Abstract
Carbon ion radiotherapy is a sophisticated radiation treatment modality because of its superiority in achieving precise dosage distribution and high biological effectiveness. However, there exist beam range uncertainties that affect treatment efficiency. This problem can be resolved if the clinical beam could be monitored precisely in real-time, such as by imaging the prompt gamma emission from the target. In this study, we performed real-time detection and imaging of 718 keV prompt gamma emissions using a Si/CdTe Compton camera. We conducted experiments on graphite phantoms using clinical carbon ion beams of 290 MeV/u energy. Compton images were reconstructed using simple back-projection methods from the energy events of 718 keV prompt gamma emissions. The peak intensity position in reconstructed 718 keV prompt gamma images was few millimeters below the Bragg peak position. Moreover, the dual- and triple-energy window images for all positions of phantoms were not affected by scattered gammas, and their peak intensity positions were approximately similar to those observed in the reconstructed 718 keV prompt gamma images. In conclusion, the findings of the current study demonstrate the feasibility of using our Compton camera for real-time beam monitoring of carbon ion beams under clinical beam intensity.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Introduction
Carbon ion radiotherapy (CIRT) is a widely accepted modality for the treatment of deeply seated tumors because of its high-dose localization around the Bragg peak (BP)1 and its biological effectiveness2,3,4,5. However, these unique properties make CIRT more sensitive to beam range uncertainties6,7,8. Unintended irradiation causes unnecessary damage to healthy tissues and reduces tumor doses9,10. One of the methods used to reduce uncertainties is the real-time monitoring of beams based on secondary radiation measurement. In this regard, a positron emission tomography (PET) system is one of the potential systems11,12,13,14,15,16. However, high-energy positrons and annihilation gamma rays emitted via pair production are less related to the range17. On the contrary, C-11, N-13, and O-15 are affected by the metabolic washout effect because of their long half-lives18. Owing to these difficulties and large detector size, in-beam PET has limited detection efficiency as a real-time beam monitoring system19,20,21.
Apart from the positron annihilation products, the relaxation of the excited nuclei also produces secondary photons, emitting gamma rays of a wide energy range22 from hundreds of keV to 10 MeV. Because prompt gamma rays are usually generated within nanoseconds, they are not prone to washout effects23. Prompt gamma-ray imaging at lower energies below 1 MeV, such as 718 keV, could be efficient in terms of washout effects, gamma detection efficiency, detector size, portability, and economic investments24 than do high-energy gamma rays. The projectile 12C ions collide with the target nuclei and produce 10B* (0.7 ns half-life) via the nuclear spallation or β+ decay of 10C25,26, and 10B* generates 718 keV gamma rays through the relaxation. Therefore, the distribution of the generated 718 keV gamma rays would correlate with the carbon beam range, although the exact distribution of the amount of gamma emissions within the track remains unclear. In addition, 718 keV prompt gamma rays are also expected to be an objective measurement for whole gamma imaging27.
Although the Compton camera was originally introduced for astronomy28,29, it shows potential for application in particle therapy for beam range monitoring30,31,32. The beam monitoring ability of our Compton camera for real-time annihilation gamma imaging has been successfully demonstrated33,34. The main advantages of a Compton camera are its compact size and its wide energy range (100 keV to several MeV)35,36 based purely on Compton kinematics without the use of mechanical collimators37. A fundamental Compton camera consists of two types of sub-detectors called scatterers and absorbers (Fig. 1a). For an individual gamma emission, Compton scattering occurs in the scatterer, and photoabsorption occurs in the absorber. We call these scattering and absorption events together with a Compton event. A cone is formed by the scattered angle θ in the scatterer. The cones are accumulated together to determine the gamma source (Fig. 1b). The scattered angle θ is calculated as follows:
where me c2 is the mass energy of an electron, E1 is the energy of the recoiled electron in the scatterer, and E2 is the energy deposited in the absorber.
Currently, we are developing a Si/CdTe-based Compton camera for medical purposes and exploring its applications33,34,38,39,40,41,42,43. Si detectors have a smaller Doppler broadening than other semiconductors44, and the employed CdTe detector comprises high-resolution Schottky diodes45. Therefore, the energy resolution of the Si/CdTe camera is high and leads to higher angular resolution than that of other commercial gamma cameras. In this study, owing to the superior features of our Si/CdTe Compton camera and the advantage of measuring the on-beam prompt gamma emissions, we performed real-time measurement of 718 keV prompt gamma rays emitted by a therapeutic carbon ion beam. This study provides valuable information regarding real-time beam range monitoring since, to date, there have been no detailed reports of beam range estimation method using 718 keV gamma rays, neither with the Compton camera nor with conventional gamma cameras.
Results
On-beam range verifications using prompt gamma emission
The 718 keV prompt gamma emissions from graphite phantom placed at different positions were measured using the Compton camera for the experimental setup illustrated in Fig. 2a. Figure 2b shows the energy spectrum measured from the graphite phantom placed at the origin using our Compton camera. The numbers of selected Compton events within the photo-peak window of 718 keV during the 10-min detection were 263, 265, and 336 from the phantom at origin (0 mm), 30 mm, and 60 mm, respectively. The detection ratio of a Compton event per projectile was approximately 5 × 10–10. The live time ratio evaluated for the phantom placed at 0 mm position was 8%. Figure 3 shows the reconstructed back-projection images of 680 keV (Fig. 3a,d,g), 718 keV (Fig. 3b,e,h), and 760 keV (Fig. 3c,f,i) for the phantoms positioned at 0 mm, 30 mm, and 60 mm, respectively. The arrow in Fig. 3c shows the direction and location of the beam entrance position in all the experiments. Although a distinct energy peak of 718 keV was not visible in the energy spectrum (Fig. 2b), peak intensity positions of 718 keV prompt gamma emissions could be successfully observed in the reconstructed Compton images. This was because the background component of the signal detected by the Compton camera was large; thus, the 718 keV signal was buried in the noise in the energy spectrum. However, the noise detected by the Compton camera from different directions had a lesser effect on the reconstructed Compton image. Because the data measured by the Compton camera is a list of all data of gamma rays generated from all the space (list-mode method) that includes the main signal and the noise from a wide background region. Thus, the pixel values related to the noise are diluted, however, the pixel values reconstructed by the signal are concentrated in the particle beam trajectory. Therefore, even though the signal-to-noise ratio is poor in the energy spectrum, the reconstructed image shows carbon beam trails. The reconstructed Compton images of 680 keV and 760 keV at respective phantom positions did not appear to have peak intensity gamma emissions (Fig. 3a,c,d,f,g,i) that trace the BP position.
The peak intensity position of the prompt gamma emissions was evaluated by plotting a y-direction projection profile on the reconstructed Compton images (Fig. 4) at the origin (0 mm, red line), 30 mm shift (blue line), and 60 mm shift (black line) positions. Y-axis values were normalized using the maximum values. The projection range of 20 mm was approximately twice the diameter of beam size, and a Gaussian function was fit within 30 mm in the x-direction. The dotted rectangular line in Fig. 3b indicates the projection range used for evaluating the x-projection profile of Fig. 3b,e,h to plot their respective x-projection profile in Fig. 4. Table 1 shows the numerical value of BP position calculated using Monte Carlo simulation, peak intensity position in 718 keV Compton image estimated by Gaussian fitting on the x-projection profile, the difference of their peak intensity position, and peak shift from the origin. The estimated experimental values may have been affected by some random and systematic errors such as those caused by a manual shifting of the phantom and projection evaluation on the reconstructed image using the appropriate Gaussian fitting method and projection range.
Dual- and triple-energy window imaging
We reconstructed the dual-energy window (DEW) and triple-energy window (TEW) images using the data with the energy window of 680 keV, 718 keV, and 760 keV to evaluate the effect of scattered gamma rays on the 718 keV prompt gamma images. Figure 5 shows the reconstructed images of 718 keV (Fig. 5a,d,g: same as Fig. 3b,e,h) and the corresponding DEW (Fig. 5b,e,h) and TEW (Fig. 5e,f,i) images for the phantom positioned at 0 mm, 30 mm, and 60 mm, respectively. The peak intensity positions in DEW images were located at − 1.0 mm, − 2.5 mm, and + 3.1 mm from the 718 keV prompt gamma images for the phantoms at 0 mm, 30 mm, and 60 mm, respectively. Furthermore, the peak intensity positions evaluated using TEW images were located at − 0.3 mm, − 1.5 mm, and + 5.9 mm from the 718 keV prompt gamma emission images for the phantoms at 0 mm, 30 mm, and 60 mm, respectively. Thus, the positions of the peak intensities of gamma emissions evaluated using DEW and TEW images were approximately at the same location as that of the imaging results with 718 keV alone.
Discussion
We studied the feasibility of Compton camera for range verification of clinical carbon ion beam. The 718 keV prompt gamma rays emitted from the graphite phantom through nuclear reactions with projectiles were measured using a Si/CdTe Compton camera. The positions of peak intensities of the gamma emissions deviated from the BP positions by 4.0–13.3 mm. With the shift of the graphite phantoms by 30 mm and 60 mm, the peak intensities of gamma emission positions were also shifted. In this study, although the 60-mm shift was highly affected by statistical errors and angular positioning, the peak intensity positions of 718 keV prompt gammas in 0-mm and 30-mm phantom positions were closer to BP than those reported in previous studies of gamma measurements were33,46. It is unclear whether this deviation is because of the resolution of the Compton camera or the distribution of 10B*. Although the quantitative deviation is unknown at this time, the position of the BP should be related to the position where the most 718 keV gamma rays are generated. In fact, we confirmed that the peak position in the Compton image with 718 keV shifted in accordance with the position of the BP. Therefore, the Compton image with 718 keV could be used to estimate the BP position. To apply the range estimation, the imaging ability should be improved, and the accuracy confirmed. Concerning the angular resolution of our Compton camera with 5° for a 700 keV source kept 150 mm from the Compton camera detector, the spatial two-point resolution was 12.2 mm. If we visualize 718 keV gamma rays from the monoenergetic carbon beams and measure the peak position, we would be able to estimate BP with a higher resolution, although significant data is required. Considering the high dose concentration ability of CIRT, the spatial resolution of 1 cm is not good enough. However, in clinical practice the tumor may move several centimeters, which results in a significant dose reduction47,48. Though, further development of Compton camera would be necessary to achieve higher spatial resolution, the Compton images at this current state has potential to contribute to the confirmation of treatment effects and adaptive treatments.
Using a broad/wide beam of σ = 11 mm, we successfully reconstructed Compton images of 718 keV prompt gamma emissions. We expected an improved image quality for cases where small-sized beams such as those used in spot scanning techniques were used because the area of the high intensity position was small.
We verified the presence of 718 keV prompt gamma rays at the position of the Compton camera by using a cadmium zinc telluride (CZT) detector. The irradiated carbon ion beam specifications and the experimental setup were the same as those used in beam range monitoring with the Compton camera. The target phantom was cuboid polymethyl methacrylate (PMMA). The energy spectrum measured by the CZT detector clearly illustrated the presence of 718 keV prompt gamma rays at the position of the Compton camera placed for beam range monitoring (Supplementary Figure). This confirms that the 718 keV peak would be hidden among the background noise and would not be visible in the energy spectrum detected using the Compton camera. However, no traces of peak intensities of prompt gamma emissions were observed in the individual reconstructed images of 680 keV (Fig. 3a,d,g) and 760 keV (Fig. 3c,f,i), similar to that experienced in our previous studies43. This indicates that the peak intensities in the 718 keV images arise from the prompt gamma emissions (not from any scattered gamma rays from high-energy secondary radiations). The effects of shifts in phantom positions on peak intensity positions were also clearly observed in DEW and TEW images (Fig. 5); the resultant peak intensity positions were consistent with those of 718 keV prompt gamma images. The image contrast was also significantly improved in the DEW and TEW images because of the suppression of the effect of scattered radiation. At the same time, the pixel values in the surrounding area were reduced to approximately zero. Therefore, we expect that the TEW and DEW methods may help quantify carbon beam intensity.
A preliminary calculation of Monte Carlo simulations had shown that the generation of 10B* fragments responsible for 718 keV prompt gamma emissions evaluated for water and PMMA phantom was slightly less (20% and 10%, respectively) than in graphite phantom. Moreover, our preliminary results on the PMMA phantom did not show a significant difference from that obtained in the graphite phantom. However, due to the complex chemical composition of a human body, 718 keV generation may slightly be lower than the graphite phantom. In addition, we also set the thickness of the phantom to 15 mm (relatively very low value with respect to human size or human size phantom) to overcome the influence of scattering and reliably capture the prompt gamma. For practical use, it will be necessary to increase the detection efficiency significantly.
The Compton camera used in this study (ASTROCAM) is a commercial product developed for the purpose of environmental measurement. ASTROCAM has a high-energy resolution, but there are some parts that are not suitable for beam monitoring. We expect that improvement in terms of optimization of detector arrangement, energy thresholds, and application-specific integrated circuits (ASIC) would deliver better results. For example, ASTROCAM is equipped with a large number of sub-detectors, but in practice, only the pair of the last layer of Si and the first layer of CdTe acquired almost all data. Moreover, the other (less useful) sub-detectors cause accidental coincidence measurements (e.g., 3-hit events), which increase dead time and reduce the signal. The efficiency of the Compton camera could be improved by aligning the detectors at the side. The size of the current sub-detectors is 50 × 50 mm2. Considering the size of single-photon emission computed tomography (SPECT) and flat-panel detector, the efficiency could be increased 100 times or more. If we could establish a precise synchronization between the beam pulses and the Compton camera operation to measure prompt gamma rays generated during the “on” stage of the pulses, the ratio of the signal could be improved and noise during the “off” stage eliminated. If we can reduce the analog to digital conversion time by improving the ASIC, we would reduce the dead time by a factor of 10. We expect that improvements in the efficiency of the Compton camera by four or more orders of magnitude would be acceptable for the passive irradiation technique for efficient beam range monitoring. However, for the spot scanning irradiation technique, the efficiency of the Compton camera should be improved by five orders of magnitude. This can be accomplished by performing electron tracking techniques in the Si detector, depth analysis inside the CdTe detector, and improvements in image reconstruction algorithms.
The range of carbon ion beam can be imaged using 718 keV prompt gamma rays. Nevertheless, the current state of Compton camera is inefficient to directly adopt this system for clinical application. However, the results of this study reveal the feasibility of Compton camera for beam range verification, especially for 718 keV prompt gammas.
Methods
Compton camera
In this study, we used the Si/CdTe semiconductor-based Compton camera (ASTROCAM 7000HS; Mitsubishi Heavy Industries Ltd., Japan)49. Figure 1a shows the schematic illustration of the Compton camera used in this study. It consists of eight layers of Si detectors and four layers of CdTe detectors. The length and width of each of the detectors were 50 mm. The energy resolution of the Compton camera was 2.2% with respect to 662 keV gamma rays at full width at half maximum (FWHM). The angular resolution measure was 5.4° (FWHM) for 662 keV.
Imaging method
Compton images in this study were reconstructed using the simple back-projection (uniformly enlarged projection) method35,50. We selected only the “two-hit events” (one hit in the Si detector and another in the CdTe detector) as a Compton event for image reconstruction. The energy window was set to 708–728 keV, and the threshold energy to determine a hit in both the detectors was set to 20 keV. The size of the field of view was 200 × 200 mm2, and the pixel size was 1 mm/px.
In some cases, the high-energy gamma rays simultaneously scatter in both the scatterer and absorber, and occasionally, we observe second scattering events of low-energy gamma rays in the absorber. The errant low-energy gamma rays generate artifacts in the reconstructed image. Conventionally, these artifacts are removed in SPECT by adopting DEW or TEW imaging methods43. In this study, we adopted DEW and TEW methods to evaluate the presence of crosstalk effects in the reconstructed 718 keV prompt gamma image from the surrounding energy windows. To reconstruct the DEW images, an upper-side energy window of 750–770 keV for 760 keV was considered, whereas, for TEW images, a lower-side energy window of 670–690 keV for 680 keV and an upper-side energy window of 750–770 keV for 760 keV were used.
Heavy-ion carbon beam
All beam monitoring experiments were conducted using carbon beams at Gunma University Heavy Ion Medical Center (GHMC), Japan51. GHMC uses the passive irradiation technique for clinical treatments. In this study, the beam used was also passive irradiation without wobbling instead of spot scanning irradiation. A horizontal carbon ion beam with a Gaussian distribution (σ = 11 mm) was delivered to the target graphite phantom for 10 min each in all experiments. The monoenergetic beam energy and beam intensity were set at 290 MeV/u and 3 × 109 particles per pulse in 3 s (i.e., 6 × 1011 carbon ions), respectively. The duration of the pulses was 3 s, comprising 1 s “on” and 2 s “off” pulses. The beam flux intensity was set to 1 × 109 per second to represent an actual clinical situation of beam delivery (“clinical” in this study was a matter of flux instead of fluence), despite the chances of large noise generations, accidental coincidences, and longer dead time ratios. In the entire beam monitoring experiment, beam delivery and operation of the Compton camera began and ended simultaneously. The BP position estimated by Monte Carlo simulation using Particle and Heavy-Ion Transport code System52 under the experimental conditions was at 98.4 mm from the beam entrance position of the graphite phantom (Table 1). All the values of BP positions mentioned in this study was the position of the highest dose delivered to the target phantom in the MC simulation calculation. The attenuation of the beam energy due to air transmission from the nozzle to the target phantom was also considered in the Monte Carlo simulation.
Experimental setup
Graphite phantoms (200 × 200 × 15 mm3, 1.71 g/cm3) were used. Figure 2a shows a schematic illustration of the experimental setup for on-beam imaging of prompt gamma emission. The precise alignment of the graphite phantom with the beam direction was confirmed using a three-dimensional laser level installed inside the experimental room. The total distance from the beam axis to the first layer of the Si detector plane in the Compton camera was 140 mm. First, the measurements were made with the graphite phantom placed at the origin (0 mm). (The center of the camera and the calculated peak position were aligned.) Subsequently, the phantom was replaced with a new one, and the installation positions were shifted 30 mm and 60 mm away from the origin in the beam direction. The shift of the graphite phantom was in a perpendicular direction to the Compton camera, as illustrated in Fig. 2a.
References
Bragg, W. & Kleeman, R. LXXIV. On the ionization curves of radium. Lond. Edinb. Dublin Philos. Mag. J. Sci. 8, 726–738 (1904).
Weyrather, S., Ritter, M. & Scholz, W. RBE for carbon track-segment irradiation in cell lines of differing repair capacity. Int. J. Radiat. Biol. 75, 1357–1364 (1999).
Kraft, G. Tumor therapy with heavy charged particles. Prog. Part. Nucl. Phys. 45, S473–S544 (2000).
Tsujii, H. & Kamada, T. A review of update clinical results of carbon ion radiotherapy. Jpn. J. Clin. Oncol. 42, 670–685 (2012).
Gwosch, K. et al. Non-invasive monitoring of therapeutic carbon ion beams in a homogeneous phantom by tracking of secondary ions. Phys. Med. Biol. 58, 3755–3773 (2013).
Schardt, D., Elsässer, T. & Schulz-Ertner, D. Heavy-ion tumor therapy: Physical and radiobiological benefits. Rev. Mod. Phys. 82, 383–425 (2010).
Knopf, A. & Lomax, A. In vivoproton range verification: A review. Phys. Med. Biol. 58, R131–R160 (2013).
Dosanjh, M., Jones, B., Pawelke, J., Pruschy, M. & Sørensen, B. Overview of research and therapy facilities for radiobiological experimental work in particle therapy. Report from the European Particle Therapy Network radiobiology group. Radiother. Oncol. 128, 14–18 (2018).
Enghardt, W. et al. Dose quantification from in-beam positron emission tomography. Radiother. Oncol. 73, S96–S98 (2004).
Howell, R. Second primary cancers and cardiovascular disease after radiation therapy. NCRP Report No. 170. Med. Phys. 39, 7729–7731 (2012).
Pawelke, J. et al. The investigation of different cameras for in-beam PET imaging. Phys. Med. Biol. 41, 279–296 (1996).
Iseki, Y. et al. Range verification system using positron emitting beams for heavy-ion radiotherapy. Phys. Med. Biol. 49, 3179–3195 (2004).
Parodi, K., Bortfeld, T. & Haberer, T. Comparison between in-beam and offline positron emission tomography imaging of proton and carbon ion therapeutic irradiation at synchrotron- and cyclotron-based facilities. Int. J. Radiat. Oncol. Biol. Phys. 71, 945–956 (2008).
Fiedler, F. et al. On the effectiveness of ion range determination from in-beam PET data. Phys. Med. Biol. 55, 1989–1998 (2010).
Yamaya, T. et al. Development of a small prototype for a proof-of-concept of OpenPET imaging. Phys. Med. Biol. 56, 1123–1137 (2011).
Ferrero, V. et al. Double-field hadrontherapy treatment monitoring with the inside in-beam PET scanner: Proof of concept. IEEE Trans. Radiat. Plasma Med. Sci. 2, 588–593 (2018).
Bongrand, A., Busato, E., Force, P., Martin, F. & Montarou, G. Use of short-lived positron emitters for in-beam and real-time β+ range monitoring in proton therapy. Phys. Med. 69, 248–255 (2020).
Parodi, K. et al. Patient study of in vivo verification of beam delivery and range, using positron emission tomography and computed tomography imaging after proton therapy. Int. J. Radiat. Oncol. Biol. Phys. 68, 920–934 (2007).
Cho, J. et al. Feasibility of proton-activated implantable markers for proton range verification using PET. Phys. Med. Biol. 58, 7497–7512 (2013).
Hirano, Y. et al. Washout rate in rat brain irradiated by a (11)C beam after acetazolamide loading using a small single-ring OpenPET prototype. Phys. Med. Biol. 61, 1875–1887 (2016).
Parodi, K. On- and off-line monitoring of ion beam treatment. Nucl. Instrum. Methods Phys. Res. A 809, 113–119 (2016).
Le Foulher, F. et al. Monte Carlo simulations of prompt-gamma emission during carbon ion irradiation. IEEE Trans. Nucl. Sci. 57, 2768–2772 (2010).
Moteabbed, M., España, S. & Paganetti, H. Monte Carlo patient study on the comparison of prompt gamma and PET imaging for range verification in proton therapy. Phys. Med. Biol. 56, 1063–1082 (2011).
Takeda, S. et al. Applications and imaging techniques of a Si/CdTe Compton gamma-ray camera. Phys. Procedia 37, 859–866 (2012).
Kozlovsky, B., Murphy, R. & Ramaty, R. Nuclear deexcitation gamma-ray lines from accelerated particle interactions. Astrophys. J. Suppl. Ser. 141, 523–541 (2002).
Pshenichnov, I., Mishustin, I. & Greiner, W. Distributions of positron-emitting nuclei in proton and carbon-ion therapy studied with GEANT4. Phys. Med. Biol. 51, 6099–6112 (2006).
Mohammadi, A. et al. Compton-PET imaging of 10C for range verification of carbon ion therapy. 2018 IEEE Nucl. Sci. Symp. Med. Imag. Conf. Proc. (Online). 1–3, 10 (2018).
Schoenfelder, V. et al. Instrument description and performance of the Imaging Gamma-Ray Telescope COMPTEL aboard the Compton Gamma-Ray Observatory. Astrophys. J. Suppl. Ser. 86, 657 (1993).
Takahashi, T. et al. Hitomi (ASTRO-H) X-ray astronomy satellite. J. Astron. Telesc. Instrum. Syst. 4, 1 (2018).
Rohling, H. et al. Clinical applicability of the Compton camera for prompt γ-ray imaging during proton therapy. Radiother. Oncol. 118, S90–S91 (2016).
Kataoka, J. et al. Ultracompact Compton camera for innovative gamma-ray imaging. Nucl. Instrum. Methods Phys. Res. A 912, 1–5 (2018).
Draeger, E. et al. 3D prompt gamma imaging for proton beam range verification. Phys. Med. Biol. 63, 035019 (2018).
Parajuli, R. et al. Annihilation gamma imaging for carbon ion beam range monitoring using Si/CdTe Compton camera. Phys. Med. Biol. 64, 055003 (2019).
Shiba, S. et al. Use of a Si/CdTe Compton Camera for in vivo real-time monitoring of annihilation gamma rays generated by carbon ion beam irradiation. Front. Oncol. https://doi.org/10.3389/fonc.2020.00635 (2020).
Takeda, S. et al. Demonstration of in-vivo multi-probe tracker based on a Si/CdTe semiconductor Compton camera. IEEE Trans. Nucl. Sci. 59, 70–76 (2012).
Suzuki, Y. et al. Three-dimensional and multienergy gamma-ray simultaneous imaging by using a Si/CdTe Compton Camera. Radiology 267, 941–947 (2013).
Phillips, G. Gamma-ray imaging with Compton cameras. Nucl. Instrum. Methods Phys. Res. B 99, 674–677 (1995).
Sakai, M. et al. In vivo simultaneous imaging with 99mTc and18F using a Compton camera. Phys. Med. Biol. 63, 205006 (2018).
Sakai, M. et al. Effect of number of views on cross-sectional Compton imaging: A fundamental study with backprojection. Phys. Med. 56, 1–9 (2018).
Sakai, M. et al. Compton imaging with 99mTc for human imaging. Sci. Rep. 9, 12906 (2019).
Nakano, T. et al. Imaging of 99mTc-DMSA and 18F-FDG in humans using a Si/CdTe Compton camera. Phys. Med. Biol. 65, 05LT01 (2020).
Sakai, M. et al. Improved iterative reconstruction method for Compton imaging using median filter. PLoS One 15, e0229366 (2020).
Sakai, M. et al. Crosstalk reduction using a dual energy window scatter correction in Compton imaging. Sensors 20, 2453 (2020).
Zoglauer, A. & Kanbach, G. Doppler broadening as a lower limit to the angular resolution of next-generation Compton telescopes. in X-Ray and Gamma-Ray Telescopes and Instruments for Astronomy 1303–1309 (International Society for Optics and Photonics, 2003).
Tanaka, T. Recent achievements of the high resolution Schottky CdTe diode for γ-ray detectors. New Astron. Rev. 48, 309–313 (2004).
Taya, T. et al. First demonstration of real-time gamma imaging by using a handheld Compton camera for particle therapy. Nucl. Instrum. Methods Phys. Res. A 831, 355–361 (2016).
Kubota, Y. et al. Robustness of daily dose for each beam angle and accumulated dose for inter-fractional anatomical changes in passive carbon-ion radiotherapy for pancreatic cancer: Bone matching versus tumor matching. Radiother. Oncol. 157, 85–92 (2021).
Sakai, M. et al. Robustness of patient positioning for interfractional error in carbon ion radiotherapy for stage I lung cancer: Bone matching versus tumor matching. Radiother. Oncol. 129, 95–100 (2018).
Matsuura, D. “ASTROCAM 7000HS” radioactive substance visualization camera. Mitsubishi Heavy Ind. Tech. Rev. 51, 68–75 (2014).
Takeda, S. et al. Experimental results of the gamma-ray imaging capability with a Si/CdTe semiconductor Compton camera. IEEE Trans. Nucl. Sci. 56, 783–790 (2009).
Ohno, T. et al. Carbon ion radiotherapy at the Gunma University Heavy Ion Medical Center: New facility set-up. Cancers 3, 4046–4060 (2011).
Sato, T. et al. Particle and heavy ion transport code system, PHITS, version 2.52. J. Nucl. Sci. Technol. 50, 913–923 (2013).
Acknowledgements
Authors thank Dr. Mitsutaka Yamaguchi of Takasaki Advanced Radiation Research Institute for technical information about Compton camera and advice. Authors also thank beam operator staffs of accelerator engineering corporation (AEC), Japan.
Funding
This work was supported by the JSPS KAKENHI (Grant number JP 19K1726100).
Author information
Authors and Affiliations
Contributions
The manuscript has been read and approved by all authors. K.A and M.S conceived and coordinated the research work. R.K.P., M.S., Y.K., N.K., and M.T. conducted experiments. R.K.P., M.S., K.A., Y.K., N.K., and M.T. analyzed the data. R.K.P., M.S., and Y.K., conducted simulations and data calibrations. R.K.P. prepared the figures and drafted the manuscript. The manuscript was reviewed by all the authors.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Parajuli, R.K., Sakai, M., Arakawa, K. et al. Carbon range verification with 718 keV Compton imaging. Sci Rep 11, 21696 (2021). https://doi.org/10.1038/s41598-021-00949-5
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-021-00949-5
- Springer Nature Limited