Abstract
Purpose
The coronavirus disease 2019 (COVID-19) pandemic has promoted the use of helmet continuous positive airway pressure (CPAP) for noninvasive respiratory support in hypoxic respiratory failure patients, despite the lack of tidal volume monitoring. We evaluated a novel technique designed to measure tidal volume during noninvasive continuous-flow helmet CPAP.
Methods
A bench model of spontaneously breathing patients undergoing helmet CPAP therapy (three positive end-expiratory pressure [PEEP] levels) at different levels of respiratory distress was used to compare measured and reference tidal volumes. Tidal volume measurement by the novel technique was based on helmet outflow-trace analysis. Helmet inflow was increased from 60 to 75 and 90 L/min to match the patient’s peak inspiratory flow; an additional subset of tests was conducted under the condition of purposely insufficient inflow (i.e., high respiratory distress and 60 L/min inflow).
Results
The tidal volumes examined herein ranged from 250 to 910 mL. The Bland‒Altman analysis showed a bias of -3.2 ± 29.3 mL for measured tidal volumes compared to the reference, corresponding to an average relative error of -1 ± 4.4%. Tidal volume underestimation correlated with respiratory rate (rho = .411, p = .004) but not with peak inspiratory flow, distress, or PEEP. When the helmet inflow was maintained purposely low, tidal volume underestimation occurred (bias − 93.3 ± 83.9 mL), corresponding to an error of -14.8 ± 6.3%.
Conclusion
Tidal volume measurement is feasible and accurate during bench continuous-flow helmet CPAP therapy by the analysis of the outflow signal, provided that helmet inflow is adequate to match the patient’s inspiratory efforts. Insufficient inflow resulted in tidal volume underestimation. In vivo data are needed to confirm these findings.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
1 Introduction
Noninvasive respiratory support is indicated for respiratory failure treatment when patients do not improve by standard oxygen supplementation (i.e., nasal prongs or oxygen mask) but tracheal intubation is not needed or contraindicated. Noninvasive respiratory support can be delivered by several devices, such as large-bore nasal cannulas, a tight face mask, or a helmet; while it is indicated to treat patients affected by cardiogenic pulmonary edema or chronic obstructive pulmonary disease exacerbation, it should be used with caution during acute hypoxemic respiratory failure [1].
Noninvasive respiratory support by head helmet is particularly useful for acute hypoxemic respiratory failure since it is associated with lower mortality and failure rates than high flow oxygen or face masks [2,3,4,5]. Helmet, as a face mask, can be used either to provide continuous positive airway pressure (CPAP) if it is connected to a continuous-flow generator or to deliver noninvasive positive pressure ventilation when it is connected to a ventilator [6]. During the recent coronavirus disease 2019 (COVID-19) pandemic, several advantages of continuous-flow helmet CPAP were described: high oxygen delivery (up to 100%), relatively high positive end-expiratory pressure (PEEP) levels (up to 10–15 cmH2O), device tolerability for prolonged continuous therapy (several consecutive days), possibility of awake prone position, simple equipment with no need for a ventilator, and use outside the intensive care unit [7,8,9,10]. However, helmet CPAP presents some limitations compared to ventilator-driven mask CPAP, such as the impossibility of use in claustrophobic patients, the need for a high amount of gas (i.e., usually 60–90 L/min) to maintain an adequate helmet flow to match the patient’s peak inspiratory flow and the lack of tidal volume monitoring.
Preserving spontaneous breathing by noninvasive support presents several advantages as well as some risks: for example, high tidal volumes are associated with worse outcomes, especially during controlled mechanical ventilation [11, 12]. Excessive lung stretch associated with high tidal volumes and elevated transpulmonary pressures could lead to patient self-inflicted lung injury, further exacerbating lung disease and leading to noninvasive treatment failure [13].
We reasoned that continuous monitoring of tidal volume would be a relevant improvement during helmet CPAP therapy, allowing clinicians to prevent lung injury associated with excessive tidal volumes, such as when respiratory support is provided by a face mask coupled with a ventilator. Ideally, the monitoring device should be embedded in the continuous-flow generator. First, we claim that gas outflow from the helmet is influenced by the patient’s respiratory activity, being decreased during inspiration and increased during expiration, despite a constant flow being blown into the helmet (inflow). Hence, we reasoned that at end-expiration (i.e., no patient’s effort) inflow and outflow are equal; when inspiration begins, a fraction of the inflow is diverted into the airways, resulting in a reduction of the outflow corresponding to the inspiratory tidal volume. Conversely, the outflow increases during expiration as the sum of the patient’s expiratory flow and the constant inflow. We hypothesized that analyses of the inflow and outflow waveforms would allow a reliable measurement of the tidal volume generated by a patient simulator. Second, we reasoned that an inspiratory flow higher than the inflow would temporarily interrupt the outflow due to diversion of the entire inflow into the airways. Thus, we hypothesized that tidal volumes would be underestimated if the patient’s peak inspiratory flow was higher than the helmet inflow.
2 Methods
The bench model designed to measure tidal volumes was based on a standard continuous-flow helmet CPAP system: a compressed gas flow regulator (Flowmeter, Levate, Italy) generating the inflow was connected to a helmet (DimAir, Dimar, Mirandola, Italy) equipped with three different mechanical nonadjustable PEEP valves (5, 10, 15 cm H2O) from the same manufacturer (Respironics Novametrix, Wallingford, CT, USA). The helmet was worn onto a head model and secured over a board; the head model was connected to a patient simulator (ASL 5000, Ingmar Medical, Pittsburg, PA, USA). A pneumotachograph (A. Fleisch, Switzerland) was positioned at the outlet port before the PEEP valve: the flow waveforms of the gas passing through the PEEP valve were recorded by an analog acquisition system (PowerLab, ADInstruments) and analyzed with dedicated software (Labchart, ADInstruments) to measure tidal volumes (Fig. 1). The possible occurrence of a pneumotachograph drift error was prevented by zero flow calibration every 5 min, if deemed necessary. Zeroing was performed with the appropriate LabChart command after temporary disconnection of the pneumotachograph from the circuit.
Pressure levels within the helmet were also continuously recorded to verify the actual peep levels by a standard invasive pressure transducer (TruWave, Edwards, Germany) connected to the helmet through a plastic tube. The patient simulator was set as follows: 60 ml/cmH2O respiratory system compliance; 3 cmH2O/L/s resistance; a heat and moisture exchanger filter connected in series to the respiratory circuit to protect the simulator; 10% inspiratory rise time, 5% inspiratory hold time, 10% inspiratory release time; and 10% pause time with passive expiration. The tidal volume generated by the simulator was considered the standard reference and recorded for comparisons; the simulated patient’s peak inspiratory flow was also recorded. Simulated respiratory efforts were managed by variations in inspiratory muscle pressure (pMusc) and respiratory rate (RR), as detailed below.
2.1 Signal processing and tidal volume measurement
Outflow waveforms were acquired at 100 Hz (Fig. 2, second row) and processed to calculate tidal volumes. First, the value of the constant helmet inflow was subtracted from the recorded outflow waveform, resulting in a waveform (named “zeroed outflow”) with a shape identical to the original but with a different off-set to obtain a null value at end expiration (e.g., 60 L/min outflow recorded at the PEEP valve – 60 L/min inflow from the compressed gas regulator; Fig. 2, third row). Second, zeroed outflow waveforms were integrated and reset at each cycle, resulting in an inspiratory and an expiratory tidal volume (Fig. 2, fourth row). Off-set of the zeroed outflow waveform was fine-tuned to equalize inspiratory and expiratory volumes. The average tidal volume obtained during three respiratory cycles was considered for analyses.
2.2 Bench tests
The first set of tests (adequate inflows) was performed by maintaining the inflow at a level higher than the patient’s peak inspiratory flow for any inspiratory effort. Mild-moderate inspiratory efforts (pMusc 10 and 15 cmH2O) were studied with 60 L/min helmet inflow, resembling clinical practice. Tidal volumes were measured at three PEEP levels (5, 10, 15 cmH2O) under different RR combinations (15, 20, 25 breaths per minutes). Strong inspiratory efforts (pMusc 20 and 25 cmH2O) were studied at 75 and 90 L/min inflows to mimic a condition in which inflow is increased to match the patient’s inspiratory effort. The three aforementioned PEEP levels were studied at different RR combinations (15, 20, 25 breaths per minutes for pMusc 20 cmH2O; 20, 25 breaths per minutes for pMusc 25 cmH2O).
In the second set of tests (insufficient inflows), we studied the patterns of strong inspiratory effort at a standard helmet inflow (60 L/min) to test how the system accuracy would be affected in conditions of insufficient inflow to match the patient’s peak inspiratory flows. In such conditions, during part of the inspiration, the inflow is completely diverted into the lung simulator due to the strong inspiratory effort, and the outflow drops to zero; lower accuracy and underestimation of tidal volume were therefore expected during this set of tests.
2.3 Statistical analysis
Data are presented as the mean ± standard deviation, as not otherwise specified. The normality of the collected variables was assessed using the Kolmogorov‒Smirnov test. Correlations between normal variables were evaluated by Pearson’s correlation coefficient; otherwise, Spearman’s rho was used. Agreement between the reference tidal volumes and the volumes measured with the novel technique was evaluated by the Bland‒Altman plot.[14] Variables collected in the same conditions of respiratory effort but with different helmet inflows were compared by the Mann‒Whitney test. Artworks were created using SigmaPlot 11.0 (Systat software, Palo Alto CA).
3 Results
A total of 63 combinations of PEEP, pMusc, RR, and helmet inflow were included in the analysis. Peak inspiratory flows ranged from 39 to 112 L/min; tidal volumes ranged from 250 to 910 mL.
3.1 Adequate inflow tests
A total of 18 combinations were studied in the mild-moderate inspiratory effort tests at standard (60 L/min) inflows; peak inspiratory flow ranged from 39 to 73 L/min, and reference tidal volumes ranged from 250 to 610 mL. Reference tidal volumes and measured volumes were tightly correlated (r = .990 p < .001); the Bland‒Altman analysis showed a bias of -9.2 ± 27.4 mL, corresponding to an average percent error of -2.3 ± 3.4% (range − 9.3 to 2.3%).
A total of 30 combinations were studied in the strong inspiratory effort tests at increased (75 and 90 L/min) inflows; peak inspiratory flow ranged from 74 to 112 L/min, and reference tidal volumes ranged from 475 to 910 mL. Reference tidal volumes and measured volumes were tightly correlated (r = .969 p < .001). The Bland‒Altman analysis showed a bias of 0.3 ± 69.2 mL, which was not different from the mild-moderate condition (p = .189 by Mann‒Whitney Test), corresponding to an average percent error of 0.2 ± 4.8% (range − 9.1 to 8.9%).
Pooling together data obtained from the adequate inflow tests, the Bland‒Altman analysis showed a bias of -3.2 ± 29.3 mL, corresponding to an average percent error of -1 ± 4.4% (Fig. 3). In univariate analyses, tidal volume underestimation correlated with RR (rho = .411, p = .004) but not with peak inspiratory flow (p = .537), pMusc (p = .989), or PEEP (p = .382).
3.2 Insufficient inflow
A total of 15 combinations were studied in the strong inspiratory effort tests at standard (60 L/min) inflow, a value purposely lower than inspiratory flow, to verify the hypothesis that tidal volume is underestimated when inflow does match the patient’s peak inspiratory effort. Reference tidal volumes ranged from 480 to 765 mL, while peak inspiratory flows exceeded the inflow ranging from 75 to 110 L/min. Reference tidal volumes were significantly correlated with measured volumes (r = .882 p < .001). However, the Bland‒Altman analysis showed a bias of -93.3 ± 83.9 mL, which was higher than the mild-moderate condition with 60 L/min inflow (p < .001 by Mann‒Whitney Test). The corresponding average relative error showed a consistent tidal volume underestimation (-14.8 ± 6.3% compared to reference, range − 24.2 to 5.6%, open circles in Fig. 4), closely related to high peak inspiratory flows (r=-.905 p < .001), pMusc (rho − 0.819 p < .001), RR (rho − 0.730 p = .002), but not PEEP (p = .841).
4 Discussion
In the present bench study, we showed that it is possible to measure tidal volumes during Helmet CPAP by simply monitoring the outlet port of the helmet. Several combinations of PEEP, respiratory rate and inspiratory effort were investigated, mimicking a wide range of clinical possibilities. Tidal volume measurements were quite accurate (error below ± 10%) along a broad spectrum of simulations, provided that the helmet inflow was adjusted to the level of the patient’s inspiratory distress. Higher patient peak inspiratory flows coupled with insufficient inflows were associated with tidal volume underestimation.
The study was designed as a proof-of-concept to show the feasibility and effectiveness of a simple technique to measure patients’ tidal volumes during helmet CPAP noninvasive respiratory support. The principle at the basis of the measurement is quite simple: inspiratory tidal volume is measured as the inlet volume not flowing out of the helmet, and expiratory volume is measured as the outlet volume in excess on top of the expected continuous outflow. The prerequisite to obtain a reliable measurement is to adjust the inflow to the patient’s inspiratory needs, as clinically indicated when strong inspiratory efforts cause relevant helmet pressure oscillations, avoiding (or minimizing) phases of zero outflow during inspiration. When outflow is reduced to zero, the amount of patient’s inspiratory flow exceeding the inflow is not detectable at the outlet port, leading to an underestimation of the measured tidal volume. It should be noted that in the case of inspiratory flow exceeding the inflow, the patient can still receive flow from the helmet compression volume, which cannot be computed by a simple outflow trace measurement. Such a limitation is intrinsic to the technique but can be easily overcome: when inspiratory effort is such that inflow becomes insufficient (as detected by tidal interruption of outflow), the clinician should increase the inflow to reduce helmet pressure swings; additionally, correct tidal volume measurements will be restored.
Tidal volume monitoring is a relevant piece of information during noninvasive respiratory support to prevent the risk of excessively large tidal volumes associated with noninvasive treatment failure and further lung damage. In common clinical practice, tidal volume is not measured during helmet CPAP therapy and is possible during mask CPAP therapy only when using a ventilator. Therefore, several tidal volume measurement techniques have been studied in the past, and others are now emerging as novel noninvasive devices. Bedside noninvasive measurement of tidal volumes by inductive plethysmography has been known for many years, but clinical application is limited to pediatric settings [15, 16]. Improved plethysmography techniques, such as optoelectronic or fiber optic, were studied but remained at a research level; thoracic impedance or exhaled gas temperature were investigated in recent years with promising results and began to be used in the clinical scenario [17,18,19,20,21]. Recently, a modified ventilator connected to a helmet equipped with an intentional leak port was used to measure tidal volumes during CPAP therapy with good results [22].
The main advantage of the described technique is its simplicity: it is applicable to a standard continuous-flow helmet CPAP system requiring only a pneumotachograph and a signal processing unit, which could be easily integrated into a continuous-flow generator machine. The trend toward an underestimation error will result in a high positive predictive value for the detection of excessive tidal volume during noninvasive respiratory support: if the tidal volume measured by the system is elevated, the actual tidal volume would be similar or higher, revealing a condition of potential harm for the patient due to self-inflicted lung injury. The tested helmet inflow levels (60, 75 and 90 L/min) resemble levels typically used for patient care: a standard inflow level of 60 L/min is commonly used for the care of mild-distress patients, while distressed patients may require inflow levels up to 90 L/min. Patient comfort is generally not affected by higher inflows because the gas freely escapes through the PEEP valve; however, increased noise may be noted [23]. The three tested PEEP levels (5, 10 and 15 cm H2O) covered the most common helmet use; the effectiveness of the proposed tidal volume monitoring system at the highest PEEP level suggests that PEEP is not a limiting factor.
However, the presented data also show limitations, mainly due to the bench nature of the study: in vivo tests are lacking, especially the presence of variable leaks, which could be a limitation; signal processing needs to be automated; and confounding factors frequently present in the clinical scenario, such as the presence of leaks or the use of different PEEP valves, need to be evaluated.
5 Conclusions
In the present proof-of-concept bench study, we showed that tidal volume measurement based on the analysis of the outflow signal is feasible and accurate during continuous-flow helmet CPAP therapy, provided that helmet inflow is maintained at a sufficient level to match the patient’s peak inspiratory flow. Insufficient inflow resulted in tidal volume underestimation. In vivo data are needed to confirm such findings.
References
Rochwerg B, Brochard L, Elliott MW, Hess D, Hill NS, Nava S, Navalesi PMOTSC, Antonelli M, Brozek J, Conti G, Ferrer M, Guntupalli K, Jaber S, Keenan S, Mancebo J, Mehta S, Raoof SMOTTF. (2017) Official ERS/ATS clinical practice guidelines: noninvasive ventilation for acute respiratory failure. Eur Respir J 50.
Patel BK, Wolfe KS, Pohlman AS, Hall JB, Kress JP. Effect of Noninvasive Ventilation delivered by Helmet vs Face Mask on the rate of endotracheal intubation in patients with Acute Respiratory Distress Syndrome: a Randomized Clinical Trial. JAMA. 2016;315:2435–41.
Grieco DL, Menga LS, Cesarano M, Rosa T, Spadaro S, Bitondo MM, Montomoli J, Falo G, Tonetti T, Cutuli SL, Pintaudi G, Tanzarella ES, Piervincenzi E, Bongiovanni F, Dell’Anna AM, Delle Cese L, Berardi C, Carelli S, Bocci MG, Montini L, Bello G, Natalini D, De Pascale G, Velardo M, Volta CA, Ranieri VM, Conti G, Maggiore SM, Antonelli M. Effect of Helmet Noninvasive Ventilation vs High-Flow Nasal Oxygen on Days Free of respiratory support in patients with COVID-19 and moderate to severe hypoxemic respiratory failure: the HENIVOT Randomized Clinical Trial. JAMA. 2021;325:1731–43.
Perkins GD, Ji C, Connolly BA, Couper K, Lall R, Baillie JK, Bradley JM, Dark P, Dave C, De Soyza A, Dennis AV, Devrell A, Fairbairn S, Ghani H, Gorman EA, Green CA, Hart N, Hee SW, Kimbley Z, Madathil S, McGowan N, Messer B, Naisbitt J, Norman C, Parekh D, Parkin EM, Patel J, Regan SE, Ross C, Rostron AJ, Saim M, Simonds AK, Skilton E, Stallard N, Steiner M, Vancheeswaran R, Yeung J, McAuley DF. Effect of noninvasive respiratory strategies on intubation or mortality among patients with Acute Hypoxemic respiratory failure and COVID-19: the RECOVERY-RS Randomized Clinical Trial. JAMA. 2022;327:546–58.
Ferreyro BL, Angriman F, Munshi L, Del Sorbo L, Ferguson ND, Rochwerg B, Ryu MJ, Saskin R, Wunsch H, da Costa BR, Scales DC. Association of Noninvasive Oxygenation Strategies with all-cause mortality in adults with Acute Hypoxemic Respiratory failure: a systematic review and Meta-analysis. JAMA. 2020;324:57–67.
Coppadoro A, Zago E, Pavan F, Foti G, Bellani G. The use of head helmets to deliver noninvasive ventilatory support: a comprehensive review of technical aspects and clinical findings. Crit Care. 2021;25:327.
Coppadoro A, Benini A, Fruscio R, Verga L, Mazzola P, Bellelli G, Carbone M, Mulinacci G, Soria A, Noe B, Beck E, Di Sciacca R, Ippolito D, Citerio G, Valsecchi MG, Biondi A, Pesci A, Bonfanti P, Gaudesi D, Bellani G, Foti G. Helmet CPAP to treat hypoxic pneumonia outside the ICU: an observational study during the COVID-19 outbreak. Crit Care. 2021;25:80.
Bellani G, Grasselli G, Cecconi M, Antolini L, Borelli M, De Giacomi F, Bosio G, Latronico N, Filippini M, Gemma M, Giannotti C, Antonini B, Petrucci N, Zerbi SM, Maniglia P, Castelli GP, Marino G, Subert M, Citerio G, Radrizzani D, Mediani TS, Lorini FL, Russo FM, Faletti A, Beindorf A, Covello RD, Greco S, Bizzarri MM, Ristagno G, Mojoli F, Pradella A, Severgnini P, Da Macalle M, Albertin A, Ranieri VM, Rezoagli E, Vitale G, Magliocca A, Cappelleri G, Docci M, Aliberti S, Serra F, Rossi E, Valsecchi MG, Pesenti A, Foti G. Noninvasive ventilatory support of patients with COVID-19 outside the Intensive Care units (WARd-COVID). Ann Am Thorac Soc. 2021;18:1020–6.
Chiumello D, Chiodaroli E, Coppola S, Cappio Borlino S, Granata C, Pitimada M, Wendel Garcia PD. Awake prone position reduces work of breathing in patients with COVID-19 ARDS supported by CPAP. Ann Intensive Care. 2021;11:179.
Retucci M, Aliberti S, Ceruti C, Santambrogio M, Tammaro S, Cuccarini F, Carai C, Grasselli G, Oneta AM, Saderi L, Sotgiu G, Privitera E, Blasi F. Prone and lateral positioning in spontaneously breathing patients with COVID-19 Pneumonia Undergoing Noninvasive Helmet CPAP Treatment. Chest. 2020;158:2431–5.
Bellani G, Laffey JG, Pham T, Madotto F, Fan E, Brochard L, Esteban A, Gattinoni L, Bumbasirevic V, Piquilloud L, van Haren F, Larsson A, McAuley DF, Bauer PR, Arabi YM, Ranieri M, Antonelli M, Rubenfeld GD, Thompson BT, Wrigge H, Slutsky AS, Pesenti A. Noninvasive ventilation of patients with Acute Respiratory Distress Syndrome. Insights from the LUNG SAFE study. Am J Respir Crit Care Med. 2017;195:67–77.
Carteaux G, Millan-Guilarte T, De Prost N, Razazi K, Abid S, Thille AW, Schortgen F, Brochard L, Brun-Buisson C, Mekontso Dessap A. Failure of noninvasive ventilation for De Novo Acute Hypoxemic Respiratory failure: role of tidal volume. Crit Care Med. 2016;44:282–90.
Brochard L, Slutsky A, Pesenti A. Mechanical ventilation to minimize progression of Lung Injury in Acute Respiratory failure. Am J Respir Crit Care Med. 2017;195:438–42.
Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.
Allen JL, Greenspan JS, Deoras KS, Keklikian E, Wolfson MR, Shaffer TH. Interaction between chest wall motion and lung mechanics in normal infants and infants with bronchopulmonary dysplasia. Pediatr Pulmonol. 1991;11:37–43.
Watson HL, Poole DA, Sackner MA. Accuracy of respiratory inductive plethysmographic cross-sectional areas. J Appl Physiol (1985). 1988;65:306–8.
Williams GW 2nd, George CA, Harvey BC, Freeman JE. A comparison of measurements of change in respiratory status in spontaneously breathing volunteers by the ExSpiron Noninvasive respiratory volume monitor Versus the Capnostream Capnometer. Anesth Analg. 2017;124:120–6.
Lerman J, Feldman D, Feldman R, Moser J, Feldman L, Sathyamoorthy M, Deitch K, Feldman U. Linshom respiratory monitoring device: a novel temperature-based respiratory monitor. Can J Anaesth. 2016;63:1154–60.
Aliverti A, Dellaca R, Pelosi P, Chiumello D, Pedotti A, Gattinoni L. Optoelectronic plethysmography in intensive care patients. Am J Respir Crit Care Med. 2000;161:1546–52.
Davis C, Mazzolini A, Murphy D. A new fibre optic sensor for respiratory monitoring. Australas Phys Eng Sci Med. 1997;20:214–9.
Kodali BS, Choi L, Chau A, Harvey BC, Brayanov J, Tsen LC, Palanisamy A. Use of a novel non-invasive respiratory monitor to study changes in pulmonary ventilation during labor epidural analgesia. J Clin Monit Comput. 2020;34:567–74.
Cortegiani A, Navalesi P, Accurso G, Sabella I, Misseri G, Ippolito M, Bruni A, Garofalo E, Palmeri C, Gregoretti C. Tidal volume estimation during Helmet Noninvasive Ventilation: an experimental feasibility study. Sci Rep. 2019;9:17324.
Lucchini A, Bambi S, Gurini S, Di Francesco E, Pace L, Rona R, Fumagalli R, Foti G, Elli S. Noise level and comfort in healthy subjects undergoing High-Flow Helmet continuous positive Airway pressure. Dimens Crit Care Nurs. 2020;39:194–202.
Funding
UNIMIB department funds.
Open access funding provided by Università degli Studi di Milano - Bicocca within the CRUI-CARE Agreement.
Author information
Authors and Affiliations
Contributions
All authors contributed to the study conception and design. Data collection and analysis were performed by AC, who drafted manuscript; GF and GB contributed to data interpretation and revised the manuscript. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Competing interests
AC and GB are co-inventors of a patent owned by Flowmeter for a project unrelated to the present work. AC and GF are co-inventors of a patent owned by Dimar. GB has relationships with Drager Medical, Getinge, and Siaretron. GF received lecture fees by Dimar. GB and GF received research funds from Dimar for a project unrelated to the present work.
Additional information
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Springer Nature or its licensor (e.g. a society or other partner) holds exclusive rights to this article under a publishing agreement with the author(s) or other rightsholder(s); author self-archiving of the accepted manuscript version of this article is solely governed by the terms of such publishing agreement and applicable law.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Coppadoro, A., Bellani, G. & Foti, G. A technique to measure tidal volume during noninvasive respiratory support by continuous-flow helmet CPAP. J Clin Monit Comput 37, 1473–1479 (2023). https://doi.org/10.1007/s10877-023-01034-0
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10877-023-01034-0