Abstract
Background
Three-dimensional surface imaging is established in many disciplines for objective facial acquisition regarding anthropometry. Former studies addressed the validation of landmark-based measurements for single race. In order to distinguish racial difference, the reproducibility of the landmark measurements must first be validated.
Objectives
Our purpose is to validate the reproducibility of 46 facial soft-tissue landmarks on x, y, z axes to prove their reliability as 3D reference points.
Methods
The study included 80 European Caucasian and 80 Chinese volunteers. Standardized 3D surface imaging was performed using Vectra 3D system. Two raters identified and defined 46 landmarks (138 coordinates), then repeatedly 3D-imaged volunteers' facial region in separate sessions. Coordinates' reproducibility of landmarks is divided into three categories (< 0.5 mm, < 1 mm, and >1 mm) for intra- and inter-rater reproducibility assessments.
Results
Coordinates' reproducibility of 160 samples was distributed as follows: Intra-rater: < 0.5 mm (45%), < 1 mm (42%), >1 mm (13%); inter-rater: < 0.5 mm (31.2%), < 1 mm (42%), > 1 mm (26.8%). The reproducibility of landmarks in nasal tip region differs slightly between Caucasians and Asians. Compared to females, males typically have higher landmark reproducibility in lip and chin region. However, there were no differences in the reproducibility ranking of landmarks by gender.
Conclusion
The majority of the 46 landmarks in the 3D plane are reproducible to 1 mm, which is clinically acceptable. All selected landmarks showed strong consistency across race and gender, suggesting their potential use as reference points in prospective clinical practice.
Level of Evidence IV
This journal requires that authors assign a level of evidence to each article. For a full description of these Evidence-Based Medicine ratings, please refer to the Table of Contents or the online Instructions to Authors www.springer.com/00266.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Introduction
Three-dimensional surface imaging (3DSI) is an established means of facial analysis for infant facial development and congenital conditions such as cleft lip and palate or alterations of the skull, in facial reconstructive surgery, and for aesthetic facial plastic surgery consultation [1,2,3]. As a noninvasive technology, it also plays an increasingly important role in evaluating facial morphology. Surface imaging is often used in conjunction with computed tomography (CT) or magnetic resonance imaging (MRI).
Although there are numerous devices for three-dimensional (3D) acquisition, the basis of anthropometric surveying is often a landmark-based approach. Facial surface landmarks are critical to the accuracy of 3D facial morphology measurement and analysis [4, 5].
Currently, 3D cameras are promising tools for the assessment of facial soft-tissue morphology with high-resolution surface textures. The advantage lies in quickly collecting information and constructing high-resolution 3D images, which can accurately capture the subject's facial skin color and surface texture [6,7,8]. Rendered 3D models may provide benefit for (1) evaluation of pediatric facial development, (2) analysis of facial morphology of patients affected by congenital and acquired pathological factors, (3) facial analysis for patients undergoing facial cosmetic, reconstructive, and orthodontic procedures, and (4) morphological studies on the normalization of facial impairments [9,10,11,12].
Facial surface landmarks are critical to the accuracy of 3D morphological assessment. Although there has been much research on the reliability and reproducibility of landmarks on facial 3D images, they are mainly focused on one race and the sample sizes are relatively small. Additionally, many of them do not assess reproducibility, and most of the selected landmarks were traditional reference points that had been validated [7, 13,14,15]. Any facial morphological analysis based on anatomical landmarks requires highly reproductive and novel reference points across different patient populations. In this study, we aimed to define a number of novel facial soft-tissue landmarks to and assessed their reproducibility using a photogrammetric 3D stereophotography system
Materials and Methods
Study Sample
The study involved 80 European Caucasians (40 males, 40 females) and 80 Asian (40 males, 40 females). The age range of Caucasian volunteers was 20 to 50 years (30.49 ± 5.52 years). The age range of Asian volunteers was 18 to 45 years (30.36 ± 2.99 years). Written informed consent was obtained prior to enrollment. The study was approved by local university ethics committee (REF: 266-13) and conducted in accordance with the Declaration of Helsinki. Exclusion criteria were facial deformities, previous facial surgery, and volunteers diagnosed with epilepsy or other seizure disorders.
3D Stereophotography Equipment and Parameters
All facial 3d scans were captured by the high-resolution Vectra XT 3D Surface Imaging System (Canfield Inc., New Jersey, USA). It is a vertical fixed photography system with six integrated cameras at different angles. Its proprietary illumination system automatically adjusts the focus for optimal face imaging and its 3.5 mm photographic time reduces artifacts of the Vectra system due to the unconscious displacement of the subject.
Image Sampling Process
Before taking the photograph, all volunteers were removed from any factors which would interfere with the image modeling: jewelry, glasses, and clothing elements (scarves, hat, etc.). They were asked to remove any hair from the face, forehead, and ears to completely expose the facial area. Male volunteers were asked to shave, while beards would cause image artifacts. The scan was performed in a well-lighted room. All volunteers were asked to sit in the same chair with a fixed backrest, lips kept closed without teeth grinding, and to look directly at the red marker dot on the 3D camera with a neutral facial expression and a natural head position. The lighting was kept under the same control conditions as in our daily work.
3D Processing
All captured images were processed, aligned, and analyzed using the proprietary Mirror® (Canfield Scientific; NJ, USA) [16]. The software was also implemented for providing reference frameworks with the x, y, z axis (sagittal, Y-Z plane; coronal, X-Y plane; and transverse, X-Z plane.) by unifying orientation for different images. The entire face was marked and symmetry of the planes was automatically adjusted by the integrated software. The midpoint between two endocanthions was chosen (mid-intercanthal point) as the origin point. The sagittal plane was referenced to the origin through the midline of the face, the coronal plane was determined to be the average natural head position, and the transverse plane was established to span the origin (Fig 1). The 3D face images were normalized on three planes to obtain comparable X, Y, and Z coordinates to assess the reproducibility of the facial landmarks [17].
Data Evaluation
In routine clinical facial surgery, surgeons will mark various anatomical landmarks for measurement. In this work, two raters identified and defined 46 landmarks based on former reports and the raters placed them separately on 3D facial images and all landmark coordinates were recorded (138 totally) for X,Y, Z axes [18,19,20]. Their name, abbreviation, and definition are displayed in Table 1. These landmarks contain some of the most commonly used classic landmarks and the novel landmarks that we found beneficial for facial measurements (Fig 2). Both raters are professional researchers in our department experienced in 3D anthropometry. One of the raters evaluates the reproducibility of all landmarks in a two-week interval to obtain intra-rater reproducibility. We then compared the two raters' results for inter-rater reproducibility. Based on the intra-rater and inter-rater assessment results of 160 volunteer, we calculated the mean and standard deviation for each landmark. The error of the landmarks is presented as the absolute difference in measurement of each landmark on three axes. It is divided into three categories (<0.5 mm, <1 mm, and >1 mm). Coordinates with a difference between the measurements of less than 0.5 mm of all samples are classified as highly reproducible; with a difference between 0.5 mm and 1 mm are determined to be moderately reproducible; with a difference above 1 mm are considered to be poorly reproducible [21, 22]. Both intra- and inter-rater evaluations assessed the reproducibility of landmarks of all samples on three planes, divided into race and gender.
Statistical Analysis
A total of 88320 variables (46 landmarks×160 subjects×3 planes×2 raters×2 measurements) were analyzed with SPSS Statistics 23.00 (IBM, Armonk, NY, USA). Based on the absolute difference of each landmark on the x, y, and z axes, we calculate the total average reproducibility difference using the following formula to: \(T=\sqrt{\frac{{(\Delta X)}^{2}+{(\Delta Y)}^{2}+{(\Delta Z)}^{2}}{3}}\), where T is the total average difference, \(\Delta X\) is the difference on the x-axis, \(\Delta Y\) is the difference on the y-axis, and \(\Delta Z\) is the difference on the z axis. Each variable is first corrected to the median for all volunteers. Bland–Altman plots were carried out for intra- and inter-rater reproducibility assessments. For each plot, the difference between the measurements of each landmark coordinate was calculated and the average measurement for that particular coordinate is generated. In Figures 3 and 4, we exhibited the representative coordinates of selected landmarks to illustrate the consistency between high, moderate, and low levels of coordinates measurements in different scenarios. The vertical axis of Bland–Altman plots shows the measurement variance of the selected landmarks, while the horizontal axis shows the average of the measurements. The zero line refers to the subject with zero measurement variance (highest reproducibility). The two dashed lines above and below the zero line indicate the subject with the highest variance between the two measurement sessions.
Results
Overall Reproducibility
Table 2 shows the overall results of coordinates' reproducibility gained from intra- and inter-rater results of 160 volunteers. In addition to the overview, we display the results separately according to race and gender. Generally, reproducibility of most assessments was less than 1 mm (intra-examiner 87%, and inter-examiner 73.2%). In the Caucasian subgroups, the intra-examiner was 83.4% and inter-examiner was 71.1%; in Asian the intra-examiner was 79.7% and the inter-examiner was 72.5%. Among females, the intra-examiner was 79.7% and inter-examiner was 69.6%; among males, the intra-examiner was 82.6% and the inter-examiner was 73.2%. The highest reproducibility (<0.5 mm) coordinates were 45% (intra-rater) and 31.2% (inter-rater) of the 160 samples. The worst reproducibility (>1 mm) coordinates accounted for 13% (intra-rater) and 26.8% (inter-rater).
The error results of all landmarks on the x, y, and z axes after all samples were grouped by gender and race are shown in Tables 3 and 4, respectively (mean and SD). Additionally, we calculated the total error based on the results on each axis for each landmark. The landmarks were ranked from most reproducible to least reproducible for both intra- and inter-examiner assessments. Compared to the intra-rater assessments, we noticed that the corresponding inter-rater results showed poorer reproducibility.
The Reproducibility of Landmarks in Caucasian and Asian Participants
Landmark accuracy for the Caucasian group ranged from 0.17 to 0.94 mm (intra-rater) and 0.20 to 1.38 mm (inter-rater) in the nose area, 0.44 to 0.61 mm (intra-rater) and 0.49 to 0.95 mm (inter-rater) in the eye area, 0.44 to 1.47 mm (intra-rater) and 0.52 to 1.75 mm (inter-rater) in the mouth area, 0.72 to 1.07 mm (intra-rater) and 1.19 to 1.63 mm (inter-rater) in the ear area, and 1.52 to 1.79 mm (intra-rater) and 1.23 to 2.04 mm (inter-rater) in the other areas.
Landmark accuracy for the Asian group ranged from 0.20 to 1.38 mm (intra-rater) and 0.25 to 1.13mm (inter-rater) in the nose area, 0.30 to 0.62 mm (intra-rater) and 0.54 to 0.85 mm (inter-rater) in the eye area, 0.47 to 1.52 mm (intra-rater) and 0.67 to 1.73 mm (inter-rater) in the mouth area, 0.86 to 0.97 mm (intra-rater) and 1.05 to 1.40 mm (inter-rater) in the ear area, and 1.23 to 2.04 mm (intra-rater) and 1.43 to 2.21 mm (inter-rater) in the other areas.
Differences in landmark reproducibility between Caucasian and Asian were concentrated in nose tip, alare, and nostril area, including Sn, Cc, Cm, Al right, Al left, Ac right, Nm of both sides, and Stb point. Several landmarks showed poor reproducibility in both Caucasians and Asians, namely Tri, Zy right, and Zy left. The landmarks with the most significant intra- and inter-group differences were Zy left with 0.8 mm in Caucasians and Pa right with 0.43 mm in Asians. Moreover, the measurement differences between intra- and inter-group assessment of landmarks in the Asian group were generally smaller than in the Caucasian group. The most and least reproducible landmarks in Asians were consistent with those in Caucasians (Table 3 & Supplement Table 1).
The Reproducibility of Landmarks in Female and Male Subgroups
The accuracy of landmarks in the female subgroup ranged from 0.24 to 1.55 mm (intra-rater) and 0.32 to 1.88 mm (inter-rater) in the nose area, 0.31 to 0.53 mm (intra-rater) and 0.45 to 0.76 mm (inter-rater) in the eye area, 0.40 to 1.33 mm (intra-rater) and 0.53 to 1.83 mm (inter-rater) in the mouth area, 0.79 to 1.22 mm (intra-rater) and 1.40 to 1.99 mm (inter-rater) in the ear area, and 1.19 to 1.97 mm (intra-rater) and 1.17 to 2.25 mm (inter-rater) in the other areas.
The accuracy of landmarks in the male subgroup ranged from 0.21 to 1.45 mm (intra-rater) and 0.31 to 1.65 mm (inter-rater) in the nose area, 0.38 to 0.68 mm (intra-rater) and 0.49 to 0.85mm (inter-rater) in the eye area, 0.41 to 1.08 mm (intra-rater) and 0.50 to 1.57 mm (inter-rater) in the mouth area, 0.83 to 1.14 mm (intra-rater) and 1.20 to 2.06 mm (inter-rater) in the ear area, and 1.28 to 1.45 mm (intra-rater) and 1.03 to 2.43 mm (inter-rater) in the other areas.
Compared to females, landmarks concentrating on the nose and mouth areas had higher reproducibility in males in intra-rater, while landmarks in the eye area had poorer reproducibility in males. Moreover, the deviations between intra- and inter-rater in males were smaller than in females.
We did not notice significant differences in the ranking of landmark reproducibility between genders overall, except for the Sellion right and Nostril base point left and right. Among both female and male groups, Pronasale (prn) was the most reproducible landmark, while Zygion left was the least reproducible landmark (Table 4 and Supplement Table 1).
Some landmarks differed in the reproducibility levels in intra-rater and inter-rater assessments as follows: Cm and Stb were highly reproducible (<0.5 mm) in intra-rater and moderately (<1 mm) in inter-rater assessment for the Caucasian sample. In Asian sample, Se right and Se left were moderately reproducible (<0.5 mm) in intra-rater and poorly reproducible (>1 mm) in inter-rater assessment.
The Representative Landmarks in Ethnic and Gender Subgroups
Bland–Altman plots are used to illustrate the consistency level between the values of each 3D coordinate (X, Y, and Z) for the facial landmarks. Some representative coordinates of facial landmarks are given in Fig. 3 to illustrate the high, moderate, and low levels of consistency between the measurements obtained from intra-rater assessment of the ethnic subgroup. Figure 3a indicates that the landmark Columella (Cm) was highly reproducible (<0.5 mm) in the X-plane for Caucasians and Asians. Figure 3b indicates that the landmark Menton (M) was moderately reproducible (>0.5 mm) in the Y-plane. Figure 3c indicates that the landmark Trichion (Tri) was poorly reproducible (>1 mm) in the Z-plane. Figure 4 exhibits some representative coordinates of the measurements obtained from the inter-rater assessment for facial landmarks in gender subgroup. Figure 4a indicates that the landmark Pronasale (Prn) was highly reproducible (<0.5 mm) in the X-plane for both females and males. Figure 4b indicates that the landmark Nasion (N) was moderately reproducible (>0.5 mm) in the Y-plane. Figure 4c indicates that the landmark Supratip break point (Stb) was poorly reproducible in the Z-plane (>1 mm).
Discussion
Facial soft-tissue landmarks and their anthropometric measurements play an important role in the clinical practice of numerous medical disciplines, particularly in reconstructive and aesthetic plastic surgery, otorhinolaryngology along with oral and maxillofacial surgery. Landmark-based cephalometric measurements facilitate diagnosis, counseling, and treatment planning, as well as an objective evaluation of a therapeutic outcome. The reproducibility of facial soft-tissue landmarks has been studied in detail on 2D photography and several classic facial landmarks have been validated for their utility in 3D surface imaging [15, 21, 23].
In the current study, we identified several non-traditional facial soft landmarks based on daily clinical experience and validated all 46 landmarks in gender-identical Caucasian and Asian samples. We explored the reproducibility of these soft-tissue landmarks on 3D facial images of our two ethnic groups. The reproducibility of facial landmarks has been validated in the three spatial planes and our measurements showed that the majority coordinates in x, y, z axes of the 46 landmarks are reproducible to less than 1 mm, which is clinically acceptable (87% intra-rater and 73.2% inter-rater). The reproducibility of the intra-rater evaluation was higher than that of the inter-rater.
Based on our measurements in different intra-rater and inter-rater sessions, we inferred the following criteria involved in facial landmarks' reproducibility. First, the clear description and definition of landmarks. Second, the morphology and contour of the facial area in which the landmark is located. Landmarks located in more projecting or well-defined areas have a higher reproducibility. For instance, Prn and TDP are more reproducible than Zy, which is located on a flatter site. Third, features and characteristics of landmarks. Landmarks with distinctive features usually have a higher reproducibility. Four, the ethnicities and genders with different facial features and characteristics. Five, rater dependency. Examples include level of attention, discipline and consistency, proficiency in 3D imaging software and knowledge of facial anatomy. Six, the quality of 3D imaging. The landmarks on the artifact-free and defect-free areas are more reproducible.
In addition, the reproducibility of our landmarks varies in the three planes. For example, certain landmarks are harder to locate accurately on one axis than on the other two. Previous studies have reported similar results [24, 25]. Medelnik et al. attribute this bias to the position of landmark relative to the individual's facial morphology [26]. The poorly reproducible landmarks are mostly concentrated on nose alare, chin, Trichion, and Zygion. Raters may not be able to find a suitable reference point in less clearly demarcated areas. Hair-bearing skin areas, such as the hairline, usually have lower reproducibility [27]. Moreover, it has been reported that the patient's head occasionally needs to be tilted back slightly to ensure data quality in nose and chin area, which complicates ensuring a consistent recording position [28]. Therefore, precautions should be taken in the preparation of the 3DSI to minimize hair and sitting-induced errors and to make landmark identification more precise.
Some landmarks and coordinates vary by race. Landmarks distributed in the nasal tip and nostrils, such as Nl, Nb, Nt, Nm, TDP, Stb, and Prn, are more reproducible in Caucasians. Nasal anatomical features of Caucasians differ from those of Asians. A previous study found that the Caucasian descent typically has relatively thick nasal skin, straighter dorsum, more pronounced nasal tip, and teardrop shape nostrils. Correspondingly, the bony vault in Asians is usually wide and short. The dorsal aesthetic lines were not clearly defined, resulting in a less well-defined TDP. The nasal tip was widened with wide alar bases. The nasal length was shortened, with diminished tip projection and horizontally oriented nostrils [29]. These factors could impact the raters to identify the nasal tip and the nostril axis and to locate the landmarks associated with these regions in Caucasians. Thus, before clinical use of a 3D landmark-based study, the reliability of the measurements for different ethnicities would need to be investigated separately.
We also observed slight gender-dependent differences in landmark placement accuracy. In intra-and inter-rater reproducibility assessment, the landmark Zy on both sides of the x-axis produced fewer errors in males. Our observation that males have larger and more pronounced zygomatic bone than females, which facilitates the rater to position Zy on the x-axis. Previous anthropometric studies have shown that males have more angular chin and jawbones than females. Male's jaws are on average 17% higher vertical and have more lateral fullness, which may affect the 3D placement of landmarks in these areas [30]. These features may make it easier to locate the landmarks Ls, Li, Me, and C.
In terms of device technology, it should be noted that Vectra XT 3D Surface Imaging System has some imperfections in 3D modeling of hair-bearing skin areas and complex structures, making it prone to artifacts and distortions. Nevertheless, despite its limitations, photogrammetry still has an irreplaceable role and potential for widespread application in predicting soft-tissue contours and monitoring treatment progress, especially for patients undergoing complex rhinoplasty and maxillofacial plastic surgery, as well as consultations for orthodontic treatment or orthognathic surgery.
Conclusion
Before being widely used in clinics, the reproducibility of each facial landmark should be verified on the x, y, and z three planes. In order to obtain good reproducibility, the rater placing landmarks must clearly define and thoroughly understand their definitions. Landmarks located at different positions on the face have broad variation in reproducible levels; the landmarks placed on clear features and boundaries area have higher reproducibility than those placed on flat or a gently curved plane. This may be related to gender and ethnic differences in facial morphology, leading to variations in the reproducibility of certain landmarks. It is also essential for raters to have sufficient knowledge of facial anatomy and proficiency in 3D images to improve the reproducibility of landmarks. In this study, the majority of the 138 coordinates from 46 facial landmarks had a reproducibility of less than 1 mm, which is clinically acceptable (87% intra-examiner and 73.2% inter-examiner). Therefore, 3D scanning with Vectra XT 3D Surface Imaging System meets the requirement of cephalometry based on facial soft-tissue landmarks in daily clinical practice. Meanwhile, race and gender reproducibility bias of the different landmarks should be taken into account during the evaluation.
References
Feng Z, Guiyang Z, Weijie F et al (2018) Application of 3D printing technology in RGPCL simulation fitting. Med Hypotheses 113(1):74–76
Garyfalia L, Peter C, Grant SH et al (2016) Evolution of preoperative rhinoplasty consult by computer imaging. Facial Plast Surg 32(1):80–87
Farkas LG (1996) Accuracy of anthropometric measurements: past, present, and future. Cleft Palate Craniofac J 33:10–18
Savran A, Alyüz N, Dibeklioğlu H et al (2008) Bosphorus database for 3D face analysis. Biometrics and identity management. Springer, Berlin Germany
Xi Z, Dell AE, Liming C et al (2011) Accurate landmarking of three—dimensional facial data in the presence of facial expressions and occlusions using a three-dimensional statistical facial feature model. IEEE Trans Syst, Man, Cybern, Part B: Cybern 41(5):1417–1428
Koban KC, Perko P, Etzel L et al (2020) Validation of two handheld devices against a non-portable three-dimensional surface scanner and assessment of potential use for intraoperative facial imaging. J Plast Reconstr Aesthet Surg 73(1):141–148
Bechtold TE, Göz TG, Schaupp E et al (2012) Integration of a maxillary model into facial surface stereophotogrammetry. J Orofac Orthop 73:126–137
White JD, Castrillon AO, Virgo C et al (2020) Sources of variation in the 3dMDface and Vectra H1 3D facial imaging systems. Sci Rep 10(1):4443
Daniele G, Valentina P, Annalisa C et al (2018) Are portable stereophotogrammetric devices reliable in facial imaging? A validation study of VECTRA H1 Device? J Oral Maxillofac Surg 76(8):1772–1784
Proffit WR, White RP, Sarver DM (2003) Contemporary treatment of dentofacial deformity. St. Louis: CV Mosby Co
Koudelovea J, Bruzek J, Caganova V et al (2015) Development of facial sexual dimorphism in children aged between 12 and 15 years: a three-dimensional longitudinal study. Orthod Craniofac Res 18:175
Pucciarelli V, Bertoli S, Codari M et al (2017) The face of Glut1-DS patients: A 3D craniofacial morphometric analysis. Clin Anat 30:644
Baik HS, Jeon JM, Lee HJ (2007) Facial soft-tissue analysis of Korean adults with normal occlusion using a 3-dimensional laser scanner. Am J Orthod Dentofacial Orthop 131:759–766
Dindaroğlu F, Kutlu P, Duran GS et al (2016) Accuracy and reliability of 3D stereophotogrammetry: a comparison to direct anthropometry and 2D photogrammetry. Angle Orthod 86(3):487–494
Khambay B, Nairn N, Bell A et al (2008) Validation and reproducibility of a high-resolution three-dimensional facial imaging system. Br J Oral Maxillofac Surg 46(1):27–32
Metzler P, Sun Y, Zemann W et al (2014) Validity of the 3D VECTRA photogrammetric surface imaging system for cranio-maxillofacial anthropometric measurements. Oral Maxillofac Surg 18:297–304
Toma AM (2007) A three-dimensional analysis of facial morphology using laser-scan imaging technology. Thesis (MPhil). Wales, UK: Applied Clinical Research and Public Health, Dental School, Cardiff University
Farkas LG (1994) Examination. In: Farkas LG (ed) Anthropometry of the Head and Face, 2nd edn. Raven Press, New York, pp 3–56
Swennen GR et al (2006) Three-dimensional cephalometry: a color altas and manual. Springer Press
Jamilian A, Darnahal A, Hamedi R et al (2016) Photogrammetric analysis of facial profile in Persian adults. Gen Dent 64(2):52–55
Toma AM, Zhurov A, Playle R et al (2009) Reproducibility of facial soft tissue landmarks on 3D laser-scanned facial image. Orthod Craniofac Res 12(1):33–42
Fink M, Medelnik J, Strobel K (2014) Metric precision via soft-tissue landmarks in three-dimensional structured-light scans of human faces. J Orofac Orthop 75(2):133–143
Ferrario VF, Sforza C, Poggio CE, Serrao G (1996) Facial three-dimensional morphometry. Am J Orthod Dentofacial Orthop 109:86–93
De Oliveira AE, Cevidanes LH, Phillips C et al (2009) Observer reliability of three-dimensional cephalometric landmark identification on cone-beam computerized tomography. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 107:256–265
Park SH, Yu HS, Kim KD et al (2006) A proposal for a new analysis of craniofacial morphology by 3-dimensional computed tomography. Am J Orthod Dentofacial Orthop 129:600.e23-600.e34
Medelnik J, Hertrich K, Steinhäuser-Andresen S et al (2011) Accuracy of anatomical landmark identification using different CBCT- and MSCT-based 3D images. An in vitro study. J Orofac Orthop 72:261–278
Berneburg M, Schubert C, Einem C et al (2010) The reproducibility of landmarks on three-dimensional images of 4- to 6-year-old children. J Orofac Orthop 71:256–264
Heike CL, Upson K, Stuhaug E, Weinberg SM (2010) 3D digital stereophotogrammetry: a practical guide to facial image acquisition. Head Face Med 6:18
Villanueva NL, Afrooz PN, Carboy JA, Rohrich Rod J et al (2019) Nasal analysis: considerations for ethnic variation. Plast Reconstr Surg 143(6):1179e–1188e
Ousterhout DK (2011) Dr. Paul tessier and facial skeletal masculinization. Ann Plast Surg 67(6):S10-5
Acknowledgements
We greatly thank the China Scholarship Council (CSC) for supporting the work of Zhouxiao Li (No. 201708320345).
Funding
Open Access funding enabled and organized by Projekt DEAL.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
All authors declare that they have no competing interests. None of the authors is shareholders in one of the named companies whose medical devices, hardware, and software were used in the study, nor did they have any other financial interest in the named company.
Informed consents
Written informed consents were obtained from all participants correspond with the Declaration of Helsinki protocols (1996). This study was performed in line with local laws and good clinical practice (1996) and approved by the Ethics Committee of the Ludwig-Maximilians-University Munich (Reference Number 266-13).
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Below is the link to the electronic supplementary material.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Li, Z., Giunta, R.E., Frank, K. et al. Reproducibility of Novel Soft-Tissue Landmarks on Three-Dimensional Human Facial Scan Images in Caucasian and Asian. Aesth Plast Surg 46, 719–731 (2022). https://doi.org/10.1007/s00266-021-02642-4
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00266-021-02642-4