Abstract
An endomicroscope opens new frontiers of non-invasive biopsy for in vivo imaging applications. Here we report two-photon laser scanning endomicroscope for in vivo cellular and tissue imaging using a Lissajous fiber scanner. The fiber scanner consists of a piezoelectric (PZT) tube, a single double-clad fiber (DCF) with high fluorescence collection, and a micro-tethered-silicon-oscillator (MTSO) for the separation of biaxial resonant scanning frequencies. The endomicroscopic imaging exhibits 5 frames/s with 99% in scanning density by using the selection rule of scanning frequencies. The endomicroscopic scanner was compactly packaged within a stainless tube of 2.6 mm in diameter with a high NA gradient-index (GRIN) lens, which can be easily inserted into the working channel of a conventional laparoscope. The lateral and axial resolutions of the endomicroscope are 0.70 µm and 7.6 μm, respectively. Two-photon fluorescence images of a stained kidney section and miscellaneous ex vivo and in vivo organs from wild type and green fluorescent protein transgenic (GFP-TG) mice were successfully obtained by using the endomicroscope. The endomicroscope also obtained label free images including autofluorescence and second-harmonic generation of an ear tissue of Thy1-GCaMP6 (GP5.17) mouse. The Lissajous scanning two-photon endomicroscope can provide a compact handheld platform for in vivo tissue imaging or optical biopsy applications.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Introduction
Advanced optical microscopy such as optical coherence tomography (OCT)1,2, confocal microscopy (CM)3,4,5,6,7, or two-photon microscopy (TPM)8,9,10,11 that enables bioimaging with intracellular or cellular level resolution has been actively miniaturized for in vivo imaging applications such as optical biopsy1,2,3,4,5,6,7,9,12,13,14 or in vivo brain imaging of freely behaving animals8,15,16. Unlike OCT with non-specificity and relatively low resolution17 or CM with low penetration and relatively high photobleaching18, TPM has several advantages such as deeper penetration depth, inherent optical sectioning capabilities, and low photobleaching and photodamage in a small volume of the focal plane18,19,20.
Endomicroscopic TPM has been recently demonstrated by using micromirrors15,21,22,23,24 and fiber scannners8,9,25,26,27,28,29,30,31,32. However, the endomicroscopic TPM are still challenging due to compact packaging, stability of scanners, high frame rate, or group delay dispersion. Micromirror based scanning probes requires additional optical components to achieve forward-viewing endomicroscopy, which precludes from reducing the diameter of the whole probes into smaller than 5 mm in a diameter21,33. A fiber-optic scanning probe can be reduced smaller than 3 mm, which facilitates the insertion into the working channel in an endoscope9,31. A fiber scanner consists of a piezoelectric (PZT) actuator and a single fiber, generating three different scanning patterns; raster27,28, spiral9,25,29,30,31, and Lissajous8,26,32. The resonant-based spiral scanners are successful regarding with small form factor and fast image acquisition25, but the fiber scanner with spiral pattern is usually venerable due to eccentricity of PZT tube, which causes a mechanical coupling phenomenon coupling between the x-axis and the y-axis34. However, Lissajous scanners distinctly emphasize high illumination uniformity and low photodamage in the center region32. Besides, the Lissajous scanning solves the mechanical coupling problem because it drives the x-axis and y-axis at different frequencies. Recent development of the micromachined tethered silicon oscillator (MTSO) attached on a fiber tip can readily generate Lissajous scanning for confocal endomicroscopy35,36,37,38. However, the Lissajous scanner based on the PZT has not been yet applied for two-photon endomicroscopy. In addition, the other conventional Lissajous two-photon endomicroscopies8,26,32 have been demonstrated at speeds below 3 frames.
Here we report a fully packaged two-photon endomicroscope for in vivo cellular and tissue imaging using Lissajous fiber scanner. Figure 1a shows the schematic diagram of Lissajous two-photon endomicroscope and its application on in vivo tissue imaging. The TPM setup couples femtosecond pulse laser into the Lissajous fiber probe and visualizes tissue images. Figure 1b also shows the schematic design of endomicroscopic probe comprising a GRIN lens and a quadrupole piezoelectric (PZT) fiber scanner based on a MTSO coupled a double-clad fiber (DCF), which delivers femtosecond pulsed laser and increases the collection of fluorescence emission signals through both the core and the inner clad. The fiber scanner obtained real-time ex vivo and in vivo images at 5 frame rate using the frequency selection rule for high definition and high frame rate (HDHF)36 Lissajous scanning.
Results and Discussions
Resonant Lissajous fiber scanner design and optomechanical properties
The MTSO attached on the DCF, which has 7 mm long, differentiates the bi-directional resonant frequencies to generate Lissajous scanning. The MTSO consists of a silicon mass and a microspring which has two silicon masses with a 3 mm long structure. The silicon mass was attached on the distal end of the optical fiber to increase effective mass, and the center of microspring was set to 2.5 mm away from the PZT tube (Fig. 2a). Figure 2b shows the frequency response and the scanning amplitudes of Lissajous fiber scanner depending on various driving voltages. The resonant frequencies of x-axis (fx) and y-axis (fy) were 882 Hz and 1,155 Hz, respectably. The MTSO separated resonant frequencies 273 Hz and eliminated the mechanical coupling phenomenon. The scanning amplitudes of x-axis and y-axis were 400 μm and 450 μm at 20 Vpp, respectively and the scanning amplitude ratio between x-axis and y-axis was 0.88 at 20 Vpp. A particular set of scanning frequencies using the selection rule can increase a frame rate and scanning density that allows the HDHF Lissajous scanning [24]. Figure 2c shows a color-map of total lobe number (N) depending on the scanning frequencies, where the maximum (N \(\ge \) 2,000) and the minimum (N \(\le \) 350) indicate in yellow and black color, respectively. Figure 2d describes a color map of the great common division (GCD) depending on the scanning frequencies, and the regions, where N is lower than 350, are filled in black color. The scanning frequencies for HDHF Lissajous scanning were selected as 885 Hz and 1,160 Hz for x-axis and y-axis, respectively for N = 409 and GCD = 5. The comparison between the conventional and the HDHF scanning methods is shown in Fig. 2e. The scanning frequencies of conventional Lissajous scanning were selected as the exact resonance frequencies (fx = 882 Hz and fy = 1,155 Hz). In contrast, HDHF Lissajous scanning can be operated at pseudo resonance frequencies (fx′ = 885 Hz and fy′ = 1,160 Hz) near the resonance. The scanning density of the HDHF Lissajous scanning (99%) was increased compared to that of conventional Lissajous scanning (56.6%) at the same measured time (0.2 s). Time-lapse images for stable Lissajous scanning of 885 Hz and 1160 Hz were captured for 2.5 msec, 10 msec, 20 mesc, 50 msec, and 100 msec as shown in Fig. 2f. The fill factors are 3.4%, 12.5%, 24.6%, 53.4%, and 81% for each time in order. The scanning speed of Lissajous scanning was in the range of approximately 0.2 pixel/μs to 1 pixel/μs. The scan speed is relatively slow at the edge of the Lissajous pattern than other areas, but only about 2% of the total pixels have the scanning speed of 0.33 pixel/μs or less (see Supplementary Fig. S1). The Lissajous fiber scanner was fully packaged with a stainless housing tube and a lens tube including the achromatic miniature GRIN lens (MO-080-018-AC450-900, GRINTECH), which has magnification of 4.76 with image NA of 0.175 and objective NA of 0.8, and the optical image is shown in Fig. 2g. Fully packaged scanning probe had 2.6 mm in diameter, 30 mm in length, and 0.3 g in weight. The probe shows relatively small filed-of-view compared to the spiral pattern endomicroscope due to the size of the silicon mass and the objective mount, but the silicon mass produced a Lissajous pattern that improved frame rate and stability over the spiral.
Lissajous two-photon endomicroscope system setup and optical resolution
The endomicroscopic system setup is shown in Fig. 3a. A Ti:sapphire (Ti:S) laser (Chameleon vision II, Coherent) provides ultrashort laser pulses with 100 fs pulse width. The excitation laser was coupled into a DCF (SMM 900, Fibercore) using an objective lens (RMS X10, Thorlabs). To compensate the chromatic dispersion of the optical fiber, the laser was pre-chirped by a prism array inside the Ti:S laser. Because the maximum negative group delay dispersion of the prism arrays was 9,500 fs2, the length of the DCF was set to 17 cm (see Supplementary, Fig. S2). The optical resolution of the endomicroscope was characterized by using the customized United States Air Force (USAF) test target patterned PDMS fluorescein isothiocyanate (FITC) coating (see Supplementary, Fig. S3) and the 10 µm (G1000, Thermo Fishers Scientific) and 200 nm (QGC-200-1, Ocean NanoTech) beads, as shown in Fig. 3b–g. Figure 3b shows the microscope (L-IM, Nikon) fluorescence image of the customized USAF test target patterned PDMS (group 6 and 7). Two-photon fluorescence (TPF) image of the group 7, elements 5 and 6, which have line widths of 2.46 μm and 2.19 μm, respectively, was discernible using the endomicroscope (Fig. 3c). Figure 3d shows the intensity profile across the vertical bars of group 7, element 6. Figure 3e shows TPF image of 10 μm beads. The lateral and axial full-width at half-maximum (FWHM) of the intensity profiles of 200 nm fluorescence beads were 0.70 μm and 7.6 μm, respectively (Fig. 3f,g).
Two-photon mouse tissue images
Two-photon mouse tissue images were obtained from a stained mouse kidney section, miscellaneous ex vivo and in vivo organs, and label free images of ex-vivo mouse ear, as shown in Fig. 4. A mouse kidney section stained with Alexa Fluor 488 wheat germ agglutinin (F-24630, Invitrogen) was used to compare two-photon tissue images of a commercial TPM (LSM 510, Zeiss) and the endomicroscope (Fig. 4a,b). Both microscopic images, which show the convoluted tubules in kidney cells, were comparable, although the image obtained by the endomicroscope had five times faster frame rate and three times lower frame averaging. Wild type (BALB/c) mice and H2B green fluorescent protein transgenic (GFP-TG) mice were prepared to obtain ex vivo (Fig. 4c–e) and in vivo tissue imaging (Fig. 4g,h). In case of the wild type mouse, 1% FITC was injected into the mouse by intravenous injection to obtain microvascular images of each organ. In case of the H2B GFP-TG mouse, the green fluorescent protein (GFP) signals were obtained from nuclei in small intestine. Figure 4c and d show microvascular ex vivo images of the wild type mouse small intestine and ear, respectively. Figure 4e shows ex vivo GFP signals of smooth muscle cells in the GFP-TG mouse small intestine. After both mice were anesthetized, the mice were placed on the XYZ translation stage for in vivo tissue imaging (Fig. 4f). Figure 4g and h show microvascular in vivo image of the wild type mouse ear and in vivo GFP signals of smooth muscle cells in the GFP-TG mouse small intestine, respectively. Thy1-GCaMP6 (GP5.17) mouse was prepared to obtain label-free imaging (Fig. 4i–l). Figure 4I,j show ex-vivo autofluorescence images of mitochondrial NADH from the mouse ear and Fig. 4k,l show ex-vivo second-harmonic generation (SHG) images from collagens of the mouse ear.
We have demonstrated the Lissajous two-photon endomicroscope for in vivo tissue imaging. The attachment of MTSO on the DCF enables Lissajous scanning, which increases illumination uniformity and reduces photodamage, and applying the frequency selection rule leads to achieve up to 5 frames/s with 99% in scanning density. The image of the kidney convoluted tubule acquired from the endomicroscope was comparable to those from conventional TPMs, although the image from the endomicroscope had five times faster frame rate and three times lower frame averaging. Furthermore, the microvascular FITC and the GFP signals of in vivo images and label-free images demonstrate that the Lissajous two-photon endomicroscope can be utilized for in vivo cellular and tissue imaging.
Methods
Lissajous fiber scanner fabrication
The Lissajous fiber scanner consists of a PZT tube, a DCF, and a MTSO. The MTSO has a silicon mass of 0.5 mm × 0.5 mm × 0.4 mm and a microspring, which has two silicon masses with a 3 mm long and 50 µm width support structure. The MTSO were micro-machined by using 6-inch SOI wafer with a 40 μm thick top silicon layer, a 2 μm thick buried oxide layer, and a 400 μm thick bottom silicon layer. First, the bottom side of the SOI wafer was passivated with a thin aluminum layer to improve heat conduction and prevent leakages of helium cooling gas during a deep reactive ion etching (DRIE) process. Second, the top side of the wafer was patterned with photoresist and etched by using DRIE. Third, the top side of the wafer was passivated with parylene to prevent thermal stress during the next DRIE process. Fourth, the bottom side of the wafer was patterned with photoresist and etched by using DRIE. Fifth, the remained photoresist and passivation layers were completely removed by using plasma ashing and aluminum wet etching. The MTSO were released by removing the buried oxide layer using vapored HF and disconnecting silicon tethers with Joule heating35.
Microscope setup
A Ti:sapphire laser (Chameleon vision II, Coherent) provided ultrashort laser pulses at 900 nm (initial pulse width ~ 100 fs) pre-chirped by an embedded prism array. The beam intensity was controlled by using an achromatic half-wave plate (AHWP05M-980, Thorlabs) and a calcite polarizer (10GL08AR.16, Newport). Dichroic beam splitter (FF705-Di01-25 × 36, Semrock) passed the laser pulses (900 nm/840 nm/880 nm) and reflected signal (509~525 nm for two-photon singal 420~440 nm for label free signal). To couple the laser to the core of DCF, the beam was initially expanded using two achromatic lenses (f = 160 mm, 67-334-INK, Edmund Optics), (f = 30 mm, 49–352-INK, Edmund Optics). The expanded beam was focused by using a microscope objective lens (RMS X10, Thorlabs) and coupled to the core of DCF (see Supplementary Fig. S4). The emission light from a sample passed a band pass filter (FF01-525/45–25, Semrock for two-photon signal and FF01-417/60–25, Semrock for label-free signal) to transmit only signal, and the signal was focused to the photomultiplier (R3788, Hamamatsu) using an achromatic lens (47–633-INK, Edmund Optics).
Mouse preparation
Wild type (BALB/c), H2B GFP-TG, and GP5.17 mice were used to obtain ex vivo, in vivo, and label-free tissue imaging. For the in vivo imaging, both mice were anesthetized by using intraperitoneal injection of a mixture of zoletil (30 mg/kg) and xylazine (10 mg/kg). The mouse skin was shaved by hair clippers and removal cream before the surgical procedure for imaging. After a small incision on the abdominal skin of the anesthetized mice, small intestine was pulled out to minimize motion artifacts and imaged by using the endomicroscope. In case of wild type mice, 1% FITC-dextran (BCBF2730V, Sigma) was intravenously injected into the mice to visualize microvascular images. In case of the GP5.17 mouse, ears were cut out and imaged after shaving. This study was carried out in accordance with the ARRIVE (Animal Research: Reporting In Vivo Experiments) guidelines. All animal experiments were performed in accordance with the protocol approved by the Animal Care Committee of KAIST (protocol no. KA2018-06). All surgical procedures were performed under anesthesia, and all efforts were made to minimize the suffering.
References
Liang, K. C. et al. Cycloid scanning for wide field optical coherence tomography endomicroscopy and angiography in vivo. Optica 5, 36–43, https://doi.org/10.1364/Optica.5.000036 (2018).
Adler, D. C. et al. Three-dimensional endomicroscopy using optical coherence tomography. Nat Photonics 1, 709–716, https://doi.org/10.1038/nphoton.2007.228 (2007).
Polglase, A. L. et al. A fluorescence confocal endomicroscope for in vivo microscopy of the upper- and the lower-GI tract. Gastrointest Endosc 62, 686–695, https://doi.org/10.1016/j.gie.2005.05.021 (2005).
Shin, H. J. et al. Fiber-optic confocal microscope using a MEMS scanner and miniature objective lens. Opt Express 15, 9113–9122, https://doi.org/10.1364/Oe.15.009113 (2007).
Kiesslich, R., Goetz, M., Vieth, M., Galle, P. R. & Neurath, M. F. Technology insight: confocal laser endoscopy for in vivo diagnosis of colorectal cancer. Nat Clin Pract Oncol 4, 480–490, https://doi.org/10.1038/ncponc0881 (2007).
Sharma, P. et al. Real-time increased detection of neoplastic tissue in Barrett’s esophagus with probe-based confocal laser endomicroscopy: final results of an international multicenter, prospective, randomized, controlled trial. Gastrointest Endosc 74, 465–472, https://doi.org/10.1016/j.gie.2011.04.004 (2011).
Hurlstone, D. P. et al. In vivo real-time confocal laser scanning endomicroscopic colonoscopy for the detection and characterization of colorectal neoplasia. Brit J Surg 95, 636–645, https://doi.org/10.1002/bjs.5988 (2008).
Helmchen, F., Fee, M. S., Tank, D. W. & Denk, W. A miniature head-mounted two-photon microscope. high-resolution brain imaging in freely moving animals. Neuron 31, 903–912 (2001).
Myaing, M. T., MacDonald, D. J. & Li, X. Fiber-optic scanning two-photon fluorescence endoscope. Opt Lett 31, 1076–1078 (2006).
Murari, K. et al. Compensation-free, all-fiber-optic, two-photon endomicroscopy at 1.55 mu m. Optics Letters 36, 1299–1301, https://doi.org/10.1364/Ol.36.001299 (2011).
Liang, W. X., Hall, G., Messerschmidt, B., Li, M. J. & Li, X. D. Nonlinear optical endomicroscopy for label-free functional histology in vivo. Light-Sci Appl 6, https://doi.org/10.1038/lsa.2017.82 (2017).
Seo, Y. H., Hwang, K. & Jeong, K. H. 1.65 mm diameter forward-viewing confocal endomicroscopic catheter using a flip-chip bonded electrothermal MEMS fiber scanner. Opt Express 26, 4780-+, https://doi.org/10.1364/Oe.26.004780 (2018).
Seo, Y. H., Hwang, K., Park, H. C. & Jeong, K. H. Electrothermal MEMS fiber scanner for optical endomicroscopy. Opt Express 24, 3903–3909, https://doi.org/10.1364/OE.24.003903 (2016).
Hwang, K., Seo, Y.-H. & Jeong, K.-H. Microscanners for optical endomicroscopic applications. Micro and Nano Systems Letters 5, 1, https://doi.org/10.1186/s40486-016-0036-4 (2017).
Piyawattanametha, W. et al. In vivo brain imaging using a portable 2.9 g two-photon microscope based on a microelectromechanical systems scanning mirror. Optics Letters 34, 2309–2311, https://doi.org/10.1364/Ol.34.002309 (2009).
Zong, W. J. et al. Fast high-resolution miniature two-photon microscopy for brain imaging in freely behaving mice (vol 14, pg 713, 2017). Nat Methods 14 (2017).
Dunkers, J. P., Sanders, D. P., Hunston, D. L., Everett, M. J. & Green, W. H. Comparison of optical coherence tomography, X-ray computed tomography, and confocal microscopy results from an impact damaged epoxy/E-glass composite. J Adhesion 78, 129–154, https://doi.org/10.1080/00218460210386 (2002).
Denk, W., Strickler, J. H. & Webb, W. W. Two-photon laser scanning fluorescence microscopy. Science 248, 73–76 (1990).
Berland, K. M., So, P. T. & Gratton, E. Two-photon fluorescence correlation spectroscopy: method and application to the intracellular environment. Biophys J 68, 694–701, https://doi.org/10.1016/S0006-3495(95)80230-4 (1995).
Gibson, E. A., Masihzadeh, O., Lei, T. C., Ammar, D. A. & Kahook, M. Y. Multiphoton microscopy for ophthalmic imaging. J Ophthalmol 2011, 870879, https://doi.org/10.1155/2011/870879 (2011).
Duan, X. et al. MEMS-based multiphoton endomicroscope for repetitive imaging of mouse colon. Biomed Opt Express 6, 3074–3083, https://doi.org/10.1364/BOE.6.003074 (2015).
Duan, X. Y., Li, H. J., Li, X., Oldham, K. R. & Wang, T. D. Axial beam scanning in multiphoton microscopy with MEMS-based actuator. Opt Express 25, 2195–2205, https://doi.org/10.1364/Oe.25.002195 (2017).
Piyawattanametha, W. et al. Fast-scanning two-photon fluorescence imaging based on a microelectromechanical systems two- dimensional scanning mirror. Opt Lett 31, 2018–2020 (2006).
Fu, L., Jain, A., Xie, H., Cranfield, C. & Gu, M. Nonlinear optical endoscopy based on a double-clad photonic crystal fiber and a MEMS mirror. Opt Express 14, 1027–1032 (2006).
Engelbrecht, C. J., Johnston, R. S., Seibel, E. J. & Helmchen, F. Ultra-compact fiber-optic two-photon microscope for functional fluorescence imaging in vivo. Opt Express 16, 5556–5564 (2008).
Flusberg, B. A., Lung, J. C., Cocker, E. D., Anderson, E. P. & Schnitzer, M. J. In vivo brain imaging using a portable 3.9 gram two-photon fluorescence microendoscope. Optics Letters 30, 2272–2274, https://doi.org/10.1364/Ol.30.002272 (2005).
Do, D., Yoo, H. & Gweon, D. G. Fiber-optic raster scanning two-photon endomicroscope using a tubular piezoelectric actuator. Journal of Biomedical Optics 19, https://doi.org/10.1117/1.Jbo.19.6.066010 (2014).
Sawinski, J. & Denk, W. Miniature random-access fiber scanner for in vivo multiphoton imaging. J Appl Phys 102, https://doi.org/10.1063/1.2763945 (2007).
Bao, H., Allen, J., Pattie, R., Vance, R. & Gu, M. Fast handheld two-photon fluorescence microendoscope with a 475 microm x 475 microm field of view for in vivo imaging. Opt Lett 33, 1333–1335 (2008).
Wu, Y., Leng, Y., Xi, J. & Li, X. Scanning all-fiber-optic endomicroscopy system for 3D nonlinear optical imaging of biological tissues. Opt Express 17, 7907–7915 (2009).
Ducourthial, G. et al. Development of a real-time flexible multiphoton microendoscope for label-free imaging in a live animal. Sci Rep-Uk 5, https://doi.org/10.1038/srep18303 (2015).
Liang, W. et al. Increased illumination uniformity and reduced photodamage offered by the Lissajous scanning in fiber-optic two-photon endomicroscopy. J Biomed Opt 17, 021108, https://doi.org/10.1117/1.JBO.17.2.021108 (2012).
Duan, C. et al. An endoscopic forward-viewing OCT imaging probe based on a two-axis scanning mems mirror. in Biomedical Imaging (ISBI), 2014 IEEE 11th International Symposium on, 1397–1400 (2014).
El Rifai, O. M. & Youcef-Toumi, K. In American Control Conference, 2001. Proceedings of the 2001. 3251–3255 (IEEE).
Park, H. C. et al. Micromachined tethered silicon oscillator for an endomicroscopic Lissajous fiber scanner. Opt Lett 39, 6675–6678, https://doi.org/10.1364/OL.39.006675 (2014).
Hwang, K., Seo, Y. H., Ahn, J., Kim, P. & Jeong, K. H. Frequency selection rule for high definition and high frame rate Lissajous scanning. Sci Rep-Uk 7, https://doi.org/10.1038/s41598-017-13634-3 (2017).
Park, H. C., Seo, Y. H. & Jeong, K. H. Lissajous fiber scanning for forward viewing optical endomicroscopy using asymmetric stiffness modulation. Opt Express 22, 5818–5825, https://doi.org/10.1364/Oe.22.005818 (2014).
Hwang, K. et al. Fully packaged confocal endomicroscopic system using Lissajous fiber scanner for indocyanine green in-vivo imaging. Proceedings of SPIE BiOS 10497 (2018).
Acknowledgements
This work was financially supported by a grant of the Korean Health Technology R&D Project, Ministry of Health & Welfare, Republic of Korea (HI13C2181), the Global Frontier Project (2011-0031848) of the Korea government, and the Ministry of Science ICT & Future Planning (2018029899).
Author information
Authors and Affiliations
Contributions
Both authors (D.Y.K and K. Hwang) contributed equally to this work. D.Y.K., K. Hwang and K.H. conceived and conducted the experiments, D.Y.K. developed the whole system setup and wrote the manuscript, K.H. developed the fiber scanner and wrote the manuscript, K.-H.J. planned and supervised the project, Y.-H.S. and J.-B.K. analyzed the results, J.A. and P.K. designed the two-photon imaging system, E.K., S.L. and S.J. prepared the mouse experiments, J.-H.Y. and Y.J. provided the commercial two-photon microscope. All authors discussed the results and commented on the manuscript.
Corresponding author
Ethics declarations
Competing Interests
The authors declare no competing interests.
Additional information
Publisher’s note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Kim, D.Y., Hwang, K., Ahn, J. et al. Lissajous Scanning Two-photon Endomicroscope for In vivo Tissue Imaging. Sci Rep 9, 3560 (2019). https://doi.org/10.1038/s41598-019-38762-w
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-019-38762-w
- Springer Nature Limited
This article is cited by
-
Two-Photon Endoscopy: State of the Art and Perspectives
Molecular Imaging and Biology (2023)
-
Novel endoscopic optical diagnostic technologies in medical trial research: recent advancements and future prospects
BioMedical Engineering OnLine (2021)
-
Converting lateral scanning into axial focusing to speed up three-dimensional microscopy
Light: Science & Applications (2020)