Abstract
Imaging metabolic dysfunction, a hallmark of solid tumors, usually requires radioactive tracers. Chemical exchange saturation transfer (CEST) imaging can potentially detect and visualize glucose uptake and metabolism, without the need for radioisotopes. Here, we tested the feasibility of using glucose CEST (glucoCEST) to image unlabeled glucose uptake in head and neck cancer by using a clinical 3T magnetic resonance imaging (MRI) scanner. The average CEST contrast between tumors and normal tissue in 17 patients was 7.58% (P = 0.006) in the 3–4 ppm offset frequency range and 5.06% (P = 0.02) in 1–5 ppm range. In a subgroup of eight patients, glucoCEST signal enhancement was higher in tumors than in normal muscle (4.98% vs. 1.28%, P < 0.021). We conclude that glucoCEST images of head and neck cancer can be obtained with a clinical 3T MRI scanner.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Introduction
Rapidly growing cancer cells typically have much higher rates of glycolysis than do their normal tissues of origin1,2,3,4. High rates of glucose uptake in tumors appear as increased activity on 18F-fluorodeoxyglucose positron emission tomography (FDG-PET) images and this is the underlying principle of PET imaging for cancer staging and treatment assessment5. Chemical exchange saturation transfer (CEST) imaging6,7, a noninvasive magnetic resonance imaging (MRI)-based technique, has been used to detect and image glucose uptake and metabolism in mice8,9,10,11,12,13 and shows promise for clinical applications14. In animal studies, CEST has been shown to detect glucose concentrations of a few millimoles in vivo. Unlike PET/computed tomography (CT), CEST imaging does not involve ionizing radiation, carries little to no risk to the patient and can be repeated as necessary for cancer diagnosis, staging and assessment of response.
Although extensive in vivo CEST studies have been done with various animal models15,16,17,18, studies of CEST using clinical MRI scanners have focused primarily on patients with brain tumors19,20,21 or stroke22. CEST imaging in patients with head and neck cancer is particularly difficult because of the complexity of the anatomic structures and the field inhomogeneity artifacts caused by susceptibility and motion23. Also, most previous glucose CEST (glucoCEST) imaging studies were performed at ultrahigh field strengths (7T or higher). Here, we adapted an immobilization device from the radiation therapy to image patients with head and neck cancer in the immobilized position (Fig. 1). To the best of our knowledge, this is the first CEST imaging study in which a standard clinical (3T) MRI scanner was used to detect unlabeled glucose uptake in a prospective cohort of patients with human papillomavirus–positive (HPV+) oropharyngeal head and neck cancer.
Results
CEST Signals in Head and Neck Cancer Tumors and Surrounding Tissues
We first performed a CEST study on 17 patients who had been treated for squamous cell carcinoma of the oropharynx. MRI scans were obtained with a Discovery 750 3T MRI scanner system, in which two 6-channel flex coils are placed laterally, centered on the tongue base (Fig. 1), while patients were immobilized in a position in which they would receive radiation treatment24. Tumors were located on the T2-weighted fast spin echo images and a slice through the largest cross-section of the tumor was selected for the CEST imaging measurements. Postprocessing of CEST data was based on asymmetry analysis of the magnetization transfer ratio (MTR) or Z-spectrum of the CEST image series, quantified as MTRasym25,26. A plot of MTRasym against the saturation offset frequency for a patient’s tumor, nodes and adjacent healthy tissue is shown in Fig. 2. As expected, the MTRasym signal was greater in the tumor regions than in the normal surrounding tissues over the entire saturation frequency offset range.
Quantitative Analysis of CEST Contrast: Tumors vs. Normal Tissue
For each CEST scan, we integrated the MTRasym spectrum over the offset frequency range of 1 to 5 ppm (the total CEST signal) and we quantified CEST contrast as the area between the MTRasym curves for tumors and healthy regions. The distribution of measured CEST contrasts in tumors for all 17 patients is shown in Fig. 3. The CEST contrast between tumor and normal tissue was evident in most patients, despite variations in the magnitude between patients. In the 1–5 ppm range, the average magnitude of CEST signal in tumor tissue from all 17 patients was found to be higher than that of normal surrounding tissue by 5.06% (P = 0.02), a statistically significant difference also noted in a previous report23.
We also measured CEST contrast in the 3–4 ppm range (which may be dominated by the amide proton transfer effect19,20,21). In that range, the average CEST contrast between tumor and normal tissue for all 17 patients was 7.58% (P = 0.006), which was larger than the average from the 1–5 ppm range. The integral of MTRasym over the 1–5 ppm frequency range is overlaid on an anatomic image in Fig. 4. The tumor showed higher CEST signal (indicated by red and yellow areas) than the surrounding tissue.
glucoCEST Enhancement (GCE) in Tumors and Normal Tissue
A pair of MTRasym curves in the area of a representative tumor (pre-injection and post-injection) are shown in Fig. 5. Elevated CEST signal in head and neck tumors relative to surrounding normal tissues was clearly seen after the injection of unlabeled glucose, particularly around 1ppm and in the 3–4 ppm range. The fact that glucoCEST appeared in a wide offset range may be due to the relatively fast exchange property of hydroxyl protons with water. The GCE can be calculated by integrating the area between the curves. The GCE values for tumors and normal tissue from all eight patients are shown in Fig. 6. Glucose uptake was reflected in significantly larger increases in GCE in the tumors versus in normal muscle tissues (4.98% in tumor regions vs. 1.28% in muscle, P < 0.021). Tumor regions had greater enhancement than the surrounding normal tissue in all patients.
Comparison of glucoCEST Enhancement and PET Enhancement
Finally, we registered patients’ glucoCEST images to their PET/CT images acquired within a week of the CEST MRI scans. For each patient, the PET image slice corresponding to the glucoCEST image was then extracted and qualitatively compared with the GCE image, as shown in Fig. 7. The GCE generally correlated with the PET results, even though the exact spatial distribution of PET activity and GCE did not match well.
Discussion
Our study showed that the CEST signal (measured in MTRasym) in head and neck cancer can be reliably detected and was comparable to that reported for brain tumors19,20,21. Although the exact mechanism underlying the elevation in CEST signal in tumors is complex and not completely understood, one possible explanation is increased cellularity in tumors. Moreover, a reduction of CEST signal (at ~3.5 ppm) has been observed in brain tumors after radiation treatment16, which suggests that CEST may be useful for assessing the response of cancer to radiation. Although the underlying mechanism of CEST contrast needs further investigation, the potential translational applications of CEST hold great promise for patients with head and neck cancer and perhaps for those with other diseases as well.
Our results demonstrate the feasibility and potential utility of glucoCEST for assessing glucose uptake and tumor metabolism in head and neck cancer on a clinical 3T MRI scanner. Indeed, glucoCEST could substantially affect the clinical care of patients with head and neck cancer (and perhaps those with other cancers) because CEST imaging is noninvasive and can be used repeatedly if necessary. If further studies prove that glucoCEST is a reliable alternative to PET, CEST imaging with unlabeled glucose could spare patients the radiation that is by necessity delivered for PET. Our findings further demonstrate that glucoCEST was sensitive enough in patients with head and neck cancer. The glucoCEST MRI scans can be obtained in conjunction with other commonly used MRI sequences within one hour. To address the challenges for clinical glucoCEST imaging from patient motion and susceptibility artifacts, we immobilized the patients in the treatment position by using a mask and dental stent and in this way we could detect changes in CEST signal on a clinical 3T scanner. Although we did not specifically study biological processes after glucose injection, it is reasonable to speculate that the change in CEST signal or contrast (ranging from 1 ppm to 5 ppm) from baseline to the signal after glucose injection results from the combined effects of glucose uptake and the byproducts of glucose metabolism8,9,10,11,12,13,14.
Our results further suggest that glucoCEST could be an alternative to FDG-PET for tumor diagnosis, staging and treatment assessment. Because CEST is noninvasive and does not use radioisotopes, CEST does not have the limitations of FDG-PET in terms of timing, logistics and infrastructure. CEST is also more readily available and less expensive than PET. The glucoCEST also differs from FDG-PET in the amount of glucose administered relative to the amount of FDG. In our study, we used 20 mL of glucose and the sensitivity of detecting glucoCEST was in the millimolar range. FDG-PET uses a much lower concentration of labeled glucose to avoid disturbing the tumor metabolic process. If the glucose concentration used in glucoCEST is large enough to disturb that process; then, it is possible that glucoCEST may not detect actual metabolic processes but instead behaves like a contrast agent. Detailed in vivo studies are needed to validate or disprove such concerns. However, many researchers agree that active tumor cells tend to accumulate and demand more glucose as an energy source. Consequently, higher glucose uptake is expected in active tumors than in surrounding normal tissues or in tumors that are less active (possibly as a result of treatment). Therefore, glucoCEST could be used to measure glucose uptake both for staging disease and for assessing its response to treatment.
Like PET images, glucoCEST images have higher contrast in tumor regions, indicating that glucoCEST and PET images likely measure similar metabolic processes in tumors. However, the differences in spatial distribution on the PET and glucoCEST images may indicate that glucoCEST does not measure glucose distribution but instead measures cascade effects and byproducts of the metabolic process after glucose uptake. Future comparative studies are needed to clarify tumor metabolism in vivo; nevertheless, the heterogeneities noted in glucoCEST images might be useful for imaging specific metabolites of interest in the future.
Although our study was limited by the relatively small number of patients, our results are still valid and consistent among patients. Our study was also limited in the H&N cancer which as an aggressive tumor, has high glucose metabolism. The exact magnitude of CEST and glucoCEST contrasts should be further quantified in future studies of a larger number of patients and in other disease sites with different glucose metabolism levels and growth rates to generalize the use of this technique. Further research also is needed to understand the contrast mechanism of glucoCEST. The dynamic characteristics of unlabeled glucose in vivo should be studied to determine the optimal amount of glucose to inject and the optimal timing of CEST imaging. A detailed dynamic study of glucose uptake at various concentrations using CEST may reveal more information about the mechanism of GCE and possibly the underlying biology of glucose uptake in head and neck tumors. Future studies also should explore CEST contrast variations in patients before, during and after treatment to determine CEST’s ability to assess treatment response. A study comparing GCE contrast with PET imaging before, during and after treatment could also reveal some mechanistic differences between the uptake of FDG and unlabeled glucose and perhaps could shed light on in vivo tumor metabolic processes.
The peak uptake time for involved nodes may be different from that of tumor; thus, the dynamic study of glucoCEST may have to be done with careful consideration of motion during a long MRI study between the pre- and post-glucose injection. In such a case, image coregistration may be necessary. In the case of tumor sites which have a lot of inherent motion (e.g. lung and liver), motion gating in the imaging acquisition (e.g. respiratory gating) in CEST and glucoCEST must be implemented to reduce motion-related artifact. In the case of small tumor and nodes, partial volume effect might reduce the glucoCEST sensitivity in which case thinner imaging slices may have to be used.
In conclusion, our results show that glucoCEST imaging of patients with head and neck cancer is feasible with a 3T clinical scanner. The fact that tumors have significantly higher CEST signal intensities and higher uptake of unlabeled glucose than does normal tissue makes CEST a potential alternative to PET.
Methods
Patients
All experiments were approved by the University of Texas MD Anderson Cancer Center institutional review board (IRB) and were performed in accordance with the IRB rules and policies. Between October 2013 and July 2015, 17 patients with head and neck squamous cell carcinoma of the oropharynx were recruited in this study. All patients gave study-specific informed consent to participate in the study and all experiments were carried out in accordance with the relevant guidelines. Other inclusion criteria were (i) histologically documented stage III or IV HPV-positive squamous cell carcinoma of the oropharynx and (ii) at least one lesion (primary tumor or metastatic node) measuring more than 1.5 cm in its maximum axial diameter.
MRI Experiments
All MRI studies were done with a Discovery 750 3T MRI scanner system (GE Healthcare, Milwaukee, WI), using two laterally placed 6-channel flex coils centered on the tongue base (Fig. 1). Patients were immobilized in the position in which they would receive radiation treatment with an intraoral tongue-immobilizing/swallow-suppressing dental stent and with a customized thermoplastic head and shoulder mask (Klarity Medical Product, Newark, OH). Patient setup and scanning technique are described in detail elsewhere24. This immobilization approach greatly improves image co-registration in longitudinal scans and reduces interval physiologic motion (e.g., swallowing). Scan volume (field of view, 25.6 cm; slice number, ~30; slab thickness, 12 cm; and pixel size, 1 mm × 1 mm in-plane) was prescribed for a standardized spatial region encompassing the palatine process region and the cricoid cartilage for all pre-, mid- and post-treatment scans. T2-weighted, pre-contrast T1-weighted and post-contrast T1-weighted with fat saturation axial images were acquired using a fast spin echo sequence.
CEST MRI
The CEST imaging pulse sequence consisted of a magnetization transfer prepared spoiled gradient echo sequence modified to saturate a range of frequency offsets, producing a spectrum for each voxel that can be used to generate CEST contrast. An off-resonance saturation pulse with a near-rectangular Fermi pulse profile and a duration of 48 ms was applied for each repetition of the sequence. The saturation pulse was applied with frequency offsets in steps of 71.4 Hz over a range of −1000 Hz to 1000 Hz for a total of 29 offsets. We acquired a single-slice CEST image (field of view, 240; slice thickness, 5 mm; repetition time/echo time, 64/4.4; flip angle, 20; number of excitations, 1 cm; acquisition matrix, 192 mm × 192 mm; and bandwidth, 244.14 Hz/pixel) for each patient and the scan time was approximately 5 minutes for a Z-spectrum of 29 saturation frequencies covering −1000 to 1000 Hz relative to the bulk water resonance frequency (defined as ∆f = 0). Tumors were located on the T2-weighted fast spin echo images and a single slice through the largest cross-section of the tumor was selected for the CEST imaging measurements.
CEST Data Analysis
Post-processing software was developed with IDL (Exelis, Boulder, CO) and was based on asymmetry analysis of the Z-spectrum as reported elsewhere24,25. The MTR curve (or Z-spectrum) was obtained from the CEST image series. In particular, the asymmetric MTR (MTRasym) at a given offset ∆f was defined as. This was calculated in regions of interest encircling the tumor, nodes and adjacent healthy tissue over the entire offset frequency range.
For each CEST scan, we obtained the total CEST signal by integrating the MTRasym spectrum over the offset frequency range of 1 to 5 ppm. We further quantified the CEST contrast as the area between the MTRasym curves for tumors and healthy regions. This value was then normalized by the length of the integration region to convert the contrast to a percentage difference:
Because previous in vivo studies indicated that amide protons resonate at a frequency of 3.5 ppm25,26, we also measured the CEST contrast in the 3–4 ppm range (which may be dominated by the amide proton transfer effect19,20,21). The CEST contrast between the tumor and the normal region in this ppm range was calculated by integrating the MRT curves as follows:
Glucose CEST MRI
During the glucose enhancement experiment, a baseline CEST image was obtained from 8 patients after the T1-weighted and T2-weighted images. Then, a bolus of unlabeled glucose (20 mL) was injected intravenously and a second CEST scan was obtained 10 minutes later to study glucose CEST (glucoCEST) enhancement. This timing encompasses the peak glucose uptake8,9,10,11,12,13,14. Our CEST sequence took 5 minutes to complete; therefore, this acquisition timing should ensure the capture of peak glucose uptake in the tumor.
Asymmetric MTR was analyzed in both sets of the acquired CEST image series (baseline and after glucose injection). As was done previously in an animal study8, glucoCEST enhancement (GCE) was defined as the area between the two MTRasym curves, before and after glucose injection. Again, we chose to integrate the MTRasym curve over the 1–5 ppm range because these offset frequencies are all relevant to the metabolic processes in a tumor8,9,10,11,12,13,14. This value was then normalized by the length of the integration region to convert the results to percentages:
Here, i denotes that this parameter can be calculated for either tumor or healthy regions.
Comparison of GCE and PET
Patients’ GCE images were first registered to their anatomic (T1/T2) images acquired in the same MRI study. Then, the anatomic MRI study (T1/T2) was registered to the patient’s PET/CT images, which had been acquired within 1 week of the MRI/CEST and the PET image slice corresponding to the GCE image was extracted. Finally, the GCE color map image was overlaid on the anatomic MRI image and qualitatively compared with the corresponding PET image.
Additional Information
How to cite this article: Wang, J. et al. Magnetic Resonance Imaging of Glucose Uptake and Metabolism in Patients with Head and Neck Cancer. Sci. Rep. 6, 30618; doi: 10.1038/srep30618 (2016).
References
Zu, X. L. & Guppy, M. Cancer metabolism: facts, fantasy and fiction. Biochem. Biophys. Res. Commun. 313, 459–465 (2004).
Hsu, P. P. & Sabatini, D. M. Cancer cell metabolism: Warburg and beyond. Cell 134, 703–707 (2008).
Vander Heiden, M. G., Cantley, L. C. & Thompson, C. B. Understanding the Warburg effect: the metabolic requirements of cell proliferation. Science 324, 1029–1033 (2009).
Ferreira, L. M. Cancer metabolism: the Warburg effect today. Exp. Mol. Pathol. 89, 372–380 (2010).
Gambhir, S. S. Molecular imaging of cancer with positron emission tomography. Nat. Rev. Cancer 2, 683–693 (2002).
Ward, K. M., Aletras, A. H. & Balaban, R. S. A new class of contrast agents for MRI based on proton chemical exchange dependent saturation transfer (CEST). J. Magn. Reson. 143, 79–87 (2000).
Sherry, A. D. & Woods, M. Chemical exchange saturation transfer contrast agents for magnetic resonance imaging. Annu. Rev. Biomed. Eng. 10, 391–411 (2008).
Walker-Samuel, S. et al. In vivo imaging of glucose uptake and metabolism in tumors. Nature Med. 19, 1067–1072 (2013).
Chan, K. W. Y. et al. Natural D-glucose as a biodegradable MRI contrast agent for detecting cancer. Magn. Reson. Med. 68, 1764–1773 (2012).
Rivlin, M., Horev, J., Tsarfaty, I. & Navon, G. Molecular imaging of tumors and metastases using chemical exchange saturation transfer (CEST) MRI. Sci. Rep. 3, 3045 (2013).
Jin, T., Mehrens, H., Hendrich, K. S. & Kim, S. G. Mapping brain glucose uptake with chemical exchange-sensitive spin-lock magnetic resonance imaging. J. Cereb. Blood Flow Metab. 34, 1402–1410 (2014).
Zu, Z. L., Spear, J., Li, H., Xu, J. Z. & Gore, J. C. Measurement of regional cerebral glucose uptake by magnetic resonance spin-lock imaging. Magn. Reson. Imaging 32, 1078–1084 (2014).
Xu, X. et al. Dynamic glucose enhanced (DGE) MRI for combined imaging of blood-brain barrier break down and increased blood volume in brain cancer. Magn. Reson. Med. 74, 1556–1563 (2015).
Xu, X. et al. Dynamic glucose-enhanced (DGE) MRI: Translation to human scanning and first results in glioma patients. Tomography 1, 105–114 (2015).
Sun, P. Z., Zhou, J., Sun, W., Huang, J. & van Zijl, P. C. M. Detection of the ischemic penumbra using pH-weighted MRI. J. Cereb. Blood Flow Metab. 27, 1129–1136 (2007).
Zhou, J. et al. Differentiation between glioma and radiation necrosis using molecular magnetic resonance imaging of endogenous proteins and peptides. Nature Med. 17, 130–134 (2011).
Cai, K. J. et al. Magnetic resonance imaging of glutamate. Nature Med. 18, 302–306 (2012).
Desmond, K. L., Moosvi, F. & Stanisz, G. J. Mapping of amide, amine and aliphatic peaks in the CEST spectra of murine xenografts at 7 T. Magn. Reson. Med. 71, 1841–1853 (2014).
Zhou, J. et al. Practical data acquisition method for human brain tumor amide proton transfer (APT) imaging. Magn. Reson. Med. 60, 842–849 (2008).
Zhou, J. et al. Three-dimensional amide proton transfer MR imaging of gliomas: Initial experience and comparison with gadolinium enhancement. J. Magn. Reson. Imaging 38, 1119–1128 (2013).
Togao, O. et al. Amide proton transfer imaging of adult diffuse gliomas: correlation with histopathological grades. Neuro-Oncology 16, 441–448 (2014).
Harston, G. W. et al. Identifying the ischaemic penumbra using pH-weighted magnetic resonance imaging. Brain 138, 36–42 (2015).
Yuan, J. et al. Amide proton transfer-weighted imaging of the head and neck at 3T: a feasibility study on healthy human subjects and patients with head and neck cancer. NMR Biomed. 27, 1239–1247 (2014).
Ding, Y. et al. Prospective observer and software-based assessment of magnetic resonance imaging quality in head and neck cancer: Should standard positioning and immobilization be required for radiation therapy applications? Pract. Radiat. Oncol. 5, e299–e308 (2015).
Zhou, J., Payen, J. F., Wilson, D. A., Traystman, R. J. & van Zijl, P. C. Using the amide proton signals of intracellular proteins and peptides to detect pH effects in MRI. Nature medicine 9, 1085–1090 (2003).
Zhou, J. & van Zijl, P. C. Chemical exchange saturation transfer imaging and spectroscopy. Progr. NMR Spectr. 48, 109–136 (2006).
Acknowledgements
We would like to thank Christine F. Wogan and Jill R. Delsigne for their helps in the preparation of this manuscript.
Author information
Authors and Affiliations
Contributions
Jihong W. study design and plan, sequence design, data acquisition and analysis, graph and manuscript preparation. Joseph W. data analysis, graph plot, manuscript editing. K.H. sequence development, data acquisition, manuscript editing. A.S.R.M. data acquisition, patient enrollment, manuscript editing. Y.D. data acquisition, patient enrollment, manuscript editing. C.D.F. study design, patient enrollment, manuscript editing. S.Y.L. data analysis, manuscript editing. S.J.F. patient protocol lead, patient enrollment, manuscript editing. J.Z. data analysis, manuscript and graph preparation.
Ethics declarations
Competing interests
The authors declare no competing financial interests.
Rights and permissions
This work is licensed under a Creative Commons Attribution 4.0 International License. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in the credit line; if the material is not included under the Creative Commons license, users will need to obtain permission from the license holder to reproduce the material. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/
About this article
Cite this article
Wang, J., Weygand, J., Hwang, KP. et al. Magnetic Resonance Imaging of Glucose Uptake and Metabolism in Patients with Head and Neck Cancer. Sci Rep 6, 30618 (2016). https://doi.org/10.1038/srep30618
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/srep30618
- Springer Nature Limited
This article is cited by
-
Improved postprocessing of dynamic glucose-enhanced CEST MRI for imaging brain metastases at 3 T
European Radiology Experimental (2023)
-
Multiparametric chemical exchange saturation transfer MRI detects metabolic changes in breast cancer following immunotherapy
Journal of Translational Medicine (2023)
-
Metabolic brain imaging with glucosamine CEST MRI: in vivo characterization and first insights
Scientific Reports (2023)
-
In vivo tracking of unlabelled mesenchymal stromal cells by mannose-weighted chemical exchange saturation transfer MRI
Nature Biomedical Engineering (2022)
-
GlucoCEST MRI for the Evaluation Response to Chemotherapeutic and Metabolic Treatments in a Murine Triple-Negative Breast Cancer: A Comparison with[18F]F-FDG-PET
Molecular Imaging and Biology (2022)