Background

Enteroviruses (EVs) are among the most common and important viruses infecting humans. EVs are associated with diverse clinical syndromes, ranging from mild febrile illness to severe central nervous system diseases, such as aseptic meningitis and encephalitis, potentially leading to paralysis [1, 2]. Neonates and young children are at the greatest risk of developing severe, and occasionally fatal, enteroviral infections [3, 4]. Serotypes of EVs have traditionally been classified into echoviruses, coxsackieviruses, groups A and B, and polioviruses [5]. Currently, EV subtypes are divided into five species (human enteroviruses [HEV] A, B, C, D, and poliovirus) with differing molecular and biological characteristics [6]. Laboratory methods used for the diagnosis of enteroviral infection have changed substantially over time [79]. Initially, EVs were detected exclusively by cell culture and identified by neutralization [10]. In the mid-1990s, polymerase chain reaction (PCR) methods that can detect all EVs were introduced and have supplanted cell culture in many diagnostic laboratories [1113]. Cell culture methods for the detection of EVs are time-consuming, requiring, on average, 7-14 days for identification. Also, most coxsackievirus group A viruses do not adapt to cells as well as other EVs [14]. Although cell culture remains the gold standard for the identification of EVs in suspected patients, molecular methods such as reverse transcription PCR (RT-PCR), real-time RT-PCR, and nucleic acid sequence-based amplification offer more sensitive, specific, and rapid results [1517]. However diagnosis by RT-PCR also has problems, including contamination from post-reaction handling and variation in results depending on laboratory staff. Several groups have described real-time RT-PCR methods for the detection of EVs in cerebrospinal fluid (CSF) [1823].

This study was developed and validated a rapid, sensitive, and reliable real-time RT-PCR assay for the routine identification of EVs using the TaqMan minor groove binder (MGB) format combined complementary locked primer (CLP) technology [24]: the TMC-PCR assay [25, 26]. After experiments to evaluate its analytical sensitivity, specificity, and reproducibility, it was used with clinical specimens from patients and the results were compared to those using a previously published TaqMan probe real-time one step RT-PCR (TTN-PCR [23]) assay.

Methods

Viruses and Controls

Five reference strains belonging to distinct genogroups [enterovirus 71 (EV71), coxsackievirus B2 (CVB2), echo 30 (E30), coxsackievirus A24 (CVA24), and poliovirus type 1 (P1)] were obtained from the American Type Culture Collection (ATCC) and were used to optimize TTN-PCR conditions and to evaluate analytical performance. Infectivity of viruses was assayed in microplates in serial 10-fold dilutions (from 10-4 to 10-10) with four wells per dilution. TCID50 titers were calculated according to the Kärber method [27]. Enteroviral isolates including 25 serotypes (12 echovirus (E1, 3, 5-7, 9, 13, 14, 16, 1, 25 and 30), four coxsackievirus A (CVA 10, 16, 22 and 24), six coxsackievirus B (CVB 1-6), poliovirus type 1 (P1) and two new enterovirus (EV71 and 74) circulating between 1997 and 2005 in Korea were used to evaluate the reactivity of the assay to various serotypes of EVs. Enterovirus Proficiency panels from Quality Control on Molecular diagnostics (QCMD-2007) were also included to compare results from both assays.

Clinical samples

In total, 158 clinical specimens, collected from patients with suspected viral meningitis between June and September 2008, were included for evaluation with both real-time PCR assays.

Extraction of viral RNA

RNA was extracted from 150 μL samples with the GM Viral Nucleic Acid Extraction Kit (GreenMate Biotech Corp, Korea), according to the manufacturer's protocol using automated machines for liquid handling (Tecan, Switzerland). The GM Viral Nucleic Acid Extraction Kit uses a silica-based extraction method [28]. RNA was then recovered in 50 μL of nuclease-free water. It was used immediately or stored at -70°C.

Primers and probe

The TMC-PCR assay for detecting EVs is based on TaqMan technology. However, to improve the sensitivity and specificity for all enteroviruses, the primers and probe described by Verstrepen et al. was modified [23]. The primer pair was modified with CLP technology (iNtRON Biotechnology, Korea); the forward and reverse primers had 5-mer nucleotides added in complementary sequences at their 5'-ends. To develop the TMC-PCR, the TaqMan probe was modified to an MGB-conjugated hybridization probe and was shorter than that described by Verstrepen et al. [23]. The TMC-PCR used two primers: NCR-cF and NCR-cR that described by Hong J et al. [24]. The TaqMan MGB probe, referred to TMP, was: FAM5'-CCGACTACTTTGGGTGTC-3'MGB-NFQ (positions 541-558 of the CVB2 sequence; GenBank accession number EF174469). It contains the reporter 6-carboxyfluorescein (FAM) and the non-fluorescent quencher (NFQ) dye.

Real-time One Step RT-PCR

TMC-PCR assay was performed using an ABI Prism 7900HT sequence detection system (Applied Biosystems). Viral RNA was amplified in 25 μL reactions using RT-PCR master mix (AgPath-ID one-step RT-PCR Kit; Ambion, CA). Reactions were incubated at 45°C for 15 min, and then at 95°C for 10 min, followed by 45 cycles of 95°C for 15 s and 60°C for 40 s.

Results

Limits of detection and reproducibility

To determine the detection limit of the TMC-PCR and TTN-PCR assays, serial dilutions of five reference strains belonging to distinct serotypes (EV71, CVB2, E30, CVA24, and P1) were tested. The highest positive dilution of CVB2, 0.5 TCID50/mL, corresponded to a Ct of 35.6, whereas the lowest dilution, 1 × 107 TCID50/mL, corresponded to a Ct of 8.38 of P1. Each of the titrated viruses was a member of a major EV group. The minimum detectable amount of enteroviral RNA was equivalent to 0.01 TCID50 in CVA24. The detection limit of the end point was the same in the TMC-PCR and TTN-PCR assays (data not shown).

Reproducibility was tested by analyzing a titrated P1 virus at least three times on different days (Table 1). In the TMC-PCR assay, the Ct mean values of the 10-2 (105 TCID50/mL) and 10-3 (104 TCID50/mL) RNA extracts, performed in duplicate in three different runs, were 21.55 and 25.02, respectively, with intra-assay standard deviations of 0.08 and 0.07, respectively. The Ct mean values of the 10-5 (100 TCID50/mL) and 10-6 (10 TCID50/mL) RNA extracts, performed in duplicate in three different runs, were 30.65 and 33.95, respectively, with inter-assay standard deviations of 0.86 and 0.80, respectively.

Table 1 Intra- and inter-assay reproducibility of the EV MGB/CLP (TMC-PCR) and TAMRA/NonCLP real time RT-PCR (TTN-PCR)

In the TTN-PCR assay, the Ct mean values of the 10-2 (105 TCID50/mL) and 10-3 (104 TCID50/mL) RNA extracts, performed in duplicate in three different runs, were 21.22 and 24.82, respectively, with intra-assay standard deviations of 0.24 and 0.16 respectively. The Ct mean values of the 10-5 (100 TCID50/mL) and 10-6 (10 TCID50/mL) RNA extracts, performed in duplicate in three different runs, were 29.05 and 32.62, respectively, with inter-assay standard deviations of 1.89 and 2.09, respectively.

Analytical sensitivity and specificity

All clinical isolates of 25 serotypes circulating between 1997 and 2007 in Korea of EVs could be detected and there was no inconsistency between the qualitative results generated by the two assays (TMC-PCR and TTN-PCR). This confirms that the modifications to the primers and probes did not affect the detectable reaction to various EV serotypes (in other words, pan-enterovirus).

From the results for the 12 proficiency panels (EV07) from QCMD (Glasgow, Scotland), including various titers of coxsackievirus B3 (CVB3), echovirus 11 (E11), enterovirus 71 (EV71), echovirus 30 (E30), and poliovirus type 3 (P3), could be detected the genomes of EVs, including three serotypes with different detection limits, depending on the assay used (Table 2). In the each case of CVB3 (EV07-07 vs. EV07-12), E11 (EV07-03 vs. EV07-04) and E30 (EV07-08 vs. EV07-09), at a 10-fold lower viral titer than the panel, TMC-PCR assay produced a positive result, whereas TTN-PCR assay produced a negative. Parechovirus type 3 in sample EV07-02 and EV07-11 was not detected by either assay, confirming that both assays were specific to EVs.

Table 2 Result with Enterovirus real time quality control panels

Comparison of TMC-PCR and TTN-PCR assays with clinical specimens

A total of 158 clinical specimens from patients diagnosed with meningitis were evaluated. Sixty were positive in both TMC-PCR and TTN-PCR assays, whereas 40 in total 158 samples were positive only in the TMC-PCR assay. The results were consistent number of negative results in both case 58/98 (Table 3). The clinical specimens consisted of 102 CSF, 48 stool, and 8 throat swab samples. To investigate assay results among different types of specimens, the TMC-PCR and TTN-PCR assays were compared using the 102 CSF samples tested; 62 and 29 samples were positive using the TMC-PCR and TTN-PCR assays, respectively. Thirty-six of the 48 stool samples were positive using the TMC-PCR assay and 29 using the TTN-PCR assay. Of the throat swab samples, 2 of 8 were positive by both assays.

Table 3 Comparison of results of MGB/CLP (TMC-PCR) and TAMRA/NonCLP real time RT-PCR (TTN-PCR) in different types of specimens

In all specimens, the positive and negative agreements of the TTN-PCR assay and the TMC-PCR assay were 100% (60/60) and 59.2% (58/98), respectively. In CSF specimens, the positive and negative agreements of the TTN-PCR assay and the TMC-PCR assay were 100% (29/29) and 54.79% (40/73), respectively. In stool specimens, the positive and negative agreements of the TTN-PCR assay and the TMC-PCR assay were 100% (29/29) and 63.16% (12/19), respectively. In throat swab specimens, the positive and negative agreements of the TTN-PCR assay and the TMC-PCR assay were 100% (2/2) and 100% (6/6), respectively.

Discussion

Several groups have developed and demonstrated molecular diagnostic methods that are highly sensitive, specific, simple, and reproducible. In the field of virology, real-time RT-PCR has become a standard diagnostic method because of its rapid turnaround time, relatively low risk of contamination, and ease of use [19, 29].

This study modified a TTN-PCR published method [23] and compared it to our modified assay, TMC-PCR, for use as a routine diagnostic assay. TMC-PCR assay detected the same region of the 5'-NCR as the traditional Verstrepen amplicon. The primers used in this assay combined previously reported CLP technology [24], a probe modified at the 5'-terminal by MGB, and a 3'-terminal NFQ. These probes provided lower backgrounds and a higher signal-to-noise ratio than other hybridization probes [30]. MGB allows the design and use of shorter probes, while maintaining a high melting temperature, which may be an advantage in designing probes for small conserved regions [25, 26].

The results of the modified primers and probe using the 2007 QCMD panel were compared. TMC-PCR assay detected all EVs in the panels, except parechovirus. However, the TTN-PCR assay could not detect some EVs, such as echovirus 11 and coxsackievirus B3, in the 2007 panel. These false negative results were in low viral titer samples (Table 1). To test TMC-PCR for EVs, EV71, CVB2, E30, CVA24, and P1, as example EV serotypes, were used to determine the dynamic range, to optimize real-time PCR conditions, and to evaluate analytical performance. The detection limits for EV in both assays were similar (data not shown). However, standard deviations (SDs) of the intra-assay and inter-assay reproducibility of the TMC-PCR and TTN-PCR assays were different, particularly at low viral dilutions (Table 2). Thus, TMC-PCR had better assay reproducibility.

To examine the analytical performances of the TMC-PCR and TTN-PCR assays, a prospective study on clinical samples was conducted. In 158 clinical samples from patients with aseptic meningitis, enterovirus was detected in 100 of 158 samples by TMC-PCR and 60 of 158 samples by TTN-PCR. A comparison of the TMC-PCR and TTN-PCR assays revealed 40 discrepant samples (Table 3), 33 (82.5%) of which were CSF specimens. When separately analyzed using CSF specimens, the positive and negative agreements between TMC-PCR and TTN-PCR were 100% and 54.79%, respectively. Then the GeneXpert Enterovirus Assay (GXEA; Cepheid, Sunnyvale, CA) were performed with the 40 discrepant samples to confirm the results as false negatives (data not shown). The results were matched to the results of the TMC-PCR assay. To check the false positive results, we also performed GeneXpert Enterovirus Assay (GXEA) with positive samples from TMC-PCR assay. There were no discord results between TMC-PCR and GXEA (data not shown).

The differences in viral titer were compared to the Ct mean values of CSF-positive results. The Ct mean values from TMC-PCR and TTN-PCR of positive CSF specimens were 34.84 and 32.82, respectively. The TTN-PCR was lower than TMC-PCR in terms of the absolute value of Ct, but the rate of positive TMC-PCR was higher than in the TMC-PCR assay. The results in stool specimens showed the same pattern, despite the absolute mean value of Ct differed between TMC-PCR (30.19) and TTN-PCR (28.91). Although TMC-PCR assay has better sensitivity compared with TTN-PCR assay, the Ct vale of TMC-PCR assay in certain sample is higher than that of TNC-PCR assay. The Ct value is not only factor to determine the sensitivity. In our results, TMC-PCR assay has slightly higher Ct than TTN-PCR assay but very lower standard deviation (SD) compared with TTN-PCR assay in low viral titers at inter assay (Table 1). It means that TMC-PCR assay can produce stable diagnostic result in the specimen with low viral titers such as CSF. We can confirm the high sensitivity in the low titer proficiency sample in the Table 2 based on stability induced by CLP and MGB technologies. For example in the Table 2, TMC-PCR assay produced a positive result, whereas TTN-PCR assay produced a negative result at a 10-fold lower viral titer than the panel: CVB3 (EV07-07 vs. EV07-12), E11 (EV07-03 vs. EV07-04) and E30 (EV07-08 vs. EV07-09). We also evaluate the sensitivity of assay developed in this study using clinical samples and confirm the high sensitivity of TMC-PCR assay (62/73) in the CSF sample than TTN-PCR assay (29/73). Conclusively TMC-PCR assay has slightly higher Ct value in the certain sample with high purity (culture supernatant or pure gene serial diluents), but can produce stable diagnostic results at the clinical sample with low viral titers.

The results reported that the TMC-PCR assay is a better routine diagnostic method for EV meningitis, because of the low quantities of viral RNA in CSF specimens, not only rapid and accurate results, and may be a useful molecular screening method.

Conclusion

This study revealed that TMC-PCR assay could give reliable results and showed better reproducibility for routine laboratory diagnosis. Also technical modification of primer provided to increase sensitivity. Our data suggest that TMC-PCR assay may be more suitable to routine diagnostic assay for EV detection from clinical specimens even though TMC-PCR assay was not the novel real-time RT-PCR assay.