value of ICPI(i) measured at four fixed positions under a two-axis crossing modiolus (Fig. 2a). As depicted in Fig. 2b, ICPI(i) and modified ICPI on the Cochlear view were measured to the nearest 0.01 mm by a neuroradiologist (Y.J.B) blinded to all subject-related information using software on the PACS workstation combined with sharpening and zooming (Infinitt, Seoul, South Korea). Values of ICPI(i) were computed twice on the same image, and the average of the two values was used for further analysis. Also, the angular depth of insertion was measured as a reference for the angle between the round window and electrode tip on the “Cochlear view” reformatted by the multiplanar reconstruction algorithm31.
Spiral diameter
The spiral diameter was measured from transorbital X-ray images taken the day after implantation. As illustrated in Fig. 3a, the spiral diameter of the electrode turn (Fig. 3b) was defined as the distance of the electrode turn measured on a horizontal line across the modiolus.
Intraoperative electrically evoked compound action potential thresholds
After the final intracochlear positioning of the electrode arrays, telemetry recordings were made under sterile conditions in the operating field. “Electrically evoked compound action potential (ECAP) thresholds” were measured in every channel for all subjects using software-based neural response telemetry (NRT) recordings (Cochlear Custom Sound 4.0) with automatic NRT mode. As outlined by guideline, stimulation rate (Hz), maximum current level (CL), and the number of sweeps were set as 250 Hz, 255 CL, and 35, respectively. Furthermore, to evaluate and describe the scalar position of the electrode arrays, ‘NRT ratio’ was proposed previously34,35. The NRT ratio was obtained by dividing the average NRT value from electrodes 18 to 16 in the apical regions by the average NRT value from electrodes 8 to 6 in the basal regions of the electrode array34,35. Specifically, values > 1.05 (i.e., cut-off value) indicate scalar translocation of the inserted electrodes34,35.
Statistical analyses
All statistical analyses were performed using R Statistical Software (R version 3.5.2: Foundation for Statistical Computing, Vienna, Austria) and RStudio (RStudio-1.2.5042, https://www.rstudio.com/). Then, all the analyses illustrated used the GraphPad Prism version 8.0.0 for Windows, GraphPad Software, San Diego, California USA (www.graphpad.com), An independent t-test (two-tailed) was used to compare the quantitative and qualitative assessments between the two different filter specifications (UR59 versus UR81). One-way ANOVA (within-subject design) and Tukey’s post-hoc test was used to determine if the ICPI(i) value at the four points differ from each other. Pearson correlation analyses were performed to identify the relationships between the modified ICPI, related parameters, and spiral diameters, because these parameters were normally distributed. P values < 0.05 were considered statistically significant.
Results
Demographic and clinical characteristics
The demographic and clinical characteristics of the patients are shown in Table 2. All patients were adults, and the mean age at implantation was 56.7 ± 19.5 years (range 18–91 years). No subject in this study exhibited inner ear anomalies based on temporal bone CT scan and/or internal acoustic canal magnetic resonance imaging (MRI). Twenty subjects (66.7%) had idiopathic progressive sensorineural hearing loss. With regard to the definite etiology of the deafness, the causative variants were observed in eight (26.7%) (Table S1), followed by otosclerosis (N = 1, 3.3%) and advanced chronic otitis media (N = 1,3.3%). Based on the high-resolution CT, we ensured the scalar tympani localization of the electrode array in all cases. Consistent with this, the NRT ratios of all patients measured three months after cochlear implantation showed less than the cut-off value (i.e., 1.05), strongly suggesting that the slim modiolar electrode array was placed within the scala tympani as previously suggested34,35. Furthermore, tip rollover of the electrode array was not observed in our cohort.
Qualitative and quantitative image analysis
The average visual scores for imaging sharpness and subjective image quality were significantly higher in the conventional CT with a high-resolution filter than with a standard-resolution filter (P < 0.001 by independent t-test) (Fig. S1). The size of the ROI was fixed as 6.00 ± 2.36 mm2 for the cochlea. The noise level was significantly lower with conventional CT with a high-resolution filter specification than with a standard-resolution filter (P < 0.001 by independent t-test) (Fig. S2).
Definition and characterization of modified intracochlear positional index (modified ICPI)
As depicted in Fig. 4, modified ICPI-related metrics on the Cochlear view could be clearly determined at four fixed positions under a two-axis crossing the modiolus in the conventional CT with a high-resolution filter. Our observed values, including those for modified ICPI and their associated metrics, showed well-preserved normality, following a Gaussian distribution (Table S2). Furthermore, the average angular depth of insertion (i.e., the final position of the electrode tip) was 375.7 ± 14.8 (range 350.3–403.3), revealing very low standard deviation of the angular depth of insertion. Thus, the electrode configuration of the cohorts included in this study was hypothesized to recapitulate the schematic illustration (see Fig. 2a) and now we see that variability of position of slim modiolar electrodes along the array due to differences in cochlear shape and size is not significant.
Under the equation, the average modified ICPI of slim modiolar electrodes implanted by the round window approach, exclusively with the pull-back technique, was 0.44 ± 0.05, ranging from 0.34 to 0.54. The distance from the modiolus to the electrode remained unchanged between point 1 (i.e., electrode No.7-8), point 2 (i.e., electrode No. 12-13), and point 3 (i.e., electrode No. 16-17), while point 4 (i.e., electrode No. 20-21) showed a sudden significant decrease (P < 0.001 by one-way ANOVA and post-hoc Tukey test) (Fig. 4b). On the other hand, the distance from the modiolus to the lateral wall gradually decreased from point 1 to point 4 with statistical significance (Fig. 4c). Overall, ICPI(i) tended to increase from the upper basal turn (i.e., electrode No. 7-8) to the apical turn (i.e., electrode No. 20-21). That is, ICPI (i) at point 1 exhibited the lowest value due to the longest distance from the modiolus to the lateral wall, while relatively maintaining the distance from the modiolus to the electrode (Fig. 4a).
Functional relevancy for our proposed CT measures
Intraoperative ECAP thresholds varied across electrode arrays, demonstrating that the ECAP thresholds tended to gradually increase from the apical to basal cochlear region (Fig. S3). By definition, ICPI values indicate the distance of each contact relative to the modiolus. This measurement is normalized by the electrode, being zero “0” the closest position to the modiolus and one “1” the closest position to the lateral wall25. Amongst four ICPI values, the ICPI at point1, which presented the lowest intracochlear position index (ICPI) value, displayed the lowest ECAP threshold (Fig. S4). In addition, the average ECAP threshold between point 1, point 2, and point 3 did not differ, suggesting that the modiolus-electrode distance presented in the current study correlates with the ECAP thresholds across electrode arrays (Fig. S4). The longer the modiolus-electrode distance, the higher the ECAP threshold. Resultantly, ICPI values presented in the current study likely show an inverse correlation with the ECAP thresholds at respective points, suggesting the effectiveness of our proposed measure of electrode array position on prediction of stimulability of the auditory nerve. Accordingly, functional relevancy justifies our CT measurement.
Correlation of modified ICPI with spiral diameter
Pearson correlation analysis was performed to identify any correlation, beyond just association, between the modified ICPI or its metrics and previously proposed parameter of modiolar proximity, the spiral diameter, measured from a post-insertion radiograph during or after cochlear implantation. The mean spiral diameter was 3.12 ± 0.38 mm, exhibiting well-preserved normality that exactly followed a Gaussian distribution. A significant positive correlation was consistently observed between each ICPI value and the spiral diameter on the transorbital view (Fig. 5a). Amongst four ICPI values, the ICPI (1) measured on electrode No. 7-8, which presented the lowest value amongst metrics (see Fig. 4a), displayed the most significant correlation with the spiral diameter (r = 0.592, P < 0.001) relative to other ICPI values, likely indicating that modiolus-electrode distance at position 1 would most significantly contribute to spiral diameter. Interestingly, the most significant correlation was obtained between average value of four ICPIs (modified ICPI) and spiral diameter (r = 0.748, P < 0.001) (Fig. 5b). This suggests that our current method of spiral diameter measurement using a post-insertion radiograph comprehensively reflects modiolus-electrode distances at point 1, 2, 3 and 4 rather than a specific point, despite being a two-dimensional evaluation. This underlines the obvious role of this indicator reflecting optimal modiolar proximity.
Discussion
To the best of our knowledge, the present study is the first to come up with a reliable and simple parameter reflecting the modiolar proximity of slim modiolar electrodes based on conventional CT. Specifically, we adopted ultra-high kernel specification with high resolution index for conventional CT, finally achieving better image sharpness and reduced metal artifacts (partial volume artifacts). This improved image quality allowed accurate assessment of the intracochlear electrode position and elaborate measurement of modiolar proximity. The vendor-specific application of noise amplitude and texture modification technique in ultra-high kernel specification may decrease noise level and improve resolution. Indeed, conventional CT-based scalar position and distance from electrodes to the modiolus were comparable to the results of cone beam CT in certain circumstances28. Further, upon ultra-high kernel specification with high-resolution index, the enhancement of intracochlear visualization of the electrode array in conventional CT could justify the newly introduced indicator (i.e., modified ICPI) specialized for modiolar proximity evaluation. The modified ICPI measured from our cohort in conventional CT ranged from 0.34 to 0.54, which showed a normal distribution. Introduction of modified ICPI would make it possible to accurately and routinely evaluate modiolar proximity for slim modiolar electrodes in cochlear implantation centers where it was previously impossible due to lack of cone-beam CT. This would popularize analysing postoperative speech outcome with relation to modiolar proximity.
Recent studies have shown a diverse range of spiral configuration of slim modiolar electrodes due to intrinsic and extrinsic factors4,7,36, raising the importance of a de facto real-time imaging modality visualizing degree of modiolar proximity on the spot. A recent study highlighted that the cochlear duct length-based customized insertion technique might potentially ensure better modiolar proximity20, which would potentially lead to better speech outcomes. Specifically, short cochlear duct length subsequently leads to less modiolar proximity, which in turn renders some subjects with short cochlear duct length less amenable to this pull-back maneuver. If that is the case, readjustment of the electrode position by pulling the electrode array further back 1–2 mm more than in the conventional pull-back maneuver, called ‘extended pull-back maneuver’, likely leads to enhanced modiolar proximity, as evidenced by our recent study20. Coupled with reloadable design of the slim modiolar electrodes, simple X-ray imaging would serve as a great alternative. Previous studies have demonstrated the excellent performance of the post-insertion simple radiographs in the evaluation of the depth and angle of insertion of electrodes in intraoperative and postoperative settings32,37. What about evaluation of modiolar proximity?
Despite previous reports addressing the parameter of modiolar proximity (e.g., spiral diameter) based on the post-insertion simple radiograph, whether or not two-dimensional spiral diameter measured at a fixed position could comprehensively reflect three-dimensional spiral configuration of slim modiolar electrodes remained elusive. In this perspective, the present study merits strong attention since the modified ICPI comprehensively reflects electrode position at four points of the first cochlear turn, exhibited much tighter correlation with the spiral diameter on the post-insertion radiograph than did any of each ICPI at four points. This, in turn, suggests that the X-ray imaging-based spiral diameter would reflect the distance between the several contacts and the modiolus. Intraoperative use of simple radiographs to evaluate the degree of modiolar proximity, angular insertion depth and intracochlear positioning of electrodes under minimal exposure to radiation38 seems judicious, particularly in children. Taken together, our results reiterate the value of X-ray-based measurement of spiral diameter, which enables immediate visual feedback and changes in insertion technique during operation, thereby ensuring optimal modiolar proximity.
However, CT measurement should not be underestimated as just an interim approach, but we would better regard it as a valuable independent approach per se. Indeed, simple X-ray is not possible to indicate the degree of modiolar proximity of the electrodes, without being equipped with cone beam CT and calculation of ICPI through it. Importantly, each ICPI (i) could provide otologists with clinically significant information on intracochlear positioning of slim modiolar electrode, which is primarily related to the distance of each contact relative to the modiolus at specific points. Importantly, these values can be measured at a fixed position based on the two lines associated with the modiolus and the round window, thus permitting excellent reproducibility. Among them, ICPI (1) measured at junction of proximal basal and distal middle region (i.e., Electrode No. 7-8) showed the lowest value due to the longest distance from the modiolus to the lateral wall, while relatively maintaining the distance from the modiolus to the electrode. The enrolled subjects were exclusively implanted using the pull-back technique, enabling closer positioning of this slim modiolar electrode array to the modiolus. Recently, a human cadaveric temporal bone study revealed that the pull-back technique led to better modiolar proximity of the slim modiolar electrode array, especially in upper basal and middle turn of the cochlea29. Indeed, our measurement herein showed no difference of the distance among the electrode contacts from modiolus at point 1 (upper basal turn), 2 (middle turn), and 3 (lower apical turn). This enhanced modiolar proximity, especially at point 1 (upper basal turn) and 2 (middle turn), could be the result of the pull-back technique. Meanwhile, the distance between the modiolus and the lateral wall, which gradually decreased along the trajectory approaching the apical region, was consistent with the histological findings of the human cochlea39. Remarkably, we noticed that ICPI (1) with the lowest value also had the most tight correlation with X-ray imaging-based spiral diameter amongst the four ICPI values. This may indicate that changes in modiolar proximity in the upper basal cochlear region corresponding to ICPI (1) may be the most sensitive in influencing the spiral configuration of the electrode array. With a reduction of the electrode volume up to 75% similar to that of the current lateral wall electrodes4, the thinner and flexible properties of the slim modiolar electrode may make it more prone to variable spiral configuration in the upper basal cochlear region, compared with the middle and apical region. Further, the highest standard deviation of ICPI (1) indicating a wide range of degree of modiolar proximity, particularly in the upper basal region, may have contributed to the most significant correlation of it with the X-ray imaging-based spiral diameter.
As with introduction of significant novel findings in literature, the present study also has limitations that should be addressed in future studies. First, the small sample size and retrospective study design could be associated with weak statistical power, despite the fact that our observed values exhibited well-preserved normality, which followed a Gaussian distribution. Second, although previous studies substantiated that the electrode to modiolus distance based on conventional CT was significantly correlated with that based on a histological evaluation, the gold standard technique28, in vivo evaluation of the modified ICPI itself, was not validated histologically in our study. We therefore suggest that future studies should include large scale histological evaluation. Third, the structure of the cochlea is relatively small and irregular; therefore, it is difficult to designate the ROI when performing quantitative and qualitative assessments between standard and high-resolution filter specifications. As a result, the average ROI was relatively small and heterogeneous. Lastly, we cannot fully explain the impact of electrode position on speech performance at present, just postulating that improved modiolar proximity might lead to better speech perception outcomes in cochlear implant recipients. Specifically, Holden et al. proposed that a tightly wrapped array (i.e., wrapping factor) elicits significantly higher word recognition scores1. Furthermore, a prediction model has revealed that the average modiolus-to-electrode distance strongly correlated with speech perception scores for perimodiolar electrodes, including slim modiolar electrodes13.
Conclusion
Introduction of conventional CT-based modified ICPI would make it possible to accurately and routinely evaluate modiolar proximity for slim modiolar electrodes in all cochlear implantation centers. Further, our results reiterate the value of X-ray-based measurement of spiral diameter which enables immediate visual feedback and changes in insertion technique intraoperatively, thereby ensuring optimal modiolar proximity. Additional pre-processing techniques, such as the metal artifact reduction and three-dimensional image fusion techniques, could further enhance the clinical significance of conventional CT-based measurement of several parameters in cochlear implantation.
Data availability
Data for all submitted results is available.
References
Holden, L. K. et al. Factors affecting open-set word recognition in adults with cochlear implants. Ear Hear. 34, 342 (2013).
Holder, J. T. et al. Matched cohort comparison indicates superiority of precurved electrode arrays. Otol. Neurotol. 40, 1160–1166 (2019).
Iso-Mustajärvi, M. et al. A new slim modiolar electrode array for cochlear implantation: a radiological and histological study. Otol. Neurotol. 38, e327–e334 (2017).
Cuda, D. & Murri, A. Cochlear implantation with the nucleus slim modiolar electrode (CI532): a preliminary experience. Eur. Arch. Otorhinolaryngol. 274, 4141–4148 (2017).
Aschendorff, A. et al. Clinical investigation of the nucleus slim modiolar electrode. Audiol. Neurotol. 22, 169–179 (2017).
McJunkin, J. L., Durakovic, N., Herzog, J. & Buchman, C. A. Early outcomes with a slim, modiolar cochlear implant electrode array. Otol. Neurotol. 39, e28–e33 (2018).
Ramos-Macías, A., Borkoski-Barreiro, S., Falcón-González, J. & Ramos-de Miguel, A. Hearing preservation with the slim modiolar electrode Nucleus CI532 cochlear implant: a preliminary experience. Audiol. Neurotol. 22, 317–325 (2017).
Hey, M. et al. Objective, audiological and quality of life measures with the CI532 slim modiolar electrode. Cochlear Implants Int. 20, 80–90 (2019).
Friedmann, D. R., Kamen, E., Choudhury, B. & Roland, J. T. Jr. Surgical experience and early outcomes with a slim perimodiolar electrode. Otol. Neurotol. 40, e304–e310 (2019).
Gomez Serrano, M., Patel, S., Harris, R. & Selvadurai, D. Initial surgical and clinical experience with the Nucleus CI532 slim modiolar electrode in the UK. Cochlear Implants Int., 20, 1–10 (2019).
Ramos-Macias, A., O’Leary, S., Ramos-deMiguel, A., Bester, C. & Falcon-González, J. C. Intraoperative intracochlear electrocochleography and residual hearing preservation outcomes when using two types of slim electrode arrays in cochlear implantation. Otol. Neurotol. 40, S29–S37 (2019).
O’Connell, B. P. et al. Insertion depth impacts speech perception and hearing preservation for lateral wall electrodes. Laryngoscope 127, 2352–2357 (2017).
Chakravorti, S. et al. Further evidence of the relationship between cochlear implant electrode positioning and hearing outcomes. Otol. Neurotol. 40, 617–624 (2019).
Shaul, C. et al. Slim, modiolar cochlear implant electrode: Melbourne experience and comparison with the contour perimodiolar electrode. Otol. Neurotol. 41, 639–643 (2020).
Durakovic, N. et al. Immediate and 1-year outcomes with a slim modiolar cochlear implant electrode array. Otolaryngol.-Head Neck Surg. 162, 731–736 (2020).
de Miguel, Á. R., Argudo, A. A., Barreiro, S. A. B., González, J. C. F. & Macías, A. R. Imaging evaluation of electrode placement and effect on electrode discrimination on different cochlear implant electrode arrays. Eur. Arch. Otorhinolaryngol. 275, 1385–1394 (2018).
Garaycochea, O. et al. Comparative study of two different perimodiolar and a straight cochlear implant electrode array: surgical and audiological outcomes. Eur. Arch. Otorhinolaryngol. 277, 69–76 (2020).
Pierre, P. V., Eklöf, M., Smeds, H. & Asp, F. Cochlear implantation with the CI512 and CI532 precurved electrode arrays: one-year speech recognition and intraoperative thresholds of electrically evoked compound action potentials. Audiol. Neurotol. 24, 299–308 (2019).
Nassiri, A. M. et al. Hearing preservation outcomes using a precurved electrode array inserted with an external sheath. Otol. Neurotol. 41, 33–38 (2020).
Lee, S. Y. et al. Modiolar proximity of slim modiolar electrodes and cochlear duct length: correlation for potential basis of customized cochlear implantation with perimodiolar electrodes. Ear Hear. https://doi.org/10.1097/AUD.0000000000000920 (2020).
Van Weert, S., Stokroos, R. J., Rikers, M. M. & Van Dijk, P. Effect of peri-modiolar cochlear implant positioning on auditory nerve responses: a neural response telemetry study. Acta Otolaryngol. 125, 725–731 (2005).
Perenyi, A. et al. Electrophysiological measurements with electrode types of different perimodiolar properties and the same cochlear implant electronics—a retrospective comparison study. J. Otolaryngol.-Head Neck Surg. 48, 1–7 (2019).
Perényi, Á., Jóri, J., Csanády, M. & Rovó, L. Dimensions of the human temporal bone that are relevant to cochlear implantation surgery in infants and toddlers. A clinical-radiological study. Orv. Hetil. 160, 936–943 (2019).
Holden, L. K. et al. Factors affecting outcomes in cochlear implant recipients implanted with a perimodiolar electrode array located in scala tympani. Otol. Neurotol. 37, 1662 (2016).
de Miguel, Á. R., Durmo, I., González, J. C. F., Barreiro, S. B. & Macías, A. R. Evaluation of intracochlear position of a slim modiolar electrode array, by using different radiological analyses. Otol. Neurotol. 40, S10–S17 (2019).
Boyer, E. et al. Scalar localization by cone-beam computed tomography of cochlear implant carriers: a comparative study between straight and periomodiolar precurved electrode arrays. Otol. Neurotol. 36, 422–429 (2015).
Jia, H. et al. Intraoperative conebeam CT for assessment of intracochlear positioning of electrode arrays in adult recipients of cochlear implants. Am. J. Neuroradiol. 39, 768–774 (2018).
Zeitler, D. M., Wang, K. H., Prasad, R. S., Wang, E. Y. & Roland, J. T. Flat-panel computed tomography versus multislice computed tomography to evaluate cochlear implant positioning. Cochlear Implants Int. 12, 216–222 (2011).
Riemann, C., Sudhoff, H. & Todt, I. The pull-back technique for the 532 slim modiolar electrode. BioMed Res. Int. 2019, 6917084 (2019).
Lee, S.-Y. et al. Appropriate imaging modality for the etiologic diagnosis of congenital single-sided deafness in children. J. Clin. Med. 7, 515 (2018).
Xu, J., Xu, S.-A., Cohen, L. T. & Clark, G. M. Cochlear view: postoperative radiography for cochlear implantation. Otol. Neurotol. 21, 49–56 (2000).
Gallant, S. et al. Comparison of skull radiograph and computed tomography measurements of cochlear implant insertion angles. Otol. Neurotol. 40, e298–e303 (2019).
Kidoh, M. et al. Reduction of dental metallic artefacts in CT: value of a newly developed algorithm for metal artefact reduction (O-MAR). Clin. Radiol. 69, e11–e16 (2014).
Mittmann, P. et al. Electrophysiological detection of intracochlear scalar changing perimodiolar cochlear implant electrodes: a blinded study. Otol. Neurotol. 36, 1166–1171 (2015).
Mittmann, P. et al. Electrophysiological detection of scalar-changing perimodiolar cochlear electrode arrays: a six-month follow-up study. Audiol. Neurotol. 20, 400–405 (2015).
Gomez Serrano, M., Patel, S., Harris, R. & Selvadurai, D. Initial surgical and clinical experience with the Nucleus CI532 slim modiolar electrode in the UK. Cochlear Implants Int. 20, 207–216 (2019).
Svrakic, M. et al. Measurement of cochlear implant electrode position from intraoperative post-insertion skull radiographs: a validation study. Otol. Neurotol. 36, 1486 (2015).
Struffert, T. et al. Imaging of cochlear implant electrode array with flat-detector CT and conventional multislice CT: comparison of image quality and radiation dose. Acta Otolaryngol. 130, 443–452 (2010).
Erixon, E., Högstorp, H., Wadin, K. & Rask-Andersen, H. Variational anatomy of the human cochlea: implications for cochlear implantation. Otol. Neurotol. 30, 14–22 (2009).
Acknowledgements
This research was supported by the SNUBH Research Fund (Nos. 00-2020-13-2018-015, 00-2020-13-2016-014, and 00-2020-13-2015-019 to B.Y.C).
Author information
Authors and Affiliations
Contributions
The final manuscript has been seen and approved by all the authors, and they have given necessary attention to the manuscript to ensure the integrity of the work. S.Y.L. and B.Y.C designed and performed experiments, analyzed data and wrote the paper; S.Y.L performed the statistical analysis; S.Y.L and J.H.H collected medical charts; J.H.H and B.Y.C collected and analyzed genetic data; Y.J.B and B.Y.C measured the radiological metrics. C.M and B.Y.C provided critical revision. All authors reviewed the manuscript.
Corresponding authors
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Lee, SY., Han, J.H., Carandang, M. et al. Simpler and effective radiological evaluations for modiolar proximity of a slim modiolar cochlear implant electrode. Sci Rep 10, 17714 (2020). https://doi.org/10.1038/s41598-020-74738-x
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-020-74738-x
- Springer Nature Limited