Abstract
Purpose
The purpose of this paper was to review the available approaches for bone strength assessment, osteoporosis diagnosis and fracture risk prediction, and to provide insights into radiofrequency echographic multi spectrometry (REMS), a non-ionizing axial skeleton technique.
Methods
A working group convened by the European Society for Clinical and Economic Aspects of Osteoporosis and Osteoarthritis met to review the current image-based methods for bone strength assessment and fracture risk estimation, and to discuss the clinical perspectives of REMS.
Results
Areal bone mineral density (BMD) measured by dual-energy X-ray absorptiometry (DXA) is the consolidated indicator for osteoporosis diagnosis and fracture risk assessment. A more reliable fracture risk estimation would actually require an improved assessment of bone strength, integrating also bone quality information. Several different approaches have been proposed, including additional DXA-based parameters, quantitative computed tomography, and quantitative ultrasound. Although each of them showed a somewhat improved clinical performance, none satisfied all the requirements for a widespread routine employment, which was typically hindered by unclear clinical usefulness, radiation doses, limited accessibility, or inapplicability to spine and hip, therefore leaving several clinical needs still unmet. REMS is a clinically available technology for osteoporosis diagnosis and fracture risk assessment through the estimation of BMD on the axial skeleton reference sites. Its automatic processing of unfiltered ultrasound signals provides accurate BMD values in view of fracture risk assessment.
Conclusions
New approaches for improved bone strength and fracture risk estimations are needed for a better management of osteoporotic patients. In this context, REMS represents a valuable approach for osteoporosis diagnosis and fracture risk prediction.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Introduction
Osteoporosis is a systemic skeletal disease characterised by the reduction in bone mass and the degeneration of bone structure that leads to an increased risk of fracture [1], with important social, medical and community care costs.
Bone strength, which is a measure of the resistance to bone fracture [2], is determined by a composite summation of numerous skeletal characteristics that can be divided into four basic components: composition, microarchitecture, size, and shape (Fig. 1). Several factors contribute to the definition of each skeletal characteristic, such as cellular density, mineralization, collagen crosslinking at a nanoscale level for the definition of bone composition; trabecular and cortical properties including porosity, thickness, connectivity at microscale level for the definition of bone microarchitecture; age, genetics, gender, and habits for the definition of bone size and shape at a macroscale level. Understanding how these variables interact and contribute to bone strength is critical in the development of fracture prediction tools [3, 4].
Available imaging techniques are able to capture a measure of the aforementioned factors and could also provide information about bone strength. Therefore, in principle, several tools are necessary to assess the different bone properties, such as spectroscopy and microscopy at nano-scale level, high-resolution quantitative computed tomography (HR-QCT) or µCT at micro-scale level, dual-energy X-ray absorptiometry (DXA) or CT at macro-scale level [3, 4]. Nowadays, in clinical practice, bone strength is indirectly estimated by bone mineral density (BMD) measured by DXA, which currently represents a well-recognized technology for the diagnosis of osteoporosis worldwide. Nonetheless, several studies demonstrated that BMD only accounts for about 50–70% of the bone strength variation, being able to describe features related to the quantity of bone tissue but overlooking information about bone quality [4,5,6].
These considerations argue for the introduction in clinical routine of further tools in order to better define and estimate bone strength and to predict the risk of fracture.
To address this, a working group convened by the European Society of Clinical and Economic Aspects of Osteoporosis and Osteoarthritis (ESCEO) met to review the data on the currently available imaging techniques for the in vivo bone strength assessment and to discuss principles, applications and perspectives of a new technology called Radiofrequency Echographic Multi-Spectrometry (REMS) on the basis of the latest available data. Specific attention has been paid to identify the clinical needs related to bone strength assessment, fracture risk prediction and osteoporotic patient management that are not completely satisfied by the techniques currently employed in clinical routine and to their possible resolution by the introduction of innovative approaches.
Clinical assessment of bone strength
Several approaches have been developed to provide an estimation of bone strength, including imaging devices, risk fracture calculators, bone biopsy techniques, and laboratory tests (Table 1). In this review, we will focus on the imaging devices: the main features of the most common approaches are described and discussed in the following paragraphs.
Dual-energy X-ray absorptiometry
The most widely used technique to obtain information about bone strength is DXA, currently considered as the gold standard densitometric technique for the diagnosis of osteoporosis. Through the acquisition of two-dimensional (2D) X-ray scans with an effective dose in the range of 1–13 µSv [7,8,9], it estimates an areal BMD expressed in g/cm2 (i.e., the ratio between bone mineral content estimated through X-ray absorption and 2D projection of the scanned bone area), which is a surrogate measure of bone strength [3]. Ex vivo studies performed on isolated human bones showed that areal BMD accounts for about two thirds of the variation of bone strength as evaluated by shear test of the femoral neck or compression of the vertebrae [5, 10].
In clinical practice, DXA is employed to categorise the patient by comparing the measured BMD value with the average BMD of a reference healthy population of young individuals, which allows to define the patient’s T-score as the number of standard deviations (SDs); the measured BMD differs from the reference value. Thus, patients are classified as healthy (normal BMD, T-score ≥ − 1), osteopenic (− 2.5 < T-score < − 1), osteoporotic (T-score ≤ − 2.5) or severely osteoporotic (T-score ≤ − 2.5 in the presence of one or more fragility fractures), according to the definitions given by the World Health Organization (WHO) [11, 12]. Moreover, DXA is employed to provide prognostic information on probability of fracture and to monitor bone status in longitudinal studies [13]. A meta-analysis of prospective cohort studies correlating the baseline BMD with the subsequent follow-up for fractures showed that the measurement at the spine and at the hip seemed to have a better predictive ability for spine and hip fractures, respectively [14]. Leslie et al. investigating whether the rate of BMD loss would predict fracture risk independently of current BMD, found that routine clinical DXA measurement in untreated women cannot accurately characterise the rate of BMD loss due to the dominating effect of measurement errors [15].
BMD alone for the assessment of fracture risk has a high specificity but low sensitivity, with many fragility fractures occurring in patients who did not have a diagnosis for osteoporosis as defined by the WHO classification based on T-score value, but whose BMD lay in the osteopenic range [16,17,18]. Moreover, in the diabetes paradigm, BMD is unable to explain the increased risk of fracture: in particular, for Type 1 diabetes, the highly increased risk of hip fracture was only partially explained by the observed BMD reduction, whereas in Type 2 diabetes the increased risk of fracture is not captured by the paradoxically higher BMD [19].
Thus, a screening program for osteoporosis based on DXA-measured BMD alone cannot be recommended for the whole population [20, 21], and further studies evaluating the most cost-effective screening strategy are warranted [22]. These findings also suggest the hypothesis that multiple factors, other than BMD, contribute to fracture risk and highlight the importance of BMD-independent determinants of bone strength and fracture risk assessment.
DXA scanners, if equipped with dedicated software modules, can also measure additional parameters related to textural features and hip geometry, as described in the subsequent paragraphs.
Trabecular bone score
The Trabecular Bone Score (TBS), calculated by the software package TBS iNsight® (Medimaps Group SA, Geneva, Switzerland), is a unit-less indirect index of trabecular microarchitecture based on pixel grey level variations in the DXA images [23, 24]. The large clinical database of the Province of Manitoba (Canada), including 29,407 women aged over 50 years at baseline, has been used for several investigations aimed at assessing the ability of TBS in fracture risk prediction. Although TBS (along with BMD and additional clinical risk factors) helps to identify individuals at high risk of fracture and to guide initiation of osteoporosis treatment [25], change in lumbar spine TBS is not a useful indicator of fracture risk regardless of osteoporosis treatment [26]. Another retrospective study on the Manitoba database showed the ability of lumbar spine TBS to predict osteoporotic fractures in patients with type 2 diabetes, whereas BMD was paradoxically increased in these patients [27].
TBS values can be also input into the Fracture Risk Assessment tool FRAX® [28]. This tool, available online at www.shef.ac.uk/FRAX [29], aims to predict osteoporotic fractures on the basis of several risk factors optionally including femoral neck BMD or spinal TBS [17]. A recent meta-analysis showed that TBS, adjusted for time since baseline measurement and age, has a significant correlation with major osteoporotic fractures, which is slightly improved when FRAX® probability is taken into account, supporting the use of TBS both as a stand-alone fracture risk prediction tool and as a valuable add-on to FRAX® [30].
Nonetheless, some limitations of TBS should be also mentioned: increased noise in DXA images results in an artefactually reduced TBS [31]; TBS, but not BMD, is dependent on the used scan mode (standard versus thick), at least for GE Lunar Prodigy scanners [32]; increases in soft-tissue thickness overlying the spine lead to diminished TBS [33]; TBS reliability in men depends on BMI [34, 35].
Furthermore, a debate is still open in the scientific literature as regards the proportion in which TBS explains the variability in vertebral strength. The very few available biomechanical studies [36, 37], which anyway represent only an indirect approach to what happens in vivo, found poor or no direct correlation between TBS and vertebral failure load, although reporting in some cases significant correlations with other elastic properties (e.g., r = 0.64 with bone stiffness [36]).
Overall, lumbar spine TBS is an evolving software that has been shown to be useful as a complementary fracture risk prediction tool. However, the magnitude of TBS increase in osteoporosis treatment is smaller than that of BMD. The relationship between change in TBS and fracture risk reduction remains to be elucidated with recent and ongoing studies that are helping to refine its clinical utility [28, 38].
Hip axis length and hip structural analysis
The hip axis length (HAL), i.e. the distance from the base of the greater trochanter to the inner pelvic rim (Fig. 2), is a further non-BMD parameter that has been considered for the assessment of hip bone strength through DXA [39]. A study by Leslie et al. [40] showed that the risk of hip fracture, adjusted for age and femoral neck BMD, increases by 3.6% in men (p = 0.022) and by 4.6% in women (p < 0.001) for every millimetre of increase in HAL. However, as pointed out by the International Society of Clinical Densitometry (ISCD) at the 2015 Position Development Conference, the association between HAL and hip fracture is actually ambiguous, with 14 published studies showing a positive association between HAL and hip fracture, 11 studies showing no association and 1 study reporting a negative association [41]. On this basis, the Hip Structural Analysis (HSA) software was developed to simultaneously evaluate both hip geometrical information and corresponding mineral mass data derived from DXA images [42]. Whether HSA could be actually useful for clinical assessment of bone fragility is currently under investigation, with few studies showing that some factors, i.e. cortical thickness of the intertrochanteric region [43], HAL and neck shaft angle [44] might have a role in fracture risk assessment for specific categories of patients.
Quantitative computed tomography
Quantitative computed tomography (QCT) allows to obtain a tomographic 3D reconstruction of the scanned bone and provides a volumetric bone mineral density (vBMD), expressed in g/cm3, which is less susceptible to osteo-arthritis artefacts than the areal BMD measured by DXA. Depending on the specific adopted protocol, the exposure dose for a lumbar QCT scan is in the range of 50–500 µSv, which is clearly lower than a standard abdominal CT scan (approximatively 8 mSv) [7, 8]. In contrast to DXA, this analysis allows separate estimation of trabecular and cortical BMD. Since trabecular bone has a substantially higher metabolic turnover than cortical bone, QCT is more sensitive to changes in BMD than DXA [45]. The American College of Radiology (ACR) has recently introduced practice guidelines for the use of QCT, with the definition of equivalent WHO classification of osteoporosis classes based on spine trabecular volumetric BMD measurements and an association between volumetric BMD and risk of fracture has also been established. Moreover, the correlation between BMD measured by QCT and derived by DXA has also been assessed. Thus, QCT might be used in case of non-availability of DXA. It has also some important advantages over DXA that are useful in specific clinical indications, i.e. abnormal body size (small/large body frame) or very obese patients, advanced degenerative spine disease (e.g. idiopathic skeletal hyperostosis, degenerative disc disease, facet arthropathy), and need for accurate monitoring of trabecular metabolic activity [46].
In addition, QCT allows the development of finite-element analysis (FEA), a computational method in which each voxel represents a finite element with assigned material properties (e.g., elastic modulus). Load simulations might be performed with the estimation of the load at which structural failure occurs, so identifying a biomechanical fracture threshold [47]. Indeed, the prospective 5-year AGES-Reykjavik case–control study of 1110 patients reported significant correlation between bone strength estimated by FEA and fracture occurrence, with age-adjusted odds ratios for low vertebral strength and incident spine fractures of 2.8 (95% CI 1.8–4.3) and 2.2 (2.2, 95% CI 1.5–3.2), and age-adjusted odds ratios for low femoral strength and incident hip fractures of 4.2 (95% CI 2.6–6.9) and 3.5 (95% CI 2.3–5.3) for women and men, respectively [48]. The FEA performed on QCT images also provides information on femoral strength changes associated with therapies, as in the “PaTH” study evaluating the effects of PTH(1–84) and/or alendronate in osteoporotic women [49]. It was shown that strength was mainly influenced by trabecular density changes, regardless of the pharmacological treatment, although changes in the cortical bone density and overall bone geometry also contributed to femoral strength changes, in that case depending on the drug used. Due to the limited clinical availability of QCT and to the associated high costs and radiation doses [50], in the past years, several studies investigated the possibility to obtain finite-element models from DXA imaging, but their clinical feasibility remains very limited and their actual usefulness should be further explored [51,52,53].
Peripheral QCT (pQCT) has a more widespread availability and examines appendicular skeleton sites with a lower effective radiation dose than central QCT (lower than 0.01 mSv [8]). Through the simultaneous calculation of BMD, bone and muscle geometrical parameters, and biomechanical parameters, it provides an evaluation of the “muscle-bone unit”, with the potential of representing a valid functional approach for bone health assessment [54]. However, with respect to DXA, pQCT has anyway a more limited accessibility, being also sensitive to motion artefacts.
High-resolution peripheral QCT
High-resolution peripheral QCT (HR-pQCT), with a single scan exposure dose of about 3–4 µSv [55, 56], allows one to obtain volumetric images of distal bones, i.e. radius and tibia, with a simultaneous measure of cortical and trabecular vBMD and a deeper representation of bone microarchitecture, thus performing the so-called virtual biopsy. A micro-FEA for peripheral bones might be thus derived from HR-pQCT scans. A cross-section of the distal radius or tibia is scanned at a resolution sufficiently high to resolve the trabecular microstructure (in the order of 80 μm). Based on these images, micro-FE models are generated and a mechanical test is simulated to derive the stiffness of the scanned region and to estimate the strength of the whole bone. In validation studies, micro-FEA showed its ability to predict bone failure load better than any density-based parameter, but this improved strength prediction did not result in a better separation of subjects with and without distal radius fractures in retrospective studies, probably due to the reduced reproducibility, and consequently reduced specificity and sensitivity, in population retrospective studies with respect to validation studies performed on cadavers [57]. Currently, several HR-pQCT devices are available worldwide [58, 59] and their use is mainly reserved for research purposes [60, 61], although there is potential to use this technique in the clinical diagnosis and management of osteoporosis irrespectiwhether its ability in fracture prediction will be fully demonstrated [59, 61].
Magnetic resonance imaging
In magnetic resonance imaging (MRI), and high-resolution MRI as well, the hydrogen atoms in water molecules are exploited to obtain a contrast between marrow signal (hyper-intense) and trabecular structure (hypo-intense). With this non-ionizing imaging modality, structural trabecular parameters (i.e., trabecular number, thickness, connectivity, anisotropy and shape) might be obtained and FEA can be applied similarly to the case of QCT [62]. However, some important limitations such as high costs, long scan time, motion artefacts and partial volume effects (in case of insufficient image resolution that results in reduced contrast between adjacent tissues and blurring effects) make this imaging technique scarcely employed and investigated in osteoporosis diagnosis [63]. Furthermore, to the best of our knowledge, no significant longitudinal studies have been conducted to assess fracture risk prediction through MRI.
Quantitative ultrasound
Quantitative ultrasound (QUS) devices measure bone properties using differential reflections and attenuation of pulsed ultrasound waves. According to the ISCD Official Positions, the only validated skeletal site for the clinical use of QUS in osteoporosis management is the heel, which, in conjunction with clinical risk factors, can be used in patients aged over 65 in order to identify low-risk populations [64, 65]. The most common parameters of interest for this non-ionizing technique are those derived from the combination of the speed of sound (SOS) and the broadband ultrasound attenuation (BUA), like the stiffness index (SI) and the quantitative ultrasound index (QUI). SOS, which relies on the accurate measurement of the time employed by a sound wave to pass through the heel, is directly proportional to the BMD. Analogously, BUA, which measures the reduction in intensity at different frequencies of a broadband ultrasound pulse sent through the bone, shows a greater attenuation of the higher frequencies in strong bones than in weak bones [66]. The SI, for instance, was empirically derived as the sum of the normalized BUA and SOS values, with a comparable contribution from the two factors [67].
Population studies have shown the capabilities of heel QUS to predict osteoporotic fractures, especially for hip, with similar sensitivity but lower specificity than DXA in discriminating fractured from non-fractured subjects [68,69,70,71]. A meta-analysis on 14 prospective studies showed that the estimated relative fracture risk ranged between 1.23 and 1.94 for each SD decrease in QUS measurements depending on the measured QUS parameter and the type of fracture, suggesting that heel QUS might be used as an alternative to DXA in the assessment of non-vertebral fracture risk [72]. However, it has been also reported that QUS predictive value wanes with time, being for instance more reliable at 1 year after baseline measurement than at 5 years [73].
The main advantages of QUS with respect to DXA are lower cost, smaller space required, portability and absence of ionizing radiation. However, QUS approaches are applicable only to peripheral skeletal sites (calcaneus, tibia, phalanges, radius), and their diagnostic performances are typically inferior to the corresponding DXA ones. Moreover, the availability of several QUS devices that differ substantially from each other in terms of measured parameters has limited their acceptance. Furthermore, QUS results are also dependent on operator, anatomical location, and relative positioning of bone and ultrasound transducer: these drawbacks currently limit the employment of QUS as a clinical diagnostic tool, except for screening purposes [47, 64].
Role of bone strength in the assessment of fracture risk: state of the art
Areal BMD
Modelling bone strength might have implications for the understanding of osteoporosis mechanism, the assessment of fracture risk and the prediction, and monitoring of drug treatments. As expressed above, the main clinically available indicator to predict fracture risk is areal BMD estimated by DXA imaging.
In a study by Kopperdahl et al. [48], derived from the AGES-Reykjavik case-control cohort, the large majority of fracture events seems to occur when femoral neck T-score of areal BMD was below about − 1.5. However, the ability of this parameter to predict hip fractures significantly improved only if combined with femoral strength estimation for a net reclassification improvement of 33%. Bone strength estimated from QCT-derived FEA was significantly correlated with fracture risk, when adjusted for age and sex, independently from baseline BMD, indicating that a reduced bone strength might be a predictive factor for high risk of osteoporotic fracture.
The predictive ability over a long period of areal BMD was evaluated in the Study of Osteoporotic Fractures (SOF) [74], which assessed BMD and risk factors in almost 8000 women between 1988 and 1990, with a clinical follow-up for fractures lasting for 25 and 20 years for hip and any non-vertebral fracture, respectively. Although BMD alone might be predictive of fracture risk, the overall prediction would be improved by taking into account other risk factors and other bone quality descriptors more closely related to bone strength.
The sensitivity of areal BMD to treatment changes was assessed in studies evaluating drug treatments, such as the FREEDOM study (Denosumab versus placebo) [75] and the HORIZON study (Zoledronic versus placebo) [76], showing that an increase in total hip BMD after 3 years of treatment was correlated with a decrease in risk of non-vertebral fracture and, interestingly, the risk of fracture showed similar relationships (slopes) with per cent change in total hip BMD for both drug and placebo groups (Fig. 3, [75]). Moreover, a meta-regression of 38 placebo-controlled trials with follow-up ranging from 1 to 8 years showed that increments of 2–6% in total hip BMD were significantly associated with a 28–66% reduction in vertebral fractures, as well as with a 16–40% reduction in hip fractures, but not with reductions in non-vertebral fractures [77]. In general, in the osteoporosis treatment, the time-interval of repeated DXA-BMD measurements must be long enough to discern real BMD increments from precision error of repeated measurements (in the order of 1–2%) [78], with reasonable interval not less than 2 years.
In summary, areal BMD estimated by DXA imaging showed reasonable diagnostic performances being discriminative, predictive and sensitive to treatment changes. However, it is widely accepted that a deeper analysis of bone biomechanics is needed to increase the accuracy of fracture risk prediction by taking into account the actual bone strength [3, 17, 79, 80].
High-resolution peripheral QCT (HR-pQCT)
The main clinical issue associated with the use of BMD alone is the twilight zone of osteopenic patients, where the majority of fractures occurs, suggesting that the standard classification may overlook some important biomechanical characteristics that could improve the assessment of fracture risk [81]. For instance, the discriminative value of parameters derived from HR-pQCT at the distal radius (i.e., total and trabecular density and heterogeneity of trabecular network) was assessed in a setting of osteopenic postmenopausal women (PMW), with and without fractures, in which BMD of the lumbar spine and femoral neck did not significantly differ [82]. Statistically significant differences were found between fractured and non-fractured subjects, with the former having − 10% in total density, − 12.3% in trabecular density, and +25.6% in heterogeneity of trabecular network [82]. This paper, on the one hand, underlined the importance of the description of trabecular microarchitecture of peripheral sites for the enhancement of the prediction of fracture risk in osteopenic patients, but, on the other hand, given the preliminary nature of the study and the small number of enrolled patients, it also emphasized the need of additional studies to better understand the actual usefulness of the mentioned bone microarchitecture parameters. A recent study by the Bone Microarchitecture International Consortium (BoMIC) [83], including 8 large cohorts of patients for a total of 7254 individuals, investigated the association between indices derived from HR-pQCT and fracture risk. During an average follow-up of 4.6 years, 86% of incident fractures were registered in patients with femoral neck BMD in the osteopenic or normal range, but in whom deficits in bone microstructure were found with HR-pQCT. The results indicate that trabecular and cortical bone density and microstructure, as well as estimated failure load measured at the peripheral skeleton, predicted incident fractures independently of femoral neck BMD and FRAX®.
Moreover, a large multimodal investigation was performed by Biver et al. [60] in order to assess the contribution of bone phenotype parameters to the prediction of low-trauma fractures in PMW: peripheral trabecular and cortical volumetric BMD and microstructure, as well as the estimated bone strength by HR-pQCT, showed high prediction rate independently of femoral neck BMD, FRAX®, and TBS. As BMD assessed by DXA at the ultra-distal radius [84] showed very good performance for fracture prediction, the authors suggested that measuring ultra-distal radius BMD in addition to femoral neck BMD could be useful to refine fracture prediction for PMW in this age range, especially when bone microstructure cannot be specifically assessed due to the limited clinical availability of HR-QCT [57,58,59]. Similarly, a cross-sectional case-control study found that bone microstructure and strength assessment were able to discriminate patients on long-term glucocorticoids with vertebral fracture independently of BMD: noteworthy, with a reduction in total volumetric BMD and cortical thickness at the distal tibia, it was able to identify patients at high risk of vertebral fractures in the subgroup considered to have low fracture risk as assessed by DXA or FRAX® [85].
The main difficulty in the full assessment of bone strength is due to the numerous bone components involved in its definition. The introduction of advanced tools in clinical practice is needed in order to collect additional information for the modelling of bone strength and fracture risk.
FEA derived from QCT
Several attempts have been performed by FEA derived from QCT [86], sometimes including bone anisotropy information [87], in order to model bone strength and to assess a personalized biomechanical fracture threshold as the ratio of external forces applied to the bone and the FE-computed bone strength [81]. Micro-FEA derived from HR-pQCT was also used in the prospective STRAMBO study, which enrolled 825 men that were followed for 8 years, showing that the distal radius trabecular number was the parameter most strongly associated with fracture risk [88].
Microindentation
The recently developed in vivo impact microindentation may have a role in the explanation of fracture propensity in case of poor bone quality associated with normal BMD, such as type 2 diabetes, atypical femoral fractures, stress fractures, glucocorticoid treatment and HIV infection [89, 90]. For instance, in a study by Rozental et al. [91], lower bone strength index measured by impact microindentation at the anterior surface of the mid-tibia diaphysis in PMW was associated with distal radius and hip fractures and, if confirmed by prospective trials, might be integrated in models for the identification of women at risk of fragility fractures.
DXA-based 3D modelling
The most promising future approach could be the integration of the available technologies, such as DXA-based 3D modelling, as shown in a retrospective open-label study aiming to assess the effects of treatments on average changes in volumetric BMD and trabecular architecture over 24 months using DXA-based 3D modelling [92]. Nevertheless, the request for a long time to carry out the FEA hampers its use in the clinical practice, which will perhaps have to wait for the improvement of the underlying technology and mathematical-physical processes.
Currently unmet clinical needs
Despite the mentioned specific added values of some imaging methods alternative to DXA, the main routinely employed parameter to estimate bone strength and to predict fracture risk is still represented by DXA-measured BMD.
In fact, currently there is not a widely accessible and cost-effective tool capable of significantly improving fracture risk estimation in individuals with osteopenic or normal BMD, which account for about 50% of fragility fractures [93].
Typically, the described approaches increase the duration of the diagnostic procedure and ionizing radiation dose without a clear improvement of the clinical outcome. For instance, a very recent study [94] has evaluated the performance of QCT-based FEA as a predictor of hip fracture risk, documenting that, compared to the traditional DXA, it would imply a 15% increment in the total costs and an over 2000 times higher radiation dose, while reducing the fractured patients by less than 5%. This is also coupled with the important accessibility issues that are common to all the QCT-based techniques, which require all the strictest precautions typical of X-ray-bearing techniques, including also HR-pQCT, the use of which is further hindered by their very limited availability in clinical contexts [57,58,59].
However, although the cost-effectiveness of osteoporosis prevention through early diagnosis has been demonstrated [95], the identification of the most cost-effective imaging technique through thorough economic analyses (including different fracture states, country-specific aspects, etc.) is still debated in literature [21, 96] and the definition of a specific cost-effectiveness model is beyond the scope of this paper.
Anyway, there is a series of clinical needs that are not effectively satisfied because of the previously mentioned limitations of the currently employed techniques and would require the introduction of reliable methods for bone strength evaluation and fracture risk estimation. The most urgent of these unmet needs include the following: osteoporosis diagnosis in patient categories for which X-ray examinations are not feasible (e.g., paediatric subjects, pregnant and breastfeeding women, etc.); under-diagnosis and late diagnosis; effective bone quality assessment integrated with a reliable fracture risk prediction available in clinical routine; short-term follow-up of patients under treatment, since the present techniques require at least 1 year between two measurements.
Radiofrequency echographic multi spectrometry (REMS)
Basic principles
REMS technology is a non-ionizing axial approach for osteoporosis diagnosis. The operating principle is based on the analysis of native raw unfiltered ultrasound signals, the so-called radiofrequency (RF) ultrasound signals, acquired during an echographic scan of lumbar vertebrae and/or femoral neck. The analysis of native unfiltered ultrasound signals allows to retain the maximum information about the characteristics of the investigated tissues, which are normally filtered out during the conventional process of B-mode image reconstruction. The bone health status is assessed through the comparison of the analysed signal spectra with previously derived reference spectral models for the considered pathological and normal conditions [97, 98]. The large amount of collected data related to internal bone structure provides both quantity- and quality-related information, being thus theoretically suitable for the estimation of bone strength and the prediction of fracture risk.
In more detail, in a REMS investigation, the probe is placed on the abdomen or on the hip in order to visualize of the target bone interface and the operator has to set the appropriate values of scan depth and transducer focus. Subsequently, the software detects the sought bone interfaces in the sequence of acquired frames and identifies the regions of interest (ROIs) for the diagnostic evaluation (Fig. 4). One key-feature lies in the exploitation of B-mode images for the identification of target bone interfaces and related ROIs, combined with the diagnostic analyses performed on the RF data. REMS analysis is then characterised by the parallel processing of the unfiltered signals of several scan lines (Fig. 5).
The analysis of single scan line spectra allows the automatic exclusion of signals corresponding to artefacts, such as calcifications or osteophytes, thanks to the identification of unexpected spectral features.
The selected measured data are finally synthetized in a patient-specific spectrum of the considered bone target, which undergoes an advanced comparison with gender-, age-, site- and BMI-matched reference spectral models extracted from a dedicated database (Fig. 6).
Actually, the spectral modifications introduced by the physical properties of the bone structure that has backscattered the ultrasound signals are identified by the comparison procedure, resulting in a BMD estimation and in the consequent diagnostic classification as healthy, osteopenic or osteoporotic. The adopted approach is based on the calculation of the Osteoporosis Score, which corresponds to the percentage of analysed spectra that were classified as “osteoporotic” through the dedicated spectral analyses [97, 98]. Linear equations are then employed to transform Osteoporosis Score into BMD values.
REMS approach has the potential to calculate also parameters different from BMD, derived from bone quality indicators and further related to bone strength. Among these, Fragility Score (FS) has been developed as an independent indicator of bone quality that provides an estimation of fracture risk independently of BMD. It is a dimensionless number in the range 0–100 obtained from the comparison between the patient-specific spectrum and reference spectral models obtained from patients with and without osteoporotic fractures. Recent preliminary studies have shown that FS has a good accuracy in the discrimination between fractured and non-fractured subjects, and also that FS is significantly correlated with the fracture risk calculated by FRAX® when the latter included also the results of the femoral neck densitometry [99,100,101].
Ongoing developments of REMS are addressed to the investigation of other musculoskeletal tissues, such as muscles and cartilage, in order to monitor the onset and evolution of diseases like sarcopenia and arthrosis by early detecting the corresponding tissue deteriorations through dedicated REMS parameters.
Clinical validation studies
The application of REMS technology for osteoporosis diagnosis has been clinically validated through an observational multicentre clinical trial involving 7 Italian centres that enrolled over 1900 PMW [102]. The aim of the study was to assess the precision and diagnostic accuracy of REMS in comparison with the clinical gold standard reference, represented by DXA. All DXA acquisitions were double-checked to avoid possible errors as identified by Messina et al. [103]. Similarly, a quality control was performed on all the REMS reports in order to verify the correct selection of scan depth and transducer focus. The analysis of the temporal distribution of the errors indicated the possibility of improving the clinical practicability of REMS through a more rigorous training of the operators, aimed at reducing the time span of the learning curves and the related initial error rates. After the cross-check, 1195 spinal and 1373 femoral comparable cases were analysed.
The REMS intra-operator precision, expressed as root-mean-square coefficient of variation (RMS-CV) was 0.38% (95% confidence interval: 0.28–0.48%) for lumbar spine and 0.32% (0.24–0.40%) for femoral neck. The corresponding least significant change for the 95% confidence level, calculated via the ISCD precision calculator (available at http://www.iscd.org/resources/calculators/), was 1.05% for lumbar spine and 0.88% for femoral neck. The inter-operator repeatability was 0.54% for spine and 0.48% for femoral neck.
Afterwards, diagnostic accuracy of REMS was assessed by determining the concordance of REMS and DXA in the discrimination between osteoporotic, osteopenic and healthy patients through the calculation of the corresponding Cohen’s kappa (k), obtaining k = 0.82 for lumbar spine and k = 0.79 for femoral neck.
Sensitivity of REMS in discriminating osteoporotic subjects from non-osteoporotic ones turned out to be 91.7% for spine and 91.5% for femur, while specificity was 92.0% and 91.8%, respectively. Furthermore, the degree of correlation between DXA and REMS T-score values resulted to be r = 0.94 for lumbar spine and r = 0.93 for femoral neck (p < 0.001 for both).
Moreover, the agreement between BMD values obtained by DXA and REMS was calculated through the Bland–Altman method [104]: the average difference (expressed as bias ± 2 SDs) was − 0.004 ± 0.088 g/cm2 for lumbar spine and − 0.006 ± 0.076 g/cm2 for femoral neck. The standard error of the estimate (SEE) resulted equal to 0.044 g/cm2 (5.3%) for lumbar spine and 0.038 g/cm2 (5.8%) for femoral neck.
The results obtained from the referred multicentre clinical study showed that REMS technology has a high sensitivity and specificity in osteoporotic patient identification, with a significant diagnostic agreement with the gold standard DXA in the classification of patients as healthy, osteopenic or osteoporotic (total diagnostic concordance was 88.8% for lumbar spine and 88.2% for femoral neck). Both intra- and inter-operator variability associated with REMS investigations were better than the corresponding values typically reported in literature for the employed comparative gold standard DXA [105, 106].
Thanks to its radiation-free approach, REMS might be applied for population mass investigations or prevention programs, early diagnosis in clinical practice and therapeutic short-term follow-up.
In the meantime, in October 2018 the first device implementing the REMS approach has received the clearance of the U.S.A. Food and Drug Administration (FDA) for the measurement of the diagnostic parameters BMD, T-score and Z-score, and for monitoring bone changes in the clinical routine.
Further studies and clinical perspectives
An International Multicentre Clinical Trial, called “Echo-Bone” and focused on an extended study of the REMS approach for osteoporosis diagnosis in a wide European clinical context through a head-to-head trial with DXA, is currently ongoing in five reference centres for osteoporosis management. At an interim analysis on 595 patients enrolled in Barcelona (Spain) [107], preliminary data confirm the recently published results [102].
Then, another recently completed longitudinal study [108] has quantified the predictive value of REMS T-score in identifying patients at risk for osteoporotic fractures by performing a follow-up of up to 5 years for the possible occurrence of fragility fractures in a population of 1370 women who had undergone baseline spinal REMS and DXA scans. After the follow-up period, all the subjects were divided into two Groups: Group A (with incident fractures) and Group B (without incident fractures). As expected, significant differences between the two groups were found in both REMS T-score (− 2.68 ± 1.28 in Group A vs − 2.03 ± 1.23 in Group B, p < 0.001) and DXA T-score (− 2.52 ± 1.20 in Group A vs − 2.08 ± 1.17 in Group B, p < 0.001). The employment of the typical threshold of T-score ≤ − 2.5 as a cut-off value for both the techniques produced the following results in the identification of Group A patients by the two techniques: sensitivity = 65% and specificity = 57% for REMS (OR = 2.6); sensitivity = 57% and specificity = 57% for DXA (OR = 1.7). This performance of REMS T-score in the identification of patients at high risk for osteoporotic fractures will have to be further verified on a wider population through the outcome of a similar study that is currently ongoing in six clinical centres.
In general, the available data suggest that REMS technology might have a beneficial impact on current diagnostic protocols and subsequent patient management in the clinical routine. Further clinical scenarios are envisaged for additional REMS applications, including in particular fracture risk assessment in paediatric patients and pregnant women, and also in patients at risk of secondary osteoporosis (e.g., diabetic, nephropathic, oncological patients).
REMS adoption in the clinical routine is expected to increase in the near future because of the mentioned evidences and advantages, and also because the technology is currently undergoing the evaluation procedures to be included in the relevant international guidelines. Noteworthy, this technology does not require radiological protection, which might have been a problem for the reception in primary care in some countries. Finally, the portability facilitates the employment on hospitalized fractured patients not-transferrable to the densitometry units and on patient follow-up at home.
Conclusions
The impact of osteoporosis and the resulting bone fracture on both patient’s life and healthcare systems is constantly increasing: means for early diagnosis and monitoring of this disease are urgently needed in order to prevent and reduce the occurrence of fractures. BMD measurement based on DXA remains the current gold standard for osteoporosis classification, as defined by WHO. However, the above-mentioned clinical situation and the reported limitations of this technique boosted the scientific research to identify reliable and accurate alternative approaches for bone strength estimation and identification of individuals at high risk of fracture. Those efforts are leading to a deeper understanding of intrinsic bone characteristics, related not only to bone quantity but also to bone quality, which might contribute to improve the accuracy of bone strength measurement and fracture risk assessment.
In this context, REMS represents the first clinically available method for direct non-ionizing measurement of lumbar and femoral BMD. Available scientific evidences describe REMS-estimated BMD as an accurate diagnostic parameter, which resulted also a predictor of incident clinical fracture risk in a representative sample of female subjects. Moreover, REMS has shown a further potential in the assessment of skeletal fragility based on bone structure quality through the Fragility Score parameter, which is independent from the densitometric evaluation.
References
Anonymous (1993) Consensus development conference: diagnosis, prophylaxis and treatment of osteoporosis. Am J Med 94:646–650
Armas LAG, Lappe JM, Heaney RP (2010) Calcium, bone strength and fractures. In: Orwoll ES, Bilezikian JP, Vanderschueren D (eds) Osteoporosis in men, 2nd edn. Academic Press, London, pp 235–241
Choksi P, Jepsen KJ, Clines GA (2018) The challenges of diagnosing osteoporosis and the limitations of currently available tools. Clin Diabetes Endocrinol 4:12. https://doi.org/10.1186/s40842-018-0062-7
Fonseca H, Moreira-Gonçalves D, Coriolano HJ et al (2014) Bone quality: the determinants of bone strength and fragility. Sports Med 44:37–53. https://doi.org/10.1007/s40279-013-0100-7
Amman P, Rizzoli R (2003) Bone strength and its determinants. Osteoporos Int 14:13. https://doi.org/10.1007/s00198-002-1345-4
Hart NH, Nimphius S, Rantalainen T et al (2017) Mechanical basis of bone strength: influence of bone material, bone structure and muscle action. J Musculoskelet Neuronal Interact 17:114–139
Fogelman I, Blake GM (2000) Different approaches to bone densitometry. J Nucl Med 41:2015–2025
Damilakis J, Adams JE, Guglielmi G et al (2010) Radiation exposure in X-ray-based imaging techniques used in osteoporosis. Eur Radiol 20:2707–2714. https://doi.org/10.1007/s00330-010-1845-0
[No authors listed] (2003) Prevention and management of osteoporosis. World Health Organ Tech Rep Ser 921:1–164
Bousson V, Le Bras A, Roqueplan F et al (2006) Volumetric quantitative computed tomography of the proximal femur: relationships linking geometric and densitometric variables to bone strength. Role for compact bone. Osteoporos Int 17:855–864
World Health Organization (1994) Assessment of fracture risk and its application to screening for postmenopausal osteoporosis: report of a WHO study group [meeting held in Rome from 22 to 25 June 1992]
Kanis JA, Melton LJ, Christiansen C et al (1994) The diagnosis of osteoporosis. J Bone Miner Res 9:1137–1141
Blake G, Adams JE, Bishop N (2013) DXA in adults and children. In: Rosen CJ, Delmas P (eds) Primer on the metabolic bone diseases and disorders of mineral metabolism, 8th edn. American Society of Bone and Mineral Research (ASBMR), Washington, pp 152–158
Marshall D, Johnell O, Wedel H (1996) Meta-analysis of how well measures of bone mineral density predict occurrence of osteoporotic fractures. BMJ 312:1254–1259
Leslie WD, Majumdar SR, Morin SN et al (2015) Why does rate of bone density loss not predict fracture risk? J Clin Endocrinol Metab 100:679–683. https://doi.org/10.1210/jc.2014-3777
Siris ES, Chen YT, Abbott TA et al (2004) Bone mineral density thresholds for pharmacological intervention to prevent fractures. Arch Intern Med 164:1108–1112
Unnanuntana A, Gladnick BP, Donnelly E et al (2010) The assessment of fracture risk. J Bone Joint Surg Am 92:743–753
Wainwright SA, Marshall LM, Ensrud KE et al (2005) Hip fracture in women without osteoporosis. J Clin Endocrinol Metab 90:2787–2793
Schacter GI, Leslie WD (2017) DXA-based measurements in diabetes: can they predict fracture risk? Calcif Tissue Int 100:150–164. https://doi.org/10.1007/s00223-016-0191-x
Kanis JA, Borgstrom F, De Laet C et al (2005) Assessment of fracture risk. Osteoporos Int 16:581–589
Müller D, Pulm J, Gandjour A (2012) Cost-effectiveness of different strategies for selecting and treating individuals at increased risk of osteoporosis or osteopenia: a systematic review. Value Health 15:284–298. https://doi.org/10.1016/j.jval.2011.11.030
Nayak S, Roberts MS, Greenspan SL (2011) Cost-effectiveness of different screening strategies for osteoporosis in postmenopausal women. Ann Intern Med 155:751–761
Pothuaud L, Carceller P, Hans D (2008) Correlations between grey-level variations in 2D projection images (TBS) and 3D microarchitecture: applications in the study of human trabecular bone microarchitecture. Bone 42:775–787. https://doi.org/10.1016/j.bone.2007.11.018
Silva BC, Leslie WD, Resch H et al (2014) Trabecular bone score: a noninvasive analytical method based upon the DXA image. J Bone Miner Res 29:518–530. https://doi.org/10.1002/jbmr.2176
Hans D, Goertzen AL, Krieg MA et al (2011) Bone microarchitecture assessed by TBS predicts osteoporotic fractures independent of bone density: the Manitoba study. J Bone Miner Res 26:2762–2769. https://doi.org/10.1002/jbmr.499
Leslie WD, Majumdar SR, Morin SN et al (2017) Change in trabecular bone score (TBS) with antiresorptive therapy does not predict fracture in women: the Manitoba BMD cohort. J Bone Miner Res 32:618–623. https://doi.org/10.1002/jbmr.3054
Leslie WD, Aubry-Rozier B, Lamy O et al (2013) TBS (trabecular bone score) and diabetes-related fracture risk. J Clin Endocrinol Metab 98:602–609. https://doi.org/10.1210/jc.2012-3118
Martineau P, Silva BC, Leslie WD (2017) Utility of trabecular bone score in the evaluation of osteoporosis. Curr Opin Endocrinol Diabetes Obes 24:402–410. https://doi.org/10.1097/MED.0000000000000365
Kanis JA, Johnell O, Oden A et al (2008) FRAX and the assessment of fracture probability in men and women from the UK. Osteoporos Int 19:385–397. https://doi.org/10.1007/s00198-007-0543-5
McCloskey EV, Odén A, Harvey NC et al (2016) A meta-analysis of trabecular bone score in fracture risk prediction and its relationship to FRAX. J Bone Miner Res 31:940–948
Winzenrieth R, Michelet F, Hans D (2013) Three-dimensional (3D) microarchitecture correlations with 2D projection image gray-level variations assessed by trabecular bone score using high-resolution computed tomographic acquisitions: effects of resolution and noise. J Clin Densitom 16:287–296
Chen W, Slattery A, Center J et al (2016) The effect of changing scan mode on trabecular bone score using lunar prodigy. J Clin Densitom 19:502–506
Amnuaywattakorn S, Sritara C, Utamakul C et al (2016) Simulated increased soft tissue thickness artefactually decreases trabecular bone score: a phantom study. BMC Musculoskelet Disord 17:17. https://doi.org/10.1186/s12891-016-0886-1
Schousboe JT, Vo TN, Langsetmo L et al (2017) Association of trabecular bone score (TBS) with incident clinical and radiographic vertebral fractures adjusted for lumbar spine BMD in older men: a prospective cohort study. J Bone Miner Res 32:1554–1558. https://doi.org/10.1002/jbmr.3130
Langsetmo L, Vo TN, Ensrud KE et al (2016) The association between trabecular bone score and lumbar spine volumetric BMD is attenuated among older men with high body mass index. J Bone Miner Res 31:1820–1826. https://doi.org/10.1002/jbmr.2867
Roux JP, Wegrzyn J, Boutroy S, Bouxsein ML et al (2013) The predictive value of trabecular bone score (TBS) on whole lumbar vertebrae mechanics: an ex vivo study. Osteoporos Int 24:2455–2460. https://doi.org/10.1007/s00198-013-2316-7
Maquer G, Lu Y, Dall’Ara E et al (2016) The initial slope of the variogram, foundation of the trabecular bone score, is not or is poorly associated with vertebral strength. J Bone Miner Res 31:341–346. https://doi.org/10.1002/jbmr.2610
Harvey NC, Glüer CC, Binkley N et al (2015) Trabecular bone score (TBS) as a new complementary approach for osteoporosis evaluation in clinical practice. Bone 78:216–224. https://doi.org/10.1016/j.bone.2015.05.016
Edmondson CP, Schwartz EN (2017) Non-BMD DXA measurements of the hip. Bone 104:73–83. https://doi.org/10.1016/j.bone.2017.03.050
Leslie WD, Lix LM, Morin SN et al (2016) Adjusting hip fracture probability in men and women using hip axis length: the Manitoba bone density database. J Clin Densitom 19:326–331. https://doi.org/10.1016/j.jocd.2015.07.004
Broy SB, Cauley JA, Lewiecki ME et al (2015) Fracture risk prediction by non-BMD DXA measures: the 2015 ISCD Official Positions Part 1: hip geometry. J Clin Densitom 18:287–308. https://doi.org/10.1016/j.jocd.2015.06.005
Beck TJ, Broy SB (2015) Measurement of hip geometry—technical background. J Clin Densitom 18:331–337. https://doi.org/10.1016/j.jocd.2015.06.006
Zhuang H, Li Y, Lin J et al (2017) Cortical thickness in the intertrochanteric region may be relevant to hip fracture type. BMC Musculoskelet Disord 18:305. https://doi.org/10.1186/s12891-017-1669-z
Lee DH, Jung KY, Hong AR et al (2016) Femoral geometry, bone mineral density, and the risk of hip fracture in premenopausal women: a case control study. BMC Musculoskelet Disord 17:42. https://doi.org/10.1186/s12891-016-0893-2
Adams JE (2009) Quantitative computed tomography. Eur J Radiol 71:415–424
Link TM, Lang TF (2014) Axial QCT: clinical applications and new developments. J Clin Densitom 17:438–448. https://doi.org/10.1016/j.jocd.2014.04.119
Manhard MK, Nyman JS, Does MD (2017) Advances in imaging approaches to fracture risk evaluation. Transl Res 181:1–14. https://doi.org/10.1016/j.trsl.2016.09.006
Kopperdahl DL, Aspelund T, Hoffmann PF et al (2014) Assessment of incident spine and hip fractures in women and men using finite element analysis of CT scans. J Bone Miner Res 29:570–580. https://doi.org/10.1002/jbmr.2069
Keaveny TM, Hoffmann PF, Singh M et al (2008) Femoral bone strength and its relation to cortical and trabecular changes after treatment with PTH, alendronate, and their combination as assessed by finite element analysis of quantitative CT scans. J Bone Miner Res 23:1974–1982. https://doi.org/10.1359/jbmr.080805
Areeckal AS, Kocher M, Sumam David S (2018) Current and emerging diagnostic imaging-based techniques for assessment of osteoporosis and fracture risk. IEEE Rev Biomed Eng 12:254–268. https://doi.org/10.1109/RBME.2018.2852620
Ruiz Wills C, Olivares AL, Tassani S et al (2019) 3D patient-specific finite element models of the proximal femur based on DXA towards the classification of fracture and non-fracture cases. Bone 121:89–99. https://doi.org/10.1016/j.bone.2019.01.001
Yang S, Leslie WD, Luo Y et al (2018) Automated DXA-based finite element analysis for hip fracture risk stratification: a cross-sectionalstudy. Osteoporos Int 29:191–200. https://doi.org/10.1007/s00198-017-4232-8
Leslie WD, Luo Y, Yang S et al (2019) Fracture risk indices from DXA-based finite element analysis predict incident fractures independently from FRAX: The Manitoba BMD Registry. J Clin Densitom. https://doi.org/10.1016/j.jocd.2019.02.001
Stagi S, Cavalli L, Cavalli T et al (2016) Peripheral quantitative computed tomography (pQCT) for the assessment of bone strength in most of bone affecting conditions in developmental age: a review. Ital J Pediatr 42:88
MacNeil JA, Boyd SK (2008) Improved reproducibility of high-resolution peripheral quantitative computed tomography for measurement of bone quality. Med Eng Phys 30:792–799. https://doi.org/10.1016/j.medengphy.2007.11.003
Burghardt AJ, Link TM, Majumdar S (2011) High-resolution computed tomography for clinical imaging of bone microarchitecture. Clin Orthop Relat Res 469:2179–2193. https://doi.org/10.1007/s11999-010-1766-x
van Rietbergen B, Ito K (2015) A survey of micro-finite element analysis for clinical assessment of bone strength: the first decade. J Biomech 48:832–841. https://doi.org/10.1016/j.jbiomech.2014.12.024
Digby MG, Bishop NJ, Paggiosi MA et al (2016) HR-pQCT: a non-invasive ‘biopsy’ to assess bone structure and strength. Arch Dis Child Educ Pract Ed 101:268–270. https://doi.org/10.1136/archdischild-2015-309455
Lespessailles E, Ibrahim-Nasser N, Toumi H et al (2018) Contribution of high resolution peripheral quantitative CT to the management of bone and joint diseases. Joint Bone Spine 85:301–306. https://doi.org/10.1016/j.jbspin.2017.04.012
Biver E, Durosier-Izart C, Chevalley T et al (2018) Evaluation of radius microstructure and areal bone mineral density improves fracture prediction in postmenopausal women. J Bone Miner Res 33:328–337. https://doi.org/10.1002/jbmr.3299
Cheung AM, Adachi JD, Hanley DA et al (2013) High-resolution peripheral quantitative computed tomography for the assessment of bone strength and structure: a review by the Canadian Bone Strength Working Group. Curr Osteoporos Rep 11:136–146. https://doi.org/10.1007/s11914-013-0140-9
Chang G, Honig S, Liu Y et al (2015) 7 Tesla MRI of bone microarchitecture discriminates between women without and with fragility fractures who do not differ by bone mineral density. J Bone Miner Metab 33:285–293. https://doi.org/10.1007/s00774-014-0588-4
Griffith JF, Genant HK (2012) New advances in imaging osteoporosis and its complications. Endocrine 42:39–51. https://doi.org/10.1007/s12020-012-9691-2
Hans D, Baim S (2017) Quantitative ultrasound (QUS) in the management of osteoporosis and assessment of fracture risk. J Clin Densitom 20:322–333. https://doi.org/10.1016/j.jocd.2017.06.018
Shepherd JA, Schousboe JT, Broy SB et al (2015) Executive summary of the 2015 ISCD position development conference on advanced measures from DXA and QCT: fracture prediction beyond BMD. J Clin Densitom 18(3):274–286. https://doi.org/10.1016/j.jocd.2015.06.013
Langton CM, Ali AV, Riggs CM et al (1990) A contact method for the assessment of ultrasonic velocity and broadband attenuation in cortical and cancellous bone. Clin Phys Physiol Meas 11:243–249
Prins SH, Jørgensen HL, Jørgensen LV et al (1998) The role of quantitative ultrasound in the assessment of bone: a review. Clin Physiol 18:3–17
Damilakis J, Maris TG, Karantanas AH (2007) An update on the assessment of osteoporosis using radiologic techniques. Eur Radiol 17:1591–1602
Chan MY, Nguyen ND, Center JR et al (2013) Quantitative ultrasound and fracture risk prediction in non-osteoporotic men and women as defined by WHO criteria. Osteoporos Int 24:1015–1022. https://doi.org/10.1007/s00198-012-2001-2
McLeod KM, Johnson S, Rasali D et al (2015) Discriminatory performance of the calcaneal quantitative ultrasound and osteoporosis self-assessment tool to select older women for dual-energy X-ray absorptiometry. J Clin Densitom 18:157–164. https://doi.org/10.1016/j.jocd.2015.02.006
Zhang L, Lv H, Zheng H et al (2015) Correlation between parameters of calcaneal quantitative ultrasound and hip structural analysis in osteoporotic fracture patients. PLoS ONE 10:e0145879. https://doi.org/10.1371/journal.pone.0145879
Marín F, González-Macías J, Díez-Pérez A et al (2006) Relationship between bone quantitative ultrasound and fractures: a meta-analysis. J Bone Miner Res 21:1126–1135
McCloskey EV, Johansson H, Kanis JA et al (2015) Predictive ability of heel quantitative ultrasound for incident fractures: an individual-level meta-analysis. Osteoporos Int 26:1979–1987. https://doi.org/10.1007/s00198-015-3072-7
Black DM, Cauley JA, Wagman R et al (2018) The ability of a single BMD and fracture history assessment to predict fracture over 25 years in postmenopausal women: the study of osteoporotic fractures. J Bone Miner Res 33:389–395. https://doi.org/10.1002/jbmr.3194
Austin M, Yang YC, Vittinghoff E et al (2012) Relationship between bone mineral density changes with denosumab treatment and risk reduction for vertebral and nonvertebral fractures. J Bone Miner Res 27:687–693. https://doi.org/10.1002/jbmr.1472
Jacques RM, Boonen S, Cosman F et al (2012) Relationship of changes in total hip bone mineral density to vertebral and nonvertebral fracture risk in women with postmenopausal osteoporosis treated with once-yearly zoledronic acid 5 mg: the HORIZON-Pivotal Fracture Trial (PFT). J Bone Miner Res 27:1627–1634. https://doi.org/10.1002/jbmr.1644
Bouxsein ML, Eastell R, Lui LY et al (2019) Change in bone density and reduction in fracture risk: a meta-regression of published trials. J Bone Miner Res 1:1. https://doi.org/10.1002/jbmr.3641
Kanis JA, Cooper C, Rizzoli R et al (2019) European guidance for the diagnosis and management of osteoporosis in postmenopausal women. Osteoporos Int. 30:3–44. https://doi.org/10.1007/s00198-018-4704-5
Osterhoff G, Morgan EF, Shefelbine SJ et al (2016) Bone mechanical properties and changes with osteoporosis. Injury 47:S11–S20. https://doi.org/10.1016/S0020-1383(16)47003-8
McCreadie BR, Goldstein SA (2000) Biomechanics of fracture: is bone mineral density sufficient to assess risk? J Bone Miner Res 15:2305–2308
Keaveny TM, Bouxsein ML (2008) Theoretical implications of the biomechanical fracture threshold. J Bone Miner Res 23:1541–1547. https://doi.org/10.1359/jbmr.080406
Boutroy S, Bouxsein ML, Munoz F et al (2005) In vivo assessment of trabecular bone microarchitecture by high-resolution peripheral quantitative computed tomography. J Clin Endocrinol Metab 90:6508–6515
Samelson EJ, Broe KE, Xu H et al (2019) Cortical and trabecular bone microarchitecture as an independent predictor of incident fracture risk in older women and men in the Bone Microarchitecture International Consortium (BoMIC): a prospective study. Lancet Diabetes Endocrinol 7:34–43. https://doi.org/10.1016/s2213-8587(18)30308-5. Erratum in: Lancet Diabetes Endocrinol 7:e1
Eastell R, Wahner HW, O’Fallon WM et al (1989) Unequal decrease in bone density of lumbar spine and ultradistal radius in Colles’ and vertebral fracture syndromes. J Clin Invest 83:168–174
Shen J, Griffith JF, Zhu TY et al (2018) Bone mass, microstructure, and strength can discriminate vertebral fracture in patients on long-term steroid treatment. J Clin Endocrinol Metab 103:3340–3349. https://doi.org/10.1210/jc.2018-00490
Johannesdottir F, Allaire B, Bouxsein ML (2018) Fracture prediction by computed tomography and finite element analysis: current and future perspectives. Curr Osteoporos Rep 16:411. https://doi.org/10.1007/s11914-018-0450-z
Panyasantisuk J, Dall’Ara E, Pretterklieber M et al (2018) Mapping anisotropy improves QCT-based finite element estimation of hip strength in pooled stance and side-fall load configurations. Med Eng Phys 59:36–42. https://doi.org/10.1016/j.medengphy.2018.06.004
Szulc P, Boutroy S, Chapurlat R (2018) Prediction of fractures in men using bone microarchitectural parameters assessed by high-resolution peripheral quantitative computed tomography—the prospective STRAMBO study. J Bone Miner Res 33:1470–1479. https://doi.org/10.1002/jbmr.3451
Unal M, Creecy A, Nyman JS (2018) The role of matrix composition in the mechanical behaviour of bone. Curr Osteoporos Rep 16:205–215. https://doi.org/10.1007/s11914-018-0433-0
Herrera S, Diez-Perez A (2017) Clinical experience with microindentation in vivo in humans. Bone 95:175–182. https://doi.org/10.1016/j.bone.2016.11.003
Rozental TD, Walley KC, Demissie S et al (2018) Bone material strength index as measured by impact microindentation in postmenopausal women with distal radius and hip fractures. J Bone Miner Res 33:621–626. https://doi.org/10.1002/jbmr.3338
Winzenrieth R, Humbert L, Di Gregorio S et al (2018) Effects of osteoporosis drug treatments on cortical and trabecular bone in the femur using DXA-based 3D modeling. Osteoporos Int 29:2323–2333. https://doi.org/10.1007/s00198-018-4624-4
Bliuc D, Alarkawi D, Nguyen TV et al (2015) Risk of subsequent fractures and mortality in elderly women and men with fragility fractures with and without osteoporotic bone density: the Dubbo Osteoporosis Epidemiology Study. J Bone Miner Res 30:637–646. https://doi.org/10.1002/jbmr.2393
Viceconti M, Qasim M, Bhattacharya P et al. (2018) Are CT-based finite element model predictions of femoral bone strengthening clinically useful? Curr Osteoporos Rep 16:216–223. https://doi.org/10.1007/s11914-018-0438-8. Review. Erratum in: Curr Osteoporos Rep. 2018 Jun 22
Kanis JA, Hiligsmann M (2014) The application of health technology assessment in osteoporosis. Best Pract Res Clin Endocrinol Metab 28:895–910. https://doi.org/10.1016/j.beem.2014.04.001
Minniti D, Davini O, Gualano MR et al (2014) Techniques for diagnosing osteoporosis: a systematic review of cost-effectiveness studies. Int J Technol Assess Health Care 30:273–281. https://doi.org/10.1017/S0266462314000257
Conversano F, Franchini R, Greco A et al (2015) A novel ultrasound methodology for estimating spine mineral density. Ultrasound Med Biol 41:281–300. https://doi.org/10.1016/j.ultrasmedbio.2014.08.017
Casciaro S, Peccarisi M, Pisani P et al (2016) An advanced quantitative echosound methodology for femoral neck densitometry. Ultrasound Med Biol 42:1337–1356. https://doi.org/10.1016/j.ultrasmedbio.2016.01.024
Caffarelli C, Tomai Pitinica MD, Francolini V et al (2018) REMS technique: future perspectives in an Academic Hospital. Clin Cases Miner Bone Metab 15:163–165
Greco A, Pisani P, Conversano F et al (2017) Ultrasound fragility score: an innovative approach for the assessment of bone fragility. Measurement 101:236–242
Pisani P, Greco A, Conversano F et al (2017) A quantitative ultrasound approach to estimate bone fragility: a first comparison with dual X-ray absorptiometry. Measurement 101:243–249
Di Paola M, Gatti D, Viapiana O et al (2018) Radiofrequency echographic multispectrometry compared with dual X-ray absorptiometry for osteoporosis diagnosis on lumbar spine and femoral neck. Osteoporos Int 30:391–402. https://doi.org/10.1007/s00198-018-4686-3
Messina C, Bandirali M, Sconfienza LM et al (2015) Prevalence and type of errors in dual-energy X-ray absorptiometry. Eur Radiol 25:1504–1511. https://doi.org/10.1007/s00330-014-3509-y
Altman DG, Bland JM (1983) Measurements in medicine: the analysis of method comparison studies. Statistician 32:307–317
Hopkins SJ, Welsman JR, Knapp KM (2014) Short-term precision error in dual energy X-ray absorptiometry, bone mineral density and trabecular bone score measurements; and effects of obesity on precision error. J Biomed Gr Comput 4:8–14
Ravaud P, Reny JL, Giraudeau B et al (1999) Individual smallest detectable difference in bone mineral density measurements. J Bone Miner Res 14:1449–1456
Ovejero Crespo D, Nogues X, Diez-Perez A (2019) The non-ionizing radiofrequency echographic multi spectrometry (REMS) applied on a Spanish cohort for the osteoporosis diagnosis on lumbar spine and femoral neck. WCO-IOF-ESCEO Abstract P795
Adami G, Arioli G, Bianchi G et al (2019) Prediction of incident fragility fractures through radiofrequency echographic multi spectrometry (REMS). Ann Rheum Dis 78(Suppl 2):933. https://doi.org/10.1136/annrheumdis-2019-eular.6256
Acknowledgements
The ESCEO Working Group was funded by the ESCEO. The ESCEO receives unrestricted educational grants to support its educational and scientific activities from non-governmental organizations, not-for-profit organizations, non-commercial or corporate partners. The choice of topics, participants, content and agenda of the Working Groups as well as the writing, editing, submission and reviewing of the manuscript are the sole responsibility of the ESCEO, without any influence from third parties.
Funding
The ESCEO Working Group was funded by the ESCEO as detailed in the “Acknowledgements section”. No other funding was received.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
A Diez-Perez has been speaker or advisor for AMGEN, UCB, Lilly, Roche, Echolight and Gilead and owns shares of Active Life Sci. ML Brandi has received honoraria from Amgen, Bruno Farmaceutici, Kyowa Kirin; Academic grants and/or speaker from: Abiogen, Alexion, Amgen, Bruno Farmaceutici, Eli Lilly, Kyowa Kirin, MSD, NPS, Servier, Shire, SPA; Consulting fees from: Alexion, Bruno Farmaceutici, Kyowa Kirin, Servier, Shire. B Cortet has received fees for occasional interventions as an expert or speaker for Amgen, Expanscience, Ferring, Lilly, Medtronic, MSD, Mylan, Novartis, Roche Diagnostics, and UCB. S Gonnelli has received lecture fees or paid advisory boards by Lilly, UCB, Abiogen, Amgen. J-Y Reginster has received Consulting fees or paid advisory boards from IBSA-GENEVRIER, MYLAN, RADIUS HEALTH, PIERRE FABRE; Lecture fees when speaking at the invitation of sponsor IBSA-GENEVRIER, MYLAN, CNIEL, DAIRY RESEARCH COUNCIL (DRC); Grant Support from Industry (All through Institution): IBSA-GENEVRIER, MYLAN, CNIEL, RADIUS HEALTH. R Rizzoli has been Speaker Bureau or Member of Scientific Advisory Boards for Danone, Echolight, Effryx, Mylan, Nestlé, ObsEva, Pfizer, Radius Health, Sandoz and TEVA/Theramex. T Thomas has received personal fees for lectures and expertises from Abbvie, Amgen, Arrow, BMS, Chugai, Expanscience, Gilead, HAC-Pharma, LCA, Lilly, Medac, MSD, Pfizer, Theramex, Thuasne, TEVA and UCB and financial support or fees for research activities from Amgen, Bone Therapeutics, Chugai, HAC-Pharma, MSD, Novartis, Pfizer, and UCB. All the other authors declare that they have no conflicts of interest concerning this article.
Statement of human and animal rights
No humans or animals have been involved in this study.
Informed consent
For this type of study formal consent is not required.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
About this article
Cite this article
Diez-Perez, A., Brandi, M.L., Al-Daghri, N. et al. Radiofrequency echographic multi-spectrometry for the in-vivo assessment of bone strength: state of the art—outcomes of an expert consensus meeting organized by the European Society for Clinical and Economic Aspects of Osteoporosis, Osteoarthritis and Musculoskeletal Diseases (ESCEO). Aging Clin Exp Res 31, 1375–1389 (2019). https://doi.org/10.1007/s40520-019-01294-4
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s40520-019-01294-4