Abstract
A conventional photoacoustic microscopy (PAM) system typically has to make tradeoffs between its spatial resolution and penetration depth, by choosing a fixed configuration of optical excitation and acoustic detection. The single-scale imaging capability of PAM may limit its applications in biomedical studies. Here, we report a quad-mode photoacoustic microscopy (QM-PAM) system with four complementary spatial resolutions and maximum penetration depths. For this we first developed a ring-shaped focused ultrasound transducer that has two independent elements with respective central frequencies at 20 MHz and 40 MHz, providing complementary acoustically-determined spatial resolutions and penetration depths. To accommodate the dual-element ultrasound transducer, we implemented two optical excitation modes to provide tightly- and weakly-focused light illumination. The dual-element acoustic detection combined with the two optical focusing modes can thus provide four imaging scales in a single imaging device, with consistent contrast mechanisms and co-registered field of views. We have demonstrated the multiscale morphological, functional, and molecular imaging capability of QM-PAM in the mouse head, leg and ear in vivo. We expect the high scale flexibility of QM-PAM will enable broad applications in preclinical studies.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Introduction
Photoacoustic microscopy1 (PAM) has been proven useful in morphological2, functional3, and molecular imaging1 with optical absorption contrast. However, conventional PAM systems, including optical-resolution photoacoustic microscopy (OR-PAM)4,5,6 and acoustic-resolution photoacoustic microscopy (AR-PAM)7,8,9,10,11,12, only work at a single scale with different type of applications. OR-PAM can provide a high lateral resolution (several micrometers) and limited penetration depth (~1 mm), and thus is widely used for single cell imaging and superficial skin imaging. AR-PAM offers relatively low resolution (tens of micrometers) and deep penetration (several millimeters), and thus can be applied for deep brain imaging. Due to the complementary characteristics of OR- and AR-PAM, it is necessary to develop a PAM system that can combine OR- and AR-PAM with flexible imaging scales for different biomedical applications13,14. However, it is not ideal to construct the OR- and AR-PAM systems separately and combine them into a single unit, as in reference15, because it involves redundant system components and the time-consuming sample translation. Recently, integrated PAM systems with tunable lateral resolutions have been reported. Estrada et al. developed a coaxial photoacoustic imaging head by integrating a focusable optical fiber with a focused ultrasonic transducer16, which can provide different resolutions at the optical and acoustic focus. Xing et al. employed an optical fiber bundle to deliver the excitation light beam and alternated the light beam size between OR and AR excitation pattern using only one or more fiber cores17. Jiang et al. improved this concept using an electrical varifocal lens to obtain continuously tunable optically-determined lateral resolution18. Nevertheless, the previous methods have only focused on altering the excitation light pattern without optimizing the acoustic detection, especially for mouse brain imaging with intact skull. In this work, we propose a new quad-mode PAM (QM-PAM) system that implements different optical excitation modes and ultrasound detection frequencies to explore improved flexibilities in both spatial resolution and maximum penetration depth, for functional and molecular photoacoustic imaging19,20,21,22.
Methods
QM-PAM system
The schematic of the proposed QM-PAM system is shown in Fig. 1. The excitation light source is a Credo dye laser (DCM, Sirah, Grevenbroich, Germany) pumped by a pulsed laser at 532 nm (IS8II-E, EdgeWave, Würselen, Germany), with a tunable wavelength range from 620 nm to 680 nm, and an Nd: YAG fiber laser (VPFL-G-20, V-Gen, Tel Aviv, Israel) at 532 nm. The dye laser beam is firstly collimated by an objective lens (PLN4X, Olympus, Tokyo, Japan) and a convex lens with a focal length of 30 mm (AC127-030-A, Thorlabs, Newton, NJ, USA). The two laser beams are then combined by a longpass dichroic mirror (DMLP605, Thorlabs, Newton, NJ, USA). Two convex lenses, L1 with a focal length of 200 mm (AC127-200-A, Thorlabs, Newton, NJ, USA) and L2 with a focal length of 50 mm (AC127-050-A, Thorlabs, Newton, NJ, USA), are 25 cm apart to provide the tightly- and weakly-focused optical modes, respectively. L2 tightly focuses the collimation light beam when L1 is flipped out the optical pathway, providing the OR optical mode, as indicated by the solid-outlined pathway in Fig. 1(a). When L1 is flipped into the optical pathway, it provides the weakly-focused optical excitation, i.e., the AR optical mode, as indicated by the dashed-outlined pathway in Fig. 1(a). Therefore, the OR and AR imaging modes are alternated by switching the optical pathway. For each mode, the two lasers can be used alone or together, depending on the wavelength requirement.
The incident light passed through the central hole of a ring-shaped focused ultrasound transducer with two independent elements before reaching the surface of the imaging target. The two transducer elements have central frequencies of 20 MHz and 40 MHz, respectively, and are concentrically and confocally arranged into a single device, as shown in Fig. 1(b,c). Both transducer elements were lapped from 500-μm lithium niobate plates (Boston Piezo-Optics, Bellingham, MA, USA) to the half-wavelength thickness. The lapped material was electroplated with chrome/gold (Cr/Au) layer on both sides. Conductive backing layer was casted using conductive silver epoxy. The materials were bonded together using epoxy to form a single cylinder, and pressed by a metal ball to form the same focal length of 11.25 mm. Both transducer elements have a −6-dB bandwidth of 78%. A central aperture with a diameter of 2 mm is used to deliver light, enabling co-axial alignment of optical excitation and acoustical detection for maximum detection sensitivity. The outer diameter of the high-frequency (40 MHz) transducer element is 7.9 mm, and the outer diameter of the low-frequency (20 MHz) transducer element is 11.2 mm. The two elements have independent amplifiers with an amplification of 51 dB. The acquired PA signals of the two elements are simultaneously sampled at 500 MHz by two independent channels of a high-speed DAQ card (ATS9350, AlazarTech, Pointe-Claire, Quebec, Canada). A two-dimensional motorized stage is used for raster scanning of the sample.
Trigger sequence
The functional and molecular imaging capabilities of QM-PAM were explored by implementing a double-wavelength illumination scheme. Both the dye laser for 640-nm light and the V-Gen laser for 532-nm light were triggered with a 2-μs delay, as shown in Fig. 1(a). The time delay between the triggers was necessary for separating the PA signals excited by the two laser pulses acquired in a single A-line. The time delay of the two laser pulses can be adjusted according to the required penetration depth.
Imaging in mice
The in vivo study for demonstrating the multiscale imaging capability of the QM-PAM was conducted on the head of a female Swiss Webster mouse (8 weeks old and 21.8 g in weight) with intact scalp and skull, and on the left hind leg of the same mouse. The in vivo study for demonstrating the functional imaging capability of the QM-PAM was conducted on the head of a female Swiss Webster mouse (10 weeks old and 26 g in weight) with the scalp removed but skull intact. The in vivo study for demonstrating the molecular imaging capability of the QM-PAM was conducted on the ear of a female Swiss Webster mouse (10 weeks old and 23.1 g in weight). The protocol was approved by the Institutional Animal Care and Use Committee (IACUC) of Duke University. All methods were performed in accordance with the relevant guidelines and regulations. The hair of the mouse head, leg and ear was shaved before imaging. During the imaging, the temperature of the mouse was held at 37 °C via a heating pad and the mouse was anesthetized via isoflurane (1.5%).
Bacteria preparation
miRFP670 expressing LMG194 (Invitrogen/ThermoFisher) Escherichia coli bacteria were prepared by co-transfection of the pWA23h plasmid encoding the heme oxygenase under rhamnose promoter23 and pBAD-miRFP670 plasmid encoding the miRFP670 near-infrared fluorescent protein under arabinose promoter24.
Data availability
The datasets generated and/or analyzed during the current study are available from the corresponding author on reasonable request.
Results and Discussions
Resolution measurements of QM-PAM
With two optical excitation modes and the dual-element ultrasound transducer, QM-PAM can operate at four different optical-acoustic configurations: OR-HF mode (optical-resolution high-frequency), OR-LF mode (optical-resolution low-frequency), AR-HF mode (acoustic-resolution high-frequency), and AR-LF mode (acoustic-resolution low-frequency). The axial resolution is 36 µm for OR-HF mode and AR-HF mode, and 72 µm for OR-LF mode and AR-LF mode (Supplementary Fig. S1.), determined by the bandwidth of the transducer elements. For OR-HF mode and OR-LF mode, the excitation light spot is much smaller than the acoustical focus spot, and the lateral resolution is optically determined by the light spot size of 6.9 µm (see Supplementary Fig. S2). For AR-HF and AR-LF modes, the lateral resolution is determined by the acoustical focus spot size of 76.3 µm (see Supplementary Fig. S2). The 20-MHz element has a twice larger acoustic numerical aperture than the 40-MHz element. Therefore, the two elements have the same acoustical focus spot size.
Multiscale imaging of QM-PAM in vitro and in vivo
The multiscale imaging capability of QM-PAM was first demonstrated on a green leaf skeleton embedded in an optically scattering medium. The scattering medium was made of 1.5% (weight/volume) agar mixed with 0.2% Intralipid (reduced scattering coefficient: ~0.4 mm−1 at 640 nm25). The green leaf skeleton was obliquely inserted into the scattering medium with an inclination angle of ~45°. The image was acquired at 640 nm. The laser pulse energy was 370 nJ for OR modes and 45 μJ for AR modes. The imaging results at four different modes are shown in Fig. 2(a–d). The maximum penetration depth was quantified as the depth at which the contrast-to-noise ratio dropped to 1. As expected, the maximum penetration depths of OR and AR modes are clearly different. For OR modes, the penetration depth was around 0.4~0.5 mm (Fig. 2(e,f)) for both elements, as the tightly-focused light was strongly attenuated before reaching the optical diffusion limit (~1 mm). Therefore, the penetration depth of OR modes was optically determined. By contrast, in AR modes, the penetration depth was jointly determined by the optical and acoustical attenuation. The weakly focused light, which had >100 times stronger pulse energy than OR modes, diffused into the scattering phantom and still generated strong photoacoustic signals at depths beyond 1 mm. The maximum penetration depth was 2.64 mm for AR-HF mode and 3.32 mm for AR-LF mode (Fig. 2(g,h)). It is worth pointing out that the 20-MHz detection is less sensitive to the limited view problem that is prevalent in high-frequency PAM systems26,27,28, as marked by the yellow rectangles in Fig. 2(g,h). The improvement in detecting vertical structures is mainly due to the wider acceptance angle of the 20-MHz transducer element.
For the mouse head imaging, the laser pulse energy at 640 nm was 370 nJ for OR modes and 45 μJ for AR modes. The head imaging results obtained at the four imaging modes are shown in Fig. 3. Blood vessels in the scalp and skull can be clearly imaged at different scales. Compared to the OR modes, the AR modes could image more blood vessels beneath the scalp, with worse resolutions. Compared to the HF modes, more vessels at larger depths were observed at LF modes due to the reduced acoustic attenuation at lower frequencies, as marked by the white arrows in Fig. 3(a–d). The side-view maximum projection images shown in Fig. 3(e–h) provide more details on penetration depths and axial resolutions. The OR modes only penetrated less than 1 mm into the scalp, while the AR modes penetrated through the scalp into the skull. Due to lower acoustic attenuation and wider detection angle, more vessels can be observed by the AR-LF mode, as marked by the yellow rectangles in Fig. 3(g,h).
Further, we imaged the mouse leg to demonstrate the maximum penetration depth in vivo. We used the 532-nm light from the EdgeWave pump laser. The laser pulse energy was 370 nJ for the OR modes and 400 μJ for the AR modes. The side-view maximum projection images of the LF modes (the larger penetration modes) are shown in Fig. 4(a,b). The penetration depth of OR-LF mode is less than 1 mm, as shown in Fig. 4(a), and the penetration depth of AR-LF mode is ~4 mm, as shown in Fig. 4(b). The outline of the leg bone can be clearly seen in the AR mode, as indicated by the dashed line in Fig. 4(b). The penetration depth of the AR mode is consistent with the previous result in the literature29. We have also compared the penetration difference between AR-HF and AR-LF modes, and studied the tissue’s acoustic attenuation dependence on the ultrasound frequency. To do so, we used the maximum laser pulse energy of 400 µJ to ensure adequate optical penetration. Figure 4(c,d) are the cross-sectional B-scan images of the mouse leg obtained by the two AR modes, which clearly show the penetration difference between the high-frequency and low-frequency detections. As the low-frequency detection benefited from less acoustic attenuation, deeper structures were obtained by the AR-LF mode. The dashed boxes in Fig. 4(d) show that more horizontal vessels were better imaged by AR-LF mode. Moreover, we quantified the average PA signal strength as a function of depth (Fig. 4(e)). The results have confirmed that (1) the low frequency detection had overall stronger signal strength over the entire depth, and (2) the low frequency signal had a slower decay and thus provided a deeper penetration.
Both the in vitro leaf phantom and the in vivo mouse results have demonstrated the multiscale imaging capability of QM-PAM, with progressive spatial resolutions and maximum penetration depths. It is clear that the OR modes can provide better spatial resolutions than the AR-modes, but less penetrations. Meanwhile, the 20-MHz transducer element can provide deeper penetration than the 40-MHz element, but worse axial resolutions.
Functional imaging of QM-PAM
The OR modes of the QM-PAM were used to image the oxygen saturation of hemoglobin (sO2) in the mouse head with the scalp removed and the skull intact. A traditional linear spectral unmixing method was used to quantify the relative concentrations of oxy-hemoglobin (HbO2) and deoxy-hemoglobin (HbR)30,31, based on the different optical absorption spectra of HbO2 and HbR (Fig. 5(a)). Figure 5(b,c) show the measured sO2 of the mouse head at two OR imaging modes. In both scales, we have observed the low sO2 levels in the mouse skull vessels, which agrees well with our previous results5. We also observed that the 20-MHz transducer element provided stronger signals from the mouse cortex than the 40-MHz transduce element, reflecting the skull’s frequency-dependent acoustic attenuation. The AR modes were not used for quantifying the sO2, due to the unknown optical fluence at different wavelengths. The less accurate sO2 quantification of the AR modes is majorly due to the volumetric averaging of PA signals from unresolved small vessels and due to the wavelength-dependent optical attenuation at deeper brain regions.
Molecular imaging of QM-PAM
The molecular imaging capability of QM-PAM was demonstrated in vivo on the mouse ear. The laser pulse energy was set to 100 nJ for the OR modes and 12 μJ for the AR modes. For this, the top and bottom surfaces of the ear were injected with bacteria expressing recently developed from bacterial phytochrome32 monomeric near-infrared fluorescent protein (FP) miRFP670 (absorbance maximum is at 641 nm and emission maximum is at 670 nm)24. miIRFP670 is one blue-shifted FP of the spectrally-distinct near-infrared FPs which can enable multi-spectral imaging across scales from cellular level to organismal level by labelling organs and tissues in vivo32,33. This FP has strong optical absorption at 640 nm (extinction coefficient is 87,000 M−1 cm−1) and has low absorption at 532 nm (~8,000 M−1 cm−1) whereas hemoglobin has the opposite trend, as shown in Fig. 5(a). Therefore, miRFP670 expressing bacteria can be identified by comparing the photoacoustic signals acquired at the two wavelengths.
The imaging results at different modes are shown in Fig. 6, in which the miRFP670 signals are displayed in red and the blood signals are shown in green. While the OR images, as shown in Fig. 6(a,b), clearly show the protein distribution on the top skin surface, the AR images show the proteins on both the top and the bottom surfaces, as shown by the yellow arrows in Fig. 6(c,d). To better show the depths of the proteins, cross-sectional B-scan images of the OR-LF mode and AR-LF mode were presented along the dashed line in Figs. 6(b,d), where the positions of the top and bottom proteins have been indicated, as shown by the dashed lines in Fig. 6(e,f). To further demonstrate the high-resolution imaging capability of QM-PAM, we also imaged a smear of mIRFP670-expressing bacteria on a glass slide, using the OR-HF mode with the 640-nm wavelength illumination (Fig. 6(g)). The result clearly shows that the OR-HF mode is cable of resolving single bacteria (Fig. 6(h)).
Conclusion
We have presented a multi-mode PAM that can provide four different length scales with complementary spatial resolutions and maximum penetration depths, as a single imaging device. The system’s function and molecular imaging capabilities have been demonstrated in the mouse head, leg and ear in vivo. We have shown that, by combining tightly and weakly focused optical excitation, QM-PAM can provide different lateral resolutions at the cost of penetration depths. By using a dual-element focused ultrasound transducer, QM-PAM can provide two different axial resolutions with different penetration depths. Combining the two maneuverers on light and sound, QM-PAM has achieved higher scalability than previous works.
However, one major drawback of QM-PAM is the worse acoustical lateral resolutions than that of conventional PAM systems using optical-acoustic beam combiners34, because the dual-element ultrasound transducer has a smaller numerical aperture. Then central aperture of the transducer also induces side-lobes in the acoustic detection. Nevertheless, the experiment results have strongly demonstrated the potential of the QM-PAM for preclinical applications, such as studying the heterogeneity of cancer hypoxic microenvironment. The acoustic resolution of QM-PAM can be improved by using ultrasound transducer with larger numerical apertures. A data fusion method that combines the acoustic signals detected by the two elements may also improve the effective detection bandwidth and reduce the side-lobes. Future studies will also include developing an image fusion algorithm that can merge images obtained at different modes.
References
Ntziachristos, V. Going deeper than microscopy: the optical imaging frontier in biology. Nat Methods 7, 603–614, https://doi.org/10.1038/nmeth.1483 (2010).
Wang, L. H. V. & Hu, S. Photoacoustic Tomography: In Vivo Imaging from Organelles to Organs. Science 335, 1458–1462, https://doi.org/10.1126/science.1216210 (2012).
Zhang, H. F., Maslov, K., Stoica, G. & Wang, L. H. V. Functional photoacoustic microscopy for highresolution and noninvasive in vivo imaging. Nat Biotechnol 24, 848–851, https://doi.org/10.1038/nbt1220 (2006).
Maslov, K., Zhang, H. F., Hu, S. & Wang, L. V. Optical-resolution photoacoustic microscopy for in vivo imaging of single capillaries. Opt Lett 33, 929–931, https://doi.org/10.1364/OL.33.000929 (2008).
Yao, J. J. et al. High-speed label-free functional photoacoustic microscopy of mouse brain in action. Nat Methods 12, 407–410, https://doi.org/10.1038/nmeth.3336 (2015).
Ma, R., Sontges, S., Shoham, S., Ntziachristos, V. & Razansky, D. Fast scanning coaxial optoacoustic microscopy. Biomed Opt Express 3, 1724–1731, https://doi.org/10.1364/BOE.3.001724 (2012).
Wang, L. H. V. & Yao, J. J. A practical guide to photoacoustic tomography in the life sciences. Nat Methods 13, 627–638, https://doi.org/10.1038/nmeth.3925 (2016).
Jathoul, A. P. et al. Deep in vivo photoacoustic imaging of mammalian tissues using a tyrosinase-based genetic reporter. Nature Photonics 9, 239–246, https://doi.org/10.1038/Nphoton.2015.22 (2015).
Omar, M., Schwarz, M., Soliman, D., Symvoulidis, P. & Ntziachristos, V. Pushing the Optical Imaging Limits of Cancer with Multi-Frequency-Band Raster-Scan Optoacoustic Mesoscopy (RSOM). Neoplasia 17, 208–214, https://doi.org/10.1016/j.neo.2014.12.010 (2015).
Laufer, J., Zhang, E., Raivich, G. & Beard, P. Three-dimensional noninvasive imaging of the vasculature in the mouse brain using a high resolution photoacoustic scanner. Appl Optics 48, D299–D306, https://doi.org/10.1364/AO.48.00D299 (2009).
Taruttis, A., van Dam, G. M. & Ntziachristos, V. Mesoscopic and Macroscopic Optoacoustic Imaging of Cancer. Cancer Res 75, 1548–1559, https://doi.org/10.1158/0008-5472.CAN-14-2522 (2015).
Dean-Ben, X. L., Lopez-Schier, H. & Razansky, D. Optoacoustic micro-tomography at 100 volumes per second. Sci Rep-Uk 7, 6850, https://doi.org/10.1038/s41598-017-06554-9 (2017).
Wang, L. V. Multiscale photoacoustic microscopy and computed tomography. Nat Photonics 3, 503–509, https://doi.org/10.1038/Nphoton.2009.157 (2009).
Soliman, D., Tserevelakis, G. J., Omar, M. & Ntziachristos, V. Combining microscopy with mesoscopy using optical and optoacoustic label-free modes. Sci Rep 5, 12902, https://doi.org/10.1038/srep12902 (2015).
Upputuri, P. K., Krisnan, M. S., Moothanchery, M. & Pramanik, M. Photonic nanojet engineering to achieve super-resolution in photoacoustic microscopy - a simulation study. Photons Plus Ultrasound: Imaging and Sensing 10064: Unsp 100644s https://doi.org/10.1117/12.2250483 (2017).
Estrada, H., Turner, J., Kneipp, M. & Razansky, D. Real-time optoacoustic brain microscopy with hybrid optical and acoustic resolution. Laser Phys Lett 11, 045601, https://doi.org/10.1088/1612-2011/11/4/045601 (2014).
Xing, W. X., Wang, L. D., Maslov, K. & Wang, L. H. V. Integrated optical- and acoustic-resolution photoacoustic microscopy based on an optical fiber bundle. Opt Lett 38, 52–54, https://doi.org/10.1364/OL.38.000052 (2013).
Jiang, B. W., Yang, X. Q., Liu, Y. Y., Deng, Y. & Luo, Q. M. Multiscale photoacoustic microscopy with continuously tunable resolution. Opt Lett 39, 3939–3941, https://doi.org/10.1364/OL.39.003939 (2014).
Kim, C. et al. In Vivo Molecular Photoacoustic Tomography of Melanomas Targeted by Bioconjugated Gold Nanocages. Acs Nano 4, 4559–4564, https://doi.org/10.1021/nn100736c (2010).
Pu, K. Y. et al. Semiconducting polymer nanoparticles as photoacoustic molecular imaging probes in living mice. Nature Nanotechnology 9, 233–239, https://doi.org/10.1038/nnano.2013.302 (2014).
Razansky, D., Vinegoni, C. & Ntziachristos, V. Multispectral photoacoustic imaging of fluorochromes in small animals. Opt Lett 32, 2891–2893, https://doi.org/10.1364/OL.32.002891 (2007).
Wang, X. D. et al. Noninvasive laser-induced photoacoustic tomography for structural and functional in vivo imaging of the brain. Nat Biotechnol 21, 803–806, https://doi.org/10.1038/nbt839 (2003).
Shcherbakova, D. M. & Verkhusha, V. V. Near-infrared fluorescent proteins for multicolor in vivo imaging. Nat Methods 10, 751, https://doi.org/10.1038/nmeth.2521 (2013).
Shcherbakova, D. M. et al. Bright monomeric near-infrared fluorescent proteins as tags and biosensors for multiscale imaging. Nat Commun 7, 12405, https://doi.org/10.1038/ncomms12405 (2016).
Du Le, V. N., Nie, Z., Hayward, J. E., Farrell, T. J. & Fang, Q. Measurements of extrinsic fluorescence in Intralipid and polystyrene microspheres. Biomed Opt Express 5, 2726–2735, https://doi.org/10.1364/BOE.5.002726 (2014).
Dean-Ben, X. L. & Razansky, D. On the link between the speckle free nature of optoacoustics and visibility of structures in limited-view tomography. Photoacoustics 4, 133–140, https://doi.org/10.1016/j.pacs.2016.10.001 (2016).
Guo, Z. J., Li, L. & Wang, L. H. V. On the speckle-free nature of photoacoustic tomography. Med Phys 36, 4084–4088, https://doi.org/10.1118/1.3187231 (2009).
Liu, W. et al. Correcting the limited view in optical-resolution photoacoustic microscopy. J Biophotonics. https://doi.org/10.1002/jbio.201700196 (2017).
Wang, L., Maslov, K., Xing, W., Garcia-Uribe, A. & Wang, L. V. Video-rate functional photoacoustic microscopy at depths. J Biomed Opt 17, 106007, https://doi.org/10.1117/1.JBO.17.10.106007 (2012).
Horecker, B. L. The absorption spectra of hemoglobin and its derivatives in the visible and near infra-red regions. J Biol Chem 148, 173–183 (1943).
Yao, J. J. et al. Double-illumination photoacoustic microscopy. Opt Lett 37, 659–661, https://doi.org/10.1364/OL.37.000659 (2012).
Shcherbakova, D. M., Baloban, M. & Verkhusha, V. V. Near-infrared fluorescent proteins engineered from bacterial phytochromes. Curr Opin Chem Biol 27, 52–63, https://doi.org/10.1016/j.cbpa.2015.06.005 (2015).
Chernov, K. G., Redchuk, T. A., Omelina, E. S. & Verkhushaa, V. V. Near-Infrared Fluorescent Proteins, Biosensors, and Optogenetic Tools Engineered from Phytochromes. Chem Rev 117, 6423–6446, https://doi.org/10.1021/acs.chemrev.6b00700 (2017).
Wang, L. V. & Yao, J. A practical guide to photoacoustic tomography in the life sciences. Nature methods 13, 627–638, https://doi.org/10.1038/nmeth.3925 (2016).
Acknowledgements
We thank Emelina Vienneau for editing the manuscript. This work was supported by Duke MEDx fund and AHA collaborative sciences grant 18CSA34080277 (to J.Y.), by grants GM122567 and NS103573 from the US National Institutes of Health and ERC-2013-ADG-340233 from the EU FP7 program (to V.V.V.).
Author information
Authors and Affiliations
Contributions
V.V.V. and J.Y. initialized and supervised the overall project. W.L., D.M.S., V.V.V. and J.Y. designed the experiments. W.L., N.K., Y.L., Q.Z. and J.Y. constructed the QM-PAM system, W.L. performed the photoacoustic experiments, and W.L. and J.Y. analyzed the photoacoustic data. D.M.S. and V.V.V. developed the miRFP670 fluorescent protein and miRFP670-expressing bacteria. W.L., V.V.V. and J.Y. wrote the manuscript. All authors reviewed the manuscript.
Corresponding authors
Ethics declarations
Competing Interests
The authors declare no competing interests.
Additional information
Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Electronic supplementary material
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Liu, W., Shcherbakova, D.M., Kurupassery, N. et al. Quad-mode functional and molecular photoacoustic microscopy. Sci Rep 8, 11123 (2018). https://doi.org/10.1038/s41598-018-29249-1
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-018-29249-1