Abstract
A method for the determination of uranium and 210Po in high salinity water samples has been elaborated. Both radionuclides are preconcentrated from 0.5 dm3 saline media by co-precipitation with hydrated manganese dioxide, followed by dissolution of the precipitate in 200 mL of 1 M HCl. Uranium isotopes 235U and 238U can be directly determined by ICP MS method with a detection limit of 0.01 ppb for 238U. Prior to a selective determination of 210Po, the majority of other naturally occurring α-emitting radionuclides (uranium, thorium and protactinium) can be stripped from this solution by their extraction with a 50% solution of HDEHP in toluene. Finally, 210Po is simply separated by direct transfer to an extractive scintillator containing 5% of trioctylphosphine oxide in Ultima Gold F cocktail and determined by an α/β separation liquid scintillation technique with detection limit below 0.1 mBq/dm3.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Introduction
According to the WHO protocol for radioactivity monitoring in drinking waters, in the first step the total α and β activities are determined [1]. Gross α, β and 226Ra activity can be simply appraised by direct adding of the water samples to scintillation cocktails followed by a simultaneous counting of α and β particles in the liquid scintillation devices with an α/β separation option [2]. However, if the total α activity exceeds 0.5 Bq/dm3, the activities of the particular α emitters should be measured. In practice, for natural water it concerns the radionuclides of radium, uranium, 210Po and 228Th. Since thorium compounds are generally poorly soluble in water, the activities of 232Th and 230Th radioisotopes are extremely low in the order of μBq/dm3, and only the activity of 228Th is observable, due to the contribution of 228Ra decay to its total budget in natural water [3].
A number of analytical methods are available for the detection of radium isotopes [4]. However, liquid scintillation counting is often used for fast and reliable determination of radium and 222Rn determination in different kinds of natural water [5–8].
Military use of depleted uranium in wars in the Middle East, Afghanistan and the Balkans, and its environmental and human health impact [9, 10] are reasons for an increasing interest in the fast and routine methods for uranium determination in environmental samples. On the other hand, a terrorist poisoning by using 210Po drew attention to its high radiotoxicity, even in extremely low concentrations in water solutions [11].
Average 238U and 234U activities in surface and ground water are in the range of 1–20 mBq/dm3, which corresponds to a concentration of elemental uranium from 0.1 to 1.5 μg/dm3 [12]. These two uranium radionuclides in natural water are not often in a radioactive equilibrium. A major source of uranium is seawater, where its average concentration is equal to 3.4 μg/dm3 [13]. However, higher concentrations of uranium are also detected in the drinking water in different countries, for example up to 150 Bq/dm3 in some private wells in Finland [14].
210Po radionuclide in a water environment occurs usually in much lower activities <1 Bq/dm3. Typical activity concentrations of 210Po in surface coastal seawater are in the range of 0.06–0.84 mBq/dm3 [15, 16].
According to WHO recommendation, taking into account uranium chemical toxicity, its level in drinking water should not exceed 2 μg/dm3 [1]. A more tolerable concentration of 30 μg/dm3 for uranium is proposed by the US EPA [17]. Both these organizations recommend close values of permissible activity concentrations of 210Po in drinking water: 0.1 and 0.148 Bq/dm3, respectively, which are among the lowest limits for any natural or anthropogenic radionuclides in drinking water.
Uranium can be measured in diverse matrices by radiometric, laser fluorimetry and mass spectrometric methods. Advances in the methods of uranium determination have been recently overviewed [18, 19]. There is no doubt that in recent years inductively coupled mass spectrometry (ICP MS) has been increasingly used for the accurate determination of trace levels of long lived radionuclides, including uranium and even radium isotopes [20]. Thanks to the very low detection limit of the ICP MS method (<0.01 ppb), the majority of water samples can be instrumentally measured directly without any pretreatment. However, an important limitation of this method is the fact that the concentration of the total dissolved solids (TDS) present in the analyzed solution cannot exceed 0.1% (1 g/dm3) [21]. Unfortunately, in some kinds of environmental samples such as seawater, geothermal or natural mineral drinking water, the TDS values are often higher than 1 g/dm3.
210Po is an α-emitter with weak gamma-rays abundance (803.1 keV – 0.0011%) and can be measured by α-spectrometry with solid detectors or by liquid scintillation counting (LSC) with α/β separation using special extractive scintillation cocktails [22]. Usually a simple separation of 210Po from other interfering radionuclides present in mineral and drinking water is achieved by its spontaneous deposition on silver discs. Such disks can be measured by an α-spectrometry systems with 30% efficiency and a very low background < 10−4 cps [23–26]. The alternative method, LSC allows 100% detection efficiency to be achieved, but the background is higher (>10−3 cps) and the resulting limits of detection are also slightly higher [27].
The aim of this study was to develop a fast method for routine determination of uranium and 210Po radionuclides in high salinity water samples (mineral, geothermal and seawater) after its preconcentration on hydrated manganese dioxide. It is well known that in such media uranium occurs mainly as a UO2(CO3) −22 anions [28], which are quantitatively scavenged by precipitated manganese dioxide [29]. Sorption of 210Po from seawater by this procedure is also well documented [30]. After dissolving the obtained precipitate in a sufficient volume of 1 M HCl to get the final TDS < 1 g/dm3, one part of solution can be used for direct determination of uranium by ICP MS, whereas from the second portion of solution 210Po can be transferred to a silver disc deposition device or to an extractive scintillation solution after removal of the other interfering α-emitting radionuclides.
Experimental
Materials and method
A standard solution of 210Pb in equilibrium with 210Bi and 210Po was purchased from Amersham (RBZ.44). A secondary uranium standard (238U), 200 ppm, was prepared by dissolution of 0.422 g of uranyl nitrate (UO2)(NO3)2·6H2O (p.a. Chemapol) in 1 dm3 of high purity deionizing water. All other chemicals used in this work were analytical grade purity compounds.
The standard reference material, Irish Sea Water: IAEA-381—“Radionuclides in seawater”, was supplied by the International Atomic Energy Agency (IAEA). For determination of the 210Po and 238U radionuclide recoveries from high salinity water, the artificial seawater solution was prepared by dissolving 27.2 g NaCl, 3 g MgSO4, 2.65 g MgCl2, 1.1 g CaCl2 and 0.755 g KCl in 5 dm3 of high purity deionizing water. Bottled mineral water with a high salt content were purchased from local stores.
Co-precipitation of 210Po and uranium from saline solutions
Simultaneous co-precipitation of 210Po and uranium from 0.5 dm3 solutions was carried out by a slightly modified method, described elsewhere [26, 31, 32]. The artificial seawater solution was acidified to pH = 1 by concentrated nitric acid, and 2.5 mL of 0.1 M diethylenetriaminepentaacetic acid (DTPA) was added to form complexes of Ca+2 and Mg+2 ions followed by a further adding of 7 mL of 0.2 M solution of potassium permanganate with intense 1 h mixing. After this time the solution was alkalized by concentrated ammonia, and 2.5 mL of 0.3 M manganese chloride was added. The mixture was heated for 1 h and left for overnight precipitation. The precipitate was filtered, dried and weighed. The addition of DTPA almost completely prevents the precipitation of calcium and magnesium hydroxides from saline solutions. Therefore, the total mass of dried manganese hydroxides precipitated from different water solutions, even those with a high salt content, varied in the narrow range from 0.1 to 0.2 g.
The 0.5 L of artificial seawater was spiked with uranyl nitrate and 210Po markers for the determination of their recoveries after the preconcentration step. For these purposes a hydrated manganese oxide precipitate with preconcentrated radionuclides was dissolved in 5 mL of 1 M HCl. One mL of this solution was taken for polonium recovery determination and the remaining 4 mL of solution was diluted to 200 mL for uranium recovery determination by ICP-MS.
Instruments
Uranium determination
Elemental uranium concentration was routinely measured by ICP MS method using a Thermo Electron Corporation X-Series device. For analytical purposes the device was calibrated with a set of secondary standard uranium solutions covering the range up to 100 ppb, prepared from the Merck standard uranium solution.
Radioactivity measurements
The separation efficiency of U, Th and Pa radionuclides from 210Po by 50% HDEHP (bis(2-ethyhexyl)hydrogen phosphate) in toluene extraction was checked by measurement of 234Th, 234Pa, 235U and 238U radionuclides in organic phase by HPGe γ spectrometry system described elsewhere [32].
The activity of 210Po transferred to an extractive scintillator was measured using a BetaScout counter (Perkin Elmer). The optimal counting parameters for α-particle measurement were settled after a preliminary 1-h count. The background in this region was equal to 0.2 cpm and the standard counting time was 80,000 s.
Quality assurance of the elaborated method
The accuracy of the developed method for uranium was evaluated for the standard reference materials IAEA-381 Irish Sea Water. The results for uranium concentration are presented in Table 1. The results of 210Po determination in the standard reference materials are also included although there are no reference values for this radionuclide.
Therefore, the accuracy of the method for 210Po determination was checked by measuring 210Po in the water samples supplied by IAEA in the frames of their intercomparison studies. Five different water samples with 210Po were provided by IAEA of Seibersdorf (Vienna). In the samples 01–04 210Po activities were in the range of 1 to 10 Bq/dm3, whereas for sample #05 the 210Po concentration activity was <0.1 Bq/dm3. The results of these measurements are presented in Table 2.
Results and discussion
Choice of the optimal condition for preconcentration of uranium prior to its determination by ICP-MS
In order to avoid an additional precipitation of calcium and manganese hydroxides, during the co-precipitation procedure with manganese hydrated oxide, two well-known complexing agents, EDTA and DTPA, were checked. The results of the uranium recoveries from artificial sea water with and without use of these complexing agents are presented in Table 3.
Using 5 mL of 0.1 M EDTA is connected with a decrease of uranium recovery in comparison to untreated sea water, while the addition of 2.5 mL of 0.1 M DTPA increases this recovery to 84.7%. Simultaneously, a reduction of the total amount of precipitate to 0.2 g was observed. Therefore, in the standard procedure for high saline water samples, the DTPA solution was used as a masking agent for Ca+2 and Mg+2 ions.
In order to achieve the final solution with TDS below 1 d/dm3, the hydrated manganese dioxide (0.2 g) with adsorbed radionuclides should be dissolved in 200 mL of 1 M HCl with 3% H2O2. A portion of this solution was taken for direct uranium determination by ICP-MS, while the rest was used for 210Po determination.
Removing of U, Th, Pa and 210Po from 1 M HCl solution by HDEHP in toluene
The standard solution containing 303 μg/mL UO2(NO3)2 with 234Th and 234Pa radionuclides in a radioactive equilibrium was prepared for U, Th and Pa separation studies from 210Po. The thorium, uranium and protactinium nuclides were extracted from 20 mL and 100 mL samples of this solution by twofold extraction of 10 mL 50% solution HDEHP in toluene. The results of these experiments are shown in the Table 4.
As is evident from Table 4, all radionuclides of Th, U and Pa are completely removed from the solution containing 210Po, however, the 210Po remains almost quantitatively in the water phase.
Choice of optimal beta scout parameters for 210Po determination by LSC with an extractive scintillator
In order to ensure optimal counting conditions for complete separation of α from β pulses in the BetaScout device, the total radioactivity in scintillator solution should be checked first. On the basis of the 3D spectrum, the proper PLI (Pulse Length Index) parameter of 440 for 210Po was settled on (see Fig. 1).
After twofold extraction of 210Po from dissolved manganese precipitate, the total volume of extractive scintillator rose to 20 mL. Therefore, the influence of the scintillator volume on the detection efficiency of 210Po in the Beta Scout device was also checked. The results of these measurement showed that the increase of the volume of scintillation solution does not influence detection efficiency, and the two-step extraction by 10 mL of an extractive scintillator ensures the complete transfer of 210Po.
Extraction of 210Po by extractive scintillation (5% TOPO in Ultima Gold F)
Selective determination of 210Po by direct extraction into liquid scintillation solution has been reported recently [22, 27, 34]. Extraction into 5% trioctylphosphine oxide in toluene from an HCl solution of acid strength above 0.1 M, allowing more than 98% extraction of 210Po with separation from U [22], is to be noted. Therefore in our experiments a solution 5% (w/v) TOPO in immiscible with water Ultima Gold F scintillation cocktail was checked for extraction of 210Po from 1 M HCl solution.
In order to avoid additional loses of the extractive scintillator during its transfer from the extraction bottle to the scintillation vials, the extraction procedure was repeated with a second 10 mL portion of extractive scintillator. The results of the total recovery of 210Po after the twofold extraction from 100 mL 1 M HCl solution are present in Table 5.
Diagram scheme of the elaborated procedure for radionuclide determination in the high salinity water samples is shown on Fig 2.
210Po and uranium in mineral water
The method has been used for the determination of the 238U and 210Po activities in some bottled mineral water samples distributed in Poland. The concentration of uranium in the majority of Polish mineral water samples was reported previously [7, 26, 34], while 210Po was determined only in a few types of bottled mineral water [26].
In our experiments, in three kinds of mineral water TDS exceeded 1 g/dm3, and in the remaining three samples mineralization was lower. Our results (see Table 6) are generally in sufficient accordance with the previous reported data for these mineral water samples.
Conclusions
The elaborated method for 238U and 210Po determination consists on co-precipitation with hydrated manganese dioxide. Uranium radionuclides can be measured by ICP-MS technique after reducing the total TDS below 1 g/dm3. Uranium recovery, even from high salinity water samples, averaged ~86%. 210Po was extracted by 5% (w/v) solution TOPO in scintillation cocktail Ultima Gold F, after removing all interfering α-emitting radionuclides by twofold extraction with 50% solution of HDEHP in toluene. The obtained detection limit for a routine liquid scintillation counting of 210Po is equal to 1 mBq/dm3, and is sufficient for determining the natural levels of this radionuclide in the majority of natural drinking water samples.
References
WHO (2006) Guidelines for drinking-water quality, vol 1. World Health Organization, Genewa
Bakir YY, Bem H (1994) Application of pulse decay discrimination liquid scintillation counting for routine monitoring of radioactivity in drinking water. Liquid Scintillation Spectrometry Glasgow, Scotland
Jia G, Torri G, Ocone R, Di Lullo A, De Angelis A, Boschetto R (2008) Determination of thorium isotopes in mineral and environmental water and soil samples by α-spectrometry and the fate of thorium in water. Appl Radiat Isotop 66:1478–1487
Köhler M, Preuße W, Gleisberg B, Schäfer I, Heinrich T, Knobus B (2002) Comparison of methods for the analysis of 226Ra in water samples. Appl Radiat Isotop 56:387–392
Salonen L (1993) A rapid method for monitoring of uranium and radium in drinking water. Sci Total Environ 130(131):23–35
Bem H, Bem EM, Majchrzak I (1998) Comparison of two methods for 226Ra determination in mineral water. Nukleonika 43(4):459–468
Nguyen Dinh Chau, Michalec B (2009) Natural radioactivity in bottled natural spring, mineral and therapeutic waters in Poland. J Radioanal Nucl Chem 279(1):121–129
Blanco Rodriguez MP, Vera Tome F, Lozano JC, Gomez Escobar V (2000) Sequential method for the determination of uranium, thorium and 226Ra by liquid scintillation alpha spectrometry. Appl Radiat Isotop 52:705–710
Bem H, Bou-Rabee F (2004) Environmental and health consequences of depleted uranium use in the 1991 Gulf War. Environ Internat 30:123–134
United Nations Environment Programme (May 2003) Depleted uranium in Bosnia and Herzegovina. Post-conflict environmental assessment. Revised edition, pp 88–89
Harrison J, Leggett R, Lloyd D, Phipps A, Scott B (2007) Polonium-210 as a poison. J Radiol Protect 27:17–40
UNESCAR (2000) United Nations Committee on the Effects of Atomic Radiation. Sources, effects, and risks of ionizing radiation. United Nations, New York
Hossain MMM, Ohde S (2002) Uranium in pore water from coastal sediments determined by chemical activation analysis. J Radioanal Nucl Chem 251(2):333–336
Vesterbacka P (2005) 238U–series radionuclides in Finnish groundwater-based drinking water and effective doses. STUK, Helsinki
Stępnowski S, Skwarzec B (2000) A comparison of 210Po accumulation in molluscs from the southern Baltic, the coast of Spitsbergen and Sasek Wielki Lake in Poland. J Environ Radioact 49:201–208
Theng TL, Mohamed CAR (2005) Activities of 210Po and 210Pb in the water column at Kuala Selangor, Malaysia. J Environ Radioact 80:273–286
US EPA (2002) Radionuclides in drinking water: a small entity compliance guide, 2002, 4
Rathore DPS (2008) Advanced in technologies for the measurement of uranium in diverse matrices. Talanta 77:9–20
Hou X, Roos P (2008) Critical comparison of radiometric and mass spectrometric method for the determination of radionuclides in environmental, biological and nuclear waste samples. Anal Chim Acta 608:105–139
Becker JS (2005) Inductively coupled plasma mass spectrometry (ICP-MS) and laser ablation ICP-MS for isotope analysis of long-lived radionuclides. Int J Mass Spectrom 242:183–195
Pawlak Z, Rabiega G (2002) Comparison of inductively coupled plasma-mass spectrometry and radiochemical techniques for total uranium in environmental water samples. Environ Sci Technol 36:5395–5398
Matthews KM, Chang-Kyu Kim, Martin P (2007) Determination of 210Po in environmental materials: a review of analytical methodology. Appl Radiat Isotop 65:207–279
Desideri D, Meli MA, Feduzi L, Roselli C, Rongoni A, Saetta D (2007) 238U, 234U, 226Ra, 210Po concentration of bottled mineral waters in Italy and their dose contribution. J Environ Radioact 94:86–97
Jia G, Belli M, Blasi M, Marchetti A, Rosamilia S, Sansone U (2000) 210Pb and 210Po determination in environmental samples. Appl Radiat Isotop 53:115–120
Wallner G, Wagner R, Katzlberger Ch (2008) Natural radionuclides in Austrian mineral water and their sequential measurement by fast methods. J Environ Radioact 99:1090–1094
Skwarzec B, Strumińska DI, Boryło A (2003) Radionuclides of 210Po, 234U and 238U in drinking bottled mineral water in Poland. J Radioanal Nucl Chem 256(2):361–364
Veronneau C, Aupiais J, Dacheux N (2000) Selective determination of polonium by photon rejecting alpha liquid scintillation (PERALS System). Anal Chim Acta 415:229–238
Langmuir D (1978) Uranium solution–mineral equilibria at low temperature with applications to sedimentary ore deposits. Geochim Cosmochim Acta 42:547–569
Varga Z (2007) Preparation and characterization of manganese dioxide impregnated resin for radionuclide pre-concentration. Appl Radiat Isotop 65:1095–1100
Skwarzec B, Strumińska DI, Boryło A (2006) Radionuclides of iron (55Fe), nikel (63Ni), polonium(210Po), uranium (234U, 235U, 238U) and plutonium (238Pu, 239+240Pu, 241Pu) in Poland and Baltic Sea environment. Nukleonika 51(2):45–51
Eikenberg J, Tricca A, Vezzu G, Bajo S, Ruethi M, Surbeck H (2001) Determination of 228Ra, 226Ra and 224Ra in natural water via adsorption on MnO2-coated discs. J Environ Radioact 54:109–131
Bem H, Olszewski M, Kaczmarek A (2004) Concentration of selected natural radionuclides in the thermal groundwater of Uniejow. Nukleonika 49(1):1–5
Jia G, Torri G, Petruzzi M (2004) Distribution coefficient of polonium between 5% TOPO in toluene and aqueous hydrochloric and nitric acid. Appl Radiat Isotop 61:279–282
Kozłowska B, Walencik A, Dorda J, Przylibski TA (2007) Uranium, radium and 40K isotopes in bottled mineral water from Outer Carpathians, Poland. Radiat Meas 42:1380–1386
Open Access
This article is distributed under the terms of the Creative Commons Attribution Noncommercial License which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
Open Access This is an open access article distributed under the terms of the Creative Commons Attribution Noncommercial License (https://creativecommons.org/licenses/by-nc/2.0), which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.
About this article
Cite this article
Grabowski, P., Bem, H. Determination of 210Po and uranium in high salinity water samples. J Radioanal Nucl Chem 286, 455–460 (2010). https://doi.org/10.1007/s10967-010-0720-4
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10967-010-0720-4