Abstract
Background
The THUNDERBEAT is a multi-functional energy device which delivers both ultrasonic and bipolar energy, but there are no randomized trials which can provide more rigorous evaluation of the clinical performance of THUNDERBEAT compared to other energy-based devices in colorectal surgery. The aim of this study was to compare the clinical performance of THUNDERBEAT energy device to Maryland LigaSure in patients undergoing left laparoscopic colectomy.
Methods
Prospective randomized trial with two groups: Group 1 THUNDERBEAT and Group 2 LigaSure in a single university hospital. 60 Subjects, male and female, of age 18 years and above undergoing left colectomy for cancer or diverticulitis were included. The primary outcome was dissection time to specimen removal (DTSR) measured in minutes from the start of colon mobilization to specimen removal from the abdominal cavity. Versatility (composite of five variables) was measured by a score system from 1 to 5 (1 being worst and 5 the best), and adjusted/weighted by coefficient of importance with distribution of the importance as follow: hemostasis 0.275, sealing 0.275, cutting 0.2, dissection 0.15, and tissue manipulation 0.1. Other variables were: dryness of surgical field, intraoperative and postoperative complications, and mortality. Follow-up time was 30 days.
Results
60 Patients completed surgery, 31 in Group 1 and 29 in Group 2. There was no difference in the DTSR between the groups, 91 min vs. 77 min (p = 0.214). THUNDERBEAT showed significantly higher score in dissecting and tissue manipulation in segment 3 (omental dissection), and in overall versatility score (p = 0.007) as well as versatility score in Segment 2 (retroperitoneal dissection p = 0.040) and Segment 3 (p = 0.040). No other differences were noted between the groups.
Conclusions
Both energy devices can be employed effectively and safely in dividing soft tissue and sealing mesenteric blood vessels during laparoscopic left colon surgery, with THUNDERBEAT demonstrating some advantages over LigaSure during omental dissection and tissue manipulation.
ClinicalTrial.gov # NCT02628093.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
The growing use of laparoscopic surgery for benign and malignant colorectal diseases during the last two decades has prompted development of new small caliber (5 mm) electrosurgical instruments for safe and effective hemostasis, vessel sealing, and tissue dissection [1,2,3]. Furthermore, the aim of these energy devices has also been to shorten operative time, lessen the thermal spread, and reduce the need for instrument exchange. The currently available energy devices used during laparoscopic colorectal surgery utilize three different energy-based methods: monopolar (ME), bipolar (BE) and ultrasonically activated electrosurgery (UAS) (4-Milsom). Amongst the most generally available energy devices in the USA are LigaSure™ and Sonicision, Medtronic, USA; Harmonic Ace, Ethicon Endo-Surgery, USA; and THUNDERBEAT (TB), Olympus, Japan. All are insertable into the abdominal cavity via 5 mm port. The TB is a multi-functional energy device which delivers simultaneously ultrasonic and bipolar energy. This allows surgeons to seal and safely divide blood vessels up to 7 mm, cut and dissect omental and mesentery tissue, and potentially reduce the need for instrument exchange [2, 4, 5]. UAS devices are multi-functional similarly to TB, but are approved to only seal vessels up to 4–5 mm in diameter [6, 7]. Bipolar electrosurgical technology has been widely used in laparoscopic bowel resection and considered a safe method for dissection and vessel ligation [1, 8,9,10]. Recent studies comparing TB to other energy devices suggest that all are safe and effective to use in laparoscopic colorectal surgery and report similar intraoperative and postoperative outcomes [2, 11, 12]. However, most of these studies are retrospective cohort studies or small prospective cohorts. There are no randomized controls trials which can provide more rigorous evaluation of the clinical performance of TB compared to other energy-based devices in laparoscopic colorectal surgery.
Aims/objective
The aim of this randomized trial was to compare the clinical performance between the THUNDERBEAT and Maryland LigaSure Energy Devices in performing soft tissue dissection, dividing and sealing blood vessels in patients undergoing left laparoscopic colectomy.
Materials and methods
Study design
This was a pilot prospective randomized trial in a single academic institution. Patients undergoing laparoscopic left colectomy for their medical condition were randomized with equal chances into one of two groups: Group 1—THUNDERBEAT and Group 2—LigaSure. After randomization, the study involved prospective data collection before, during and after surgery and video recording of the surgery. All surgeries were carried out according to the regular surgical and anesthesia care. The follow up after surgery was 30 days. The study was approved by the Institutional Review Board.
Population
Sixty patients, male and female, age 18 years and above and American Society of Anesthesiologists (ASA) class between 1 and 3, undergoing elective left laparoscopic colectomy for neoplasm or diverticulitis, were included in the study after providing research informed consent. Patients with morbid obesity (body mass index, BMI > 35), multiple previous abdominal surgeries, on anticoagulants prior to surgery, coagulopathy disorders, pregnant women, and those to whom electrosurgery is contraindicated were excluded from the study.
Outcomes
The primary outcomes were: (1) dissection time to specimen removal (DTSR), and (2) versatility score. The primary outcome DTSR was defined as time to specimen removal measured in minutes from the start of colon mobilization to specimen removal from the abdominal cavity. Versatility (composite of five variables) was measured by a score system from 1 to 5 (1 being worst and 5 the best) for each of the six specific segments (Fig. 1) and adjusted/weighted by coefficient of importance with distribution of the importance as follow: hemostasis 0.275, sealing 0.275, cutting 0.2, dissection 0.15, and tissue manipulation 0.1. A mean score of 3.5 and above was considered a high versatility and below 3.5 and lower as low versatility (4). The overall versatility score is presented as an average of the mean versatility scores from each of the six surgical segments each evaluated using the score displayed on Table 1. The Versatility score was developed before the trial and based on the surgeon’s experience about the relative importance of the five variables included in the versatility score.
The ease of use of instruments was evaluated with a survey evaluating the surgeon’s opinion on the instrument immediately after surgery consisting of 8 questions on the scale from 1 to 10, where a score of 1 is worst and 10 is the best (Table 2). The secondary outcome, drier surgical field was an evaluation of the entire surgical field for overall oozing of blood or any other body fluids using a scoring system from 1 to 5 as described on Table 3.
Other outcome data collected were failure of the energy instruments to control bleeding, complications related to use of the instrument, visible thermal spread, postoperative bleeding requiring intervention, thermal injuries manifestation after surgery, reoperation, readmission, length of hospital stay and mortality. Patients were assessed daily after surgery until discharge and at 30 days following surgery.
Study instruments
THUNDERBEAT 5 mm to 35 cm (Olympus, Japan)
The TB device has been cleared under 510 (K) by FDA and currently used for regular care. The surgeons were able to coagulate blood vessels up to 7 mm, cut and dissect during surgery. The device consists of: THUNDERBEAT device and generator (Fig. 2a). The device is provided sterile and intended for single use only. The generators settings were the same for all cases in this study.
Maryland LigaSure™ 5 mm to 37 cm (Medtronic, USA)
Maryland LigaSure device is an FDA approved device and currently is used throughout the USA in surgery for tissue dissection and vessel ligation. The device consists of the LigaSure device and Generator-Force Triad (Fig. 2b). The device is provided sterile and intended for single use only.
All participating surgeons were trained in using TB and Maryland LigaSure devices as they are used daily for regular surgical care, and surgeons used the devices on at least 10 patients prior to this study.
Equipment
All surgeries were performed using the same laparoscopic equipment from Olympus and all cases were recorded with the Olympus video recording systems. The equipment included, High Definition LCD Monitor, Xenon Light Source, C02 Insufflator, laparoscopic camera 5 Endoeye Flex Deflectable scope, and Video System Center. Also, for all study cases the same THUNDERBEAT and Force Triad Generator were used with preset mode.
Statistical considerations
Blocked randomization using the method of random permuted blocks was used in the trial. A series of randomized blocks of four was generated with a 1:1 allocation ratio to allow for an equal number of patients in the two groups, by independent statistician to minimize bias. Group status was defined by the two different instruments: Group 1 THUNDERBEAT and Group 2 LigaSure. The randomization was generated by a biostatistician, and subjects were assigned to the groups at the start of the surgery using randomization envelopes.
This study was designed as a pilot (exploratory) randomized study. Because at the time this study was planned there were no other studies in colorectal surgery evaluating THUNDERBEAT that could have been used for estimation of dissection time, the study was not powered to detect a specific difference in DTSR between the groups. Post hoc analysis was performed to compare the primary outcome DTSR between the groups.
The primary outcome DTSR between the groups was evaluated using Wilcoxon rank-sum test. Demographic, preoperative, and postoperative variables were compared between groups by the Wilcoxon rank-sum test for continuous variables and the χ2 test/Fisher’s exact test for categorical variables, as appropriate. The versatility score was calculated as defined. The surgeon’s opinion evaluation survey is presented as number and percentages, and then compared between the two instrument groups. All p-values are two-sided with statistical significance at the 0.05 α level. All analyses were performed in IBM SPSS Version 25 (Armonk, NY: IBM Corp).
Results
For presenting the result in this manuscript, we used the CONSORT (Consolidated Standards of Reporting Trials) protocol guidance for more transparency and better quality of the report on RCT [13]. Seventy three patients were enrolled in the study from February 2016 to April 2019, and sixty of them were randomized into the two groups: n = 31 in Group 1: THUNDERBEAT and n = 29 in Group 2: LigaSure and included in the analyses. The other 13 patients were withdrawn from the study prior to the randomization for variety of reasons such as surgery cancellation, change of surgical plan or change in medical conditions excluding the subjects. All patients completed the 30 days follow up time. There were no statistical differences between the two groups in terms of demographics—age, sex, BMI, ASA, preoperative diagnosis, and preoperative comorbidities (Table 4). No significant difference was found between the groups in the primary outcomes, DTSR (p = 0.214) and the total time of the surgical procedure (p = 0.311). Detailed results of the intraoperative outcomes are listed in Table 5. Significant differences in versatility overall score and in surgical segment 3 versatility score were observed between the groups, with TB device having a higher score than LigaSure device (p = 0.045 and 0.041 respectively). In addition, TB demonstrated significantly higher scores in tissue dissection and tissue manipulation in surgical segment 3, and higher sealing score in segment 6 with p = 0.011, 0.026 and 0.023, correspondingly (Table 5). There were no significant differences between the groups regarding the coagulation tests PT, PTT, and INR (p = 0.463, 0.122, and 0.069, respectively).
Postoperatively, there were no significant differences between the groups in restoration of the gastrointestinal tract, postoperative complication within 30 days, and other hospital parameters of surgical recovery (Table 6). There was no mortality observed. Surgeon survey results are listed in Table 7. Surgeon’s overall satisfaction with the instrument was significantly higher for the THUNDERBEAT device (p = 0.015). LigaSure demonstrated a higher score for “less pain and less fatigue” from use the of instrument handle but it was not statistically significant (p = 0.08) (Table 7).
Discussion
The goal of this randomized trial was to evaluate the clinical performance between TB and Maryland LigaSure Energy Devices with primary outcome DTSR. Surgical time has become an important intraoperative factor along with safe and effective hemostasis and tissue dissection, as it is directly related to the cost-effectiveness of the surgical procedure. Development of new advanced energy based surgical instruments for laparoscopic surgery have shortened the surgical procedure time and have replaced the conventional hemostasis tools such as sutures, clips and staples with fast, safe and effective hemostasis. Our study did not find significant difference between the TB and LigaSure devices, consistent with the finding of other recent publications [2, 12]. A retrospective study from Italy compared outcomes and cost between TB and three other energy-based devices in patients undergoing laparoscopic colorectal resection, did not find any significant difference in the operative time and other outcomes between the groups [2]. Similar findings were reported by Shuradja in 2018 from a prospective study with retrospective cohort for comparison of TB to LigaSure and Harmonic Ace. The study reported no significant difference between the devices in operative time and suggested that they are equally safe and effective [12]. In our study, the 12-min difference in DTSR between the groups (higher in TB group) was neither clinically nor statistically significant. A post hoc power calculation shows that a statistically significant 12-min difference between the two groups could only be detected with a power of 24%. The current study, with 30 patients per group, if adequately powered at 80%, would be able to detect a difference of 27 min or greater between the two groups. Nonetheless, since 12 min may affect the cost in the operating room, the effect of the surgeon’s experience with the energy device on the operative time was evaluated. We found significantly less time to specimen removal between one experienced surgeon compared to the rest of the surgeons, which may explain the 12 min difference between the groups. Since there is no significant difference between the devices in the time to specimen removal and the overall total time of the surgery, number of applications until vessel coagulation achieved, and number of instruments exchanges in/out of the abdomen, we did not conduct a cost analysis. The actual cost between the devices is also similar (TB $475 vs. LigaSure $495). However, the device prices may differ between different institutions. Allaix et al. also did not find any cost difference between TB and other energy devices [11].
The versatility of the new surgical devices allowing multiple functions of the device is important for shortening the surgical time and minimizing the instruments exchange while providing safe and effective tissue dissection and vessel sealing. In order to have more objective evaluation of the clinical usefulness of the energy devices in the study, we used a “versatility score” (see methods) [4, 5]. In this study, the TB demonstrated significantly higher overall versatility score vs. LigaSure (4.8 vs. 4.7, p = 0.045), clinically inconsequential, as both instruments had high versatility scores. Two previous studies have reported significantly higher versatility score for TB device compared to other surgical devices, but this versatility difference was not reported to affect the clinical performance of the devices [4, 5].
The TB tip (Fig. 2a, b) allows making a swift entry into a new surgical plane, which is a technical differentiation between it and the LigaSure. Both instruments demonstrated safe and effective clinical performance in vessel sealing of left colic artery, sigmoid arteries, superior rectal arteries and large vessels such as IMA and IMV. There was no significant difference in the number of applications until vessel coagulation was achieved for sealing of all vessels including IMA and IMV. In two cases, with calcified vessels (one in TB group and one in LigaSure group), the numbers of applications were higher, and clips had to be used to achieve vessel sealing and hemostasis. This is an important concept in all compression energy devices.
One of the main concerns when working with ultrasonic energy is thermal spread. With TB and other energy devices like Harmonic Ace (Johnson & Johnson), there is a high temperature at the vibrating jaws, which can reach up 200 °C compared to 100 °C in bipolar instruments, theoretically increasing the risk of thermal injuries [4, 14]. Seehofer et al. using a Thermocamera reported that after repeated activation TB reaches 172 ± 7 °C, and 60 seconds are needed to decline to a safer 60 C [15, 16]. Despite these data, no intraoperative or postoperative delayed thermal injuries occurred in our study.
This study also evaluated the “dryness” of the surgical field, assessing for generalized oozing during mesenteric and vessel dissection, and lysis of adhesions [17]. Both groups had similar scores (Table 5). Patient’s coagulation factors PT, PTT and INR were evaluated prior to surgery, and patients on anticoagulants prior to surgery or coagulopathy disorders were excluded from the study in order not to affect the instruments hemostasis and sealing evaluation. During surgery all patients had warm air blanket and were well monitored for hypothermia.
Postoperatively, the groups did not differ significantly in complication rates. Three patients in Group 1 and four in Group 2 experienced rectal bleeding not related to the energy devices but rather to the intestinal anastomosis. No patients experienced abdominal bleeding postoperatively requiring reoperation. One patient required transfusion postoperatively following rectal bleeding from their anastomosis. Two patients were treated for anastomotic leak and recovered successfully, with one requiring reoperation and temporary stoma. Neither event appeared to have any relation to use of an energy device.
Surgeon’s overall satisfaction with the instruments showed a significant preference for the TB regarding ease of opening and closing the handle, ease to maneuver the instrument, and the ease in pushing the handle seal and cut buttons (Table 7). Less fatigue and pain were reported with the LigaSure, but it was not statistically significant. While the survey was designed to evaluate the surgeon’s satisfaction with the energy devices, the results are surely influenced by the surgeon’s individual preference and experience, as only five surgeons participated in the study and half of the patients were operated on by one experienced surgeon. Therefore, the significant difference between the groups should not be interpreted as clinically important or be generalized to other surgeons’ experience with LigaSure or TB energy devices. However, this information may be useful for improvement of the device’s features. A few pros and cons about the technical aspect of the devices are presented in Fig. 3.
Limitations of the study include the following:
-
(1)
It was conducted in a single academic institution with highly experienced colorectal surgeons in laparoscopy and surgical energy devices.
-
(2)
The study used the Maryland LigaSure Energy Device for comparison, which also was a new device at that time this study was designed. It is possible that other style LigaSure energy devices (e.g., blunt tipped) may have performed better than the one used in this study.
-
(3)
At the time we embarked on this study, there were no other studies in colorectal surgery evaluating THUNDERBEAT to use for estimation of dissection time. Post hoc analysis was performed to compare the DTSR between the groups, thus the results from this study may not be generalizable.
In summary, the TB and LigaSure devices compared in this randomized trial did not show significant difference between the study groups in the primary outcome, time to specimen removal. No patient in either group had any complication related to the energy devices, most importantly bleeding or thermal injury. A versatility score comparing the instruments is significantly different favoring TB, but not clinically significant as both groups have very high versatility scores. The study did not find any significant difference in intraoperative or postoperative complications or any complications related to the energy devices.
In conclusion, in a randomized control trial comparing THUNDERBEAT and LigaSure compressive energy devices, both were effective and safe in dividing soft tissue and sealing mesenteric blood vessels during laparoscopic left colon surgery, with THUNDERBEAT demonstrating some advantages over LigaSure during omental dissection and tissue manipulation.
References
Targarona EM, Balague C, Marin J et al (2005) Energy sources for laparoscopic colectomy: a prospective randomized comparison of conventional electrosurgery, bipolar computer-controlled electrosurgery and ultrasonic dissection. Operative outcome and costs analysis. Surg Innov 12:339–344
Allaix ME, Arezzo A, Giraudo G, Arolfo S, Mistrangelo M, Morino M (2017) The THUNDERBEAT and other energy devices in laparoscopic colorectal resections: analysis of outcomes and costs. J Laparoendosc Adv Surg Tech 27:1225–1229
Devassy R, Hanif S, Krentel H, Verhoeven HC, Torres-de la Roche LA, De Wilde RL (2019) Laparoscopic ultrasonic dissectors: technology update by a review of literature. Med Devices (Auckl NZ) 12:1
Milsom J, Trencheva K, Monette S, Pavoor R, Shukla P, Ma J, Sonoda T (2012) Evaluation of the safety, efficacy, and versatility of a new surgical energy device (THUNDERBEAT) in comparison with Harmonic ACE, LigaSure V, and EnSeal devices in a porcine model. Laparoendosc Adv Surg Tech A 22(4):378–386
Milsom JW, Trencheva K, Sonoda T, Nandakumar G, Shukla PJ, Lee S (2015) A prospective trial evaluating the clinical performance of a novel surgical energy device in laparoscopic colon surgery. Surg Endosc 29(5):1161–1166
Bergamaschi R, Yavuz Y, Marvik R (2003) Laparoscopic bowel resection: a comparison of three ultrasonically activated devices. JSLS 7(1):19–22
Gossot D, Buess G, Cuschieri A et al (1999) Ultrasonic dissection for endoscopic surgery. The E.A.E.S. Technology Group. Surg Endosc 13(4):412–417
Campbell PA, Cresswell AB, Frank TG et al (2003) Real-time thermography during energized vessel sealing and dissection. Surg Endosc 17(10):1640–1645
Takada M, Ichihara T, Kuroda Y (2005) Comparative study of electrothermal bipolar vessel sealer and ultrasonic coagulating shears in laparoscopic colectomy. Surg Endosc 19(2):226–228
Campagnacci R, de Sanctis A, Baldarelli M et al (2007) Electrothermal bipolar vessel sealing device vs. ultrasonic coagulating shears in laparoscopic colectomies: a comparative study. Surg Endosc 21(9):1526–1531
Allaix ME, Furnée EJ, Arezzo A, Mistrangelo M, Morino M (2016) Energy sources for laparoscopic colorectal surgery: is one better than the others? J Laparoendosc Adv Surg Tech 26(4):264–269
Suhardja TS, Norhadi S, Ee E, Hodgkins B (2018) Comparison of the THUNDERBEAT and other energy devices in laparoscopic colorectal resection: a single-center experience. J Laparoendosc Adv Surg Tech 28(12):1417–1421
Grant S, Mayo-Wilson E, Montgomery P et al (2018) CONSORT-SPI 2018 Explanation and Elaboration: guidance for reporting social and psychological intervention trials. Trials 19:406. https://doi.org/10.1186/s13063-018-2735-z
Ceccanti S, Falconi I, Frediani S, Boscarelli A, Catani M, Cozzi DA (2017) The THUNDERBEAT system for tissue dissection and vascular control in laparoscopic splenectomy. Minim Invasive Ther Allied Technol 26(4):249–252
Seehofer D, Mogl M, Boas-Knoop S, Unger J, Schirmeier A, Chopra S, Eurich D (2012) Safety and efficacy of new integrated bipolar and ultrasonic scissors compared to conventional laparoscopic 5-mm sealing and cutting instruments. Surg Endosc 26(9):2541–2549
Kwak HY, Dionigi G, Kim D, Lee HY, Son GS, Lee JB et al (2016) Thermal injury of the recurrent laryngeal nerve by THUNDERBEAT during thyroid surgery: findings from continuous intraoperative neuromonitoring in a porcine model. J Surg Res 200(1):177–182
Rajagopalan S, Mascha E, Na J, Sessler DI (2008) The effects of mild perioperative hypothermia on blood loss and transfusion requirement. Anesthesiology: J Am Soc Anesthesiol 108(1):71–77
Acknowledgements
The authors would like to thank and acknowledge that this study was supported by Research Grant from Olympus. Also, the authors thank the medical illustrator Yuko Tonohira for drawing the procedure illustration.
Author information
Authors and Affiliations
Contributions
Conception and design: JWM, KT, PJS, PC, KG. Acquisition of data: JWM, KT, PJS, KG, MPP, KM. Analysis and interpretation of data: KT, PC, JWM, PJS, KG. Drafting the article or revising it critically for important intellectual content: KT, JWM, PC, PJS, KG, MPP, KM. Final approval of the version to be published: JWM, KT, PC, PJS, KG, MPP, KM.
Corresponding author
Ethics declarations
Disclosures
This study was investigator-initiated and industry sponsored trial. supported by Research Grant from Olympus. The authors Koianka Trencheva, Dr. PH, MS, Kota Momose, MD, Miroslav P. Peev, MD, Paul Christos, Dr. PH, MS, Parul J. Shukla, MD, and Kelly Garrett, MD have no conflicts of interest or financial ties to disclose. Dr. Jeffrey Milsom was the primary investigator receiving the Olympus Grant and has had a consulting agreement with Olympus in the past.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
This study has been accepted as a poster at ASCRS 2019.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Milsom, J.W., Trencheva, K., Momose, K. et al. A pilot randomized controlled trial comparing THUNDERBEAT to the Maryland LigaSure energy device in laparoscopic left colon surgery. Surg Endosc 36, 4265–4274 (2022). https://doi.org/10.1007/s00464-021-08765-8
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00464-021-08765-8