In this review, we provide a comprehensive description of the principles behind recent technological innovations, elucidate the current evidence, and point to specific fields of clinical care in which adoption of the most advanced technology, without careful thought, may put the previously established clinical role and benefits of PET at risk. In doing so, the authors will not argue that the introduction of new PET technology is in any way undesirable, but rather that during this process more emphasis should be placed on its ultimate effect on patient-relevant outcomes―on the possible limitations as well as the desired advantages.
This review belongs to a two-part series of reviews published in EJNMMI addressing the pros and cons of new PET technologies. The complementary review by Nicolas Aide et al. covers the pros [1].
How should we evaluate the clinical benefits of PET and PET technologies?
The fundamental role of PET in clinical oncology is firmly established. This recognition of its clinical value has been achieved stepwise over time, by providing patients and clinicians with crucial additional information concerning tumor characterization, biology, and metastatic spread. However, this has been paralleled by concerns that in the average patient, PET may lead to overdiagnosis and overtreatment as a more sensitive imaging method. Before PET was introduced and before the concept of evidence-based medicine was widely adopted in the 1990s [2], there was generally no demand for a systematic and standardized assessment of the clinical benefit of any one imaging procedure. However, in evaluating the importance of PET imaging and PET technology today, medical societies and regulatory authorities ask for proof of actual clinical benefit to the individual patient―just as they do for development and approval of therapeutic drugs.
Fryback and Thornbury have proposed six hierarchical levels of evidence for the clinical efficacy of diagnostic imaging (Fig. 1) [3]. In accordance with this system, for regulatory purposes, evidence of improved patient outcomes usually implies level 5 evidence. This requires proof of an advantage with regard to patient morbidity (e.g., as a consequence of a reduced rate of invasive procedures or futile surgeries), mortality, or quality of life [4]. In specific cases, proof of superior diagnostic accuracy may be sufficient if there is prior evidence from (randomized controlled) trials that the improvement in lesion/disease detection will lead to changes in treatment management that benefit patient outcome (e.g., early detection of disease relapse) [5]. However, such “linked evidence” is often difficult to establish, either because the study quality is too low or because there is no effective treatment to offer the patient [4, 6]. In the initial evaluation of a new test, patient-relevant endpoints may be replaced by surrogate endpoints if their direct relationship has been demonstrated based on biological plausibility and empirical evidence [7].
In a different scenario, use of a new diagnostic method can be justified if it has shown equal diagnostic accuracy in a well-designed study and provides other advantages over the standard method used [5] such as lower invasiveness, lower radiation exposure, or lower costs.
Success stories in nuclear medicine: Introduction of PET in oncology
Prospective randomized controlled studies are regarded as the optimal method to demonstrate evidence for a clinical benefit by regulatory authorities. Accordingly, such studies have been major drivers in the past for the broad adoption of new radiopharmaceuticals and nuclear medicine techniques in routine clinical practice. However, as PET radiopharmaceuticals are often not patented, the perceived costs of patient scanning are high, and radiopharmaceutical availability may be limited, it remains especially difficult to find industrial sponsors willing to cover the costs of such expensive trials. Furthermore, not every clinical application of PET is equally susceptible to variability in imaging technology and requires highest-level evidence for its safe use. Nevertheless, there are fields in which seminal studies have been able to provide high levels of evidence for the additional value of PET, and selected studies are briefly outlined in the following paragraphs. The third section discusses the possible effects of “new” PET technology in these areas of successful application.
PET in lung cancer
In a prospective, randomized controlled trial, [18F]fluorodeoxyglucose (FDG)-PET for staging patients with non-small cell lung cancer (NSCLC) revealed extensive mediastinal lymph node or distant metastases more frequently than conventional staging methods alone. Patients who did not undergo PET had progression or relapse within a year after surgery in 41% of cases as compared to 21% in patients who had undergone PET for staging. Thus, the addition of PET to the conventional workup prevented unnecessary surgery in one out of five patients (level 5 evidence) [8]. This study was confirmed by a randomized controlled trial by Fischer et al. [9] and led to the introduction of FDG-PET into standard protocols for the care of patients with NSCLC and potentially curative treatment [10, 11].
PET in prostate cancer
[68Ga]Ga-prostate specific membrane antigen (PSMA)-11 was recently approved by the US Food and Drug Administration for PET in patients with prostate cancer [12]. This was based on two prospective studies showing that PSMA-11 PET has a high positive predictive value in assessment of lymph node metastases prior to surgery and in the detection of lesions in biochemical relapse. In either situation, this could have an important influence on therapeutic decisions (surrogate endpoint) [12, 13]. Hofman et al. recently reported a randomized controlled trial in 302 men with first diagnosis of prostate cancer before planned surgery or radiotherapy with curative intent. The authors showed that PSMA-11 PET/CT resulted in significantly more relevant changes in treatment management than conventional imaging (surrogate for level 5 evidence) [14].
PET in lymphoma
FDG-PET has become the standard for staging and response assessment in patients with Hodgkin lymphoma (HL) and FDG-avid non-Hodgkin lymphoma. In the case of HL, the analysis of prospective, randomized controlled trials showed that bone marrow biopsy could be safely omitted if involvement had already been excluded by FDG-PET (level 5 evidence) [15,16,17]. In advanced stage HL, the prospective, randomized controlled HD15 trial demonstrated that radiotherapy can be restricted to 11% of patients after effective first-line chemotherapy based on the presence or absence of FDG-avid residual disease (level 5 evidence) [18]. Subsequent randomized controlled trials showed that a favorable PET result also allows a reduction of chemotherapy in advanced-stage HL [15, 19], may allow further de-escalation in selected patients in early stages of HL [20, 21], and can reliably guide selective administration of consolidative radiotherapy in advanced-stage diffuse large B-cell lymphoma (each level 5 evidence) [22].
Summary
In the tumor entities and clinical settings described above, PET has been clearly shown to have a direct, beneficial influence on patient outcomes. In this way, nuclear medicine has succeeded in contributing to the treatment and cure of many oncological patients. However, the examples presented here have one important feature in common: the added benefit of PET imaging was primarily a consequence of its qualitative value, namely the functional information provided by the radiopharmaceutical and based on visual assessment, i.e., information on the presence of lesions, their location, and their spread (staging), not of its quantitative accuracy or precision or because of incremental improvements in image quality in an existing modality. It should be noted, however, that lack of imaging procedure guidelines and PET/CT system performance harmonization might have hampered the success of quantitative reads. Moreover, quantitative reads were mainly restricted to simple lesional uptake metrics, such as the SUVmax, while ignoring information on the location and spread of the disease. Radiomics analysis and quantitative dissemination features [23] can capture information on uptake variability (intra- and inter-lesional) as well as the spatial spread of the disease over the body and may thus provide a better and more clinically relevant quantitative surrogate for the disease aspects typically considered by visual inspection, as explained above.
“New” PET technologies
What is meant by “new” technologies?
New technologies for the purpose of this review comprise PET image reconstruction algorithms with point spread function (PSF) compensation and Bayesian penalized likelihood (PL) reconstruction algorithms as well as “digital PET” (i.e., using silicon photomultipliers (SiPM) instead of photomultiplier tubes (PMT)).
Point spread function reconstruction
The most common “conventional” PET reconstruction algorithm, ordered subset expectation maximization (OSEM), is iterative and involves a stepwise approximation of the tracer distribution that would have resulted in an image that is closest to the observed image. For this purpose, OSEM employed in PET imaging includes scatter and random correction, attenuation correction, normalization, dead time, and decay correction. This allows more accurate modeling of the observed distribution, especially in the low activity background [24].
PSF reconstruction algorithms are commonly based on the OSEM principle, but include an additional corrective term, which compensates for the scanner’s specific PSF. The latter describes the system’s depiction of point sources depending on their location in the field of view (FOV) [25, 26]. Due to the finite size of the detector elements, point sources at the periphery of the FOV are not perceived as a Gaussian activity distribution (as―ideally―in the FOV center) but as a skewed and degraded distribution curve. In non-PSF PET, this results in a steady decrease in spatial resolution from the FOV center to its periphery, which is compensated for by PSF reconstruction [26]. Notably, PSF reconstruction usually refers to the correction for F-18 while optimal PSF compensation for other radionuclides would require integrating their different PSF based on specific positron ranges [27, 28].
Bayesian penalized likelihood reconstruction
Standard OSEM reconstruction may not fully converge, i.e., the estimated probability of iterated images may never reach its maximum [29], especially as the iterative process is often stopped earlier (using a small number of iterations) to prevent excessive noise. An additional post-reconstruction filter is usually applied to achieve more visually appealing, smoothed images [30], which further counteracts the anticipated convergence of focal activity maxima.
PL reconstruction is also iterative, but the endpoint of iterations is determined by a penalization/regularization term instead of a fixed number of iterative steps. This term increases with increasing noise and regulates the overall function so that it reaches full convergence before developing excessive noise. Its relative strength is determined by a user-defined penalization/regularization factor β, which sets the tolerated level of noise in the final image (a higher β will result in lower noise level) [29, 31]. Another fundamental difference to OSEM is that currently applied PL reconstruction algorithms account for the relative differences between neighboring voxels, which enables a differentiated approach to be taken for voxels in the (ideally homogenous) background as compared to voxels in a “hot” or “cold” lesion [31, 32]. This aims at providing low background variability/noise while still achieving high contrast recovery (CR) of focal activity maxima. Since PL includes PSF compensation, the characteristics of this relative difference function regulate the amount of edge preservation (or smoothing) at sharp transitions between different activity levels. This balance is predefined by the developer [31,32,33]. PL reconstruction algorithms use the Bayesian principle by integrating estimates about the physical properties of the unknown image as a prior probability [34]. They can also be characterized as block sequential algorithms, as the sinogram is separated into blocks during the iterative process [29].
Silicon photomultipliers
The photomultiplier converts the photon emitted from the scintillation crystal into an electronic signal. The ideal photomultiplier would be of small size, provide full coverage of the scintillation crystal area (with no dead space between channels), be perfectly sensitive to single photons without the interference of dark current, and provide high single photon timing resolution <50 ps [35]. Compared to conventional photomultiplier tubes (PMT), SiPM in state-of-the-art PET/CT systems are smaller (enabling higher spatial resolution) and provide up to 100% coverage of the crystal area, as well as offering high sensitivity, low noise, and fast timing resolution [35]. PET/CT systems equipped with SiPM are commonly named “digital PET” although, strictly speaking, only SiPM with digital photon counting are fully digital [36]. Most current systems use analog SiPM, which generate a Geiger avalanche with a current signal that is proportional to the number of activated microcells [35, 37, 38]. Moreover, while all these current PET/CT systems use lutetium-based scintillation crystals, the SiPM design varies. This results in detector area coverage ranging from only 40% [39] to a full 100% [36, 40, 41], and a timing resolution ranging from 382 ps [42] down to 214 ps [40] as well as high ratios of SiPM channels per crystal element [39, 42] to the ideal 1-to-1 coupling of crystal to SiPM elements [36, 41].
What was expected of these new technologies?
All three technological innovations aim at improving image quality (Table 1) and were introduced as game-changing technologies. The increase in image quality was expected to greatly increase diagnostic accuracy and giving clinicians and patients greater assurance that lesions would be detected with optimal quantitative accuracy [43,44,45,46]. Equally, the improvement in image quality (namely in noise properties) would, it was hoped, translate directly into a shortening of acquisition time by up to 90% [36, 47] or reduction in injected activity of up to 50% [36, 48,49,50]. Notably, this approach could be balanced in a way that can retain the accustomed image characteristics while achieving higher patient comfort or lower effective dose.
Does the current evidence support new PET technologies?
PSF reconstruction
PSF reconstruction improves the reconstructed spatial resolution [51, 52] and CR in small lesions [52,53,54,55]. However, it introduces so-called edge artifacts (edge elevations or Gibbs’ artifacts) that present as an overshoot (and undershoot) of activity at the sphere surface (illustration in Fig. 2 and [52]). Consequently, the true activity concentration can be overestimated over a range of different sphere sizes [52, 53, 56]. If sphere or lesion contrast is low, these edge artifacts are practically absent, but so is the potential improvement in reconstructed spatial resolution [52, 57]. To date, there is no consistent solution to these edge artifacts in conventional reconstruction algorithms, apart from applying a post-reconstruction filter to smooth the final data (and thereby negating the improved spatial resolution) [56, 58, 59].
Several studies have shown that the increasing lesion contrast improves lesion conspicuity/detectability (level 1 evidence; Table 2). However, lesion detectability does not necessarily imply improved sensitivity, specificity, or accuracy (see section “Effects on PET success stories”). Two studies about diagnostic accuracy in lymph node staging (level 2 evidence) have shown that PSF increases sensitivity while (slightly) reducing specificity [63, 64].
PL reconstruction
Several phantom and patient studies have shown that PL reconstruction can improve the trade-off between image noise and accurate CR of focal activity maxima compared to OSEM reconstruction. This could be interpreted as an improvement in overall image quality (level 1 evidence; Table 3). However, these studies also showed that the net benefit of PL reconstruction regarding image quality depends strongly on the β value chosen as well as the reconstruction parameters selected for the comparative conventional reconstruction algorithm [72]. A single study of level 2 evidence evaluated the diagnostic accuracy of OSEM with time of flight (TOF) compared to PL reconstruction, using a common β value of 400. In 121 histologically proven pulmonary lesions (either lung cancer, lung metastases, or benign lesions), diagnostic accuracy of visual assessment was unaffected by PL reconstruction in lesions ≤10 mm as well as >10 mm. Using an SUVmax cut-off optimized for each reconstruction, diagnostic accuracy was higher with PL reconstruction in lesions <10 mm. This was a result of increased sensitivity while specificity tended to decrease―a similar pattern as observed with PSF reconstruction [94].
Identification of a diagnostic benefit of PL reconstruction is further complicated by the requirement that an optimized β value must be obtained for different clinical scenarios and radiopharmaceuticals or activity levels [79,80,81, 86, 96]. To achieve a significant advantage in reconstructed spatial resolution and small lesion conspicuity may require β values as low as 150 [72, 86]. This will result in comparably high image noise, which may render the method inadequate for whole body imaging [72] unless PET systems with especially high sensitivity are used [42]. In clinical practice, β values between 300 and 600 have been recommended for whole body assessment with 18F- or 68Ga-labeled compounds [72, 73, 77, 79, 81,82,83,84].
Few patient studies have directly compared the image quality of PL reconstruction and OSEM-based reconstruction in order to evaluate their potential to reduce acquisition time. A potential to reduce acquisition time by 25 to >50% has been reported [81, 93]. Other authors have only investigated PL reconstruction at different β values [80, 84]. It is generally recognized that a reduction in acquisition time could equally well translate into reduced injected activity [97] and therefore into minimized radiation exposure. However, the studies cited here have mostly used noise level as a surrogate for image quality (evidence level 1). Reliable evidence of a secure reduction in acquisition time or activity would require proof of non-inferiority in diagnostic accuracy or other relevant clinical outcomes and, where relevant, non-inferiority in quantitative accuracy.
SiPM
Intra-individual comparison of SiPM and PMT in patients is complicated by the requirement to perform separate PET examinations as part of a prospective study, possibly combined with a higher radiation exposure due to a second CT scan. If both scans are performed using a single injection of activity, unbiased comparison implies a randomized scan order, because the later time point may be associated with systematically higher standardized uptake values (SUV), higher sensitivity yet higher image noise. Some previous studies have employed such randomized order, but no study beyond evidence level 1 has been performed in an unbiased protocol (Table 4). So far, no study has reported any possible disadvantages of SiPM technology regarding image quality or lesion detectability, which suggests that SiPM may constitute a systematic improvement over PMT. However, the actual clinical benefit of SiPM has not been studied. Further investigation is also required to assess whether there is a general, significant benefit associated with SiPM technology or whether this advantage is restricted to specific PET detector designs with high relative coverage of the detector area and especially favorable TOF characteristics [35, 38]. Zhang et al. estimated an increase in the effective sensitivity by a factor of 1.3 to 5.5 for activities of 7.4 to 337 MBq in a commercial SiPM PET system with a TOF resolution of 322 ps compared to a PMT system with comparable axial FOV length and 550 ps [47]. The anticipated further improvements in TOF resolution associated with SiPM [35] enable a further increase in effective sensitivity. The improved TOF performance of SiPM-based systems will likely also show benefit in image quality for patients with a regular weight (<90 kg) [105, 106]. This could be translated into an equivalent reduction in injected activity or acquisition time at equal image quality. Alternatively, the TOF gain could be exploited for an image quality or SNR gain equivalent to the square root of the relative improvement in timing resolution [105, 107].
Effects on PET success stories
In section ““New” PET technologies,” we discussed that the level of evidence for the clinical efficacy of new PET technologies is generally low, especially when compared to the body of evidence for the overall clinical value of PET imaging as outlined in section “Success stories in nuclear medicine: Introduction of PET in oncology.” In such specific clinical settings, the sought-for advantage of new technologies in lesion contrast and lesion detectability must be weighed against the risk of changing diagnostic and therapeutic strategies that have proven to be beneficial for patients, but where the evidence-base was established using established PET technology. The more susceptible the clinical consequence is to changes in image quality and lesion contrast, the more thorough the investigation of any potential undesired effects should be. Strictly speaking, the introduction of new methods and technologies would presuppose that they have shown non-inferiority―if not superiority―in the clinical endpoints routinely addressed by PET.
Lung cancer
In patients with lung cancer, a possible advantage for patient outcomes could be established if new PET technologies helped to obviate invasive diagnostic procedures by improving differentiation between benign and malignant pulmonary lesions or by increasing the reliability of non-invasive lymph node staging. Furthermore, such imaging could benefit patients if futile surgery could be avoided through more accurate distinction of locally unresectable from resectable disease or through improved detection of otherwise unsuspected distant metastases.
However, there are currently no data on the effects of new PET technologies on lung cancer staging beyond level 2 (diagnostic accuracy). As described in section ““New” PET technologies,” PL reconstruction using a β of 400 did not improve diagnostic accuracy in the characterization of pulmonary lesions [94]. As reported by Lasnon et al., PSF reconstruction may improve sensitivity in lymph node staging; however, despite only a slight simultaneous increase in false positive findings, this could ultimately negate a potentially beneficial effect on lymph node staging [63].
The effect on patient-relevant outcomes remains equally unclear. In a recent single-center analysis [108], two PET systems were compared for pretherapeutic thoracic lymph node staging with FDG-PET/CT in patients with NSCLC, one scanner using PMT and OSEM with TOF reconstruction while the other was equipped with analog SiPM and PL reconstruction (β = 450). In addition to diagnostic accuracy per lymph node station, the potential to omit confirmatory invasive procedures (transbronchial biopsy) by modifying the diagnostic algorithm was investigated (surrogate endpoint for level 5 evidence). However, the analysis was retrospective, and the two scanners were used in separate patient cohorts. Therefore, neither intrapatient comparison nor randomization was available (although clinical key parameters were comparable in the two cohorts). Diagnostic accuracy of PET/CT per lymph node station was similar in the conventional scanner cohort (74.3%; n = 448 lymph node stations) and “digital” scanner cohort (79.8%; n = 252; chi-squared test, p = 0.1). More importantly, the frequency of invasive procedures that could have been avoided was similar in both cohorts (79.2% vs. 82.1%; p = 0.75) [108]. Consequently, in this context, SiPM and PL reconstruction did not appear to be superior. However, prospective intra-individual comparison or a randomized design would be required for a definite conclusion.
Prostate cancer
In patients with prostate cancer, a higher sensitivity and diagnostic accuracy of PSMA-PET are eagerly anticipated considering its crucial role in detecting small PSMA-positive lesions in initial staging or in biochemical relapse. PSMA-guided treatment of single-lesion or oligometastatic relapse is being increasingly applied and may achieve favorable outcomes [109,110,111,112]. In this context, improved accuracy in detecting such lesions may result in superior progression-free survival after targeted treatment. In PSMA-PET, the relatively low injected activity can result in unfavorable noise levels, especially in overweight patients, which can complicate the detection of small lesions. Different groups have demonstrated an improvement in image quality for PSMA-PET using PL reconstruction (level 1 evidence) [81, 96, 113]. Alberts et al. recently compared a PMT-based and an analog SiPM-based scanner for [68Ga]Ga-PSMA-11 PET/CT in two matched retrospective cohorts with biochemical relapse of 88 patients each. Images from the conventional scanner were reconstructed with OSEM while the SiPM scanner used OSEM with PSF and TOF. Detection rates for both malignant (local recurrence, lymph nodes, bone) and benign lesions (e.g., ganglia or urinary tract activity) were about two times as high in the SiPM cohort compared to the PMT cohort. Pathological scans were significantly more frequent in the SiPM cohort compared to the conventional cohort in patients with prostate specific antigen (PSA) <2.0 ng/ml (level 1 evidence). Based on phantom measurements, the authors further showed that lesion contrasts and lesion discernibility were higher with the SiPM-based scanner especially in smaller lesions with 0.25 to 0.5 ml [98]. Such results constitute an important first step toward understanding the effect of new PET technology on patient care in prostate cancer and the emerging role of PET imaging in this context. However, as Fig. 1 illustrates, we are still a long way from demonstrating higher lesion counts as proof of a true benefit to patient outcomes.
Lymphoma
The use of FDG-PET in lymphoma has brought the highest-level evidence of its benefit to patients. The unique feature of PET imaging in these patients is that it provides a more reliable assessment of chemosensitivity during treatment and residual viable disease after treatment (especially after induction chemotherapy) than clinical criteria and CT. More reliable assessment allows adoption of a tailored, less intensive, and less toxic treatment for the majority of patients who respond to therapy. Demonstrating this direct connection between the occasionally subtle PET findings and fundamental treatment decisions has required laborious and large-scale prospective studies as well as the development and acceptance of common criteria for PET interpretation.
Between 2003 and 2008, the HD15 trial for advanced stage HL showed that radiotherapy after effective chemotherapy can be restricted to the 11% of patients who showed PET positive residual lesions [18]. The cut-off used for PET positivity was the uptake in mediastinal blood pool structures (“Deauville score 3”) [114]. In the subsequent HD18 trial for advanced stages starting in 2010, the restriction of radiotherapy to PET positive residua was the new standard carried over from HD15. Unexpectedly, the proportion of patients showing PET positive residua and consequently undergoing radiotherapy increased from 11% in HD15 to 30% in HD18. What went wrong? In HD15, mostly stand-alone PET scanners were used, whereas by the time of HD18, technology had improved, and combined PET/CT scanners had become the standard. The practical solution for keeping the rate of patients undergoing radiotherapy close to 11% was to switch the visual cut-off for PET positivity to uptake higher than that of the liver (“Deauville score 4”) [15].
Later, Barrington et al. also reported a shift toward more positive reads for PET studies with subsequent clinical consequences [115]. This shift was found to coincide with the introduction of a new generation of PET/CT systems with PSF reconstruction. A systematic increase in Deauville scores can be expected as PSF increases the uptake, especially in small lesions [53, 57] such as residual tumor tissue but not in large, homogenous organs such as the liver and blood pool [60]. Reliable use of the well-established Deauville score could thereby be hampered. Enilorac et al. [71] compared PET scans in diffuse large B-cell lymphoma reconstructed with PSF by applying either no post-reconstruction filter or a Gaussian filter that complied with the EANM Research Ltd. (EARL) harmonization standard [116, 117]. The authors concluded that the influence of PSF on imaging results was limited, because major discordances (PET positive vs. negative) at interim staging affected only 5% of all cases. However, focusing on cases with Deauville score 3 based on the EARL-compliant reconstruction, 4 of these 22 patients (18%) with residual tissue were given a divergent rating of Deauville score 4 using the unfiltered PSF reconstruction [71]. Similarly, 3 of 18 patients (17%) with Deauville score 3 were upgraded to Deauville score 4 by unfiltered PSF reconstruction in end-of-treatment scans [118]. Ly et al. used the PET data from 52 patients with lymphoma to compare PL reconstruction (β = 500; compliant with the updated EARL specifications [56, 119]) with OSEM (without PSF or TOF; compliant with the first EARL standard). PL reconstruction led to a Deauville score 4 in 4 of 31 patients (13%; scoring based on the SUVmax) or 4 of 30 patients (13%; SUVpeak) who were rated as Deauville score 3 with OSEM [95]. In both the above studies, these observations were of immediate clinical importance because the cutoff between Deauville scores 3 and 4 is currently used to distinguish responders from non-responders. To show that such changes do not alter the progression-free and overall survival, large trials with sufficiently long follow-up will be needed to provide sufficient statistical power.
Summary and outlook
In this review, we have argued that a true advance in imaging technology should eventually be shown to improve patient outcome or other measures of patient benefit and that this is yet to be demonstrated for the latest PET technologies. If they introduce variability and decrease standardization, as illustrated for the Deauville score, without a net improvement in clinical outcome, there are good grounds for remaining cautious about their role in clinical care.
Comparability of image characteristics and quantitative results is a general issue in medical imaging, and is not confined to PET imaging, let alone to the “new” technologies. Unfortunately, SUV differences between different scanner generations and reconstruction algorithms cannot simply be eliminated by normalization to a reference region such as the liver [60, 102]. Retrospective harmonization of quantitative reads may be achieved by a data transformation method, called ComBat [120]. This method uses the quantitative reads within a multicenter setting to derive a data-driven transformation and aligns the distribution of data values among centers. This method may be useful when developing or validating evaluation criteria when data is collected in a non-harmonized fashion. However, due to the nature of data-driven transformations, the method needs to be trained for each specific combination of multicenter sites, PET/CT systems, and for each dataset (i.e., disease type and stage). Therefore, the obtained transformation cannot be directly or prospectively applied to other multicenter datasets, diseases, and stages nor can it be derived from phantom studies. Furthermore, covariates need to be considered carefully [120].
Fortunately, the nuclear medicine community has been eager to characterize and address this variability by introducing PET/CT performance standards aiming at realizing upfront and generalizable harmonization. For conventional scanners with OSEM-based reconstruction, the first EARL standard ensured that all systems achieved sphere recovery coefficients (RC) within a common range [116, 117]. Following the same approach, Kaalep et al. have demonstrated that a new standard can be defined based on a different range of sphere RC that is now satisfied by PET/CT scanners with PSF or PL reconstruction, some of them equipped with SiPM [56]. Currently, the safest way of applying new technologies in clinical routine practice is to have two distinct datasets available―one that is compliant with the conventional standard and one that fulfills the new standard at least until their effects on image interpretation and subsequent patient management are better understood (Fig. 3). Ideally, these datasets should be reconstructed prospectively. As an alternative, retrospective smoothing of the new datasets, usually with a Gaussian filter of specific width, can be employed to achieve image characteristics that are comparable to the conventional standard [58, 59, 119]. It may be noted that the EARL standards were introduced to harmonize data for visual assessment and SUV calculations in oncological whole-body F-18 PET with the goal of achieving sphere CR at high contrast that are comparable within a range of about ±10% [56]. However, different clinical tasks, target lesion characteristics, or radionuclides will warrant validation of the methodology and―potentially―different or additional steps to achieve the necessary level of harmonization. As an example, a much smaller range of CR of about ±3% between state-of-the-art PET scanners has recently been demonstrated in F-18 studies using a Hoffman 3D brain phantom. This required task-specific quantitative measures, reconstruction settings, and image-based normalization [121]. Regarding radiomics features, Pfaehler et al. showed using patient-derived 3D prints that considerable variability in features remained between different PET scanners―despite the improved agreement achieved with EARL-compliant reconstruction [122]. This suggests that such applications may warrant additional efforts, such as the named ComBat approach [120]. EARL-compliant reconstruction has been successfully transferred from F-18 to Ga-68 and Zr-89. However, thorough correction for radionuclide-specific cross-calibration mismatch was necessary to ensure compliance [123, 124].
If an appropriate new standard has been established, the next step is to characterize the implications of this new standard for clinical decisions in appropriately designed clinical trials for specific clinical scenarios. Apart from laborious randomized trials, this might also be done as post hoc analyses such as those performed in HD17 and HD18 to explore thresholds that discriminate response better using the different technologies/datasets. This would include determining the degree to which the established and the new dataset differ―especially for those patients where uptake is close to the decision threshold―and evaluating the potential for management change. New metrics that may be more suited for advanced reconstructions, e.g., based on SUVpeak for lesional uptake [56, 72], should also be explored in such clinical trial datasets. Besides Deauville scores and other established diagnostic cut-offs [65], such studies may address the influence on prognostic cut-offs [58] or on normal databases [91, 92]. Using either approach, non-inferiority of the new standard regarding clinical endpoints must be well proven, before it can be used safely to guide crucial clinical decisions. Notably, post hoc analyses may be of limited validity as the effect of using the alternative, new dataset for therapeutic decisions regarding patient outcome may only be assessed indirectly―if at all.
Recently there has been interest in use of virtual (or in silico) imaging trials whereby the patients, imaging system, and clinical interpretation processes are emulated using computational models [125,126,127]. If properly validated, virtual imaging trials could have a role in evaluating the impact of new technologies on diagnostic and quantitative accuracy, thus decreasing the required size and length of real clinical trials and reducing the length of the transition period.
References
Aide N. Review on new PET technologies - pros. Eur J Nucl Med Mol Imaging. 2021.
Guyatt G, Cairns J, Churchill D, Cook D, Haynes B, Hirsh J, et al. Evidence-based medicine: a new approach to teaching the practice of medicine. JAMA. 1992;268:2420–5. https://doi.org/10.1001/jama.1992.03490170092032.
Fryback DG, Thornbury JR. The efficacy of diagnostic imaging. Med Decis Mak. 1991;11:88–94. https://doi.org/10.1177/0272989x9101100203.
Janatzek S. Nutzen diagnostischer Tests – vom Surrogat zur Patientenrelevanz. Earth Planet Sci Lett. 2011;105:504–9. https://doi.org/10.1016/j.zefq.2011.07.027.
Lord SJ, Irwig L, Simes RJ. When is measuring sensitivity and specificity sufficient to evaluate a diagnostic test, and when do we need randomized trials? Ann Intern Med. 2006;144:850–5. https://doi.org/10.7326/0003-4819-144-11-200606060-00011.
Schünemann HJ, Oxman AD, Brozek J, Glasziou P, Jaeschke R, Vist GE, et al. Grading quality of evidence and strength of recommendations for diagnostic tests and strategies. Bmj. 2008;336:1106–10. https://doi.org/10.1136/bmj.39500.677199.AE.
EUnetHTA. Guideline. Endpoints used in REA of pharmaceuticals: surrogate endpoints. European Network for Health Technology Assessment (EUnetHTA); 2015.
van Tinteren H, Hoekstra OS, Smit EF, van den Bergh JH, Schreurs AJ, Stallaert RA, et al. Effectiveness of positron emission tomography in the preoperative assessment of patients with suspected non-small-cell lung cancer: the PLUS multicentre randomised trial. Lancet. 2002;359:1388–93. https://doi.org/10.1016/s0140-6736(02)08352-6.
Fischer B, Lassen U, Mortensen J, Larsen S, Loft A, Bertelsen A, et al. Preoperative staging of lung cancer with combined PET-CT. N Engl J Med. 2009;361:32–9. https://doi.org/10.1056/NEJMoa0900043.
De Leyn P, Dooms C, Kuzdzal J, Lardinois D, Passlick B, Rami-Porta R, et al. Revised ESTS guidelines for preoperative mediastinal lymph node staging for non-small-cell lung cancer. Eur J Cardiothorac Surg. 2014;45:787–98. https://doi.org/10.1093/ejcts/ezu028.
Postmus PE, Kerr KM, Oudkerk M, Senan S, Waller DA, Vansteenkiste J, et al. Early and locally advanced non-small-cell lung cancer (NSCLC): ESMO Clinical Practice Guidelines for diagnosis, treatment and follow-up. Ann Oncol. 2017;28:iv1–iv21. https://doi.org/10.1093/annonc/mdx222.
FDA. FDA News Release: FDA approves first PSMA-targeted PET imaging drug for men with prostate cancer. U.S. Food and Drug Administration (FDA); 2020.
Fendler WP, Calais J, Eiber M, Flavell RR, Mishoe A, Feng FY, et al. Assessment of 68Ga-PSMA-11 PET accuracy in localizing recurrent prostate cancer: a prospective single-arm clinical trial. JAMA Oncol. 2019;5:856–63. https://doi.org/10.1001/jamaoncol.2019.0096.
Hofman MS, Lawrentschuk N, Francis RJ, Tang C, Vela I, Thomas P, et al. Prostate-specific membrane antigen PET-CT in patients with high-risk prostate cancer before curative-intent surgery or radiotherapy (proPSMA): a prospective, randomised, multicentre study. Lancet. 2020;395:1208–16. https://doi.org/10.1016/s0140-6736(20)30314-7.
Voltin CA, Goergen H, Baues C, Fuchs M, Mettler J, Kreissl S, et al. Value of bone marrow biopsy in Hodgkin lymphoma patients staged by FDG PET: results from the German Hodgkin Study Group trials HD16, HD17, and HD18. Ann Oncol. 2018;29:1926–31. https://doi.org/10.1093/annonc/mdy250.
El-Galaly TC, d’Amore F, Mylam KJ, de Nully BP, Bøgsted M, Bukh A, et al. Routine bone marrow biopsy has little or no therapeutic consequence for positron emission tomography/computed tomography-staged treatment-naive patients with Hodgkin lymphoma. J Clin Oncol. 2012;30:4508–14. https://doi.org/10.1200/jco.2012.42.4036.
Barrington SF, Kirkwood AA, Franceschetto A, Fulham MJ, Roberts TH, Almquist H, et al. PET-CT for staging and early response: results from the Response-Adapted Therapy in Advanced Hodgkin Lymphoma study. Blood. 2016;127:1531–8. https://doi.org/10.1182/blood-2015-11-679407.
Engert A, Haverkamp H, Kobe C, Markova J, Renner C, Ho A, et al. Reduced-intensity chemotherapy and PET-guided radiotherapy in patients with advanced stage Hodgkin’s lymphoma (HD15 trial): a randomised, open-label, phase 3 non-inferiority trial. Lancet. 2012;379:1791–9. https://doi.org/10.1016/s0140-6736(11)61940-5.
Johnson P, Federico M, Kirkwood A, Fosså A, Berkahn L, Carella A, et al. Adapted treatment guided by interim PET-CT scan in advanced Hodgkin’s lymphoma. N Engl J Med. 2016;374:2419–29. https://doi.org/10.1056/NEJMoa1510093.
Radford J, Illidge T, Counsell N, Hancock B, Pettengell R, Johnson P, et al. Results of a trial of PET-directed therapy for early-stage Hodgkin’s lymphoma. N Engl J Med. 2015;372:1598–607. https://doi.org/10.1056/nejmoa1408648.
Fuchs M, Goergen H, Kobe C, Kuhnert G, Lohri A, Greil R, et al. Positron emission tomography-guided treatment in early-stage favorable Hodgkin lymphoma: final results of the international, randomized phase III HD16 trial by the German Hodgkin Study Group. J Clin Oncol. 2019;37:2835–45. https://doi.org/10.1200/jco.19.00964.
Freeman CL, Savage KJ, Villa D, Scott DW, Srour L, Gerrie AS, et al. Long-term results of PET-guided radiation in advanced-stage diffuse large B-cell lymphoma patients treated with R-CHOP. Blood. 2020. https://doi.org/10.1182/blood.2020005846.
Cottereau AS, Nioche C, Dirand AS, Clerc J, Morschhauser F, Casasnovas O, et al. (18)F-FDG PET dissemination features in diffuse large B-cell lymphoma are predictive of outcome. J Nucl Med. 2020;61:40–5. https://doi.org/10.2967/jnumed.119.229450.
Qi J, Leahy RM. Iterative reconstruction techniques in emission computed tomography. Phys Med Biol. 2006;51:R541–78. https://doi.org/10.1088/0031-9155/51/15/r01.
Alessio A, Kinahan P, Harrison RL, Lewellen TK. Measured spatially variant system response for PET image reconstruction; 2005.
Rapisarda E, Bettinardi V, Thielemans K, Gilardi MC. Image-based point spread function implementation in a fully 3D OSEM reconstruction algorithm for PET. Phys Med Biol. 2010;55:4131–51. https://doi.org/10.1088/0031-9155/55/14/012.
Carter LM, Kesner AL, Pratt EC, Sanders VA, Massicano AVF, Cutler CS, et al. The impact of positron range on PET resolution, evaluated with phantoms and PHITS Monte Carlo simulations for conventional and non-conventional radionuclides. Mol Imaging Biol. 2020;22:73–84. https://doi.org/10.1007/s11307-019-01337-2.
Herraiz JL, Bembibre A, López-Montes A. Deep-learning based positron range correction of PET images. Appl Sci. 2021;11:266. https://doi.org/10.3390/app11010266.
Ahn S, Fessler J. Globally convergent image reconstruction for emission tomography using relaxed ordered subsets algorithms. IEEE Trans Med Imaging. 2003;22:613–26.
Liow JS, Strother SC. Practical tradeoffs between noise, quantitation, and number of iterations for maximum likelihood-based reconstructions. IEEE Trans Med Imaging. 1991;10:563–71. https://doi.org/10.1109/42.108591.
Ross S. White paper Q.Clear. GE Healthcare. General Electric Company; 2014.
Ahn S, Ross SG, Asma E, Miao J, Jin X, Cheng L, et al. Quantitative comparison of OSEM and penalized likelihood image reconstruction using relative difference penalties for clinical PET. Phys Med Biol. 2015;60:5733–51. https://doi.org/10.1088/0031-9155/60/15/5733.
Asma E, Ahn S, Ross S, Chen A, Manjeshwar R. Accurate and consistent lesion quantitation with clinically acceptable penalized likelihood images; 2012.
Mumcuoğlu EU, Leahy RM, Cherry SR. Bayesian reconstruction of PET images: methodology and performance analysis. Phys Med Biol. 1996;41:1777–807. https://doi.org/10.1088/0031-9155/41/9/015.
Surti S, Karp JS. Update on latest advances in time-of-flight PET. Phys Med. 2020;80:251–8. https://doi.org/10.1016/j.ejmp.2020.10.031.
Rausch I, Ruiz A, Valverde-Pascual I, Cal-González J, Beyer T, Carrio I. Performance evaluation of the Vereos PET/CT system according to the NEMA NU2-2012 standard. J Nucl Med. 2019;60:561–7. https://doi.org/10.2967/jnumed.118.215541.
Frach T, Prescher G, Degenhardt C, Gruyter R, Schmitz A, Ballizany R. The digital silicon photomultiplier - principle of operation and intrinsic detector performance; 2009.
Jiang W, Chalich Y, Deen MJ. Sensors for positron emission tomography applications. Sensors (Basel). 2019;19:5019. https://doi.org/10.3390/s19225019.
Chen S, Hu P, Gu Y, Yu H, Shi H. Performance characteristics of the digital uMI550 PET/CT system according to the NEMA NU2-2018 standard. EJNMMI Phys. 2020;7:43. https://doi.org/10.1186/s40658-020-00315-w.
van Sluis J, de Jong J, Schaar J, Noordzij W, van Snick P, Dierckx R, et al. Performance characteristics of the digital biograph vision PET/CT system. J Nucl Med. 2019;60:1031–6. https://doi.org/10.2967/jnumed.118.215418.
Li X, Qi W, Miyahara M, Kolthammer J. Performance characterization of an SiPM-based time-of-flight canon PET/CT scanner. J Nucl Med. 2020;61:14.
Pan T, Einstein SA, Kappadath SC, Grogg KS, Lois Gomez C, Alessio AM, et al. Performance evaluation of the 5-ring GE discovery MI PET/CT system using the national electrical manufacturers association NU 2-2012 standard. Med Phys. 2019;46:3025–33. https://doi.org/10.1002/mp.13576.
Tierney MC. Raising the value of PET: technology, reliability and reporting take it up a notch. Radiology Business.
GE. Q.SUV. Quantitative SUV you and your patients can trust. GE Healthcare; 2014.
Siemens. Inside Biograph TruePoint PET•CT. Siemens Medical Solutions USA, Inc.; 2009.
Philips. Proven accuracy inspires confidence. Philips Vereos Digital PET/CT. Koninklijke Philips N.V.; 2019.
Zhang J, Maniawski P, Knopp MV. Performance evaluation of the next generation solid-state digital photon counting PET/CT system. EJNMMI Res. 2018;8:97. https://doi.org/10.1186/s13550-018-0448-7.
Knopp M, Binzel K, Bardos P, Knopp M, Wright C, Zhang J, et al. Next generation digital PET/CT: a phase I intra-individual comparison with current photomultiplier TOF PET/CT. Radiological Society of North America 2015 Scientific Assembly and Annual Meeting, November 29 - December 4, 2015. Chicago, IL, USA; 2015.
Sekine T, Delso G, Zeimpekis KG, de Galiza BF, ter Voert EEGW, Huellner M, et al. Reduction of 18F-FDG dose in clinical PET/MR imaging by using silicon photomultiplier detectors. Radiology. 2017;286:249–59. https://doi.org/10.1148/radiol.2017162305.
Siemens. Biograph Vision. Siemens Healthcare GmbH; 2018.
Martí-Climent JM, Prieto E, Domínguez-Prado I, García-Velloso MJ, Rodríguez-Fraile M, Arbizu J, et al. [Contribution of time of flight and point spread function modeling to the performance characteristics of the PET/CT Biograph mCT scanner]. Rev Esp Med Nucl Imagen Mol 2013;32:13–21. doi:https://doi.org/10.1016/j.remn.2012.07.001.
Rogasch JM, Hofheinz F, Lougovski A, Furth C, Ruf J, Großer OS, et al. The influence of different signal-to-background ratios on spatial resolution and F18-FDG-PET quantification using point spread function and time-of-flight reconstruction. EJNMMI Phys. 2014;1:12. https://doi.org/10.1186/2197-7364-1-12.
Munk OL, Tolbod LP, Hansen SB, Bogsrud TV. Point-spread function reconstructed PET images of sub-centimeter lesions are not quantitative. EJNMMI physics. 2017;4:5. https://doi.org/10.1186/s40658-016-0169-9.
Prieto E, Domínguez-Prado I, García-Velloso MJ, Peñuelas I, Richter J, Martí-Climent JM. Impact of time-of-flight and point-spread-function in SUV quantification for oncological PET. Clin Nucl Med. 2013;38:103–9. https://doi.org/10.1097/RLU.0b013e318279b9df.
Knäusl B, Rausch IF, Bergmann H, Dudczak R, Hirtl A, Georg D. Influence of PET reconstruction parameters on the TrueX algorithm. A combined phantom and patient study. Nuklearmedizin. 2013;52:28–35. https://doi.org/10.3413/Nukmed-0523-12-07.
Kaalep A, Sera T, Rijnsdorp S, Yaqub M, Talsma A, Lodge MA, et al. Feasibility of state of the art PET/CT systems performance harmonisation. Eur J Nucl Med Mol Imaging. 2018;45:1344–61. https://doi.org/10.1007/s00259-018-3977-4.
Rogasch JM, Steffen IG, Hofheinz F, Großer OS, Furth C, Mohnike K, et al. The association of tumor-to-background ratios and SUVmax deviations related to point spread function and time-of-flight F18-FDG-PET/CT reconstruction in colorectal liver metastases. EJNMMI Res. 2015;5:31. https://doi.org/10.1186/s13550-015-0111-5.
Houdu B, Lasnon C, Licaj I, Thomas G, Do P, Guizard AV, et al. Why harmonization is needed when using FDG PET/CT as a prognosticator: demonstration with EARL-compliant SUV as an independent prognostic factor in lung cancer. Eur J Nucl Med Mol Imaging. 2019;46:421–8. https://doi.org/10.1007/s00259-018-4151-8.
Tsutsui Y, Daisaki H, Akamatsu G, Umeda T, Ogawa M, Kajiwara H, et al. Multicentre analysis of PET SUV using vendor-neutral software: the Japanese Harmonization Technology (J-Hart) study. EJNMMI Res. 2018;8:83. https://doi.org/10.1186/s13550-018-0438-9.
Kuhnert G, Boellaard R, Sterzer S, Kahraman D, Scheffler M, Wolf J, et al. Impact of PET/CT image reconstruction methods and liver uptake normalization strategies on quantitative image analysis. Eur J Nucl Med Mol Imaging. 2016;43:249–58. https://doi.org/10.1007/s00259-015-3165-8.
Akamatsu G, Ishikawa K, Mitsumoto K, Taniguchi T, Ohya N, Baba S, et al. Improvement in PET/CT image quality with a combination of point-spread function and time-of-flight in relation to reconstruction parameters. J Nucl Med. 2012;53:1716–22. https://doi.org/10.2967/jnumed.112.103861.
Schaefferkoetter J, Casey M, Townsend D, El Fakhri G. Clinical impact of time-of-flight and point response modeling in PET reconstructions: a lesion detection study. Phys Med Biol. 2013;58:1465–78. https://doi.org/10.1088/0031-9155/58/5/1465.
Lasnon C, Hicks RJ, Beauregard JM, Milner A, Paciencia M, Guizard AV, et al. Impact of point spread function reconstruction on thoracic lymph node staging with 18F-FDG PET/CT in non-small cell lung cancer. Clin Nucl Med. 2012;37:971–6. https://doi.org/10.1097/RLU.0b013e318251e3d1.
Hotta M, Minamimoto R, Yano H, Gohda Y, Shuno Y. Diagnostic performance of (18)F-FDG PET/CT using point spread function reconstruction on initial staging of rectal cancer: a comparison study with conventional PET/CT and pelvic MRI. Cancer Imaging. 2018;18:4. https://doi.org/10.1186/s40644-018-0137-9.
Rogasch JMM, Albers J, Steinkrüger FL, Lukas M, Wedel F, Amthauer H, et al. Point spread function reconstruction for integrated 18F-FET PET/MRI in patients with glioma: does it affect SUVs and respective tumor-to-background ratios? Clin Nucl Med. 2019;44:e280–e5. https://doi.org/10.1097/rlu.0000000000002423.
Akamatsu G, Mitsumoto K, Taniguchi T, Tsutsui Y, Baba S, Sasaki M. Influences of point-spread function and time-of-flight reconstructions on standardized uptake value of lymph node metastases in FDG-PET. Eur J Radiol. 2014;83:226–30. https://doi.org/10.1016/j.ejrad.2013.09.030.
Armstrong IS, Kelly MD, Williams HA, Matthews JC. Impact of point spread function modelling and time of flight on FDG uptake measurements in lung lesions using alternative filtering strategies. EJNMMI Physics. 2014;1:99. https://doi.org/10.1186/s40658-014-0099-3.
Aklan B, Oehmigen M, Beiderwellen K, Ruhlmann M, Paulus DH, Jakoby BW, et al. Impact of point-spread function modeling on PET image quality in integrated PET/MR hybrid imaging. J Nucl Med. 2016;57:78–84. https://doi.org/10.2967/jnumed.115.154757.
Kurita Y, Ichikawa Y, Nakanishi T, Tomita Y, Hasegawa D, Murashima S, et al. The value of Bayesian penalized likelihood reconstruction for improving lesion conspicuity of malignant lung tumors on (18)F-FDG PET/CT: comparison with ordered subset expectation maximization reconstruction incorporating time-of-flight model and point spread function correction. Ann Nucl Med. 2020;34:272–9. https://doi.org/10.1007/s12149-020-01446-x.
Jansen BHE, Jansen RW, Wondergem M, Srbljin S, de Klerk JMH, Lissenberg-Witte BI, et al. Lesion detection and interobserver agreement with advanced image reconstruction for (18)F-DCFPyL PET/CT in patients with biochemically recurrent prostate cancer. J Nucl Med. 2020;61:210–6. https://doi.org/10.2967/jnumed.118.222513.
Enilorac B, Lasnon C, Nganoa C, Fruchart C, Gac AC, Damaj G, et al. Does PET reconstruction method affect Deauville score in lymphoma patients? J Nucl Med. 2018;59:1049–55. https://doi.org/10.2967/jnumed.117.202721.
Rogasch JM, Suleiman S, Hofheinz F, Bluemel S, Lukas M, Amthauer H, et al. Reconstructed spatial resolution and contrast recovery with Bayesian penalized likelihood reconstruction (Q.Clear) for FDG-PET compared to time-of-flight (TOF) with point spread function (PSF). EJNMMI physics. 2020;7:2. https://doi.org/10.1186/s40658-020-0270-y.
Teoh EJ, McGowan DR, Macpherson RE, Bradley KM, Gleeson FV. Phantom and clinical evaluation of the Bayesian penalized likelihood reconstruction algorithm Q.Clear on an LYSO PET/CT System. J Nucl Med. 2015;56:1447–52. https://doi.org/10.2967/jnumed.115.159301.
Reynés-Llompart G, Gámez-Cenzano C, Romero-Zayas I, Rodríguez-Bel L, Vercher-Conejero JL, Martí-Climent JM. Performance characteristics of the whole-body discovery IQ PET/CT system. J Nucl Med. 2017;58:1155–61. https://doi.org/10.2967/jnumed.116.185561.
Lindström E, Sundin A, Trampal C, Lindsjö L, Ilan E, Danfors T, et al. Evaluation of penalized-likelihood estimation reconstruction on a digital time-of-flight PET/CT scanner for (18)F-FDG whole-body examinations. J Nucl Med. 2018;59:1152–8. https://doi.org/10.2967/jnumed.117.200790.
Te Riet J, Rijnsdorp S, Roef MJ, Arends AJ. Evaluation of a Bayesian penalized likelihood reconstruction algorithm for low-count clinical (18)F-FDG PET/CT. EJNMMI Phys. 2019;6:32. https://doi.org/10.1186/s40658-019-0262-y.
Otani T, Hosono M, Kanagaki M, Onishi Y, Matsubara N, Kawabata K, et al. Evaluation and optimization of a new PET reconstruction algorithm, Bayesian penalized likelihood reconstruction, for lung cancer assessment according to lesion size. AJR Am J Roentgenol. 2019;213:W50–w6. https://doi.org/10.2214/ajr.18.20478.
Caribé P, Koole M, D’Asseler Y, Van Den Broeck B, Vandenberghe S. Noise reduction using a Bayesian penalized-likelihood reconstruction algorithm on a time-of-flight PET-CT scanner. EJNMMI Phys. 2019;6:22. https://doi.org/10.1186/s40658-019-0264-9.
Seo Y, Khalighi MM, Wangerin KA, Deller TW, Wang YH, Jivan S, et al. Quantitative and qualitative improvement of low-count [(68)Ga]citrate and [(90)Y]microspheres PET image reconstructions using block sequential regularized expectation maximization algorithm. Mol Imaging Biol. 2020;22:208–16. https://doi.org/10.1007/s11307-019-01347-0.
Trägårdh E, Minarik D, Brolin G, Bitzén U, Olsson B, Oddstig J. Optimization of [(18)F]PSMA-1007 PET-CT using regularized reconstruction in patients with prostate cancer. EJNMMI Phys. 2020;7:31. https://doi.org/10.1186/s40658-020-00298-8.
Lindström E, Velikyan I, Regula N, Alhuseinalkhudhur A, Sundin A, Sörensen J, et al. Regularized reconstruction of digital time-of-flight (68)Ga-PSMA-11 PET/CT for the detection of recurrent disease in prostate cancer patients. Theranostics. 2019;9:3476–84. https://doi.org/10.7150/thno.31970.
Messerli M, Stolzmann P, Egger-Sigg M, Trinckauf J, D’Aguanno S, Burger IA, et al. Impact of a Bayesian penalized likelihood reconstruction algorithm on image quality in novel digital PET/CT: clinical implications for the assessment of lung tumors. EJNMMI Phys. 2018;5:27. https://doi.org/10.1186/s40658-018-0223-x.
Guo B, Wu Z, Zhao B, Huang B, Li X, Zhao J, et al. Quantification accuracy using Bayesian penalized likelihood based reconstruction 68Ga PET-CT. J Nucl Med. 2020;61:162.
Shkumat N, Vali R, Shammas A. Clinical evaluation of reconstruction and acquisition time for pediatric 18F-FDG brain PET using digital PET/CT. Pediatr Radiol. 2020;50. https://doi.org/10.1007/s00247-020-04640-1.
Reynés-Llompart G, Sabaté-Llobera A, Llinares-Tello E, Martí-Climent JM, Gámez-Cenzano C. Image quality evaluation in a modern PET system: impact of new reconstructions methods and a radiomics approach. Sci Rep. 2019;9:10640. https://doi.org/10.1038/s41598-019-46937-8.
Howard BA, Morgan R, Thorpe MP, Turkington TG, Oldan J, James OG, et al. Comparison of Bayesian penalized likelihood reconstruction versus OS-EM for characterization of small pulmonary nodules in oncologic PET/CT. Ann Nucl Med. 2017;31:623–8. https://doi.org/10.1007/s12149-017-1192-1.
Wu Z, Qin Z, Huang B, Guo B, Hao X, Wu P, et al. Improved absolute quantification using Bayesian penalized likelihood reconstruction on a digital PET/CT – towards true uptake measurement. Research Square; 2020.
Schwyzer M, Martini K, Benz DC, Burger IA, Ferraro DA, Kudura K, et al. Artificial intelligence for detecting small FDG-positive lung nodules in digital PET/CT: impact of image reconstructions on diagnostic performance. Eur Radiol. 2020;30:2031–40. https://doi.org/10.1007/s00330-019-06498-w.
Svirydenka H, Muehlematter UJ, Nagel HW, Delso G, Ferraro DA, Kudura K, et al. (68)Ga-PSMA-11 dose reduction for dedicated pelvic imaging with simultaneous PET/MR using TOF BSREM reconstructions. Eur Radiol. 2020;30:3188–97. https://doi.org/10.1007/s00330-020-06667-2.
Bjöersdorff M, Oddstig J, Karindotter-Borgendahl N, Almquist H, Zackrisson S, Minarik D, et al. Impact of penalizing factor in a block-sequential regularized expectation maximization reconstruction algorithm for (18)F-fluorocholine PET-CT regarding image quality and interpretation. EJNMMI Phys. 2019;6:5. https://doi.org/10.1186/s40658-019-0242-2.
Mairal E, Doyen M, Rivasseau-Jonveaux T, Malaplate C, Guedj E, Verger A. Clinical impact of digital and conventional PET control databases for semi-quantitative analysis of brain (18)F-FDG digital PET scans. EJNMMI Res. 2020;10:144. https://doi.org/10.1186/s13550-020-00733-y.
Lindström E, Oddstig J, Danfors T, Jögi J, Hansson O, Lubberink M. Image reconstruction methods affect software-aided assessment of pathologies of [(18)F]flutemetamol and [(18)F]FDG brain-PET examinations in patients with neurodegenerative diseases. Neuroimage Clin. 2020;28:102386. https://doi.org/10.1016/j.nicl.2020.102386.
Lindström E, Lindsjö L, Sundin A, Sörensen J, Lubberink M. Evaluation of block-sequential regularized expectation maximization reconstruction of (68)Ga-DOTATOC, (18)F-fluoride, and (11)C-acetate whole-body examinations acquired on a digital time-of-flight PET/CT scanner. EJNMMI physics. 2020;7:40. https://doi.org/10.1186/s40658-020-00310-1.
Teoh EJ, McGowan DR, Bradley KM, Belcher E, Black E, Gleeson FV. Novel penalised likelihood reconstruction of PET in the assessment of histologically verified small pulmonary nodules. Eur Radiol. 2016;26:576–84. https://doi.org/10.1007/s00330-015-3832-y.
Ly J, Minarik D, Edenbrandt L, Wollmer P, Trägårdh E. The use of a proposed updated EARL harmonization of (18)F-FDG PET-CT in patients with lymphoma yields significant differences in Deauville score compared with current EARL recommendations. EJNMMI Res. 2019;9:65. https://doi.org/10.1186/s13550-019-0536-3.
Baratto L, Park SY, Hatami N, Davidzon G, Srinivas S, Gambhir SS, et al. 18F-FDG silicon photomultiplier PET/CT: a pilot study comparing semi-quantitative measurements with standard PET/CT. PLoS One. 2017;12:e0178936. https://doi.org/10.1371/journal.pone.0178936.
Boellaard R, Delgado-Bolton R, Oyen WJ, Giammarile F, Tatsch K, Eschner W, et al. FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0. Eur J Nucl Med Mol Imaging. 2015;42:328–54. https://doi.org/10.1007/s00259-014-2961-x.
Alberts I, Prenosil G, Sachpekidis C, Weitzel T, Shi K, Rominger A, et al. Digital versus analogue PET in [(68)Ga]Ga-PSMA-11 PET/CT for recurrent prostate cancer: a matched-pair comparison. Eur J Nucl Med Mol Imaging. 2020;47:614–23. https://doi.org/10.1007/s00259-019-04630-y.
Hsu DFC, Ilan E, Peterson WT, Uribe J, Lubberink M, Levin CS. Studies of a next-generation silicon-photomultiplier-based time-of-flight PET/CT system. J Nucl Med. 2017;58:1511–8. https://doi.org/10.2967/jnumed.117.189514.
Vandendriessche D, Uribe J, Bertin H, De Geeter F. Performance characteristics of silicon photomultiplier based 15-cm AFOV TOF PET/CT. EJNMMI Phys. 2019;6:8. https://doi.org/10.1186/s40658-019-0244-0.
Surti S, Viswanath V, Daube-Witherspoon ME, Conti M, Casey ME, Karp JS. Benefit of improved performance with state-of-the art digital PET/CT for lesion detection in oncology. J Nucl Med. 2020;61:1684–90. https://doi.org/10.2967/jnumed.120.242305.
Fuentes-Ocampo F, López-Mora DA, Flotats A, Paillahueque G, Camacho V, Duch J, et al. Digital vs. analog PET/CT: intra-subject comparison of the SUVmax in target lesions and reference regions. Eur J Nucl Med Mol Imaging. 2019;46:1745–50. https://doi.org/10.1007/s00259-018-4256-0.
Nguyen NC, Vercher-Conejero JL, Sattar A, Miller MA, Maniawski PJ, Jordan DW, et al. Image quality and diagnostic performance of a digital PET prototype in patients with oncologic diseases: initial experience and comparison with analog PET. J Nucl Med. 2015;56:1378–85. https://doi.org/10.2967/jnumed.114.148338.
Sonni I, Baratto L, Park S, Hatami N, Srinivas S, Davidzon G, et al. Initial experience with a SiPM-based PET/CT scanner: influence of acquisition time on image quality. EJNMMI Phys. 2018;5:9. https://doi.org/10.1186/s40658-018-0207-x.
Conti M. Focus on time-of-flight PET: the benefits of improved time resolution. Eur J Nucl Med Mol Imaging. 2011;38:1147–57. https://doi.org/10.1007/s00259-010-1711-y.
Bailey DL, Meikle SR, Jones T. Effective sensitivity in 3D PET: the impact of detector dead time on 3D system performance. 1996 IEEE Nuclear Science Symposium Conference Record; 1996, vol.2. p. 1290–4.
Cherry SR, Jones T, Karp JS, Qi J, Moses WW, Badawi RD. Total-Body PET: Maximizing sensitivity to create new opportunities for clinical research and patient care. J Nucl Med 2018;59:3–12. doi:https://doi.org/10.2967/jnumed.116.184028.
Rogasch JMM, Frost N, Bluemel S, Michaels L, Penzkofer T, von Laffert M, et al. FDG-PET/CT for pretherapeutic lymph node staging in non-small cell lung cancer: a tailored approach to the ESTS/ESMO guideline workflow. Lung Cancer. 2021.
Phillips R, Shi WY, Deek M, Radwan N, Lim SJ, Antonarakis ES, et al. Outcomes of observation vs stereotactic ablative radiation for oligometastatic prostate cancer: the ORIOLE phase 2 randomized clinical trial. JAMA Oncol. 2020;6:650–9. https://doi.org/10.1001/jamaoncol.2020.0147.
Zschaeck S, Wust P, Beck M, Wlodarczyk W, Kaul D, Rogasch J, et al. Intermediate-term outcome after PSMA-PET guided high-dose radiotherapy of recurrent high-risk prostate cancer patients. Radiat Oncol. 2017;12:140. https://doi.org/10.1186/s13014-017-0877-x.
Koerber SA, Sprute K, Kratochwil C, Winter E, Haefner MF, Katayama S, et al. Clinical outcome of PSMA-guided radiotherapy for patients with oligorecurrent prostate cancer. Eur J Nucl Med Mol Imaging. 2021;48:143–51. https://doi.org/10.1007/s00259-020-04777-z.
Vogel MME, Kroeze SGC, Henkenberens C, Schmidt-Hegemann NS, Kirste S, Becker J, et al. Prognostic risk classification for biochemical relapse-free survival in patients with oligorecurrent prostate cancer after [(68)Ga]PSMA-PET-guided metastasis-directed therapy. Eur J Nucl Med Mol Imaging. 2020;47:2328–38. https://doi.org/10.1007/s00259-020-04760-8.
Ter Voert E, Muehlematter UJ, Delso G, Pizzuto DA, Müller J, Nagel HW, et al. Quantitative performance and optimal regularization parameter in block sequential regularized expectation maximization reconstructions in clinical (68)Ga-PSMA PET/MR. EJNMMI Res. 2018;8:70. https://doi.org/10.1186/s13550-018-0414-4.
Barrington SF, Mikhaeel NG, Kostakoglu L, Meignan M, Hutchings M, Müeller SP, et al. Role of imaging in the staging and response assessment of lymphoma: consensus of the International Conference on Malignant Lymphomas Imaging Working Group. J Clin Oncol. 2014;32:3048–58. https://doi.org/10.1200/JCO.2013.53.5229.
Barrington SF, Sulkin T, Forbes A, Johnson PWM. All that glitters is not gold - new reconstruction methods using Deauville criteria for patient reporting. Eur J Nucl Med Mol Imaging. 2018;45:316–7. https://doi.org/10.1007/s00259-017-3893-z.
Lasnon C, Desmonts C, Quak E, Gervais R, Do P, Dubos-Arvis C, et al. Harmonizing SUVs in multicentre trials when using different generation PET systems: prospective validation in non-small cell lung cancer patients. Eur J Nucl Med Mol Imaging. 2013;40:985–96. https://doi.org/10.1007/s00259-013-2391-1.
Aide N, Lasnon C, Veit-Haibach P, Sera T, Sattler B, Boellaard R. EANM/EARL harmonization strategies in PET quantification: from daily practice to multicentre oncological studies. Eur J Nucl Med Mol Imaging. 2017;44:17–31. https://doi.org/10.1007/s00259-017-3740-2.
Boellaard R, Kobe C, Zijlstra JM, Mikhaeel NG, Johnson PWM, Müller S, et al. Does PET reconstruction method affect Deauville scoring in lymphoma patients? J Nucl Med. 2018;59:1167–9. https://doi.org/10.2967/jnumed.118.211607.
Kaalep A, Burggraaff CN, Pieplenbosch S, Verwer EE, Sera T, Zijlstra J, et al. Quantitative implications of the updated EARL 2019 PET-CT performance standards. EJNMMI Phys. 2019;6:28. https://doi.org/10.1186/s40658-019-0257-8.
Orlhac F, Boughdad S, Philippe C, Stalla-Bourdillon H, Nioche C, Champion L, et al. A postreconstruction harmonization method for multicenter radiomic studies in PET. J Nucl Med. 2018;59:1321–8. https://doi.org/10.2967/jnumed.117.199935.
Verwer EE, Golla SSV, Kaalep A, Lubberink M, van Velden FHP, Bettinardi V, et al. Harmonisation of PET/CT contrast recovery performance for brain studies. Eur J Nucl Med Mol Imaging. 2021. https://doi.org/10.1007/s00259-021-05201-w.
Pfaehler E, Beukinga RJ, de Jong JR, Slart R, Slump CH, Dierckx R, et al. Repeatability of (18) F-FDG PET radiomic features: a phantom study to explore sensitivity to image reconstruction settings, noise, and delineation method. Med Phys. 2019;46:665–78. https://doi.org/10.1002/mp.13322.
Kaalep A, Huisman M, Sera T, Vugts D, Boellaard R. Feasibility of PET/CT system performance harmonisation for quantitative multicentre (89)Zr studies. EJNMMI Phys. 2018;5:26. https://doi.org/10.1186/s40658-018-0226-7.
Huizing DMV, Koopman D, van Dalen JA, Gotthardt M, Boellaard R, Sera T, et al. Multicentre quantitative (68)Ga PET/CT performance harmonisation. EJNMMI Phys. 2019;6:19. https://doi.org/10.1186/s40658-019-0253-z.
Abadi E, Segars W, Tsui BM, Kinahan P, Bottenus N, Frangi A, et al. Virtual clinical trials in medical imaging: a review. J Med Imaging. 2020;7:042805.
Badano A, Graff CG, Badal A, Sharma D, Zeng R, Samuelson FW, et al. Evaluation of digital breast tomosynthesis as replacement of full-field digital mammography using an in silico imaging trial. JAMA Netw Open. 2018;1:e185474. https://doi.org/10.1001/jamanetworkopen.2018.5474.
Badano A. In silico imaging clinical trials: cheaper, faster, better, safer, and more scalable. Trials. 2021;22:64. https://doi.org/10.1186/s13063-020-05002-w.
Funding
Open Access funding enabled and organized by Projekt DEAL. JMMR is participant in the BIH-Charité Digital Clinician Scientist Program funded by the Charité –Universitätsmedizin Berlin, the Berlin Institute of Health, and the German Research Foundation (DFG).
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
This article does not contain any studies with human participants or animals performed by any of the authors.
Conflict of interest
The authors declare no competing interests.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
This article is part of the Topical Collection on Technology
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Rogasch, J.M., Boellaard, R., Pike, L. et al. Moving the goalposts while scoring―the dilemma posed by new PET technologies. Eur J Nucl Med Mol Imaging 48, 2696–2710 (2021). https://doi.org/10.1007/s00259-021-05403-2
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00259-021-05403-2