Abstract
Purpose of Review
Cardiac resynchronization therapy (CRT) represents a well-established and effective non-pharmaceutical heart failure (HF) treatment in selected patients. Still, a significant number of patients remain CRT non-responders. An optimal placement of the left ventricular (LV) lead appears crucial for the intended hemodynamic and hence clinical improvement. A well-localized target area and tools that help to achieve successful lead implantation seem to be of utmost importance to reach an optimal CRT effect.
Recent Findings
Recent studies suggest previous multimodal imaging (CT/cMRI/ECG torso) to guide intraprocedural LV lead placement. Relevant benefit compared to empirical lead optimization is still a matter of debate. Technical improvements in leads and algorithms (e.g., multipoint pacing (MPP), adaptive algorithms) promise higher procedural success. Recently emerging alternatives for ventricular synchronization such as conduction system pacing (CSP), LV endocardial pacing, or leadless pacing challenge classical biventricular pacing.
Summary
This article reviews current strategies for a successful planning, implementation, and validation of the optimal CRT implantation. Pre-implant imaging modalities offer promising assistance for complex cases; empirical lead positioning and intraoperative testing remain the cornerstone in most cases and ensure a successful CRT effect.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Introduction
Cardiac resynchronization therapy (CRT) might be judged as one of the most impressive success stories for heart failure treatment since its initial reports nearly 30 years ago.
Driven by the concept and patent by Dr. Mower in 1990 and the first encouraging case report from Serge Cazeau in 1994 using a four-chamber stimulation with two Y-adapters and an epicardial left ventricular lead, the understanding of conductance disorder, especially left bundle branch block (LBBB), grew immediately [1, 2]. Furthermore, technical tools were developed to access the coronary sinus, to reach challenging targets in the tributaries, and to ensure stable lead position and allow device programming, which is adapted to the individual electro-mechanical disorder. Based on our current knowledge, this first patient with a QRS of 200 ms, LBBB, non-ischemic cardiomyopathy, and severe heart failure is reflecting a typical “responder.”
With the extension of CRT therapy and growth of scientific knowledge, the question came up what the definition of “response” is and when and how it is determined. For years, the statement that nearly 30% of CRT patients do not respond to resynchronization therapy has become general knowledge and is discussed and partially accepted without questioning the reasons. Generally, the main goals of HF treatment are relief of symptoms, restoration of quality of life (QOL), slowing of disease progression, lowering the rates of emergency room visits and hospitalizations, and prolongation of life. Alleviation of symptoms and lower morbidity are the main objectives for patients in New York Heart Association (NYHA) HF functional Classes III and IV and preventing disease and HF progression is priority for patients in NYHA Classes I and II [3].
The further the endpoint of judgment veers away from the time of CRT implantation, the more it will be overlapped by other events, disease progression, or other concomitant factors that overlap the CRT effect. The time after implantation, which best reflects the CRT effect, has never been defined and will clearly influence the interpretation of trial data.
Furthermore, patients’ expectations in CRT therapy regarding response will differ from scientific judgment. Whereas the patient expects an improvement in quality of life, in physical activity, and decrease in hospitalizations, he will not necessarily be convinced by the improvement of echocardiographic remodeling measures or not even by prolongation of his life in poor quality. Without regard to the discussion of response or non-response, the basis for a potential improvement is always the implantation of the device and the definition of the “optimal left ventricular pacing site.”
Finding the Optimal LV Lead Position
Up to now, achieving synchrony between both ventricles as well as normalizing right and left ventricular filling is mainly facilitated by the implantation of a right atrial, right ventricular, and left ventricular transvenous epicardial lead using the coronary system.
An unchangeable limitation in selecting the optimal implantation site is the given anatomy of the coronary venous system, which varies substantially. Whereas in some patients different site branches might be technically accessible, the implanter might be glad in others if only one tributary will allow a stable lead position close to the intended target area however this is defined. Finally, the LV lead position is often a compromise between target area, lead stability, threshold, intrinsic signal, and inadvertent diaphragm stimulation. Especially in patients who already had cardiac surgery, aortal coronary venous grafts, or coronary venous lead extractions, the choice is limited. There are multiple tools and tricks to overcome these problems [4].
The process of defining the optimal site for left ventricular lead placement might be divided into different steps:
-
1.
Understanding the individual problem of conduction delay (LBBB) including underlying heart disease, medical history, and previous interventions.
-
2.
Characterizing the quality of tissue, including the detection and visualization of areas of scar, late(st) electrical and mechanical activation—PRIOR Implantation NON-INVASIVELY.
-
3.
Showing the implanter’s reality by coronary sinus venogram and estimating the accessibility of target tributaries based on experience and available tools, pure anatomical approach—DURING implant.
-
4.
Detecting and correcting electrical and mechanical left ventricular delay including acute hemodynamic optimization—DURING implantation INVASIVELY.
Ad 1: Understanding the individual problem of conduction delay (LBBB) including underlying heart disease, medical history, and previous interventions.
When considering a patient for CRT, a global understanding of the underlying mechanisms of heart failure is needed and all reversible and treatable targets as ischemia, valvular disease, metabolic disorder, arrhythmias including atrial fibrillation etc. should be eliminated first. Guideline-directed heart failure medication should be optimized for at least 3–6 months, including modern strategies as ARNI and SGLT2 inhibitors before re-evaluating the need for CRT using echocardiography and the clinical estimation [5]. As far as the patient suffers from a non-ischemic cardiomyopathy, shows a QRS complex > 150 ms with typical LBBB morphology, and has not undergone any previous cardiac surgery, the coronary venous system usually offers at least a few options of tributaries. The LV lead implant site will be less crucial in this population and a good response can be predicted.
In contrast, patients with an ischemic origin of their heart failure, multiple coronary interventions, previous coronary artery bypass grafting (CABG), QRS complex between 130 and 150 ms, or non-LBBB configuration likely have a more challenging coronary sinus anatomy with less options to optimize the LV lead position. A response in these patients is potentially difficult to predict and more preceding imaging methods such as MRI, strain analysis, and electroanatomical mapping should be integrated to identify the best pacing site. Furthermore, additional information regarding the acute response (see 4.) will be helpful to decide whether an alternative approach should be preferred.
The same strategy is recommended for CRT reimplantation or chronic non-responder.
Ad 2:Characterizing the quality of tissue, including the detection and visualization of areas of scar, late(st) electrical and mechanical activation—PRIOR Implantation NON-INVASIVELY.
Preceding non-invasive imaging is helpful to characterize the tissue quality of the myocardium, detect scar and viable myocardium, and detect regions of delayed electrical and mechanical activation. The detection of intramural scar and its distribution within and over the left ventricular wall can add additional information about the risk of sudden cardiac death in dilative cardiomyopathy and support the decision to implant a CRT-D instead of CRT-P.
Cardiac MRI (CMR) and Dual-Source CT (DSCT)
Late gadolinium enhancement (LGE) CMR is the gold standard for delineating myocardial scar with high resolution, as the superior spatial resolution of LGE-CMR permits differentiation between epicardial, transmural, and sub-endocardial infarction.
The extent, location, and transmurability of scar is of major importance for the guidance of the implantation site and the outcome independent of its ischemic or non-ischemic origin [6,7,8]. The size of scar is inversely correlated with the acute response, reverse remodeling, functional improvement, and survival. A lead implantation in the empirically targeted postero-lateral region will lead to a lower response rate and a potential increase of arrhythmias if it expands the scar [6]. Both CMR-guided last mechanical activation as well as electrical approaches have been focusing on detection of the latest left ventricular activation and considering this an optimal pacing site [9, 10]. If this is found within a larger transmural scar region, it seems questionable to be the preferable site. Sometimes electrical activation propagating from off the most delayed spot might result in an earlier mechanical activation restoring synchrony when thus avoiding a scar region [11]. This correlation between scar morphology and subendocardial, subepicardial, or transmural dispersion and mechanical response will at least help to avoid certain regions and target viable border zones. Approximately one third of CRT candidates are not eligible for an MRI due to metallic implants or other rhythmological devices, and other imaging technologies have to be discussed. Modern dual-source CT (DSCT) scans can detect scar, but its information regarding tissue characterization and viability is less compared to MRI. Besides, DSCT is not delivering superior information about delayed mechanical activation and carries the risk of contrast and radiation exposure.
Summarizing, CT is infrequently used outside from scientific questions to achieve an anatomical overview of the vascular access and the coronary venous anatomy [12, 13], especially in patients with congenital heart disease or previous cardiac surgery. Its role in multimodality imaging in planning and performing CRT implantations will be discussed later.
Trans-Thoracic Echocardiography
Transthoracic echocardiography plays a major role in imaging of heart failure patients, not only scheduled for CRT. Due to its cost-effectiveness, its general availability, and its fast information, it should be used in every CRT candidate. Scar is detected by wall thinning and abnormal motion patterns, and viable and ischemic myocardium can be identified as well.
New modalities such as 3D contrast echo and pulse cancelation have been investigated to improve scar detection [14, 15].
Furthermore, concomitant problems associated with asynchrony like mitral regurgitation can be judged immediately. Since the early days of CRT, multiple echocardiographic parameters and indices have been investigated to qualify and quantify disturbed and delayed wall motion specially to localize the area of latest mechanical activation.
Early investigations started with tissue Doppler imaging (TDI), established a dyssynchrony index (SDI), and continued detection of asynchrony with speckle tracking technology using radial and longitudinal strain. These trials were retrospective and/or single-center investigations and showed that these parameters can either identify patients for CRT or predict response. Also, prospective multicenter trials which used these parameters to identify asynchrony according to these definitions failed to show the benefit of the echocardiographic guided implantation, though. Especially in patients with a QRS < 130 ms, a CRT treatment deteriorates outcome, despite the evidence of echocardiographic proven dyssynchrony [16, 17].
There are currently no data showing that intraprocedural echocardiography either transthoracically or transesophageally applied used during the CRT implantation and during the LV lead placement is useful to select the best tributary to reduce asynchrony.
Preprocedural identification of delayed contraction by radial strain and targeting this region for the final LV lead position resulted in TARGET and STARTER in a reduction of asynchrony, reverse remodeling, and diverging results regarding all-cause mortality and CRT-D free survival compared to an empiric LV lead placement [18, 19]. However, it has to be emphasized that in none of these randomized trials the LV lead position was investigated in different positions and intra-individually adapted or optimized according to intraprocedural echo parameters. Currently, no echo parameter is accepted or guideline endorsed for the identification of CRT responders [20]. This is also reflected in the 2016 ESC guidelines for the diagnosis and treatment of acute and chronic heart failure, which do not recommend using the presence of echocardiographic dyssynchrony as selection criteria for CRT [21].
Electrocardiography
Twelve-lead ECG is the traditionally and currently mainly used non-invasive method to diagnose electrical disorders and conduction disturbances. It is the easiest tool to detect left bundle branch block (LBBB) and select patients for CRT. The expansion up to 300 electrodes allows a 3D visualization of onset of activation and propagation captured from the body surface [22]. A correlation of low voltage and scar has been shown [23, 24]. Furthermore, the change of electrical activation by CRT can be detected, but no data are published showing that the spatial and temporal resolution of the electrocardiographic imaging (ECGi) is high enough to be used as guidance for the best LV lead spot during the implantation.
Multi-Modality Imaging and Image Fusion Technology
The so-called multi-modality imaging approach is based on the fusion of different imaging modalities to match information about tissue characterization (MRI; LGE) especially scar and viability (PET-CT), late mechanical or electrical activation (MRI, Echo; speckle tracking, ECGi), and its correlation to the given coronary venous anatomy (CT; MRI). This information will be fused with the CS angiography to guide the implanter to select the optimal side branch for LV lead [9, 25, 26•] (Fig. 1).
CT is superior in delineating the coronary venous tree with a higher spatial resolution compared to MRI, especially if rapid acquisition is performed in 3D technology in whole heart cycles.
This approach optimizes the efforts of non-invasive imaging, although it is expensive and time consuming and is hardly covered by the reimbursements for CRT implantations in most countries of the world—apart from clinical trials.
Disappointingly, MRI or radial strain-guided LV lead placement, in combination with multimodality imaging, did not result in increased clinical or echocardiographic response, nor in a significant reduction of death or heart failure hospitalization. However, only 77% of all patients had an optimal or adjacent LV lead placement. Thus, cardiac CT can be valuable for identification of suitable cardiac veins at the free LV wall, but reaching the intended target branch with a stable fixation remains the limiting factor as previously shown in STARTER (86% concordant) and TARGET (63% concordant) and Borgquist et al. (77%) [18, 19, 27•].
Summarizing, standard use of a targeted multimodality imaging strategy should currently not be recommended for all CRT patients but may prove valuable in select cases. Skilled implanters, easy-to-use image integration tools, and segmental evaluation strategies are all important for multimodality imaging to reach its full potential to facilitate targeted CRT implantations (Fig. 2).
Ad 3:Showing the implanter’s reality by coronary sinus venogram and estimating the accessibility of target tributaries based on experience and available tools, pure anatomical approach—DURING implant.
Generally, the implanter of a CRT device should not restrict himself to the pure mechanical adjunct but should always have a general knowledge of the patient’s medical history, the underlying cause of heart failure, and at least basic information derived from transthoracic echo and 12-lead ECG. In most routine cases, the initial intraprocedurally performed coronary sinus venogram gives the first information (or surprise) about the anatomy and uncovers the challenges to target a certain tributary. An occlusive balloon venography delivers a more detailed angiogram and allows an estimation which tributaries might be reached based on the implanter’s skills and his knowledge about tools and tricks to finish successfully [4].
According to the anatomical classification from the implant venograms used in MADIT-CRT, the final lead tip position was allocated to basal, midventricular, and apical and anterior, lateral, and posterior and correlated with outcome parameters (see Fig. 3). This retrospective analysis clearly demonstrates that an apical LV lead position is associated with an increased heart failure hospitalization and death in all subgroups [28]. A difference between an anterior, posterior, and lateral position could not be detected. Consequently, an apical position should be avoided anyhow by selecting a lead type and a tributary in which it can be fixed in a midventricular position.
In a further data analysis of MADIT-CRT, focusing of LBBB patients’ posterior or lateral LV lead location was associated with significant long-term mortality reduction, accompanied by LV reverse remodeling. The deleterious effect of an apical position was confirmed [29••].
A recently published single-center retrospective analysis of more than 2000 CRT recipients who were followed for nearly 4 years showed a significant long-term mortality benefit for a lateral placement over an anterior or posterior lead position [30].
Derived from these studies, an anatomically guided lateral lead position is never unfavorable, even advantageous in LBBB and should be the primary intended LV lead site, always avoiding being located close to the apex.
This purely anatomically guided implant strategy is mainly used in daily routine when a straightforward quick implantation is intended.
A more individual strategy should be chosen to optimize the LV lead position in complex patients with QRS < 150 ms, scar areas, and a difficult venous anatomy offering different side branches.
Ad 4: Detecting and correcting electrical and mechanical left ventricular delay including acute hemodynamic optimization—DURING implantation INVASIVELY.
Early in CRT evolution, acute hemodynamic measurements showed that hemodynamic response is highly related to different pacing sites [31,32,33]. Stimulation of either LV alone or BIV in a VDD mode using the epicardial venous access resulted in a superior effect using the free wall (what is lateral or posterolateral) compared to an anterior (via the intervententricular branch) stimulation site [31] (Fig. 4). Different tools have been used over the years to measure either changes in LV + dp/dt, which reflects the maximum change in LV contractility or the change in pulse pressure (PP), which is a parameter of change in forward stroke volume. A dual transducer micromanometer catheter (SPC-780c; Millar Instruments) can simultaneously measure both parameters; a pressure wire protected by a guiding catheter and placed in the LV (usually used for intracoronary FFR measurement) can be used for dp/dt changes alone. The simplest and cheapest tool is a fluid-filled pigtail catheter in the LV and an arterial sheath to determine either the best pacing site or the best AV-delay. It has to be emphasized that every pacing site is associated with an optimal AV-delay, thus an individualized protocol has to be run. Every patient has an individual superior pacing site.
Whereas the average LV + dp/dt increased at the free wall by 14%, it was 6% at the anterior site. Pulse pressure improved by 8% vs. 4%. LV and BIV pacing effects were similar [31] (Fig. 4).
Notably, one third of the patients showed a deterioration of contractility when paced anterior, but in no patient a negative hemodynamic effect appeared at the free wall.
This clearly demonstrates that it would be worth to identify the individual optimal pacing site and AV-delay.
A hemodynamically optimized chronic LV pacing alone showed in a 3-month cross-over design significant functional improvements in the wide QRS group (> 150 ms). The improvements were comparable in a similar study design for BIV stimulation without preceding invasive testing [1].
A LV + dp/dt response of > 10% during implant predicts a LV reverse remodeling at 6 months and thus will help to guide the lead placement [34]. Nevertheless, the baseline LV + dp/dt seems to be a better predictor for 1-year survival than the acute response using a cut-off of 650 mmHg/s [35]. These findings are not generally conflicting because baseline LV + dp/dt is a measure of global impaired contractility before CRT and is furthermore influenced by other factors like mitral regurgitation which is also substantial for survival [36]. Still, there are no arguments imaginable not to target the site with the best relative improvement.
Based on our global understanding for conduction delay, it was predominantly the goal to identify the region of latest electrical activation (LEA) [37, 38]. Whereas electro-anatomical mapping is cost intensive, time consuming, and has not made its way into the routine CRT implantation procedure, detection of LEA is simple, cheap, and can be used intraprocedurally without any other mapping equipment [39]. The delay can be evaluated at different sites according to the accessible tributaries. The so-called Q-LV-Time is the interval from the onset of the QRS on the surface ECG to the sensed signal at the tip of the transvenous epicardial LV lead. This can be either expressed in milliseconds or as percentage of the QRS width on surface ECG. Zanon et al. evaluated whether Q-LV might be used to identify the optimal site in an individual patient by systematically screening all the suitable coronary sinus (CS) veins [40]. A strong correlation was observed between Q-LV prolongation and improvements in acute hemodynamic response. A Q-LV value of greater than 95 ms appeared significant, yielding an improvement in AHR of > 10%, a finding which has been associated with predicting long-term remodeling (Fig. 5). Certainly, hemodynamic effect could have been higher if the AV-delay were optimized in every site instead of using a fixed AV-delay of 130 ms.
Whether the study of BEHAR, performed in an exceedingly small group of patients (N = 8), all ischemic with a complex protocol and showing that the region of latest endocardial electrical activation does not reflect the best hemodynamic response, should change our implantation strategy is unclear [41].
Generally, when the latest electrical activation is detected, this spot should be the preferred implantation site except it is in a transmural scar.
In a previous study by Gold et al., the implantation of the LV lead with a favorable Q-LV interval was independently associated with a greater reverse remodeling and symptomatic improvement at 6 months [37]. In a double-blind randomized trial, it was recently shown that an electrically guided CRT implantation appeared non-inferior to an imaging-guided strategy as described before considering the outcomes of change in LVEF, LV reverse remodeling, and clinical response [42].
A general guide for successful CRT implantation can be found in Table 1.
Additional Considerations—with Respect to LV Pacing Site
LV only versus BIV
Early invasive studies intended to identify the best acute hemodynamic effect for different pacing sites did not show an acute difference for both pacing modes when stimulating the same site. LV pacing alone resulted in a significant functional improvement [31, 43].
Further trials like GREATER-EARTH and B-LEFT-HF showed neither inferiority nor superiority for LV over BIV during 6–12 months follow-up regarding echocardiographic reverse remodeling parameters, clinical and functional outcome parameters as well as biomarkers [44,45,46].
In a subgroup of CRT candidates with a normal AV conduction, an adaptive algorithm might provide intrinsic RV activation which results in a LV stimulation alone. Echocardiographic measurements showed an improved and potentially more efficient LV activation sequence without reducing dyssynchrony. Whether this has a detectable clinical benefit remains unproven [47•].
In summary, there are no acute, chronic, or clinical data which suggest that a different LV pacing site is favorable depending on an intended LV or BIV pacing mode. Thus, this is not to be considered during the implantation.
Multipoint Pacing (MPP)
Multipoint pacing (MPP) refers to pacing from more than one pole of a multipolar LV lead. Physiologically, this also results in “multifocal” or “trifocal” stimulation, albeit from only two leads.
Quadripolar CS leads offer the opportunity to program different vectors and activate larger areas of myocardium by cathodal stimulation from two widely separated electrodes. Besides the greater opportunities regarding threshold, intrinsic sensing and phrenic nerve stimulation multipoint pacing might be of superior benefit in extremely enlarged left ventricles and in non-responders as it might be suspected by retrospective analysis [48]. In acute hemodynamic measurements in typical LBBB patients, an adjunct benefit over conventional BIV pacing was proven at any site, but after optimization of BIV, MPP was not always superior. Thus, primarily the optimization of the LV pacing site is of greater relevance than MPP itself [49, 50•, 51].
Finally, a meta-analysis comparing MPP with BIP group suggested that MPP seems to be superior and decreased heart failure hospitalizations, improved LVEF, increased CRT response, and decreased all-cause morbidity and cardiovascular death rate [52].
Conduction System Pacing, HIS, and LBB Pacing
Biventricular pacing using the epicardial venous system is the most established approach to resynchronize the atrioventricular filling and the interaction between both ventricles. This approach has been studied extensively over more than two decades and the results in heart failure patients with a QRS > 130 ms and typical LBBB configuration are convincing and have been implemented in all guidelines. Nevertheless, biventricular pacing in other conduction disorders such as prolonged PR interval, QRS < 130 ms, and atypical LBBB or RBBB has shown a diverging or even deleterious effect. Right ventricular pacing itself has also a negative impact on cardiac function and survival in selected patients. Thus, the idea came up to access the original conduction system at different levels to allow a physiological electrical propagation.
Already in 1977, Narula et al. described a narrowing of QRS in a LBBB patient when the HIS bundle was stimulated [53]. Upadyay et al. found, using intracardiac septal mapping, that conduction block within the left‐sided His fibers or proximal portion of the left bundle branch was present in 64% of patients with a 12‐lead ECG appearance of LBBB [54]. These patients have the highest chance of successful resynchronization with His bundle pacing (HBP). Surface 12-lead ECG is not predictive to localize the region of propagation disturbance which makes patient selection critical without invasive evaluation [54, 55].
A short left ventricular activation (LVAT95) time as invasively investigated by Arnold et al. identifies patients in whom HBP does not result in a normalization of electrical propagation. Before suggesting HBP as a serious alternative to epicardial transvenous BIVP, it is important to distinguish disease conceptually and anatomically within the His–Purkinje conduction system from the appearance of propagation discontinuity on the LV epicardial surface [56].
Although the technique of HBP can be readily learned with a high level of success after an appropriate learning curve, its success is still adversely impacted by the atrioventricular (AV) nodal and infra-Hisian conduction disease [57,58,59]. Still, HBP is limited by poor ventricular sensing, elevated acute and chronic pacing thresholds necessitating lead reintervention and rapid battery replacement, and failure to reliably normalize the QRS complex in patients with bundle branch block. The benefits and limitations of HBP were evident in the pilot His-SYNC trial (His Bundle Pacing Versus Coronary Sinus Pacing for Cardiac Resynchronization Therapy) [7], which demonstrated in an on-treatment analysis that HBP resulted in superior electrical resynchronization with a trend toward improved echocardiographic response as compared with BIV; however, nearly one half of the patients randomized to the HBP arm required crossover to BIV, most commonly due to a failure to correct the wide QRS duration (> 150 ms) [60]. Targeting the region below HIS bundle is emerging as left bundle branch pacing (LBP). The idea behind this is to screw a lead through the interventricular septum until the left bundle can be stimulated, but without perforation of the septum. This method seems technically feasible with a success above 90%, roughly 10% injury of the right bundle branch and increase of tricuspid regurgitation initially described by Huang et al. and expanded by an observational prospective experience of more than 600 patients with a pacemaker indication [61, 62].
Summarizing, technically it is feasible to stimulate either the HIS bundle or the left bundle branch with improved tools, but it is too early to dethrone conventional transvenous biventricular or left ventricular pacing. The current evidence for HBP is superior to LBP, but both are still in their infancy compared to the long robust scientific evaluation of conventional BVP. Furthermore, it is still unclear which conduction disorder is treated best by which stimulation approach.
Epicardial versus Endocardial (WISE) Pacing—Retrograde versus Transseptal
Based on the physiological electrical propagation from the endocardial to the epicardial space, a primarily endocardial stimulation will be superior and might not be limited by the coronary venous anatomy. Whereas modern leadless pacing is mainly used in single chamber pacing of the right ventricle, recently allowing an AV-synchronous pacing triggered by heart sounds and not by the electrical atrial signal, left ventricular endocardial pacing is more challenging. In this context, the prospective, multi-center ALSYNC study proved feasibility of a transseptal approach to implant an endocardial LV lead with a considerable clinical outcome after 6 months. There are a few limitations, though. Transseptal puncture via the SVC may be challenging; a permanent lead in the LV is associated with a significant risk of thromboembolic events requiring long-term anticoagulation and mitral valve damage due to mechanical irritation or endocarditis. Therefore, a transvenous permanent endocardial approach should be limited to exceptional cases. To overcome the risk of thromboembolic events and to avoid permanent anticoagulation, a tiny receiver electrode of less than 10 mm length and 2.7 mm diameter was developed for the WISE-CRT system (EBR Systems, Sunnyvale, California). The receiver transforms ultrasound energy delivered from the transmitter which is implanted at the intercostal space and is connected to the battery. The system routinely uses an arterial femoral access with retrograde crossing of the aortic valve and needs a mandatory RV pacing electrode which is required for triggering. Further limitations are the distance between transmitter and receiver, the angulation between both, and an early battery depletion. Predominant complications are related to the femoral access and its closure. Based on the closure experience after TAVR procedures, this problem can be overcome in an interventional team, but a transseptal access has also been described if an arterial access is impossible [63].
The SELECT-LV trails as well as the real-world post-market registry have proven the technical feasibility but have also shown the potential risks and complications as perforation of the LV, induction of arrhythmias, and peripheral embolization of the receiver in one patient [64,65,66]. The clinical and reverse remodeling response was comparable to the standard biventricular pacing with approximately 70%. However, it has to be kept in mind that patients selected for the leadless ultrasound stimulation have been either clinical non-responder to standard BIV pacing or had no accessible cardiac veins.
Generally, this innovative technology offers promising LV stimulation options, but requires further improvement as the lack of RV pacing, simplification and miniaturization of delivery tools, and improvement of battery longevity.
How To Place a Left Ventricular Lead for Cardiac Resynchronization Therapy
Corrective pacing to improve the activation of electrically delayed left ventricular wall segments with subsequent dyssynchronous mechanical deformation is the cornerstone of cardiac resynchronization. In most patients with LBBB, the electro-mechanical delay is found most pronounced at the lateral left ventricular wall. Placement of left ventricular pacing leads for CRT is therefore mainly aimed at attaining a stable and electrically appropriate left ventricular lateral position of the pacing electrode(s). At the same time, intolerable unintended phrenic nerve stimulation must absolutely be avoided.
Several routes are possible to reach these goals. Since more than two decades, the standard way to obtain a left ventricular lateral electrode position is transvenous lead placement. The LV lateral wall is reached using lateral coronary venous branches draining to the main coronary sinus or the great cardiac vein [67].
Before the introduction of transvenous lead placement, open surgery and left ventricular epicardial lead implantation has been the standard way to pace the left ventricle [68]. This method is still in use, although no longer the standard technique. Epicardial leads are sutured to the epicardial surface or screwed into the myocardium. Alternative approaches to LV pacing that are still awaiting thorough evaluation comprise ultrasound-activated LV endocardial electrode and transseptal left bundle branch pacing (Fig. 6f, g) [69, 70].
Transvenous LV Lead Placement over Coronary Sinus Branches
Transvenous LV lead placement may be complex and requires specific training and experience. This has been addressed by structured education programs [71].
After introducing a 0.035 wire and a pre-shaped guiding sheath into the right heart, engagement of the coronary sinus ostium with an atraumatic wire followed by the guiding catheter is the first step to successful lead implantation. The CS ostium can be located by pulling back the catheter from the RV across the tricuspid valve while applying some counterclockwise torque. Immediately after crossing the tricuspid valve, the catheter moves posteriorly, and the tip reaches the CS ostial region. In many patients, the entry into the CS ostium is complicated by the presence of an ostial (“Thebesian”) valve (Fig. 6d), which enables blood drainage from the CS to the right atrium but impedes backflow to the CS and hinders the entry of the (frequently upward oriented) catheter tip. Usually, the guiding catheters have a multipurpose-like distal shape, but different catheter curves are available to fit individual anatomical properties and should be tried after unsuccessful attempts to cannulate the ostium. In many cases, a diagnostic left coronary Amplatz 1 or Amplatz 2 catheter (inserted over the 0.035-in. wire) helps to orientate the tip downward and allows getting beyond the Thebesian valve with the wire followed by the catheter. A second venous valve (Vieussens valve, Fig. 6a) is encountered at the mid-coronary sinus region in the majority of patients and may be a relevant obstacle to transvenous LV lead placement in up to 20% of patients [72]. The guiding catheter should only be passed through these valves on a guidewire, as blunt retrograde manipulation to overcome an obstructing Vieussens valve will likely result in extensive CS dissection (Fig. 6e). Sometimes the successful passage of a rigid valve may afford considerable effort and sophisticated experienced telescoping technique (Fig. 6b, c) resembling the treatment of a coronary artery chronic total occlusion. LV transvenous leads are usually implanted over a 0.014-in. guidewire. The angle of LV lateral veins with reference to their orifice to the CS is frequently sharp and may make it hard to advance a guidewire into the side branch and to overcome the kinked implantation route with the LV lead over the wire. Possible strategies to mitigate this problem include the use of firm guidewires and the placement of a second (“buddy”) wire. The most promising approach to lead implantation over sharp angles is the use of a sub-selective inner catheter with an angled tip. These catheters are advanced distally to the side branch orifice over a guide wire with or without a dilator. After drawback of the wire (and the dilator, if used), the sub-selective catheter enables side branch engagement through gentle pull-back and rotation. The threshold to use sub-selective inner catheters to cannulate angulated side branches should be low, as these catheters not only enable wire placement in difficult anatomies, but also provide the required back-up to advance the lead over the wire to the distal CS side branch.
The LV lead should be selected according to the side branch anatomy. Various hook-like or spiral shapes are available in different lengths (of the entire lead and the distal straight part). Stabilization within the vein is accomplished by advancing the tip into a distal wedge position. Guidewire withdrawal within the lead to the main CS releases the pre-shaped lead configuration and increases the lead adherence to the venous wall. Quadripolar leads should be preferred over bipolar or unipolar leads in order to increase the variability of pacing configurations, to avoid phrenic nerve stimulation, and to enable simultaneous multi-point pacing (MPP, see above) of the LV lateral myocardium.
In large straight vein branches or in redo procedures after lead dislodgement, with anticipated poor lead stability, a special active fixation lead (e.g., Medtronic Attain Stability Quad) is recommended. This lead is equipped with a sharp hook that is attached to the lead body at some distance proximal to the lead tip. By clockwise rotation, the lateral hook engages the venous wall firmly and stabilizes the electrode in a desired position.
An interesting additional strategy is the use of collateral veins to reach a desired pacing region. In many patients, the size of venous collaterals allows implanting conventional 4F leads over this way. A new promising technology, called LV micro-lead implantation, has been reported to enable implant quadripolar very thin leads over very thin collaterals using a dedicated wire core with four electrodes (1,2F) and a specific thin catheter (tip 2,4F) [73•].
Interventional techniques like balloon dilatation and snaring of a guide wire may also help to overcome difficult coronary venous anatomies [74]. A schematic representation of pitfalls of LV lead implantation is shown in Fig. 7.
Conclusion
Practical recommendations—workflow:
-
Transvenous epicardial approach:
Non-ischemic DCM with LBBB > 150 ms without previous interventions
LV implantations site less critical, but
-
Based on venogram: avoid apical position, avoid anterior position.
-
Target postero-lateral or lateral position (confirm in 2 angulations in fluoro).
Ischemic cardiomyopathy with QRS 130–150 ms and not typical LBBB, re-implantations, chronic “non-responder”
LV lead implant site crucial for response
-
If implantation is only anatomical guided, target the same regions as in DCM, but
-
Expend pre-interventional imaging (MRI, tissue Doppler imaging, CT) to detect scar and late mechanical activation.
-
Avoid placing in a scar region.
-
Identify target region by delayed electrical activation and verify with CS venogram the availability of tributary in aimed site.
-
If more than one tributary will touch the target region, verify the superiority by acute hemodynamic response or shortening of the electrical delay using at least different AV-intervals.
-
Narrowing of the paced surface QRS will additionally be helpful.
Generally, an individualized approach based on the acute improvement of hemodynamics and/or electrical parameters would be the goal to achieve the most benefit in each patient, but in routine these options are barely used.
Alternate pacing options
If a target vein is not accessible, an alternate stimulation approach via the conduction system as HIS bundle or left bundle brunch pacing should be considered depending on the conduction disorder. A leadless endocardial left ventricular stimulation might be a further option.
Change history
05 March 2022
The original version of this paper was updated due to the insertion of Open Access funding note.
Abbreviations
- BIV/BIVP:
-
Biventricular pacing
- CMRI:
-
Cardiac magnetic resonance imaging
- CRT:
-
Cardiac resynchronization therapy
- CS:
-
Coronary sinus
- CSP:
-
Conduction system pacing
- DSCT:
-
Dual-source computed tomography
- ECGI:
-
Electrocardiographic imaging
- HBP:
-
His bundle pacing
- LBBB:
-
Left bundle branch block
- LBP:
-
Left bundle pacing
- LEA:
-
Latest electrical activation
- LGE:
-
Late gadolinium enhancement
- LV:
-
Left ventricle
- MPP:
-
Multipoint pacing
- NYHA:
-
New York Heart Association
- QOL:
-
Quality of life
- PP:
-
Pulse pressure
- SDI:
-
Systolic dyssynchrony index
- TDI:
-
Tissue Doppler imaging
References
Papers of particular interest, published recently, have been highlighted as: • Of importance •• Of major importance
Cazeau S, Leclercq C, Lavergne T, Walker S, Varma C, Linde C, et al. Effects of multisite biventricular pacing in patients with heart failure and intraventricular conduction delay. N Engl J Med. 2001;344(12):873–80. https://doi.org/10.1056/NEJM200103223441202.
Mower M. Method and apparatus for treating hemodynamic dysfunction by simultaneous pacing of both ventricles. US Patent #4,928,688, 29 May 1990.
Daubert C, Behar N, Martins RP, Mabo P, Leclercq C. Avoiding non-responders to cardiac resynchronization therapy: a practical guide. Eur Heart J. 2017;38(19):1463–72. https://doi.org/10.1093/eurheartj/ehw270.
Pothineni NVK, Supple GE. Navigating challenging left ventricular lead placements for cardiac resynchronization therapy. J Innov Card Rhythm Manag. 2020;11(5):4107–17. https://doi.org/10.19102/icrm.2020.110505.
Writing C, Maddox TM, Januzzi JL Jr, Allen LA, Breathett K, Butler J, et al. 2021 Update to the 2017 ACC expert consensus decision pathway for optimization of heart failure treatment: answers to 10 pivotal issues about heart failure with reduced ejection fraction: a report of the American College of Cardiology Solution Set Oversight Committee. J Am Coll Cardiol. 2021;77(6):772–810. https://doi.org/10.1016/j.jacc.2020.11.022.
Leyva F, Foley PW, Chalil S, Ratib K, Smith RE, Prinzen F, et al. Cardiac resynchronization therapy guided by late gadolinium-enhancement cardiovascular magnetic resonance. J Cardiovasc Magn Reson. 2011;13:29. https://doi.org/10.1186/1532-429X-13-29.
White JA, Yee R, Yuan X, Krahn A, Skanes A, Parker M, et al. Delayed enhancement magnetic resonance imaging predicts response to cardiac resynchronization therapy in patients with intraventricular dyssynchrony. J Am Coll Cardiol. 2006;48(10):1953–60. https://doi.org/10.1016/j.jacc.2006.07.046.
Ypenburg C, Roes SD, Bleeker GB, Kaandorp TA, de Roos A, Schalij MJ, et al. Effect of total scar burden on contrast-enhanced magnetic resonance imaging on response to cardiac resynchronization therapy. Am J Cardiol. 2007;99(5):657–60. https://doi.org/10.1016/j.amjcard.2006.09.115.
Sommer A, Kronborg MB, Norgaard BL, Poulsen SH, Bouchelouche K, Bottcher M, et al. Multimodality imaging-guided left ventricular lead placement in cardiac resynchronization therapy: a randomized controlled trial. Eur J Heart Fail. 2016;18(11):1365–74. https://doi.org/10.1002/ejhf.530.
Yagishita D, Shoda M, Yagishita Y, Ejima K, Hagiwara N. Time interval from left ventricular stimulation to QRS onset is a novel predictor of nonresponse to cardiac resynchronization therapy. Heart Rhythm. 2019;16(3):395–402. https://doi.org/10.1016/j.hrthm.2018.08.035.
Spragg DD, Dong J, Fetics BJ, Helm R, Marine JE, Cheng A, et al. Optimal left ventricular endocardial pacing sites for cardiac resynchronization therapy in patients with ischemic cardiomyopathy. J Am Coll Cardiol. 2010;56(10):774–81. https://doi.org/10.1016/j.jacc.2010.06.014.
Bai W, Xu X, Ji H, Liu J, Ma H, Xie H, et al. Evaluation of the anatomical variations of the coronary venous system in patients with coronary artery calcification using 256-slice computed tomography. PLoS ONE. 2020;15(11):e0242216. https://doi.org/10.1371/journal.pone.0242216.
Markstad H, Bakos Z, Ostenfeld E, Geijer M, Carlsson M, Borgquist R. Preoperative CT of cardiac veins for planning left ventricular lead placement in cardiac resynchronization therapy. Acta Radiol. 2019;60(7):859–65. https://doi.org/10.1177/0284185118803796.
Gaibazzi N, Bianconcini M, Marziliano N, Parrini I, Conte MR, Siniscalchi C, et al. Scar detection by pulse-cancellation echocardiography: validation by CMR in patients with recent STEMI. JACC Cardiovasc Imaging. 2016;9(11):1239–51. https://doi.org/10.1016/j.jcmg.2016.01.021.
Montant P, Chenot F, Goffinet C, Poncelet A, Vancraeynest D, Pasquet A, et al. Detection and quantification of myocardial scars by contrast-enhanced 3D echocardiography. Circ Cardiovasc Imaging. 2010;3(4):415–23. https://doi.org/10.1161/CIRCIMAGING.109.912477.
Chung ES, Leon AR, Tavazzi L, Sun JP, Nihoyannopoulos P, Merlino J, et al. Results of the Predictors of Response to CRT (PROSPECT) trial. Circulation. 2008;117(20):2608–16. https://doi.org/10.1161/CIRCULATIONAHA.107.743120.
Ruschitzka F, Abraham WT, Singh JP, Bax JJ, Borer JS, Brugada J, et al. Cardiac-resynchronization therapy in heart failure with a narrow QRS complex. N Engl J Med. 2013;369(15):1395–405. https://doi.org/10.1056/NEJMoa1306687.
Adelstein E, Alam MB, Schwartzman D, Jain S, Marek J, Gorcsan J, et al. Effect of echocardiography-guided left ventricular lead placement for cardiac resynchronization therapy on mortality and risk of defibrillator therapy for ventricular arrhythmias in heart failure patients (from the Speckle Tracking Assisted Resynchronization Therapy for Electrode Region [STARTER] trial). Am J Cardiol. 2014;113(9):1518–22. https://doi.org/10.1016/j.amjcard.2014.01.431.
Khan FZ, Virdee MS, Palmer CR, Pugh PJ, O’Halloran D, Elsik M, et al. Targeted left ventricular lead placement to guide cardiac resynchronization therapy: the TARGET study: a randomized, controlled trial. J Am Coll Cardiol. 2012;59(17):1509–18. https://doi.org/10.1016/j.jacc.2011.12.030.
Spartalis M, Tzatzaki E, Spartalis E, Damaskos C, Athanasiou A, Livanis E, et al. The role of echocardiography in the optimization of cardiac resynchronization therapy: current evidence and future perspectives. Open Cardiovasc Med J. 2017;11:133–45. https://doi.org/10.2174/1874192401711010133.
Ponikowski P, Voors AA, Anker SD, Bueno H, Cleland JG, Coats AJ et al. 2016 ESC Guidelines for the diagnosis and treatment of acute and chronic heart failure: The Task Force for the Diagnosis and Treatment of Acute and Chronic Heart Failure of the European Society of Cardiology (ESC). Developed with the special contribution of the Heart Failure Association (HFA) of the ESC. Eur J Heart Fail. 2016;18(8):891–975. https://doi.org/10.1002/ejhf.592.
Pereira H, Niederer S, Rinaldi CA. Electrocardiographic imaging for cardiac arrhythmias and resynchronization therapy. Europace. 2020. https://doi.org/10.1093/europace/euaa165.
Bank AJ, Gage RM, Curtin AE, Burns KV, Gillberg JM, Ghosh S. Body surface activation mapping of electrical dyssynchrony in cardiac resynchronization therapy patients: potential for optimization. J Electrocardiol. 2018;51(3):534–41. https://doi.org/10.1016/j.jelectrocard.2017.12.004.
Gage RM, Curtin AE, Burns KV, Ghosh S, Gillberg JM, Bank AJ. Changes in electrical dyssynchrony by body surface mapping predict left ventricular remodeling in patients with cardiac resynchronization therapy. Heart Rhythm. 2017;14(3):392–9. https://doi.org/10.1016/j.hrthm.2016.11.019.
Nguyen UC, Cluitmans MJM, Strik M, Luermans JG, Gommers S, Wildberger JE, et al. Integration of cardiac magnetic resonance imaging, electrocardiographic imaging, and coronary venous computed tomography angiography for guidance of left ventricular lead positioning. Europace. 2019;21(4):626–35. https://doi.org/10.1093/europace/euy292.
• Salden OAE, van den Broek HT, van Everdingen WM, Mohamed Hoesein FAA, Velthuis BK, Doevendans PA, et al. Multimodality imaging for real-time image-guided left ventricular lead placement during cardiac resynchronization therapy implantations. Int J Cardiovasc Imaging. 2019;35(7):1327–37. https://doi.org/10.1007/s10554-019-01574-0. Shows the feasability of real-time image-guided LV lead placement including fused CT and cMR images on the flurouscopic image during CRT implantation. This is a promising (although expensive) alternative for complex implant cases.
• Borgquist R, Carlsson M, Markstad H, Werther-Evaldsson A, Ostenfeld E, Roijer A, et al. Cardiac resynchronization therapy guided by echocardiography, MRI, and CT imaging: a randomized controlled study. JACC Clin Electrophysiol. 2020;6(10):1300–9. https://doi.org/10.1016/j.jacep.2020.05.011. In this unselected group of 102 patients eligible for CRT, a multimodal imaging implantation approach did not result in an improved clinical outcome after 6 months.
Singh JP, Klein HU, Huang DT, Reek S, Kuniss M, Quesada A, et al. Left ventricular lead position and clinical outcome in the multicenter automatic defibrillator implantation trial-cardiac resynchronization therapy (MADIT-CRT) trial. Circulation. 2011;123(11):1159–66. https://doi.org/10.1161/CIRCULATIONAHA.110.000646.
•• Kutyifa V, Kosztin A, Klein HU, Biton Y, Nagy VK, Solomon SD, et al. Left ventricular lead location and long-term outcomes in cardiac resynchronization therapy patients. JACC Clin Electrophysiol. 2018;4(11):1410–20. https://doi.org/10.1016/j.jacep.2018.07.006. This study emphasizes the importance of the LV lead location site. In the long-term observation of the MADIT CRT cohort, a mortality benefit assosiated with CRT-D was only seen in patients with lateral/posterior LV lead location whereas an apical position was associated with an early risk for death.
Behon A, Schwertner WR, Merkel ED, Kovacs A, Lakatos BK, Zima E, et al. Lateral left ventricular lead position is superior to posterior position in long-term outcome of patients who underwent cardiac resynchronization therapy. ESC Heart Fail. 2020. https://doi.org/10.1002/ehf2.13066.
Butter C, Auricchio A, Stellbrink C, Fleck E, Ding J, Yu Y, et al. Effect of resynchronization therapy stimulation site on the systolic function of heart failure patients. Circulation. 2001;104(25):3026–9. https://doi.org/10.1161/hc5001.102229.
Butter C, Auricchio A, Stellbrink C, Schlegl M, Fleck E, Horsch W, et al. Should stimulation site be tailored in the individual heart failure patient? Am J Cardiol. 2000;86(9A):144K-K151. https://doi.org/10.1016/s0002-9149(00)01385-0.
Stellbrink C, Breithardt OA, Franke A, Sack S, Bakker P, Auricchio A, et al. Impact of cardiac resynchronization therapy using hemodynamically optimized pacing on left ventricular remodeling in patients with congestive heart failure and ventricular conduction disturbances. J Am Coll Cardiol. 2001;38(7):1957–65. https://doi.org/10.1016/s0735-1097(01)01637-0.
Duckett SG, Ginks M, Shetty AK, Bostock J, Gill JS, Hamid S, et al. Invasive acute hemodynamic response to guide left ventricular lead implantation predicts chronic remodeling in patients undergoing cardiac resynchronization therapy. J Am Coll Cardiol. 2011;58(11):1128–36. https://doi.org/10.1016/j.jacc.2011.04.042.
Bogaard MD, Houthuizen P, Bracke FA, Doevendans PA, Prinzen FW, Meine M, et al. Baseline left ventricular dP/dtmax rather than the acute improvement in dP/dtmax predicts clinical outcome in patients with cardiac resynchronization therapy. Eur J Heart Fail. 2011;13(10):1126–32. https://doi.org/10.1093/eurjhf/hfr094.
Kosmidou I, Lindenfeld J, Abraham WT, Kar S, Lim DS, Mishell JM, et al. Transcatheter mitral valve repair in patients with and without cardiac resynchronization therapy: the COAPT Trial. Circ Heart Fail. 2020;13(11): e007293. https://doi.org/10.1161/CIRCHEARTFAILURE.120.007293.
Gold MR, Birgersdotter-Green U, Singh JP, Ellenbogen KA, Yu Y, Meyer TE, et al. The relationship between ventricular electrical delay and left ventricular remodelling with cardiac resynchronization therapy. Eur Heart J. 2011;32(20):2516–24. https://doi.org/10.1093/eurheartj/ehr329.
Singh JP, Fan D, Heist EK, Alabiad CR, Taub C, Reddy V, et al. Left ventricular lead electrical delay predicts response to cardiac resynchronization therapy. Heart Rhythm. 2006;3(11):1285–92. https://doi.org/10.1016/j.hrthm.2006.07.034.
Rad MM, Blaauw Y, Dinh T, Pison L, Crijns HJ, Prinzen FW, et al. Left ventricular lead placement in the latest activated region guided by coronary venous electroanatomic mapping. Europace. 2015;17(1):84–93. https://doi.org/10.1093/europace/euu221.
Zanon F, Baracca E, Pastore G, Fraccaro C, Roncon L, Aggio S, et al. Determination of the longest intrapatient left ventricular electrical delay may predict acute hemodynamic improvement in patients after cardiac resynchronization therapy. Circ Arrhythm Electrophysiol. 2014;7(3):377–83. https://doi.org/10.1161/CIRCEP.113.000850.
Behar JM, Jackson T, Hyde E, Claridge S, Gill J, Bostock J, et al. Optimized left ventricular endocardial stimulation is superior to optimized epicardial stimulation in ischemic patients with poor response to cardiac resynchronization therapy: a combined magnetic resonance imaging, electroanatomic contact mapping, and hemodynamic study to target endocardial lead placement. JACC Clin Electrophysiol. 2016;2(7):799–809. https://doi.org/10.1016/j.jacep.2016.04.006.
Stephansen C, Sommer A, Kronborg MB, Jensen JM, Norgaard BL, Gerdes C, et al. Electrically vs. imaging-guided left ventricular lead placement in cardiac resynchronization therapy: a randomized controlled trial. Europace. 2019;21(9):1369–77. https://doi.org/10.1093/europace/euz184.
Auricchio A, Stellbrink C, Butter C, Sack S, Vogt J, Misier AR, et al. Clinical efficacy of cardiac resynchronization therapy using left ventricular pacing in heart failure patients stratified by severity of ventricular conduction delay. J Am Coll Cardiol. 2003;42(12):2109–16. https://doi.org/10.1016/j.jacc.2003.04.003.
Boriani G, Kranig W, Donal E, Calo L, Casella M, Delarche N, et al. A randomized double-blind comparison of biventricular versus left ventricular stimulation for cardiac resynchronization therapy: the Biventricular versus Left Univentricular Pacing with ICD Back-up in Heart Failure Patients (B-LEFT HF) trial. Am Heart J. 2010;159(6):1052-8 e1. https://doi.org/10.1016/j.ahj.2010.03.008.
Skaf S, Thibault B, Khairy P, O’Meara E, Fortier A, Vakulenko HV, et al. Impact of left ventricular vs biventricular pacing on reverse remodelling: insights from the Evaluation of Resynchronization Therapy for Heart Failure (EARTH) Trial. Can J Cardiol. 2017;33(10):1274–82. https://doi.org/10.1016/j.cjca.2017.07.478.
Thibault B, Ducharme A, Harel F, White M, O’Meara E, Guertin MC, et al. Left ventricular versus simultaneous biventricular pacing in patients with heart failure and a QRS complex >/=120 milliseconds. Circulation. 2011;124(25):2874–81. https://doi.org/10.1161/CIRCULATIONAHA.111.032904.
• Burns KV, Gage RM, Curtin AE, Gorcsan J 3rd, Bank AJ. Left ventricular-only pacing in heart failure patients with normal atrioventricular conduction improves global function and left ventricular regional mechanics compared with biventricular pacing: an adaptive cardiac resynchronization therapy sub-study. Eur J Heart Fail. 2017;19(10):1335–43. https://doi.org/10.1002/ejhf.906. Adaptive CRT algorithms with LV only pacing may improve LV echocardiographic parameters compared with biventricular pacing in patients with intact AV conduction.
Varma N, Baker J 2nd, Tomassoni G, Love CJ, Martin D, Sheppard R, et al. Left ventricular enlargement, cardiac resynchronization therapy efficacy, and impact of multipoint pacing. Circ Arrhythm Electrophysiol. 2020;13(11):e008680. https://doi.org/10.1161/CIRCEP.120.008680.
Engels EB, Vis A, van Rees BD, Marcantoni L, Zanon F, Vernooy K, et al. Improved acute haemodynamic response to cardiac resynchronization therapy using multipoint pacing cannot solely be explained by better resynchronization. J Electrocardiol. 2018;51(6S):S61–6. https://doi.org/10.1016/j.jelectrocard.2018.07.011.
• van Everdingen WM, Zweerink A, Salden OAE, Cramer MJ, Doevendans PA, Engels EB, et al. Pressure–volume loop analysis of multipoint pacing with a quadripolar left ventricular lead in cardiac resynchronization therapy. JACC Clin Electrophysiol. 2018;4(7):881–9. https://doi.org/10.1016/j.jacep.2018.02.005. MPP is a valuable tool that may improve acute hemodynamic outcome in selected patients, although the optimization of the LV pacing site remains the cornerstone of LV lead implantation.
Zanon F, Baracca E, Pastore G, Marcantoni L, Fraccaro C, Lanza D, et al. Multipoint pacing by a left ventricular quadripolar lead improves the acute hemodynamic response to CRT compared with conventional biventricular pacing at any site. Heart Rhythm. 2015;12(5):975–81. https://doi.org/10.1016/j.hrthm.2015.01.034.
Hu F, Zheng L, Ding L, Du Z, Liang E, Wu L, et al. Clinical outcome of left ventricular multipoint pacing versus conventional biventricular pacing in cardiac resynchronization therapy: a systematic review and meta-analysis. Heart Fail Rev. 2018;23(6):927–34. https://doi.org/10.1007/s10741-018-9737-5.
Narula OS. Longitudinal dissociation in the His bundle. Bundle branch block due to asynchronous conduction within the His bundle in man. Circulation. 1977;56(6):996–1006. https://doi.org/10.1161/01.CIR.56.6.996.
Upadhyay GA, Cherian T, Shatz DY, Beaser AD, Aziz Z, Ozcan C, et al. Intracardiac delineation of septal conduction in left bundle-branch block patterns. Circulation. 2019;139(16):1876–88. https://doi.org/10.1161/CIRCULATIONAHA.118.038648.
Upadhyay GA, Vijayaraman P, Nayak HM, Verma N, Dandamudi G, Sharma PS, et al. His corrective pacing or biventricular pacing for cardiac resynchronization in heart failure. J Am Coll Cardiol. 2019;74(1):157–9. https://doi.org/10.1016/j.jacc.2019.04.026.
Arnold AD, Shun-Shin MJ, Keene D, Howard JP, Chow JJ, Lim E, et al. Electrocardiographic predictors of successful resynchronization of left bundle branch block by His bundle pacing. J Cardiovasc Electrophysiol. 2021;32(2):428–38. https://doi.org/10.1111/jce.14845.
Bhatt AG, Musat DL, Milstein N, Pimienta J, Flynn L, Sichrovsky T, et al. The efficacy of His bundle pacing: lessons learned from implementation for the first time at an experienced electrophysiology center. JACC Clin Electrophysiol. 2018;4(11):1397–406. https://doi.org/10.1016/j.jacep.2018.07.013.
Keene D, Arnold AD, Jastrzebski M, Burri H, Zweibel S, Crespo E, et al. His bundle pacing, learning curve, procedure characteristics, safety, and feasibility: insights from a large international observational study. J Cardiovasc Electrophysiol. 2019;30(10):1984–93. https://doi.org/10.1111/jce.14064.
Sharma PS, Dandamudi G, Herweg B, Wilson D, Singh R, Naperkowski A, et al. Permanent His-bundle pacing as an alternative to biventricular pacing for cardiac resynchronization therapy: a multicenter experience. Heart Rhythm. 2018;15(3):413–20. https://doi.org/10.1016/j.hrthm.2017.10.014.
Upadhyay GA, Vijayaraman P, Nayak HM, Verma N, Dandamudi G, Sharma PS, et al. On-treatment comparison between corrective His bundle pacing and biventricular pacing for cardiac resynchronization: a secondary analysis of the His-SYNC Pilot Trial. Heart Rhythm. 2019;16(12):1797–807. https://doi.org/10.1016/j.hrthm.2019.05.009.
Huang W, Wu S, Vijayaraman P, Su L, Chen X, Cai B, et al. Cardiac resynchronization therapy in patients with nonischemic cardiomyopathy using left bundle branch pacing. JACC Clin Electrophysiol. 2020;6(7):849–58. https://doi.org/10.1016/j.jacep.2020.04.011.
Su L, Wang S, Wu S, Xu L, Huang Z, Chen X, et al. Long-term safety and feasibility of left bundle branch pacing in a large single-center study. Circ Arrhythm Electrophysiol. 2021;14(2):e009261. https://doi.org/10.1161/CIRCEP.120.009261.
Sieniewicz BJ, Gould JS, Rimington HM, Ioannou N, Rinaldi CA. Transseptal delivery of a leadless left ventricular endocardial pacing electrode. JACC Clin Electrophysiol. 2017;3(11):1333–5. https://doi.org/10.1016/j.jacep.2017.04.020.
Butter C, Fehrendt S, Moller V, Seifert M. Leadless endocardial ultrasound based left ventricular stimulation: WISE CRT System: alternative to conventional methods. Herzschrittmacherther Elektrophysiol. 2018;29(4):340–8. https://doi.org/10.1007/s00399-018-0605-8.
Reddy VY, Miller MA, Neuzil P, Sogaard P, Butter C, Seifert M, et al. Cardiac resynchronization therapy with wireless left ventricular endocardial pacing: the SELECT-LV study. J Am Coll Cardiol. 2017;69(17):2119–29. https://doi.org/10.1016/j.jacc.2017.02.059.
Sieniewicz BJ, Betts TR, James S, Turley A, Butter C, Seifert M, et al. Real-world experience of leadless left ventricular endocardial cardiac resynchronization therapy: a multicenter international registry of the WiSE-CRT pacing system. Heart Rhythm. 2020;17(8):1291–7. https://doi.org/10.1016/j.hrthm.2020.03.002.
Daubert JC, Ritter P, Le Breton H, Gras D, Leclercq C, Lazarus A, et al. Permanent left ventricular pacing with transvenous leads inserted into the coronary veins. Pacing Clin Electrophysiol. 1998;21(1 Pt 2):239–45. https://doi.org/10.1111/j.1540-8159.1998.tb01096.x.
Auricchio A, Stellbrink C, Sack S, Block M, Vogt J, Bakker P, et al. The Pacing Therapies for Congestive Heart Failure (PATH-CHF) study: rationale, design, and endpoints of a prospective randomized multicenter study. Am J Cardiol. 1999;83(5B):130D-D135. https://doi.org/10.1016/s0002-9149(98)01014-5.
Seifert M, Butter C. Evaluation of wireless stimulation of the endocardium, WiSE, technology for treatment heart failure. Expert Rev Med Devices. 2016;13(6):523–31. https://doi.org/10.1080/17434440.2016.1187559.
Rademakers LM, van Hunnik A, Kuiper M, Vernooy K, van Gelder B, Bracke FA, et al. A possible role for pacing the left ventricular septum in cardiac resynchronization therapy. JACC Clin Electrophysiol. 2016;2(4):413–22. https://doi.org/10.1016/j.jacep.2016.01.010.
Mascheroni J, Mont L, Stockburger M, Patwala A, Retzlaff H, Gallagher AG. A validation study of intraoperative performance metrics for training novice cardiac resynchronization therapy implanters. Int J Cardiol. 2020;307:48–54. https://doi.org/10.1016/j.ijcard.2020.02.003.
Randhawa A, Saini A, Aggarwal A, Rohit MK, Sahni D. Variance in coronary venous anatomy: a critical determinant in optimal candidate selection for cardiac resynchronization therapy. Pacing Clin Electrophysiol. 2013;36(1):94–102. https://doi.org/10.1111/pace.12026.
• Anselme F, Albatat M, Marquie C, Leclercq C, Ritter P, Ollivier JF, et al. Exploring a new systematic route for left ventricular pacing in cardiac resynchronization therapy. Circ J. 2021;85(3):283–90. https://doi.org/10.1253/circj.CJ-20-0266. Gives a perspective on LV micro-lead implantation as a novel alternative to overcome some of the implantation problems of conventional LV leads.
Worley S, Ellenbogen KA. Application of interventional procedures adapted for device implantation: new opportunities for device implanters. Pacing Clin Electrophysiol. 2007;30(8):938–41. https://doi.org/10.1111/j.1540-8159.2007.00789.x.
Funding
Open Access funding enabled and organized by Projekt DEAL.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
NA
Conflict of Interest
The authors declare no competing interests.
Human and Animal Rights and Informed Consent
This article does not contain any studies with human or animal subjects performed by any of the authors.
Additional information
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
This article is part of the Topical Collection on Devices
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Butter, C., Georgi, C. & Stockburger, M. Optimal CRT Implantation—Where and How To Place the Left-Ventricular Lead?. Curr Heart Fail Rep 18, 329–344 (2021). https://doi.org/10.1007/s11897-021-00528-9
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11897-021-00528-9