Abstract
This study compared standard of care testing (SOC) to BioFire® FilmArray® Pneumonia plus Panel (PNplus). PNplus detects 15 bacteria with semiquantitative log bin values, 7 antibiotic resistance markers, three atypical bacteria (AB), and eight viral classes directly from bronchoalveolar lavage-like specimens (BLS) and sputum-like specimens (SLS). Fifty-two laboratories from 13 European countries and Israel tested 1234 BLS and 1242 SLS with PNplus and SOC. Detection rates and number of pathogens/samples were compared for PNplus pathogens. PNplus bin values and SOC quantities were compared. Three thousand two hundred sixty-two bacteria in PNplus were detected by PNplus and/or SOC. SOC detected 57.1% compared to 95.8% for PNplus (p ≤ 0.0001). PNplus semiquantitative bin values were less than SOC, equal to SOC, or greater than SOC in 5.1%, 25.4%, and 69.6% of results, respectively. PNplus bin values were on average ≥ 1 log than SOC values (58.5% 1–2 logs; 11.0% 3–4 logs). PNplus identified 98.2% of MRSA and SOC 55.6%. SOC detected 73/103 AB (70.9%) and 134/631 viruses (21.2%). PNplus detected 93/103 AB (90.3%) and 618/631 viruses (97.9%) (p ≤ 0.0001). PNplus and SOC mean number of pathogens/samples were 1.99 and 1.44, respectively. All gram-negative resistance markers were detected. PNplus and SOC results were fully or partially concordant for 49.1% and 26.4% of specimens, respectively. PNplus was highly sensitive and detected more potential pneumonia pathogens than SOC. Semiquantification may assist in understanding pathogen significance. As PNplus generates results in approximately 1 h, PNplus has potential to direct antimicrobial therapy in near real time and improve antimicrobial stewardship and patient outcomes.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Determining the etiology of community-acquired pneumonia (CAP), hospital-associated pneumonia (HAP), and ventilator-associated pneumonia (VAP) can be complicated [1]. Based on traditional test methods such as Gram stain and culture, diagnostic yield can be low [2,3,4]. Additionally, poor specimen quality may yield inconclusive or difficult to interpret results, sampling may require an invasive procedure, and/or patients may be on empiric therapy prior to specimen collection, reducing diagnostic yield [5]. Although international guidelines from the Infectious Diseases Society of America (IDSA)/American Thoracic Society (ATS) and the European Respiratory Society (ERS) recommend diagnostic testing for moderate to severe CAP [6, 7], actual compliance with guidelines can be low (16.7% and 23.9%, respectively) and differs by geography and disease severity [8]. Aside from influenza A/B, often no viral testing is performed, despite evidence that other viruses are a significant cause of CAP in all age patients [9,10,11]. Consequently, broad use of empiric antibiotic treatment for undiagnosed viral infections has been associated with increased antibiotic resistance globally [12,13,14].
In HAP and VAP, empiric therapy often includes broad-spectrum antibiotics for both gram-positive and gram-negative bacteria due to the risk of infection with multidrug-resistant pathogens [15, 16]. Identification of specific pathogen(s) is of increasing importance to allow prompt initiation of targeted and effective therapy and should have a positive impact on antimicrobial resistance and consequently on healthcare expenditure, while reducing risks for adverse events such as renal impairment and development of Clostridium difficile disease [17,18,19].
To address these unmet diagnostic needs and to provide a solution for prompt initiation of targeted therapy, the BioFire® FilmArray® Pneumonia (PN)/Pneumonia plus (PNplus) Panels (BioFire Diagnostics, LLC, Salt Lake City, UT) were developed. BioFire PN/PNplus Panels are US Food and Drug Administration (FDA) cleared and CE-IVD marked highly multiplexed nested polymerase chain reaction tests that identify the common causes of CAP, HAP, and VAP [20]. BioFire PN/PNplus Panels contain identical test reagents; however, results for Middle East respiratory syndrome coronavirus (MERS-CoV) are masked by the software for BioFire PN Panel. BioFire PN/PNplus Panels detect 15 bacteria with a semiquantitative log value (104, 105, 106, ≥ 107), three atypical bacteria (Legionella pneumophila [all serotypes], Mycoplasma pneumoniae, Chlamydia pneumoniae), viruses reported as eight groups (adenovirus, coronaviruses [OC43, NL63, HKU-1, 229E], MERS-CoV [BioFire PNplus Panel], human metapneumovirus, human rhinovirus/enterovirus [HRV/EV], influenza A, influenza B, parainfluenza viruses, respiratory syncytial virus), and 7 genetic markers of antibiotic resistance (mecA/C and MREJ, blaCTX-M, blaKPC, blaVIM, blaOXA-48-like, blaIMP, blaNDM). BioFire PN/PNplus Panels are intended for use in persons of all ages and in various settings, including outpatient, emergency department, and hospitalized. The tests are validated for lower respiratory tract specimens, including sputum-like specimens (SLS) (induced, expectorated sputa, endotracheal aspirates [ETA]) and bronchoalveolar lavage (BAL)-like specimens (BLS) (BAL, mini-BAL). Specimen preprocessing is not required, test setup is approximately 5–10 min, and time to results is approximately 1 h.
This multicenter study assessed BioFire PNplus Panel performance as compared to standard of care testing (SOC) performed at a variety of institutions with variable test utilization and laboratory practices.
Materials and methods
Clinical sites, specimens, and SOC testing
Forty-eight academic medical center laboratories and four independent medical laboratories from 13 countries (Austria, Belgium, Denmark, Israel, Italy, France, Germany, Netherlands, Portugal, Spain, Sweden, Switzerland, and the UK) tested 2476 unique specimens (1234 BLS; 1242 SLS) from adult and pediatric patients suspected of pneumonia (Supplemental Table 7). Specimen selection was at discretion of study site. SOC was performed per institutional policies and healthcare provider prescription. SOC varied by site and all included bacterial culture and phenotypic susceptibility testing when indicated. Additional test methods were at the discretion of the laboratory and may have included additional cultures (fungal, viral, mycobacterial cultures) as needed, urinary antigen testing (Streptococcus pneumoniae, L. pneumophila), immunofluorescence tests, and nucleic acid amplification tests (NAATs) for selected bacteria, viruses, and antibiotic resistance markers. Methicillin-susceptible Staphylococcus aureus (MSSA) was differentiated from methicillin-resistant S. aureus (MRSA) using phenotypic methods, mecA NAAT, and/or PBP2a latex agglutination at the discretion of the laboratory. The investigators were instructed to perform the BioFire PNplus Panel in accordance with the manufacturer’s instruction for use. Specimens were split and BioFire PNplus Panel testing was performed using either fresh specimens or from a frozen aliquot. Data was deidentified, no protected health information was provided, and participation was in accordance with local institutional ethical guidelines.
BioFire PNplus and SOC results were compared for the following:
-
1.
Detection of BioFire PNplus Panel pathogens by BioFire PNplus Panel and SOC: Results were evaluated for all specimens and by specimen types. SOC detected or not detected was determined in consideration of all test results reported. SOC C. pneumoniae, L. pneumophila, M. pneumoniae, and viral results were counted as not detected when a negative or no result was provided as patterns of testing varied extensively by institution. All SOC results were considered true positive. Results of BioFire PNplus Panel were considered true positive or true negative based on performance data established in US FDA clinical studies [20]. Mean numbers of BioFire PNplus Panel pathogens per specimen, rate, and composition of coinfections were compared to SOC. No discordant analyses were performed due to large numbers of isolates that would have been referred to a supplemental PCR. Additionally, most laboratories would not have another culture-independent method to confirm or refute results and the study was designed to mimic real-life interpretation of data.
-
2.
Number and type of microorganisms reported by SOC not on BioFire PNplus Panel: Organisms considered always as normal flora (ex viridans streptococci) were excluded from analysis, and remaining organisms were classified as potential pathogens and questionable pathogens.
-
3.
BioFire PNplus Panel and SOC results were evaluated for overall concordance. For common bacteria identified by routine culture, positive percent agreement (PPA) and negative percent agreement (NPA) for on-panel pathogens were calculated. The negative predictive value (NPV) for BioFire PNplus on panel pathogens and off panel pathogens was calculated.
-
4.
Comparison of semiquantitative BioFire PNplus Panel and SOC values for bacteria: SOC reporting varied, including no quantification, semiquantitative descriptive (rare, few, moderate, many, etc.), semiquantitative numerical (1+, 2+, 3+, 4+), and quantitative culture in log values (102–108 CFU/mL). To equate SOC to BioFire PNplus Panel bin values, standardized reference values (SRV) were established as defined in Supplemental Table 8. Data was evaluated for all SOC and BioFire PNplus Panel comparisons (n = 1297) and for a subset of comparisons where quantitative numerical culture values were provided (n = 903).
Statistical analysis
p values were calculated using a 2-specimen test for equality of proportions with a continuity correction. BioFire PNplus Panel PPA, NPA, and NPV for the common bacterial pathogens were calculated using the Clopper-Pearson method.
Results
Detection of BioFire PNplus Panel pathogens by BioFire PNplus and SOC
Of the 2476 specimens tested with BioFire PNplus, 13 specimens (0.53%) gave invalid results (5 SLS; 8 BLS) leaving 2463 specimens (1237 SLS, 1226 BLS) evaluable. BioFire PNplus Panel detected one or more pathogens in 1875 specimens (76.13%) and SOC in 1380 specimens (56.03%) (p ≤ 0.0001) (Fig. 1). In total, 3893 bacteria and viruses were detected by at least one method (Table 1). BioFire PNplus Panel detected 3743/3893 pathogens (96.15%) compared to 1995/3893 pathogens (51.25%) for SOC (p ≤ 0.0001). BioFire PNplus Panel and SOC concordance for all pathogens detected was 47.39%. Sensitivities of BioFire PNplus Panel detections ranged from 85.54 to 100%, and sensitivities of SOC detections ranged from 8.89 to 100%.
For all specimen types, 3262 bacteria were detected by at least one method (Table 1). BioFire PNplus Panel detected 3125/3262 bacteria (95.77%) and SOC 1861/3262 (57.05%) (p ≤ 0.0001). BioFire PNplus Panel and SOC concordance was 52.85%. The most frequently detected bacteria per sample by BioFire PNplus Panel were S. aureus (21.15%), Haemophilus influenzae (19.69%), and Pseudomonas aeruginosa (15.63%) and by SOC were P. aeruginosa (12.83%), S. aureus (12.71%), and H. influenzae (8.65%).
For BLS, 1358 bacteria were detected by at least one method (Supplemental Table 9). BioFire PNplus Panel identified 1288/1358 bacteria (94.85%) compared to 841/1358 (61.93%) for SOC (p ≤ 0.0001). BioFire PNplus Panel and SOC concordance was 56.77%. For SLS, 1904 bacteria were detected by at least one method (Supplemental Table 10). BioFire PNplus identified 1836/1904 (96.43%) bacteria and SOC 1020/1904 (53.57%) (p ≤ 0.0001). BioFire PNplus Panel and SOC concordance was 50%. PPAs for the individual PNplus common bacterial pathogens ranged from 68.42 to100% (mean 92.90%; 95% confidence interval [CI] 91.61 to 94.04%) for all specimens combined; 50.0 to 100% (mean 91.77%; 95% CI 89.66 to 93.56%) for BLS; and 80.95 to 100% (mean 93.74%; 95% CI 92.03 to 95.18% ) for SLS.
NPAs for the individual BioFire PNplus Panel common bacterial pathogens ranged from 87.07 to 99.31% (mean 96.10%; 95% CI 95.89 to 96.30%) for all specimens combined; 90.18 to 99.43% (mean 97.14%; 95% CI 96.89 to 97.39%) for BLS; and 83.86 to 99.19% (mean 94.98%; 95% CI 94.73 to 95.37%) for SLS. NPVs for the individual BioFire PNplus Panel common bacterial pathogens ranged from 99.04 to 99.96% (mean 99.63%; 95% CI 99.56 to 99.68%) for all specimens combined; 99.0 to 99.92% (mean 99.61%; 95% CI 99.51 to 99.69%) for BLS; and 99.15 to 99.94% (mean 99.64%; 95% CI 99.54 to 99.71%) for SLS. This evaluation did not consider the quantity of the common bacterial pathogens reported and the NPA and NPV may in practice be higher based on laboratory sample type and pathogen reporting thresholds. Overall for specimens negative by BioFire PNplus Panel but positive by SOC culture for which either a semiquantitative or quantitative result was provided, 45.1% had quantities reported by SOC (e.g., 1+, rare, 1 colony, < 103) that would be considered below the level of detection of the BioFire PNplus Panel.
C. pneumoniae, M. pneumoniae, and L. pneumophila were detected in 0.28%, 1.66%, and 2.23% of all specimens, respectively. Sensitivities of C. pneumoniae, M. pneumoniae, and L. pneumophila detections by BioFire PNplus Panel were 100% (7/7), 90.24% (37/41), and 89.09% (49/55) respectively, and SOC were 57.14% (4/7), 48.78% (20/41), and 89.09% (49/55), respectively (Table 1). Detection of M. pneumoniae was significantly better by BioFire PNplus Panel than SOC (p = 0.0001). BioFire PNplus Panel and SOC detection rates for atypical bacteria combined were 90.29% (93/103) and 70.87% (73/103), respectively (p = .0008). BioFire PNplus Panel detected 97.94% (618/631) of the total viral detections and SOC 21.24% (134/631) (p ≤ 0.0001). RV/EV was the predominant virus group detected (58.32%), followed by PIVs (10.78%) and the non-SARS-CoVs (8.24%).
Single and codetections for BioFire PNplus pathogens identified by BioFire PNplus and SOC
Distribution of negative and positive results for all specimens and breakdown of composition for BioFire PNplus Panel and SOC are shown in Figs. 2. and 1. Majority of BioFire PNplus Panel results (70.57%) and SOC results (55.75%) contained at least one bacteria, and at least one virus was detected in 20.49% and 2.93% of specimens, respectively. Mean number of bacteria and viruses detected by BioFire PNplus Panel in all specimens, BLS, and SLS were 1.99, 1.8, and 2.15, respectively, compared to 1.44, 1.44, and 1.8 respectively, by SOC (Table 2). For all specimens, 73.81% of BioFire PNplus Panel positive specimens (n = 1875) contained 1 (45.22%) or 2 (28.59%) pathogens compared to SOC positive specimens (n = 1380) where 92.32% contained 1 (65.51%) or 2 (26.81%) BioFire PNplus Panel pathogens.
Among the 1027 BioFire PNplus Panel specimens with codetections, 63.0% (647/1027) had only multiple bacteria (2 to 8); 35.44% (364/1027) had both bacteria (1 to 7) and virus(es) (1 to 3); and 1.33% (16/1207) had only 2 viruses detected. Among the 476 SOC specimens with codetections, 86.34% (411/476) had only multiple bacteria (2 to 5); 13.03% (62/476) had both bacteria (1 to 5) and virus(es) (1 to 2); and 0.63% (3/476) had 3 viruses only detected.
Pathogens detected by SOC not included in BioFire PNplus Panel
Excluding normal oral flora, 649 additional potential pathogens were detected by SOC including 230 gram-negative bacteria, the most prevalent being Stenotrophomonas maltophilia (n = 70), Citrobacter spp. (n = 31), Acinetobacter spp. (n = 28), Haemophilus parainfluenzae (n = 22), Morganella morganii (n = 15), and Hafnia alvei (n = 11). Another 28 gram-negative bacterial spp., representing 18 genera, were reported less than 10 times. Yeasts were reported for 283 specimens, Candida spp. (n = 123), Candida albicans (n = 113), C. glabrata (n = 27), C. parapsilosis (n = 6), C. krusei (n = 5), C. tropicalis (n = 5), C. kefyr (n = 2), and C. lusitaniae (n = 2). Pneumocystis sp. was reported for 11 specimens, molds for 28 specimens, including 11 Aspergillus fumigatus, and five anaerobes including three Fusobacterium spp. There were 42 gram-positive cocci of questionable significance including 24 Enterococcus spp. and seven Streptococcus C/G groups. There were 24 gram-positive rods, including 15 Corynebacterium spp. and 1 Nocardia sp. and four Mycobacterium (two M. tuberculosis, one M. avium, and one M. fortuitum). Additional viral targets (n = 22) included cytomegalovirus (n = 10), herpes simplex virus (n = 5), Epstein-Barr virus (n = 4), human herpes virus 6 (n = 2), and varicella zoster virus (n = 1).
Including BioFire PNplus Panel pathogens and additional SOC pathogens not in BioFire PNplus Panel, the number of potential pathogens for all specimens totaled 4542, of which BioFire PNplus Panel detected 82.41% (3743/4542) and SOC detected 58.21% (2644/4542) (p ≤ 0.0001). Excluding pathogens of questionable significance that may be considered colonizers (yeasts, gram-positive cocci, gram-positive rods [excluding Mycobacteria spp., Nocardia spp.], molds [due to difficulty in interpretation even in immunocompromised persons]), the total number of potential pathogens was 4274 of which BioFire PNplus Panel detected 87.58% (3743/4274) and SOC detected 55.59% (2376/4274) (p ≤ 0.0001). The NPVs for individual pathogens not included in the BioFire PNplus Panel ranged from 97.16% for S. maltophilia to 99.96% for organisms reported only once. The composite NPV of the BioFire PNplus Panel for all off-panel pathogens minus those of questionable significance was 89.12%. This evaluation did not consider the quantity of the off-panel pathogens reported and the NPV may in practice be higher based on laboratory sample type and pathogen reporting thresholds.
Concordance of BioFire PNplus Panel and SOC results
Overall, 1209/2463 specimens (49.10%) demonstrated total concordance between BioFire PNplus Panel and SOC results (Table 3). This included positive concordance (27.0%), negative concordance (5.36%) for BioFire PNplus Panel bacteria and viruses, and 16.73% for SOC specimens positive for non-BioFire PNplus Panel bacteria and viruses. 26.43% (651/2463) of specimens demonstrated partial concordance and 24.48% (603/2463) no concordance. Positive, partial, and negative concordances for BLS were 56.85%, 20.55%, and 22.59%, respectively, and 41.13%, 32.34%, and 26.35%, respectively, for SLS.
Bacterial quantification by BioFire PNplus Panel and SOC
Semiquantitative results for BioFire PNplus Panel and SOC bacteria were compared for 1297 matched detections (Table 4). Mean differences between all three levels of SOC SRVs and BioFire PNplus Panel SRVs for all bacteria, BLS bacteria, and SLS bacteria were 1.06 (range 0.96–1.14), 1.34 (range 1.19 to 1.52), and 0.90 (range 0.71 to 1.14), respectively. For a subset of specimens (n = 903) with SOC quantitative culture results, mean difference between SOC SRV and BioFire PNplus Panel SRV for bacteria was 1.18. For specimens BioFire PNplus Panel positive and SOC negative, mean BioFire PNplus Panel SRVs for all bacteria, BLS bacteria, and SLS bacteria were 3.21, 3.10, and 3.27, respectively. Using this subset of BioFire PNplus Panel positive/SOC negative results, if a threshold for BLS reporting was set at BioFire PNplus Panel bin values of 104, 105, 106, or ≥ 107, then 482, 290, 158, and 77 BioFire PNplus Panel bacteria results, respectively, would be reported as positive and discordant with SOC. If a threshold for SLS reporting was set at BioFire PNplus Panel bin values of 104, 105, 106, or ≥ 107, then 834, 540, 336, and 177 bacteria results, respectively, would be reported as discordant with SOC. Therefore, increasing the BioFire PNplus Panel bin threshold for reporting bacteria as detected would result in increased concordance with SOC reported as culture negative. Establishment of reporting guidelines in consideration of both BioFire PNplus Panel bin values and the inherent log value differences between BioFire PNplus Panel results and SOC results improve overall concordance of bacteria reporting.
SOC SRVs and BioFire PNplus Panel SRVs were equivalent for 25.37% of all bacteria, 19.38% of BLS bacteria, and 31.92% of SLS bacteria (Table 5). BioFire PNplus Panel SRVs were greater than SOC SVRs for 69.55% of all bacteria, 76.28% of BLS bacteria, and for 62.88% of SLS bacteria, with majority of BioFire PNplus Panel SRVs 1–2 logs greater than SOC SVRs. SOC SVRs were greater than BioFire PNplus Panel SRVs for 5.09% of all bacteria, 4.34% of BLS bacteria, and 5.83% of SLS bacteria, with majority demonstrating a one log difference.
Detection of MSSA and MRSA
A total of 531 specimens contained S. aureus, 97.93% (520/531) detected by BioFire PNplus Panel and 58.95% (313/531) detected by SOC (p ≤ 0.0001) with a concordance of 56.87% (Table 6). BioFire PNplus Panel did not detect 11 S. aureus (2.07%, all MSSA). A total of 24 specimens were reported to contain MSSA by SOC but were reported to contain MRSA by BioFire PNplus Panel and two specimens reported to contain MRSA by SOC were reported as MSSA by BioFire PNplus Panel. Mean SRVs for S. aureus detected by SOC, BioFire PNplus Panel, and detected only by BioFire PNplus Panel were 3.14, 4.18, and 2.92, respectively.
Detection of gram-negative resistance markers
BioFire PNplus Panel detected in 1537 specimen bacteria (Acinetobacter baumannii-calcoaceticus complex, Enterobacter cloacae complex, Escherichia coli, Klebsiella aerogenes, Klebsiella oxytoca, Klebsiella pneumoniae group, Proteus spp., P. aeruginosa, Serratia marcescens) for which ESBL and/or carbapenemase resistance genes would be reported. Of these, 185 (12.04%) had a total of 229 resistance genes (blaCTX-M [n = 133)], blaKPC [n = 67], blaIMP [n = 1], blaNDM [n = 2], blaVIM [n = 26]). Six blaOXA-48-like genes were detected in 1029 specimens (0.58%) positive by BioFire PNplus Panel for one or more of the following: E. cloacae complex, E. coli, K. aerogenes, K. oxytoca, K. pneumoniae group, Proteus spp., or S. marcescens. One resistance gene was detected in 76.76% (142/185) of specimens, two in 20.0% (37/185), and three in 3.24% (6/185). Due to large variations and inconsistencies in reporting of SOC phenotypic/genetic susceptibility data, no comparisons could be made.
Discussion
This multinational study is the largest to date that compared BioFire PNplus Panel to SOC testing, which varied extensively by site and physician prescribing practices. BioFire PNplus Panel identified significantly more positive specimens (76.13%) than SOC (56.03%) (p ≤ 0.0001) and more potential pathogens than SOC (p ≤ 0.0001) independent of specimen type. Largest discrepancies for bacterial detections were for fastidious pathogens, which may be concealed by overgrowth of normal flora, or may be non-viable, including Streptococcus pyogenes (77.27%), Streptococcus agalactiae (78.85%), Moraxella catarrhalis (61.7%), H. influenzae (53.97%), and S. pneumoniae (48.07%). Despite detection differences, the three most common bacteria identified by BioFire PNplus Panel and SOC were similar, S. aureus, H. influenzae, and P. aeruginosa. Lower SOC bacteria detection may relate to local reporting guidelines and testing of specimens from patients on antimicrobials. The most prevalent pathogen not included on BioFire PNplus Panel detected by SOC was S. maltophilia; however, the incidence was low (2.84%), similarly with other various gram-negative rods (6.5%). Despite the low prevalence of pathogens not detected by BioFire PNplus Panel, it is essential to perform culture or other ancillary testing to detect off-panel pathogens and to provide susceptibility results.
Our rate of bacteria detection (95.77%) as compared to culture was consistent with Mitton et al. (92.0%) [21] and Yoo et al. (99.3%) [22]. A study by Webber et al. demonstrated that the BioFire PNplus Panel identified most bacteria (98.4%) detected by SOC and additionally 92 bacteria including more S. aureus (23.9%) and H. influenza (27.2%) [23]. Murphy et al. identified potential pathogens in 48.82% of BLS and in 72.01% of SLS with BioFire PN Panel, detecting more S. aureus, H. influenzae, M. catarrhalis, and P. aeruginosa [20]. In this study, the overall BioFire PNplus Panel PPA and NPA for common bacteria were 92.90% and 96.10%, respectively, and similar to other studies. Clinical studies for FDA clearance demonstrated a PPA of 96.2% and NPA of 98.3% for BLS and a PPA of 96.3% and NPA of 97.2% for SLS after discordant resolution using a molecular comparator. Lee et al. demonstrated a PPA of 90% and NPA of 97.4% as compared to SOC bacteria detection, with BioFire PN Panel identifying a pathogen in 47.4% of BLS and in 60% of SLS for an overall positivity rate of 55.93% [24]. Edin demonstrated for on-panel pathogens a PPA of 100% and an NPA of 73.2% [25], and Gastli demonstrated a PPA of 94.4% and an NPA of 96.0% as compared to culture [26]. Similarly, a VAP study using a research use only version of BioFire PNplus Panel demonstrated for bacteria an 89.0% PPA and 95.9% NPA with SOC [27]. BioFire PNplus Panel reflects high performance compared to SOC, yielding additional clinically actionable results that may be missed by SOC.
Prevalence of atypical pathogens in this specimen set was low (4.17%), with L. pneumophila the most frequently detected. Low SOC percent detections of M. pneumoniae (48.78%) and C. pneumoniae (57.14%) compared to BioFire PNplus Panel (90.24% and 100%, respectively) were mainly due to lack of testing and therefore a missed opportunity in CAP to either limit treatment to a macrolide or fluoroquinolone or stop treatment if tested negative. This missed opportunity for applying antimicrobial stewardship principles and streamlining therapy is of particular importance due to adverse effects of fluoroquinolones [28]. Of the 41 patients M. pneumoniae positive, age was available for 31 (range 7 to 88 years), including 12 (range 23 to 80 years) with no SOC result. Lack of testing may be a study artifact or be indicative of local testing practices. Adults may not be tested as M. pneumoniae is often viewed as an illness of school age children, adolescents, and young adults, although studies demonstrate infections in all aged adults, with up to a 15% prevalence in persons aged 56 or older [29]. Conversely, percentage of L. pneumophila detected by BioFire PNplus Panel and SOC were similar (89.09%). There were only 6/55 specimens L. pneumophila positive with no SOC result, indicating a higher awareness compared to M. pneumoniae and C. pneumoniae.
BioFire PNplus Panel detections were limited for certain viruses (example, influenza A, influenza B) due to seasonality and time of specimen collection. SOC viral detections were additionally limited by lack of testing. Webber et al. compared standard viral testing with BioFire PNplus Panel and demonstrated a 99.2% correlation [23]. Hughes et al. found an 87% PPA and 100% NPA for BioFire PNplus Panel viral detections compared to BioFire® FilmArray® Respiratory 2 (RP2) Panel (BioFire Diagnostics, LLC) [30]. Virus detection in absence of a bacterial pathogen and in conjunction with clinical presentation, chest radiograph, and other diagnostic tests, such as a low procalcitonin, could support antimicrobial stewardship and discontinuation of antibiotics in the setting of CAP [31].
BioFire PNplus Panel identified more codetections (41.85%) compared to SOC (19.21%), which was mainly influenced by lack of SOC viral testing. Codetections were commonly identified by Webber et al. (25%) [23], Murphy et al. (29.49%) [20], and Lee et al. (42.3%) [24]. All three studies found the majority of codetections contained 2 pathogens, but could rarely contain 5 to 6+ pathogens, similar to our results. In this study, BioFire PNplus Panel on average identified more potential pathogens (1.99) per specimen than SOC (1.44). Detection of multiple pathogens raises interpretation questions that need to be viewed in light of clinical parameters, pathogens detected, and abundances.
Use of BioFire PNplus Panel has led to concerns that identification of more bacteria than SOC, which may be colonizers, could lead to antibiotic overtreatment. Specimen types should be considered. SLS are prone to more oropharyngeal contamination compared to BLS. H. influenzae, S. pneumoniae, and M. catarrhalis can be normal flora, and hospitalized or ventilated patients may be colonized with gram-negative bacilli and S. aureus. Although laboratory reporting varies depending on specimen type, patient populations, and clinical need, clinical guidelines recommend using different reporting thresholds for different specimen types. Generally, SLS bacteria are considered, in light of other variables, significant at ≥ 106 or ≥ 107 (moderate, numerous, 3+, 4+) and in BLS ≥ 103 or ≥ 104 (few, moderate, 2+ or greater). For this reason, the BioFire PNplus Panel lower limit of reporting for bacteria was set at 103.5 genomes/mL. The bin values may allow for different interpretations based on specimen type. For example, SOC and BioFire PNplus Panel concordance for bacteria in SLS was 51.23% but when only evaluating BioFire PNplus Panel results reported at 106 and ≥ 107, correlation increased to 66.8%. Additionally, more BioFire PNplus Panel negative/SOC negative results would be obtained when higher bin thresholds for reporting BioFire PNplus Panel bacteria are used. Overall, the mean SRVs for SOC were consistently ~ 1 log lower than BioFire PNplus Panel SRV when using three different interpretation schemes. Mean SRV for specimen culture negative, BioFire PNplus Panel positive, was 3.21 and would potentially equate to SOC SRV of 2.21, which likely is below the limit of detection or limit of culture reporting. Similarly, SOC S. aureus-negative specimens but BioFire PNplus positive had a SRV of 2.92.
Murphy et al. demonstrated that BioFire PN Panel bin values were accurate and reproducible within ± 0.5 log10 copies/mL and correlated with another quantitative molecular method [20]. Despite low concordance with quantitative culture, particularly when values were < 106 (3.1–38.9%), concordance improved to 90.9–100% when quantitative culture values were > 106. There were few instances when BioFire PN Panel did not detect a bacterium or reported values lower than quantitative culture, which is similar to what we report in this study. Lee et al. also demonstrated an overestimation of quantification by BioFire PN Panel [24]. Buchan et al. demonstrated that PN values were frequently higher than culture values, resulting in semiquantitative agreement (within the same log10 value) of 43.6% [32]. Gastli et al. reported that 90.1% of organisms with a BioFire PNplus Panel result of ≥ 106 grew significantly in culture [26] and Yoo et al. reported that 86% of bacteria considered significant by culture (moderate or many quantities) yielded BioFire PNplus Panel results of ≥ 107 [22 not 26] 22.
Although BioFire PNplus Panel does not make claims as to the significance of the bin value, knowing the relative bacteria abundance may be helpful in understanding coinfections and differentiating colonization versus infection. BioFire PNplus Panel bin values should be interpreted in consideration of specimen type, Gram stain, types and bin values for other pathogens detected, type of pneumonia, presence of resistance markers, biomarkers, and clinical risk factors. Laboratories should establish reporting guidelines and provide physician education in conjunction with a multidisciplinary team consisting of ID, critical care specialists, pulmonologists, stewardship committee, and infection control practitioners.
Major limitations of this study were a lack of clinical information and comprehensive gram-negative phenotypic antibiotic susceptibility data, such as that described by Murphy et al. [20], needed to better understand the relevance of resistance marker results and impact on patient care. Discordant analyses were not performed due to the large number of participating sites and the number of specimens tested. Discordant analyses would have required some type of confirmatory testing for 150 culture isolates (BioFire PNplus Panel negative) and 1898 individual PCRs or direct sample sequencing for the confirmation of pathogens only detected by BioFire PNplus Panel (SOC negative). However, this study therefore does highlight the need for microbiologists and clinicians to address issues relating to discordant results and test interpretation. Additionally, detailed information regarding testing for viral or atypical bacterial pathogens for each sample was not available so no direct performance comparison could be made. Gram stains were not systematically reported and it is not known if a quality score was a testing requirement. However, considering the high sensitivity and specificity of BioFire PNplus Panel as demonstrated in the US FDA clinical studies [20], in combination with detection of key gram-positive and gram-negative resistance genes, BioFire PNplus Panel may facilitate decisions to optimize therapy, including discontinuation, de-escalation, and escalation, which should be evaluated on a patient-by-patient basis. Depending on local epidemiology, negative results for drug-resistant pathogens may yield a high negative predictive value and may allow for therapy de-escalation in the right clinical context, an approach endorsed by IDSA CAP guidelines and in CDC Core Elements of Hospital Antibiotic Stewardship Programs [6, 33]. Similarly, several hypothetical therapeutic reviews based on BioFire PN Panel results as compared to actual treatment revealed potential for antibiotic adjustment. Buchan et al. found, based on BioFire PN Panel results, 70.7% of patients could have had an antibiotic adjustment, including discontinuation or de-escalation in 48.2% resulting in an average savings of 6.2 antibiotic days/patient [32]. Lee et al. identified a potential for BioFire PN Panel to alter antibiotic prescription in 40.7% of patients [24]. A study by Monard et al. demonstrated that a multidisciplinary committee proposed modifications of empiric therapy in 77% of pneumonia episodes, including de-escalation (40%) and escalation (22%), and in microbiologic documented cases, the BioFire PNplus Panel increased appropriateness of therapy in 87% of cases as compared to 77% in routine care [34]. This data provides an early indication that proper use and interpretation of BioFire PNplus Panel could lead to targeted therapy not an increase in inappropriate antimicrobial usage. Prospective interventional studies in progress will provide data on interpretation of bin values, detection of resistance genes, and clinical impact of a rapid diagnosis. Finally, the BioFire PNplus bin comparison to SOC reporting was difficult to standardize as culture reporting can vary from technologist to technologist and laboratory to laboratory, which could lead to interpretive error. However, despite using 3 different interpretative criteria, results did not significantly differ. Strengths of the study include the large specimen size, the even distribution of specimen types, and geographical diversity of testing sites with differences in both ordering practices and results reporting.
The clinical laboratory plays a vital role in diagnosis of CAP, HAP, and VAP but faces numerous challenges due to testing complexity [1. 5, 8-10]. Often poor-quality sputum specimens are submitted and without quality rejection screening by Gram stain, culture results can be misleading or negative. HAP and VAP patients pose a different dilemma since these patients quickly become colonized with S. aureus and various gram-negative bacilli which may lead to pneumonia with multidrug-resistant strains [15, 16]. Specimens can contain a diversity of pathogens including bacteria, viruses, and fungi [4, 6,7,8,9,10, 34,35,36,37,38]. Time to traditional bacterial detection is 24–72 h and antibiotic susceptibility data takes an additional 24–48 h. Comprehensive viral diagnostics are often not performed aside from influenza A/B testing, may not be performed 24/7, or require referral to a reference laboratory delaying time to results. Urinary antigen tests for S. pneumoniae provide results in < 30 min but can be false negative and false positive, particularly in children [39]. Urinary antigen tests for L. pneumophila are restricted in serotype detection and can have sensitivities of < 50% [40, 41]. Serology can be difficult to interpret and may require an acute and convalescent serum collected weeks apart. NAATs are the gold standard for the detection of the atypical bacteria but may not be routinely performed. Consequently, testing imitations lead to empiric HAP/VAP treatment with broad-spectrum antibiotics, especially in regions with high antimicrobial resistance rates. The switch to targeted therapy can take days, increasing risk of antimicrobial resistance and adverse events such as acute kidney injury and C. difficile disease [18]. Finally, if specimens are obtained after the start of antibiotic treatment, results may be altered or negative, without identifying the etiologic agent [5].
In conclusion, the BioFire PNplus Panel meets the challenges associated with routine test methods including poor pathogen recovery, lack of diagnostic comprehensiveness, and delayed time to result [42]. However, several factors need to be considered including the lack of a specimen quality marker and the inability to report the presence or absence of normal flora [43]. Although a Gram stain is not required prior to testing, good laboratory practice should still be followed to insure sample quality [43]. Pretreatment or dilution of samples would affect both the sensitivity of the assay and semiquantitative results and therefore is not recommended in the manufacturer’s instructions for use. Interpretation challenges include understanding the increased detection rates, significance of the bin value, the differentiation between colonization and infection, and the presence of gram-negative resistance markers without direct linkage to a specific pathogen. However, approved for use with BLS and SLS, BioFire PNplus allows for easy specimen testing for CAP [6, 7] and meets IDSA/ATS recommendations [13] for non-invasive diagnostic testing as a preferred method for VAP and ERS guidelines to test distal quantitative specimens [16]. Additionally, studies that used BioFire PNplus panels in COVID-19 patients demonstrated not only improved diagnosis of bacterial coinfections but enhanced options for appropriate therapy [35, 44, 45]. Verroken et al. demonstrated that BioFire PNplus Panel speeded up therapeutic changes in 46.9% of COVID-19 patients, five patients having antibiotics stopped and one third remained antibiotic free [35]. BioFire PNplus is rapid, simple to perform, and highly robust, with only 0.53% of the specimens in this study yielding invalid results. Detection of pathogens and antibiotic resistance markers can be used to inform immediate treatment decisions and improve patient outcomes.
Data availability
The datasets generated during and/or analyzed during the current study are not publicly available due to restrictions by individual contributors but are available as composite data from the corresponding author on reasonable request.
References
Campbell S, Forbes BA (2011) The clinical microbiology laboratory in the diagnosis of lower respiratory tract infections. J Clin Microbiol 49(9 Suppl):S30–S33. https://doi.org/10.1128/JCM.00789-11
Bartlett JG (2011) Diagnostic tests for agents of community-acquired pneumonia. Clin Infect Dis 52(Suppl 4):S296–S304
Feikin DR, Hammitt LL, Mukrdoch DR, O’Brien KL, Scott AG (2017) The enduring challenges of determining pneumonia etiology in children. Consideration for future research priorities. Clin Infect Dis 64(S3):5188–5196
Jain S, Self WH, Wunderink RG, Fakhran S, Balk R, Bramley AM, Reed C, Grijalva CG, Anderson EJ, Courtney DM, Chappell JD, Qi C, Hart EM, Carroll F, Trabue C, Donnelly HK, Williams DJ, Zhu Y, Arnold SR, Ampofo K, Waterer GW, Levine M, Lindstrom S, Winchell JM, Katz JM, Erdman D, Schneider E, Hicks LA, McCullers JA, Pavia AT, Edwards KM, L. Finelli L for the CDC EPIC Study Team (2015) Community-acquired pneumonia requiring hospitalization among U.S. adults. N Engl J Med 373:415–427
Harris AM, Bramley AM, Jain S, Arnold SR, Ampofo K, Self WH, Williams DJ, Anderson EJ, Grijalva CG, McCullers JA, Pavia AT, Wunderink RG, Edwards KM, Winchell JM, Lauri A, Hicks LA (2017) Influence of antibiotics on the detection of bacteria by culture-based and culture-independent diagnostic tests in patients hospitalized with community-acquired pneumonia. Open Forum Infect Dis. https://doi.org/10.1093/ofid/ofx014
Metlay JP, Waterer GW, Long AC, Anzueto A, Brozek J, Crothers K, Cooley LA, Dean NC, Fine MJ, Flanders SA, Griffin MR, Metersky ML, Musher DM, Restrepo MI, Cynthia Whitney CG, on behalf of the American Thoracic Society and Infectious Diseases Society of America (2019) Diagnosis and treatment of adults with community-acquired pneumonia. An official clinical practice guideline of the American Thoracic Society and Infectious Diseases Society of America. Am J Respir Crit Care Med 200(7):e45–e67. https://doi.org/10.1164/rccm.201908-1581ST
Woodhead M, Blasi F, Ewig S, Garau J, Huchon G, Ieven M, Ortqvist A, Schaberg T, Torres A, van der Heijden G, Read R, Verheij TJM, Joint Taskforce of the European Respiratory Society and European Society for Clinical Microbiology and Infectious Diseases (2011) Guidelines for the management of adult lower respiratory tract infections—full version. Clin Microbiol Infect 17(Suppl 6):E1–E59
Carugati M, Aliberti S, Reyes LF, Sadud RF, Irfan M, Prat C, Soni NJ, Faverio P, Gori A, Blasi F, Restrepo MI (2018) Microbiological testing of adults hospitalised with community-acquired pneumonia: an international study. ERJ Open Res 4:00096–02018. https://doi.org/10.1183/23120541.00096-2018
Charlton CL, Babady E, Ginocchio CC, Hatchette TF, Jerris RC, Li Y, Loeffelholz M, McCarter YS, Miller MB, Novak-Weekley S, Schuetz AN, Tang Y-W, Widen R (2018) Drews SJ (2019) Practical guidance for clinical microbiology laboratories: viruses causing acute respiratory tract infections. Clin Microbiol Rev 32(1):e00042–e00018. https://doi.org/10.1128/CMR.00042-18
Gadsby NJ, Russell CD, McHugh MP, Mark H, Morris AC, Laurenson IF, Hill AT, Templeton KE (2016) Comprehensive molecular testing for respiratory pathogens in community-acquired pneumonia. Clin Infect Dis 62:817–823
Zhan Y, Yang Z, Chen R, Wang Y, Guan W, Zhao S (2014) Respiratory virus is a real pathogen in immunocompetent community-acquired pneumonia: comparing to influenza like illness and volunteer controls. BMC Pulm Med 14:144. https://doi.org/10.1186/1471-2466-14-144
Centers for Disease Control (2019) Antibiotic resistance threats in the United States, 2019. Atlanta, GA: U.S. Department of Health and Human Services, CDC; 2019. www.cdc.gov/DrugResistance/Biggest-Threats.html. DOI: https://doi.org/10.15620/cdc:82532
Ray MJ, Tallman GB, Bearden DT, Elman MR, McGregor JC (2019) Antibiotic prescribing without documented indication in ambulatory care clinics: national cross sectional study. BMJ 367:l6461. https://doi.org/10.1136/bmj.l6461
World Health Organization https://www.who.int/health-topics/antimicrobial-resistance. Accessed 27 Sep 2020
Kalil AC, Metersky ML, Klompas M, Muscedere J, Sweeney DA, Palmer LB, Napolitano LM, O’Grady NP, Bartlett JG, Carratalà J, El Solh AA, Ewig S, Fey PD, File TM Jr, Restrepo MI, Roberts JA, Waterer GW, Cruse P, Knight SL, Brozek JL (2016) Management of adults with hospital-acquired and ventilator-associated pneumonia: 2016 clinical practice guidelines by the Infectious Diseases Society of America and the American Thoracic Society. Management of adults with HAP/VAP. Clin Infect Dis 63(5):e61–e111. https://doi.org/10.1093/cid/ciw353
Torres A, Niederman MS, Chastre J et al (2017) International ERS/ESICM/ESCMID/ALAT guidelines for the management of hospital-acquired pneumonia and ventilator-associated pneumonia: guidelines for the management of hospital-acquired pneumonia (HAP)/ventilator-associated pneumonia (VAP) of the European Respiratory Society (ERS), European Society of Intensive Care Medicine (ESICM), European Society of Clinical Microbiology and Infectious Diseases (ESCMID) and Asociación Latinoamericana del Tórax (ALAT). Eur Respir J 50:1700582
Smits WK, Lyras D, Borden Lacy DB, Wilcox MH, Kuijper EJ (2016) Clostridium difficile infection. Nat Rev Dis Primers 2:16020
Tamma PD, Avdic E, Li DX, Dzintars K, Cosgrove SE (2017) Association of adverse events with antibiotic use in hospitalized patients. JAMA Intern Med 177(9):1308–1315. https://doi.org/10.1001/jamainternmed.2017.1938
Webb BJ, Sorensen J, Jephson A, Mecham I, Dean NC (2019) Broad-spectrum antibiotic use and poor outcomes in community-onset pneumonia: a cohort study. Eur Respir J 54(1):1900057. https://doi.org/10.1183/13993003.00057-2019
Murphy C, Fowler R, Balada-Llasa JM et al (2020) Multicenter evaluation of the BioFire FilmArray Pneumonia Panel for the detection and quantification of agents of lower respiratory tract infection. J Clin Microbiol 58(7):e00128–e00120. https://doi.org/10.1128/JCM.00128-20
Mitton B, Rule R, Said M (2021) Laboratory evaluation of the BioFire FilmArray Pneumonia plus panel compared to conventional methods for the identification of bacteria in lower respiratory tract specimens: a prospective cross-sectional study from South Africa. Diagn Microbiol Infect Dis 99(2):115236. https://doi.org/10.1016/j.diagmicrobio.2020.115236
Yoo Y, Huh K, Shim HJ, Yun SA, Chung YN, Kang OK, Huh HJ, Lee NY (2020) Evaluation of the BioFire FilmArray Pneumonia Panel for rapid detection of respiratory bacterial pathogens and antibiotic resistance genes in sputum and endotracheal aspirate specimens. Int J Infect Dis 95:326–331. https://doi.org/10.1016/j.ijid.2020.03.024
Webber DM, Wallace MA, Carey-Ann D, Burnham C-AD, Neil W, Anderson NW (2020) Evaluation of the BioFire® FilmArray® Pneumonia Panel for detection of viral and bacterial pathogens in lower respiratory tract specimens in the setting of a tertiary care academic medical center. J Clin Microbiol posted Online 22 April. https://doi.org/10.1128/JCM.00343-20
Lee SH, Ruan S-Y, Pan S-C, Lee T-F, Chien J-Y, Hsueh P-R (2019) Performance of a multiplex PCR pneumonia panel for the identification of respiratory pathogens and the main determinants of resistance from the lower respiratory tract specimens of adult patients in intensive care units. J Microbiol Immunol Infect 52(6):920–928. https://doi.org/10.1016/j.jmii.2019.10.009
Edin A, Eilers H, Allard A (2020) Evaluation of the Biofire Filmarray Pneumonia panel plus for lower respiratory tract infections. Infect Dis Ther 52(7):479–488. https://doi.org/10.1080/23744235.2020.1755053
Gastli N, Loubinoux J, Daragon M et al (2020) Multicentric evaluation of BioFire FilmArray Pneumonia Panel for rapid bacteriological documentation of pneumonia. Clin Microbiol Infect. https://doi.org/10.1016/j.cmi.2020.11.014.Gastli
Iannello A, Dubost C, Weber C, Alberti-Segui C, Mousset C, Ginocchio C, Rogatcheva M, Barraud O, François B, Moucadel V, Yugueros-Marcos J (2018) 2002. Evaluation of the BioFire® Pneumonia Panel in ICU patients with suspected ventilator-associated pneumonia. Open Forum Infect Dis 5(Suppl 1):S583. https://doi.org/10.1093/ofid/ofy210.1658
FDA Drug Safety Communication: FDA advises restricting fluoroquinolone antibiotic use for certain uncomplicated infections 2016. http://www.fda.gov/Drugs/DrugSafety/ucm500143.htm. Accessed 12 May 2016
Waites KB, Xiao L, Liu Y, Balish MF, Atkinson TP (2017) Mycoplasma pneumoniae from the respiratory tract and beyond. Clin Microbiol Rev 30:747–809. https://doi.org/10.1128/CMR.00114-16
Hughes AEO, Webber DM, Wallace MA, Johnson C, Burnham C-A D, Neil W, Anderson NW (2020) Comparable detection of viral pathogens in lower respiratory tract specimens with the BioFire Respiratory Panel 2 and BioFire Pneumonia Panel. J Clin Microbiol 58(6):e00254–e00220. https://doi.org/10.1128/JCM.00254-20
Schuetz P, Wirz Y, Sager R et al (2017) Procalcitonin to initiate or discontinue antibiotics in acute respiratory tract infections. Cochrane Database Syst Rev 10:CD007498
Buchan BW, Windham S, Balada-Llasat J-M, et al. (2020) Practical comparison of the BioFire® FilmArray® Pneumonia Panel to routine diagnostic 1 methods and potential impact on antimicrobial stewardship in adult hospitalized patients with lower respiratory tract infections. Posted Online 29 April 2020 J. Clin. Microbiol. doi:https://doi.org/10.1128/JCM.00135-20.
Centers for Disease Control and Prevention https://www.cdc.gov/antibiotic-use/core-elements/hospital.html. 2019
Monard C, Pehlivan J, Auger G, Alviset S, Dinh AT, Duquaire P, Gastli N, d’Humières C, Maamar A, Boibieux A, Baldeyrou M, Loubinoux J, Dauwalder O, Cattoir V, Armand-Lefèvre L, Kernéis S, and the ADAPT study group (2020) Multicenter evaluation of a syndromic rapid multiplex PCR test for early adaptation of antimicrobial therapy in adult patients with pneumonia. Crit Care 24:434. https://doi.org/10.1186/s13054-020-03114-
Verroken A, Scohy A, Gérard L, Wittebole X, Collienne C, Laterre P-F (2020) Co-infections in COVID-19 critically ill and antibiotic management: a prospective cohort analysis. Crit Care 24:410. https://doi.org/10.1186/s13054-020-03135-7
Falsey AR, Becker KL, Swinburne AJ, Nylen ES, Formica MA, Hennessey PA, Criddle MM, Peterson DR, Baran A (2013) Walsh EE (2013) Bacterial complications of respiratory tract viral illness: a comprehensive evaluation. J Infect Dis 208:432–441
Jain S, Williams DJ, Arnold SR, for the Centers for Disease Control and Prevention (CDC) Etiology of Pneumonia in the Community (EPIC) Study Team et al (2015) Community-Acquired Pneumonia Requiring Hospitalization Among U.S. Children. N Engl J Med 372(9):835–845. https://doi.org/10.1056/NEJMoa1405870
Vos LM, Bruning AHL, Reitsma JB, Schuurman R, Riezebos-Brilman A, Hoepelman AIM, Oosterheert JJ (2019) Rapid molecular tests for influenza, respiratory syncytial virus, and other respiratory viruses: a systematic review of diagnostic accuracy and clinical impact studies. Clin Infect Dis 69:1243–1253
Molinos L, Zalacain R, Menéndez R, Reyes S, Capelastegui A, Cillóniz C, Rajas O, Borderías L, Martín-Villasclaras JJ, Bello S, Alfageme I (2015) Sensitivity, specificity, and positivity predictors of the pneumococcal urinary antigen test in community-acquired pneumonia. Ann Am Thorac Soc 12(10):1482–1489
Svarrer CW, Lück C, Elverdal PL, Uldum S (2012) Immunochromatic kits Xpect Legionella and BinaxNOW Legionella for detection of Legionella pneumophila urinary antigen have low sensitivities for the diagnosis of Legionnaires’ disease. J Med Microbiol 61:213–221
Pierre DM, Baron J, Yu VL, Stout JE (2017) Diagnostic testing for Legionnaires’ disease. Ann Clin Microbiol Antimicrob 16:59. https://doi.org/10.1186/s12941-017-0229-6
Jain S, Pavia AT (2016) The modern quest for the “holy grail” of pneumonia etiology. Clin Infect Dis 62:826–828
Alby K, Mitchell SL (2018) Lower respiratory multiplex panels for the detection of bacterial and viral infections. Clin Microbiol Newsl 40(16):131–136. https://doi.org/10.1016/j.clinmicnews.2018.07.005
Dhesi Z, Enne VI, Brealey D, Livermore DM, High J, Russell C, Colles A, Kandil H, Mack D, Martin D, Page V, Parker R, Roulston K, Singh S, Wey E, Swart AM, Stirling S, Barber JA, O’Grady J, Gant V. Organisms causing secondary pneumonias in COVID-19 patients at 5 UK ICUs as detected with the FilmArray test. medRxiv preprint doi: https://doi.org/10.1101/2020.06.22.20131573
Kreitmann L, Monard C, Dauwalder O, Simon M (2020) Argaud L (2020) Early bacterial co-infection in ARDS related to COVID-19. Intensive Care Med 13:1–3. https://doi.org/10.1007/s00134-020-06165-5
Acknowledgements
The members of the EME Evaluation Program Collaborative for their support and contribution to this study: Hospital Universitario Central de Asturias, Asturias, Spain, Forcelledo L, Fernández J; ADMED Laboratoires - Analyses et Diagnostics Médicaux, La Chaux-de-Fond, Switzerland, Lienhard R; Analyse Biolab GmbH, Linz, Austria, Kerschner H; AOUC Azienda Ospedaliero Universitaria Careggi, Florence, Italy, Rossolini GM; AP-HP Nord Université de Paris, Hôpital Bichat, Paris, France, Armand-Lefèvre L, d’Humières C; APHP-Hôpital Lariboisière-Fernand Widal, Université de Paris, Paris, France, Cambau E, Benmansour H; Azienda Ospedaliero Universitaria San Giovanni Battista Molinete, Torino, Italy, Cavallo R; Bioanalytica, Lucerne, Switzerland, Altwegg M, Berlinger L; Centre Hospitalier Universitaire de Clermont-Ferrand, Clermont, France, Bonnet R, Saint-Sardos P; Centre Hospitalier Universitaire de Liège - Sart Tilman, Liège, Belgium, Meex C; Centre Hospitalier Universitaire de Nimes, Nimes, France, Lavigne JP; Centre Hospitalier Universitaire de Poitiers, Poitiers, France, Leveque N, Broutin L; Centre Hospitalier Universitaire de Rennes, Rennes, France, Cattoir V, Auger G; Centro Hospitalar e Universitário de Coimbra, Coimbra, Portugal, Pereira H; Clinical Microbiology Lab, Meir Medical Center, Kfar Saba, Israel, and Department of Clinical Microbiology and Immunology, Sackler Faculty of Medicine, Tel Aviv University, Israel, Paitan Y; Cliniques Universitaires Saint-Luc - UCL, Louvain, Belgium, Verroken A; Département de Biologie des Agents Infectieux , UF de bactériologie, CHU Angers, Angers, France, Pailhoriès H, Lemarié C; Ente Ospedaliero Cantonale - Servizio di Microbiologia EOLAB, Bellinzona, Switzerland, Martinetti-Lucchini G, Frigerio Malossa S; Fondazione Policlinico Universitario Agostino Gemelli IRCCS, Roma, Italy, Sanguinetti M, Spanu T; HCL - Hospices Civils de Lyon, Lyon, France, Vandenesch F; Hôpital Cochin Paris, Paris, France, Poyart C, Loubinoux J, Mira JP; Hôpital Robert-Debré AP-HP, Paris, France, Bonacorsi S, Cointe A; Hospital General Universitario Gregorio Marañón, Madrid, Spain, Muñoz P, Kestler M; Hospital Sant Joan de Déu, Barcelona, Spain, Esteva C; Hospital Universitari de Girona, Doctor Josep Trueta, Girona, Spain, Queralt X; Hospital Universitario La Paz, Madrid, Spain, García-Rodríguez J; Hospital Universitario y Politécnico La Fe, Valencia, Spain, Gómez MD, López-Hontangas JL; Laboratory of Microbiology and Virology, Ospedale Amedeo di Savoia, Torino, Italy, Ghisetti V, Burdino E; Max von Pettenkofer Institute of Hygiene and Medical Microbiology, Faculty of Medicine, LMU Munich, Germany, Schubert S; Medical Microbiology, Department of Medicine, University of Perugia, Perugia, Italy and Microbiology, Perugia General Hospital, Perugia, Italy, Mencacci A, Allegrucci F; Noordwest Ziekenhuisgroep, Alkmaar, The Netherlands, Rozemeijer W, Paternotte N; Norrlands Universitetssjukhus, Umeå, Sweden, Allard A; Operative Unit of Clinical Microbiology, Policlinico Sant'Orsola Malpighi, Bologna, Italy, Re MC, Ambretti S; OUH Odense Universitethospital, Odense, Denmark, Skov M, Agergaard, CN; Royal Bolton Hospital, Bolton, England, Subudhi P; Universitäts Klinikum Frankfurt, Frankfurt, Germany, Wichelhaus TA; University Hospital Basel, Basel, Switzerland, Egli A, Hinic V; University Hospitals of North Midlands, Stoke-on-Trent, England, Alcock A, Banavathi K; UOC Microbiologia e Virologia, P.O. Cotugno, Azienda Ospedaliera Specialistica dei Colli, Napoli, Italy, Tiberio C, Ruocco G, Atripaldi L
Code availability
Not applicable.
Funding
BioFire Pneumonia plus Panels were provided free of charge by bioMérieux, France. No other compensation was provided to the test sites.
Author information
Authors and Affiliations
Consortia
Contributions
CCG, CG-M, and BM contributed to the study conception and design. Material preparation and data collection were performed by CCG, CG-M, BM, and the EME Evaluation Program Collaborative. All members of the EME Evaluation Program Collaborative contributed equally to the study. Data analysis was performed by CCG, CG-M, and CR. The manuscript was written by CCG and all authors commented on previous versions of the manuscript. All authors read and approved the final manuscript. All members of the EME Evaluation Program Collaborative reviewed and approved the manuscript.
Corresponding author
Ethics declarations
Competing interests
Christine C. Ginocchio, Carolina Garcia-Mondragón, and Barbara Mauerhofer are employees of bioMérieux. Cory Rindlisbacher is an employee of BioFire Diagnostics. EME Evaluation Program Collaborative sites did not receive any other resources to perform this study other than the test panels. The study design, authors, data analysis, interpretation of data, and writing of report were performed by bioMérieux employees as per author contribution listed above. Data was reviewed and approved by the EME Evaluation Program Collaborative sites.
Ethics approval
Ethical approval was in accordance with individual institution requirements.
Consent to participate
Not applicable.
Consent for publication
All authors and contributors have read and approved the manuscript.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Ginocchio, C.C., Garcia-Mondragon, C., Mauerhofer, B. et al. Multinational evaluation of the BioFire® FilmArray® Pneumonia plus Panel as compared to standard of care testing. Eur J Clin Microbiol Infect Dis 40, 1609–1622 (2021). https://doi.org/10.1007/s10096-021-04195-5
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10096-021-04195-5