Abstract
Background
Four-dimensional (4D) ultrasound scanning (3D real-time mode) can improve the orientation of the anatomy of the area of interest and navigation by controlling the needle position. The objectives of this study were to identify the optimal technique for navigation and to assess clinically the efficacy of 4D ultrasound navigation for epidural anaesthesia at lower thoracic and lumbar levels.
Design
Single-centre case series study was performed.
Methods
Sixteen patients were included. First, conventional 2D scanning was performed, followed by 4D reconstruction, and the basic tissues with high acoustic impedance (bone structures) and available acoustic windows were determined. Movement of the needle was controlled on the sagittal plane in 2D mode and at the same time in 4D mode (3D real-time mode). To improve the visibility of the needle, the 3D reconstruction was rotated during manipulation.
Results
The 4D scanning mode provided 100 % visibility of compact bone tissues and 93 % visibility of the posterior complex. Needle visualisation strongly depended on the rotation of the reconstructed image with the sensor remaining motionless. The needle was redirected in one patient (7 %) because it was in contact with the vertebral lamina. Dilation of the epidural space during saline injection was observed in five patients (36 %). A change in the puncture level was not required any patients; no complications associated with epidural puncture were observed.
Conclusions
Ultrasound navigation in 4D could improve epidural anaesthesia due to the enhanced spatial orientation of the operator. The technique of “position contrast” should be used for reliable needle visualisation.
Sommario
Scopo
L’ecografia in quattro dimensioni (4D) (modalità 3D in tempo reale) può migliorare l’orientamento anatomico nella zona d’interesse e la navigazione, controllando la posizione dell’ago. Gli obiettivi di questo studio sono stati di identificare la tecnica ottimale per la navigazione e valutare l’efficacia clinica della navigazione ecografica 4D per l’anestesia epidurale a livello toracico e lombare inferiore. Sono stati studiati una serie di casi di un singolo centro.
Materiali e Metodi
Sedici pazienti sono stati inclusi nello studio. Inizialmente, è stato eseguito un esame convenzionale 2D, seguito da ricostruzione 4D e sono stati individuati tessuti con alta impedenza acustica di base (strutture ossee) e finestre acustiche disponibili. Il movimento dell’ago era controllato sul piano sagittale in modalità 2D e allo stesso tempo in modalità 4D (modalità 3D in tempo reale). Per migliorare la visibilità dell’ago, la ricostruzione 3D è stata ruotata durante la manipolazione.
Risultati
La modalità di scansione 4D ha permesso il 100 % di visibilità dei tessuti ossei compatti e il 93 % del complesso posteriore. La visualizzazione dell’ago dipendeva fortemente dalla rotazione delle ricostruzioni, con il sensore immobile. L’ago è stato reindirizzato in un paziente (7 %) perché era a contatto con la lamina vertebrale. La dilatazione dello spazio epidurale durante l’iniezione di soluzione salina è stata osservata in cinque pazienti (36 %). In nessun paziente è stato necessario un cambiamento nel livello della puntura; non sono state osservate complicanze associate alla puntura epidurale.
Conclusioni
La navigazione ecografia in 4D potrebbe migliorare l’anestesia epidurale per il miglioramento della risoluzione spaziale offerta all’operatore. La tecnica opera di “contrasto di posizione” deve essere utilizzato per la visualizzazione affidabile dell’ago.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Introduction
Epidural anaesthesia is widely used in clinical practise [1–6]. However, the success of this method of anaesthesia depends on the ability to correctly estimate the placement of the epidural space to puncture and catheterise it. Currently, the most common method for determining the level of puncture and correct needle direction is based on anatomical landmarks, and the depth of the epidural space placement is assessed by the “loss of resistance” test [7]. The level of puncture as determined by anatomical landmarks is an inaccurate technique, with an error rate as high as 70 % [8]. It can be difficult to determine anatomical landmarks in patients with obesity, indeterminate landmarks, and anatomical variations or anomalies, all of which are factors that can complicate epidural anaesthesia [9]. The technical complexities of the epidural puncture method can increase the risks of neurological complications, unsuccessful blocks, and patient dissatisfaction [10–13].
Ultrasound guidance in nerve blocks is widely used in clinical practice. However, the application of ultrasound during neuraxial anaesthesia has remained limited [14] due to the difficulty of visualising the dense structures surrounding the epidural space and spinal canal [15].
The aims of this study were to identify the optimal technique for navigation and clinical evaluation of 4D (3D real-time mode) ultrasound navigation efficacy during epidural anaesthesia at the lower thoracic and lumbar levels and to assess clinically the efficacy of 4D ultrasound navigation for epidural anaesthesia at these levels.
Methods
This study was approved by the local ethics committee of First Moscow State Medical University, and written informed consent was obtained from all of the subjects. The study included 16 patients undergoing abdominal surgery.
Equipment
The ultrasound scan was performed using a Voluson-I (GE HealthCare, USA), equipped with a multi-frequency volume convex transducer of 4–8.5 MHz (RAB4-8-RS). Volumetric scanning was performed by vibration of the linear plane using the transducers (mechanical scanning principle).
Patients
All the patients received midazolam 0.05–0.07 mg/kg as premedication. Venous access was secured in all of the patients before the epidural puncture for the infusion of crystalloid solutions, and non-invasive monitoring of the patients’ blood pressure, pulse oximetry, respiration rate, and electrocardiogram (ECG) was maintained. The patients were placed in the lateral position.
Scanning
Before the 4D reconstruction, traditional 2D scanning was performed. The placement of the lumbar epidural was identified by counting up the vertebrae from the sacrum; the thoracic epidural placement was determined by counting down from the twelfth rib. Thereafter, the outlines of the intervertebral joints, vertebral plates, and transverse processes were defined. The dorsal complex was identified by the paramedian oblique access (ligamentum flavum, epidural space, and posterior dura mater), and if possible, the front complex was also determined (anterior dura mater, posterior longitudinal ligament, and vertebral body). Next, a reconstruction frame was constructed. During the 4D reconstruction, tissues with high acoustic impedance (bone structures) and an acoustic window were determined. Because there were no data on the effects of the gel on the neuraxial structures, the transducer was covered with a sterile material to maintain the sterile field; the gel was placed between the transducer surface and the material, and sterile saline, instead of the gel, was applied to the skin.
Puncture
Two types of needles were used to perform the epidural punctures: Tuohy 18 G needles with ultrasound notches (Tuohy Sono, Pajunk Gmbh) and standard Tuohy 18 G epidural needles (Smith). All attempts at ultrasound scanning and needle insertion were made by one operator. The needle advancement was monitored from two perspectives at the same time: on the sagittal plane in 2D mode, and in volumetric reconstruction mode. The “loss of resistance” test and catheter insertion were conducted by the second anaesthetist. If resistance due to bone contact was encountered during needle insertion, the operator redirected the needle. The number of needle insertions (including needle redirections without removal) necessary to achieve the epidural space in the course of the anaesthesia procedure was counted.
Anaesthesia
A positive “loss of resistance” test and a negative aspiration test were required before inserting the catheter into the epidural space 5 cm from the tip of the needle in the cranial direction. To avoid intrathecal catheter placement, a standard test dose of lidocaine was infused into the epidural catheter; thereafter, epidural anaesthesia was performed in accordance with the planned operation.
Data analysis
This study was not powered for statistical analysis. Nevertheless, we calculated the 95 % confidence interval for the probability of success visualisation of needle movement in 4D mode.
Results
Epidural anaesthesia was performed under 4D ultrasound guidance in 16 patients. The characteristics of these patients are presented in Table 1.
During 4D scanning, 100 % visualisation of the dense bone structures and 93 % visualisation of the neuraxial structures (the dorsal complex) were accomplished. The “loss of resistance” test was successful in 100 % of cases. The epidural catheter was successfully inserted in 100 % of cases. Some resistance during catheter insertion occurred in one patient. The needle was redirected in one patient (7 %) because it encountered the vertebral bone plate. Dilatation of the epidural space during saline administration was observed in five patients (36 %). The steps of the procedure are presented in Table 2. Needle visualisation strongly depended on the rotation of the reconstructed image while the sensor remained motionless (Table 3).
None of the patients required changes in the level of puncture placement, and no perforations of the dura mater or any other complications related to the epidural puncture were noted. The clinical efficacy during the procedure was sufficient; none of the patients had any complications associated with the epidural anaesthesia technique.
Discussion
There are limited data about real-time ultrasound guidance, i.e., when the needle is inserted “in-plane” with the ultrasound beam [16–18]. This method has two limitations: first, the two-dimensional (2D) image obtained provides data about a very restricted puncture area; and second, the needle must be directed along the narrow plane of the ultrasound beam (1–2 mm). Thus, to realise the potential of this method fully, enhanced requirements for performing blockade techniques and personal experience must be employed. The dimensional representation of the image can improve orientation in the anatomical zone of interest, including guiding of the needle position in space.
Only one pilot experimental study assessing the potential of 4D ultrasound (3D real-time mode) has previously been conducted. The authors demonstrated the concept of vertebral structure sonography with dimensional reconstruction, but the needle visualisation, and thus real-time guidance, was complicated. Among their reasons for these complications, the authors listed the embalming of the corpses used, the high density of the surrounding structures, and the low transducer resolution [19].
Optical mapping structures in 4D reconstruction are based on the layers of the scanned 2D images and, therefore, they do not fundamentally differ in interpretation. The border of tissue (e.g., bone/muscle) perpendicular to the sensor appears as a hyperechoic area (lamina, articular and transverse processes). Areas located at an acute angle are visualised slightly less clearly. Areas lying at an obtuse angle relative to the sensor could appear as anechoic areas (spinous processes) (Fig. 1). In contrast to traditional 2D ultrasound, with which only a hyperechoic strip is visible, and the operator must form a coherent picture virtually moving the sensor on the skin on the transverse plane, with 3D imaging, volumetric areas appear. Because 3D (4D) modelling consists of mathematical processing of a multitude of dimensional images, for the best results, good 2D image scanning should be achieved, and the appropriate soft program should be used. The front and dorsal complexes appear as hyperechoic structures located under the lamina. The size of these structures could predict the size of the acoustic window and, therefore, the complexity of the puncture of the selected interval. The resolution of the equipment did not allow for differentiation of the equipment structures included in the dorsal (the ligamentum flavum and posterior epidural space of the dura mater) or especially the frontal system (front dura of the posterior longitudinal ligament and the posterior surface of the vertebral body).
An essential benefit of 3D modelling, except for volumetric representation, is that it can rotate an image at any angle, approximate and move away it, and focus on areas that are of interest at each time point, without changing the position of the transducer and needle (Fig. 2). This benefit allows the sensor not to follow the needle, to move over the skin, and to choose the most beneficial foreshortening at every moment for better control of the needle tip. We noted from this perspective that regarding the image of the needle, seen as a hyperechoic strip superimposed on the underlying hyperechoic formation, its visibility could deteriorate significantly. Therefore, for reliable control of the needle in the space, the operator should dynamically change the angle and zoom navigation. To clarify the position of the needle relative to the spinous and articular processes, the “above view” is optimal (from the sensor, Fig. 3). Then, to verify the depth of the needle passing through the tissues, until it is near the interlaminar space, the above-side view is optimal, with rotation of the image so that the needle’s hyperechoic shadow is superimposed on the acoustic shadow of the spinous process, which clearly stands out as a relatively large anechoic structure (Fig. 4). When approaching a planned interval, control of the needle tip from this angle is difficult. For maximum detail, the operator should reasonably rotate the image on the control from the medial side, “through the spinous process”, focusing the view on the target interspace (Fig. 5). From this position, the operator can see and directly contact the needle tip with the dorsal epidural space. Thus, the method of “positional contrast”—the dynamic change in viewing angle of reconstruction in real time for the best visual inspection of the needle position, combining the hyperechoic shadow of needle with underlying relatively hypoechoic forms—is optimal for puncture control during each stage.
In our study, an above view was successful in 30 % of cases, and the technique of “dynamic contrast” was effective in 85–100 % of attempts (Table 3).
To perform anaesthesia, the “Epi Long Soft Sono Set” (Pajunk Gmbh) and “Minipack” (Smith) sets for epidural anaesthesia were used. The “Minipack” consists of an 18 G Tuohy needle and a polyether block amide 110-cm catheter. The “Epi Long Soft Sono Set” is a TuohySono needle set intended for ultrasound guidance. The distal end of the needle has special reflectors that are designed to reflect ultrasonic waves with maximum intensity under different needle tilts to improve ultrasound visualisation of the needle. A catheter reinforced with a steel wire to protect against inflection was also used with these needles.
The notched needles with improved ultrasound beam visibility had the same visibility over their entire lengths as the Tuohy needles. Although the circularly arranged incisions on the needle created the optical illusion of needle thickening in some cases, they provided better control of the depth of needle insertion (Fig. 4).
The epidural space is filled with adipose tissue, and it has low acoustic impedance, which was decreased after saline administration. Despite using a reinforced catheter, attempts to visualise the catheter passing into the accessible part of the epidural space were unsuccessful. This inability was potentially due to the insufficient resolution power of the device, the small number of observations, or operator inexperience with dimensional ultrasound.
The need for the participation of two anaesthesiologists at the initial stage of development of the methods was dictated by the necessity of reliable verification of the epidural space with the traditional method (LOS). In the absence of a special device ensuring the identification of the epidural space [17], one anaesthesiologist cannot perform gentle advancement of the needle, control on 4D imaging for device reliability, and LOS. Participation of a specialist will be possible to consider further enhanced navigation by ultrasonic techniques and the use of special devices to facilitate puncture.
Conclusion
4D ultrasound guidance can facilitate epidural anaesthesia by improving the dimensional orientation of the operator. For reliable needle visualisation, a “positional contrast” method should be used, and special needles for ultrasound guidance can facilitate the control of the depth of needle insertion. Further studies are required to assess the efficacy of dimensional ultrasound guidance during neuraxial blockade in patients with expected difficulties, such as obesity or altered anatomy.
References
Jørgensen H, Wetterslev J, Møiniche S, Dahl JB (2000) Epidural local anaesthetics versus opioid-based analgesic regimens on postoperative gastrointestinal paralysis, PONV and pain after abdominal surgery. Cochrane Database Syst Rev 4: CD001893
Holte K, Kehlet H (2001) Epidural analgesia and risk of anastomotic leakage. Reg Anesth Pain Med 26:111–117
Block BM, Liu SS, Rowlingson AJ, Cowan AR, Cowan JA, Wu CL (2003) Efficacy of postoperative epidural analgesia: a meta-analysis. JAMA 290:2455–2463
Werawatganon T, Charuluxanun S (2005) Patient controlled intravenous opioid analgesia versus continuous epidural analgesia for pain after intra-abdominal surgery. Cochrane Database Syst Rev 1: CD004088
Marret E, Remy C, Bonnet F, Postoperative Pain Forum Group (2007) Meta-analysis of epidural analgesia versus parenteral opioid analgesia after colorectal surgery. Br J Surg 94:665–673
Pöpping DM, Elia N, Marret E, Remy C, Tramèr MR (2008) Protective effects of epidural analgesia on pulmonary complications after abdominal and thoracic surgery: a meta-analysis. Arch Surg 143:990–999
Barash P, Cullen BF, Stoelting RK, Cahalan M, Stock C, Ortega R (2013) Clinical anesthesia, 7th edn. Lippincott Williams & Wilkins, Philadelphia
Furness G, Reilly MP, Kuchi S (2002) An evaluation of ultrasound imaging for identification of lumbar intervertebral level. Anaesthesia 57:277–280
Grau T, Leipold RW, Conradi R, Martin E (2001) Ultrasound control for presumed difficult epidural puncture. Acta Anaesthesiol Scand 45:766–771
de Filho GR, Gomes HP, da Fonseca MH, Hoffman JC, Pederneiras SG, Garcia JH (2002) Predictors of successful neuraxial block: a prospective study. Eur J Anaesthesiol 19:447–451
Auroy Y, Narchi P, Messiah A, Litt L, Rouvier B, Samii K (1997) Serious complications related to regional anesthesia: results of a prospective survey in France. Anesthesiology 87:479–486
Harrison DA, Langham BT (1992) Spinal anaesthesia for urological surgery. A survey of failure rate, postdural puncture headache and patient satisfaction. Anaesthesia 47:902–903
Horlocker TT, McGregor DG, Matsushige DK, Schroeder DR, Besse JA (1997) A retrospective review of 4,767 consecutive spinal anesthetics: central nervous system complications. Perioperative Outcomes Group. Anesth Analg 84:578–584
Chin KJ, Karmakar MK, Peng P (2011) Ultrasonography of the adult thoracic and lumbar spine for central neuraxial blockade. Anesthesiology 114:1459–1485
Tran D, Kamani AA, Lessoway VA, Peterson C, Hor KW, Rohling RN (2009) Preinsertion paramedian ultrasound guidance for epidural anesthesia. Anesth Analg 109:661–667
Grau T, Leipold RW, Fatehi S, Martin E, Motsch J (2004) Real-time ultrasonic observation of combined spinal-epidural anaesthesia. Eur J Anaesthesiol 21:25–31
Karmakar MK, Li X, Ho AM, Kwok WH, Chui PT (2009) Real-time ultrasound-guided paramedian epidural access: evaluation of a novel in-plane technique. Br J Anaesth 102:845–854
Tran D, Kamani AA, Al-Attas E, Lessoway VA, Massey S, Rohling RN (2010) Single-operator real-time ultrasound-guidance to aim and insert a lumbar epidural needle. Can J Anaesth 57:313–321
Belavy D, Ruitenberg MJ, Brijball RB (2011) Feasibility study of real-time three-/four-dimensional ultrasound for epidural catheter insertion. Br J Anaesth 107:438–445
Acknowledgments
The author wishes to thank GE Healthcare Russia, in particular Alexey Kiselev, for providing the ultrasound unit.
Conflict of interest
The authors declare that they have no conflict of interest.
Ethical approval
All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.
Informed consent
Informed consent was obtained from all individual participants included in the study.
Author information
Authors and Affiliations
Corresponding author
Electronic supplementary material
Below is the link to the electronic supplementary material.
Supplementary material 1 (MP4 5394 kb)
Supplementary material 2 (MP4 7126 kb)
Supplementary material 3 (MP4 9420 kb)
Rights and permissions