Abstract
Biosensors based on field-effect transistors (FETs) are suitable for use in miniaturized and cost-effective healthcare devices. Various semiconductive materials can be applied as FET channels for biosensing, including one- and two-dimensional materials. The signal transduction interface between the biosample and the channel of FETs plays a key role in translating electrochemical reactions into output signals, thereby capturing target ions or biomolecules. In this Review, distinctive signal transduction interfaces for FET biosensors are introduced, categorized as chemically synthesized, physically structured, and biologically induced interfaces. The Review highlights that these signal transduction interfaces are key in controlling biosensing parameters, such as specificity, selectivity, binding constant, limit of detection, signal-to-noise ratio, and biocompatibility.
Similar content being viewed by others
Introduction
A platform based on a solution-gated field-effect transistor (FET), which originates from electronics, is suitable for use in miniaturized and cost-effective systems to directly measure biological samples as the FET biosensor in the field of in vitro diagnostics1. Such miniaturized electronic devices can be easily equipped with a wireless function and attached to the body, which is available for wearable biosensors to detect biomarkers in tears, sweat, and saliva, that is, for diagnostics in a blood-sampling free manner2,3,4. In general, the gate insulator surface (e.g., SiO2) is directly in contact with a measurement solution in the FET biosensor without a metal gate electrode, which is different from a metal-oxide-semiconductor (MOS) transistor, for which the potential of the measurement solution is controlled by the reference electrode5,6,7, as shown in Fig. 1a. When ions or biomolecules with charges are adsorbed on the gate insulator surface, their charges electrostatically interact with electrons across the gate insulator, resulting in a change in the conductivity of the channel of the FET. That is, the drain–source current (IDS) at the channel changes with the change in the density of ions or biomolecules with charges adsorbed on the gate insulator. That is, such charged species induce a change in the potential (Vout) of the gate insulator/electrolyte solution interface at a constant IDS, which is potentiometrically detected5,8. Oxide and nitride membranes (e.g., Ta2O5 and Si3N4) used as the gate insulator contribute to the detection of a change in pH with the FET on the basis of the reaction in equilibrium between hydrogen ions and hydroxy groups at their surfaces5. Afterwards, various ion-sensitive membranes (ISMs)2,9,10,11,12,13,14,15 and biological receptors with enzymes, antibodies, and single-stranded DNAs16,17,18,19,20,21 were modified on the gate electrode of the FET biosensors to specifically and selectively detect target ions and biomolecules, and further cellular activities, which were accompanied by changes in ion concentration, were monitored in real-time with the FET biosensors22,23,24,25. Thus, the detection principle of FET biosensors is based on the potentiometric measurement of the changes in ionic and biomolecular charges or membrane capacitances at the gate electrode/electrolyte solution interface. Moreover, one-dimensional (1D) and two-dimensional (2D) semiconductive materials have been recently proposed for the channel of FETs for biosensing devices15,26,27,28,29,30. In particular, a solution-gated 1D or 2D-channel FET biosensor with a steep subthreshold slope (SS) contributes to an ultrahigh sensitive biosensing, owing to a relatively large electric double-layer capacitance at the electrolyte solution/channel interface27,31. Moreover, thin-film transistors with transparent amorphous oxide semiconductors can be applied as one of the FET biosensors, which are deposited on transparent substrates such as glass and plastics32. Thus, the number of studies on the FET biosensors is increasing yearly (Fig. 1b).
Considering the concept of biosensors, which consist of three components, namely, a biological sample, a signal transduction interface, and a detection device (Fig. 1a)1,33, a specific biological target detected should be considered in developing FET biosensors for their applications in various fields such as clinical diagnostics and pharmaceutical discovery. In particular, the signal transduction interface is designed and positioned between the biological sample and the detection device to specifically and selectively detect the target biomarker. Functionalized interfaces are chemically synthesized, physically and chemically structured, and biologically induced, considering the diversity of material properties (Fig. 1a). Moreover, as a detection device, various sensing principles such as electrochemical and optical methods are proposed for the detection of biomarkers; particularly, the solution-gated FETs for biosensing are focused on in this review, considering the topical backgrounds mentioned above.
In electrochemical biosensors such as FET biosensors, the signal transduction interface between the electrode surface and the biological sample plays a key role in capturing target ions or biomolecules, which then transduce the electrical properties of ions or biomolecules and electrochemical reactions into output signals (Fig. 1a). In general, biological receptors such as enzymes and antibodies are often utilized on the electrode as a biologically induced electrical interface to transduce their reactions with target biomarkers into electrical signals, showing high specificity and selectivity in their detections16,34,35,36,37. However, the use of such biological materials is problematic because there are no enzymes or antibodies applicable to every target biomarker, as well as other reasons such as their poor long-term stability, high cost and time-consuming production, and the difficulty of quality control of their production. On the other hand, polymer membranes such as molecularly imprinted polymers (MIPs)38,39,40,41,42,43 and aptamers44,45,46 that are artificially designed and simply fabricated are expected as a platform of the signal transduction interfaces of FET biosensors, thereby not only overcoming the above issues of biological materials, but also enhancing the binding constant (Ka) and thereby the limit of detection (LOD). These polymer membranes are arranged on the gate electrode surface as a chemically synthesized electrical interface43,47,48,49,50,51,52,53,54,55. Such flexible polymeric membranes are also available for wearable biosensors. That is, it is better to use soft materials such as polymeric membranes or a thin film as the signal transduction interfaces to decrease the stiffness of the wearable and flexible biosensors. Moreover, living cells themselves, which are cultured on the gate electrode surface of FET biosensors, can also work as a biologically induced electrical interface. For instance, mast cells with IgE antibodies cultured on the gate electrode induce electrical signals on the basis of the change in pH, which is caused by the immunological reactions of the antibodies with the target antigen56,57. Moreover, vascular endothelial cells with a basement membrane cultured on the gate electrode are induced to invade cancer cells, resulting in the generation of electrical signals of FET biosensors58,59. Such living cells with distinctive functions can be utilized as one of the biologically induced electrical interfaces similar to immobilized enzyme membranes. Thus, we can design diverse signal transduction interfaces for electrochemical biosensing at not only the molecular level but also the cellular level.
However, we have to address concerns on how to increase the signal-to-noise ratio (S/N) for the FET biosensors, based on not only the effect of counterions in sample solutions capable of shielding targeted biomolecular charges (Debye length limitation) thereby reducing the output signals18,20,60,61,62,63,64,65,66,67,68, but also noise signals that are unexpectedly caused by the fouling and nonspecific adsorptions of interfering species in biological sample solutions69,70,71,72,73. The Debye length λ depends on the ionic strength of the electrolyte solution used and is expressed as λ = (ε0εrkBT/2NAe2I)1/2, where I is the ionic strength of the electrolyte solution, ε0 is the permittivity of free space, εr is the dielectric constant, kB is the Boltzmann constant, T is the absolute temperature, NA is the Avogadro number, and e is the elementary charge. The Debye length limitation is controlled by changing the ionic strength of a measurement solution, that is, diluted measurement solutions are useful for improving the detection sensitivity of the FET biosensors to charged biomolecules because of the reduction of the shielding effect by counterions. Furthermore, in terms of increasing S/N, convex (e.g., nanopillars)74,75,76 and concave (e.g., nanofilters and nanopores)70,71,72,73,75 electrode surfaces may be effective as physically and chemically structured electrical interfaces. The increase in electrode surface area is expected to enhance the output signals of FET biosensors, whereas it may generate noise signals simultaneously. In this case, the electrical noise signals can be suppressed in the FET biosensors when small and large interfering species are trapped outside the Debye length by a polymeric nanofilter interface, blocking their approach to the electrode surface, which means the suppression of nonspecific noise signals. That is, the polymeric nanofilter structurally leads electrical signals to a small target biomarker, which passes through the nanofilter interface to the electrode while preventing small interfering species from approaching the electrode70,71,72; that is, S/N can be increased.
Considering the application of electric devices with FETs to biosensors, it is very important to design devices such that the change in the charge density based on biomolecular recognition events is directly transduced into electrical signals at the signal transduction interfaces without relying on enzymes and antibodies. Therefore, the signal transduction interfaces classified as chemically synthesized, physically and chemically structured, and biologically induced interfaces become a key element in controlling the performances of FET biosensors, which determine their future applications. However, there are no reports on the signal transduction interfaces as classified in the above. In this paper, we review some distinctive signal transduction interfaces classified in relation to FET biosensors, which transduce biomolecular recognition events into electrical signals. In particular, the diverse signal transduction interfaces in FET biosensors are noted to become a key element in controlling biosensing parameters with respect to biological targets, such as specificity, selectivity, Ka, LOD, S/N, and biocompatibility. Therefore, the diversity of signal transduction interfaces should broaden the possibility of developing novel biosensing devices, in parallel with the development of new channel semiconductors of FET biosensors.
Chemically synthesized electrical interfaces for FET-based biosensing
Polymeric membranes can be artificially functionalized by copolymerizing characteristic monomers and entrapping distinctive chemicals in a biological receptor-free manner. In particular, functional polymeric membranes should have the ability to transduce the interactions of such chemicals with charged ions and biomolecules into electrical signals. The chemically synthesized electrical interfaces must be accordingly designed, considering the physicochemical characteristics and components of polymeric membranes. As the typical ISMs, hydrophobic polymers are mostly utilized to entrap poorly water-soluble ionophores such as crown ethers that coordinate ions with charges, which dissolve in a plasticizer2,9,10,11,12,13,14,15, whereas water-soluble biomolecules such as glucose are easily incorporated into hydrophilic polymeric membranes with functional monomers such as phenylboronic acids (PBAs), which expectedly induce electrical charges on the basis of diol binding to some biomolecules owing to the formation of deprotonated boronic acid diol esters with negative charges49,50,51,52,77,78. These chemical interactions, which contribute to the change in the density of ionic and molecular charges, on/in such polymeric membranes, are output as the electrical signals of FET biosensors.
Ion-sensitive membranes and their biocompatibility
Most popular ion sensors are composed of hydrophobic and flexible polymeric membranes such as polyvinyl chloride (PVC), including ionophores coated on electrodes for the potentiometric measurements of target ions. Artificial ionophores such as crown ethers and calixarenes are dissolved in a plasticizer to enhance their mobilities in the flexible PVC membrane. Various ions such as Na+, K+, Ca2+, NH4+, and Cl- can be detected using the potentiometric ISM-coated electrodes, including FETs (Fig. 2a)2,9,10,11,12,13,14,15. Thus, flexible ISMs enable the detection of inorganic ions in biological and environmental samples. The difference in the potential at the ISM/electrolyte solution interface (ΔE) changes with the change in ion concentrations, that is, the density of ionic charges captured on/in the ISMs under equilibrium. That is, the detection sensitivity of potentiometric sensors with ISMs ideally follows the slope obtained on the basis of the Nernstian equation (e.g., 59.2/n mV/decade for n-valent ions at 298 K). However, the detection sensitivity may be lower, depending on the density of ionophores entrapped in the ISMs. Here, ΔE is expressed as Eqs. 1 and 2 on the basis of the Nernstian equation in relation to the logarithm of the n-valent cation concentration [Mn+]:7
where k is the Boltzmann constant, T is the absolute temperature, q is the elementary charge, and CDL is the capacitance of the electric double layer. The parameter β reflects the chemical sensitivity of ISMs, which depends on the site density NS of ionophores, and the surface reactivity is characterized on the basis of Ka of the ionophore–ion coordination bond. Thus, the site density of ionophores is a major factor that determines ΔE, that is, the detection sensitivity of ISMs. Moreover, the selectivity of a crown ether L for ions \({{{{{{\rm{M}}}}}}}_{1}^{{{{{{\rm{n}}}}}}+}\) and \({{{{{{\rm{M}}}}}}}_{2}^{{{{{{\rm{n}}}}}}+}\) is expressed as the ratio of each binding constant Ka(1)/Ka(2); e.g., Ka(1) = [M1Ln+]/(\([{{{{{{\rm{M}}}}}}}_{1}^{{{{{{\rm{n}}}}}}+}]\) + [L]). The complexing ability, that is, Ka of crown ether with ions, is dependent on the relative sizes of the cavity of the crown ether and the ions79. In addition, poly- and bis (crown ether)s are favorable for the formation of 2:1 (crown ether unit/ion) sandwich-type complexes with ions, which contribute to the increase in Ka, that is, the selectivity to a specific ion80.
Moreover, the biocompatibilities of ISMs with flexibility should be considered for wearable biosensors, focusing on cytotoxicity and biofouling. Plasticizers used in ISMs may be cytotoxic when used in direct contact with the human body as a wearable biosensor. In fact, plasticizers were found to be cytotoxic to living cells13. The plasticizers and ionophores included in ISMs may leak into the sample solution and then result in it becoming cytotoxic (Fig. 2a). That is, we had better develop a plasticizer-free ISM, from which the ionophores do not leak, to apply to biocompatible ion sensors. In this case, the mobility and dispersibility of ionophores in a plasticizer-free polymeric material should be considered, and then the matrix material that has polarity and a glass transition temperature below room temperature is preferred. Moreover, the dielectric constant of the polymer selected is desired to be in the range of 4–15 so as not to increase membrane resistance [e.g., fluoropolysilicone (FPS)]81,82. A plasticizer-free Na+-sensitive FPS-gate FET with calix[4]arene actually exhibited higher biocompatibility (i.e., noncytotoxicity) and sufficient sensitivity to and selectivity for Na+, as expected (Fig. 2b)83. On the other hand, we have another issue with ISMs that show nonspecific adsorption of, for example, proteins in whole samples such as blood, that is, biofouling of ISMs, resulting in the reduction in sensitivity. As an anti-biofouling treatment, ISMs are modified with some hydrophilic polymers (Fig. 2c)84,85. Hydrophilic polydopamine (poly-DA) modification actually enhanced the anti-adhesive properties of ISMs by increasing the surface hydrophilicity85. This is why a poly-DA film prevented the nonspecific adsorptions of proteins that may generate noise signals on the ISMs, while the detection sensitivity for monovalent ions did not deteriorate, keeping it near the Nernstian response. Thus, the modified ISM-based FETs can be utilized more safely and precisely in actual biological samples with interfering species as wearable biosensors with flexibility.
MIP-based electrical interfaces
As described in Introduction, there are some issues with biological receptors such as enzymes and antibodies; therefore, there is a need to develop versatile polymeric membranes to capture various biomarkers as intended. In the case of FET biosensors, such functional polymeric membranes should induce electrical signals. An MIP is a soft and biomimetic material developed to artificially realize the selective detection of targets38,39,40,41,42. A rigid and highly crosslinked polymer matrix contributes to the selective molecular recognition, which generates a strong target–functional monomer interaction (Fig. 3a). Thanks to their simple preparation and versatility, molecular imprinting technologies have been applied to various biosensing devices to selectively detect targets86,87,88,89. In particular, MIP films have been modified onto the gate electrode of FET biosensors for the specific and selective detection of small biomolecules to form a chemically synthesized electrical interface43,47,48,49,50,51,52,53. PBA is often utilized as the target-interacting functional monomer for a MIP-based electrical interface for the FET biosensor (Fig. 3b)43,49,50,51,52,53, because it covalently forms stable esters with diol biomolecules (e.g., saccharides, catechol amines, and lactic acid), which are water-soluble chemicals (Fig. 3b). This is why PBA has attracted a lot of attention in the field of molecular recognition. Furthermore, the esterification of PBA–diol compound binding is based on a reversible reaction, depending on pH78. Therefore, the template molecules can be eluted from the polymerized matrix by simply controlling pH. In particular, PBA is in equilibrium at an anionic form in the esterification (Fig. 3b). For instance, glucose molecules used as the template, which bind to PBA, are removed under acidic conditions, whereas glucose boronate esters induce negative charges in the MIP matrices under relatively basic conditions because the pKa of a glucose boronate ester is 6.878. Hence, FET biosensors can detect the change in the density of negative charges generated by the PBA–diol compound binding on the basis of the detection principle.
Here, a quantitative analysis is needed to grasp the chemical basis of MIP–target interactions underlying the electrical responses of MIP-based FET biosensors. The binding characteristics of a target to MIP are often quantified with adsorption isotherm equations in the case that the binding process involves the reversible reaction of the target to the target-selective MIP. According to a previous study90, the Langmuir adsorption isotherm equation is described as
where A is a signal observed at equilibrium for the MIP-bound template, [c] the free concentration of the template at equilibrium, N the number of available active centers in the MIP per unit volume, and Ka the binding constant. Equation 1 assumes homogeneously distributed binding sites with a constant binding constant Ka.
In the unsaturated region, the operation of a silicon-based FET is represented as
where μ is the electron mobility in the channel, COX is the gate oxide capacitance, \(\frac{W}{L}\) is the channel width-to-length ratio, VDS and VGS are the applied drain–source and gate–source voltages, respectively, and VT is the threshold voltage, which is described as ref. 7
where Eref refers to the potential of a reference electrode relative to a vacuum, \((-{\psi }_{0}+{{{{{{\rm{\chi }}}}}}}^{{{{{{\rm{sol}}}}}}})\) to the interfacial potential between the electrolyte solution and the gate oxide electrode (the factor χsol is the surface dipole moment of the solution, which can be considered constant), \(\frac{{\phi }_{{{{{{\rm{si}}}}}}}}{{{{{{\rm{q}}}}}}}\) to the work function for silicon electron, Qit, Qf, and QB to the charge of the interface traps, the fixed oxide charge, and the bulk depletion charge per unit area, respectively, and \({\phi }_{{{{{{\rm{f}}}}}}}\) to the Fermi potential difference between the doped bulk silicon and the intrinsic silicon.
When the MIP film is coated on the gate oxide electrode of the FET, the capacitance and charge in the MIP film should be added to Eq. 3, which is modified as
with
where QMIP is the charge of the MIP film, and CCom is the combined capacitance of COX and the MIP film (CMIP) on the gate oxide electrode. Considering that CMIP would hardly change after the addition of targeted molecules43,91 and COX is also constant, CCom is almost constant regardless of the adsorption of biomolecules, especially small molecules. In addition, the change in \({\psi }_{0}\) is assumed to remain unchanged because pH can be basically maintained by using a buffer solution. Eref, \(\frac{{\phi }_{{{{{{\rm{si}}}}}}}}{q}\), Qit, Qf, QB, and \({\phi }_{{{{{{\rm{f}}}}}}}\) are the factors that do not change even after molecular recognition events at the MIP interface. Considering the above, the electrical response output with the MIP-based FET biosensor, that is, the change in VT (ΔVT) is involved in ΔQMIP on the basis of Eq. 4.
The binding affinity of PBA to a diol compound depends on pH in a solution, in particular, the B(OH)3- complex is much stabler than the B(OH)2 complex92. That is, the reversible reaction between the target diol biomolecule (T) and PBA in the MIP film is described as
the formation rate of the\({{{{{\rm{T}}}}}}\cdot {{{{{{\rm{PBA}}}}}}}^{-}\) complex at time t is written as
where ka is the association rate constant and kd is the dissociation rate constant. At time t, \([{{{{{\rm{PBA}}}}}}]={[{{{{{\rm{PBA}}}}}}]}_{0}-[{{{{{\rm{T}}}}}}\cdot {{{{{{\rm{PBA}}}}}}}^{-}]\), where \({[{{{{{\rm{PBA}}}}}}]}_{0}\) is the concentration of PBA at t = 0. This is substituted into Eq. 7 to give
Here, the charge density of the MIP film (QMIP) is proportional to the formation of the\({{{{{\rm{T}}}}}}\cdot {{{{{{\rm{PBA}}}}}}}^{-}\) complex in the MIP film. Then, the maximum QMIP, that is, Qmax is dependent on the PBA concentration in the MIP film (\({[{{{{{\rm{PBA}}}}}}]}_{0}\) at t = 0), which is based on the capacity of the incorporated ligand. Therefore, Eq. 8 is modified to
where \(\frac{d{Q}_{{{{{{\rm{MIP}}}}}}}}{dt}\) is the formation rate of the associated complex (\({{{{{\rm{T}}}}}}\cdot {{{{{{\rm{PBA}}}}}}}^{-}\)) in the MIP film on the gate oxide surface and [c] is the concentration of the analyte (T) in a solution. In addition, integrating Eq. 9 gives
where Ka is the binding constant of T and PBA (ka/kd) in the MIP film. From Eq. 10, \({Q}_{{{{{{\rm{MIP}}}}}}}^{t=0}=0\). Considering Eq. 4,
which is estimated at an equilibrium time t. Here, \(\varDelta {V}_{{{{{{\rm{out}}}}}}}^{\max }\) is the maximum change in potential induced by ΔQmax at the MIP interface, which is proportional to the density of binding sites. Actually, ΔVout at the MIP interface is measured as the change in the surface potential against the reference electrode at a constant IDS using the source follower circuit8, as shown in Fig. 3c. That is, the detected ΔVout, which is induced by ΔQ, is consistent with –ΔVT obtained in the VGS and IDS transfer characteristics.
Considering the above, the electrical signals of MIP-based FET biosensors obey the Langmuir adsorption model. Therefore, the adsorption isotherm equation for the MIP-FET system is obtained from Eqs. 1 and 11 as
where [c] is the target concentration at equilibrium. Furthermore, the homogeneity and heterogeneity of binding sites distributed in MIPs are critical to the effective increase in selectivity. The binding sites in MIPs are assumed to be heterogeneously distributed owing to the randomness of polymerization and the insufficient elution of template molecules from the matrixes, depending on the polymerization processes93; that is, MIPs may unintentionally contain both nonselective and highly selective binding sites at a certain ratio. In that case, the bi-Langmuir adsorption isotherm equation (Eq. 13) may be utilized for the heterogeneous binding model of the MIP-FET system, instead of Eq. 12.
Equation 13 assumes two main types of binding sites with different affinities in the heterogeneous MIP film. However, Ka can be simply compared among MIPs for various target biomolecules using Eq. 12, as far as the results analyzed in most previous works are concerned.
In addition, stiffer MIP matrices may be more selective; that is, it should be beneficial to prepare flexible polymeric membranes as rigid as possible94. Actually, electroactive polymeric membranes with relatively high Young’s moduli, such as polypyrrole (PPy) and poly(o-phenylenediamine) (PoPDA) appear to be utilized as the MIP matrices, preserving the macromolecular arrangements after the template extraction95,96,97,98,99,100,101. On the other hand, 2-hydroxyethylmethacrylate (HEMA), ethylene glycol dimethacrylate (EGDMA), and so forth are copolymerized to improve the hydrophilicity of MIP matrices with flexibility, which include small biomolecules and electrolytes and may show relatively low Young’s moduli, when PBA derivatives (e.g., 4-vinyl-PBA) are copolymerized in the MIP matrices on the gate electrode (Fig. 3b)43,49,50,51,52. Such hydrophilic MIPs contribute to the reduction in the intensity of noise signals based on the nonspecific adsorptions of interfering biomacromolecules such as proteins in sample solutions102. Moreover, basic N-[3-(dimethylamino)propyl]methacrylamide (DMAPM) is often incorporated in the MIP matrices, including PBA to control pH43,49,50,51,52,103. Furthermore, methacrylic acid (MAA) can be used as a functional monomer in the MIP matrices because of its versatility in interactions43,50,53,104,105,106,107. MAA is used to electrostatically interact with amino groups in target biomolecules, to further increase the selectivity of biosensors. Note that MIP matrices should be crosslinked to maintain the cavity size and shape, which are associated with the specificity and selectivity for target biomolecules. The rigidity of MIPs would also block nontarget biomacromolecules such as proteins from coming in contact with the gate electrode, which contributes to the suppression of nonspecific signals. In particular, the swelling–deswelling behavior of such hydrogels after the reaction with target biomolecules may cause the change in the capacitance of MIP films and then cancel out the electrical signals on the basis of the change in the density of molecular charges. Therefore, the density of cross-linking in the MIP matrices on the gate electrode should be controlled to minimize the swelling–deswelling behavior52,91, although such capacitive signals may be useful for the detection of target biomolecules. Actually, a glucose-selective MIP, which was randomly copolymerized as poly(HEMA-ran-DMAPM-ran-VPBA-ran-MAA) and cross-linked with MBA on the gate surface of FET, worked well as a chemically synthesized electrical interface of FET biosensors (Figs. 3b, c)52. At pH 7.4, the binding constant for glucose (\({{{{{{\rm{K}}}}}}}_{{{{{{\rm{a}}}}}}}^{{{{{{\rm{glucose}}}}}}}\)) with the glucose-selective MIP-FET was estimated as 1.2 × 103 M−1 by fitting the Langmuir isotherm equation (Fig. 3c), which was approximately 260 times higher than that of the pristine PBA–glucose binding (4.6 M−1), whereas the binding constant for fructose (\({{{{{{\rm{K}}}}}}}_{{{{{{\rm{a}}}}}}}^{{{{{{\rm{fructose}}}}}}}\)) in the glucose-selective MIP (2.2 × 102 M−1) hardly changed from that for the pristine PBA–fructose binding (1.6 × 102 M−1)78. This means that not only did the imprinting for the template molecule in the polymeric membrane increase Ka for the target biomolecule, but the molecular charges induced by the diol compound/PBA binding also contributed to the generation of electrical signals of the FET biosensor. In this case, the selectivity (S) for glucose in the MIP film was calculated from the ratio of \({{{{{{\rm{K}}}}}}}_{{{{{{\rm{a}}}}}}}^{{{{{{\rm{glucose}}}}}}}\) to the Ka of other sugars and PBA (\({{{{{{\rm{K}}}}}}}_{{{{{{\rm{a}}}}}}}^{{{{{{\rm{sugars}}}}}}}\)), that is, \({{{{{{\rm{S}}}}}}}_{{{{{{\rm{glucose}}}}}}/{{{{{\rm{fructose}}}}}}}^{{{{{{\rm{MIP}}}}}}}={{{{{{\rm{K}}}}}}}_{{{{{{\rm{a}}}}}}}^{{{{{{\rm{glucose}}}}}}}{/{{{{{\rm{K}}}}}}}_{{{{{{\rm{a}}}}}}}^{{{{{{\rm{fructose}}}}}}}=5.6\) 52. From these calculations, the detection selectivity for glucose to fructose using the glucose-selective MIP-FET was about 200 times higher than that in the pristine PBA–sugar binding (\({{{{{{\rm{S}}}}}}}_{{{{{{\rm{glucose}}}}}}/{{{{{\rm{fructose}}}}}}}^{{{{{{\rm{pristine}}}}}}}=2.9\times {10}^{-2}\)). In this glucose-selective MIP-FET, moreover, LOD was determined to be ~3 μM for glucose detection in accordance with the Kaiser limit theory108, which can be compared with that in glucose-responsive MIPs obtained by other readout technologies109. Here, LOD would be associated with Ka in Eq. 12. When ΔVout at the concentration \({[{{{{{\rm{c}}}}}}]}_{{{{{{\rm{LOD}}}}}}}\) is \({\Delta {{{{{\rm{V}}}}}}}_{{{{{{\rm{out}}}}}}}^{{{{{{\rm{LOD}}}}}}}\), Eq. 12 is modified to
Then, Eq. 14 is rearranged as
Actually, the glucose-selective MIP-FET shown above sufficiently satisfies Eq. 15. Furthermore, this trend can also be found in other MIPs, regardless of the readout technologies (Fig. 3d)51,52,53,96,97,99,100,105,106,107,110,111,112,113,114. Therefore, Ka is a parameter used to control LOD. Note that Ka of the target/MIP interaction may be inherently derived from that of the target/functional monomer (e.g., PBA–diol compounds and host–guest interactions). Moreover, as another parameter, the thickness of MIP films appears to be related to the LOD (Fig. 3e)47,48,51,52,97,99,107,112,114,115,116,117. Thin-film MIPs with a thickness less than ca. 50 nm would contribute to the enhancement of LOD, which may result from a high Ka. This means that thicker MIP films probably include more heterogeneous sites for binding to target biomolecules50. The thickness and adhesiveness of MIP films at substrates can be precisely controlled to make them thinner (<50 nm) by some grafting methods, such as surface-initiated atom transfer radical polymerization (SI-ATRP), which results in a higher Ka51,112.
In contrast, a nonimprinted polymer (NIP) should be synthesized on the gate electrode as a control polymer by the same method as that for MIP except for adding a template molecule. Even NIPs may show some affinities with target biomolecules owing to their nonspecific adsorptions and other properties, resulting in a high Ka. The imprinting factor (IF) is often evaluated as one of the imprinting parameters for target biomolecules, \({{{{{\rm{IF}}}}}}={{K}}_{{{{{{\rm{a}}}}}}}^{{{{{{\rm{MIP}}}}}}}/{{K}}_{{{{{{\rm{a}}}}}}}^{{{{{{\rm{NIP}}}}}}}\)47,51,94,96,101,105,109,111,114,115. That is, a higher IF indicates better MIP performance. For instance, a non-glucose-selective NIP-FET against the glucose-selective MIP-FET mentioned above hardly showed glucose responsivity within the relatively wide range of concentrations, the \({{{{{{\rm{K}}}}}}}_{{{{{{\rm{a}}}}}}}^{{{{{{\rm{NIP}}}}}}}\) of which was difficult to analyze52; that is, \({{{{{{\rm{K}}}}}}}_{{{{{{\rm{a}}}}}}}^{{{{{{\rm{NIP}}}}}}}\) was low, resulting in a very high IF. This may be because the NIP suppresses not only the uptake of glucose molecules into itself owing to the higher cross-linking density but also the nonspecific adsorptions owing to its hydrophilicity based on HEMA. Note that a PBA-containing hydrogel-coated FET with a lower cross-linking density, the composition of which was almost the same as that of the non-glucose-selective NIP-FET, showed some electrical signals for the change in glucose concentrations91. Thus, the Ka values of MIPs and NIPs are the important parameters for controlling S, LOD and IF in the chemically synthesized electrical interface. The number of studies on MIPs is increasing yearly (Figure S1); their further applications to FET biosensors are expected in the future, the number of which is still small (Figure S2).
In this section, the chemically synthesized interfaces have been introduced to enhance the performance of FET biosensors, focusing on ISMs and MIPs. Functional compounds such as crown ether and PBA should be included in such polymeric membranes to induce the specificity of targets. In addition, poly- and bis (crown ether)s and the molecular imprinting increase Ka, which should result in a lower LOD and a higher S for targets. Note that such chemically synthesized interface materials should be designed with consideration for biocompatibility when used in direct contact with the human body as a wearable biosensor.
Physically and chemically structured electrical interfaces for FET-based biosensing
Nanostructured electrodes with nanopillars and nanopores may be effective for increasing S/N for a physically structured electrical interface (Fig. 1a). The increase in the surface area of a nanostructured electrode is expected to increase the intensity of output signals, but may also generate noise signals simultaneously. However, a previous paper showed that electrostatic screening is weaker in the vicinity of concave surfaces and stronger in the vicinity of convex surfaces75. That is, molecular charges tethered at the convex gate surface (e.g., nanopillars) are shielded by counterions in a sample solution, and thus no electrical signals may be induced in the FET biosensors. On the other hand, porous membranes that allow the filtration of biomolecules are available for size-exclusion-based separation methods in electrochemical and optical biosensing73,118. In addition to such methods, the concept of polymeric porous nanofilters on the FET biosensors is introduced here as a strategy for physically and chemically structured electrical interfaces, which allow the reduction in the intensity of nonspecific electrical signals from small biomolecules as well as biomacromolecules as interfering species, which results in the increase in S/N70,71,72.
Polymeric nanofilter-based structural interfaces
An Au electrode, which has a strong catalytic action, enhances the oxidation of organic compounds, including small biomarkers such as glucose119,120,121. Also, the Au surface should be oxidized upon exposure to UV/ozone122, so that the oxidized surface is easily reduced through the redox reaction at the surface with oxidizable biomolecules. Such Au-thin film is often utilized as the gate electrode of an extended-Au-gate–FET (EG-Au–FET), in which the gate electrode is extended from the metal gate of FET and can be handled separately from the FET. This is why not only are probe molecules easily modified on the Au gate electrode by −S–Au binding and so forth, but the Au electrode also enables the highly sensitive detection of small biomolecules owing to the redox reaction but without detection specificity or selectivity. Therefore, it is important to impart the selectivity for target detections on the Au surface; that is, only a small target biomolecule reaches and interacts with the Au surface, whereas interfering species should be filtered by some membranes on the Au surface. Here, a polymeric nanofilter interface is designed and synthesized on the Au gate electrode of an EG-Au–FET in accordance with this strategy, as shown in Fig. 4a70,71,72. The polymeric nanofilter interface, that is, the physically and chemically structured electrical interface, is composed of two layers: an anchor layer and a filter layer. The anchor layer is coated on the Au gate surface to be thinner under the filter layer, which determines the gap of the sensing capability at the Au substrate (X) and the density of the polymeric nanofilter (Y). The filter layer has the chromatographic effect, in which small interfering species are trapped but through which a small target biomolecule can reach the Au gate surface; that is, a higher S/N is obtained for the detection of a small target biomolecule. Note that potentiometric biosensors such as the FET biosensors detect changes in the density of ionic and molecular charges within the diffusion layer of a few nanometers on the detection surface, considering the Debye length limitation18,29,60,61,62,63,64,65,66,67,68. This is why the filter layer should be located at a distance larger than the Debye length. Thus, nonspecific signals generated in the filter layer are not physicochemically output as actual signals.
Surface modification by electrografting aryl diazonium salts on electrodes such as Au has attracted considerable attention because of the high chemical stability, the ease of functionalization, and the flexibility of surface design123,124,125. Aryl diazonium salts are reduced to the corresponding radicals at a reductive potential applied at the Au electrode. The generated radicals covalently react with the Au electrode surface to form a stable film, which tolerates UV irradiation126. Moreover, a multilayered aryl film is formed by chain reactions between unreacted radicals and benzene rings at the aryl film grafted on the Au electrode, the thickness of which can be controlled to be a few nanometers,70,71,125. The multilayered aryl film provides significant advantages in arranging the filter layer, contributing to the controls of not only the anchor layer thickness to be about the same as the Debye length but also the appropriate gap between grafted aryl molecules that allows a small target biomolecule to reach the Au surface. Also, the aryl films with different functionalities can be designed by modifying aryl diazonium salts with various functional groups. As an example, the aryl diazonium salt, which has a hydroxyl group at the para-substituent of the benzene ring, was utilized to develop a multilayered film for the polymeric nanofilter interface70. In this case, for polymerizing the filter layer, the surface was further functionalized by modifying the initiator of SI-ATRP via esterification reactions, although even only a multilayered aryl film grafted via aryl diazonium salt chemistry sufficiently suppressed the nonspecific electrical signals generated by the adhesion of large and small biomolecules to the Au surface71.
As shown in Fig. 4b, a polymeric nanofilter interface as a model was proposed to specifically detect l-cysteine as a small target on the Au surface and to effectively trap l-DOPA as a small interfering species in the filter layer70. In particular, PBA was incorporated in the filter layer to capture l-DOPA but not l-cysteine. PBA forms stable esters with diol compounds such as l-DOPA at equilibrium, as described in the above section. The binding affinity of catechol to PBA is 830 M–1 larger than that of other diol compounds at pH 7.478; thus, l-DOPA has a high affinity to PBA at pH 7.4. That is, l-DOPA is trapped in the filter layer with PBA, whereas l-cysteine reaches the Au surface. In this case, the potential may change owing to the anionic form of PBA–l-DOPA complex. However, the filter layer is located outside the detection range (i.e., the Debye length) of the EG-Au–FET sensor; thus, such unexpected noise components are not detected. Here, two types of polymeric nanofilters without and with PBA were designed to investigate the filtration ability of the polymeric nanofilter. First, an MAA-based polymer (control nanofilter) was designed, and then PBA was incorporated into the filter via amidation reactions (PBA-based nanofilter). In this model, a photo mediated SI-ATRP method was employed to control the thickness and composition because it was more tolerant to MAA polymerization than the conventional ATRP method127. In addition, HEMA was copolymerized to enhance the hydrophilicity of the filter layer, so that l-cysteine reached the Au electrode surface through the nanofilter as expected, whereas the nonspecific adsorption of biomacromolecules such as interfering proteins (e.g., albumin) was effectively suppressed. Lastly, the filter layer was cross-linked to form a rigid polymer structure using the hydrophilic cross-linker MBA. This is because the absence of cross-linking would generate electrical noise arising from changes in the conformation of the polymeric nanofilter induced by the binding of l-DOPA with PBA. Actually, the nonspecific signals based on l-DOPA were effectively reduced by ~70–90% (☐→ Δ), whereas l-cysteine was specifically detected at nM levels (♢) using the polymeric nanofilter-coated EG-Au-FET with PBA, as shown in Fig. 4c70. Thus, l-DOPA was successfully trapped by the nano filter outside the detection range of the EG-Au–FET; therefore, the polymeric nanofilter contributed to the increase in S/N for the detection of l-cysteine. Additionally, the polymeric nanofilter with hydrophilicity structurally excluded nonspecific noise signals from the interaction between a few biomacromolecules such as albumin and the Au electrode70,71,72.
Considering the above concept of a structural polymeric nanofilter interface, the relatively thicker MIPs mentioned in the above section may also function as the nanofilter at a distance from the gate surface above the Debye length. Moreover, they supply electrical charges based on the interaction with target biomolecules to the FET biosensor, which is captured in the cavities in the vicinity of the gate surface. This means that the thinner MIP film grafted by SI-ATRP on the gate surface should effectively contribute to the generation of electrical signals of the FET biosensor based on target biomolecules without any loss of undetected signals captured in it.
Aptamer nanofilter-based structural interfaces
Aptamers are also candidates for a polymeric nanofilter. For instance, DA and l-DOPA were discriminated form the electrical signals with the aptamer nanofilter-modified FET biosensor despite their similar chemical structures (Fig. 4d)55. That is, PBA, which has been described in the above section, cannot be used as a receptor in the nanofilter because it binds to catechol amines such as DA and l-DOPA. This is why an aptamer molecule such as deoxyribonucleic acid (DNA), ribonucleic acid (RNA), or a peptide can be used as one of the structural nanofilter interfaces, although it is mostly utilized as a receptor to detect a target. Some aptamers with different base sequences for DA recognition have been reported (e.g., 44 mer)54, whereas the base sequence of the l-DOPA aptamer is probably undetermined. The size of aptamer molecules can be relatively large (>a few nm), although it depends on the base length. Therefore, the DA aptamer can be modified on an anchor layer as the nanofilter to capture DA molecules in the undetectable region, where the height exceeds the Debye length on the EG-Au–FET70,124,125. Here, 1× PBS (pH 7.4) was employed as the measurement solution to demonstrate the concept of aptamer nanofilter as the chemically and physically structured interface55, for which the Debye length is calculated to be less than 1 nm. Actually, using the EG-Au–FET, the DA aptamer layer served as the nanofilter interface to prevent DA molecules from coming in contact with the sensing surface, and l-DOPA molecules passed through the DA aptamer nanofilter and then were specifically detected by the redox reactions with the sensing surface (Fig. 4d). Thus, the structural aptamer nanofilter interface led to the increase in S/N for the detection of l-DOPA.
On the other hand, aptamers are utilized as a receptors on electrode surfaces to specifically and selectively trap and detect target biomolecules by overcoming the Debye length limitation with the FET biosensors54. For instance, the negatively charged backbones of aptamers with a stem–loop structure are able to approach the gate surface owing to structural reorientation based on the selective binding with a target biomolecule; that is, the negative charges of aptamers enter the diffusion layer (i.e., the Debye length) that is less affected by counterions, resulting in the generation of electrical signals of the FET biosensors. Whether aptamers are utilized as a receptor to trap and detect target biomolecules at the electrode surface or to trap but disable interfering species in the nanofilters, we need to consider the Debye length limitation. That is, not only should the ionic strength in the measurement solutions be controlled to be lower or higher, but the thickness of the anchor layer should also be designed to be thinner or thicker.
In this section, the chemically and physically structured interfaces have been introduced to decrease nonspecific noise signals, that is, increase S/N in the FET biosensors. In general, hydrophilic polymers are coated to suppress nonspecific adsorptions of large interfering species such as proteins on the gate electrode. On the other hand, filtering membranes such as polymeric nanofilters can prevent even small interfering species from approaching the gate electrode surface. Increasing S/N is very important for the FET biosensors that directly detect targets with charges in real samples, including various interfering species with charges.
Biologically induced cellular electrical interfaces for FET-based biosensing
When living cells are confluently cultured on an electrode surface, the cell layer itself can serve as a biologically induced cellular electrical interface to detect analytes (Fig. 5). That is, target biomolecules specifically and selectively react with cell membrane proteins such as antibodies or are taken into cells through membrane proteins such as transporters. Then, such cells are mostly activated, depending on each cell function. Here, we need to consider what is transduced into electrical signals from such activated cells. As an example, rat basophilic leukemia (RBL-2H3) cells were cultured as a signal transduction interface to induce immunological reactions on the gate oxide surface of a FET biosensor (Fig. 5a)56,57. This is because IgE antibodies, which bind to Fcε receptors at the RBL-2H3 cell membrane, are specifically cross-linked with allergens (antigens), resulting in the immunological response of RBL-2H3 cells128. The specific binding of antigens to IgE is simply detected by immunological assays. However, such methods cannot be employed to determine whether the antigen-bound IgE activates basophils in patients. This means that what is more important in the diagnosis of type I allergy is to evaluate the potential of antigen-specific IgE to activate basophils from a patient rather than the detection of IgE–antigen binding only. In fact, secretions mainly including histamine, which were released during immunological reactions based on the IgE–antigen reaction, were indirectly detected as the change in pH with the FET biosensor56,57. Histamine is mainly released from RBL-2H3 cells in a relatively short time through immunological reactions and shows basicity (pKa 9.75) in particular. An immunological response was observed within 10 min upon adding the antigen [dinitrophenyl-conjugated human serum albumin (DNP-HSA)] to RBL-2H3 cells with the FET biosensor; therefore, basic histamine had the greatest impact on electrical signals. This is why secreted histamines are considered to be close to the cell/gate interface, resulting in an increase in pH around it. The FET biosensor used was based on a pH-sensitive FET, the gate of which was composed of Ta2O5 with hydroxyl groups in a solution1,5,7. The pH-sensitive FETs used showed a response of 55.8 mV/pH, which was near the Nernstian response. Here, RBL-2H3 cells were cultured on the gate oxide surface, so we need to consider a gap of approximately 50–150 nm between the cells and the gate oxide. The pH behavior at the cell/substrate nanogap was observed by laser scanning confocal microscopy using phospholipid fluorescein inserted at the cell membrane (Fig. 5b)25,57. In addition, the cell culture gate FET biosensors continuously monitored the change in pH on the basis of cellular respiration (metabolism) at the cell/gate nanogap interface for a few days to one month24,31,32,129,130. Some proteins contained in a cell culture medium are nonspecifically adsorbed at the gate oxide surface during preculture, contributing to the adhesion of cells at the gate oxide surface. These biomacromolecules prevent target biomolecules with charges from approaching the gate oxide surface, but small hydrogen ions easily pass through such biomacromolecules and then react hydroxy groups at the gate oxide surface. This means that the change in pH at the cell/gate nanogap interface contributes to the change in Vout according to the principle of FET biosensors. In addition, hydrogen ions should be effectively concentrated in the closed nanogap space between the cell membrane and the gate surface, which can be easily detected. This is the simplest way to use pH-sensitive FETs129.
As another example of biologically induced cellular electrical interfaces, vascular endothelial cells were confluently cultured as a cell layer on the gate electrode surface of a FET biosensor to monitor the invasion of cancer cells (Fig. 5c)59. Metastatic cancer cells adhere to vascular endothelial cells and then invade into surrounding tissues through basement membrane, which results in metastasis58. Analytical methods to monitor metastasis in real time contribute to not only the early diagnosis of cancer but also the clarification of the mechanism of metastasis. Here, vascular endothelial cells were utilized as a biologically induced interface material on the FET biosensor to simulate the interaction with cancer cells in blood. Actually, the invasion process of cancer cells (HeLa cells) into the vascular endothelial cell [human coronary artery endothelial (HCAE) cell] layer was continuously monitored in the cell culture medium (pH 7.4) using the FET biosensor59. Note that the basement membrane, which is a thin and fibrous extracellular matrix of tissue composed of collagen IV and so forth and negatively charged at pH 7.4, should be formed on the gate oxide surface by HCAE cells131. The basement membrane is also known to be decomposed by collagenase secreted by cancer cells during the invasion process132. As a result, Vout gradually increased after adding HeLa cells onto the HCAE-cell–FET, which resulted from the change in the density of some kind of electrical charge on the gate electrode, but Vout for the HCAE-cell–FET without HeLa cells hardly changed because only the culture medium was added (right graph in Fig. 5c)59. From the change in the fluorescence image of collagen IV in the basement membrane, furthermore, collagen IV was observed under the HCAE cell layer before adding Hela cells, whereas it was partially decomposed after introducing HeLa cells onto HCAE cells. Thus, HeLa cells are considered to have passed between HCAE cells and then decomposed the basement membrane by secreting a collagen-IV-decomposing enzyme. Considering the electrical signal measurement and fluorescence observation results, Vout must have increased by the decomposition of the negatively charged basement membrane, that is, the invasion of cancer cells, corresponding to the decrease in the density of negative charges. Thus, vascular endothelial cells with the basement membrane serve as the biologically induced cellular electrical interface that provides the specificity and S for the detection of cancer cells as a target and induces the generation of electrical signals.
In this section, we have introduced that unique cellular functions contribute to the specificity to and S for targets detected. Such biologically derived functional materials are useful as the biologically induced interfaces in the FET biosensors, including enzymes and antibodies.
Conclusion and outlook
In developing a biosensor, we consider the design criteria based on its three components, namely, the biological target, signal transduction interface, and detection device. Among the detection devices, a platform based on an electronic device with the FET biosensors is suitable for use in miniaturized and cost-effective systems to directly measure biological samples because the FET biosensors allow the direct detection of intrinsic ionic and biomolecular charges in principle, which contributes to label- and enzyme-free biosensing. Such miniaturized electronic devices can be easily equipped with a wireless function and attached to the body, which is available for wearable biosensors to detect biomarkers in a blood-sampling-free manner (i.e., tears, sweat, and saliva). Here, it is very important to determine how the change in the density of charges based on biomolecular recognition events is directly transduced into electrical signals at the signal transduction interfaces, regardless of the wearability of FET biosensors. Such bio/device interfaces are chemically synthesized, physically and chemically structured, and biologically induced to control the biosensing parameters such as specificity, S, Ka, LOD, S/N, and biocompatibility with respect to the biological target, although the chemically synthesized electrical interfaces are also useful as the signal transduction interfaces for the flexible wearable biosensors. In particular, the increase in Ka for the target biological target, which results in the enhancement of S and LOD, becomes a key challenge for enzyme-free interfaces, and then biocompatible materials may be chosen for the signal transduction interfaces. On the other hand, various semiconductor materials have been recently applied as the channel of FET biosensors. Therefore, the diversity of signal transduction interfaces broadens the possibility of developing novel biosensing devices, in parallel with the development of new channel materials for the FET biosensors. In addition, the steep SS, which is one of the transistor characteristics, should lead to an increase in the sensitivity of biosensing, although it is a key parameter of FETs (see 1. Introduction)27,31, and then the optimal biosensing parameters are effectively provided by the diverse signal transduction interfaces for the practical use of FET biosensors.
In addition, the FET biosensors can be applied to semiconductor integrated circuits to measure multiple samples simultaneously. This is one of the advantages of utilizing semiconductor technology and also the unique feature of FETs because other biosensors (e.g., surface plasmon resonance and quartz crystal microbalance sensors) hardly enable the integration of electrodes as in complementary metal oxide semiconductor sensors21. That is, it will be a challenge to coat and arrange different signal transduction interface materials for various biomarkers on the individual gate electrodes in the arrayed devices in the future. This means that the methods to analyze a huge amount of data, including complicated information, must be required, such as the omics approaches.
References
Sakata, T. Biologically coupled gate field-effect transistors meet in vitro diagnostics. ACS Omega 4, 11852–11862 (2019).
Kim, J. et al. Wearable salivary uric acid mouthguard biosensor with integrated wireless electronics. Biosens. Bioelectron. 74, 1061–1068 (2015).
Gao, W. et al. Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis. Nature 529, 509–514 (2016).
Park, J. et al. Soft, smart contact lenses with integrations of wireless circuits, glucose sensors, and displays. Sci. Adv. 4, 9841 (2018).
Matsuo, T. & Wise, K. D. An integrated field-effect electrode for biopotential recording. IEEE Trans. Biomed. Eng. BME-21, 485–487 (1974).
Buck, R. P. & Hackleman, D. E. Field effect potentiometric sensors. Anal. Chem. 49, 2315–2321 (1977).
Bergveld, P. The impact of MOSFET-based sensors. Sens. Actuators 8, 109–127 (1985).
Sakata, T., Kamahori, M. & Miyahara, Y. DNA analysis chip based on field effect transistors. Jpn. J. Appl. Phys. 44, 2854–2859 (2005).
Ammann, D., Pretsch, E. & Simon, W. A sodium ion-selective electrode based on a neutral carrier. Anal. Lett. 7, 23–32 (1974).
Moss, S. D., Janata, J. & Johnson, C. C. Potassium ion-sensitive field effect transistor. Anal. Chem. 47, 2238–2243 (1975).
Oesch, U., Caras, S. & Janata, J. Field effect transistors sensitive to sodium and ammonium ions. Anal. Chem. 53, 1983–1986 (1981).
Zdrachek, E. & Bakker, E. Potentiometric sensing. Anal. Chem. 91, 2–26 (2019).
Ito, K., Satake, H., Mori, Y., Tseng, A. C. & Sakata, T. Biocompatible and Na+-sensitive thin-film transistor for biological fluid sensing. Sci. Technol. Adv. Mater. 20, 917–926 (2019).
Fakih, I. et al. Selective ion sensing with high resolution large area graphene field effect transistor arrays. Nat. Commun. 11, 3226 (2020).
Shao, Y., Ying, Y. & Ping, J. Recent advances in solid-contact ion-selective electrodes: functional materials, transduction mechanisms, and development trends. Chem. Soc. Rev. 49, 4405–4465 (2020).
Danielsson, B., Lundström, I., Mosbach, K. & Stiblert, L. On a new enzyme transducer combination: the enzyme transistor. Anal. Lett. 12, 1189–1199 (1979).
Souteyrand, E. et al. Direct detection of the hybridization of synthetic homo-oligomer DNA sequences by field. Eff. J. Phys. Chem. B 101, 2980–2985 (1997).
Sakata, T. & Miyahara, Y. DNA sequencing based on intrinsic molecular charges. Angew. Chem. Int. Ed. 45, 2225–2228 (2006).
Ingebrandt, S. et al. Label-free detection of single nucleotide polymorphisms utilizing the differential transfer function of field-effect transistors. Biosens. Bioelectron. 22, 2834–2840 (2007).
Stern, E. et al. Importance of the Debye screening length on nanowire field effect transistor sensors. Nano Lett. 7, 3405–3409 (2007).
Rothberg, J. M. et al. An integrated semiconductor device enabling non-optical genome sequencing. Nature 475, 348–352 (2011).
Fromherz, P., Offenhäusser, A., Vetter, T. & Weis, J. A neuron-silicon junction: a Retzius cell of the leech on an insulated-gate field-effect transistor. Science 252, 1290–1293 (1991).
Offenhäusser, A. & Knoll, W. Cell-transistor hybrid systems and their potential applications. Trends Biotechnol. 19, 62–66 (2001).
Sakata, T. et al. Single embryo-coupled gate field effect transistor for elective single embryo transfer. Anal. Chem. 85, 6633–6638 (2013).
Satake, H., Saito, A. & Sakata, T. Elucidation of interfacial pH behaviour at cell/substrate nanogap for in situ monitoring of cellular respiration. Nanoscale 10, 10130–10136 (2018).
Chen, R. J. et al. Noncovalent functionalization of carbon nanotubes for highly specific electronic biosensors. Proc. Nat. Acad. Sci. USA 100, 4984–4989 (2003).
Patolsky, F. et al. Detection, stimulation, and inhibition of neuronal signals with high-density nanowire transistor arrays. Science 313, 1100–1104 (2006).
Nebel, C. E., Rezek, B., Shin, D., Uetsuka, H. & Yang, N. Diamond for bio-sensor applications. J. Phys. D: Appl. Phys. 40, 6443 (2007).
Ohno, Y., Maehashi, K., Yamashiro, Y. & Matsumoto, K. Electrolyte-gated graphene field-effect transistors for detecting ph and protein adsorption. Nano Lett. 9, 3318–3322 (2009).
Sarkar, D. et al. MoS2 field-effect transistor for nextgeneration label-free biosensors. ACS Nano 8, 3992–4003 (2014).
Sakata, T., Nishitani, S., Saito, A. & Fukasawa, Y. Solution-gated ultrathin channel indium tin oxide-based field-effect transistor fabricated by one-step procedure that enables high-performance ion sensing and biosensing. ACS Appl. Mater. Interfaces 13, 38569–38578 (2021).
Sakata, T. et al. Ion sensitive transparent-gate transistor for visible cell sensing. Anal. Chem. 89, 3901–3908 (2017).
Turner, A. P. F., Kaube, I. & Wilson, G. Biosensors: fundamentals and applications. Oxford University Press (1987).
Updike, S. J. & Hicks, G. P. The enzyme electrode. Nature 214, 986–988 (1967).
Bergenstal, R. M. Continuous glucose monitoring: transforming diabetes management step by step. Lancet 391, 1334–1336 (2018).
Free, A. H., Adams, E. C., Kercher, M. L., Free, H. M. & Cook, M. H. Simple specific test for urine glucose. Clin. Chem. 3, 163–168 (1957).
Charlton, D. E. Test device and method for colored particle immunoassay. U. S. Patent, US6485982B1 (2002).
Wulff, G. & Sarhan, A. Über die anwendung von enzymanalog gebauten polymeren zur racemattrennung. Angew. Chem. 84, 364 (1972).
Andersson, L., Ekberg, B. & Mosbach, K. Imprinting of amino acid derivatives in macroporous polymers. Tetrahedron Lett. 26, 3623–3624 (1985).
Sellergren, B., Ekberg, B. & Mosbach, K. Molecular imprinting of amino acid derivatives in macroporous polymers: demonstration of substrate- and enantio-selectivity by chromatographic resolution of racemic mixtures of amino acid derivatives. J. Chromatogr. A 347, 1–10 (1985).
Wulff, G. & Schauhoff, S. Racemic resolution of free sugars with macroporous polymers prepared by molecular imprinting. selectivity dependence on the arrangement of functional groups versus. Spat. Requir. J. Org. Chem. 56, 395–400 (1991).
Takeuchi, T. et al. Molecularly imprinted tailor-made functional polymer receptors for highly sensitive and selective separation and detection of target molecules. Chromatography 37, 43–64 (2016).
Sakata, T., Nishitani, S. & Kajisa, T. Molecularly imprinted polymer-based bioelectrical interfaces with intrinsic molecular charges. RSC Adv. 10, 16999–17013 (2020).
Tuerk, C. & Gold, L. Systematic evolution of ligands by exponential enrichment: RNA ligands to bacteriophage T4 DNA polymerase. Science 249, 505–510 (1990).
Ellington, A. D. & Szostak, J. W. In vitro selection of RNA molecules that bind specific ligands. Nature 346, 818–822 (1990).
Bock, L. C., Griffin, L. C., Latham, J. A., Vermaas, E. H. & Toole, J. J. Selection of single-stranded DNA molecules that bind and inhibit human thrombin. Nature 355, 564–566 (1992).
Iskierko, Z. et al. Extended-gate field-effect transistor (EG-FET) with molecularly imprinted polymer (MIP) film for selective inosine determination. Biosens. Bioelectron. 74, 526–533 (2015).
Dabrowski, M. et al. Early diagnosis of fungal infections using piezomicrogravimetric and electric chemosensors based on polymers molecularly imprinted with d-arabitol. Biosens. Bioelectron. 79, 627–635 (2016).
Nishitani, S., Kajisa, Y. & Sakata, T. Development of molecularly imprinted polymer-based field effect transistor for sugar chain sensing. Jpn. J. Appl. Phys. 56, 04CM02 (2017).
Nishitani, S. & Sakata, T. Potentiometric adsorption isotherm analysis of a molecularly imprinted polymer interface for small-biomolecule recognition. ACS Omega 3, 5382–5389 (2018).
Kajisa, T., Li, W., Michinobu, T. & Sakata, T. Well-designed dopamine-imprinted polymer interface for selective and quantitative dopamine detection among catecholamines using a potentiometric biosensor. Biosens. Bioelectron. 117, 810–817 (2018).
Kajisa, T. & Sakata, T. Molecularly imprinted artificial biointerface for an enzyme-free glucose transistor. ACS Appl. Mater. Interfaces 10, 34983–34990 (2018).
Yang, H., Nishitani, S. & Sakata, T. Potentiometric langmuir isotherm analysis of histamine-selective molecularly imprinted polymer-based field-effect transistor. ECS J. Solid State Sci. Technol. 7, Q3079–Q3082 (2018).
Nakatsuka, N. et al. Aptamer-field-effect transistors overcome debye length limitations for small-molecule sensing. Science 362, 319–324 (2018).
Himori, S., Nishitani, S. & Sakata, T. Aptamer-based nanofilter interface for small-biomarker detection with potentiometric biosensor. Electrochim. Acta 368, 137631 (2021).
Yang, H. et al. Non-optical detection of allergic response with a cell-coupled gate field-effect transistor. Anal. Chem. 89, 12918–12923 (2017). An example of biologically induced interfaces is introduced in this paper.
Satake, H. & Sakata, T. Interfacial pH behavior at cell/gate insulator nanogap induced by allergic responses. ACS Omega 4, 14255–14260 (2019).
Chambers, A. F., Groom, A. C. & MacDonald, I. C. Dissemination and growth of cancer cells in 9 metastatic sites. Nat. Rev. Cancer 2, 563–572 (2002).
Sakata, T. & Matsuse, Y. In situ electrical monitoring of cancer cells invading vascular endothelial cells with semiconductor-based biosensor. Genes Cells 22, 203–209 (2017).
Schasfoort, R. B. M., Bergveld, P., Kooyman, R. P. H. & Greve, J. Possibilities and limitations of direct detection of protein charges by means of an immunological field-effect transistor. Anal. Chim. Acta 238, 323–329 (1990).
Poghossian, A., Cherstvy, A., Ingebrandt, S., Offenhäuser, A. & Schöning, M. J. Possibilities and limitations of label-free detection of DNA hybridization with field-effect-based devices. Sens. Actuators B Chem. 111–112, 470–480 (2005).
Sakata, T. & Miyahara, Y. Direct transduction of primer extension into electrical signal using genetic field effect transistor. Biosens. Bioelectron. 22, 1311–1316 (2007).
Maekawa, Y., Shibuta, Y. & Sakata, T. Distinctive potential behavior at the oxidized surface of a semiconductor device in a concentrated aqueous salt solution. ChemElectroChem 1, 1516–1524 (2014).
Maekawa, Y., Shibuta, Y. & Sakata, T. Effect of Double-Stranded DNA on Electrical Double Layer Structure at Oxide/Electrolyte Interface in Classical Molecular Dynamics Simulation. Chem. Phys. Lett. 619, 152–157 (2015).
Maekawa, Y., Shibuta, Y. & Sakata, T. Effect of ionic atmosphere around DNA/electrolyte interface on potentiometric signal. J. Electrochem. Soc. 164, B548–B552 (2017).
Lowe, B. M. et al. Dynamic behaviour of the silica-water-bio electrical double layer in the presence of a divalent electrolyte. Phys. Chem. Chem. Phys. 19, 2687–2701 (2017).
Lowe, B. M., Skylaris, C.-K., Green, N. G., Shibuta, Y. & Sakata, T. Molecular dynamics simulation of potentiometric sensor response: the effect of biomolecules, surface morphology and surface charge. Nanoscale 10, 8650–8666 (2018).
Kim, Y., Inoue, Y., Hasegawa, H., Yoshida, Y. & Sakata, T. In situ electrical monitoring of methylated DNA based on its conformational change to G-quadruplex using solution-gated field-effect transistor. Anal. Chem. 93, 16709–16717 (2021).
Banerjee, I., Pangule, R. C. & Kane, R. S. Antifouling coatings: recent developments in the design of surfaces that prevent fouling by proteins, bacteria, and marine organisms. Adv. Mater. 23, 690–718 (2011).
Nishitani, S. & Sakata, T. Polymeric nanofilter biointerface for potentiometric small-biomolecule recognition. ACS Appl. Mater. Interfaces 11, 5561–5569 (2019).
Nishitani, S. & Sakata, T. Enhancement of signal-to-noise ratio for serotonin detection with well-designed nanofilter-coated potentiometric electrochemical biosensor. ACS Appl. Mater. Interfaces 12, 14761–14769 (2020).
Nishitani, S., Man, Y., Noguchi, T. & Sakata, T. Polyaniline-based nanostructure interface for signal-to-noise ratio enhancement in potentiometric enzyme-free biosensors. Adv. Mater. Interfaces 9, 2201029 (2022).
del Río, J. S., Henry, O. Y. F., Jolly, P. & Ingber, D. E. An antifouling coating that enables affinity-based electrochemical biosensing in complex biological fluids. Nat. Nanotechnol. 14, 1143–1149 (2019).
Anandan, V., Rao, Y. L. & Zhang, G. Nanopillar array structures for enhancing biosensing performance. Int. J. Nanomed. 1, 73–79 (2006).
Shoorideha, K. & Chui, C. O. On the origin of enhanced sensitivity in nanoscale FET-based biosensors. Proc. Natl Acad. Sci. 111, 5111–5116 (2014).
Park, Y. M. et al. Flexible nanopillar-based immunoelectrochemical biosensor for noninvasive detection of Amyloid beta. Nano Converg. 7, 29 (2020).
Lorand, J. P. & Edwards, J. O. Polyol Complexes and Structure of the Benzeneboronate Ion. J. Org. Chem. 24, 769–774 (1959).
Springsteen, G. & Wang, B. A detailed examination of boronic acid–diol complexation. Tetrahedron 58, 5291–5300 (2002).
Kolthoff, I. M. & Chantooni, M. K. Jr Transfer activity coefficients in various solvents of several univalent cations complexed with dibenzo-18-crown-6. Anal. Chem. 52, 1039–1044 (1980).
Kimura, K., Tsuchida, T., Maeda, T. & Shono, T. Properties of cis- and trans-bis(crown ether)s for complexation and extraction of alkali metal picrates. Talanta 27, 801–805 (1980).
Högg, G., Lutze, O. & Cammann, K. Novel membrane material for ion-selective field-effect transistors with extended lifetime and improved selectivity. Anal. Chim. Acta 335, 103–109 (1996).
Cazalé, A., Sant, W., Launay, J., Ginot, F. & Temple-Boyer, P. Study of field effect transistors for the sodium ion detection using fluoropolysiloxane-based sensitive layers. Sens. Actuators B Chem. 177, 515–521 (2013).
Sakata, T. et al. Biocompatible and flexible paper-based metal electrode for potentiometric wearable wireless biosensing. Sci. Technol. Adv. Mater. 21, 379–387 (2020).
Pawlak, M., Grygolowicz-Pawlak, E., Crespo, G. A., Mistlberger, G. & Bakker, E. PVC-based ion-selective electrodes with enhanced biocompatibility by surface modification with “click” chemistry. Electroanalysis 25, 1840–1846 (2013).
Qia, L., Jianga, T., Lianga, R. & Qin, W. Polymeric membrane ion-selective electrodes with anti-biofouling properties by surface modification of silver nanoparticles. Sens. Actuators B Chem. 328, 129014 (2021).
Lotierzo, M. et al. Surface plasmon resonance sensor for domoic acid based on grafted imprinted polymer. Biosens. Bioelectron. 20, 145–152 (2004).
Luo, Q., Yu, N., Shi, C., Wang, X. & Wu, J. Surface plasmon resonance sensor for antibiotics detection based on photo-initiated polymerization molecularly imprinted array. Talanta 161, 797–803 (2016).
Diltemiz, S. E., Keçili, R., Ersöz, A. & Say, R. Molecular imprinting technology in quartz crystal microbalance (QCM). Sens. Sens. 17, 454 (2017).
Okan, M., Sari, E. & Duman, M. Molecularly imprinted polymer based micromechanical cantilever sensor system for the selective determination of ciprofloxacin. Biosens. Bioelectron. 88, 258–264 (2017).
García-Calzón, J. A. & Díaz-García, M. E. Characterization of binding sites in molecularly imprinted polymers. Sens. Actuators B Chem. 123, 1180–1194 (2007).
Kajisa, T. & Sakata, T. Glucose responsive hydrogel electrode for biocompatible glucose transistor. Sci. Technol. Adv. Mater. 18, 26–33 (2017).
Furikado, Y. et al. Universal reaction mechanism of boronic acids with diols in aqueous solution: kinetics and the basic concept of a conditional formation constant. Chem. –Eur. J. 20, 13194–13202 (2014).
Stanley, B. J., Szabelski, P., Chen, Y.-B., Sellergren, B. & Guiochon, G. Affinity distributions of a molecularly imprinted polymer calculated numerically by the expectation-maximization method. Langmuir 19, 772–778 (2003).
Curk, T., Dobnikar, J. & Frenkel, D. Rational design of molecularly imprinted polymers. Soft Matter 12, 35–44 (2016).
Malitesta, C. et al. MIP sensors – the electrochemical approach. Anal. Bioanal. Chem. 402, 1827–1846 (2012).
Waffo, A. F. T. et al. Fully electrochemical MIP sensor for artemisinin. Sens. Actuators B Chem. 275, 163–173 (2018).
Sehit, E. et al. Ultrasensitive nonenzymatic electrochemical glucose sensor based on gold nanoparticles and molecularly imprinted polymers. Biosens. Bioelectron. 165, 112432 (2020).
Shen, M. & Kan, X. Aptamer and molecularly imprinted polymer: synergistic recognition and sensing of dopamine. Electrochim. Acta 367, 137433 (2021).
Yeasmin, S., Wu, B., Liu, Y., Ullah, A. & Cheng, L.-J. Nano gold-doped molecularly imprinted electrochemical sensor for rapid and ultrasensitive cortisol detection. Biosens. Bioelectron. 206, 114142 (2022).
Goyal, A. & Sakata, T. Development of a redox-label-doped molecularly imprinted polymer on β-cyclodextrin/reduced graphene oxide for electrochemical detection of a stress biomarker. ACS Omega 7, 33491–33499 (2022).
Majd, S. M. et al. Design of a novel aptamer/molecularly imprinted polymer hybrid modified Ag–Au@Insulin nanoclusters/Au-gate-based MoS2 nanosheet field-effect transistor for attomolar detection of BRCA1 gene. Talanta 257, 124394 (2023).
Ota, S., Yui, Y., Sato, T., Yoshimoto, N. & Yamamoto, S. Rapid purification of immunoglobulin G using a protein a-immobilized monolithic spin column with hydrophilic polymers. Anal. Sci. 37, 985–990 (2021).
Aoki, T. et al. Glucose-Sensitive lower critical solution temperature changes of copolymers composed of N-isopropylacrylamide and phenylboronic acid moieties. Polym. J. 28, 371–374 (1996).
Piletska, E., Piletsky, S., Karim, K., Terpetschnig, E. & Turner, A. Biotin-specific synthetic receptors prepared using molecular imprinting. Anal. Chim. Acta 504, 179–183 (2004).
da Silva, M. S., Viveiros, R., Aguiar-Ricardo, A., Bonifácio, V. D. B. & Casimiro, T. Supercritical fluid technology as a new strategy for the development of semi-covalent molecularly imprinted materials. RSC Adv. 2, 5075–5079 (2012).
Pesavento, M., Marchetti, S., de Maria, L., Zeni, L. & Cennamo, N. Sensing by molecularly imprinted polymer: evaluation of the binding properties with different techniques. Sensors 19, 1344 (2019).
Sönmezler, M., Özgür, E., Yavuz, H. & Denizli, A. Quartz crystal microbalance based histidine sensor. Artif. Cells Nanomed. Biotechnol. 47, 221–227 (2019).
Kaiser, H. Die Berechnung der nachweisempfindlichkeit. Spectrochim. Acta 3, 40–67 (1947).
Caldara, M. et al. Recent advances in molecularly imprinted polymers for glucose monitoring: from fundamental research to commercial application. Chemosensors 11, 32 (2023).
Ersöz, A., Denizli, A., Özcan, A. & Say, R. Molecularly imprinted ligand-exchange recognition assay of glucose by quartz crystal microbalance. Biosens. Bioelectron. 20, 2197–2202 (2005).
Dai, C.-M., Geissen, S.-U., Zhang, Y.-L., Zhang, Y.-J. & Zhou, X.-F. Selective removal of diclofenac from contaminated water using molecularly imprinted polymer microspheres. Environ. Pollut. 159, 1660–1666 (2011).
Suda, N., Sunayama, H., Kitayama, Y., Kamon, Y. & Takeuchi, T. Oriented, molecularly imprinted cavities with dual binding sites for highly sensitive and selective recognition of cortisol. R. Soc. Open Sci. 4, 170300 (2017).
El-Sharif, H. F., Turner, N. W., Reddy, S. M. & Sullivan, M. V. Application of thymine-based nucleobase-modified acrylamide as a functional co-monomer in electropolymerised thin-film molecularly imprinted polymer (MIP) for selective protein (Haemoglobin) binding. Talanta 240, 123158 (2022).
Bartold, K. et al. Molecularly imprinted polymer-based extended-gate field-effect transistor (EG-FET) chemosensor for selective determination of matrix metalloproteinase-1 (MMP-1) protein. Biosens. Bioelectron. 208, 114203 (2022).
Casey, C. N., Campbell, S. E. & Gibson, U. J. Phenylalanine detection using matrix assisted pulsed laser evaporation of molecularly imprinted amphiphilic block copolymer films. Biosens. Bioelectron. 26, 703–709 (2010).
Jolly, P. et al. Aptamer–MIP hybrid receptor for highly sensitive electrochemical detection of prostate specific antigen. Biosens. Bioelectron. 75, 188–195 (2016).
Kim, D.-M. et al. A potentiometric non-enzymatic glucose sensor using a molecularly imprinted layer bonded on a conducting polymer. Biosens. Bioelectron. 91, 276–283 (2017).
Bonanno, L. M. & DeLouise, L. A. Whole blood optical biosensor. Biosens. Bioelectron. 23, 444–448 (2007).
Hutchings, G. Vapor phase hydrochlorination of acetylene: correlation of catalytic activity of supported metal chloride catalysts. J. Catal. 96, 292–295 (1985).
Hashmi, A. S. K. Homogeneous gold catalysis – still accelerating. J. Organomet. Chem. 694, 481–598 (2009).
Zope, B. N., Hibbitts, D. D., Neurock, M. & Davis, R. J. Reactivity of the gold/water interface during selective oxidation catalysis. Science 330, 74–78 (2010).
King, D. E. Oxidation of gold by ultraviolet light and ozone at. 25°C. J. Vac. Sci. Technol. A 13, 1247–1253 (1995).
Cao, C., Zhang, Y., Jiang, C., Qi, M. & Liu, G. Advances on aryldiazonium salt chemistry based interfacial fabrication for sensing applications. ACS Appl. Mater. Interfaces 9, 5031–5049 (2017).
Himori, S., Nishitani, S. & Sakata, T. Control of potential response to small biomolecules with electrochemically grafted aryl-based monolayer in field-effect transistor-based sensors. Langmuir 35, 3701–3709 (2019).
Nishitani, S. et al. Densification of diazonium-based organic thin film as bioelectrical interface. Langmuir 37, 14369–14379 (2021).
Adali-Kaya, Z., Tse Sum Bui, B., Falcimaigne-Cordin, A. & Haupt, K. Molecularly imprinted polymer nanomaterials and nanocomposites: atom-transfer radical polymerization with acidic monomers. Angew. Chem. Int. Ed. 54, 5192–5195 (2015).
Fors, B. P. & Hawker, C. J. Control of a living radical polymerization of methacrylates by light. Angew. Chem. Int. Ed. 51, 8850–8853 (2012).
Yanase, Y. et al. Application of SPR imaging sensor for detection of individual living cell reactions and clinical diagnosis of type I allergy. Allergol. Int. 62, 163–169 (2013).
Sakata, T., Saito, A. & Sugimoto, H. Live monitoring of microenvironmental ph based on extracellular acidosis around cancer cells with cell-coupled gate ion-sensitive field-effect transistor. Anal. Chem. 90, 12731–12736 (2018).
Satake, H. & Sakata, T. Estimation of extracellular matrix production using cultured-chondrocyte-based gate ion-sensitive field-effect transistor. Anal. Chem. 91, 16017–16022 (2019).
Orkin, R. W. et al. A murine tumor producing a matrix of basement membrane. J. Exp. Med. 145, 204–220 (1977).
Liotta, L. A. et al. Biochemical mechanisms of tumor invasion and metastasis. Anticancer Drug Des. 2, 195–202 (1987).
Acknowledgements
We would like to thank all the members of Sakata Laboratory for their help and useful discussion.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Competing interests
The author declares no competing interests.
Peer review
Peer review information
Communications Chemistry thanks Jianfeng Ping and the other anonymous, reviewers for their contribution to the peer review of this work. A peer review file is available.
Additional information
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Sakata, T. Signal transduction interfaces for field-effect transistor-based biosensors. Commun Chem 7, 35 (2024). https://doi.org/10.1038/s42004-024-01121-6
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s42004-024-01121-6
- Springer Nature Limited