Abstract
In recent years, there has been an increased interest in continuous monitoring of patients and their Implanted Medical Devices (IMDs) with different wireless technologies such as ultrasounds. This paper demonstrates a high data-rate intrabody communication link based on Lithium Niobate (LN) Piezoelectric Micromachined Ultrasonic Transducers (pMUTs). The properties of the LN allow to activate multiple flexural mode of vibration with only top electrodes. When operating in materials like the human tissue, these modes are merging and forming a large communication bandwidth. Such large bandwidth, up to 400 kHz, allows for a high-data rate communication link for IMDs. Here we demonstrate a full communication link in a tissue phantom with a fabricated LN pMUT array of 225 elements with an area of just 3 by 3 mm square, showing data-rates up to 800 kbits/s, starting from 3.5 cm and going up to 13.5 cm, which covers the vast majority of IMDs.
Similar content being viewed by others
Introduction
The Internet of Medical Things (IoMT) has been the subject of wide research aimed at creating networks of medical devices that continuously acquire medical data from patients and make it available to both the user and healthcare providers1. Wearable devices such as smartwatches and wrist bands2 are examples of commercially available devices that provide a diverse collection of health information (heart rate, exercise, sleep patterns, etc.) and can easily connect to the internet3. Implantable Medical Devices (IMDs)4 are a second category of IoMT devices than can provide data that is more specific to certain organs and parts of the body given their closer proximity to the site of interest due to implantation. These devices also offer site-specific stimulation capabilities for treatment purposes. However, the transmission of this data from within the body to the external world is challenging. The goal of this work is to bridge the communication technology gap between the IMDs and the final beneficiaries of the data—the patients and the doctors. This is achieved by creating innovative technologies for intrabody communication links5,6,7 that allow wireless communication between implanted and non-implanted devices, and at the same time keep the patient safe by complying with the Food and Drug Administration (FDA) regulations8.
Ultrasonic technology has emerged as a viable candidate for implementing intrabody communication, given the wide variety of implantation depths IMDs require9. The reason being that ultrasonic waves are mechanical in nature and therefore propagate with lower attenuation in human tissue at the frequencies of interest, around 1 MHz. The first demonstration of intrabody communication utilizing ultrasound technology was performed with bulk piezoelectric transducers based on Lead Zirconate Titanate (PZT)10. Good performance in terms of data rate and penetration of the acoustic waves into the body was shown. However, PZT transducers contain lead which make these devices non-bio-compatible11 and requires extra packaging for implanting as part of an IMD. Moreover, their bulky form factor is a further disadvantage in terms of invasiveness for the patients, which also introduces a high risk of infection and rejection from the body. Thus, recent research has been focused on Micromachined Electro-Mechanical Systems (MEMS), such as Capacitive Micromachined Ultrasonic Transducers (cMUTs)12 (working principle in Methods) and Piezoelectric Micromachined Ultrasonic Transducers (pMUTs)13 (working principle in Methods) as shown in Fig. 1. For both devices, the key advantage is their miniaturization capability through standardized semiconductor micro-fabrication processes. The main advantage of pMUTs over cMUTs is that they don’t require a DC bias to be actuated, therefore making them more appealing for CMOS integration (less complex design and lower power consumption). On the other hand, the key advantage of the pMUTs over bulk transducers is the biocompatibility of the main piezoelectric material they use, the Aluminum Nitride (AlN)14 as opposed to the PZT11 used in the bulk transducers. This allows for the implantation of pMUTs alongside IMDs without the need of additional packaging. A second advantage is the possibility to integrate the pMUT fabrication directly in CMOS foundries, lowering the final product cost and shortening the time to market. Nonetheless, the AlN has a smaller electro-mechanical coupling coefficient compared to other piezoelectric materials such as PZT, meaning that it generates less output pressure when applying the same AC voltage. Recent work has replaced the AlN with micromachined PZT showing better performance but losing the advantage of biocompatibility15. The latest promising option is to dope the AlN with Scandium (ScAlN)16, which allows to increase the electromechanical coupling factor. Preliminary results have shown similar performances to that of PZT while maintaining the biocompatibility of the pMUT.
A key challenge for intrabody communication is to increase the data rate. In fact, operating at ultrasound frequencies such as 1 MHz avoids signal attenuation17, but in return, it limits the data rate. This is opposed to what can be achieved with RF antennas operating at much higher frequency but having a dramatic increase of attenuation in tissue as shown in Fig. 2c. In water, the EM waves attenuate with a rate of 0.2 dB/cm while the US waves attenuate at a rate of 0.002 dB/cm; even though the US waves attenuate much less, it only introduces a difference of 3 dB at 15 cm, which is beyond the maximum range of our experiment. However, when considering the tissue phantom, the EM waves attenuate at a rate of 4 dB/cm while the US waves attenuate only at a rate of 0.7 dB/cm; this introduces a difference of 50 dB at 15 cm when operating in tissue phantom, which is closer to the human body in terms of attenuation and acoustic impedance, hence making the US waves a better fit for intrabody communication18,19,20 such as in our application with the LN pMUTs. The main concept, to increase the data-rate, is to design devices with a large operation bandwidth. PMUTs based on AlN and ScAlN already present a wider bandwidth compared to bulk PZT transducers, cMUTs, or even PZT-based pMUTs3,15,16.
In this work, we show a better alternative by designing arrays of pMUTs that include elements centered around different frequencies. These frequencies are closely spaced to cover a certain desired bandwidth. Once the pMUT array is implanted or submerged in a liquid medium, the different resonance frequencies will merge together, due to the damping effect, and achieve a large bandwidth. However, this poses challenges at the fabrication level to precisely tune the resonance frequency of each individual pMUT. In order to ultimately increase the communication bandwidth, the proposal of this work is to introduce a different piezoelectric material, the X-cut Lithium Niobate (LN), which allows to harness multiple resonance frequencies in one device which can therefore merge into a larger bandwidth when operating in a body-like medium. At the same time, by harnessing stronger piezoelectric coefficients21 compared to AlN and ScAlN, the LN pMUTs are able to cover a broad range of implantation depths and result in higher data rates for the communication schemes.
Results
Experimental setup
In this work, we are demonstrating an X-cut LN-based pMUTs. Implemented at the system level, this allows the creation of wide band and high-data-rate intrabody communication links with a high implantation depth range. In Fig. 1 we are showing the design of a single LN pMUT (a), the design of an entire array of pMUT (b), the Scanning Electron Microscope (SEM) image of a fabricated device (c), and ultimately an optical image of a fabricated array (d). For more details, the step-by-step fabrication process is described in Methods. In order to determine the remarkable properties of the LN pMUTs, we are characterizing these in air and in tissue-like media. The characterization in air consists of measuring the 3D membrane displacement of the pMUT’s membrane with a Digital Holographic Microscope (DHM). The measurements match the resonant modes predicted by a Finite Element Simulation (FEM) as shown in Fig. 2a. More details are found in Methods. These results show that, with only two top electrode activation, we can excite multiple resonance modes around the main peak and that these are very distinctively separated for in-air operation. Following, the characterization in tissue-like media (de-ionized water, silicone oil, castor oil, or tissue phantom), consists of transmitting a high intensity and short duration signal with a reference hydrophone and then receiving this ultrasonic signal at a certain distance with an LN pMUT array. More details about the characterization setup are included in Methods. This allows to recover the intensity of the received signal and to determine the bandwidth and maximum achievable distance. In Fig. 2b, we are showing the extracted bandwidth from the sound pressure level (SPL) received by the pMUT array. Finally, the communication setup through a body tissue phantom, implemented with the aid of two Universal Software Radio Peripheral (USRPs), is shown in Fig. 3. The communication results, representing the quality of the channel, are shown in Fig. 4 for different communication distances or depths. More details about the realized communication scheme in MATLAB Simulink and synthesized with the USRPs are shown in Methods.
Lithium Niobate pMUTs
The LN pMUTs are innovative since they harness, for the main resonance mode, a different piezoelectric coefficient compared to traditional pMUTs. The PZT, AlN, and ScAlN-based pMUTs consist of a piezoelectric thin-film sandwiched between a top and a bottom electrode in order to activate the d31 piezoelectric coefficient22. The vertical electric field excites an in-plane displacement. Instead, in the case of the LN, we are taking advantage of a combination of piezoelectric coefficients (i.e., d31 and d11) that strongly couple the electric field to displacement21. These coefficients can be activated by electrodes that are on the same plane, thus the need of only one metal layer. The full fabrication process is shown in Methods. In Fig. 1a we are showing the design elements of a single pMUT. This consists of the two activation electrodes, signal, and ground, of width Wel, and the releasing cavity that defines the vibrating membrane, of length Lcav and width Wcav. More details concerning the array of pMUTs are shown in Methods. Additionally, in Fig. 1b we are showing the design of a 15 × 15 pMUT array. In Fig. 1c we are showing a Scanning Electron Microscope (SEM) image of a fabricated LN pMUT (unreleased). Here we can see the different layers such as the Si substrate, SiO2 support layer, the Pt bottom electrode layer (not used for this device), the LN thin-film piezoelectric layer, and finally the top electrode Al (in perspective). Lastly, in Fig. 1d we are showing the optical image of the fabricated array. Here we can distinguish all the individual pMUTs and their released cavities (different shades of green), the routing traces connecting all the elements in parallel to combine the received and transmitted pressure, and the gold pads that are needed to wire-bond the array to a Printed Circuit Board (PCB). The whole array fits into a 3 × 3 mm2 area, making it ideal as a miniaturized ultrasound antenna to implant into the human body alongside IMDs.
The displacement of the pMUT membrane is measured over the frequency range with a Digital Holographic Microscope (DHM) as shown in Fig. 2a (more details in Methods). The DHM creates a full 3D reconstruction of the membrane which allows for precise measurement, both in terms of peak displacement and vibration mode. The figure shows the maximum membrane displacement of several pMUT samples and their mean value profile curve (“black-bold” curve). At this point, from the same curve, we can determine the main resonance frequency to be around fres = 660 kHz and the maximum displacement dmax = 240 nm when applying an input voltage of Vin = 3 V, resulting in a displacement sensitivity of Sdisp = 80 nm/V. Moreover, each major resonance peak is matched with a Finite Element Analysis (FEA) simulation in COMSOL Multiphysics (more details in Methods). The additional resonances are due to the strong anisotropic properties of the piezoelectric coefficient’s matrix of the LN, which in this case play a key role in achieving a large operation bandwidth. This effect happens because the electric field generated by the top electrodes excites multiple modes through different piezoelectric coefficients21, which are strong in multiple directions, as opposed to other materials that have strong coefficients only in one direction (AlN, ScAlN, and PZT). At this point, when submerging the pMUTs in a damping media such as tissue phantom or silicon oil, several adjacent resonance modes will merge together23,24, resulting in an overall increased large bandwidth for the LN pMUTs. The details of the mode-merging modeling are described in Methods.
A time-domain measurement of the membrane displacement is performed with a Laser Doppler Vibrometer (LDV), as shown in Methods. The resonance frequency obtained from the LDV is fres ≈ 699 kHz and the peak displacement is dmax ≈ 88 nm for an input signal of Vin = 1 V, resulting in a displacement sensitivity Sdisp ≈ 88 nm/V (confirming the 3D DHM measurement). While, on one hand, this technique allows to measure the displacement in only one point in the pMUT membrane, it also allows for a time-domain characterization. The time-domain approach allows to drive the pMUT membrane with several sine wave cycles (N = 800) and measure the ring-up and the ring-down of the membrane displacement, which is a function of the resonance quality factor25, which results to be Q ≈ 381. At this point, an array of 15 × 15 LN pMUTs is wire-bonded to a Printed Circuit Board (PCB) and covered in a thin Parylene layer. This layer has a dual function, one is protecting the wire bonds and the second is to act as a matching layer between the pMUT acoustic impedance and the operating medium impedance. At this point the LN array is submerged in a de-ionized (DI) water tank, to closely match the acoustic impedance of the human tissue and tested for its ultrasound transmission capabilities (more details in Methods). Finally, the Fast Fourier Transform (FFT) can be applied to the received pulse on this measurement, to extract the frequency domain response shown in Fig. 2b. From the graph, we can extract the operation bandwidth of the pMUT at –6 dB, which corresponds to a large bandwidth BW−6 dB ≈ 401 kHz at a center frequency of fc ≈ 630 kHz. It is important to notice that the resonance frequency was lowered in the DI water due to the damping effect and at the same time, the modes shown in Fig. 2a are merged inside one large band. Similarly, we can extract the quality factor in water which corresponds to Q ≈ 2, which is two orders of magnitude lower than in air, indicating good ultrasound radiation in water or tissue-like media.
Communication link
Once the LN pMUT arrays have been characterized for ultrasound transmission, we are ready to set up an entire communication link that emulates as closely as possible an intrabody communication scenario as shown in Fig. 3. The intrabody setup consists mainly of a body tissue phantom from 3B Scientific26 that mimics the human body properties, such as propagation speed of the ultrasound waves, an average body density, and its corresponding acoustic impedance. An incision is performed on the phantom in order to implant the pMUT array receiver. At this point, in order to avoid air gaps around the implanted array, we apply an ultrasound gel27, similar to the one used for ultrasound imaging, which gives continuity to the ultrasound transmission. Furthermore, the communication link is implemented with the use of Universal Software Radio Peripherals (USRP)28. This allows us to implement a communication scheme of choice in MATLAB Simulink (more details in Methods) and test it in real-time in the above described intrabody scenario. In a real-life scenario, the entire electronics can be implemented in CMOS circuitry or miniaturized Field Programmable Gate Arrays (FPGA). While other groups have demonstrated miniaturized electronics for intrabody communication29,30, this goes beyond the purpose of this paper. The main focus of this work is to demonstrate the high performance of the LN pMUT arrays, in an intrabody scenario, in terms of large bandwidth and long intrabody communication range. In order to take full advantage of the large 400 kHz bandwidth provided by the LN pMUTs, we choose a Quadrature Phase Shift Keying (QPSK) modulation scheme as the communication protocol. Here we are choosing to transmit the row pixels of an image over the ultrasound link and then the information is serialized into a bitstream. More details about the encoding and the QPSK modulation are included in the Methods. At this point, the bitstream can be fed directly into a QPSK modulator object provided in MATLAB Simulink which will interface with the USRP Software Defined Radio (SDR) transmitter. The SDR transmitter is in charge of up converting the modulated data from baseband (DC center frequency) up to the RF frequency corresponding to the central frequency of the pMUT array transmitter fc ≈ 630 kHz. Finally, once the data payload is ready, we can proceed with a decoding MATLAB script (the reverse procedure used to encode) and re-assemble the received data into an image. At this point, we are able to compare the received image with the originally transmitted one in terms of BER at several communication distances or implanted device penetration depths, which will degrade with lower SNR at longer distances. By choosing an image as transmitted data we can visually interpret on-the-fly the quality of the ultrasonic channel in terms of lost pixels (“black”) or degraded pixels (“un-real colors”).
Once the communication setup is ready to transmit and receive, testing can be performed. The main test consists of characterizing the quality of the ultrasonic channel in the tissue phantom at different distances, starting from a minimum of Dmin = 3.5 cm and a maximum distance in the phantom of Dmax = 13.5 cm. The results are shown in Fig. 4. The Signal-to-Noise Ratio (SNR) obtained for such communication depth range has a maximum of SNRmax = 9.5 dB and reaches a minimum of SNRmin = 1.5 dB as shown in the power spectrum of the QPSK modulation. Under these SNR conditions, we are computing the amount of pixel errors when receiving the encoded image, a metric which is commonly known as BER. In order to have a direct interpretation of the BER and thus of the quality of the ultrasonic channel based on the LN pMUT arrays, in Fig. 4 we are showing the decoded image for each distance. Here we can appreciate how some pixels got “lost” or have the wrong information resulting in abrupt color changes compared to other adjacent pixels. On one hand, the BER had a minimum of BERmin = 3 × 10−5 where real-time data transmission can be directly implemented on the communication channel without additional coding or electronics31. On the other hand, the BER has a maximum of BERmax = 5 × 10−2 which is the limit for being able to correctly reconstruct the original information by implementing bit error detection and correction algorithms32. The BERmax sets the limit to the implantation depth of the IMD in order to maintain reliable ultrasonic data transmission. Additionally, the experimental BER vs SNR is plotted in the initial theoretical curve for a QPSK link in Fig. 2d, showing the logarithmic decrease of the communication errors while the signal level decreases at longer implantation depths. These results are promising toward allowing a broad implantation range Drange = 3.5–13.5 cm which enables the implantation of all three categories of IMD discussed in the introduction and at the same time offers a large communication bandwidth of BW = 400 kHz. Ultimately, this visual interpretation of the ultrasonic channel quality can be useful for applications such as scanning for multiple IMDs in order to find an optimal location for the external transmitter33.
In conclusion, given the results in terms of large bandwidth and deep implantation range, we believe that the LN-based pMUT technology has the potential to significantly improve the wireless communication links for implanted medical devices for real-time monitoring. The LN piezoelectric thin film shows promising insights on how to achieve a large band thanks to the combination of spurious modes under the damping of the acoustic medium such as water and tissue. This leaves open the research to explore different actuation designs of the pMUT with the X-cut LN layer, different device orientations, as well as exploring different crystal cuts, such as Y- and Z- cuts.
Methods
Micromachined Ultrasonic Transducers (MUT) categories and fabrication of LN pMUTs
The first category of transducers is the capacitive MUTs (cMUTs), which generally consist of a suspended membrane on top of a silicon substrate that is actuated (i.e., put in vibration mode) by an electric field as shown in Fig. 5a. This electric field is generated by applying a high DC bias to the membrane (it helps make the membrane more compliant or softer) and an AC driving voltage signal that sets the membrane into motion at its resonance frequency. When in motion, the membrane will displace the particles in the surrounding medium and generate ultrasonic waves that can propagate in the far field of the device and be used as carrier for an ultrasonic communication link.
The second category is the piezoelectric MUTs (pMUTs), which consist of a membrane that moves at its resonance frequency, generating ultrasonic waves as shown in Fig. 5b. The simplest pMUT structure consists of a thin-film piezoelectric membrane, for example, made of Aluminum Nitride (AlN), that is sandwiched between two electrodes (i.e., top and bottom electrodes). When applying a driving AC voltage signal to the electrodes, the AlN will start expanding and contracting in the lateral dimension due to the inverse piezoelectric effect. Since the membrane is clamped and suspended on top of a cavity, this will act as a boundary condition that will force the membrane to vibrate in the Z-axis during its expansion and contraction in the lateral dimension.
In this work, we are interested in LN-based pMUTs, and their step-by-step fabrication process shown in Fig. 5c. The first part of the process starts with a thin film X-cut LN wafer of 300 μm on which a Pt layer of 200 nm is sputtered (step 1). Following, a SiO2 layer of 1 μm is added through chemical vapor deposition (step 2). The presence of this layer shifts the neutral bending axis of the pMUT device to the middle of the piezoelectric layer, which maximizes the membrane displacement. So far, the LN acts as the handling wafer for the Pt and the SiO2 layers, but the LN needs to act as a device layer instead. Thus, it needs to be transferred to another handling wafer. In order to achieve this passage, the LN wafer is flip bonded to a Double Side Polished (DSP) Si wafer of 300 μm thickness, through a surface-activated bonding technique (step 3). In the same step, the LN wafer is reduced to a device thickness of 1 μm through a Chemical and Mechanical Polishing (CMP) process. Now the LN is a device layer, and the Si wafer acts as a handling wafer. In the second part of the fabrication process, patterning masks are used to lithographically implement the design of the desired pMUTs, including both single-elements and array layouts. In particular, there is an electrode definition mask and a cavity releasing mask. The first one will define the 150 nm Al top electrode that is electron-beam sputtered and shaped through a lift-off process (step 4). Finally, the second mask defines the pMUT cavities during the releasing process (step 5). This step consists of a Deep Reactive Ion Etching (DRIE) process to etch straight trenches (i.e., cavities) from the back of the Si wafer and stop on the SiO2 layer, thus releasing the pMUT membrane.
COMSOL simulation and air measurements of fabricated devices
To simulate the pMUTs, COMSOL Multiphysics is used to generate a 3D model of a single LN pMUT and to run a Finite Element Analysis (FEA) to find its behavior in the frequency domain. Besides providing several physics domains for the models, the advantage of using COMSOL is that is also provides coupling modules between these different domains. In fact, to simulate a pMUT element we make use of two multiphysics modules: the piezoelectric module, which couples the electrical domain with the mechanical one, and the ultrasonic module, which couples the mechanical domain with the sound-pressure domain in different materials. This simulation tool allows us to set up the application medium such as air and tissue-like media (oil, water, tissue phantom, etc.) and select a particular thin-film layer as the piezoelectric layer. This allows to select a LN cut (i.e., crystal orientation and piezoelectric coefficient matrix), in our case the X-cut, and the angle of orientation at which the electric field is applied. In Fig. 6a we are showing the sound pressure level (SPL) generated by the pMUT while sweeping the operation frequency. Here we are detecting a resonance frequency of 657 kHz in the air where the SPL is maximum. Finally, the three inserts show the geometry mashing, the main mode of vibration, and the membrane displacement over frequency.
Once the devices are fabricated, these are characterized in the air with a Digital Holographic Microscope (DHM) to measure the full 3D displacement of the pMUT’s membrane as shown in the reconstruction in Fig. 6b and detect the mode shape. Here we measured a main resonance of fair ≈ 669 kHz (close match to simulation) and an average peak displacement sensitivity of Sdisp ≈ 93 nm/V. Following, the peak displacement is measured with a Laser Doppler Vibrometer (LDV) while driving the pMUT with N = 800 sine wave cycles at the resonance frequency in air. This allows to extract its quality factor Q with the ring-down technique. This consists of counting how many cycles it takes to halve the displacement amplitude once the driving signal is turned off, and then multiplying this number by 4.53 to find the quality factor Q. In our case, the ring-down decay takes Ndecay = 84, as shown in Fig. 6c, thus Qair ≈ 381.
Modes merging in air vs water/tissue
The LN pMUTs present interesting piezoelectric properties for which multiple resonance modes can be activated around the main resonance frequency. With a frequency sweep on the DHM, we can detect the additional resonance modes based on the shape of the 3D displacement of the membrane as shown in Fig. 6b. The modes are then extracted from the measurement and fitted to a Butterworth–Van Dyke (BVD) model based on their frequency, quality factor, coupling factor, and peak displacement, as shown in Fig. 7a. Since these measurements are performed in air the quality factor of each resonance is high, making them very distinctive from one another. This is due to the high acoustic impedance mismatch to the air which prevents the ultrasonic radiation and keeps most of the energy on displacing the pMUT’s membrane. Similarly, the air has also a low density, thus the damping of the displacement is low as well. Following, Fig. 7b shows the simulated combined SPL at the output of the LN pMUT surface based on the above measurements. As it can be noticed, the peaks do not merge together but rather form multiple fractional bands, because of the high-quality factor of each peak in air.
Finally, we model what happens to the peak displacements of all the resonance modes and to the combined output SPL of the LN pMUT when exposed to a denser external load, such as water or a tissue phantom. First, in Fig. 7c the displacement attenuation of each resonance peak and the lowering of their corresponding quality factor can be noticed. This is due to the more efficient radiation of the energy from the pMUT surface into the medium by generating ultrasonic waves. Moreover, given the high density of the new medium, this also has a damping effect on the displacement amplitude of each resonance peak, allowing them to merge into a single output pressure band as shown in Fig. 7d. In this case, the resulting large bandwidth of BW−6 dB ≈ 400 kHz is a very close match to the measured results in Fig. 2b with a reference hydrophone at a radiation distance of 3.3 cm.
Ultrasonic measurement setup, bandwidth extraction, and comparison
A 15×15 LN pMUT array is coated with Polydimethylsiloxane (PDMS) and submerged in a de-ionized water tank and tested for ultrasonic transmission as shown in Fig. 8a. This shows a record bandwidth of BW−6 dB ≈ 401 kHz and a center frequency fwater ≈ 630 kHz. In addition, Qwater ≈ 2 is lower than in Qair, implying that most of the input energy is radiated into the medium, making the LN pMUT ideal for underwater and intrabody communication. In Fig. 8b we are showing the driving signal (“blue” curve) of a hydrophone and the received pulse by an LN pMUT array (“orange” curve). The hydrophone is excited with a high-intensity pulse to emulate a Dirac pulse, VDirac = 9 V and fDirac = 5 MHz, which allows to extract the step response of the system. It is interesting to notice that the received signal is delayed by 21 μs at a distance of 3.3 cm, which allows us to estimate the sound velocity in the water to be c ≈ 1450 m/s. Finally, the received pulse has an intensity of VRX ≈ 4.2 mV.
At this point, we can compare pMUTs based on other materials, such as AlN as shown in Fig. 8c and ScAlN doped at 36% as shown in Fig. 8d, and then compare it to the bandwidth obtain in this work for LN pMUTs as shown in Fig. 8e. The impulse response is measured in a water tank as in Fig. 8b and FFT is applied for bandwidth extraction at −6 dB from the peak as shown in Fig. 2b. The LN-based pMUTs of this work show the highest bandwidth of 401 kHz, while the 36% Sc-doped AlN shows large bandwidth of 375 kHz as well, which makes it a good material to explore. However, a disadvantage of the Sc is the reproductivity of the sputtering process, while the advantage of the LN is that during the bonding process to the substrate it maintains the same crystal structure and piezoelectric properties, hence making it more reliable and predictable.
Data serialization into a bitstream and modulation scheme implementation
First, a raw image of 100 × 100 pixels is serialized in MATLAB to create a bitstream for the communication scheme as shown in Fig. 9a. Each pixel consists of an RGB vector of three integers (0–255) that can be converted into an 8-bit string, for a total of 24 bits for the vector. At this point, all the pixels are concatenated in a bitstream resulting in a total raw data of DataRAW = 240 kbits. Then, the bitstream is encoded with a Quadrature Phase Shift Keying (QPSK) modulation, which allows to encode 2 bits per second as shown in Fig. 9b. The modulation is done asynchronously eliminating the need for a clock. On the other hand, there is the need to add an overhead to the raw data in order for the receiver to detect it. This increase in data length is of ~10%, resulting a final DataQPSK ≈ 264 kbits. Finally, the QPSK data is up-converted by the USRP at the operation frequency of the pMUT array and transmitted through a tissue phantom that mimics the human tissue properties as shown in Fig. 9c. The pMUT array bandwidth BW ≈ 400 kHz, translating in a data rate ≈ 800 kbits/s, meaning that the time to transmit one image or frame is TTX ≈ 0.33 s. On the receiving side of the intrabody ultrasonic transmission link, the signal is down-converted to baseband by another USRP and sampled at twice the bandwidth for perfect reconstruction (Nyquist theorem). The signal requires constant frequency and frame synchronization as discussed in the paper. At this point, the data bitstream is demodulated from the QPSK scheme and re-assembled in an RGB pixels matrix.
The generated bitstream can be fed directly into a QPSK modulator object provided in MATLAB Simulink which will interface with the USRP Software Defined Radio (SDR) transmitter. The SDR transmitter is in charge of up converting the modulated data from baseband (DC center frequency) up to the RF frequency corresponding to the central frequency of the pMUT array transmitter fc = 630 kHz, as shown in Fig. 9d. The encoded bitstream is received by a pMUT array receiver with the same central frequency as shown in Fig. 9e. At this point, since the receiver and the transmitter don’t share a common clock, the major issue is synchronizing the SDRs at the two ends of the communication link. For this reason, we are using three standard synchronization blocks for the QPSK modulation, such as frequency offset compensation, symbol, and carrier synchronization. This will allow to maintain a stable ultrasound link and even stream the data in real-time.
Time-domain comparison to avoid capacitive coupling
When in water, we have that the driving signal couples directly into the received signal through a capacitive coupling effect, while the received ultrasonic pulse is received with a delay of 20 μs, which indicated a distance of around 3 cm as shown in Fig. 10a. Then the ultrasonic path is blocked, and the capacitive coupling is still present in water as shown in Fig. 10b. At this point, when repeating the same experiments in a tissue phantom, the capacitive coupling is not present anymore as shown in Fig. 10c. In this plot, we can see the ultrasonic pulse delayed by 20 μs at around 3 cm without any of the driving signal leaking into the received signal. Additionally, in Fig. 10d we are blocking once again the direct path in between the TX and RX and the ultrasonic pulse is not present anymore. This shows us that by using a tissue phantom in the experimental setup we can avoid capacitive coupling in the ultrasonic pulse and hence implement a reliable communication link.
Data availability
Data are available upon request.
References
Chen, M., Gonzalez, S., Vasilakos, A., Cao, H. & Leung, V. C. Body area networks: a survey. Mob. Netw. Appl. 16, 171–193 (2011).
Salem, E. Y., Zain, M. Y. & Alfons, M. Smart wearable health device for heart rate and temperature measurements. Int. J. Technol. Diffus 11, 17–28 (2020).
Demirors, E., Alba, G., Santagati, G. E. & Melodia, T. High data rate ultrasonic communications for wireless intra-body networks. in 2016 IEEE International Symposium on Local and Metropolitan Area Networks (LANMAN) 1–6 (2016).
Kiourti, A. & Nikita, K. S. A review of in-body biotelemetry devices: Implantables, ingestibles, and injectables. IEEE Trans. Biomed. Eng. 64, 1422–1430 (2017).
Herrera, B., Pop, F., Cassella, C. & Rinaldi, M. AlN PMUT-based ultrasonic power transfer links for implantable electronics. In 2019 20th International Conference on Solid-State Sensors, Actuators and Microsystems and Eurosensors XXXIII, TRANSDUCERS 2019 and EUROSENSORS XXXIII 861–864 (IEEE, Germany, 2019).
Pop, F. et al. Zero-power acoustic wake-up receiver based on DMUT transmitter, PMUTS arrays receivers and MEMS switches for intrabody links. In 2019 20th International Conference on Solid-State Sensors, Actuators and Microsystems and Eurosensors XXXIII, TRANSDUCERS 2019 and EUROSENSORS XXXIII (IEEE, Germany, 2019).
Pop, F. V. et al. Novel pMUT-based acoustic duplexer for underwater and intrabody Communication. IEEE Int. Ultrason. Sympos. https://doi.org/10.1109/ULTSYM.2018.8579709 (2018).
Food, Administration, D. et al. Guidance for Industry and FDA staff Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers (MD FDA, 2008).
Pop, F., Herrera, B., Cassella, C. & Rinaldi, M. Enabling real-time monitoring of intrabody networks through the acoustic discovery architecture. IEEE Trans. Ultrason. Ferroelectr. Freq. Control. https://doi.org/10.1109/TUFFC.2020.3002973 (2020).
Santagati, G. E. & Melodia, T. Experimental evaluation of impulsive ultrasonic intra-body communications for implantable biomedical devices. IEEE Trans. Mob. Comput. 16, 367–380 (2017).
Moulin, E., Assaad, J., Delebarre, C., Kaczmarek, H. & Balageas, D. Piezoelectric transducer embedded in a composite plate: application to Lamb wave generation. J. Appl. Phys. 82, 2049–2055 (1997).
Nan, H. et al. Non-contact thermoacoustic detection of embedded targets using airborne-capacitive micromachined ultrasonic transducers. Appl. Phys. Lett. 106, 84101 (2015).
Guedes, A. et al. Aluminum nitride pMUT based on a flexurally-suspended membrane. In 2011 16th International Solid-State Sensors, Actuators and Microsystems Conference 2062–2065 (IEEE, China, 2011).
Berg, N. G., Paskova, T. & Ivanisevic, A. Tuning the biocompatibility of aluminum nitride. Mater. Lett. 189, 1–4 (2017).
Kusano, Y. et al. Effects of DC bias tuning on air-coupled PZT piezoelectric micromachined ultrasonic transducers. J. Microelectromech. Syst. 27, 296–304 (2018).
Kusano, Y., Luo, G.-L., Horsley, D., Taru Ishii, I. & Teshigahara, A. 36% Scandium-doped Aluminum Nitride Piezoelectric Micromachined Ultrasonic Transducers. 2018 IEEE Int. Ultrason. Symp. https://doi.org/10.1109/ULTSYM.2018.8579694 (2018).
Nam, K. et al. Comparison of ultrasound attenuation and backscatter estimates in layered tissue-mimicking phantoms among three clinical scanners. Ultrason. Imaging 34, 209–221 (2012).
Kaushal, H. & Kaddoum, G. Underwater optical wireless communication. IEEE Access 4, 1518–1547 (2016).
Lacovara, P. High-bandwidth underwater communications. Mar. Technol. Soc. J. 42, 93–102 (2008).
Cho, J.-H. & Hyo, S. Communication strategies for various types of swallowable telemetry capsules. Mod. Telemet. https://doi.org/10.5772/23924 (2011).
Pop, F. V., Kochhar, A. S., Vidal-Álvarez, G. & Piazza, G. Laterally vibrating lithium niobate MEMS resonators with 30% electromechanical coupling coefficient. Proc. IEEE Int. Conf. Micro Electro Mech. https://doi.org/10.1109/MEMSYS.2017.7863571 (2017).
Wang, Q., Lu, Y., Mishin, S., Oshmyansky, Y. & Horsley, D. A. Design, fabrication, and characterization of scandium aluminum nitride-based piezoelectric micromachined ultrasonic transducers. J. Microelectromechanical Syst. 26, 1132–1139 (2017).
Sun, C. et al. Investigation of broadband characteristics of multi-frequency piezoelectric micromachined ultrasonic transducer (MF-pMUT). IEEE Sens. J. 19, 860–867 (2018).
Wang, T., Kobayashi, T. & Lee, C. Broadband piezoelectric micromachined ultrasonic transducer (pMUT) using mode-merged design. In 10th IEEE International Conference on Nano/Micro Engineered and Molecular Systems 238–242 (IEEE, Xi'an, 2015).
Rozen, O. et al. Monolithic MEMS-CMOS ultrasonic rangefinder based on dual-electrode PMUTs. Proc. IEEE Int. Conf. Micro Electro Mech. Syst. 2016, 115–118 (2016).
Abayazid, M., Vrooijink, G. J., Patil, S., Alterovitz, R. & Misra, S. Experimental evaluation of ultrasound-guided 3D needle steering in biological tissue. Int. J. Comput. Assist. Radiol. Surg. 9, 931–939 (2014).
Wang, X., Fowlkes, J. B., Cannata, J. M., Hu, C. & Carson, P. L. Photoacoustic imaging with a commercial ultrasound system and a custom probe. Ultrasound Med. Biol. 37, 484–492 (2011).
Sklivanitis, G. et al. Proc. of the NATO Symp. on “Cognitive Radio and Future Networks” (IST-123) (STO, 2014).
Charthad, J., Weber, M. J., Chang, T. C. & Arbabian, A. A mm-sized implantable medical device (IMD) with ultrasonic power transfer and a hybrid bi-directional data link. IEEE J. Solid-State Circuits 50, 1741–1753 (2015).
Chang, T. C., Wang, M. & Arbabian, A. Multi-access networking with wireless ultrasound-powered implants. IEEE Biomed. Circuits Syst. Conf. 2019, 1–4 (2019).
Herrera, B. et al. Pmut-based high data rate ultrasonic wireless communication link for intra-body networks. Hilton Head MEMS Workshop https://doi.org/10.31438/trf.hh2018.31 (2017).
Van Nee, R. D. J. OFDM codes for peak-to-average power reduction and error correction. in Proceedings of GLOBECOM’96. 1996 IEEE Glob. Telecommun. Conf. 1, 740–744 (1996).
Pop, F. V., Herrera, B., Cassella, C. & Rinaldi, M. IFCS/EFTF 2019—Joint Conference of the IEEE International Frequency Control Symposium and European Frequency and Time Forum, Proceedings (IEEE, 2019)
Acknowledgements
This work was supported by the National Science Foundation (NSF) through the following grants: 1618731, 1726512, and 1763964.
Author information
Authors and Affiliations
Contributions
F.P and M.R. conceived the idea; F.P. designed the device and the experimental setups and performed modeling and simulations; F.P and B.H. fabricated and tested the devices and performed the experiments and analyzed the data; M.R. initiated, coordinated, and supervised the research. F.P., B.H., and M. R. contributed to the preparation of the manuscript.
Corresponding authors
Ethics declarations
Competing interests
The authors declare no competing interests.
Peer review
Peer review information
Nature Communications thanks the anonymous reviewer(s) for their contribution to the peer review of this work. Peer reviewer reports are available.
Additional information
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Pop, F., Herrera, B. & Rinaldi, M. Lithium Niobate Piezoelectric Micromachined Ultrasonic Transducers for high data-rate intrabody communication. Nat Commun 13, 1782 (2022). https://doi.org/10.1038/s41467-022-29355-9
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41467-022-29355-9
- Springer Nature Limited
This article is cited by
-
An implantable piezoelectric ultrasound stimulator (ImPULS) for deep brain activation
Nature Communications (2024)
-
Surface and bulk acoustic wave resonators based on aluminum nitride for bandpass filters
AAPPS Bulletin (2024)
-
Thin-film PMUTs: a review of over 40 years of research
Microsystems & Nanoengineering (2023)
-
Soft ferroelectret ultrasound receiver for targeted peripheral neuromodulation
Nature Communications (2023)
-
The effects of three profiles on the mechanical properties and grain size of self-piercing riveting joints using ultrasonic welding
The International Journal of Advanced Manufacturing Technology (2023)