Abstract
1H Magnetic Resonance Spectroscopic imaging (SI) is a powerful tool capable of investigating metabolism in vivo from mul- tiple regions. However, SI techniques are time consuming, and are therefore difficult to implement clinically. By applying non-uniform sampling (NUS) and compressed sensing (CS) reconstruction, it is possible to accelerate these scans while re- taining key spectral information. One recently developed method that utilizes this type of acceleration is the five-dimensional echo planar J-resolved spectroscopic imaging (5D EP-JRESI) sequence, which is capable of obtaining two-dimensional (2D) spectra from three spatial dimensions. The prior-knowledge fitting (ProFit) algorithm is typically used to quantify 2D spectra in vivo, however the effects of NUS and CS reconstruction on the quantitation results are unknown. This study utilized a simulated brain phantom to investigate the errors introduced through the acceleration methods. Errors (normalized root mean square error >15%) were found between metabolite concentrations after twelve-fold acceleration for several low concentra- tion (<2 mM) metabolites. The Cramér Rao lower bound% (CRLB%) values, which are typically used for quality control, were not reflective of the increased quantitation error arising from acceleration. Finally, occipital white (OWM) and gray (OGM) human brain matter were quantified in vivo using the 5D EP-JRESI sequence with eight-fold acceleration.
Similar content being viewed by others
Introduction
One of the major goals of in vivo 1H magnetic resonance spectroscopy (MRS) is to quantify the concentration of metabolites in certain tissues and investigate how these concentrations relate to pathology. Single voxel approaches, such as the point-resolved spectroscopy (PRESS)1 sequence, provide a means to quantify several metabolites from a single location. PRESS acquires a one dimensional (1D) spectrum from a volume of interest (VOI), which can subsequently be fit using prior-knowledge 1D fitting algorithms2,3,4,5,6,7, to yield absolute and relative concentration values. Due to severe spectral overlap of certain metabolite resonances in 1D acquisition, techniques have been developed to spread these signals over a second spectral dimension8,9,10,11,12, One of these two-dimensional (2D) spectroscopy methods, called J-resolved spectroscopy (JPRESS)9, 13, 14, acquires a second spectral dimension by introducing a time increment (t 1) into the PRESS sequence. By acquiring spectra from several t 1 values, JPRESS allows for the detection of both chemical shift in the direct spectral dimension (F 2) and J-coupling in the indirect spectral dimension (F 1). Similar to 1D spectral quantitation, this 2D JPRESS spectrum can be fit using 2D fitting algorithms such as prior-knowledge fitting (ProFit)15, 16, to yield metabolite concentrations.
Unfortunately, the main drawback of the JPRESS method is the long acquisition time necessary to obtain the second temporal (spectral) dimension. This becomes an even greater issue when combining JPRESS with spectroscopic imaging (SI)17, 18, to obtain metabolic information from multiple spatial dimensions. Several methods have been developed to significantly reduce acquisition time for SI, one of which is the echo planar spectroscopic imaging (EPSI) sequence19, 20, EPSI utilizes an echo planar bipolar gradient to simultaneously encode one spatial and one spectral dimension per TR. Different types of k-space readout gradients have also been incorporated into SI in order to accelerate data acquisition21,22,23,24, Another method capable of reducing scan duration is non-uniform sampling (NUS) combined with compressed sensing (CS) reconstruction25,26,27, CS has been applied to the combined JPRESS and SI method, called J-resolved echo planar spectroscopic imaging (EP-JRESI), for both four dimensional (4D - 2 spatial, 2 spectral)28 and five dimensional (5D - 3 spatial, 2 spectral)29 acquisitions. In particular, the accelerated 5D EP-JRESI method is capable of up to sixteen-fold acceleration (16x), yielding (k x , k y , k z , t 2, t 1) data in a clinically feasible time. However, the effects of the NUS and iterative, nonlinear reconstruction process on prior-knowledge quantitation results remain unknown. This is especially important with regards to the Cramér Rao lower bounds (CRLBs)30, 31, since many investigators utilize these CRLB values for the purposes of analyzing metabolite concentrations. Thus, it is important to determine whether CRLB% values are reflective of the true uncertainty in metabolite concentrations when using accelerated methods.
Therefore, the goal of this study was to investigate the effects of NUS and CS reconstruction on ProFit quantitation of accelerated 5D EP-JRESI data. Synthetic 5D EP-JRESI brain phantom measurements were simulated, and several voxels from these data were then fit using ProFit. Metabolite concentration values, as well as CRLBs, were investigated at several acceleration factors: four-fold (4x), eight-fold (8x), twelve-fold (12x), and 16x acceleration. Additionally, in vitro phantom measurements were acquired and were retrospectively undersampled and reconstructed in order to assess the performance of ProFit quantitation. Finally, to demonstrate the application of this method, ProFit was used to quantify occipital white matter (OWM) and gray matter (OGM) voxels from in vivo healthy brain measurements obtained using 8x NUS.
Methods
5D EP-JRESI acquisition
The 5D EP-JRESI sequence utilizes both EPSI and NUS to obtain 5D acquisitions, (k x , k y , k z , t 2, t 1) = (16, 16, 8, 256, 64), in a clinically feasible scan time28, 29, A fully sampled 5D EP-JRESI acquisition with TR = 1.2 s takes approximately 164 minutes, making NUS a necessity for in vivo applications. The sequence was developed by incorporating two major modifications to the standard chemical shift imaging (CSI) method17: (1) an echo planar bipolar gradient for readout and (2) a time increment (t 1) inserted between the two 180° pulses. A water acquisition was acquired with only one t 1 point, which was used for eddy current corrections32. Water suppression was performed using three WET33 pulses for metabolite measurement. Since the echo planar readout gradient is used to acquire the (k x , t 2) dimensions simultaneously, the NUS scheme is applied along the remaining, incremented dimensions: (k y , k z , t 1). The 3D volume produced by (k y , k z , t 1) was sampled according to the following density function based on exponential sampling34:
where \({\mathscr{P}}({k}_{y},{k}_{z},{t}_{1})\) is the probability of sampling a point in the (k y , k z , t 1) volume, and k y , k z , and t 1 were determined by the desired spatial and indirect spectral points, (k y , k z , t 1) = (16, 8, 64): k y = [−8, −7, −6, …, 7], k z = [−4, −3, −2, …, 3], and t 1 = [0, 1, 2, …, 64]. This type of non-uniform sampling scheme ensures that points containing higher signal, such as central k-space and earlier t 1 points, are sampled adequately. Sampling masks were generated for each NUS factor (4x, 8x, 12x, and 16x) and were implemented as an option on the console.
Prior to reconstruction, binary filtering was applied, and the spectral range outside 1.2–4.3 ppm in the F 2 dimension was set to zero while the 1.2–4.3 ppm region was unchanged, as previously described29, 35, This was performed in order to decrease the dynamic range resulting from residual water/fat signals. In order to reconstruct the missing data points29, the following optimization problem was solved in MATLAB using the split Bregman algorithm36:
Here, u is the reconstructed data, R is the sampling mask, \( {\mathcal F} \) is the Fourier transform operator applied across the (k y , k z , t 1) dimensions, f is the acquired data, and σ 2 is an estimate of the noise variance. The goal of the optimization was to minimize the total variation (TV) of the reconstructed data, and was implemented on a coil-by-coil basis. Here, TV is applied on the three undersampled dimensions, (y, z, t 1). The reconstruction took approximately 30–45 minutes per coil on a workstation equipped with a 3.1 GHz Intel Xeon dual processor and 128 GB RAM. Finally, coil combination was performed after applying frequency drift and phase corrections as previously described29, 37. A general schematic for the accelerated 5D EP-JRESI acquisition method can be seen in Fig. 1.
ProFit quantitation
For this study, the new version of ProFit16 was used to fit the 2D J-resolved spectra obtained using the 5D EP-JRESI technique. ProFit uses non-linear least squares fitting to fit an actual spectrum using prior-knowledge spectra obtained either experimentally or through simulation. The degrees of freedom for fitting are increased from iteration to iteration, allowing for measurement of line-broadening factors, frequency shifts, concentrations, and other parameters for several metabolites. First, a basis set was created using GAMMA simulation38 and previously reported chemical shifts and J-coupling constants39 of 22 metabolites: creatine 3.0 (Cr3.0), creatine 3.9 (Cr3.9), n-acetyl aspartate (NAA), phosphorylcholine (PCh), choline (Ch), glycerylphosphoryl choline (GPC), aspartate (Asp), γ-aminobutyric acid (GABA), glucose (Glc), glutamine (Gln), glutamate (Glu), glutathione (GSH), lactate (Lac), myo-inositol (mI), n-acetyl aspartyl glutamate (NAAG), phosphoethanolamine (PE), taurine (Tau), threonine (Thr), scyllo-inositol (Scy), alanine (Ala), glycine (Gly), and ascorbic acid (Asc). Small modifications were made to the fitting algorithm: (1) only three of the four outer iterations were used, foregoing spline baseline modeling and (2) the inner iterations were limited to decrease fitting time. The second modification was accomplished by simply changing the maximum number of iterations that the lsqnonlin function in MATLAB used in the optimization options. In addition to concentration values, ProFit also calculates the CRLB values for each metabolite30, 31, from the Fisher matrix:
Eq. (3) uses the basis matrix, \( {\mathcal B} \), and an estimate of noise variance, σ 2, to yield the Fisher matrix (F). \( {\mathcal B} \) is a vectorization of all the 2D spectra in the basis set, and therefore each row in \( {\mathcal B} \) corresponds to the prior-knowledge of each metabolite in the (F 2,F 1) domain. Noise variance is estimated in ProFit from a spectral region far removed from the diagonal peaks. CRLB values for the metabolites are readily obtained from the following:
The diagonal elements from F −1, denoted as diag(), yield CRLB values for each individual metabolite. Relative CRLBs, CRLB%, can be obtained by dividing the CRLB by the corresponding metabolite concentration and multiplying by 100:
Typically, 20–30% is the acceptable CRLB% used in a variety of in vivo studies.
Virtual Phantoms
A three dimensional Shepp-Logan brain phantom was simulated to replicate the 5D EP-JRESI experiment with the following scan parameters: B 0 = 2.89 T (123.23 MHz), (k x , k y , k z , t 2, t 1) = (16, 16, 8, 256, 64), direct spectral bandwidth = 1190 Hz, indirect spectral bandwidth = 500 Hz, and TE = 30 ms. The two-dimensional spectra were simulated using GAMMA simulation38 with concentration values similar to previously reported in vivo values39. The virtual phantom, or vphantom, included NAA (10.5 mM), NAAG (1.7 mM), GABA (1.3 mM), Ala (0.27 mM), Asc (0.6 mM), Asp (3.4 mM), Ch (0.25 mM), PCh (0.8 mM), GPC (2 mM), Cr (8 mM), Glc (0.3 mM), Glu (11.3 mM), Gln (5.3 mM), GSH (2.3 mM), mI (6 mM), Lac (1.8 mM), PE (1.6 mM), Scy (0.6 mM), Tau (2.7 mM), and Thr (0.5 mM). The signal intensities of the spectra were modulated based on the spatial structure of the Shepp-Logan phantom.
After simulation, the vphantom was retrospectively undersampled using the NUS schemes for 4x, 8x, 12x, and 16x acceleration. These data were subsequently reconstructed to produce accelerated 5D EP-JRESI data, and seventy-six voxels from each vphantom measurement were quantified using ProFit. Metabolite means, standard deviations, CRLB% values, and normalized root mean square error (nRMSE) values were investigated using these 76 voxels. The nRMSE was calculated using the fully sampled vphantom measurement as the true value:
where c f,n is the ProFit result from the fully sampled vphantom measurement, c us,n is the ProFit result from the undersampled vphantom measurement, n is the number of voxels (n = 76), and \({\bar{c}}_{f}\) is the average ProFit result from the fully sampled vphantom. Considering typical in vivo metabolite standard deviations, an error of 15% was chosen as the cutoff for acceptable quality. In order to examine the spatial effects of the reconstruction, metabolite maps were produced using peak integration for NAA (1.9–2.1 ppm) and mI (3.4–3.6 ppm). Eq. (6) was applied voxel by voxel to calculate nRMSE for both the NAA and mI spatial maps.
Additionally, a second vphantom was simulated in order to investigate the effects of NUS and reconstruction on the spatial signal leakage more carefully. The second vphantom was simulated with the same metabolite concentrations described above, however signal was only placed in a 3 × 3 voxel region in the central slice (slice 5). The second vphantom underwent the same undersampling and reconstruction process that was described for the first vphantom. The spatial leakage from the 3 × 3 region was assessed both qualitatively and quantitatively by measuring the NAA and Asp signals across the (y,z) dimensions.
In Vitro Phantom
Nine gray matter brain phantom measurements were obtained using the fully sampled 5D EP-JRESI sequence on a Siemens 3 T Trio (Siemens Healthcare, Erlangen, Germany) scanner. The phantom contained the following metabolites: NAA(8.9 mM), NAAG(0.51 mM), GABA(0.7 mM), Asp(2.1 mM), Ch(0.9 mM), PCh(0.6 mM), Cr(7 mM), Glc(1 mM), Glu(12.5 mM), Gln(2.5 mM), GSH (2 mM), mI(4.4 mM), Lac(1 mM), PE(1 mM), Tau(1.8 mM), and Thr(0.3 mM). The following scan parameters were used for the 5D EP-JRESI acquisition: field of view (FOV) = 16 × 16 × 12 cm3, (k x , k y , k z , t 2, t 1) = (16, 16, 8, 256, 64), direct spectral bandwidth = 1190 Hz, indirect spectral bandwidth = 500 Hz, and TR/TE = 1200/30ms. Retrospective undersampling using the NUS schemes for 4x, 8x, 12x, and 16x acceleration was performed for each phantom and these data were subsequently reconstructed to produce accelerated 5D EP-JRESI data. Eight voxels were chosen from each phantom measurement, and were quantified using ProFit for all of the acceleration factors used. Metabolite means, standard deviations, and CRLB% values were investigated using these 72 voxels.
In Vivo
Ten healthy volunteers (mean age = 25 years old) were scanned using the 8x accelerated 5D EP-JRESI sequence prospectively on a Siemens 3 T Trio (Siemens Healthcare, Erlangen, Germany) scanner. Outer volume suppression (OVS) bands were applied to suppress skull marrow lipids and contaminating signal outside the PRESS localization in addition to global water suppression33 for all in vivo measurements. The acquisitions were performed using the same parameters listed for the vphantom and in vitro measurements. Additionally, the TR used for acquisition was 1200 ms, the field of view (FOV) was set to 24 × 24 × 12 cm3, and the PRESS VOI was changed for each volunteer but was typically 9 × 12 × 6 cm3. Occipital white matter (OWM) and gray matter (OGM) voxels were identified for each volunteer and were fit using ProFit. In all cases, the medial occipital gray matter region was first identified, and then the voxel left of the OGM was assumed to be the OWM. Metabolite means and standard deviations were used to compare the metabolite concentrations for the two brain regions.
Statistical Analysis
For the virtual phantoms, all errors were calculated using standard nRMSE calculations as described in Eq. (6). However, the differences between the retrospective in vitro phantom measurements were assessed using means and standard deviations of the ProFit results. Furthermore, a Student’s t-test was used to determine if any significant (p < 0.05) differences were found between the fully sampled and accelerated phantom data for each metabolite. Similarly, multiple Student’s t-tests were also used to compare the OGM and OWM voxels from the in vivo measurements. To be more stringent when analyzing the in vivo results, a Bonferroni correction40 was applied, and significance was therefore determined to be p < 0.004.
Results
Virtual Phantoms
Metabolite maps of NAA and mI can be seen in Fig. 2 for the fully sampled simulation, and also for the 4x, 8x, 12x, and 16x accelerated vphantoms. The average nRMSE values for NAA were 1.34% for 4x, 6.47% for 8x, 13.4% for 12x, and 11.6% for 16x acceleration. Similarly, the average nRMSE values for mI were 2.07% for 4x, 10.6% for 8x, 18.7% for 12x and 20.6% for 16x acceleration. The reconstructed signal error increased throughout the entire vphantom as the acceleration factor increased, as demonstrated by the higher nRMSE values.
ProFit quantitation of 76 voxels from the fully sampled and accelerated 5D EP-JRESI vphantom acquisitions yielded several important findings. Figure 3 shows a bar graph with the concentration values for the metabolites fit using ProFit. The standard deviations for these concentration values increased as the acceleration factor increased. Most of the metabolites did not exceed the nRMSE threshold of 15% until 12x acceleration. However, metabolites that were simulated with lower concentrations such as Ala, Asc, Glc, and Thr exhibited nRMSE values greater than 15% at 4x and 8x acceleration. Figure 4 shows the nRMSE% values as a function of simulated concentration values (mM) to better highlight this result. The mean CRLB% values did not exceed 20% for most metabolites even at 16x acceleration, as seen in Table 1. However, the standard deviation of the CRLB% values for the metabolites steadily increased as the acceleration factor increased. At 4x acceleration, the standard deviation for the CRLB% values for most metabolites was 20% of the mean CRLB% value, whereas for 8x it was 40%, for 12x it was 80%, and for 16x acceleration it was also 80%.
Figure 5 shows the results of the second vphantom study. Three slices (slice 4 - slice 6) are shown for the fully sampled simulation, as well as the simulated results using NUS and non-linear reconstruction for 4x, 8x, 12x, and 16x acceleration. While there is signal leakage present for all of the accelerations, the maximum of this signal leakage is less than 1% of the original signal. While only the NAA results are shown, the signal leakage is still under 1% for much lower concentrated resonances as well, including Asp and mI. The leakage is exclusive to the undersampled spatial dimensions, (y,z), and is more apparent in the z direction, which is exemplified in Fig. 5.
In Vitro Phantom
Concentrations and CRLB% values for several metabolites calculated from the 5D EP-JRESI data acquired in the phantom can be seen in Fig. 6. For quantitation purposes, certain metabolites were combined: tNAA (NAA + NAAG), tCh (Ch + PCh + Gpc), and Glx (Glu + Gln), similar to what has been widely practiced in 1D MRS. This combination was performed because ProFit would over-estimate or under-estimate some of these metabolite components, resulting in values with very high standard deviations. This over/under estimation was present in both fully sampled and accelerated measurements. Due to the binary filtering before reconstruction, Lac was not reliably quantified. The average CRLB% values remained relatively low, and only exceeded the cutoff of 30% for GABA and Tau. Using 8x acceleration yielded no significant differences from the fully sampled results, except for Tau. However, several significant differences did exist for 12x and 16x NUS.
In Vivo
An example in vivo 5D EP-JRESI data set can be seen in Fig. 7. A metabolite map of NAA, which was produced by displaying the signal intensity calculated using peak integration over three spatial dimensions, is shown. Signal is only present in 3–4 slices due to the choice of the VOI and FOV. The fit results using ProFit show low residuals for both the OGM and OWM. The mean values and standard deviations over all ten healthy volunteers can be seen in Fig. 8.
Discussion
From the vphantom, in vitro, and in vivo results presented, it is clear that the ProFit algorithm is capable of fitting 2D J-resolved spectra acquired using the accelerated 5D EP-JRESI method. The vphantom results demonstrated nRMSE values of less than 15% for most metabolites when using 4x and 8x acceleration. NAA, Cr3.0, and tCh, which were both simulated with higher concentrations, had nRMSE values less than 7% when using 16x acceleration. The signal leakage for all metabolites was less than 1% of the original signal, as determined by the results from the second vphantom. Additionally, the retrospective in vitro phantom measurements demonstrated significant differences (p < 0.05) for several metabolites at 12x and 16x acceleration. For the in vivo measurements, voxels were chosen containing a majority of either white or gray matter consistent with a reference T 1-weighted MRI. Due to the spatial resolution of 1.5 × 1.5 × 1.5 cm3, quantified OWM and OGM voxels were not purely white or gray matter, but were instead a mixture of the two. This could be the major reason why no significant differences were found between the two spatial regions. Nonetheless, ProFit results from the 8x accelerated 5D EP-JRESI in healthy volunteers demonstrated metabolic trends for white and gray matter that were consistent with previous literature values41: elevated tNAA, tCh, and mI in the OWM, and elevated Glx in the OGM. These findings show the potential of ProFit quantitation in combination with the accelerated 5D EP-JRESI acquisition to differentiate tissues in vivo.
Spatially, higher nRMSE values were present as the acceleration factor increased. For the 8x accelerated metabolite maps displayed in Fig. 2, mI was simulated at 6 mM and had a spatial nRMSE of 10.6%, whereas NAA was simulated at 10.5 mM and had a spatial nRMSE of only 6.5%. This implies that lower concentration metabolites will have even larger spatial profile errors at 8x acceleration. The 12x and 16x reconstructions suffered from large errors in the central regions of the phantom. From Fig. 5, it is clear that there is minimal signal leakage present within the same slice, however this leakage is increased in neighboring slices. This is primarily due to the fact that y = 16, while z = 8 for the acquisition. Since y has more points, signal leakage or aliasing along this dimension is suppressed, while signal aliasing is more problematic in the z dimensions because of the limited number of points. The in vivo NAA metabolite map shown in Fig. 7 does not show noticeable aliasing artifacts, however. Therefore, for accurate spatial reconstruction, 8x acceleration appears to be the limit for this simulated brain phantom.
Both the vphantom and in vitro phantom measurements demonstrated that higher acceleration factors (12x and 16x) increased concentration error results as quantified by ProFit for several metabolites: Ala, Asc, Asp, Cr3.9, GABA, Glc, Glu, GSH, Lac, mI, PE, Tau, and Thr. Furthermore, lower concentration metabolites yielded higher errors even when using 4x and 8x acceleration. Since these signals were present very close to the spectral baseline, the reconstruction method was not capable of faithfully reconstructing these metabolites. In general, metabolites that had concentrations greater than 0.5 mM were accurate (nRMSE < 15%) until 12x acceleration.
Surprisingly, CRLB% values did not change drastically for most metabolites, even at 16x acceleration. From Eqs (3) and (4), it can be shown that the relationship between CRLB and noise standard deviation is: CRLB ∝ σ. Since the CRLB% did not change drastically, it is implied that noise was accurately reconstructed within the 2D spectra. From Table 1, it is clear that CRLB% is not a good indicator of reconstruction error, since several metabolites with nRMSE values greater than 15% would pass the CRLB% < 20% criterion. The CRLB% metric is more indicative of the orthogonality between basis sets for the metabolites, and using this as a quality control metric for ProFit quantitation of 2D J-resolved spectra from the accelerated 5D EP-JRESI acquisition may not be appropriate. In fact, using the CRLB% value as a quality filtering metric introduces bias into the quantitative results42, 43.
Aside from the NUS and iterative, nonlinear reconstruction process, other scan parameters and processing methods are also important for in vivo ProFit quantitation of the 5D EP-JRESI data. In order to shorten the acquisition duration, a TR of 1200ms was chosen as one of the scan parameters. This choice led to T 1 weighting of the spectra, which is a common problem in MRSI with one spectral dimension as well, and affected in vivo quantitation of metabolites44. Another scan parameter, t 2, may have also played a role in quantitation results. ProFit may have been able to distinguish metabolites with greater ease if spectral resolution was improved. However, acquiring more t 2 points leads to larger data sizes, greatly increasing the amount of memory necessary for reconstruction. With higher computational capabilities, however, t 2 = 512 or 1024 may be more feasible. In addition to WET pulses and OVS bands, which were used to suppress water and lipid signals, binary filtering was also applied to the direct spectral domain before reconstruction. This made it difficult to quantify metabolites close to or outside the 1.2–4.3 ppm range, such as Lac. With improvements to the sequence design, including better water45 and fat46 suppression pulses, or perhaps including metabolite cycling methods47,48,49, other metabolites may be quantified using ProFit in vivo.
Finally, taking into consideration the current scan parameters and processing methods, the vphantom, in vitro and in vivo results suggest that 8x acceleration is the upper limit for decreasing scan time while retaining accurate concentration values from ProFit. Metabolite ratios quantified in vivo as less than 0.06 with respect to Cr3.0 may have significant error even if the reported CRLB% is less than 20%. Application of ProFit quantitation to different accelerated 2D spectral SI methods, such as the accelerated echo planar correlated spectroscopic imaging (EP-COSI) method50,51,52, still needs to be evaluated. Utilization of higher acceleration factors may be achieved for different applications, including muscle and breast SI, since the primary resonances of interest (fat) have higher concentrations in vivo. Future studies will focus on implementing the 8x accelerated 5D EP-JRESI sequence in different pathologies and quantifying the results using ProFit.
Conclusion
This study demonstrates that the ProFit algorithm can be used to quantify 2D J-resolved spectra obtained from accelerated 5D EP-JRESI acquisitions. With the current implementation of the sequence, 8x acceleration is the upper limit for NUS while retaining accurate (nRMSE < 15%) ProFit quantitation results for many metabolites. Furthermore, quantitation of occipital white and gray matter regions in the brain using ProFit demonstrated regional trends consistent with previous in vivo results.
References
Bottomley, P. A. Spatial localization in nmr spectroscopy in vivo. Annals of the New York Academy of Sciences 508, 333–348 (1987).
Provencher, S. W. Estimation of metabolite concentrations from localized in vivo proton nmr spectra. Magnetic Resonance in Medicine 30, 672–679 (1993).
Ratiney, H. et al. Time-domain semi-parametric estimation based on a metabolite basis set. NMR in Biomedicine 18, 1–13 (2005).
Naressi, A. et al. Java-based graphical user interface for the mrui quantitation package. Magnetic Resonance Materials in Physics, Biology and Medicine 12, 141–152 (2001).
Vanhamme, L., van den Boogaart, A. & Van Huffel, S. Improved method for accurate and efficient quantification of mrs data with use of prior knowledge. Journal of Magnetic Resonance 129, 35–43 (1997).
Wilson, M., Reynolds, G., Kauppinen, R. A., Arvanitis, T. N. & Peet, A. C. A constrained least-squares approach to the automated quantitation of in vivo 1 h magnetic resonance spectroscopy data. Magnetic Resonance in Medicine 65, 1–12 (2011).
Chong, D. G., Kreis, R., Bolliger, C. S., Boesch, C. & Slotboom, J. Two-dimensional linear-combination model fitting of magnetic resonance spectra to define the macromolecule baseline using fitaid, a fitting tool for arrays of interrelated datasets. Magnetic Resonance Materials in Physics, Biology and Medicine 24, 147–164 (2011).
Aue, W., Bartholdi, E. & Ernst, R. R. Two-dimensional spectroscopy. application to nuclear magnetic resonance. The Journal of Chemical Physics 64, 2229–2246 (1976).
Ryner, L. N., Sorenson, J. A. & Thomas, M. A. Localized 2d j-resolved 1 h mr spectroscopy: strong coupling effects in vitro and in vivo. Magnetic Resonance Imaging 13, 853–869 (1995).
Kreis, R. & Boesch, C. Spatially localized, one-and two-dimensional nmr spectroscopy andin vivoapplication to human muscle. Journal of Magnetic Resonance, Series B 113, 103–118 (1996).
Thomas, M. A. et al. Localized two-dimensional shift correlated mr spectroscopy of human brain. Magnetic Resonance in Medicine 46, 58–67 (2001).
Dreher, W. & Leibfritz, D. Detection of homonuclear decoupled in vivo proton nmr spectra using constant time chemical shift encoding: Ct-press. Magnetic resonance imaging 17, 141–150 (1999).
Thomas, M. A., Hattori, N., Umeda, M., Sawada, T. & Naruse, S. Evaluation of two-dimensional l-cosy and jpress using a 3 t mri scanner: from phantoms to human brain in vivo. NMR in biomedicine 16, 245–251 (2003).
Schulte, R. F., Lange, T., Beck, J., Meier, D. & Boesiger, P. Improved two-dimensional j-resolved spectroscopy. NMR in Biomedicine 19, 264–270 (2006).
Schulte, R. F. & Boesiger, P. Profit: two-dimensional prior-knowledge fitting of j-resolved spectra. NMR in Biomedicine 19, 255–263 (2006).
Fuchs, A., Boesiger, P., Schulte, R. F. & Henning, A. Profit revisited. Magnetic Resonance in Medicine 71, 458–468 (2014).
Brown, T., Kincaid, B. & Ugurbil, K. Nmr chemical shift imaging in three dimensions. Proceedings of the National Academy of Sciences 79, 3523–3526 (1982).
Maudsley, A., Hilal, S., Perman, W. & Simon, H. Spatially resolved high resolution spectroscopy by “four-dimensional” nmr. Journal of Magnetic Resonance (1969) 51, 147–152 (1983).
Mansfield, P. Spatial mapping of the chemical shift in nmr. Magnetic Resonance in Medicine 1, 370–386 (1984).
Posse, S., Tedeschi, G., Risinger, R., Ogg, R. & Bihan, D. L. High speed 1 h spectroscopic imaging in human brain by echo planar spatial-spectral encoding. Magnetic Resonance in Medicine 33, 34–40 (1995).
Adalsteinsson, E. et al. Volumetric spectroscopic imaging with spiral-based k-space trajectories. Magnetic resonance in medicine 39, 889–898 (1998).
Furuyama, J. K., Wilson, N. E. & Thomas, M. A. Spectroscopic imaging using concentrically circular echo-planar trajectories in vivo. Magnetic resonance in medicine 67, 1515–1522 (2012).
Schirda, C. V., Tanase, C. & Boada, F. E. Rosette spectroscopic imaging: Optimal parameters for alias-free, high sensitivity spectroscopic imaging. Journal of Magnetic Resonance Imaging 29, 1375–1385 (2009).
Cunningham, C. H. et al. Design of flyback echo-planar readout gradients for magnetic resonance spectroscopic imaging. Magnetic resonance in medicine 54, 1286–1289 (2005).
Lustig, M., Donoho, D. & Pauly, J. M. Sparse mri: The application of compressed sensing for rapid mr imaging. Magnetic Resonance in Medicine 58, 1182–1195 (2007).
Hu, S. et al. Compressed sensing for resolution enhancement of hyperpolarized 13 c flyback 3d-mrsi. Journal of magnetic resonance 192, 258–264 (2008).
Geethanath, S. et al. Compressive sensing could accelerate 1 h mr metabolic imaging in the clinic. Radiology 262, 985–994 (2012).
Furuyama, J. K. et al. Application of compressed sensing to multidimensional spectroscopic imaging in human prostate. Magnetic Resonance in Medicine 67, 1499–1505 (2012).
Wilson, N. E., Iqbal, Z., Burns, B. L., Keller, M. & Thomas, M. A. Accelerated five-dimensional echo planar j-resolved spectroscopic imaging: Implementation and pilot validation in human brain. Magnetic Resonance in Medicine 75, 42–51 (2016).
Rao, C. R. Minimum variance and the estimation of several parameters. In Mathematical Proceedings of the Cambridge Philosophical Society, vol. 43, 280–283 (Cambridge Univ Press, 1947).
Cavassila, S., Deval, S., Huegen, C., Van Ormondt, D. & Graveron-Demilly, D. Cramer–rao bounds: an evaluation tool for quantitation. NMR in Biomedicine 14, 278–283 (2001).
Klose, U. In vivo proton spectroscopy in presence of eddy currents. Magnetic Resonance in Medicine 14, 26–30 (1990).
Ogg, R. J., Kingsley, R. & Taylor, J. S. Wet, a t 1-and b 1-insensitive water-suppression method for in vivo localized 1 h nmr spectroscopy. Journal of Magnetic Resonance, Series B 104, 1–10 (1994).
Barna, J., Laue, E., Mayger, M., Skilling, J. & Worrall, S. Exponential sampling, an alternative method for sampling in two-dimensional nmr experiments. Journal of Magnetic Resonance 73, 69–77 (1987).
Eslami, R. & Jacob, M. Robust reconstruction of mrsi data using a sparse spectral model and high resolution mri priors. Medical Imaging. IEEE Transactions on 29, 1297–1309 (2010).
Goldstein, T. & Osher, S. The split bregman method for l1-regularized problems. SIAM Journal on Imaging Sciences 2, 323–343 (2009).
Iqbal, Z., Wilson, N. E. & Thomas, M. A. 3d spatially encoded and accelerated te-averaged echo planar spectroscopic imaging in healthy human brain. NMR in Biomedicine 3, 329–339 (2016).
Smith, S., Levante, T., Meier, B. H. & Ernst, R. R. Computer simulations in magnetic resonance. an object-oriented programming approach. Journal of Magnetic Resonance, Series A 106, 75–105 (1994).
Govindaraju, V., Young, K. & Maudsley, A. A. Proton nmr chemical shifts and coupling constants for brain metabolites. NMR in Biomedicine 13, 129–153 (2000).
Van Belle, G., Fisher, L. D., Heagerty, P. J. & Lumley, T. Biostatistics: a methodology for the health sciences (John Wiley & Sons, 2004).
Pouwels, P. J. & Frahm, J. Regional metabolite concentrations in human brain as determined by quantitative localized proton mrs. Magnetic Resonance in Medicine 39, 53–60 (1998).
Tisell, A., Leinhard, O. D., Warntjes, J. B. M. & Lundberg, P. Procedure for quantitative 1 h magnetic resonance spectroscopy and tissue characterization of human brain tissue based on the use of quantitative magnetic resonance imaging. Magnetic resonance in medicine 70, 905–915 (2013).
Kreis, R. The trouble with quality filtering based on relative cramér-rao lower bounds. Magnetic resonance in medicine 75, 15–18 (2016).
Mlynárik, V., Gruber, S. & Moser, E. Proton t1 and t2 relaxation times of human brain metabolites at 3 tesla. NMR in Biomedicine 14, 325–331 (2001).
Tkáč, I., Starčuk, Z., Choi, I.-Y. & Gruetter, R. In vivo 1 h nmr spectroscopy of rat brain at 1 ms echo time. Magnetic Resonance in Medicine 41, 649–656 (1999).
Spielman, D. M., Pauly, J. M., Macovski, A., Glover, G. H. & Enzmann, D. R. Lipid-suppressed single-and multisection proton spectroscopic imaging of the human brain. Journal of Magnetic Resonance Imaging 2, 253–262 (1992).
Dreher, W. & Leibfritz, D. New method for the simultaneous detection of metabolites and water in localized in vivo 1 h nuclear magnetic resonance spectroscopy. Magnetic resonance in medicine 54, 190–195 (2005).
Hock, A. et al. Non-water-suppressed proton mr spectroscopy improves spectral quality in the human spinal cord. Magnetic resonance in medicine 69, 1253–1260 (2013).
Dong, Z. Proton mrs and mrsi of the brain without water suppression. Progress in nuclear magnetic resonance spectroscopy 86, 65–79 (2015).
Burns, B. L., Wilson, N. E. & Thomas, M. A. Group sparse reconstruction of multi-dimensional spectroscopic imaging in human brain in vivo. Algorithms 7, 276–294 (2014).
Burns, B., Wilson, N. E., Furuyama, J. K. & Thomas, M. A. Non-uniformly under-sampled multi-dimensional spectroscopic imaging in vivo: maximum entropy versus compressed sensing reconstruction. NMR in Biomedicine 27, 191–201 (2014).
Wilson, N. E., Burns, B. L., Iqbal, Z. & Thomas, M. A. Correlated spectroscopic imaging of calf muscle in three spatial dimensions using group sparse reconstruction of undersampled single and multichannel data. Magnetic resonance in medicine 74, 1199–1208 (2015).
Acknowledgements
The authors would like to acknowledge Dr. Paul Macey for donating a PC workstation, Dr. Anke Henning and associates from ETH Zurich, Switzerland, for sharing the earlier versions of the ProFit algorithm, the National Institute of Health R21 Grants (NS080649-02, EB020883-02, NS 086449-02), R56 Grant (HL 131010-02), and the UCLA Dissertation Year Fellowship (2016–2017).
Author information
Authors and Affiliations
Contributions
Z.I., N.E.W., and M.A.T. conceived the experiments, Z.I. and N.E.W. conducted the experiments, Z.I. analyzed the results. All authors reviewed the manuscript.
Corresponding author
Ethics declarations
Competing Interests
The authors declare that they have no competing interests.
Additional information
Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Iqbal, Z., Wilson, N.E. & Thomas, M.A. Prior-knowledge Fitting of Accelerated Five-dimensional Echo Planar J-resolved Spectroscopic Imaging: Effect of Nonlinear Reconstruction on Quantitation. Sci Rep 7, 6262 (2017). https://doi.org/10.1038/s41598-017-04065-1
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-017-04065-1
- Springer Nature Limited