Abstract
Aim of this study was to assess the impact of virtual monoenergetic images (VMI) in combination and comparison with iterative metal artifact reduction (IMAR) on hip prosthesis-associated artifacts in photon-counting detector CT (PCD-CT). Retrospectively, 33 scans with hip prosthesis-associated artifacts acquired during clinical routine on a PCD-CT between 08/2022 and 09/2022 were analyzed. VMI were reconstructed for 100–190 keV with and without IMAR, and compared to polychromatic images. Qualitatively, artifact extent and assessment of adjacent soft tissue were rated by two radiologists using 5-point Likert items. Quantitative assessment was performed measuring attenuation and standard deviation in most pronounced hypodense and hyperdense artifacts, artifact-impaired bone, muscle, vessels, bladder and artifact-free corresponding tissue. To quantify artifacts, an adjusted attenuation was calculated as the difference between artifact-impaired tissue and corresponding tissue without artifacts. Qualitative assessment improved for all investigated image reconstructions compared to polychromatic images (PI). VMI100keV in combination with IMAR achieved best results (e.g. diagnostic quality of the bladder: median PI: 1.5 (range 1–4); VMI100keV+IMAR: 5 (3–5); p < 0.0001). In quantitative assessment VMI100keV with IMAR provided best artifact reduction with an adjusted attenuation closest to 0 (e.g. bone: PI: 302.78; VMI100keV+IMAR: 51.18; p < 0.0001). The combination of VMI and IMAR significantly reduces hip prosthesis-associated artifacts in PCD-CT and improves the diagnostic quality of surrounding tissue.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Introduction
Hip replacements, ranging among the most common orthopedic prostheses, typically lead to extensive artifacts in CT and therefore pose a common challenge in diagnostic imaging1,2. The severity of artifacts is influenced by image acquisition, image reconstruction parameters and composition of the metal implants3. Optimization of acquisition protocols and reconstruction parameters with use of high tube voltage, high tube current, narrow collimation and increased slice thickness allows for significant reduction of artifacts, however frequently does not suffice3. To enable diagnostic readability of artifact-impaired images further artifact reduction is needed.
Dual-energy CT (DECT) has been proven to reduce artifacts caused by arthroplasty4. Additionally, DECT-derived virtual monoenergetic images (VMI) have been shown to reduce artifact extent, however, lead to reduced contrast at higher keV5,6. Still, high keV VMI offer only limited artifact reduction in larger implants with the concurrent disadvantage of reduced contrast. Another possible adaption for reduction of metal artifacts are iterative metal artifact reduction algorithms (IMAR)7,8. Especially the combination of VMI and IMAR showed excellent results in previous DECT studies9. In 2021 the first PCD-CT was licensed for clinical use10. With direct conversion of every single photon into energy signals and the omittance of scintillators and detector septa, PCD-CT achieves higher spatial resolution, reduced radiation dose and less artifacts compared to energy-integrating detector CT (EID-CT)11,12. Especially high threshold images with tin filtration offer the prospect of substantial artifact reduction13,14.
Despite these promises, there are no clinical studies on the performance of VMI and IMAR in PCD-CT. Therefore, the aim of our study was to assess the impact of VMI and IMAR on image quality improvement in PCD-CT with hip prostheses associated artifacts.
Materials and methods
The local institutional review board of the University Hospital Bonn approved this study (Ethics Committee of the Medical Faculty of the University of Bonn; reference number 426/22). The Ethics Committee of the Medical Faculty of the University of Bonn waived the need of informed consent. All scans were acquired during clinical routine and no scan was performed solely for research purposes. The study was conducted in accordance with the Declaration of Helsinki and its amendments.
Patients with hip implants and associated artifacts receiving abdominal/pelvic scan between 08/2022 and 09/2022 were included in this retrospective, single-center study. Scans were performed on a clinical PCD-CT (NAEOTOM Alpha, Siemens Healthcare GmbH, Erlangen, Germany). Inclusion criteria included the presence of hip prostheses and the acquisition of a spectral post-processing (SPP) enabled image data set with standard protocol as described below. Scans were performed for various indications.
Imaging protocol
A weight-adapted volume of iodine based contrast agent (Accupaque 300 mg/ml, GE Healthcare Buchler GmbH & Co. KG, Braunschweig, Germany) was applied intravenously with a flow rate of 3 ml/s followed by a bolus of 40 ml of physiologic saline solution.
Scan parameters were a tube voltage of 120 kVp with activated automatic tube current modulation, a pitch of 0.8 and a gantry rotation time of 0.5 s. Detector collimation was 144 × 0.4 mm. Reconstruction parameters were 1 mm slice thickness with an increment of 0.7 mm. Scans were performed in a head-first supine position. A regular kernel (QR40; Siemens Healthcare GmbH, Erlangen, Germany) and Quantum Iterative Reconstruction (QIR Level 3; Siemens Healthcare GmbH, Erlangen, Germany) was used for image reconstruction.
VMI were reconstructed in axial view for 100–190 keV in an interval of 10 keV using dedicated software (syngo.via VB 60, Monoenergetic Plus; Siemens Healthcare GmbH, Erlangen, Germany). Iterative metal artifact reduction optimized for hip prosthesis (iMAR, Siemens Healthcare GmbH, Erlangen, Germany) is commercially available and was used as supplied by the vendor.
Quantitative image analysis
Quantitative assessment of polychromatic and virtual monochromatic images was performed by region of interest (ROI) based attenuation analysis. ROIs were placed in the most pronounced hypodense and hyperdense artifacts using a conventional clinical DICOM viewer (Deep Unity R20 XX; Dedalus HealthCare GmbH, Bonn, Germany). Furthermore, values and standard deviation of X-ray attenuation was evaluated in muscle tissue, bone and bladder tissue with and without presence of artifacts. Hereby, differences in attenuation and gradient of attenuation of tissue were measured and an adjusted attenuation and image noise assessed as proposed previously15. Adjusted attenuation was assessed to differentiate artifact reduction from regular changes in Hounsfield units (HU) for differing VMI energy levels. Adjusted attenuation was calculated as the difference of tissue impaired by artifacts and without artifact impairment. Image noise is higher in images with presence of artifacts and VMI of high keV, therefore we calculated an adjusted image noise using the difference of standard deviation between muscle tissue in HU values in areas with and without artifacts. Adjusted attenuation of 0 indicates optimal artifact reduction, values above 0 indicate insufficient artifact reduction and values below 0 an overcorrection of the artifact correction. All measurements were performed for polychromatic and virtual monochromatic images from 100 to 190 keV in steps of 10 keV.
Qualitative image analysis
Two radiologists with two (YCL) and eleven (DK) years of experience in abdominopelvic CT evaluated the CT images independently regarding artifact extent of hyperdense and hypodense artifacts as well as assessment of surrounding tissue using a five-point Likert grading scale. Rating of artifacts was defined as follows: (1) excessive artifacts; (2) pronounced artifacts; (3) moderate artifacts; (4) minor artifacts; and (5) artifacts are absent. For assessment of diagnostic quality of bone, muscle, vessels, bladder and soft tissue, the following Likert scale was used: (1) highly restricted diagnostic interpretability; (2) restricted diagnostic interpretability; (3) moderate diagnostic interpretability; (4) minor restrictions on diagnostic interpretability; and (5) unrestricted diagnostic interpretability. Additionally, both readers had to select the reconstruction with the best diagnostic quality for each patient. Polychromatic images as well as VMI with 100 keV, 130 keV, 160 keV and 190 keV were rated.
Statistical analysis
All statistical analyses were conducted using IBM SPSS Version 27 (IBM Corp., Armonk, NY, USA). Graphs were carried out using the software GraphPad PRISM Version 6.02 (GraphPad Software, San Diego, CA, USA). Quantitative results are stated as mean and standard deviation. Wilcoxon signed-rank test was used for statistical analysis of quantitative image parameters. Qualitative results are expressed as median with interquartile range (IQR). Interrater reliability was assessed using the intraclass correlation coefficient (ICC). ICC estimates and their 95% confident intervals (CI) were calculated based on a mean-rating (k = 2), consistency, two-way mixed-effects model. p values below 0.05 were considered significant.
Results
Participant characteristics
Between 08/2022 and 09/2022, a total of 130 patients received scans of the abdominal/pelvic region on a PCD-CT at our institute. 97 of these patients were excluded from analysis as they had no hip implants or were missing a complete SPP image data set (Fig. 1). Overall, 33 patients were included in the analysis (17 female, 51.52%) with an average age of 76 (range 61–91) years. Mean DLP was 515.58 mGy*cm and mean CTDIvol was 7.78 mGy.
Quantitative image analysis
Results of quantitative analysis are summarized in Supplemental Information 1 and Fig. 2; Fig. 3 shows typical reconstructions of a pelvic CT with unilateral hip replacement and Fig. 4 shows representative images of a pelvic CT with bilateral hip replacement using VMI and IMAR. VMI100keV+IMAR showed the best adjusted attenuation overall compared to PI with the closest value to 0 (Overall PI: 634.66; VMI100keV+IMAR: 102.81; p < 0.0001). Mean attenuation of hyperattenuating artifacts lowered in VMI compared to polychromatic images (PI: 661.58 ± 159.78; VMI190keV: 56.79 ± 96.82; p < 0.00001), in particular for the combination VMI and IMAR (VMI190keV+IMAR: 6.23 ± 69.91; p < 0.0001) (Supplemental Information 1).
Hypoattenuating artifacts showed an elevation of HU values in VMI (PI: − 424.97 ± 79.16; VMI190keV: − 71.63 ± 75.53; p = 0.0003), with increasingly higher values for higher keV reconstructions. The combination of VMI and IMAR showed most extensive artifact reduction with positive CT numbers for hypodense artifacts (VMI100keV+IMAR: 10.69 ± 27.12; VMI190keV+IMAR: 18.83 ± 26.35; p < 0.0001).
Adjusted image noise decreased for all reconstructions with the lowest value for VMI110keV+IMAR (PI: 48.15; VMI110keV+IMAR: 8.37; p < 0.0001), as stated in Table 1. Mean ROI size was 50.34 mm2.
Qualitative image analysis
Qualitative rating results are summarized in Supplemental Information 2. Qualitative assessment of muscle (M), bones (BO), bladder (BL) and iliac vessels (IV) significantly improved compared to conventional images (median PI: M 1 (1–3), BO 1 (1–3), BL 1.5 (1–4); IV 2 (1–4); median VMI100keV+IMAR: M 4 (3–5), BO 4 (3–5), BL 5 (3–5); IV 5 (2–5); p < 0.05; Supplemental Information 2). Hyperdense artifacts were subjectively reduced, scoring a 5 (3–5, p < 0.05) for all reconstructions combining VMI and IMAR. For the PI the score was 1 (1–3) and VMI190keV scored 2 (1–4; p < 0.05). For hypodense artifacts, PI scored 1 (1–2), VMI190keV scored 2 (1–4; p < 0.05) and combination of VMI100keV and IMAR scored 5 (3–5; p < 0.05). In many cases especially hypodense artifacts adjacent to the prosthesis increased with higher keV VMI (Fig. 5). Rating results for all criteria by each rater are shown in Fig. 6. Overall, raters favored VMI100keV+IMAR in 62.12% of all scans (41/66) and VMI130keV+IMAR in 28.79% of the examinations (19/66).
Interrater agreement was excellent with an overall ICC value of 0.944 [95% confidence interval (CI) 0.939–0.949]. Respective ICC was 0.963 (95% CI 0.955–0.970) for the extent of hypodense artifacts, 0.956 (95% CI 0.946–0.965) for the extent of hyperdense artifacts, 0.953 (95% CI 0.942–0.962) for assessment of muscle tissue, 0.976 (95% CI 0.970–0.981) for bone, 0.915 (95% CI 0.895–0.931) for assessment of the bladder and 0.909 (95% CI 0.889–0.926) for assessment of the iliac vessels.
Discussion
This study evaluates PCD-CT artifact reduction capabilities regarding hip implants in clinical imaging. The main findings are that VMI100keV in combination with IMAR deliver best results regarding artifact-reduction and diagnostic quality. Every combination of VMI and IMAR led to artifact reduction compared to conventional polychromatic images, with low keV VMI images demonstrating best contrast, especially for vascular delineation in scans with iodine contrast.
As hip arthroplasty and associated revision surgery procedures are projected to increase, there is a growing need for precise, fast and artifact-free imaging including dedicated scans for evaluation of hip arthroplasty-associated pathologies16. PCD-CT is an emerging technology with numerous potential advantages compared with EID-CT based on the direct detection and weighting of every photon17,18. Employing energy-thresholds allows for elimination of image background noise; virtual higher-energy image reconstructions are less affected by artifacts typically originating from lower energy photons19. High photon sensitivity is a further inert advantage of PCD-CT; diagnostic image quality is achievable with lower tube current, than necessary with EID CT.
Several studies have investigated the effects of VMI and IMAR for reduction of artifacts in dual energy EID-CT9,20,21,22,23,24,25,26. A recent study investigating split-filter EID-CT with images reconstructed at 120 kVp-equivalent, reported that PI images in combination with IMAR leads to best artifact reduction15. In contrast, most previous EID-CT studies reported best artifact reduction for high-energy VMI in combination with IMAR23,24,25,26.
Even though this study did not compare PCD-CT to EID-CT, images achieved were of high diagnostic quality and therefore offer a feasible solution for patients with hip prosthesis suspected of pathologies masked by implant-associated artifacts (Fig. 7).
As signal acquisition of PCD-CT differs from dual energy EID-CT, with PCD-CT technology enabling imaging with lower noise and higher spatial resolution, lower energy PCD-CT VMI images are expected to have less noise and higher image quality. In fact, in PCD-CT artifact extent remains relatively stable between 120 and 190 keV, with 190 keV images showing severely reduced contrast; Images reconstructed at 100 keV combined with IMAR resulted in best artifact reduction and showed superior contrast compared to high-keV images.
A recent phantom study on PCD-CT reported best artifact reduction for 100 keV at scans performed with 120 kVp, which is in line with the results of this study27. As recently reported for EID-CT, in PCD-CT IMAR is superior to VMI for the reduction of extensive artifacts but may lead to distortion and blurring. VMI effectively reduces minor artifacts without distorting images15,23.
Laukamp et al. reported optimal keV reconstructions for artifact reduction in EID-CT depending on the affected tissue; 200 keV for bone, 140 keV for soft tissue9. This could not be confirmed for PCD-CT in the current study. Quantitatively, there were minor, insignificant deviations of adjusted attenuation for muscle and bladder towards VMI with higher keV. Nevertheless, VMI100 keV in combination with IMAR led to optimal results for both quantitative and qualitative assessment of artifact reduction and image quality.
As reported previously15,23,28, combining IMAR and high keV VMI led to the appearance of new hypodense artifacts adjacent to implants, suggesting a local overcorrection of artifacts. These new artifacts intensify with higher keV VMI and thereby limit diagnostic assessment of surrounding tissue.
There are limitations to the study. Study design was monocentric and therefore only a small cohort of patients was retrospectively included. No distinction was made between unilateral and bilateral hip prostheses. Furthermore, prosthesis composition was not taken into account. As stated before, based on the material and shape varying artifact extent can be assumed3,29. Further multi-center evaluation on larger cohorts, especially focusing on patients needing a hip prosthesis revision should be part of further subsequent studies.
Overall, the current results show that the combination of IMAR and VMI are the most potent tool for reduction of even extensive hip prosthesis associated artifacts in PCD-CT. The combination of IMAR and VMI improves diagnostic readability and thus may enable the detection of pathologies that would otherwise be concealed by artifacts in conventional CT images.
Data availability
The anonymized datasets generated during and analyzed during the current study are available from the corresponding author on reasonable request. Due to local privacy laws, CT images can not be provided as theoretically there is a risk of identification of personal information in pseudo-anonymized CT datasets.
References
Ferguson, R. J. et al. Hip replacement. Lancet (London, England) 392, 1662–1671 (2018).
Schwartz, A. M., Farley, K. X., Guild, G. N. & Bradbury, T. L. Projections and epidemiology of revision hip and knee arthroplasty in the United States to 2030. J. Arthroplasty 35, S79-85 (2020).
Lee, M.-J. et al. Overcoming artifacts from metallic orthopedic implants at high-field-strength MR imaging and multi-detector CT. Radiographics 27, 791–803 (2007).
Kasparek, M. F. et al. Dual-energy CT and ceramic or titanium prostheses material reduce CT artifacts and provide superior image quality of total knee arthroplasty. Knee Surg. Sports Traumatol. Arthrosc. ESSKA 27, 1552–1561 (2019).
Neuhaus, V. et al. Metal artifact reduction by dual-layer computed tomography using virtual monoenergetic images. Eur. J. Radiol. 93, 143–148 (2017).
Wellenberg, R. H. H. et al. Quantifying metal artefact reduction using virtual monochromatic dual-layer detector spectral CT imaging in unilateral and bilateral total hip prostheses. Eur. J. Radiol. 88, 61–70 (2017).
Neroladaki, A. et al. Metallic artifact reduction by evaluation of the additional value of iterative reconstruction algorithms in hip prosthesis computed tomography imaging. Medicine 98, e14341 (2019).
Bongers, M. N. et al. Comparison and combination of dual-energy- and iterative-based metal artefact reduction on hip prosthesis and dental implants. PLoS One 10, e0143584 (2015).
Laukamp, K. R. et al. CT metal artifacts in patients with total hip replacements: For artifact reduction monoenergetic reconstructions and post-processing algorithms are both efficient but not similar. Eur. Radiol. 28, 4524–4533 (2018).
McCarthy, L.-J. FDA Clears First Major Imaging Device Advancement for Computed Tomography in Nearly a Decade (2021).
Flohr, T. et al. Photon-counting CT review. Phys. Med. 79, 126–136 (2020).
Willemink, M. J., Persson, M., Pourmorteza, A., Pelc, N. J. & Fleischmann, D. Photon-counting CT: Technical principles and clinical prospects. Radiology 289, 293–312 (2018).
Zhou, W. et al. Reduction of metal artifacts and improvement in dose efficiency using photon-counting detector computed tomography and tin filtration. Invest. Radiol. 54, 204–211 (2019).
Lee, C.-L. et al. Metal artifact reduction and tumor detection using photon-counting multi-energy computed tomography. PLoS One 16, e0247355 (2021).
Wichtmann, H. M. et al. Metal implants on abdominal CT: Does split-filter dual-energy CT provide additional value over iterative metal artifact reduction?. Abdominal Radiol. (New York) 48, 424–435 (2023).
Mealy, A. & Sorensen, J. Effects of an aging population on hospital costs related to elective hip replacements. Public Health 180, 10–16 (2020).
Baffour, F. I. et al. Photon-counting detector CT for musculoskeletal imaging: A clinical perspective. Am. J. Roentgenol. https://doi.org/10.2214/AJR.22.28418 (2022).
Kijowski, R. & Fritz, J. Emerging technology in musculoskeletal MRI and CT. Radiology 306, 6–19 (2023).
Leng, S. et al. Photon-counting detector CT: System design and clinical applications of an emerging technology. Radiographics 39, 729–743 (2019).
Schmidt, A. M. A. et al. Combination of iterative metal artifact reduction and virtual monoenergetic reconstruction using split-filter dual-energy CT in patients with dental artifact on head and neck CT. Am. J. Roentgenol. 218, 716–727 (2022).
Higashigaito, K., Angst, F., Runge, V. M., Alkadhi, H. & Donati, O. F. Metal artifact reduction in pelvic computed tomography with hip prostheses: Comparison of virtual monoenergetic extrapolations from dual-energy computed tomography and an iterative metal artifact reduction algorithm in a phantom study. Invest. Radiol. 50, 828–834 (2015).
Kanani, A., Yazdi, M., Owrangi, A. M., Karbasi, S. & Mosleh-Shirazi, M. A. Metal artifact reduction in cervix brachytherapy with titanium applicators using dual-energy CT through virtual monoenergetic images and an iterative algorithm: A phantom study. Brachytherapy 21, 933–942 (2022).
Yoo, H. J., Hong, S. H., Choi, J.-Y. & Chae, H. D. Comparison of metal artifact reduction algorithms in patients with hip prostheses: Virtual monoenergetic images vs orthopedic metal artifact reduction. J. Korean Soc. Radiol. 83, 1286–1297 (2022).
Selles, M. et al. Quantitative analysis of metal artifact reduction in total hip arthroplasty using virtual monochromatic imaging and orthopedic metal artifact reduction, a phantom study. Insights Imaging 12, 171 (2021).
Park, J., Kim, S. H. & Han, J. K. Combined application of virtual monoenergetic high keV images and the orthopedic metal artifact reduction algorithm (O-MAR): Effect on image quality. Abdom. Radiol. (New York) 44, 756–765 (2019).
Neuhaus, V. et al. Reducing artifacts from total hip replacements in dual layer detector CT: Combination of virtual monoenergetic images and orthopedic metal artifact reduction. Eur. J. Radiol. 111, 14–20 (2019).
Anhaus, J. A., Schmidt, S., Killermann, P., Mahnken, A. & Hofmann, C. Iterative metal artifact reduction on a clinical photon counting system-technical possibilities and reconstruction selection for optimal results dependent on the metal scenario. Phys. Med. Biol. 67, 25 (2022).
Shim, E. et al. Metal artifact reduction for orthopedic implants (O-MAR): Usefulness in CT evaluation of reverse total shoulder arthroplasty. Am. J. Roentgenol. 209, 860–866 (2017).
Huber, F. A. et al. Comparison of different CT metal artifact reduction strategies for standard titanium and carbon-fiber reinforced polymer implants in sheep cadavers. BMC Med. Imaging 21, 29 (2021).
Funding
Y. C. L. is supported by a research grant from Siemens Healthcare GmbH. Open Access funding enabled and organized by Projekt DEAL.
Author information
Authors and Affiliations
Contributions
D.K. and Y.C.L. created the study concept. Y.C.L. and D. K. wrote the main manuscript text. Y.C.L. prepared the figures. Y.C.L. provided data collection. D.K. and Y.C.L. provided samples. Y.C.L. performed statistical analyses. U.A., N.M., P.A.K., A.I., J.A.L., and T.D. reviewed the manuscript.
Corresponding author
Ethics declarations
Competing interests
Y.C.L. is supported by a research grant from Siemens Healthcare. The other authors declare no conflicts of interest.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Layer, Y.C., Mesropyan, N., Kupczyk, P.A. et al. Combining iterative metal artifact reduction and virtual monoenergetic images severely reduces hip prosthesis-associated artifacts in photon-counting detector CT. Sci Rep 13, 8955 (2023). https://doi.org/10.1038/s41598-023-35989-6
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-023-35989-6
- Springer Nature Limited
This article is cited by
-
Influence of helical pitch and gantry rotation time on image quality and file size in ultrahigh-resolution photon-counting detector CT
Scientific Reports (2024)
-
Photon-counting detector computed tomography for metal artifact reduction: a comparative study of different artifact reduction techniques in patients with orthopedic implants
La radiologia medica (2024)
-
Photon-Counting Computed Tomography for Microstructural Imaging of Bone and Joints
Current Osteoporosis Reports (2024)