Abstract
Muscle strength assessments are vital in rehabilitation, orthopedics, and sports medicine. However, current methods used in clinical settings, such as manual muscle testing and hand-held dynamometers, often lack reliability, and isokinetic dynamometers (IKD), while reliable, are not easily portable. The aim of this study was to design and validate a wearable dynamometry system with high accessibility, accuracy, and reliability, and to validate the device. Therefore, we designed a wearable dynamometry system (WDS) equipped with knee joint torque sensors. To validate this WDS, we measured knee extension and flexion strength in 39 healthy adults using both the IKD and WDS. Comparing maximal isometric torque measurements, WDS and IKD showed strong correlation and good reliability for extension (Pearson’s r: 0.900; intraclass correlation coefficient [ICC]: 0.893; standard error of measurement [SEM]: 9.85%; minimal detectable change [MDC]: 27.31%) and flexion (Pearson’s r: 0.870; ICC: 0.857; SEM: 11.93%; MDC: 33.07%). WDS demonstrated excellent inter-rater (Pearson’s r: 0.990; ICC: 0.993; SEM: 4.05%) and test–retest (Pearson’s r: 0.970; ICC: 0.984; SEM: 6.15%) reliability during extension/flexion. User feedback from 35 participants, including healthcare professionals, underscores WDS's positive user experience and clinical potential. The proposed WDS is a suitable alternative to IKD, providing high accuracy, reliability, and potentially greater accessibility.
Similar content being viewed by others
Introduction
Muscle strength is an important clinical outcome in rehabilitation, orthopedics, and sports medicine. Muscular weakness independently predicts all-cause mortality in healthy populations1,2. Furthermore, increased muscle strength in older adults has been associated with longer survival, improved functional outcomes, and enhanced quality of life3. In addition, reduced muscle strength has been identified as a contributing factor to cognitive decline and Alzheimer’s disease and seems to influence the long-term risk of stroke4,5. For muscle strength evaluation, the quadriceps and hamstrings, which are involved in knee extension and flexion, have been identified as effective muscle groups to assess declining muscle strength in a variety of chronic disorders6,7.
In clinical settings, the strength of the quadriceps and hamstrings is assessed using manual muscle tests (MMT) or handheld dynamometers (HHDs) that can quantitatively represent muscle strength8. However, MMT cannot provide a quantitative measure of muscle strength, making it difficult to objectively record changes in strength over time. In addition, MMTs and HHDs are susceptible to errors introduced by raters, which affect their accuracy and reliability9. In particular, when assessing powerful muscle groups, such as the quadriceps, challenges arise if the force to be measured exceeds the rater's weight and muscle strength. In such cases, the rater may be unable to stabilize the limbs and/or device, leading to difficulties obtaining accurate measurements10,11.
Isokinetic dynamometry (IKD) equipment, such as the Biodex System 4 Pro (Biodex Medical System Inc., Shirley, NY, USA), is considered the gold standard for measuring muscle strength and demonstrates high accuracy and reliability12. Despite the accuracy and reliability of IKDs, their limited portability and cost limit their use in regular hospital monitoring and athletic field evaluations13,14. Various muscle strength evaluation devices have been proposed to address the limitations of MMT, HHDs, and IKDs.
Sung et al. proposed an anchoring frame to assist in stabilizing HHD when measuring in the supine position15, and Hogrel et al. introduced a portable system for measuring knee extension in a seated position16. Padulo et al. designed an isometric bench that, although not highly portable, allowed the measurement of knee muscle strength in a seated position17. Compared with traditional methods, these inventions have the advantage of improving reliability and portability. However, they still present limitations, such as being bulky and inconvenient for transportation or requiring anchoring to bed frames, walls, or benches; therefore, they are influenced by environmental conditions17,18,19.
Myong et al.20 and Brookshaw et al.21 proposed devices that utilize load cells attached to the limb, allowing the measurement of knee extension, elbow extension, and elbow flexion strength without the need to install equipment in the surrounding environment. Cho et al.22 proposed a device that used load cells to measure the dorsiflexion and plantar flexion of the ankle. These devices demonstrated portability, validity, and reliability comparable to those of IKD. The portable articulated dynamometry system (PADS) proposed by Myong et al. is a device which facilitates the assessment of knee extension strength by securing load cells to the limbs, without requiring the installation or attachment of any equipment to the surrounding environment. This device functioned as a precursor to the current research20. Even though the PADS has enhanced accessibility for measuring maximum knee extension strength in daily life compared to the IKD, it has two primary limitations: Firstly, previous studies utilizing the PADS have only focused on assessing knee extension strength. As the hamstrings-to-quadriceps (H:Q) torque ratio is frequently used to detect muscle imbalances or as an indicator for injury prevention, prior studies have indicated that it is important to measure both knee extension and flexion strength together23,24,25. Secondly, in preliminary research using the PADS, the ability to adjust the length of the device based on limb length was restricted. Further, when measuring the force with a load cell to calculate the torque occurring at a knee joint, effort is required to measure the distance from the axis of rotation to the point of force application for each measurement. AS such, the PADS was intentionally designed without a feature for length adjustment.
In the present study, a wearable design for a muscle strength measurement system, in which torque sensors are positioned on the knee joint at the axis of rotation, is proposed. This study serves as a follow-up to previous PADS research using load cells and aims to improve the fixation force on the limb using a wearable design20,26. With the enhancement of fixation force, simultaneously measuring knee extension and flexion has become easier. The measurement method using torque sensors minimizes the distortion in the measurement values, even with deformations in the limb fixation and frame, and the symmetric design allows the intuitive setting of joint angles for measurement. In addition, the fixation is adjustable in length, enabling placement on high-stiffness areas with low-sensitive tissue, reducing pain or discomfort during measurement, and facilitating the measurement of maximum muscle strength27. The aim of this study was to design a wearable dynamometry system (WDS) and to evaluate its accuracy and reliability through cross-validation with measurements taken using the Biodex System 4 Pro in healthy adult subjects. Further, we report the results of a user experience questionnaire survey (UEQ-S).
Results
In this study, the inter-rater and test–retest reliabilities of knee flexion and extension strength measurements were assessed in 39 healthy adults. Each participant underwent two individual evaluation sessions for intersession reliability conducted by two independent raters for both knees, resulting in 312 measurements for extension and flexion for each device (a wearable dynamometry system [WDS] and the IKD).
The cross-validation results for the WDS and IKD are presented in Table 1. The maximum extension and flexion torques measured using the WDS and IKD showed a strong correlation (Pearson’s r: 0.900 in extension; Pearson’s r: 0.870 in flexion). The maximal strength measured with the WDS and IKD revealed no statistically significant differences in both extension (mean difference: − 2.00 Nm; 95% CI − 13.75 to 9.74 Nm; p-value: 0.955) and flexion (mean difference: 4.54 Nm; 95% CI − 0.97 to 10.06 Nm; p-value: 0.758). The intraclass correlation coefficient (ICC) between the two devices was 0.893 for extension and 0.857 for flexion, indicating excellent reliability. The standard error of measurement (SEM) for extension and flexion was 17.12 Nm (9.85%) and 9.31 Nm (11.93%), respectively, and the minimal detectable change (MDC) at 95% confidence interval (CI) for extension and flexion was 47.47 Nm (27.31%) and 25.81 Nm (33.07%), respectively. The linear relationship and Bland–Altman analysis between the measurements taken with the two devices are demonstrated in Fig. 1. The measurements of extension and flexion strength by the two devices exhibited strong linearity (Fig. 1a and c). The mean difference between the extension strength measurements of the two devices was -2.00 Nm, with most differences falling within the 95% limits of agreement (LOA range: − 49.53 to 45.52 Nm) (Fig. 1c). The mean difference in flexion strength measurements between the two devices was 4.55 Nm, indicating a higher bias for the measurements using WDS; however, most differences remained within the 95% limits of agreement (LOA range: -19.99 to 29.08 Nm) (Fig. 1d).
The WDS was further validated by evaluating inter-rater reliability. Two independent raters assessed each participant, and the data obtained from each rater showed a strong interrater correlation (Pearson’s r: WDS, 0.990; IKD, 0.980) (Fig. 2). No significant inter-rater differences were observed for the WDS (mean difference: 1.21 Nm; 95% CI − 8.13, 10.55 Nm; p-value: 0.972) and IKD (mean difference: 0.34 Nm; 95% CI − 9.56, 10.23 Nm; p-value: 0.993). The ICC demonstrated excellent reliability for both the WDS and IKD (0.993 and 0.991, respectively). The mean SEM was 4.98 Nm (4.05%) for the WDS and 5.98 Nm (4.86%) for the IKD. The MDC was 13.79 Nm (11.21%) for the WDS and 16.56 Nm (13.48%) for the IKD (Table 1). The linear relationship and Bland–Altman analysis between the measurements of raters are shown in Fig. 2. Both the WDS and IKD showed strong linearity (Fig. 2a and c). The mean difference in measurements between raters using the WDS was 1.21 Nm, with the majority of differences falling within the 95% limits of agreement (LOA range: − 18.11 to 20.54 Nm) (Fig. 2c). The mean difference between the measurements of raters using the IKD was 0.34 Nm, with most differences also falling within the 95% limits of agreement (LOA range: − 23.01 to 23.68 Nm), although a higher occurrence of outliers was observed as the measurement values increased with the IKD (Fig. 2d).
All participants attended a second measurement session at least 24 h after the first session, and test–retest reliability was assessed by comparing the data from the two sessions. Data obtained from both sessions showed strong test–retest correlations (Pearson’s r: WDS, 0.970; IKD, 0.970) (Fig. 3). No significant differences were observed between the sessions for the WDS (mean difference: 3.03 Nm; 95% CI − 6.30, 12.37 Nm; p-value: 0.937) and IKD (mean difference: 1.90 Nm; 95% CI − 8.00, 11.80 Nm; p-value: 0.964). The ICC demonstrated excellent reliability for both the WDS and IKD (0.984 and 0.986, respectively). The mean SEM was 7.56 Nm (6.15%) for the WDS and 7.54 Nm (6.13%) for the IKD. The MDC was 20.96 Nm (17.04%) for the WDS and 20.89 Nm (17.00%) for the IKD (Table 1). The linear relationship and Bland–Altman analysis between session measurements are presented in Fig. 3. Both the WDS and IKD demonstrated strong linearity (Fig. 3a and c). The mean difference in session measurements using the WDS was 3.03 Nm, suggesting a potential learning effect from the test, with most differences within the 95% limits of agreement (LOA range: − 26.38 to 32.44 Nm) (Fig. 3c). The mean difference in session measurements using the IKD was 1.90 Nm, with most differences also within the 95% limits of agreement (LOA range: − 27.43 to 31.23 Nm) (Fig. 3d).
Of the 39 participants, 35 participated in a user experience survey using the UEQ-S. The UEQ-S results showed a significant difference in the overall user experience for the WDS (p < 0.001). Notably, the participants indicated that the WDS outperformed IKD in terms of ease of use, efficiency, clarity, inventiveness, and being on the leading edge. The results for each item are presented in Table 2.
Discussion
The quadriceps and hamstrings play important roles, such as walking and balance, in daily life. Measuring knee extension and flexion strength is an effective method to assess the strength of the quadriceps and hamstrings. IKD, known as a benchmark for assessing muscle strength, is distinguished by its exceptional precision and reliability12. However, its high price, large dimensions (132 × 165 × 152 cm), and weight (612 kg) have restricted its widespread use in regular testing owing to its limited accessibility14,15 (see Supplementary Table S1 online). In addition, both MMT and HHDs are convenient for use in clinical settings and are commonly used for routine assessments; however, both methods are susceptible to errors introduced by raters28,29. In particular, measurement becomes challenging and unreliable if the force being assessed exceeds the rater's weight and muscular strength, causing the rater to be pushed back or unable to maintain the resistance10,30. Therefore, the research team recognized the need to develop a dynamometer system with good portability and high reliability.
The PADS previously developed by our research team was foldable and easy to carry and showed high validity and reliability in measuring the extension torque of the knee20. However, previously developed devices have limitations. First, for convenience in the torque calculation, the length of the lever arm in the PADS was fixed, making it impossible to adjust the fixation point to match the length of the leg. Regarding IKD, when measuring the maximum torque, the fixation point was positioned approximately 5 cm proximal to the distal aspect of the lateral malleolus to reduce discomfort and accurately transfer force31. However, with PADS, there was no adjustment for the fixation point's location, which occasionally caused discomfort or pain20. Second, the PADS had an aluminum frame in contact with the lower thigh, popliteal fossa, and calf. The fixation part was covered with silicone but was made of a rigid plastic material. Owing to the placement of the rigid material in the direction of the applied force, forces > 300 Nm led to pain caused by pressure from the fixation part32. In addition, there were cases in which the frame's contact with the patient's thigh or popliteal fossa caused discomfort.
In the present study, an ergonomically designed device was developed to measure the maximum extension and flexion torques of the knee. The proposed device is designed with an exoskeleton structure in which a torque sensor is placed concentrically on the knee joint, and a rigid frame acting as a lever arm is located on the left and right sides of the limb to surround the limb. This arrangement ensures that when assessing the force exerted by the knee joint, the participants can apply pressure without experiencing significant constriction or impediment from structural elements such as a frame. In addition, PADS using the load cell made it challenging to adjust the position of the fixation point because of the importance of the lever arm length for torque calculation. However, by incorporating the torque sensor, adjusting the fixation point to fit the leg length became more straightforward. This facilitates the attainment of maximal isometric extension and flexion torque, resulting in accurate and reliable measurements by the device33,34.
The results of the present study showed that the WDS demonstrated good agreement with the IKD in measuring knee isometric extension and flexion torque. In a study that used a load cell to measure knee isometric extension torque using PADS, the inter-rater SEM was 5.51%20. The WDS proposed in the present study showed an inter-rater SEM of 3.72%, which is comparable to previous results20. In addition, the intersession SEM for PADS was 8.38% in extension20, and that for WDS in the present study was 5.47% in extension, which is comparable to previous results (see Supplementary Table S2 online).
The ICC of the device proposed by Hogrel et al. and IKD was 0.973, and the SEM showed a good agreement of approximately 12%16. In addition, the ICC of the device proposed by Myong et al. and IKD was 0.935, and the SEM showed a good agreement of 9.75%20. For knee extension, the outcomes of the WDS proposed in the present study were comparable to those of existing devices, with an ICC of 0.893 and SEM of 9.85% for extension and an ICC of 0.857 and SEM of 11.93% for flexion.
In a user experience study, respondents expressed that the WDS was easier to use, more efficient, clearer, more innovative, and had a greater leading edge than the IKD. Approximately 25% of the participants (n = 9) were healthcare professionals such as doctors, nurses, physical therapists, and occupational therapists working at a university hospital. Healthcare professionals evaluated the system as easy to use, efficient, and clear (see Supplementary Table S3 online). Positive evaluations from these individuals are expected to play a significant role in the subsequent implementation of the system in a clinical setting.
The WDS is a wearable device that allows patients to measure maximal isometric knee extension and flexion torque in various positions. Weighing approximately 2.85 kg, it is portable and features an intuitive design with a high-precision torque sensor positioned at the axis of the knee joint rotation that exhibits low crosstalk35. To minimize bending or distortion, a stainless steel frame was used to withstand high forces. The WDS has yet to be commercialized; however, patents for WDS are currently being processed through the PCT system for international patents. In addition, the WDS is being prepared for product certification by obtaining Korean good manufacturing practice approval, making it easily accessible to researchers, healthcare professionals, and general users.
This study has some limitations; therefore, the interpretation of its findings should be approached with caution. First, because only healthy young adults aged 20–39 years were enrolled, the proposed device was not verified for individuals with relatively weak quadriceps and hamstrings, such as patients with musculoskeletal disorders or older adults. Additionally, this system has not yet been tested on stronger individuals, or those who have undergone strength training, such as athletes. Therefore, further research targeting patient populations or specifically focusing on older adults and athletes could validate the device in the future.
Second, when measuring knee extension and flexion torque using the WDS developed in this study, a learning effect may be observed. Although participants received training and performed practice attempts prior to undergoing testing, a learning effect could have been reflected in the measurements. Indeed, the change in the mean between sessions for both WDS and IKD was greater than zero, indicating an increasing trend for both devices.
Third, the device proposed in this study was developed as an isometric measurement system primarily aimed at enhancing portability. In addition to isometric methods, typical isokinetic and isotonic methods are used for measuring muscle strength. Isometric methods measure muscle strength while the joint angle is fixed, whereas isokinetic and isotonic methods measure muscle strength while moving the joint angle. Because movement speed is also an important factor in evaluating daily living performance, it should be considered when measuring muscle strength36. Isokinetic testing is extensively used to assess muscle strength while minimizing the risk of muscle injury37. However, applying it to portable devices is not easy because a driving system that includes an actuator is essential for isokinetic measurements. Therefore, our research team aims to conduct future research to develop a module that includes not only a torque sensor but also an actuator. Therefore, we expect to expand the application range of this device to isokinetic measurements and passive motion in knee rehabilitation.
In conclusion, the precise measurement of muscle strength and tracking of its trends are important in a clinical setting. However, conventional methods for assessing muscle strength, such as MMT and HHDs with limited reliability and reproducibility or the less accessible IKD, have been the primary approaches. In this study, the WDS demonstrated substantial reliability in measuring knee extension and flexion in healthy adults, with good accessibility and cost efficiency. The proposed WDS is an efficient alternative to established MMT and IKD.
Methods
Participants
This study included 39 healthy adults. Pregnant patients, those with a recent history of cardiovascular, musculoskeletal, neurological, or infectious diseases, and those who could not provide consent were excluded. In the study, the participants were Koreans aged between 20 and 39 years. Measurements of height, weight, and maximal grip strength were collected for all participants to evaluate physical characteristics (see Supplementary Table S4 online). This study followed the Declaration of Helsinki and was approved by the Institutional Review Board of Seoul National University Hospital (IRB no. 2201-025-1288). Written informed consent was obtained from all participants. The participant appearing in Fig. 6 provided written informed consent to publish the figure in an online open-access publication.
Sampling size and methods
Based on previous research data20, a minimum of 16 participants were required. This calculation was performed to determine the sample size necessary to observe a maximum 95% limit of agreement (LOA) in Bland–Altman analysis38. The sample size was calculated with a statistical power of 90% and a significance level of 0.05 (two-tailed). Additionally, the convenience sampling method was applied to recruit participants who were easily accessible within the research setting and conformed to the study criteria.
WDS
The WDS is a portable device featuring an improved exoskeleton design applied to the PADS20. The exoskeleton design facilitates joint angle adjustment and allows the measurement of extension and flexion forces in the same posture by aligning the knee joint and torque sensor axes39.
The main frame was composed of stainless steel (SUS304) with a thickness of 5 mm, which provided sufficient rigidity for muscle strength measurements. The design model used in the present study enabled angle adjustment at 5° intervals based on the combination of the main frame and the torque sensor.
The main frame was connected to an adjustable fixture tailored to the leg length and featured pads made of cotton and silicone (Ecoflex; Smooth-On, Inc., Macungie, PA, USA). It was designed to be secured to the thigh and calf using Velcro belts.
The torque sensor, which is responsible for measuring the rotational force of the knee joint, was positioned on the same axis as the rotational axis of the knee joint to ensure accurate torque measurement. In this study, a knee joint torque measurement system was used with two torque sensors (ATS200-D100; Aidin Robotics, Gyeonggi-do, South Korea) positioned bilaterally (Fig. 4). Each sensor could measure the torque within the range of 0–200 Nm with a resolution of 0.25 Nm. The sensor is thin and lightweight and effectively minimizes crosstalk using an Artificial Neural Network-based calibration method to obtain torque values excluding the moment35,40. The signal outputs from both sensors were transmitted to a laptop for data visualization and recorded through a universal serial bus to a controller area network transceiver (Fig. 5).
Validation was performed using a Class M1 weight set ranging from 0.1 to 100 kg within the measurement range (see Supplementary Fig. S1 online).
Experimental design
Maximal isometric extension and flexion torques of the knee joint were measured at 90° using the isometric setup of the WDS and IKD (Biodex System 4 Pro; Biodex Medical Systems Inc., Shirley, NY, USA) (Fig. 6). We aimed to evaluate the test–retest reliability of the two devices through two measurements (sessions) at least 24 h apart for each participant. In each session, measurements for both knees were carried out by two independent raters. A single trial consisted of extension and flexion movements, each lasting 3 s. Between each trial, a 30-s rest period was provided, while each rater recorded the maximum value measured across the two trials as the measurement outcome. When switching raters or changing the device, participants were allowed a 3-min rest period (see Supplementary Fig. S2 online). One rater was a board-certified physiatrist with 6 years of clinical experience, whereas the other was an engineer who was part of the development team. Before each trial, the participants were instructed to exert maximal effort and were provided consistent verbal encouragement throughout the trial. All participants performed a 3-min stretching session with the physiatrist before each assessment, and all assessments were performed under the direct supervision of the same physiatrist.
User experience study
After all measurements were completed, participants were asked to fill out the UEQ-S form for both devices. The UEQ-S is a simple evaluation method that measures users' subjective perceptions of a product, such as attractiveness and efficiency. Each item in the UEQ-S is rated on a 7-point Likert scale ranging from − 3 (very negative) to + 3 (very positive)41. All participants completed the questionnaire voluntarily and anonymously.
Data analysis
Due to the difference in the leg fixing positions between the WDS and the IKD, which resulted in different lever arm lengths, a constant corresponding to the ratio of the lever arm lengths was applied to the WDS to allow an absolute comparison of the measurements acquired using both devices. The constant value was 25/16 and was applied after the device calibration phase. The Kolmogorov–Smirnov test was applied to assess data normality, and a permutation test was performed to further assess whether there were any significant differences in measurements between the two devices, between raters, or between sessions. The inter-rater and test–retest reliabilities of each device were confirmed to assess the reliability of the WDS and IKD. The validity of the WDS was verified by considering IKD measurements as the gold standard and by comparing the values of the two devices42. The ICC established by Shrout and Fleiss was used to confirm the inter-rater and test–retest reliabilities and evaluate validity43. ICC model 2 (a two-way random effects model) was used to assess both the agreement between the two devices and the inter-rater reliability, whereas ICC model 3 (a two-way mixed effects model) was used to evaluate the test–retest reliability. The SEM was computed using the standard deviation (SD), where \(SEM = SD \sqrt{1-ICC}\). Furthermore, the MDC at a 95% CI was determined using the formula: \(MDC=1.96\cdot \sqrt{2}\cdot SEM\). To examine the inter-rater and test–retest reliabilities and the inconsistency of the values measured by the two devices, a Bland–Altman plot was drawn, and a scatter plot and Pearson’s correlation were shown for the two measured values44. All analyses were performed using R version 4.2.3.
Data availability
The data acquired in this study are not openly available due to the sensitive nature of human data (e.g. age, sex, height and weight). A de-identified dataset containing the full demographic and clinimetric data is available from the corresponding author (S.K.) upon reasonable request.
References
Metter, E. J. et al. Skeletal muscle strength as a predictor of all-cause mortality in healthy men. J. Gerontol. Ser. A Biol. Sci. Med. Sci. 57(10), B359–B365 (2002).
García-Hermoso, A. et al. Muscular strength as a predictor of all-cause mortality in an apparently healthy population: A systematic review and meta-analysis of data from approximately 2 million men and women. Arch. Phys. Med. Rehabil. 99(10), 2100–2113 (2018).
Katula, J. A., Jack Rejeski, W. & Marsh, A. P. Enhancing quality of life in older adults: A comparison of muscular strength and power training. Health Qual. Life Outcomes 6(1), 1–8 (2008).
Boyle, P. A., Buchman, A. S., Wilson, R. S., Leurgans, S. E. & Bennett, D. A. Association of muscle strength with the risk of Alzheimer disease and the rate of cognitive decline in community-dwelling older persons. Arch. Neurol. 66, 1339–1344 (2009).
Åberg, N. D. et al. Influence of cardiovascular fitness and muscle strength in early adulthood on long-term risk of stroke in Swedish men. Stroke 46, 1769–1776 (2015).
Ruas, C. V. et al. Alternative methods of determining hamstrings-to-quadriceps ratios: A comprehensive review. Sports Med. Open 5, 1–14 (2019).
Eriksrud, O. & Bohannon, R. W. Relationship of knee extension force to independence in sit-to-stand performance in patients receiving acute rehabilitation. Phys. Ther. 83(6), 544–551 (2003).
Saygin, D. et al. Hand-held dynamometry for assessment of muscle strength in patients with inflammatory myopathies. Rheumatology 60(5), 2146–2156 (2021).
Deones, V. L., Wiley, S. C. & Worrell, T. Assessment of quadriceps muscle performance by a hand-held dynamometer and an isokinetic dynamometer. J. Orthopaedic Sports Phys. Ther. 20(6), 296–301 (1994).
Kolber, M. J. & Cleland, J. A. Strength testing using hand-held dynamometry. Phys. Ther. Rev. 10(2), 99–112 (2005).
Bohannon, R. W. et al. Isometric knee extension force measured using a handheld dynamometer with and without belt-stabilization. Physiother. Theory Pract. 28(7), 562–568 (2012).
Palmer, T. B. et al. Real-time measurement of isometric peak torque and rate of torque development using a novel strength testing device: A validity and reliability study. Physiol. Meas. 41(11), 115005 (2020).
Romero-Franco, N. et al. Validity and reliability of a low-cost dynamometer to assess maximal isometric strength of upper limb: Low cost dynamometry and isometric strength of upper limb. J. Sports Sci. 37(15), 1787–1793 (2019).
Stark, T. et al. Hand-held dynamometry correlation with the gold standard isokinetic dynamometry: A systematic review. PM&R 3(5), 472–479 (2011).
Sung, K.-S., Yi, Y. G. & Shin, H.-I. Reliability and validity of knee extensor strength measurements using a portable dynamometer anchoring system in a supine position. BMC Musculoskelet. Disord. 20(1), 1–8 (2019).
Hogrel, J.-Y., Benveniste, O. & Bachasson, D. Routine monitoring of isometric knee extension strength in patients with muscle impairments using a new portable device: Cross-validation against a standard isokinetic dynamometer. Physiol. Meas. 41(1), 015003 (2020).
Padulo, J. et al. Validity and reliability of isometric-bench for knee isometric assessment. Int. J. Environ. Res. Public Health 17(12), 4326 (2020).
Kollock, J. R., Roger, O., Onate, J. A. & Van Lunen, B. The reliability of portable fixed dynamometry during hip and knee strength assessments. J. Athl. Train. 45(4), 349–356 (2010).
Fransen, M., Crosbie, J. & Edmonds, J. Isometric muscle force measurement for clinicians treating patients with osteoarthritis of the knee. Arthritis Care Res. Off. J. Am. Coll. Rheumatol. 49(1), 29–35 (2003).
Myong, Y. et al. Development and validation of a portable articulated dynamometry system to assess knee extensor muscle strength. Sci. Rep. 13(1), 11887 (2023).
Brookshaw, M., Sexton, A. & McGibbon, C. A. Reliability and validity of a novel wearable device for measuring elbow strength. Sensors 20(12), 3412 (2020).
Cho, S. Y. et al. A portable articulated dynamometer for ankle dorsiflexion and plantar flexion strength measurement: A design, validation, and user experience study. Sci. Rep. 13, 22221 (2023).
Kellis, E. et al. Hamstring to quadriceps strength ratio and cross-sectional area of the quadriceps and hamstrings muscles assessed using extended field-of-view ultrasonography. Res. Sports Med 29(1), 25–42 (2021).
Andrade, M. S. et al. Age-related differences in torque in angle-specific and peak torque hamstring to quadriceps ratios in female soccer players from 11 to 18 years old: Α cross-sectional study. Res. Sports Med. 29(1), 77–89 (2021).
Stastny, P., Michal, L. & James, J. T. Muscle imbalances: Testing and training functional eccentric hamstring strength in athletic populations. JoVE (J. Vis. Exp.) 135, e57508 (2018).
Sarkisian, S. V. et al. Design, development, and validation of a self-aligning mechanism for high-torque powered knee exoskeletons. IEEE Trans. Med. Robot. Bionics 2(2), 248–259 (2020).
Bohannon, R. W. Test-retest reliability of hand-held dynamometry during a single session of strength assessment. Phys. Ther. 66(2), 206–209 (1986).
Janssen, J. C. & Le-Ngoc, L. Intratester reliability and validity of concentric measurements using a new hand-held dynamometer. Arch. Phys. Med. Rehabil. 90(9), 1541–1547 (2009).
Wadsworth, C. T. et al. Intrarater reliability of manual muscle testing and hand-held dynametric muscle testing. Phys. Ther. 67(9), 1342–1347 (1987).
Valentin, G. & Maribo, T. Hand-held dynamometry fixated with a tripod is reliable for assessment of back extensor strength in women with osteoporosis. Osteoporos. Int. 25, 2143–2149 (2014).
Hansen, E. M. et al. Hand-held dynamometer positioning impacts discomfort during quadriceps strength testing: A validity and reliability study. Int. J. Sports Phys. Ther. 10(1), 62 (2015).
Suzuki, T. Reliability of measurements of knee extensor muscle strength using a pull-type hand-held dynamometer. J. Phys. Ther. Sci. 27(3), 967–971 (2015).
Sposito, M. et al. Exoskeleton kinematic design robustness: An assessment method to account for human variability. Wear. Technol. 1, e7 (2020).
Tang, X. et al. A study of knee exoskeleton configuration based on lower limb motion characteristics analysis. Machines 11(7), 709 (2023).
Seok, D.-Y. et al. Ultra–thin joint torque sensor with enhanced sensitivity for robotic application. IEEE Robot. Autom. Lett. 5(4), 5873–5880 (2020).
Yoshioka, S. et al. Biomechanical analysis of the relation between movement time and joint moment development during a sit-to-stand task. Biomed. Eng. Online 8(1), 1–9 (2009).
Parraca, J. A. et al. Test-retest reliability of isokinetic strength measurements in lower limbs in elderly. Biology 11(6), 802 (2022).
Lu, M.-J. et al. Sample size for assessing agreement between two methods of measurement by Bland−Altman method. Int. J. Biostat. https://doi.org/10.1515/ijb-2015-0039 (2016).
Pons, J. L. Wearable Robots: Biomechatronic Exoskeletons (Wiley, 2008).
Kim, Y. B. et al. 6-axis force/torque sensor with a novel autonomous weight compensating capability for robotic applications. IEEE Robot. Autom. Lett. 5(4), 6686–6693 (2020).
Schrepp, M., Hinderks, A. & Thomaschewski, J. Design and evaluation of a short version of the user experience questionnaire (UEQ-S). Int. J. Interact. Multimed. Artif. Intell. 4(6), 103–108 (2017).
Martin, H. J. et al. Is hand-held dynamometry useful for the measurement of quadriceps strength in older people? A comparison with the gold standard Biodex dynamometry. Gerontology 52(3), 154–159 (2006).
Koo, T. K. & Mae, Y. L. A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J. Chiropract. Med. 15(2), 155–163 (2016).
Giavarina, D. Understanding Bland Altman analysis. Biochem. Med. 25(2), 141–151 (2015).
Acknowledgements
This work was supported by the National Research Foundation of Korea (NRF) grant funded by the Korean government (MIST) (No-2021R1C1C2095529), the Interdisciplinary Research Initiatives Program from College of Engineering and College of Medicine, Seoul National University (grant no. 800-20230385), and the Seoul National University Hospital Research Fund (grant no. 1920230020).
Author information
Authors and Affiliations
Contributions
Concept/idea/research design: S.P., Y.M., M.C., S.K.; Acquisition of data: S.P., Y.M., M.C., S.Y.C.; Analysis and interpretation of data: S.P., Y.M., M.C., W.H.L., B.O., S.K.; Writing/review/editing of manuscript: S.P., Y.M., M.C., W.H.L., B.O., S.K.; Acquisition of funding: S.P., Y.M.; Providing facilities/equipment: B.O., S.K.; Final approval of the manuscript: all authors.
Corresponding authors
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Park, S., Myong, Y., Cho, M. et al. Design and validation of a wearable dynamometry system for knee extension-flexion torque measurement. Sci Rep 14, 10428 (2024). https://doi.org/10.1038/s41598-024-60985-9
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-024-60985-9
- Springer Nature Limited