Abstract
Accurate and rapid point-of-care (PoC) diagnostics are critical to the control of the COVID-19 pandemic. The current standard for accurate diagnosis of SARS-CoV-2 is laboratory-based reverse transcription polymerase chain reaction (RT-PCR) assays. Here, a preliminary prospective performance evaluation of the QuantuMDx Q-POC SARS-CoV-2 RT-PCR assay is reported. Between November 2020 and March 2021, 49 longitudinal combined nose/throat (NT) swabs from 29 individuals hospitalised with RT-PCR confirmed COVID-19 were obtained at St George’s Hospital, London. In addition, 101 mid-nasal (MN) swabs were obtained from healthy volunteers in June 2021. These samples were used to evaluate the Q-POC SARS-CoV-2 RT-PCR assay. The primary analysis was to compare the sensitivity and specificity of the Q-POC test against a reference laboratory-based RT-PCR assay. The overall sensitivity of the Q-POC test compared with the reference test was 96.88% (83.78– 99.92% CI) for a cycle threshold (Ct) cut-off value for the reference test of 35 and 80.00% (64.35–90.95% CI) without altering the reference test’s Ct cut-off value of 40. The Q-POC test is a sensitive, specific and rapid PoC test for SARS-CoV-2 at a reference Ct cut-off value of 35. The Q-POC test provides an accurate option for RT-PCR at PoC without the need for sample pre-processing and laboratory handling, enabling rapid diagnosis and clinical triage in acute care and other settings.
Similar content being viewed by others
Introduction
Severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) is an enveloped, positive sense, single-stranded RNA virus of the genus betacoronavirus that caused the corona virus disease 2019 (COVID-19) pandemic with devastating public health impact. By the end of 2022, there have been reported more than 650 million individuals infected and more than 6.5 million deaths globally1. The first diagnostic assays for this disease were developed shortly after the initial publication of the viral genome in January 20202. These were nucleic acid amplification tests (NAATs) based on reverse transcriptase polymerase chain reaction (RT-PCR), where target viral RNA sequences are converted to DNA and exponentially amplified to allow detection with fluorescent probes. The lower the concentration of target RNA sequence in a clinical sample, the higher the cycle threshold (Ct) value at which fluorescence is detectable.
Most convenient assays use upper respiratory (e.g. nasopharyngeal or oropharyngeal or combined nose/throat (NT)) swab samples to detect SARS-CoV-2 RNA and are highly sensitive3. Viral load is highest in the upper respiratory tract in the first week of illness4. However, standard laboratory-based RT-PCR assays are relatively complex and time consuming, requiring highly trained personnel and expensive equipment based in a centralised laboratory. They take several hours to perform in batches, and, when time is added for sample transport, results often take 24 h or more to be reported. This can result in delayed diagnosis, which can complicate clinical triage.
Point-of-care (PoC) diagnostics were identified as one of eight key research priorities to tackle the pandemic5. Rapid, PoC or near-patient testing for active infections confers several benefits over laboratory-based testing6,7,8 including faster triage, which in turn aids infection control and clinical management. Rapid PoC tests can also enable effective community testing and alleviate pressure on overburdened centralised labs. While lateral flow-based rapid antigen tests (RDTs) are now in widespread use, they are less sensitive than NAATs9. Thus, development of rapid PoC NAAT-based tests are a priority, especially for use when immediate actions are needed for positive cases such as in accident and emergency departments, pre-surgical or chemotherapy day unit admissions, care homes, airports, prisons and low resource healthcare settings. Several versions of PoC NAATs have been developed10,11,12,13, but all have one or more limitations including long sample-to-result times, complexity (steps required) and cost.
To address the urgent need for a rapid and sensitive COVID-19 diagnostic, QuantuMDx have repurposed their RT-PCR PoC diagnostics platform, the Q-POC, to run a SARS-CoV-2 assay. Here, we present a detailed description of this technology and a preliminary prospective performance evaluation of Q-POC compared with validated laboratory-based RT-PCR assays.
Methods
Ethics statement
The study was conducted in accordance with relevant UK guidelines and regulations. Ethical approval for clinical samples collected at St George’s Hospital was provided by the Institutional Review Board (Integrated Research Application System project ID: 282104; South Central—Oxford C Research Ethics Committee reference: 20/SC/0171; registered at clinicaltrials.gov NCT04351646) as part of the “Development and Assessment of Rapid Testing for SARS-CoV-2 outbreak” (DARTS) study, sponsored by St George’s Hospital NHS Foundation Trust. The samples collected at QuantuMDx were from healthy individuals undergoing testing in a pre-existing staff COVID-19 testing programme. The samples were collected for the purposes of determining test performance (test performance evaluation) and were anonymised, therefore ethical approval was not deemed necessary. All participants for this study were recruited following informed consent.
Swab samples
Patients with COVID-19 hospitalised at St George’s Hospital were recruited into the prospective study between 19th November 2020 and 3rd March 2021 (inclusive). Prior to recruitment, they were confirmed as positive for SARS-CoV-2, following NT swabs (in Sigma Virocult, Corsham, UK) taken as standard of care at hospital admission. These were processed on site at South West London Pathology, using Roche RNA extraction kits (Magnapure, West Sussex, UK) followed by RT-PCR with altona Diagnostics RealStar SARS-CoV-2 RT-PCR (S and E target genes, Hamburg, Germany) or Roche cobas SARS-CoV-2 Test (E and ORF target genes) kits. All patients consented to NT swabs being taken. Following recruitment to DARTS, longitudinal NT swab samples were collected, using a Copan swab (Cat number: 503CS01), at days 0, 3, 5, 7, 10 and 14 (± 2 days) or until they withdrew from the study. Anonymised samples were transferred to long-term storage within 2 h of collection on the wards, without viral transport medium, at -80 C until use. In addition, anonymised mid-nasal (MN) swab samples were collected between 15 and 18th June 2021 (inclusive) from healthy staff volunteers at QuantuMDx following informed consent and tested immediately.
Q-POC testing
The Q-POC device is a PoC diagnostics platform, which runs a rapid molecular test in a single use microfluidic cassette (Q-CAS) by incorporating sample processing, DNA amplification (with RT step, if required), and downstream qualitative detection of pathogens in small volumes of sample (Fig. 1). The total hands-on time is approximately one minute. It involves two main steps, (1) loading a 400 µl sample into a test cassette and (2) inserting the test cassette into the instrument. Results can be interpreted after ~ 30 min.
For detection of SARS-CoV-2 by RT-PCR at PoC a flexible mini tip Copan FLOQSwab (designed for nasopharyngeal sampling) was used to take a NT or a MN swab. The tip (fresh or defrosted) was then broken off into a sample tube containing 3 ml of MSwab sample collection, transport and preservation medium (Copan Diagnostics, Italy). Following sample tube inversion (× 5), a 400 µl aliquot is pipetted into a test cassette, Q-CAS, and sealed in with a cap.
The loaded cassette is then scanned and loaded into the Q-POC device. Within the Q-CAS, reconstitution of lyophilised RT-PCR reagents, held within sealed ‘Reagent Fuses’ for increased shelf-life, occurs using 50 µl of the swab eluate containing released RNA. Following an RT step, rapid PCR is achieved by shuttling fluid between heating zones (shuttle flow PCR, sfPCR—Fig. 1B). This allows target amplification and detection in 32 min and equates to 48 cycles (with a threshold for positive samples of ≤ 45). SARS-CoV-2 gene targets (Orf1ab, N and S) are all detected in a single fluorescence channel (FAM), while an RNase P target (used as a control) is detected in a second channel (HEX). Fluorescence readings are taken after each cycle (qPCR reader) and are analysed to provide a qualitative result. The use of individual heating zones combined with shuttling of the fluid, unlike the temperature ramping used by conventional PCR, is far more energy efficient allowing for multiple assay runs even when the Q-POC is battery operated. The Q-POC is front-loading and intuitive, with a simple touch screen interface designed for PoC testing in resource limited settings. It contains all the necessary mechanics, electronics and optics to drive the cassette-based assays, in addition to reading, interpreting and presenting test results. With regards to analysis of data, a calling algorithm gives a positive, negative or invalid result with no need for user interpretation of curves. It is portable and operable either on mains power or by an exchangeable battery, which can last up to 4 h. Data are recorded and initially stored on the Q-POC within a secured database, with a storage capacity of > 1000 patient results. In addition, the Q-POC will connect to a specified receipt-style printer. The Q-POC operating system runs on an ARM-based processing board integrating connectivity options: GPS, Bluetooth, Wi-Fi and GPRS (2G/3G) connectivity. Operation data are also retained within the Q-POC to allow for tracking and trending of instrument performance.
Laboratory-based reference testing
An aliquot (140 µl) of each swab sample in 3 ml of MSwab was used with a QIAamp Viral RNA Mini Kit (Qiagen, Germany) for RNA extraction, as per the manufacturer’s instructions. The CE IVD-marked SARS-CoV-2 RT-PCR Detection Assay (QuantuMDx, UK) was used as the reference test and gold standard for assessing the Q-POC SARS-CoV-2 RT-PCR assay. It is based on TaqMan chemistry and utilises FAM for the detection of the three SARS-CoV-2 loci, Orf1ab, N gene and the S gene and HEX for the detection of the RNase P gene (a specimen and process control). The assay takes approximately 75 min to run, undertaking 45 cycles of PCR. The cycle threshold (Ct) for detection of all SARS-CoV-2 targets is 40, with ≤ 40 indicating a valid positive result. The Ct for detection of the RNase P target is 35, with ≤ 35 with no SARS-CoV-2 signal (or Ct > 40) indicating a valid negative result. Runs are invalid if there is no SARS-CoV-2 signal (or Ct > 40) and no RNase P signal (or Ct > 35).
Statistical analysis
Operators were not blinded to the status of patients/volunteers from which the samples were taken prior to Q-POC testing but did not have access to the results of the subsequent reference testing; however, the Q-POC requires no subjective interpretation of a result.
Data were analysed with GraphPad Prism (version 6.07 for Windows). Diagnostic sensitivity, specificity, positive predictive value and negative predictive value were determined, using the laboratory-based RT-PCR as a reference standard (Supplementary Data S1). Samples that were invalid on the PoC platform were not included in the primary sensitivity analysis. The manuscript was prepared in accordance with the EQUATOR Network’s STARD guidelines (Supplementary Data S2).
Results
Swab samples (n = 150) were collected and tested, using the Q-POC (Fig. 2). Two cohorts were recruited. In the first, hospitalised patients at St George’s Hospital with an RT-PCR confirmed COVID-19 diagnosis at admission were clinically assessed between the 19th November 2020 and 3rd March 2021 as part of DARTS. The recruits provided longitudinal samples predicted to have a wide range of viral loads thereby providing a challenging sample set for initial evaluation of the Q-POC. Patients (n = 29) who provided a NT sample, using a mini tip Copan swab, were eligible for inclusion in the Q-POC evaluation. The median age of COVID-19 positive study participants included in the study was 64 (IQR 52–74), 20 (69%) were male, 16 (55%) were symptomatic and the majority (24; 83%) had mild disease at enrolment (Table 1).
Two samples from this cohort produced invalid results on the Q-POC test and were not retested. One of these samples came from a patient who provided only one sample so only their invalid result was included in the analysis (Fig. 2). Therefore, complete clinical data paired with laboratory reference test data were available for 28 patients and 47 samples, with 40 samples testing positive and 7 testing negative using the reference test (along with the specified Ct cut-off of 40). Participants formed a convenience series. No adverse events from performing the Q-POC or the reference test were recorded.
In the second cohort, 101 MN Copan swab samples were collected from healthy volunteer staff to provide a SARS-CoV-2 negative dataset. Of these, 6 samples returned an invalid Q-POC test result. One was not retested and 5 were retested, producing valid results. Therefore, test data for further analysis were available for 100 samples from this cohort, all of which tested negative with the reference test (Fig. 2).
Overall, a total of 155 sample runs between the 2 cohorts were undertaken on the Q-POC. Of these, there were 8 invalid runs, 5 of which were retested to produce valid results, and 3 of which were not retested as the samples were destroyed (Fig. 2). This gives an invalid rate over the study period of 5.2%. Of the 147 samples with paired Q-POC and reference test data, 40 were positive and 107 negative for SARS-CoV-2 with the reference test.
Figure 3 presents the Ct values from the reference test used with longitudinal NT swab samples from COVID-19 positive participants. As predicted, this sample set provided a wide range of viral loads (using Ct values as a basic proxy), including multiple samples between Ct values of 30 and 40. Included in the graph are the available Ct data from the hospital test undertaken prior to enrolment (for 22 of the 28 participants analysed). The day the swab was taken for this test was used as day 0. Two different validated assays (see “Methods”) were used at St George’s Hospital over the study period. Irrespective of the test used, there is a general increase in Ct values with time after the first test. No participants with a hospital test Ct value below 26 produced a negative result within the short (12 day) study period, while several participants with a starting hospital Ct value above 26 were negative at subsequent time points.
Prior to the reference RT-PCR testing undertaken to derive the data presented in Fig. 3, samples were run on the Q-POC SARS-CoV-2 RT-PCR assay. While the assay is designed to generate a SARS-CoV-2 positive or negative result, Ct values can also be derived. Figure 4 compares Ct values for the reference versus the Q-POC test. Data are from samples that were called positive by both tests (n = 32) and thus have paired Ct data. The two datasets passed normality tests (P = 0.86 and P = 0.69 for the reference and Q-POC tests, respectively; D’Agostino & Pearson omnibus normality test). Interestingly, there is very good concordance between the two paired datasets (as judged by eye, Fig. 4A). Linear regression analysis of Q-POC and reference datasets (Fig. 4B) show a strong correlation (r2 = 0.76). A Bland–Altman plot demonstrated very good agreement between the 2 tests, with a bias of − 6.7 Ct units meaning the Q-POC test averages a 6.7 higher Ct value with mean (± SEM) Ct values of 25.8 ± 0.9 versus 32.5 ± 0.9 for the reference and Q-POC tests respectively (Fig. 4C). This measured bias is consistent with the Ct cut-off value of 40 (above which a result is called negative) for the reference assay compared with the Ct cut-off value of 45 for the Q-POC assay.
With the inclusion of data from the 47 samples from COVID-19 patients and from single samples from 100 healthy volunteers, the performance characteristics of the Q-POC test were determined (Table 2). The overall sensitivity (95% CI) of the Q-POC test was 80.0% (64.4–91.0%) against the SARS-CoV-2 RT-PCR laboratory reference test with a specificity and 99.1% (94.9–100%), with PPV and NPV of 97.0% and 93.0%, respectively. Interestingly, the sensitivity of the Q-POC test was 96.9% (31/32) at a reference test Ct ≤ 35 and only 12.5% (1/8) at a reference test Ct > 35. Thus, if the reference test Ct cut-off was adjusted to 35, sensitivity increased significantly to 96.9% (83.8–99.9%), while specificity dropped only slightly to 98.3% (82.2–99.9%).
Of note, while not evaluated in this study, the instructions for use (IFU; Supplementary Data S3) for the Q-POC test states that the limit of detection is 1000 genome equivalent copies per ml with a standard deviation of 0.8 and average Ct of 41.89, which is in line with the results presented above. Furthermore, in vitro analysis for exclusivity or cross-reaction was undertaken with no cross-reactivity observed under the tested conditions. All micro-organisms were used at a concentration of 1 × 105 colony forming units (cfu) per reaction. Finally, as recommended by the Medicines & Healthcare products Regulatory Agency (MHRA), UK, QuantuMDx continuously monitor and assess the assay design in silico for newly emerging variants of SARS-CoV-2 with regular reports which are published by the MHRA.
Discussion
In a pandemic that causes a Public Health Emergency of International Concern, one of the earliest interventions is accurate diagnosis of an infection. At the start of a pandemic, this will depend largely on clinical evaluations (including imaging, biochemical and haematological assessments) that guide immediate containment measures. NAATs can be developed with urgency to diagnose a highly transmissible infectious disease accurately and assist in managing patients and breaking transmission chains (as happened with COVID-192). Accurate diagnostics are also critical to support the development of new therapeutic interventions, and to assess public health measures to contain infections, including the development and implementation of vaccination programmes.
There are several challenges for developers of diagnostics in the early stages of a pandemic. Regulatory authorities will be as unfamiliar with a disease such as COVID-19 as are other experts and this may delay appropriate evaluation of tests under development. Urgent need coupled with limited supplies and a motive to profit can encourage the proliferation of diagnostic tests that are easy to produce in bulk (such as lateral flow assays) but that fail to meet minimum requirements for accuracy (even if these requirements have been agreed, which itself can take too long procedurally). One way to circumvent some of these limitations is to repurpose high quality NAAT-based testing platforms to deal with new infections such as those caused by SARS-CoV-2.
The COVID-19 pandemic highlighted an urgent need for PoC methods to detect SARS-CoV-2, including its variants. To address this, QuantuMDx initially developed a laboratory-based multiplex assay with simple-to-use lyophilised reagents that are stable when stored at room temperature, and quicker to run than many others3. However, these tests require automated equipment and batch testing, and cannot give rapid results for individual assessment.
QuantuMDx therefore repurposed cassettes, the portable analytical platform and software to provide an accurate, affordable and simple-to-use PoC test for SARS-CoV-2, as detailed in the Methods. Here it is demonstrated that, using a reference test Ct cut-off of 35, the Q-POC test for SARS-CoV-2 had a sensitivity of 96.88% (83.78–99.92% CI) and a specificity of 98.3% (82.2–99.9%). These performance characteristics are comparable to or better than other PoC NAAT devices10,11,12,13.
Centralised RT-PCR testing has the advantage of higher throughput capability with Q-POC being able to process only one cartridge at a time. This can be mitigated to an extent depending on the clinical setting and clinical testing algorithm by using multiple processing units with random access for each. An advantage of the Q-POC includes a rapid run-time (32 min) when compared with the longer running times of laboratory-based diagnostic platforms and some PoC NAAT devices also10,11. A possible disadvantage is the need to add the swab to a buffer and a following pipetting step to introduce the sample into the cassette, rather than placing the sample swab directly into the cassette. Given that MSwab does not inactivate viruses, this may require the use of a safety hood for sample flow in a laboratory context, although in non-laboratory contexts PoC devices including lateral flow tests have been used globally without requiring specialised handling facilities. The flexibility in sourcing samples into a buffer before processing retains the ability to re-test on the same instrument, the ability to retest on a different instrument/platform and to sequence retained samples. These are advantages of the described Q-POC/Q-CAS system that has been developed as a platform diagnostic.
Within this study, the Q-POC test failure rate was 5.2% (8/155). Of the 8 invalid results all were QC errors where 2 were invalid as the RNase P target was detected at > 40, 3 were invalid as the RNase P target was not detected at all, and 3 were invalid because both the internal process control (IPC; detected in the ROX channel) and the RNase P target were not detected. Consistent with improvements in manufacturing, following this preliminary evaluation the cartridge failure rate (‘invalid result’ category) is expected to be < 2% and within specifications and industry standards.
Ct values as an estimate of viral load and/or infectiousness can be influenced by many variables such as the site and methods for sample collection, the choice of diagnostic targets within the viral genome and sample quality14. In establishing cut-off values a key principle is that the clinical risk of transmission must be minimised, so that chains of transmission are effectively interrupted. Several reports have suggested a Ct cut-off for reference RT-PCR tests of < 3515,16,17. When Ct values are assessed in relation to the ability to culture virus, there are < 3% culture positive samples with Ct = 3518. This Ct cut-off value is the most clinically relevant, because samples with higher cut-off Ct values are non-infectious and therefore could be termed ‘biological false positives’ despite being frequently detected (with up to 50% of positive results in assays having higher Ct cut-off values)19. Interestingly, the reference test’s analytical sensitivity data demonstrate that a Ct of 35 equates to approximately 10 genome copies per reaction (Supplementary Date S4). If as the data suggest, the Q-POC test can detect similar viral loads but no lower, this would equate to a lower limit of detection (LLOD) of 200 copies/ml (with a 50 µl reaction volume in the cassette), which is in keeping with data in the IFU and a recently published study from Spain that evaluated the Q-POC test and estimated a LLOD of 100 copies/ml20. In this latter study, a cut-off Ct of 30 was suggested at which the sensitivity of the Q-POC test was 100% and in line with Spanish guidelines20.
At a reference test cut-off Ct of 35, the Q-POC test has excellent performance characteristics, that can be further examined in larger clinical trials. Testing strategies can be optimised for individual use cases, such as for travellers, health-care workers, prison services, triage of patients coming to emergency departments, wherever there is no laboratory and for contacts at risk of infection by timing testing strategies appropriately and by repeating testing at intervals when indicated.
Our study has several limitations. Although the sample size was small, it has been intended to assess and optimise performance of this novel technology not to develop individual use cases. Secondly, paired swab samples were not used for the reference and the Q-POC test rather both tests were run from a single swab sample. While this might remove the variable associated with multiple sampling from a participant subject at any given time point, the samples were placed into 3 ml of MSwab buffer (required for the Q-POC test) rather than the normal 1 ml of viral transport medium used in general for laboratory-based RT-PCR reference tests. This may have the effect of decreasing the reference test sensitivity by a small margin. Thirdly, the samples collected from the COVID-19 patients were not accordance with the IFU for the Q-POC test, which uses MN swabs. Further comparative studies would be needed to determine if NT swabs can be used without loss of sensitivity, although it is worth noting that a small study found no difference in diagnostic sensitivity when comparing NT and nasopharyngeal swabs21.
Data availability
The datasets generated during and/or analysed during the current study are available in the supplementary information.
References
WHO. Coronavirus (COVID-19) Dashboard. https://covid19.who.int/. Accessed 02 Jan 2023.
Corman, V. M. et al. Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCR. Euro Surveill 25, 2000045. https://doi.org/10.2807/1560-7917.ES.2020.25.3.2000045 (2020).
UKHSA. QuantuMDx RT-PCR: SARS-CoV-2 Nucleic Acid Detection. https://www.gov.uk/government/publications/coronavirus-covid-19-serology-and-viral-detection-tests-technical-validation-reports/desktop-validation-report-for-q-poc. Accessed 02 Jan 2023.
Cevik, M. et al. SARS-CoV-2, SARS-CoV, and MERS-CoV viral load dynamics, duration of viral shedding, and infectiousness: A systematic review and meta-analysis. Lancet Microbe 2, e13–e22. https://doi.org/10.1016/S2666-5247(20)30172-5 (2021).
WHO. COVID 19 Public Health Emergencyof International Concern (PHEIC) - Global research andinnovation forum: towards a research roadmap. https://www.who.int/publications/m/item/covid-19-public-health-emergency-of-international-concern-(pheic)-global-research-and-innovation-forum. Accessed 02 Jan 2023
Bouzid, D. et al. Added value of rapid respiratory syndromic testing at point of care versus central laboratory testing: A controlled clinical trial. J. Antimicrob. Chemother 76, iii20–iii27. https://doi.org/10.1093/jac/dkab241 (2021).
Brendish, N. J., Malachira, A. K. & Clark, T. W. Molecular point-of-care testing for respiratory viruses versus routine clinical care in adults with acute respiratory illness presenting to secondary care: A pragmatic randomised controlled trial protocol (ResPOC). BMC Infect. Dis. 17, 128. https://doi.org/10.1186/s12879-017-2219-x (2017).
Shengchen, D. et al. Evaluation of a molecular point-of-care testing for viral and atypical pathogens on intravenous antibiotic duration in hospitalized adults with lower respiratory tract infection: A randomized clinical trial. Clin. Microbiol. Infect. 25, 1415–1421. https://doi.org/10.1016/j.cmi.2019.06.012 (2019).
Dinnes, J. et al. Rapid, point-of-care antigen and molecular-based tests for diagnosis of SARS-CoV-2 infection. Cochrane Database Syst. Rev. 8, CD013705. https://doi.org/10.1002/14651858.CD013705 (2020).
Assennato, S. M. et al. Performance evaluation of the SAMBA II SARS-CoV-2 test for point-of-care detection of SARS-CoV-2. J. Clin. Microbiol. 59, e01262-e11220. https://doi.org/10.1128/JCM.01262-20 (2020).
Gibani, M. M. et al. Assessing a novel, lab-free, point-of-care test for SARS-CoV-2 (CovidNudge): A diagnostic accuracy study. Lancet Microbe 1, e300–e307. https://doi.org/10.1016/S2666-5247(20)30121-X (2020).
Hansen, G. et al. Clinical performance of the point-of-care cobas liat for detection of SARS-CoV-2 in 20 minutes: A multicenter study. J. Clin. Microbiol. 59, e02811–02820. https://doi.org/10.1128/JCM.02811-20 (2021).
Zhen, W., Smith, E., Manji, R., Schron, D. & Berry, G. J. Clinical evaluation of three sample-to-answer platforms for detection of SARS-CoV-2. J. Clin. Microbiol. 58, e00783-e1720. https://doi.org/10.1128/JCM.00783-20 (2020).
Rabaan, A. A. et al. Viral dynamics and real-time RT-PCR Ct values correlation with disease severity in COVID-19. Diagnostics (Basel) 11, 1091. https://doi.org/10.3390/diagnostics11061091 (2021).
Chang, M. C., Hur, J. & Park, D. Interpreting the COVID-19 test results: A guide for physiatrists. Am. J. Phys. Med. Rehabil. 99, 583–585. https://doi.org/10.1097/PHM.0000000000001471 (2020).
Chen, C. J. et al. Optimization of the CDC protocol of molecular diagnosis of COVID-19 for timely diagnosis. Diagnostics (Basel) 10, 333. https://doi.org/10.3390/diagnostics10050333 (2020).
La Scola, B. et al. Viral RNA load as determined by cell culture as a management tool for discharge of SARS-CoV-2 patients from infectious disease wards. Eur. J. Clin. Microbiol. Infect. Dis. 39, 1059–1061. https://doi.org/10.1007/s10096-020-03913-9 (2020).
Jaafar, R. et al. Correlation between 3790 quantitative polymerase chain reaction-positives samples and positive cell cultures, including 1941 severe acute respiratory syndrome coronavirus 2 isolates. Clin. Infect. Dis. 72, e921. https://doi.org/10.1093/cid/ciaa1491 (2021).
Yang, Q. et al. Just 2% of SARS-CoV-2-positive individuals carry 90% of the virus circulating in communities. Proc. Natl. Acad. Sci. USA 118, e2104547118. https://doi.org/10.1073/pnas.2104547118 (2021).
Fernandez-Pittol, M. et al. Assessment of QuantuMDx Q-POC assay for rapid detection of SARS-CoV-2 using middle turbinate swabs. Microbiol. Spectr. 11, e0425622. https://doi.org/10.1128/spectrum.04256-22 (2023).
Vlek, A. L. M., Wesselius, T. S., Achterberg, R. & Thijsen, S. F. T. Combined throat/nasal swab sampling for SARS-CoV-2 is equivalent to nasopharyngeal sampling. Eur. J. Clin. Microbiol. Infect. Dis. 40, 193–195. https://doi.org/10.1007/s10096-020-03972-y (2021).
Acknowledgements
The authors thank the volunteers and site personnel for participating in the studies.
Funding
HMS is supported by the Wellcome Trust Institutional Strategic Support Fund [204809] awarded to St. George’s University of London. The DARTS study is supported by a DFID/Wellcome Trust Epidemic Preparedness coronavirus grant [220764]. This work was also supported by the St George’s Hospital Charity. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
Author information
Authors and Affiliations
Contributions
Conceptualization: J.C., M.S., P.S., C.T., T.P., H.M.S., S.K. Methodology: J.C., M.S., K.B., G.M., P.S., D.E.K., Y.A., C.T., T.P., H.M.S., S.K. Investigation: K.B., G.M., D.M., C.P., A.-M.C., E.S., J.M., D.J.C., B.M.O.D., N.M.E., E.G., D.E.K., I.M. Formal analysis: J.C., M.S., D.M., K.B., Y.A., T.P., H.M.S., S.K. Writing—original draft: J.C., Y.A., H.M.S., S.K. Writing—review and editing: all.
Corresponding authors
Ethics declarations
Competing interests
JC, MS, KB, GM, DM, PS, CP, A-MC, ES, JM and CT are employees of QuantuMDx. SK and HMS are shareholders and SK is an advisor to QuantuMDx. SK, HMS and EG are in receipt of funds from QuantuMDx to develop diagnostic technologies and assays (that have supported DJC). SK is also member of the Scientific Advisory Committee for the Foundation for Innovative New Diagnostics (FIND), a not-for-profit organisation that produces global guidance on affordable diagnostics, views expressed here do not represent FIND. All other authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Caffry, J., Selby, M., Barr, K. et al. The QuantuMDx Q-POC SARS-CoV-2 RT-PCR assay for rapid detection of COVID-19 at point-of-care: preliminary evaluation of a novel technology. Sci Rep 13, 9827 (2023). https://doi.org/10.1038/s41598-023-35479-9
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-023-35479-9