Abstract
Background
The all-inside cruciate ligament graft preparation technique has become popular due to its utility in sparing a growing physis, preserving a tendon in ACL surgery, and/or reduction of pain. However, few studies have compared graft preparation techniques to determine the ideal construct for cruciate ligament reconstruction. We sought to compare biomechanical properties of two quadrupled all-inside cruciate ligament graft preparation techniques and three alternative all-inside graft preparation techniques that may be used when the available tendon is too short to be quadrupled.
Methods
Fifty porcine extensor tendons were evenly divided into five groups (n = 10) representing all-inside graft preparation techniques, including two quadrupled (Quad-A, Quad-B) and three alternative methods (Tripled, Folded, Two-Doubled). Each graft construct underwent preconditioning (10 loading cycles from 20 to 50 N at 0.1 Hz), cyclic loading (500 loading cycles from 50 to 250 N at 1.0 Hz) and load-to-failure (tension applied at 20 mm/min).
Results
Quad-A and Quad-B demonstrated no significant differences in cyclic displacement (10.5 ± 0.3 vs 11.7 ± 0.4 mm; p = 0.915), cyclic stiffness (1086.2 ± 487.3 vs 460.4 ± 71.4 N/mm; p = 0.290), pullout stiffness (15.9 ± 4.3 vs 7.4 ± 4.4 N/mm; p = 0.443), ultimate failure load (641.2 ± 84.7 vs 405.9 ± 237.4 N; p = 0.672), or ultimate failure displacement (47.3 ± 6.7 vs 55.5 ± 0.7 mm; p = 0.778). The mean cyclic displacement of the Two-Doubled group was significantly greater than the Quad-A (29.7 ± 2.2 vs 10.5 ± 0.3 mm; p < 0.001), Quad-B (29.7 ± 2.2 vs 11.7 ± 0.4 mm; p < 0.001), Tripled (29.7 ± 2.2 vs 11.3 ± 0.2 mm; p < 0.001), and Folded group (29.7 ± 2.2 vs 13.3 ± 0.2 mm; p < 0.001). There were no other statistically significant differences between the three alternative all-inside graft preparation techniques.
Conclusion
The current study demonstrates the biomechanical properties of two quadrupled all-inside graft constructs, Quad-A and Quad-B, are not significantly different. When the available tendon is of insufficient length, the Two-Doubled group demonstrated more than twice the cyclic displacement of all other graft preparation techniques, and is therefore not recommended for use in all-inside cruciate ligament reconstruction.
Similar content being viewed by others
Background
The all-inside technique is a relatively new approach to anterior cruciate ligament (ACL) and posterior cruciate ligament (PCL) reconstruction that has gained popularity in recent years, in part because of its minimally invasive approach, preservation of cortical bone, and potential sparing of tendon tissue (autograft) or cost (allograft) compared to traditional techniques (Benea et al., 2014; Smith et al., 2008; Connaughton et al., 2017; Jones and Schuett, 2018). This technique can also be employed to preserve the integrity of the tibial and femoral physes by use of sockets that are restricted to the epiphysis, a feature that is particularly important for ACL reconstruction in skeletally immature patients (Stadelmaier et al., 1995; Kocher et al., 2002; Kercher et al., 2009; Frosch et al., 2010; Cordasco et al., 2016). Fashioned with either an autograft or allograft tendon, which is prepared and secured within femoral and tibial sockets using cortical suspension devices, the all-inside ACL and PCL reconstruction technique incorporates a construct with multiple components susceptible to failure. Accordingly, careful research on the relative efficacy and biomechanical properties of these techniques is needed.
While various graft fixation constructs have been studied in depth (Fabbri et al., 2016; Lubowitz, 2012; Fritsch et al., 2017; Sasho et al., 2018; Tiefenboeck et al., 2018), current biomechanical literature falls short in offering relative comparisons between graft preparation techniques. Although the quadrupled graft preparation technique is the preferred method of graft preparation, multiple techniques have been described (Lubowitz, 2012; Fritsch et al., 2017; Sasho et al., 2018; Tiefenboeck et al., 2018; McCarthy et al., 2012). Additionally, insufficient tendon length can become a problem for ACL reconstruction in petite patients or in the case of inadvertent amputation of a tendon during harvest, such as in the setting of previous surgery and resultant scar. In the setting of PCL reconstruction, tendon length is a relatively consistent concern because a longer graft is required to approximate the deficient ligamentous structure. In these cases, the harvested tendon or allograft tendon may be of insufficient length to create a quadrupled graft of appropriate length, and one of several alternative methods for graft preparation may be used. In one commonly used alternative, the graft is tripled, and a whipstitch placed at each end as a result. Another involves selection of two shorter grafts, and doubling them. Finally, a different tendon graft might be used and folded back onto itself (such as Achilles tendon).
The purpose of the current study was to examine the biomechanical properties of five different methods used to prepare grafts for all-inside ACL and PCL reconstruction: two types of quadrupled constructs (Quad-A and Quad-B), a tripled construct (Tripled), a folded construct (Folded), and two grafts in doubled form (Two-Doubled) which are illustrated in Fig. 1. The two quadrupled constructs represent the preferred techniques used for graft preparation in all-inside cruciate ligament reconstruction. The remaining three constructs are potential alternative methods used for construction of grafts when the available tendon is not long enough to produce a quadrupled graft of adequate length. We hypothesized that, while the two quadrupled techniques would demonstrate similar biomechanical properties, the three alternative techniques would demonstrate significant differences in their biomechanical characteristics.
Methods
Porcine extensor tendons
Fifty fresh frozen (−23o C) porcine extensor tendons, acquired from J & J Packing Co. Inc. (Brookshire, TX) were randomly divided into five groups. Specimens were thawed in a refrigerator within 48 h of use and underwent only one freeze-thaw cycle. Once thawed, the bifurcate tendon was divided, all adherent muscle tissue was removed, and grafts were trimmed to achieve average graft construct lengths of 50–70 mm for all groups (Table 1). In an effort to maintain structural integrity of the extensor tendons, individual tendons were not trimmed longitudinally to create smaller tendon widths. Specimens were kept hydrated by being wrapped in a moist paper towel during preparation and were sprayed with water approximately every two minutes during testing.
Surgical techniques
The tendons were prepared using one of the five techniques illustrated in Fig. 1. All graft constructs were prepared either by, or under the direct supervision of, the same faculty orthopaedic surgeon. One of the coauthors served as an assistant during graft preparation, to ensure consistency of technique. The graft constructs were suspended with double-looped No. 5 Fiberwire suture on both the femoral and tibial ends. The final dimensions of the graft constructs in each of the groups are summarized in Table 1.
Quadrupled techniques
The first quadrupled technique, Quad-A, was prepared by wrapping the tendon around the No. 5 suture in the same fashion that has been demonstrated by McCarthy and colleagues (McCarthy et al., 2012). The tendon was folded in half around one No. 5 Fiberwire suture loop at the tibial end, the free ends were then whip-stitched together using a No.2 Fiberwire suture and passed together in the same direction around the second No. 5 suture loop at the femoral end. The free ends were tucked into the fold on the tibial end, with the whip stitch suture wrapped and tied around the tibial suspensory suture. The graft was secured with four cerclage stitches, two on the femoral end and two on the tibial end, passing through each of the four tendon limbs and secured with buried knots.
The second quadrupled technique, Quad-B, was prepared by wrapping the tendon around the No. 5 suture in the same fashion that has been demonstrated by Lubowitz (Lubowitz, 2012). The tendon was folded in half around a No. 5 suture loop at the tibial end and the free ends were passed in opposite directions around the No. 5 suture loop on the femoral end. The free tendon ends were then whip-stitched together with a No. 2 suture and tucked into the folded tendon at the tibial end, with the whip stitch suture wrapped and tied around the tibial suspensory suture. The graft was secured with four cerclage stitches, two on the femoral end and two on the tibial end, passing through each of the four tendon limbs and secured with buried knots.
Tripled technique
The Tripled graft construct was created by whip-stitching each of the free ends with a No. 2 suture, passing each free end around a No. 5 suture and then tucking the ends into the opposite femoral and tibial folds. The No. 2 suture from the whip-stitched ends was then wrapped and knotted around the suspensory suture at the tibial and femoral ends, respectively. The graft was secured with four cerclage stitches, two on the femoral end and two on the tibial end, passing through each of the three tendon limbs, and secured with buried knots.
Folded technique
The Folded graft construct was created by passing one of the free ends of the tendon around the femoral or tibial-sided No. 5 suture, and wrapping around so the free ends met in the center of the graft. Each free end was then whip-stitched to the tendon on the opposing long side of the graft using a No. 2 suture. The graft was secured with four cerclage stitches, two on the femoral end and two on the tibial end, passing through both of the tendon limbs and secured with buried knots.
Two-doubled technique
The Two-Doubled technique was prepared using two tendons, each folded over a No. 5 looped-suture to produce a total of four tendon limbs. The free-ends were whip-stitched together and the suture connecting the two tendons was passed through a second No. 5 looped-suture on the tibial end. The graft was secured with four cerclage stitches, two on the femoral end and two on the tibial end, passing through each of the four tendon limbs, and secured with buried knots.
Biomechanical testing
Graft constructs were mounted onto an 858 Mini Bionix Material Testing System (MTS Systems Corp., Eden Prairie, MN) by passing the doubled No. 5 looped-suture in a figure-of-eight pattern around 4 screws which were used to lightly compress aluminum clamps on the suture on both the femoral and tibial ends (Fig. 2).
Load and displacement data were collected at a frequency of 20 Hz during all phases of testing and the testing protocol was executed as previously described by Walsh and colleagues (Walsh et al., 2008). Displacement was measured by the elongation of the distance between clamps during testing. Prior to testing each specimen, the load cell was zeroed without tension. Preloading occurred as tension was applied to each specimen at a rate of 1 N/sec up to 20 N and displacement was zeroed while the load was held at 20 N for 10–20 s. Preconditioning was performed with 10 loading cycles from 20 to 50 N at 0.1 Hz and subsequent cyclic loading occurred for 500 loading cycles from 50 to 250 N at 1.0 Hz. Cyclic displacement was defined as the displacement between clamps occurring during cyclic testing. Cyclic stiffness was defined as the slope of the load versus displacement curve in the elastic region during the cyclic testing.
To test load-to-failure, tension was applied to each specimen at a rate of 20 mm/min until the tendon or the suture failed, producing a rapid decrease in load. Pullout stiffness was defined as the slope of the load versus displacement curve in the elastic region during the load-to-failure testing. Ultimate failure load was defined as the load at the onset of plastic deformation within the system, based on the change in slope of the load versus displacement curve, regardless of the source of that failure. Ultimate failure displacement was defined as the amount of displacement within the system at the point of ultimate failure load.
Statistical analysis
All outcomes were assessed using generalized estimating equation (GEE) models to compare the differences of the graft types while controlling for the clustered nature of the data. GEE parameter estimates were generated using quasi-likelihood estimation. Post hoc tests were conducted to determine differences between graft types. Reported p-values were Bonferroni-adjusted to account for the type I error from multiple comparisons. Statistical significance was defined as a p-value < 0.05. All statistical analyses were performed using SAS version 9.2 (SAS Institute; Cary, NC).
Results
Tendon and graft construct dimensions
The dimensions of the tendons and graft constructs for each of the preparation techniques, prior to testing, are reported in Table 1 with the associated post hoc pairwise comparisons reported in Table 2. No significant differences were found in the dimensions of the Quad-A vs Quad-B groups in terms of tendon length (p > 0.999), tendon width (p = 0.373), graft construct length (p = 0.439), or graft construct width (p > 0.999).
After being trimmed to achieve final graft construct lengths of 50–70 mm, the tendon lengths of the Tripled (196.3 mm), Folded (148.5 mm) and Two-Doubled group (115.5 mm) demonstrated significant differences when compared to one another (p < 0.001). There were no statistically significant differences (p > 0.05) between the individual tendon widths of the Tripled, Folded and Two-Doubled groups. The graft construct length of the Tripled vs Folded groups was not significantly different (p > 0.999), nor was the difference between the Tripled vs Two-Doubled groups (p = 0.132). There was a statistically significant difference between the Folded vs Two-Doubled group (68.9 ± 1.0 vs 60.3 ± 1.0 mm; p = 0.003), however. The final graft construct widths, prior to testing, demonstrated statistically significant differences in the Tripled vs Folded (1p = 0.004), Tripled vs Two-Doubled (1p = 0.010) and Folded vs Two-Doubled (p < 0.001) group comparisons.
Biomechanical testing
The biomechanical parameter estimates of each of the graft preparation techniques are reported in Table 3. Quad-A vs Quad-B comparisons demonstrated no significant differences in cyclic displacement (p = 0.915), cyclic stiffness (p = 0.290), pullout stiffness (p = 0.443), ultimate failure load (p = 0.672), or ultimate failure displacement (p = 0.778). When compared to the Tripled group, the Quad-A group demonstrated significantly greater ultimate failure load (p = 0.046) and less ultimate failure displacement (p = 0.010). The post hoc pairwise comparisons for each of the biomechanical parameters are organized in Table 4.
The mean cyclic displacement of the Two-Doubled group was the largest amongst all groups, and was significantly greater than the Quad-A group (p < 0.001), Quad-B group (p < 0.001), Tripled group (p < 0.001), and Folded group (p < 0.001). No other statistically significant differences were observed between the groups. However, the pullout stiffness of the Quad-A group was nearly statistically significantly greater than the Tripled group (p = 0.059).The mechanisms of failure of the various graft types were also observed. No slipping of the suspensory suture at the clamp-suture junction was observed, nor was stretching of the Fiberwire suture. Quad-A and Quad-B grafts failed when the whipstitched free end of the graft pulled from its tibial end and through the nearest cerclage suture, elongating the graft significantly. The Tripled group showed a similar mechanism of failure as the free, whipstitched ends at the femoral and tibial sides of the graft pulled through the cerclage sutures.
The mechanism of failure in the Folded group was more variable: four grafts failed when the whipstitch suture itself broke, three grafts failed when the whipstitched end pulled through the cerclage suture, and three grafts failed when the cerclage sutures broke at the tibial end of the graft.
The mechanism of failure of the Two-Doubled group was consistently located at the tibial side of the graft where no tendon tissue was wrapped around suspensory suture, instead, the whip stitch was looped around the suspensory suture. In seven cases, the whipstitch broke at the tibial end of the graft. Two grafts failed when the whipstitch slid through the tendon, and one graft failed when the whipstitch broke at its interface with the suspensory suture.
Discussion
We determined that the biomechanical properties of two quadrupled cruciate ligament graft constructs, Quad-A and Quad-B, are not significantly different. To our knowledge, this is the first biomechanical comparison of these two commonly used techniques. We also demonstrated that these two quadrupled techniques achieve ultimate failure loads that are sufficient to withstand the forces (150–590 N) encountered during early rehabilitation protocols following ACL and PCL reconstruction (Shelburne et al., 2004). Thus, our results support either of the quadrupled techniques being the preferred method of graft preparation for all-inside ACL and PCL reconstruction.
Importantly, our results indicate that the least robust portion of an all-inside graft construct is the suture used to secure it and/or suspend it. This is in contrast to other graft preparation techniques, in which the graft itself is the first to fail when tested in a biomechanical environment. The all-inside graft preparation technique seems to transfer the loads from the fixation itself to the whipstitch.
At present, the literature remains unclear on which graft preparation technique should be used if the tendon is of insufficient length to produce a quadrupled graft. In the current study, we compared three alternative methods of graft preparation to the quadrupled techniques to evaluate their relative biomechanical strength. During cyclic loading, cyclic displacement of the Two-Doubled group was greater than twice that of every other group, which makes this technique the least appealing option for all-inside ACL and PCL reconstruction. Although not statistically significant, due likely to the high variability of the Qual-A and Quad-B groups relative to the other groups, the Tripled technique demonstrated the least amount of cyclic displacement of the three alternative methods. Thus the Tripled group may be the preferred technique to employ if the tendon is of insufficient length to be quadrupled. However, further research is needed to clarify if this is truly the case. The Tripled group failed at significantly lower ultimate failure loads when compared to the Quad-A group, but not the Quad-B group; a significance that was lost in the Folded and Two-Doubled groups as a result of higher variability. Ultimate failure displacement was also significantly greater in the Tripled group compared to the Quad-A group; however, this parameter has less clinical significance due to the anatomical limitations of the knee joint itself.
While the ultimate failure displacement in the Two-Doubled group did not differ significantly, the variability of this group was relatively high when compared to the Tripled and Folded groups. This appeared to be due to failure of the suture material at the tibial end of the graft where the free ends were whipstitched together, and this whipstitch was utilized to incorporate the suspensory suture. This may indicate that wrapping the tendon itself, rather than suture, around the suspensory device lends inherent strength to the graft construct. Therefore, we question the integrity of graft preparation in single tendon, non-folded constructs such as quadriceps tendon mounted on a suspensory device via suture alone on the both the femoral and tibial ends (Slone et al., 2016).
Ensuring that various graft fixation methods are sufficient to withstand the forces experienced during early post-operative rehabilitation has been an active area of research in recent years. Fortunately, biomechanical studies have demonstrated that many suspensory devices have sufficient strength to withstand forces that exceed even the highest loads experienced during early exercise protocols (Bartz et al., 2007; Johnson et al., 2015; Rylander et al., 2014). Additionally, there are several different approaches used for graft fixation in all-inside reconstruction, including interference screws, suture buttons and a combination of interference screws and suture buttons together (Frosch et al., 2010). As cruciate ligament reconstruction trends towards less invasive techniques such as all-inside ACL and PCL replacement, it becomes increasingly important to evaluate the biomechanical properties of the graft constructs as well as the fixation devices.
Johnson et al. demonstrated that modern cortical suspension devices are capable of withstanding ultimate forces of 784–2231 N (Johnson et al., 2015). These forces are much greater than those tolerated by the tissue grafts in our study, which suggests that the suture by which the graft is secured, or to the fixation device is the weakest component of the graft-link construct. This extrapolation is complicated by a variety of factors including dynamic in vivo force vectors and loads as well as by supplemental fixation techniques such as cortical suspension devices used in combination with an interference screw. Nevertheless, it is essential to identify and minimize those components which are susceptible to rupture or failure.
Fabbri et al. used a similar model to evaluate whether graft length is critical for success in all-inside reconstruction, including a quadrupled model similar to Quad-A in our study, as well as a tripled model similar to that in our study, and a half-quadrupled model that was unique to their study (Fabbri et al., 2016). The quadrupled method achieved the highest ultimate failure load, followed by the tripled method The mechanism of failure seen in their study was consistent with that seen in our study, with failure occurring by slippage of suture within the graft. This study, along with ours, provides more comprehensive investigation of quadrupled graft preparation techniques and of alternative techniques.
Mayr et al. compared several all-inside graft preparation techniques with varying degrees of suture fixation (Mayr et al., 2016). This included one model with four buried cerclage sutures (two sutures at both ends of the graft as described originally by (Lubowitz, 2012)), one with two buried cerclage sutures at the tibial end, and the final with two cerclage sutures at the tibial end with additional suspension on the tibial cortical button. The four-suture and hybrid two-suture with additional suspension model had comparable graft elongation and ultimate failure loads, thus proposing a faster graft preparation technique with comparable biomechanical strength to the current standard. Similarly, we used a hybrid suspensory device at the tibial end of the grafts by wrapping and tying the whipstitched suture around the suspensory suture. However, we chose not to focus on the level of suture burden of the various techniques, but instead the different methods of folding the grafts into various constructs.
Barbosa et al. described an alternative technique which minimizes suture in the bony socket, proposing that suture wrapping the tendon in the bony socket reduces the bone-tendon contact area crucial for biologic integration of the graft (Barbosa et al., 2017). Their technique eliminates the cerclage suture at both ends of the graft in favor of directly suturing the whip-stitched free ends to the Fiber Loop. Unfortunately, a biomechanical evaluation of the technique was not conducted, which would be an important consideration in the clinical use of this model.
Our study is not without its limitations. First, a study of porcine tendons cannot be directly compared with studies using human tissue, although our graft preparation technique was equivalent to the clinical setting. Porcine tendon specimens have been reported to have similar properties to human tendons in previous biomechanical studies (Yamanaka et al., 1999; Dargel et al., 2009) and the porcine knee is widely used as a model to investigate ACL reconstruction (Dargel et al., 2009; Debandi et al., 2011; Zhu et al., 2018). Second, the length of the prepared grafts was not uniform. While the displacement of the constructs could theoretically be affected by significant differences in width and/or length, we observed that the majority of displacement and subsequent failure occurred at the tendon/suture interface in each of the five groups. This is especially true for the Two-Doubled group, where the constructs elongated as a consequence of the whipstitch being pulled away from the free ends. Thus, it stands to reason that differences in displacement, stiffness, and load to failure are more the result of differences in the interaction between the tendon and suture rather than the material properties themselves.
Another limitation is that this study was designed to test the relative biomechanical properties of different graft preparation techniques in isolation, which does not provide information about the graft and fixation device interaction or complex as an entire unit. Additionally, the testing protocol provides information about the relative displacement, stiffness and load tolerance in a linear fashion, which limits us from making inferences based upon the external forces and dynamic vectors that are placed upon the ACL or PCL graft as it occurs in vivo. Further research is necessary to investigate the biomechanical properties of the entire graft constructs with various fixation devices in vitro in addition to appropriate prospective studies designed to identify clinically relevant outcomes in the long term.
Conclusion
The two quadrupled techniques for all-inside cruciate ligament reconstruction demonstrated no significant difference in any of the primary endpoints measured, suggesting that either technique could be the preferred method for all-inside ACL and PCL reconstruction. The Tripled technique is preferred if a graft is of insufficient length for quadrupled preparation. The Two-Doubled technique was observed to have greater than twice the cyclic displacement of every other preparation technique tested, and is therefore not recommended for use in all-inside cruciate ligament reconstruction.
References
Barbosa NC, Guerra-Pinto F, Cabeleira C, Beja da Costa P (2017) Hamstring graft biological preparation for anterior cruciate ligament reconstruction. Arthrosc Tech 6:e245–e248. https://doi.org/10.1016/j.eats.2016.09.031
Bartz RL, Mossoni K, Tyber J, Tokish J, Gall K, Siparsky PN (2007) A biomechanical comparison of initial fixation strength of 3 different methods of anterior cruciate ligament soft tissue graft tibial fixation: resistance to monotonic and cyclic loading. Am J Sports Med 35:949–954
Benea H, d’Astorg H, Klouche S, Bauer T, Tomoaia G, Hardy P (2014) Pain evaluation after all-inside anterior cruciate ligament reconstruction and short term functional results of a prospective randomized study. Knee 21:102–106. https://doi.org/10.1016/j.arthro.2008.06.004
Connaughton AJ, Geeslin AG, Uggen CW (2017) All-inside ACL reconstruction: how does it compare to standard ACL reconstruction techniques? J Orthop 14:241–246. https://doi.org/10.1016/j.jor.2017.03.002
Cordasco FA, Mayer SW, Green DW (2016) All-inside, all-epiphyseal anterior cruciate ligament reconstruction in skeletally immature athletes: return to sport, incidence of second surgery, and 2-year clinical outcomes. Am J Sports Med 45:856–863. https://doi.org/10.1177/0363546516677723
Dargel J, Koebke J, Brüggemann G-P, Pennig D, Schmidt-Wiethoff R (2009) Tension degradation of anterior cruciate ligament grafts with dynamic flexion-extension loading: a biomechanical model in porcine knees. J Arthros & Rel Surg 25:1115–1125
Debandi A, Maeyama A, Lu S, Hume C, Asai S, Goto B et al (2011) Biomechanical comparison of three anatomic ACL reconstructions in a porcine model. Knee Surg Sports Traumatol Arthrosc 19:728–735. https://doi.org/10.1007/s00167-010-1338-3
Fabbri M, Monaco E, Lanzetti RM, Perugia D, Guzzini M, Labianca L et al (2016) Single harvesting in the all-inside graft-link technique: is the graft length crucial for success? A biomechanical study. J Orthop Traumatol 18:17–22. https://doi.org/10.1007/s10195-016-0420-0
Fritsch B, Figueroa F, Semay B (2017) Graft preparation technique to optimize hamstring graft diameter for anterior cruciate ligament reconstruction. Arthrosc Tech 6:e2169–e2175. https://doi.org/10.1016/j.eats.2017.08.011
Frosch K-H, Stengel D, Brodhun T, Stietencron I, Holsten D, Jung C et al (2010) Outcomes and risks of operative treatment of rupture of the anterior cruciate ligament in children and adolescents. Arthrosc J Arthrosc Relat Surg 26:1539–1550. https://doi.org/10.1016/j.arthro.2010.04.077
Johnson JS, Smith SD, LaPrade CM, Turnbull TL, LaPrade RF, Wijdicks CA (2015) A biomechanical comparison of femoral cortical suspension devices for soft tissue anterior cruciate ligament r0econstruction under high loads. Am J Sports Med 43:154–160. https://doi.org/10.1177/0363546514553779
Jones PE, Schuett DJ (2018) All-inside anterior cruciate ligament reconstruction as a salvage for small or attenuated hamstring grafts. Arthrosc Tech 7:e453–e457. https://doi.org/10.1016/j.eats.2017.11.007
Kercher J, Xerogeanes J, Tannenbaum A, Al-Hakim R, Black JC, Zhao J (2009) Anterior cruciate ligament reconstruction in the skeletally immature: an anatomical study utilizing 3-dimensional magnetic resonance imaging reconstructions. J Pediatr Orthop 29:124–129. https://doi.org/10.1097/BPO.0b013e3181982228
Kocher MS, Saxon HS, Hovis WD, Hawkins RJ (2002) Management and complications of anterior cruciate ligament injuries in skeletally immature patients: survey of the Herodicus society and the ACL study group. J Pediatr Orthop 22:452–457
Lubowitz JH (2012) All-inside anterior cruciate ligament graft link: graft preparation technique. Arthrosc Tech 1:e165–e168. https://doi.org/10.1016/j.eats.2012.06.002
Mayr R, Heinrichs CH, Eichinger M, Smekal V, Schmoelz W, Attal R (2016) Preparation techniques for all-inside ACL cortical button grafts: a biomechanical study. Knee Surg Sports Traumatol Arthrosc 24:2983–2989. https://doi.org/10.1007/s00167-015-3605-9
McCarthy MM, Graziano J, Green DW, Cordasco FA (2012) All-epiphyseal, all-inside anterior cruciate ligament reconstruction technique for skeletally immature patients. Arthrosc Tech 1:e231–e239. https://doi.org/10.1016/j.eats.2012.08.005
Rylander L, Brunelli J, Taylor M, Baldini T, Ellis B, Hawkins M et al (2014) A biomechanical comparison of anterior cruciate ligament suspensory fixation devices in a porcine cadaver model. Clin Biomech 29:230–234. https://doi.org/10.1016/j.clinbiomech.2013.11.001
Sasho T, Sasaki T, Hoshi H, Akagi R, Enomoto T, Sato Y, et al. (2018) Evaluating different closed loop graft preparation technique for tibial suspensory fixation in ACL reconstruction using TightRopeTM. Asia-Pacific J Sport Med Arthrosc Rehabil Technol 12:5–11. doi:https://doi.org/10.1016/j.asmart.2018.01.002
Shelburne K, Pandy M, Anderson F, Torry M (2004) Pattern of anterior cruciate ligament force in normal walking. J Biomech 37:797–805
Slone HS, Ashford WB, Xerogeanes JW (2016) Minimally invasive quadriceps tendon harvest and graft preparation for all-inside anterior cruciate ligament reconstruction. Arthrosc Tech 5:e1049–e1056
Smith PA, Schwartzberg RS, Lubowitz JH (2008) No tunnel 2-socket technique: all-inside anterior cruciate ligament double-bundle retroconstruction. Arthrosc J Arthrosc Relat Surg 24:1184–1189. https://doi.org/10.1016/j.arthro.2008.06.004
Stadelmaier DM, Arnoczky SP, Dodds J, Ross H (1995) The effect of drilling and soft tissue grafting across open growth plates. A histologic study. Am J Sports Med 23:431–435
Tiefenboeck TM, Hirtler L, Winnisch M, Joestl J, Koch T, Komjati M et al (2018) A bigger suture diameter for anterior cruciate ligament all-inside graft link preparation leads to better graft stability: an anatomical specimen study. Knee 25:427–433. https://doi.org/10.1016/j.knee.2018.03.010.7
Walsh MP, Wijdicks CA, Parker JB, Hapa O, LaPrade RF (2008) A comparison between a retrograde interference screw, suture button, and combined fixation on the tibial side in an all-inside anterior cruciate ligament reconstruction: a biomechanical study in a porcine model. Am J Sports Med 37:160–167
Yamanaka M, Yasuda K, Yohyama H, Nakano H, Wada T (1999) The effect of cyclic displacement on the biomechanical characteristics of anterior cruciate ligament reconstructions. Am J Sports Med 27:772–777
Zhu J, Dong J, Marshall B, Linde MA, Smolinski P, Fu FH (2018) Medial collateral ligament reconstruction is necessary to restore anterior stability with anterior cruciate and medial collateral ligament injury. Knee Surg Sports Traumatol Arthrosc 26:550–557. https://doi.org/10.1007/s00167-017-4575-x
Author information
Authors and Affiliations
Contributions
CRW, KCY, JOS, AKB, and JMB performed the biomechanical testing of the ACL graft constructs and co-wrote the manuscript. JTN provided statistical analysis of the data that was gathered. AH provided expertise regarding the idea's inception, and contributed to the writing of the manuscript. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Competing interests
The authors declare that they have no competing interests
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
About this article
Cite this article
Wichern, C.R., Skoglund, K.C., O’Sullivan, J.G. et al. A biomechanical comparison of all-inside cruciate ligament graft preparation techniques. J EXP ORTOP 5, 42 (2018). https://doi.org/10.1186/s40634-018-0158-0
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s40634-018-0158-0