Abstract
In patients with end-stage heart failure, advanced therapies such as heart transplantation and long-term mechanical circulatory support (MCS) with a left ventricular assist device (LVAD) have to be considered. LVADs can be implanted as a bridge to transplantation or as an alternative to heart transplantation: destination therapy. In the Netherlands, long-term LVAD therapy is gaining importance as a result of increased prevalence of heart failure together with a low number of heart transplantations due to shortage of donor hearts. As a result, the difference between bridge to transplantation and destination therapy is becoming more artificial since, at present, most patients initially implanted as bridge to transplantation end up receiving extended LVAD therapy. Following LVAD implantation, survival after 1, 2 and 3 years is 83%, 76% and 70%, respectively. Quality of life improves substantially despite important adverse events such as device-related infection, stroke, major bleeding and right heart failure. Early referral of potential candidates for long-term MCS is of utmost importance and positively influences outcome. In this review, an overview of the indications, contraindications, patient selection, clinical outcome and optimal time of referral for long-term MCS is given.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
-
Since the introduction of continuous-flow LVADs in the Netherlands in 2006, the number of implantations has increased substantially, outnumbering heart transplantation as treatment for advanced heart failure.
-
The results after LVAD implantation justify the use as an alternative to heart transplantation.
-
Currently, the four implanting centres (UMCU, EMC, UMCG and LUMC) have sufficient capacity for the LVAD implantations needed.
-
Early referral to an LVAD-implanting centre is mandatory for optimal timing and outcome of LVAD implantation.
Introduction
Patients suffering from advanced heart failure despite individualised optimal medical treatment, with or without cardiac resynchronisation therapy, should be considered for heart transplantation or long-term mechanical circulatory support (MCS) [1]. Currently, heart transplantation is still considered to be the gold standard, showing a relatively good median survival of 15 years [2,3,4,5]. Meanwhile, long-term MCS is becoming more and more important due to the growing number of heart failure patients together with the decline in the number of donor hearts. First generation left ventricular assist devices (LVADs) were big pulsatile devices with limited durability. Already in 1993, LVADs were used as bridge to transplantation in the University Medical Center Utrecht [6]. From 2006, smaller and more reliable continuous flow devices became available. The short-term outcome was very promising with a 2-year survival of 76% [7]. Since that time, MCS has become an important part of therapy in advanced heart failure and the number of centres in the Netherlands implanting LVADs has increased to four. Outcome parameters are registered per centre and reported yearly to a central European database (EUROpean registry for patients with Mechanical Assisted Circulatory Support, EUROMACS).
LVADs can be used as bridge to transplantation, or as an alternative to heart transplantation, which is known as destination therapy and in some patients as a bridge to decision in case of temporary contraindications. The present situation in the Netherlands is that most patients with an LVAD as bridge to transplantation will have to wait several years before a donor heart becomes available and many patients will never be transplanted at all. In that way the difference between bridge to transplantation and destination therapy is becoming more and more artificial.
Currently, the HeartWare Ventricular Assist Device (HVAD) (Medtronic, Framingham, MA, USA) and the Heartmate 3 (HM3, Abbott, St. Paul, MN, USA) are the most frequently used devices for long-term MCS (Fig. 1). The HM3 replaced the Heartmate II (HM-II, Abbott, St. Paul, MN, USA) some years ago, resulting in less need for pump replacements and improved survival free of disabling stroke or reoperation for malfunction than its predecessor [8]. Both HVAD and HM3 are small centrifugal pumps implanted in the pericardial cavity showing very low rates of haemolysis, but necessitating intensive anticoagulation. The percutaneous abdominal driveline is still one of the shortcomings in the design, potentially leading to recurrent or persistent infections.
Indications for long-term MCS
Indications for long-term MCS generally follow those of heart transplantation. In case of contraindications for heart transplantation, MCS may be considered as an alternative to transplantation in selected patients for which all the below-mentioned criteria also apply:
-
Advanced heart failure with a low left ventricular ejection fraction <30% despite optimal therapy consisting of maximally tolerable medication with or without resynchronisation therapy and other interventions to optimise the cardiac condition, as indicated by the current heart failure guideline [2];
-
Exercise tolerance, assessed by cardiopulmonary exercise testing, reveals a peak VO2 <12 ml/min/kg (<14 ml/kg/min if intolerant to beta blocker) or <50% of the predicted value for age and sex in ambulatory patients: strong intrinsic motivation and Inter-agency Registry for Mechanically Assisted Circulatory Support (INTERMACS) profile 2–6 (Tab. 1; [12])
Contraindications for long-term MCS
Patients in cardiogenic shock (INTERMACS profile 1) despite an intra-aortic balloon pump, temporary MCS and/or inotropic support are not eligible for long-term MCS. In addition, a life expectancy of less than 2 years, based on extracardiac disease, is a contraindication for long-term MCS. Furthermore, severe comorbidities may be temporary or persistent contraindications (Tab. 2).
Patient selection
Patient selection is of utmost importance for outcome after LVAD implantation and is performed by a specialised, multidisciplinary team in LVAD-implanting centres, who take the above-mentioned indications and contraindications into consideration [9].
As mentioned previously, patients in INTERMACS profile 1 (refractory cardiogenic shock) are generally not candidates for long-term MCS directly, but require stabilisation on temporary MCS first, to see if organ function recovers. Primary LVAD implantation in these patients has a proven worse outcome in comparison with patients in INTERMACS profile 2–4 [10, 11].
Besides INTERMACS classification, the evaluation of right ventricular function is very important as there are no reliable options for long-term right ventricular support and right heart failure (RHF) is one of the main complications after LVAD implantation. It is thought to occur in 20–30% of patients, especially early postoperatively after LVAD implantation and is the primary cause of death in 10% [8, 10, 11]. Many criteria are formulated to try to predict perioperative RHF after LVAD implantation. No single criterion suffices, but recently a risk score based on the EUROMACS data was developed, in which invasive pressure measurements, echocardiographic and clinical parameters were combined [12]. Based on this score, a reasonable prediction of early postoperative RHF can be made (C-index of 0.70).
The final decision on LVAD implantation is made by the MCS team (consisting of at least a cardiologist, cardiothoracic surgeon and specialised nurses and technicians) weighing indication, contraindications, right ventricular function, age, previous operations and the ability and willingness of the patient to comply to a complex medical regime against a prospect of potential improvement after LVAD implantation.
With respect to contraindications, potential reversibility has to be analysed, especially with regard to renal insufficiency and hepatic failure [13,14,15]. Age has to be judged as a biological component in the decision to implant an LVAD. Although there is no absolute upper limit, given the poorer results in elderly people, it is generally not advisable to proceed in patients older than 75 years [1].
Survival
EUROMACS data, including 2113 patients, demonstrated a survival of 69% (CI 66–71%), 55% (CI 52–58%) and 44% (CI 40–47%) at 1, 2 and 3 years after continuous-flow LVAD implantation, respectively [16]. In the Netherlands, 496 patients (72% male, median age 55 (range 16–74) years) received MCS between 2006 and 2019. Current survival of the four LVAD centres combined is 83%, 76% and 70% after 1, 2 and 3 years, respectively (Fig. 2), with heart transplantation (26%), death (28%), explantation of LVAD (2%) and alive on LVAD on 31 December 2018 (44%) as the endpoint. These data are quite promising given the poor prognosis of the patients before LVAD implantation [17]. Following LVAD implantation, not only survival, but also quality of life and exercise capacity improves impressively, allowing a return to a normal life, including sports activities and even resumption of work [18,19,20]. Despite this promising survival, morbidity after LVAD implantation remains substantial, as was confirmed in a recent publication showing major bleeding and ventricular tachycardia as the most commonly encountered adverse events [21].
Adverse events
Infection
Device-related infections might be limited to the exit site of the driveline but may also extend to other parts of the system. Incidence rates are highest in the first 3 months postoperatively, namely 0.25 events per patient-year. Thereafter, incidence is 0.17 events per patient-year [10]. Patients often require longstanding antibiotic and/or surgical treatment. A recent study identified that the risk for LVAD-associated infections is increased in HM-II when compared with HVAD and in patients who need post-LVAD ICD-related procedures [22]. In the MOMENTUM 3 trial, comparing outcome in HM3 versus HM-II, device-related infections occurred equally in HM3 and HM-II [8].
Right heart failure
RHF is defined by INTERMACS as increased central venous pressure (>15 mm Hg) with echocardiographic (right heart dysfunction, dilatation and/or significant tricuspid regurgitation) and clinical signs of venous congestion [11]. This may require an increased dose of diuretics and/or inotropics and/or nitric oxide ventilation and/or temporary mechanical support. RHF can occur in the early postoperative phase, but may also develop later in the course of the disease. Patients with late RHF have a worse prognosis in terms of survival and functional capacity, and are more frequently readmitted in comparison with patients without late RHF [23].
Device malfunction
Device malfunction, including pump thrombosis and driveline-related problems, were most often seen in the HM-II resulting in the need for LVAD replacement. Technical improvement led to almost elimination of pump thrombosis in HM3, as shown in the MOMENTUM 3 trial [8]. However, in HVAD patients, pump thrombosis is still an important problem [24]. In HM3, rare cases of outflow graft twisting have been reported, resulting in decreased pump flow and the need for reparative treatment [25].
Bleeding
Major bleeding is defined as a suspected internal or external bleeding, resulting in death, rethoracotomy, hospitalisation and/or transfusion of red blood cells (within the first 7 days after the implantation requiring transfusion ≥4 units of packed red blood cells, or any transfusion beyond 7 days postoperatively) [11].
Bleeding is related to the use of anticoagulation and antiplatelet therapy in combination with acquired Von Willebrand syndrome after LVAD implantation as a result of decreased pulsatility [26,27,28]. This may result in recurrent episodes of gastrointestinal bleeding and nose bleeds.
Stroke
Patients on MCS may suffer from ischaemic and/or haemorrhagic stroke. In the MOMENTUM 3 trial strokes occurred equally (0.10 and 0.26 events per patient-year, p = 0.09, respectively) in both devices during short-term follow-up (31–180 days postoperatively), but beyond this period strokes were 3.3 times less frequently seen with HM3 [28]. Stroke is not only an important cause of morbidity, but also a predictor of mortality [29, 30]. In case of ischaemic or haemorrhagic stroke, the anticoagulation regimen often needs to be revised, thereby increasing the risk for either a haemorrhagic transformation of the ischaemic stroke or pump thrombosis, respectively. This delicate balance between thrombosis and bleeding, known as haemocompatibility, remains one of the major challenges in MCS management.
Arrhythmias
Ventricular arrhythmias are highly prevalent during MCS (30%), both in the early postoperative phase and later in the course of the disease [31]. Ventricular arrhythmias might be tolerated relatively well (i.e. no loss of consciousness) because output is preserved by the LVAD. However, clinically patients may present with RHF. Most often, the underlying cardiomyopathy leads to ventricular arrhythmias, especially in those patients who already had ventricular arrhythmias prior to the LVAD implantation [31]. There is no consensus about ICD tachytherapy in MCS patients, where a shock in conscious patients is unfortunate, while on the other hand ventricular arrhythmias are detected early to prevent RHF and hypoperfusion. Most often, ICD settings are adapted to only treat very fast ventricular arrhythmias including ventricular fibrillation. Apart from ventricular arrhythmias, atrial fibrillation is also common in MCS, and depending on the clinical effect, might require rhythm control [2].
Referral
Given the fact that the optimal timing of LVAD implantation is crucial and that the outcome after LVAD implantation in patients with rapidly progressive heart failure (INTERMACS I) is far inferior to outcome in patients with less severe heart failure, early referral to a transplant and MCS centre is mandatory. Several characteristics suggesting referral are:
-
Severely symptomatic: NYHA III+ to IV despite optimal heart failure treatment;
-
Relatively young patients with symptomatic heart failure;
-
Genetic cardiomyopathies with a likelihood of rapid progression of disease (e.g. PLN mutation);
-
Recurrent admissions for heart failure;
-
Inotrope dependency;
-
Difficulties in titration of heart failure medication (as a result of hypotension, renal failure, intolerance);
-
The need for high-dose diuretics (arbitrary >4 mg bumetanide/>160 mg furosemide).
The mnemonic ‘I Need Help’, is a helpful tool for timely referral (Tab. 3; [32]).
Conclusions and future directions
All patients with advanced heart failure that proves refractory to optimal conventional therapy have to be considered for heart transplantation and/or long-term MCS. Early consultation and referral to a tertiary centre for evaluation of treatment options and the correct timing of advanced therapies is mandatory. In this analysis, many factors have to be weighed, including prognosis without heart transplantation/MCS, outcome after heart transplantation/MCS with regard to mortality and morbidity as well as an idea on potential improvement after heart transplantation/MCS implantation.
Currently, survival after LVAD therapy in the Netherlands approximates 83%, 76% and 70% after 1, 2 and 3 years, respectively. However, this therapy is still associated with substantial morbidity. The intensive management of LVAD patients is restricted to implanting centres, but in case of adverse events, these patients may present to other hospitals. Therefore, all cardiologists need to be aware of the management of adverse events in MCS patients [33]. Outcome after LVAD therapy can be improved by technical adjustments in the design; infectious complications surely will be diminished if there is no longer a need for a driveline to deliver energy to the pump [34]. Personalised anticoagulation may decrease bleeding problems as well as thrombosis. In this way outcome after LVAD implantation will improve even more. Therefore, it has to be expected that long-term MCS will become more and more important as a generally accepted, frequently applied therapy in advanced heart failure.
References
Crespo-Leiro MG, Metra M, Lund LH, et al. Advanced heart failure: a position statement of the Heart Failure Association of the European Society of Cardiology. Eur J Heart Fail. 2018;20:1505–35.
Ponikowski P, Voors AA, Anker SD, et al. 2016 ESC Guidelines for the diagnosis and treatment of acute and chronic heart failure: The Task Force for the diagnosis and treatment of acute and chronic heart failure of the European Society of Cardiology (ESC) Developed with the special contribution of the Heart Failure Association (HFA) of the ESC. Eur Heart J. 2016;37:2129–200.
Sammani A, Wind AM, Kirkels JH, et al. Thirty years of heart transplantation at the University Medical Centre Utrecht. Neth Heart J. 2017;25:516–23.
Zijlstra LE, Constantinescu A, Manintveld O, et al. Heart transplantation in the 21st century in Netherlands: improved survival in the last decade. Ned Tijdschr Geneeskd. 2015;159:A9346.
Damman K, Brugemann J, De Boer RA, et al. Heart transplantation in the Netherlands: A national achievement. Neth Heart J. 2018;26:223–4.
Lahpor JR, de Jonge N, van Swieten HA, et al. Left ventricular assist device as bridge to transplantation in patients with end-stage heart failure: Eight-year experience with the implantable HeartMate LVAS. Neth Heart J. 2002;10:267–71.
Lok SI, Martina JR, Hesselink T, et al. Single-centre experience of 85 patients with a continuous-flow left ventricular assist device: clinical practice and outcome after extended support. Eur J Cardiothorac Surg. 2013;44:e233–e8.
Mehra MR, Uriel N, Naka Y, et al. A fully magnetically levitated left ventricular assist device—final report. N Engl J Med. 2019;380:1618–27.
Potapov EV, Antonides C, Crespo-Leiro MG, et al. 2019 EACTS Expert Consensus on long-term mechanical circulatory support. Eur J Cardiothorac Surg. 2019;56:230–70.
Goldstein DJ, Meyns B, Xie R, et al. Third annual report from the ISHLT mechanically assisted circulatory support registry: a comparison of centrifugal and axial continuous-flow left ventricular assist devices. J Heart Lung Transplant. 2019;38:352–63.
Kirklin JK, Pagani FD, Kormos RL, et al. Eighth annual INTERMACS report: Special focus on framing the impact of adverse events. J Heart Lung Transplant. 2017;36:1080–6.
Soliman OII, Akin S, Muslem R, et al. Derivation and validation of a novel right-sided heart failure model after implantation of continuous flow left ventricular assist devices: the EUROMACS (European registry for patients with mechanical circulatory support) right-sided heart failure risk score. Circulation. 2018;137:891–906.
Hasin T, Topilsky Y, Schirger JA, et al. Changes in renal function after implantation of continuous-flow left ventricular assist devices. J Am Coll Cardiol. 2012;59:26–36.
Brisco MA, Kimmel SE, Coca SG, et al. Prevalence and prognostic importance of changes in renal function after mechanical circulatory support. Circ Heart Fail. 2014;7:68–75.
Damman K, Testani JM. The kidney in heart failure: an update. Eur Heart J. 2015;36:1437–44.
de By TMMH, Mohacsi P, Gahl B, et al. The European registry for patients with mechanical circulatory support (EUROMACS) of the European Association for Cardio-Thoracic surgery (EACTS): second report. Eur J Cardiothorac Surg. 2018;53:309–16.
Rose EA, Gelijns AC, Moskowitz AJ, et al. Long-term use of a left ventricular assist device for end-stage heart failure. N Engl J Med. 2001;345:1435–43.
Pruijsten RV, Lok SI, Kirkels HH, Klöpping C, Lahpor JR, de Jonge N. Functional and haemodynamic recovery after implantation of continuous flow left ventricular assist devices in comparison with pulsatile left ventricular assist devices in patients with end-stage heart failure. Eur J Heart Fail. 2012;14:319–25.
Estep JD, Starling RC, Horstmanshof DA, et al. Risk assessment and comparative effectiveness of left ventricular assist device and medical management in ambulatory heart failure patients: results from the ROADMAP study. J Am Coll Cardiol. 2015;66:1747–61.
Brouwers C, de Jonge N, Caliskan K, et al. Predictors of changes in health status between and within patients 12 months post left ventricular assist device implantation. Eur J Heart Fail. 2014;16:566–73.
Felix SEA, Ramjankhan FZ, Buijsrogge MP, et al. Outcome of mechanical circulatory support at the University Medical Centre Utrecht. Neth Heart J. 2020;28:210–8.
Tattevin P, Flécher E, Auffret V, et al. Risk factors and prognostic impact of left ventricular assist device-associated infections. Am Heart J. 2019;214:69–76.
Rich JD, Gosev I, Patel CB, et al. The incidence, risk factors, and outcomes associated with late right-sided heart failure in patients supported with an axial-flow left ventricular assist device. J Heart Lung Transplant. 2017;36:50–8.
Glass CH, Christakis A, Fishbein GA, et al. Thrombus on the inflow cannula of the HeartWare HVAD: an update. Cardiovasc Pathol. 2019;38:14–20.
Mehra MR, Salerno C, Naka Y. A tale of the twist in the outflow graft: An analysis from the MOMENTUM 3 trial. J Heart Lung Transplant. 2018;37:1281–4.
Shah P, Tantry US, Bliden KP, et al. Bleeding and thrombosis associated with ventricular assist device therapy. J Heart Lung Transplant. 2017;36:1164–73.
Netuka I, Kvasnicka T, Kvasnicka J, et al. Evaluation of von Willebrand factor with a fully magnetically levitated centrifugal continuous-flow left ventricular assist device in advanced heart failure. J Heart Lung Transplant. 2016;35:860–7.
Muslem R, Caliskan K, Leebeek FWG. Acquired coagulopathy in patients with left ventricular assist devices. J Thromb Haemost. 2018;16:429–40.
Colombo PC, Mehra MR, Goldstein DJ, et al. Comprehensive analysis of stroke in the long-term cohort of the MOMENTUM 3 study. Circulation. 2019;139:155–68.
Frontera JA, Starling R, Cho SM, et al. Risk factors, mortality, and timing of ischemic and hemorrhagic stroke with left ventricular assist devices. J Heart Lung Transplant. 2017;36:673–83.
Yap SC, Ramjankhan F, Muslem R, et al. Ventricular arrhythmias in patients with a continuous-flow left ventricular assist device. J Am Coll Cardiol. 2016;68:323–5.
Baumwol J. “I Need Help”—A mnemonic to aid timely referral in advanced heart failure. J Heart Lung Transplant. 2017;36:593–4.
Felix SE, Martina JR, Kirkels JH, et al. Continuous-flow left ventricular assist device support in patients with advanced heart failure: points of interest for the daily management. Eur J Heart Fail. 2012;14:351–6.
Sidhu K, Lam PH, Mehra MR. Evolving trends in mechanical circulatory support: Clinical development of a fully magnetically levitated durable ventricular assist device. Trends Cardiovasc Med. 2020;30:223–9.
Acknowledgements
The authors are grateful for the use of EUROMACS data to provide the Kaplan-Meier survival analysis of patients with an LVAD in the Netherlands.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
S.E.A. Felix, N. de Jonge, K. Caliskan, O. Birim, K. Damman, M. Kuijpers, L.F. Tops, M. Palmen and F.Z. Ramjankhan declare that they have no competing interests.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Felix, S.E.A., de Jonge, N., Caliskan, K. et al. The role of long-term mechanical circulatory support in patients with advanced heart failure. Neth Heart J 28 (Suppl 1), 115–121 (2020). https://doi.org/10.1007/s12471-020-01449-3
Published:
Issue Date:
DOI: https://doi.org/10.1007/s12471-020-01449-3