Proton MRI can provide detailed morphological images, but it reveals little information about cell homeostasis. On the other hand, sodium MRI can provide metabolic information but cannot resolve fine structures. The complementary nature of proton and sodium MRI raises the prospect of their combined use in a single experiment. In this work, we assessed the repeatability of normalized proton density (PD), T1, T2, and normalized sodium density-weighted quantification measured with simultaneous 3D 1H MRF/23Na MRI in the brain at 7 T, from ten healthy volunteers who were scanned three times each. The coefficients of variation (CV) and the intra-class correlation (ICC) were calculated for the mean and standard deviation (SD) of these 4 parameters in grey matter, white matter, and cerebrospinal fluid. As result, the CVs were lower than 3.3% for the mean values and lower than 6.9% for the SD values. The ICCs were higher than 0.61 in all 24 measurements. We conclude that the measurements of normalized PD, T1, T2, and normalized sodium density-weighted from simultaneous 3D 1H MRF/23Na MRI in the brain at 7 T showed high repeatability. We estimate that changes > 6.6% (> 2 CVs) in mean values of both 1H and 23Na metrics could be detectable with this method.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Introduction
Proton (1H) MRI can provide images that reveal detailed anatomical information in vivo. Furthermore, it allows the measurement of physical properties such as proton density (PD), longitudinal relaxation time (T1), and transversal relaxation time (T2), which can be helpful to reveal and study pathologies1,2,3. Recently, magnetic resonance fingerprinting (MRF)4 made possible the generation of multi-parametric maps (PD, T1, T2, among others) efficiently and precisely in a single scan. While standard 1H MRF generally cannot directly probe the metabolic state of tissue, several recent works tried to address this limitation by incorporating new proton metabolic information, such as single-voxel proton spectroscopy5, chemical exchange saturation transfer (CEST), and semi-solid macromolecule magnetization transfer (MT)6, into MRF protocols. In our method, we propose to assess a new complementary metabolic information related to cellular ionic homeostasis and tissue viability, and that is not directly detectable using 1H MRI or MRS, using a simultaneous acquisition of sodium (23Na) MRI along 1H MRF.
Sodium ions (Na+) play a fundamental role in the human brain, and sodium homeostasis between the intra- and extracellular compartments is tightly coupled with potassium ions (K+) concentrations through Na+/K+-ATPase (sodium–potassium pump) activity7. This pumping process maintains a constant gradient of sodium concentration across the cell membrane, which is used to control cell volume, pH balance, glucose and neurotransmitter transport, membrane electrical potential and pulse transmission, and protect the cells from swelling8. Consequently, variations in intra- and extracellular sodium concentrations reflect important metabolic information that could help with the diagnosis and prognosis of many different pathologies related to dysregulation of ion homeostasis (impairment of Na+/K+-ATPase or ion channels, cell membrane damage), or to energetic processes occurring within the cell and that are required to maintain this ion homeostasis9. However, distinguishing intra- and extracellular sodium concentrations is still very challenging, and in general, most sodium MRI studies aim at detecting variations in the total sodium concentration (TSC) in tissues, which is a weighted average of intra- and extracellular sodium concentrations, or in normalized sodium density-weighted (where a gel or fluid phantom can serve as an external reference, or cerebrospinal fluid or vitreous humor signals are used as stable internal references for sodium signal)8,10,11.
The 23Na nucleus has 100% natural abundance and a spin 3/2, and is therefore MR visible in vivo12. However, it has a low gyromagnetic ratio compared to 1H (1H γ/2π ≈ 42.6 MHz/T versus 23Na γ/2π ≈ 11.3 MHz/T) and the average Na+ concentration in brain tissue is approximately 2,000 times lower than water concentration10,11. Hence, in brain, the sodium MRI signal-to-noise ratio (SNR) is about 20,000 times lower than that of proton MRI13. In practice, these challenges result in low-resolution images and long scan times (due to data averaging) required to increase SNR, and necessitates supplementary high-resolution proton (1H) scans for anatomical reference14.
The idea of simultaneous multinuclear MRI was proposed in 198615, but the first truly simultaneous implementations did not appear until the last decade16,17,18,19,20. Recently, we presented the first multinuclear method that simultaneously acquires sodium images and proton multi-parametric maps (normalized PD, T1, T2, and B1+) based on MRF21,22. The simultaneous acquisition of 1H and 23Na allows for a natural co-registration between images with high-resolution structural information from 1H and images with low-resolution metabolic information from 23Na.
After the development of a novel method, it is fundamental to realize a repeatability study to determine the sensitivity of the method to detect changes over time in longitudinal studies, or between subjects in transversal studies. Due to the unique characteristics of our method: simultaneous acquisition, pulse sequences and k-space sampling (MRF full radial for 1H and MRI center-out radial for 23Na), image reconstruction (dictionary and non-uniform FFT), and MRI equipment (magnet, coils, receiver), this study is still necessary even when previous works already assessed the repeatability of MRF23,24 and sodium MRI25 separately and in different data acquisition circumstances.
In this work, we assessed the repeatability of the quantification of normalized PD, T1, T2, and normalized 23Na density-weighted measured from simultaneous 3D 1H MRF/23Na MRI acquisitions22 in the brain at 7 T.
Results
Figure 1 shows one slice of the 1H maps and 23Na images from the 3 scans of subject 1 after co-registration and masking. Figure 2 shows one selected slice of the 1H maps and 23Na images from the first scan of each subject after co-registration, using the maps from the subject 1 as a reference, and masking. Figure 3 shows the brain segmentation in gray matter (GM), white matter (WM), and cerebrospinal fluid (CSF) of scan 1 for subject 1. Table 1 summarizes the results of the statistical analysis for all tissues and all scans, where Meanall and SDall are the mean value and standard deviation calculated over all the data, Inter-Var is the inter-subject variation, Intra-Var is the mean intra-subject variation, CV is the mean coefficient of variation and ICC is the intra-class variation. Figure 1S in supplementary information shows images from subject 2 along the 3 axes.
As a general result, we can highlight that the mean CV was lower than 6.9% and the ICC was higher than 0.61 for all the 24 statistical results, mean and standard deviation (SD) of all 4 measurements in 3 brain regions.
Normalized PD
As the PD was normalized by the mean CSF intensity, the normalized PD value for the CSF was defined as 1.00. We observed that mean normalized PD over the 30 scans had a meanall ± SDall of 0.87 ± 0.04 for GM and 0.66 ± 0.03 for WM. The CV values were lower than 2.6% for the mean values, and within the range 4.4–4.7% for the SD values. The estimated ICC values were within the range of 0.62–0.74.
T1
The mean T1 over the 30 scans had a meanall ± SDall of 2570 ± 170 ms for CSF, 1450 ± 40 ms for GM and 940 ± 20 ms for WM. The CV values were lower than 2.6% for the mean values, and within the range of 1.5–5.2% for the SD values. ICC values were all within the range 0.78–0.99.
T2
The mean T2 over the 30 scans had a meanall ± SDall of 102 ± 19 ms for CSF, 40 ± 2 ms for GM and 32 ± 1 ms for WM. The CV values were lower than of 3.2% for the mean values, and within the range 1.5–3.4% for the SD values. ICC values were all within the range 0.82–0.98.
Normalized 23Na density-weighted
The mean normalized 23Na density-weighted over the 30 scans had a meanall ± SDall of 0.50 ± 0.04 for CSF, 0.35 ± 0.02 for GM and 0.31 ± 0.02 for WM. The CV values were lower than 3.3% for the mean values, within the range 4.6–6.9% for the SD values. ICC values were all within the range of 0.61–0.88.
Discussion
As already observed in Yu et al.22, the measured mean T1 and T2 values in this study showed discrepancies with other results from the literature26,27,28,29. In particular, the T1 values are approximately 20% lower than usually measured. A possible explanation for this discrepancy could be due to magnetization transfer (MT) effects30, which might be addressed by including MT as an additional dimension in the MRF dictionary. The T2 values were also approximately 20% lower than usually measured. Systematic reductions in T2 values have been reported in many previous MRF implementations29,30. Nonetheless, these discrepancies should not affect the ability of the method to detect intra-subject or inter-subject variations when the same sequence is applied to all subjects.
Due to our specific normalization, it is unfeasible to compare the calculated values from the normalized PD and normalized Na23 density-weighted with the literature. However, the fact that the proton and sodium density of the CSF shows a higher concentration, followed by the GM, and WM is consistent with previous works26,32.
The CVs associated with the mean values were much lower than the CVs obtained for the SD values. This suggests that the mean value is a more sensitive variable to detect changes over time or between subjects. Moreover, the CVs in CSF were higher than the CVs in GM and WM. This can be related with the fact that the CSF is more sensitive to segmentation errors due to the coarse slice thickness and partial volume effects. This behavior was also observed by Leroi et al.26.
The CVs obtained for the mean values of the T1, T2, and normalized PD were in the same range as the CVs measured in previous repeatability studies on 3D MRF methods. For example, Buonincontri et al.24 measured CVs in the range of 07–1.3% for T1, 2.0–7.8% for T2, and 1.4–2.5% for normalized PD for the repeatability of 3D MRF in the healthy human brain at 1.5 T and 3 T. This study also showed the highest CVs for CSF, similarly to our current findings.
In summary, CVs and ICCs showed good to very good results (CV values lower than 6.9% and ICCs values higher than 0.61) over the 24 measurements. All of the variables (mean and SD of normalized PD, T1, T2, normalized 23Na density-weighted in all 3 tissues) could therefore be considered for detecting changes over time in individuals (intra-subject variations) and differences between subjects (inter-subject variations). To be on the conservative side, we can estimate that this method should be able to detect changes greater than the double of the CVs. Considering only the mean value, which is the most sensitive variable, our method should detect variations > 5.2% for PD in GM and WM, > 1.0% for T1 in GM and WM, > 5.2% for T1 in CSF, > 1.2 for T2 in GM and WM, > 6.4% for T2 in CSF, > 5.0% for Na23 density in GM and WM, and > 6.6% for 23Na density in CSF.
We found that simultaneous 3D 1H MRF/23Na MRI is highly repeatable for T1 and T2 measurements, but less repeatable for normalized PD and normalized 23Na density-weighted. This is most likely due to the fact that T1 and T2 were estimated based on the unique shape of signal dynamics (fingerprint), whereas PD and sodium density were estimated from the signal amplitude. Although the method accounts for transmit field (B1+) inhomogeneity, other inhomogeneities from the receive field (B1-) or from B0, which were not corrected in the present study, as well as pre-amplifier gain variations, can still induce non-negligible bias in the signal amplitudes.
Conclusion
In this work, we assessed the repeatability of the mean value and SD of normalized PD, T1, T2, and normalized 23Na density-weighted measurements in GM, WM and CSF, measured from simultaneous 3D 1H MRF/23Na MRI acquisition in 10 different subjects at 7 T (3 scans/subject). We showed that the overall repeatability was deemed very good, where CVs were lower than 6.9% and ICCs were higher than 0.61 in all 24 statistical measurements. We found out that the mean value of the measurements is a more sensitive metric than their respective SD (CV of mean values ≤ 3.3% for all measurements), and that this method should therefore be able to measure changes (inter- and intra-subject variations) > 6.6% (> 2 × CV) in normalized PD, T1, T2, and normalized 23Na density-weighted images. In future works, we will implement the method to study patients with neuropathologies compared to healthy controls in transversal studies, and over time in the same subjects in longitudinal studies.
Materials and methods
Volunteers and scanning protocol
Ten healthy volunteers (5 men, 5 women, mean age = 34.6 ± 10.4 years) were scanned three times in two sessions within a week. In the first session, the volunteers were scanned twice, with a short break between the scans. During the break, the volunteers were asked to move the head position. The B0 shim was recalibrated before the second scan. In the second session, the volunteers were scanned only once. The study was approved by the New York University Grossman School of Medicine institutional review board and was performed in accordance with the relevant guidelines and regulations set forth by the Human Research Protections Program. Informed consent was obtained before each scan session.
MRI hardware
All experiments were performed at 7 T (MAGNETOM, Siemens, Erlangen, Germany) using a 16-channel-Transmit/Receive (8 proton channels + 8 sodium channels) dual-tuned 1H/23Na radiofrequency (RF) coil developed in-house33. An external frequency generator was inserted in the RF cabinet of the system to demodulate the sodium signal with a proper local oscillator34. This modification in the receiver chain allowed simultaneous acquisition of both proton and sodium signals, as described in more details in Yu et al.21.
Pulse sequence
The simultaneous 3D 1H MRF/23Na MRI sequence22 was based on a “stack-of-stars” sampling scheme35. The nuclear spins were sequentially excited every TR (7.5 ms) for 1H and every 2 TRs (15 ms) for 23Na using non-selective pulses followed by one simultaneous readout for both nuclei. The 23Na nuclei were excited every 2 TRs to make sure that a large effective spoiling moment can be obtained for the sodium acquisition part22. The phase encoding gradient moments were distributed such that images from both nuclei had the same slice thickness. The frequency encoding gradient moments were distributed such that a full radial trajectory for 1H and a center-out radial trajectory for 23Na were obtained in k-space, leading to a ratio of ~ 1.9 in in-plane resolution between the 1H and 23Na images. The full radial trajectory was chosen to minimize the effects of B0 inhomogeneities in 1H22. The SAR calculation contemplates both nucleus irradiations. More details about the simultaneous 3D 1H MRF/23Na MRI sequence can be found in Yu et al.22. A diagram of the sequence is shown in Fig. 4.
The 3D simultaneous 1H MRF/23Na MRI sequence parameters were: FOV 240 × 240 × 168 mm3, 1H 160 × 160/23Na 84 × 84 matrix, 1H 1.5 × 1.5 mm2/23Na 2.85 × 2.85 mm2 in-plane resolution, 1H TR = 7.5 ms/23Na TR = 15 ms, 1H TE = 2 ms/23Na TE = 1 ms, 1 slab of 56 slices, 3 mm slice thickness for both 1H and 23Na, 6 shots per slab, total scan time 21 min.
Data processing
The images were reconstructed and processed offline in MATLAB (Mathworks, Natick, MA, USA). The full-radial proton data and center-out sodium data were processed separately.
For proton MRF reconstruction, images were reconstructed with CG-SENSE36 in order to reduce the radial artifacts37. The MRF dictionary was grouped and averaged with the same sliding window as CG-SENSE along the time domain38. The dictionary was computed using the extended phase graph (EPG) formalism39 implemented in C++ 38. Different step sizes were used for T1, T2, and B1+: T1 ranged from 150 to 4347 ms, T2 ranged from 15 to 435 ms, both incremented in steps of 5%; B1+ranged from 10° to 130°, in steps of 1°.
The sodium MRI reconstruction was performed using non-uniform fast Fourier transform (NUFFT)40 from all center-out radial samples combined into one single k-space dataset. A phase correction was applied to remove the phase drift between the MR system and the external frequency generator22.
All the images from the 3 scans for each volunteer were segmented in grey matter (GM), white matter (WM), and cerebrospinal fluid (CSF) tissue compartments with SPM 12 (UCL, London, UK)41. Tissue segmentation was performed using the PD, T1, and T2 maps of each scan as input images. The tissue probability map outputs were then normalized and binarized with a threshold of 0.9 to minimize the number of pixels with multiple tissue components to generate non-overlapping GM, WM and CSF masks.
In order to realize a quantitative analysis, we normalized the proton and sodium density-weighted images. The PD map was normalized by the mean intensity of the CSF measured over the pixels of the CSF binarized mask to minimize partial volume effects (as described previously). The 23Na density-weighted was normalized by the mean sodium intensity in the eyes (vitreous humor), which exhibited the maximum signal intensity in the sodium images. A manual ROI over the whole image volume that contains only the eyes was applied to calculate the mean eye intensity. The final sodium density-weighted image is therefore a normalized sodium density-weighted image that is proportional to some extent to the TSC. Nevertheless, this image should not be confounded with a TSC map since the effect of sodium relaxation times in different tissues were not measured nor mitigated by the acquisition (TE = 1 ms, TR = 15 ms, FA = 30°) in this case.
Statistical analysis
As a first step in the statistical analysis, the masks were applied to all proton and sodium maps to calculate the mean value and the standard deviation of the normalized PD, T1, T2, and normalized 23Na density-weighted for each tissue (WM, GM and CSF) and each scan of each subject. We defined meanall and SDall as the mean and SD over all the data for each measurement respectively. We then calculated the mean intra-subject variance (the mean value of the variance between results from scans of the same subject) and the inter-subject variance (the variance between results from different subjects) of each measurement (mean and SD), see Eqs. 1 and 2.
where \(Intra\_var\left(a\right)\) is the mean intra-subject variance for the measurement \(a\), \(i\) represents a subject (i = 1 to N, with N = 10 subjects in this study), \({var}_{i}(a)\) is the variance of the measurement \(a\) for subject \(i\), and \(Inter\_var\left(a\right)\) is the inter-subject variance for the measurement \(a\), \({mean}_{i}(a)\) is the mean value of the measurement \(a\) for subject \(i\), and \(var\) is the variance among all the subjects.
Finally, we computed the coefficient of variation (CV, in %) as expressed in Eq. 342, and the intra-class correlation (ICC) as the inter-subject variance (inter-var) divided by the sum of the intra and inter variances43, see Eq. 4.
where \(CV\left(a\right)\) is the mean coefficient of variation for the measurement \(a\) expressed as percentage, \({SD}_{i}(a)\) is the standard deviation of the measurement \(a\) for subject \(i\), and \(ICC\left(a\right)\) is the intra-class correlation for the measurement \(a\).
The CV is considered an indicator of the utility of a measure for detecting within-subject changes over time. An ideal set of measurements has a CV equal to 0%. The ICC is a measure of the repeatability of the method. It has values between 0 and 1, where higher values are associated with more repeatable measurements. From the literature44, we can interpret that CV was regarded as very good if CV ⩽ 10%, good if 10% < CV ⩽ 20%, moderate if CV > 20%, and poor if CV > 30%. On the other hand, the ICC was regarded as very good if ICC ⩾ 0.8, good if 0.6 ⩽ ICC < 0.8, fair/moderate if 0.4 ⩽ ICC < 0.6, and poor if ICC < 0.4.
Data availability
The MRI datasets in this study are available upon request to the corresponding author. All measurements of normalized PD, T1, T2 and normalized sodium density weighted data in GM, WM and CSF, for all scans and all subjects, are included in Supplementary material.
References
Larsson, H. B. W. et al. Assessment of demyelination, edema, and gliosis by in vivo determination of T1 and T2 in the brain of patients with acute attack of multiple sclerosis. Magn. Reson. Med. 11, 337–348 (1989).
Williamson, P. et al. Frontal, temporal, and striatal proton relaxation times in schizophrenic patients and normal comparison subjects. Am. J. Psychiatry. 149, 549–551 (1992).
Pitkanen, A. et al. Severity of hippocampal atrophy correlates with the prolongation of MRI T2 relaxation time in temporal lobe epilepsy but not in Alzheimer’s disease. Neurology 46, 1724–1730 (1996).
Ma, D. et al. Magnetic resonance fingerprinting. Nature 495, 187–193 (2013).
Kulpanovihc, A. & Tal, A. The application of magnetic resonance fingerprinting to single voxel proton spectroscopy. NMR Biomed. 31, 4001 (2018).
Pelrman, O., Farrar, C. T. & Heo, H. MR fingerprinting for semisolid magnetization transfer and chemical exchange saturation transfer. NMR Biomed. 9, 4710 (2022).
Rose, A. & Valdes, R. Understanding the sodium pump and its relevance to disease. Clin. Chem. 40, 1674–1685 (1994).
Ouwerkerk, R. Sodium MRI. Methods Mol. Biol. 711, 175–201 (2011).
Erecinska, M., Cherian, S. & Silver, I. A. Energy metabolism in mammalian brain during development. Prog. Neuro Gibol. 73, 397–445 (2004).
Madelin, G. & Regatte, R. R. Biomedical applications of sodium MRI in vivo. JMagn. Reson. Imag. 38, 511–529 (2013).
Madelin, G., Lee, J. S., Regatte, R. R. & Jerschow, A. Sodium MRI: Methods and applications. Prog. Nucl. Magn. Reson. Spectrosc. 79, 14–47 (2014).
Berendsen, H. J. C. & Edzes, H. T. The observation and general interpretation of sodium magnetic resonance in biological material. Ann. NY Acad. Sci. 204, 459–485 (1973).
Rooney, W. D. & Springer, C. S. Jr. The molecular environment of intracellular sodium: 23Na NMR relaxation. NMR Biomed. 4, 227–245 (1991).
Ouwerkerk, R. Sodium magnetic resonance imaging: from research to clinical use. J. Am. Coll. Radiol. 4, 739–741 (2007).
Lee, S. W., Hilal, S. K. & Cho, Z. H. A multinuclear magnetic resonance imaging technique simultaneous proton and sodium imaging. Magn. Reson. Imag. 4, 343–350 (1986).
Keupp, J. et al. Simultaneous dual-nuclei imaging for motion corrected detection and quantification of 19F imaging agents. Magn. Reson. Med. 66, 1116–1122 (2011).
Gordon, J. W. et al. Simultaneous imaging of 13C metabolism and 1H structure: technical considerations and potential applications. NMR Biomed. 28, 576–582 (2015).
de Bruin, P. W. et al. Time-efficient interleaved human 23Na and 1H data acquisition at 7 T. NMR Biomed. 28, 1228–1235 (2015).
Meyerspeer, M. et al. Simultaneous and interleaved acquisition of NMR signals from different nuclei with a clinical MRI scanner. Magn. Reson. Med. 76, 1636–1641 (2016).
Kaggie, J. D. et al. Synchronous radial 1H and 23Na dual-nuclear MRI on a clinical MRI system, equipped with a broadband transmit channel. Concepts Magn. Reson. Part B Magn. Reson. Eng. 46, 191–201 (2016).
Yu, Z., Madelin, G., Sodickson, D. K. & Cloos, M. A. Simultaneous proton magnetic resonance fingerprinting and sodium MRI. Mag. Res. Med. 83, 2232–2242 (2020).
Yu, Z. et al. Simultaneous 3D acquisition of 1H MRF and 23Na MRI. Magn. Res. Med. 00, 1–14 (2021).
Jiang, Y. et al. Repeatability of magnetic resonance fingerprinting T1 and T2 estimates assessed using the ISMRM/NIST MRI system phantom. Magn. Res. Med. 78, 1452–1457 (2017).
Buonincontri, G. et al. Three dimensional MRF obtains highly repeatable and reproducible multi-parametric estimations in the healthy human brain at 15T and 3T. Neuroimage 226, 117573 (2021).
Riemer, F. et al. Measuring tissue sodium concentration: Cross-vendor repeatability and reproducibility of 23Na-MRI across two sites. JMagn Reson. Imag. 50, 1278–1284 (2019).
Leroi, L. et al. Simultaneous proton density, T1, T2, and flip-angle mapping of the brain at 7 T using multiparametric 3D SSFP imaging and parallel-transmission universal pulses. Mag. Reson. Med. 84, 3286–3299 (2020).
Marques, J. P. et al. MP2RAGE, a self-bias-field corrected sequence for improved segmentation and T1-mapping at high field. Neuroimage 49, 1271–1281 (2010).
Rooney, W. D. et al. Magnetic field and tissue dependencies of human brain longitudinal 1H2O relaxation in vivo. Magn. Reson. Med. 57, 308–318 (2007).
Emmerich, J. et al. Rapid and accurate dictionary-based T2 mapping from multi-echo turbo spin echo data at 7 Tesla. JMagn Reson Imaging. 49, 1253–1262 (2019).
Hilbert, T. et al. Magnetization transfer in magnetic resonance fingerprinting. Magn. Reson Med. 84, 128–141 (2020).
Bustin, A. et al. Highdimensionality undersampled patch-based reconstruction (HD-PROST) for accelerated multi-contrast MRI. Magn. Reson. Med. 81, 3705–3719 (2019).
Lu, A., Atkinson, I. C., Claiborne, T. C., Damen, F. C. & Thulborn, K. R. Quantitative Sodium Imaging With a Flexible Twisted Projection Pulse Sequence. Magn. Reson Med. 63, 1583–1593 (2010).
Wang, B. et al. A radially interleaved sodium and proton coil array for brain MRI at 7T. NMR Biomed. e4608 (2021).
Meyerspeer, M. et al. Simultaneous and interleaved acquisition of NMR signals from different nuclei with a clinical MRI scanner. Magn. Reson. Med. 76, 1636–1641 (2016).
Block, K. T. et al. Towards routine clinical use of radial stack-of-stars 3D gradient-echo sequences for reducing motion sensitivity. J. Kor. Soc. Magn. Reson. Med. 18, 87–106 (2014).
Pruessmann, K. P., Weiger, M., Börnert, P. & Boesiger, P. Advances in sensitivity encoding with arbitrary k-space trajectories. Magn. Reson. Med. 46, 638–651 (2001).
Kara, D. et al. Parameter map error due to normal noise and aliasing artifacts in MR fingerprinting. Magn. Reson. Med. 81, 3108–3123 (2019).
Cloos, M. A. et al. Multiparametric imaging with heterogeneous radiofrequency fields. Nat. Commun. 7, 12445 (2016).
Weigel, M. Extended phase graphs: dephasing, RF pulses, and echoes-pure and simple. JMagn Reson Imaging. 41, 266–295 (2015).
Fessler, J. A. & Sutton, B. P. Nonuniform Fast Fourier Transforms Using Min-Max Interpolation. IEEE Trans Sign Process. 51, 560–574 (2003).
Ashburner, J. et al. SPM12 Manual The FIL Methods Group (and honorary members), 2015.
Brown, C. E. Coefficient of Variation. In: Applied Multivariate Statistics in Geohydrology and Related Sciences. Springer, Berlin, Heidelberg.1998.
Wolak, M., Fairbairn, D. & Paulsen, Y. Guidelines for estimating repeatability. Methods Ecol Evol. 3, 129–137 (2012).
Fleiss, J., Design and analysis of clinical experiments. John Wiley & Sons, 73, (2011).
Acknowledgements
The research reported in this publication was supported by the NIH/NIBIB grant R01 EB026456, and performed under the rubric of the Center for Advanced Imaging Innovation and Research, a NIBIB Biomedical Technology Resource Center (P41 EB017183).
Author information
Authors and Affiliations
Contributions
G.G.R. acquired and processed the images, performed the statistical analysis and wrote the manuscript, Z.Y. developed the pulse sequence and acquired the images, L.F.O. acquired the images, L.C. recruited the volunteers, M.A.C designed the sequence and experiment, and G.M. designed the sequence and experiment, and wrote the manuscript. All the authors revised the article and approved the final version.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Rodriguez, G.G., Yu, Z., O′Donnell, L.F. et al. Repeatability of simultaneous 3D 1H MRF/23Na MRI in brain at 7 T. Sci Rep 12, 14156 (2022). https://doi.org/10.1038/s41598-022-18388-1
Received:
Accepted:
Published: